Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Similar documents
DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Methods DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

The introduction of oscillometric automated blood pressure

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

External Oscillatory Blood Pressure - EOBPTM

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

In clinical practice, blood pressure (BP) treatment

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

A GUIDE TO OUR NIBP TECHNOLOGY

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death

Home blood pressure measurement with oscillometric upper-arm devices

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

doi: /fampra/cmw053

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

THE ACCURATE ASSESSMENT

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure?

THE NEW ARMENIAN MEDICAL JOURNAL DISTRIBUTION, AWARENESS, TREATMENT, AND CONTROL OF ARTERIAL HYPERTENSION IN YEREVAN (ARMENIA)

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Validation Study Result Form

Blood pressure measurement in the year 2008: revival of oscillometry?

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Atrial Fibrillation. Why It s Important. Updated on 2007 / 11 / 23. Microlife Corp ww.microlife.com. Page 1

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Validation Study Result Form

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

greater than 90% of patients having blood pressure measured consistently according to the protocol.

Auscultatory versus oscillometric blood pressure measurement in patients with atrial fibrillation and arterial hypertension

There is convincing evidence in clinical studies

AFFORDABLE TECHNOLOGY

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Accuracy of Digital Arm and Wrist Manometers: Clinical Implications for the Dental Hygienist. Abstract

The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

STATE OF THE ART BP ASSESSMENT

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

Arm position and blood pressure: a risk factor for hypertension?

Online Supplementary Appendix

Protocol. Automated Ambulatory Blood Pressure Monitoring for the Diagnosis of Hypertension in Patients with Elevated Office Blood Pressure

Assessing Blood Pressure Accuracy of an Aneroid Sphygmomanometer in a National Survey Environment

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

4/4/17 HYPERTENSION TARGETS: WHAT DO WE DO NOW? SET THE STAGE BP IN CLINICAL TRIALS?

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM

SIGNAL PROCESSING and KNOWLEDGE ACQUISITION in OBJECTIVE BLOOD PRESSURE MEASUREMENT with CONVENTIONAL CLINICAL SPHYGMOMANOMETRY

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Biomedical Instrumentation E. Blood Pressure

The Hypertension Clinic is a part of the Internal Medicine

Χρήση έξυπνων τεχνολογιών στην ανίχνευση κολπικής μαρμαρυγής Use of smart technology in atrial fibrillation detection

Automatic oscillometric blood pressure devices are frequently

Is Traditional Clinic Blood Pressure Dead?

Blood Pressure Estimation Using Photoplethysmography (PPG)

Comparison of the Microlife A6 PC (BP3GU1-8Y) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Maker Spacelabs Healthcare Manufacturer Spacelabs Healthcare. Brand Spacelabs Model OnTrak

How Low Do We Go? Update on Hypertension

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Atrial fibrillation Etiology and complications - A descriptive study

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

An Automated System for Blood Pressure Determination During Exercise

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Prevalence, awareness of hypertension in rural areas of Kurnool

Effect of respiration on Korotkoff sounds and oscillometric cuff pressure pulses during blood pressure measurement

Blood Pressure Competency Training

Special Lecture 10/28/2012

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE General practice Indicators for the NICE menu

K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease

SECTION 16: BLOOD PRESSURE MEASUREMENT

Continuous-flow (CF) left ventricular assist devices. Original Article

FOR many years, casual blood pressure (BP)

a central pulse located at the apex of the heart Apical pulse Apical-radial pulse a complete absence of respirations Apnea

The magnitude and duration of ambulatory blood pressure reduction following acute exercise

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee

Essential Hypertension Management Considerations By Elaine Lewis, ND

Data Collection Worksheet

Transcription:

ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson MD, FRCPC Department of Medicine and Cardiovascular Risk Factor Reduction Unit, University of Saskatchewan, Saskatoon, SK, Canada. Manuscript submitted 18th November, 2009 Manuscript accepted 18th January, 2010 Clin Invest Med 2010; 33 (1): E54-E62. Abstract Objective: To compare blood pressure readings obtained with two commonly used oscillometric monitors: Omron HEM 711 AC (OM) and Welch-Allyn 52000 series NIBP/ oximeter (WA) with mercury sphygmomanometers (Merc) in subjects with atrial fibrillation. Methods: We recruited 51 hemodynamically stable subjects with atrial fibrillation. Fifty four subjects in normal sinus rhythm served as controls. Supine blood pressure readings in each arm were recorded simultaneously using one monitor and Merc. The second monitor then replaced the first and readings were repeated. Merc was then switched to the opposite arm, and both monitors retested. Apical heart rates were ascertained with a stethoscope. We used the averaged, same arm Merc readings as gold standard. Results: Automated blood pressure readings were obtained in all control subjects and in all but three of those with atrial fibrillation. Both monitors, and operators, noted a difference between apical and radial/brachial pulse rates: apical-recorded: Merc 6.1±15.0; OM 5.5±13.7; WA 10.0±21.2 beats per minute. Both monitors were accurate in controls: over 90% of readings were within 10 mmhg of averaged Merc, and both achieved European Hypertension Society standards. In subjects with atrial fibrillation, about one quarter of all oscillometric readings differed from Merc by more than 10 mmhg. Both falsely high and falsely low readings occurred, some up to 30 mmhg. There was no relation between accuracy and heart rate. Conclusions: Single blood pressure readings, taken with oscillometric monitors in subjects with atrial fibrillation differ, often markedly, from those taken manually. Health care professionals should record multiple readings manually, using validated instruments when making therapeutic decisions. Atrial fibrillation is the most common chronic arrhythmia; occurring in up to 9% of people aged 65 or more 1. It increases the risk of stroke and death by at least 1.5 fold. Atrial fibrillation is present in up to 17% of hypertensives. 2 Hypertension has been implicated as a causative factor in 14% of cases of atrial fibrillation. Accurate blood pressure recording in subjects with atrial fibrillation is problematic. Not only does the blood pressure itself vary over time, but Karotkoff sounds may vary in pitch and amplitude, and some may be well outside the average systolic or diastolic pressure. Guidelines recognize this problem but are generally silent on the proper way to record such readings. 3 Some recommend recording the average of multiple, auscultatory readings, 4,5,6 while other make no specific recommendation. 7 Oscillometric blood pressure monitors have become increasingly common both in hospitals and in physicians offices. 8 To our knowledge, none have 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E54.

been validated via recognized protocols in subjects with atrial fibrillation. Furthermore, often a single blood pressure reading is recorded, and may lead to therapeutic decisions. If there was a large error in such a single reading, inappropriate treatment could result. We recently encountered a patient with longstanding hypertension and atrial fibrillation, dyspnea on exertion, left ventricular hypertrophy and marked diastolic dysfunction. His home monitor readings averaged less than 130/80, but showed considerable variation (116/74 to 210/90). Our averaged clinic readings were 165/84, leading us to intensify his regimen. 9 Ordinarily, we place considerable weight on accurate home readings. We therefore decided to test the accuracy of automated monitors. Specifically, we compared blood pressure readings obtained with a mercury sphygmomanometer with those of two commonly used oscillometric monitors, in subjects with and without atrial fibrillation. We chose the Omron HEM 711 AC which is commonly used in the home, and the Welsh-Allyn 52000/NIBP, which is used in our hospital. Methods Subjects Our subjects were men or women, 18 years or older, either outpatients or those hospitalized at Royal University Hospital, with stable heart rate and blood pressure for 24 hours. We excluded inpatients whose recorded heart rate or blood pressure varied by more than 50%, who had had surgery within 24 hours, or had other severe illness. Outpatients were asymptomatic and attending the Cardiovascular Risk Factor Reduction Clinic. 10 Patients with atrial fibrillation were identified by routine electrocardiography or telemetry, and confirmed clinically, at the time of measurement. Control subjects were in sinus rhythm on clinical examination and had no history of atrial fibrillation. We excluded those with more than 2 extrasystoles per minute. All subjects gave informed consent and the protocol was approved by the University of Saskatchewan Biomedical Ethics Committee (EC 2006-229). We used the Omron HEM 711AC home monitor, which, according to the manufacturer has met criteria established by the European Hypertension Society, the British Hypertension Society and the Association for the Advancement of Medical Instrumentation {http://www.healthcare.omron.co.jp/english/vali_us.p hp} and the Welch-Allyn 52000 series NIBP/oximeter, a commonly used model in our hospital. Both are selfinflating digital, oscillometric monitors. Our comparator was a standard mercury sphygmomanometer (Py- Mah Trimline, Trimline Medical Products, Branchberg, NJ). Protocol We collected demographic and anthropomorphic data for each subject. Persons recording manual blood pressure had completed our locally developed Blood Pressure Measurement course. 11 They measured apical heart rate via auscultation. Because there was no published, validated blood pressure measurement protocol for patients with atrial fibrillation, we designed the following. All measurements were taken in the supine position, the patients had voided, had not smoked, eaten or consumed caffeinated beverages for 30 minutes, and had rested 5 minutes. Cuff size was determined by arm circumference according to the manufacturers. Two operators obtained readings on each patient. We used the supine position to ensure that both arms were supported and that both operators had room to record their readings. A mercury sphygmomanometer cuff was applied to one arm and one of the automated monitors (in random order) on the other. Blood pressure was recorded simultaneously, with the manual operator using Canadian Hypertension Education Program technique (other than the seated position). 5 Then, the other 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E55.

TABLE 1. Subject characteristics AFib (n=51) Cntrl (n=54) p* Sex (m/f) 30/21 35/19 0.55 Age 73.8 (12.4) 58.3 (12.7) 0.85 Apical HR 82.9 (15.5) 70.4 (13.4) 0.303 SBP 124.0 (22.9) 129.1 (17.3) 0.046 DBP 68.7 (10.1) 77.0 (11.7) 0.736 AFib: atrial fibrillation; Cntrl: subjects in normal sinus rhythm; m/f: male, female; HR: heart rate; SBD, DBP: systolic and dia- stolic blood pressure (mean of 4 mercury sphygmomanometry readings); figures in parentheses are standard deviation. * calculated with Fishers exact test for age, one way ANOVA for others automated monitor was substituted for the first one, and the procedure repeated. Finally, the mercury sphygmomanometer cuff was transferred to the other arm, and the two automated monitors tested in similar fashion. Thus we had 8 blood pressure and heart rate recordings for each subject: 2 on each arm with the mercury sphygmomanometer, and one on each arm with each of the automated monitors. For subjects with atrial fibrillation, we asked observers to record systolic blood pressure when most beats were heard and diastolic blood pressure when most beats disappeared. We included the group of patients in sinus rhythm in order to test the validity of the protocol. Statistical analyses We used the average of the two mercury readings on an arm as our gold standard and compared each manometer s reading on that arm. We set clinically reasonable accuracy to demonstrate that 90 percent of the automated readings would fall within 15 mmhg of our gold standard 12. Assuming that the average systolic blood pressure was 130 mmhg in each group we required 39 arms. However, because of the uncertainty of these assumptions we planned to enroll 50 subjects (100 arms) in each group. TABLE 2. Heart rates and pulse deficits Apex-pulse Apex-OM Apex-WA AFib 6.1 (15.0) 5.5 (13.7) 10.0 (21.2) Cntrl 0.7 (5.4) 0.3 (5.4) 0.1 (4.7) Figures in parentheses are standard deviation. OM: Omron HEM 711 AC; WA: Welsh-Allyn 520000 series NIBP/oximeter. Other abbreviations same as Table 1. Standard descriptive statistics and linear regression were performed using StatMost 3.2 for Windows (Datamost Corp, Sandy, UT) Results Subject characteristics are shown in Table 1. There were more men than women in each group. Subjects with atrial fibrillation were numerically older and had lower systolic and diastolic pressures. For systolic blood pressure the difference was statistically significant. Adequate readings with both automated monitors were obtained on all control subjects, but not on all subjects with atrial fibrillation. In two of the latter group (4%) we could not obtain a reading with either monitor, while we failed to obtain readings with one of the two in two other subjects. Accuracy of heart rate recording is shown in Table 2. In general all methods underestimated the apex rate (the pulse deficit ) in subjects with atrial fibrillation, but were quite precise in control subjects. Figures 1 and 2 are Bland-Altman type plots for systolic and diastolic readings with each monitor, compared to those taken with the mercury sphygmomanometer, for each group. There appears to be greater deviation from mercury readings with both monitors in subjects with atrial fibrillation. Table 3 shows the proportion of readings within defined ranges from the averaged mercury readings for subjects with sinus rhythm. Both devices would likely be rated pass ; the small deviation at! 5 mmhg for the Welsh-Allyn monitor (57% observed vs. " 60% expected) would not likely be clinically significant. Indeed it was explained by 2 subjects 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E56.

FIGURE 1. Bland-Altman Type Plots for Systolic Blood Pressure (SBP). AFib: subjects in atrial fibrillation; NSR: subjects in normal sinus rhythm; OM Omron HEM 711 AC monitor; WA: Welsh-Allyn 52000 series NIBP/oximeter monitor. Values are mmhg. TABLE 3. Percent of oscillometric readings within 5, 10 or 15 mmhg of averaged mercury eadings r in control subjects (n=108 arms).! 5 mmhg! 10 mmhg! 15 mmhg ESH standard 12 60 75 90 OM SBP 65 89 96 DBP 65 91 100 WA SBP 57 90 95 DBP 70 98 99 ESH: European Hypertension Society. Other abbreviations same as Table 2. TABLE 4. Percent of oscillometric readings within 5, 10 or 15 mmhg of averaged mercury eadings r in subjects with atrial fibrillation (n=96 arms).! 5 mmhg! 10 mmhg! 15 mmhg ESH standard 12 60 75 90 Omron SBP 49 72 84 DBP 47 77 92 WA SBP 46 72 81 DBP 57 86 96 Abbreviations same as Table 3. 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E57.

FIGURE 2. Bland-Altman Type Plots for Diastolic Blood Pressure (DBP). Abbreviations same as Figure 1. whose readings were 6 and 7 mmhg different than the average mercury reading. Table 4 shows similar data in subjects with atrial fibrillation. Both monitors would likely be rated fail. About one quarter of readings differed from mercury readings by more than 10 mmhg and 2 subjects showed a difference of over 30 mmhg systolic and 20 mmhg diastolic pressures with one or other of the monitors. The monitors appeared to achieve about the same level of accuracy. When atrial fibrillation subjects were divided into those above and below the median heart rate, there was no difference in the accuracy of the readings (data not shown). Figure 3 shows regression lines with 95% confidence intervals, for simultaneous readings in arms of subjects with atrial fibrillation. For the Omron monitor the regression coefficients for systolic and diastolic pressure were 0.835 and 0.739; for the Welsh-Allyn, 0.768 and 0.688. Figure 4 shows similar regression lines for subjects in sinus rhythm. The correlation coefficients were, numerically greater: ranging from 0.848 to 0.9946. 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E58.

FIGURE 3. Simultaneous readings on opposite arms for subjects in atrial fibrillation. Abbreviations same as Figure 1. Discussion These data show that blood pressure readings, obtained with oscillometric monitors, are likely to be similar to those taken manually in subjects in sinus rhythm, but not in those with atrial fibrillation. Indeed, only about 1 in 10 single oscillometric readings differed from averaged mercury readings by more than 10 mmhg in subjects in sinus rhythm, while 1 in 4 did so in subjects with atrial fibrillation. Both falsely high and falsely low readings were recorded so that the error appears to be random and unpredictable. We found no relation between the size of the error and apical heart rate, absolute blood pressure or any other easily measurable clinical parameter. A limitation of our study is the lack of a true gold standard for blood pressure measurements in subjects with atrial fibrillation. Experts recommend several readings in order to estimate the true blood pressure. We took the average of only two mercury readings from each arm. We feel that more readings would have caused discomfort for very little extra accuracy. Another limitation is our comparison of only one oscillometric reading with the average of the mercury readings. Perhaps more oscillometric readings would have approached the mean. However, in practice, de- 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E59.

FIGURE 4. Simultaneous readings on opposite arms for subjects in normal sinus rhythm. Abbreviations same as Figure 1. cisions are often made to change or withhold therapy on the basis of single readings. Our sinus rhythm group showed good agreement between the two methods using this protocol. Oscillometric recorders use a maximal amplitude algorithm, in which the pressure at maximum oscillation of the arm is assumed to equal mean arterial pressure. 13 Systolic and diastolic pressures are then calculated using proprietary algorithms. An irregular pulse may make it more difficult to ascertain the pressure at maximal oscillation, as the pulse pressure will change from beat to beat. Furthermore, the calculations for systolic and diastolic pressure assume that the R-R interval is constant. 14 It is not surprising therefore, that such monitors are different than mercury sphygmomanometry readings in subjects with atrial fibrillation. Other attempts to assess the accuracy of oscillometric recorders in atrial fibrillation have used different protocols. Stewart and colleagues compared five monitors in 30 patients with atrial fibrillation, using a same-arm, sequential method with a mercury-based sphygmomanometer as gold standard. 15 They found that one of the tested monitors using a microphone to detect Korotkoff sounds was within 5 mmhg of the mercury based monitor in about 70% of readings. All 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E60.

other monitors were less accurate. They did not include subjects in normal sinus rhythm and none of the monitors had been validated with accepted protocols. Jani and colleagues compared four mercury readings with four oscillometric readings in the same arm in 20 subjects with atrial fibrillation and 20 in normal sinus rhythm. 16 Although the variation among readings was greater in atrial fibrillation, all readings were within 15 mmhg. Given the small number of subjects, the possibility of a type 2 error exists. Finally, Anastas et al compared a Welch-Allyn Vital Signs monitor with an aneroid instrument in 53 subjects with atrial fibrillation. 17 They found no significant difference in systolic blood pressure but a 3.3 mmhg difference in diastolic blood pressure. They did not use a comparison group in normal sinus rhythm. Ambulatory monitors using oscillometry have been used in patients with atrial fibrillation. Blood pressure variability, as assessed by the standard deviation from the mean seems greater than in subjects in sinus rhythm. 18 Whether the readings are as accurate in predicting complications of hypertension, as in patients in sinus rhythm is unknown. Our protocol compared widely used regular monitors, includes a sinus rhythm control group and is adequately powered to detect significant differences. Neither monitor performed up to current standards. It may be argued that, had we taken more readings with the automated monitors, their accuracy would have approached that of mercury sphygmomanometry. Nevertheless, in clinical practice, therapeutic decisions will often be made on the basis of a single reading. Advances in oscillometry will undoubtedly occur. 19 However, for patients with irregular heart rhythms, it appears likely that multiple manual blood pressure recordings using standard Riva-Rocci technique will continue to be required. Acknowledgments We thank Ms. Fran Doyle, RN for helping with blood pressure recordings. References 1. Kannel WB, Wolf PA, Benjamin EJ et al. Prevalence, incidence, prognosis, and predisposing conditions for atrial fibrillation: population-based estimates. Am J Cardiol 1998; 82: 2N-9N. 2. Hennersdorf MG, Schueller PO, Steiner S et al. Prevalence of paroxysmal atrial fibrillation depending on the regression of left ventricular hypertrophy in arterial hypertension. Hypertens Res 2007; 30: 535-540. 3. Anon. Summary of the 2007 European Society of Hypertension (ESH) and European Society of Cardiology (ESC) guidelines for the management of arterial hypertension. Vasc Health Risk Manag 2007; 3: 783-795. 4. Williams B, Poulter NR, Brown MJ et al. Guidelines for management of hypertension: report of the fourth working party of the British Hypertension Society, 2004-BHS IV. J Hum Hypertens 2004; 18: 139-185. 5. Padwal RS, Hemmelgarn BR, Khan NA et al. The 2009 Canadian Hypertension Education Program recommendations for the management of hypertension: Part 1--blood pressure measurement, diagnosis and assessment of risk. Can J Cardiol 2009; 25: 279-286. 6. Pickering TG, Hall JE, Appel LJ et al. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005; 45: 142-161. 7. Chobanian AV, Bakris G, Black H, et al. The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 report. - JAMA 2003 May 21;289 (19 ):2560-72 :2560-2572.8. Pickering TG. What will replace the mercury sphygmomanometer? Blood Press Monit 2003; 8: 23-25. 8. Pickering TG. What will replace the mercury sphygmomanometer? Blood Press Monit 2003; 8: 23-25. 9. Lamb T, Wilson M, Wilson T. AF and hypertension: is my home monitor fibbing? Perspectives in Cardiology 2009; 24: 19-22. 10. Wilson TW, Quest DW, Wilson M et al. A cardiovascular risk factor reduction clinic. Can J Cardiol 1999; 15: 887-891. 11. Wilson M, Card SE, Buxton AC et al. A randomized trial of teaching blood pressure measurement to medical students. Annals Royal College Physicians and Surgeons of Canada 2002; 35: 15-18. 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E61.

12. O'Brien E, Pickering T, Asmar R et al. Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press Monit 2002; 7: 3-17. 13. Geddes LA, Voelz M, Combs C et al. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng 1982; 10: 271-280. 14. van Montfrans GA. Oscillometric blood pressure measurement: progress and problems. Blood Press Monit 2001; 6: 287-290. 15. Stewart MJ, Gough K, Padfield PL. The accuracy of automated blood pressure measuring devices in patients with controlled atrial fibrillation. J Hypertens 1995; 13: 297-300. 16. Jani B, Bulpitt CJ, Rajkumar C. Blood pressure measurement in patients with rate controlled atrial fibrillation using mercury sphygmomanometer and Omron HEM-750CP deice in the clinic setting. J Hum Hypertens 2006; 20: 543-545. 17. Anastas ZM, Jimerson E, Garolis S. Comparison of noninvasive blood pressure measurements in patients with atrial fibrillation. J Cardiovasc Nurs 2008; 23: 519-524. 18. Olsen R, Amlie A, Omvik P. Twenty-four-hour ambulatory blood pressure monitoring in atrial fibrillation. Blood Press Monit 2002; 7: 149-156. 19. Thien T, Adiyaman A, Staessen JA et al. Blood pressure measurement in the year 2008: revival of oscillometry? Neth J Med 2008; 66: 453-456. Correspondence to: Dr. TW Wilson Department of Medicine Royal University Hospital 103 Hospital Drive SASKATOON, SK, Canada, S7N 0W8 Tel: 306 966 7967 Fax: 306 966 797 Email: thomas.wilson@usask.ca 2010 CIM Clin Invest Med Vol 33, no 1, February 2010 E62.