ph fax

Similar documents
Unrivaled, End-to-End

REF # DOC #80XXX-00. STANDARD IMAGING, INC Deming Way Middleton, WI TEL TEL FAX

REF # DOC #80XXX-00. STANDARD IMAGING, INC Deming Way Middleton, WI TEL TEL FAX

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

Nuclear Associates

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

SRS Uncertainty: Linac and CyberKnife Uncertainties

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

EORTC Member Facility Questionnaire

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Quality Assurance in Stereotactic. Radiotherapy. Swedish Cancer Institute Seattle, WA

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

X-Ray Guided Robotic Radiosurgery for Solid Tumors

EXACTRAC HIGHLY ACCURATE PATIENT MONITORING

Volumetric Modulated Arc Therapy - Patient Specific QA

Work partially supported by VisionRT

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

such as lung cancer or breast cancer

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

A new geometric and mechanical verification device for medical LINACs

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Protura Robotic Patient Positioning System. for efficiency + performance

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Case Study. Institution Farrer Park Hospital

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

Universal Couchtop. for safe, efficient treatment planning + imaging. Long Extension. 1 Piece

The Journey of Cyberknife Commissioning

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

SBRT REQUIRES: STEREOTACTIC BODY RADIOTHERAPY STEREOTACTIC BODY RADIOTHERAPY (SBRT) (SBRT) What s s in a name? Stereotactic Body Radiotherapy

Quality assurance in external radiotherapy

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

In-Room Radiographic Imaging for Localization

Quality assurance program for prototype stereotactic system developed for Neptun 10 PC linac

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Dosimetry of the Gamma Knife TM

Cyberknife Stereotactic Treatment

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

Small field diode dosimetry

In-Room Radiographic Imaging for Localization

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Dosisverifikation mit Delta 4 Discover während der Behandlung

Limits of Precision and Accuracy of Radiation Delivery Systems

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Elekta Synergy Digital accelerator for advanced IGRT

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Basic dosimetric data for treatment planning system

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Eric E. Klein, Ph.D. Chair of TG-142

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Image Registration: The Challenge for QA Centers

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Credentialing for the Use of IGRT in Clinical Trials

Diode calibration for dose determination in total body irradiation

IMRT QA: Point Dose Measurements or 2D Array?

A Patient s Guide to SRS

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

AbstractID: 8073 Title: Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Seattle, WA May 2008

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Additional Questions for Review 2D & 3D

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

Transcription:

This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY USA 2008 Standard Imaging, Inc. 10523 800-221-0111 1294-21 (12.08)

PHANTOMS COMPLETE, stereotactic radiosurgery/srt QA The Lucy 3D QA Phantom performs image transfer QA, dosimetry QA, and machine QA within exact coordinate system of commercially available stereotactic head frames

The Lucy 3D QA Phantom verifies and assures stereotactic (SRS/SRT) precision Current advances in SRS/SRT incorporate many steps in the imaging, planning, and treatment chain. At each step, it is critical to minimize the chance of error. Although errors in each step may be small, cumulative errors over the entire SRS/SRT chain may become clinically significant. To assess the overall performance of an SRS system, the versatile Lucy 3D QA Phantom is essential to assure the precision and accuracy of each step, and of the entire treatment process. Imaging and Dose Assessment The Lucy 3D QA Phantom is compatible with CT and MRI imaging modalities providing assessment of TPS calculations for image fusion, distance measurements, volume measurements, and dose. These parameters are evaluated within the exact coordinate system of commercially available stereotactic head frames. Critical dimensions of the Lucy 3D QA Phantom are made with a tolerance of 0.1 mm or less1. The Lucy 3D QA Phantom will accomplish many of the commissioning requirements of your TPS2. Ideal for commissioning and stereotactic treatment verification. 1. R.W. Schulte, T.S. Lee, K.E. Schubert, MRI Distortion Correction for Submillimeter-Precision Functional Radiosurgery, Med. Phys. 66(3), 5252 (2006). 2. AAPM Radiation Therapy Committee TG 53, Quality assurance for clinical radiotherapy treatment planning, Med. Phys. 25(10), October 1998.

Image Transfer QA Minimize clinically significant errors by quantifying system variances in the imaging chain to within 0.1 mm. Increase your confidence in the capability of your system to correctly image known volumes, distances, and geometries. Confirm that CT/MRI/Angio scans and image fusion are accurately recreated within the TPS therefore assuring accurate target dose delivery. Locks into SRS Frames Quickly and precisely position the Lucy 3D QA Phantom within the manufacturer s exact coordinate system with the Stereotactic Frame Interface, available for commercial SRS systems. SRS Patient Dosimetry QA Obtain absolute dose, relative dose, and point dose dosimetry QA measurements at isocenter and at exact positions off isocenter, using standard ion chambers, film, diodes, gels, MOSFETs, and TLDs you currently use. Radiation Alignment Machine QA Assess optical and geometric isocenter, laser alignment, indexed table positioning alignment, and positioning coordinates for a comprehensive evaluation of geometric accuracy. The Lucy phantom, with the new MRI Isocentric Volume Insert, is the only option I have found that provides for a true end-to-end test of radiosurgery accuracy, addressing both frame-based and frameless image-guided localization. It is an essential tool for any physicist responsible for quality assurance of a radiosurgery service. Sam S. Hancock, Ph.D., DABR, DABMP Chief Medical Physicist Southeast Missouri Hospital Frameless Stereotactic Radiosurgery The Lucy 3D QA Phantom mounted on the Precision Leveling and Rotational Alignment Base simulates patient positioning for frameless cranial or body SRS. The 3D geometry of the patient set-up can be examined throughout 36O in the. transverse, sagittal, and coronal planes

Locks into SRS Frames Make QA setup easy by locking the Lucy 3D QA Phantom to an exact SRS coordinate system T he Lucy 3D QA Phantom can be mounted on commercial SRS frames and positioned for CT, MRI, and angiographic imaging. The SRS frame with Lucy 3D QA Phantom is then mounted on the treatment couch and treated to simulate a patient. Quickly and precisely position the Lucy 3D QA Phantom within the manufacturer s exact coordinate system. The Lucy 3D QA Phantom positioned on a TomoTherapy couch using a Varian / Radionics / CRW/BRW frame BrainLAB Stereotactic Frame Interface The Lucy 3D QA Phantom in a Radionics localizer box being positioned for a CT scan Varian / Radionics / CRW/BRW Stereotactic Frame Interface Leksell Gamma Knife Stereotactic Frame Interface Lock the Lucy 3D QA Phantom into your SRS coordinate system

The QA Process Verify and control critical elements of the SRS process Image Fusion Planned Dose to Delivered Dose If both CT and MRI scans are imported into the treatment planning To compare the delivered radiation pattern with the planned software, the images can be fused using the tools within the dose distribution, the film density pattern must be registered treatment planning software. The accuracy of the fused images with the planned dose distribution4. To compare the planned can be evaluated by comparing them to the known measurements dose to the delivered dose, an isodose distribution must be described in the Lucy 3D QA Phantom specifications. produced from the film and overlaid on the treatment plan Distance Measurements Exact manufacturing tolerances allow you to compare the Lucy 3D QA Phantom distance measurements to the distance measurements of images in treatment planning software3. Distance measurements should be evaluated at every step of isodose distribution. In a similar manner, the Dosimetry Insert for Ion Chamber may be used to confirm prescribed dose at the exact center of the Lucy 3D QA Phantom. The TLD Dosimetry Cassette and the MOSFET Dosimetry Cassette may also be used for dosimetry tests. the imaging process as errors may occur whenever images are Additional Information transferred from one program to another. The steps after which For a 14 step general guideline for using the Lucy 3D QA Phantom distance accuracy should be evaluated include: as a tool for stereotactic treatment verification, contact Standard 1. Acquisition of CT and MRI images Imaging and ask for Technical Note 4647 Stereotactic Treatment 2. Transfer of the CT and MRI images to treatment Verification with the Lucy 3D QA Phantom or visit our website: planning software www.standardimaging.com 3. All fusion and planning steps 4. Saving and recalling plans from record and verify software Distance measurements should also be verified in all planes used for treatment planning, transverse, sagittal, and coronal. Applications Make QA set up easy by locking the Lucy 3D QA Phantom to your exact SRS coordinate system Evaluate manufacturer s automatic patient positioning systems with film, ion chambers, diodes, gels, MOSFETs, and TLDs Commission your system quickly by using the Lucy 3D QA Phantom to confirm specifications upon delivery of equipment Check cone beam CT performance in 3D space Use as a QA tool as described in the performance validation references (see end of brochure) and in AAPM Task Group reports Use for QC and dosimetry measurements with framed and frameless SRS systems Evaluate CT/MRI/Angio fusion at 20 points in 3D space Effectively evaluate six degrees of freedom positioning systems in 3D space with convenient image markers 3. R. Ramani, et.al., A QA phantom for dynamic stereotactic radiosurgery: Quantitative measurements, Med. Phys. 22(8), August 1995. 4. R. Ramani, et.al., The use of Radiochromic film in treatment verification of dynamic stereotactic radiosurgery, Med. Phys. 21(3), 1994. Verify and control critical elements of the SRS process with the Lucy 3D QA Phantom

Comprehensive QA Packages All of the accessories you need to perform Dosimetric, CT, and MRI QA Dosimetry QA Accessory Package Dosimetry Insert for Ion Chamber Several Chamber Cavity Options Available Target/Treatment Verification Film Cassette REF 70078 Dosimetry Film Cassette for Three 2.5 x 2.5 Inch Films REF 70084 The ultimate goal of SRS is to be able to deliver a prescribed dose to within an accuracy of ± 2 %. A treatment can be planned to easily measure the absolute dose with the ion chamber. This cassette is made of black acrylic to accommodate radiochromic or conventional therapy film. It positions one 3 inch by 3 inch film at the exact center of the Lucy 3D QA Phantom. The insert is made so the center of the ion chamber s active volume is positioned at the geometric center of the Lucy 3D QA Phantom. A clamp is provided to fix the chamber in position. There are sharp markers in the Target/ Treatment Verification Film Cassette which produce four impressions forming a square on the film equidistant from the center used for isocentricity and distance measurement tests. This cassette positions three films at the exact center of the Lucy 3D QA Phantom for film dosimetry measurements. The two films on either side of the central film are separated by 2.25 mm acrylic spacers. The cassette and spacers are made of clear acrylic. Each insert accommodates a specific model ion chamber, however multiple inserts for different ion chambers can be used interchangeably with one Lucy 3D QA Phantom. Model A16 Exradin MicroChamber Collecting Volume: 0.007 cc Capable of measuring extremely small field sizes of 3.4 mm by 3.4 mm, allowing it to fit inside a 4 mm beam. sold seperately The Lucy 3D QA Phantom showing the Exradin A16 Microchamber in position Planned dose to delivered dose verification

CT Imaging QA Accessory Package CT Marker Cylinders, set of Four REF 70072 Each marker cylinder contains five 2 mm diameter aluminum spheres which are spaced 5 mm center to center. The targets within the four marker cylinders create a rectangle 30 mm on each side. The CT Market Cylinders, REF 70072, when used in conjunction with the MRI Marker Cylinders, REF 70074, are used to evaluate the fusion function of treatment planning programs. CT Grid Insert for Spatial Distortion REF 70075 CT Volume Insert with Three Known Geometries Electron Density QA Insert This insert is a two dimensional metallic grid specifically designed for checking image distortion and symmetry. Grid lines are visible at conventional settings for CT scanners. The grid wires are 0.5 mm aluminum and are spaced 5 mm apart. The wires originate at the center of the insert. REF 70076 The Electron Density Insert is designed to be a quick and convenient check of the CT density tables utilized by the CT imaging and treatment planning systems. The insert is comprised of five materials of differing density: Blue Water (a water equivalent material), air, trabecular bone, cortical bone, and adipose. This insert has three irregularly shaped known air volumes of 250, 750, and 1750 mm3. The volume insert is used to evaluate the integrity of images as they are moved from one imaging system to another, challenging the TPS to accurately recreate. REF 70104 MRI Imaging QA Accessory Package MRI Marker Cylinders, set of Four REF 70074 Each marker cylinder contains five 2 mm diameter spheres filled with mineral oil. The spheres are spaced 5 mm center to center. The targets within the four marker cylinders create a rectangle 30 mm on each side. The MRI Marker Cylinders, REF 70074, when used in conjunction with the CT Marker Cylinders, REF 70072, are used to evaluate the fusion function of treatment planning programs. MRI Volume Insert with Three Known Geometries REF 70077 This insert has three irregularly shaped volumes which are filled with mineral oil. The volume insert is used to evaluate the integrity of images as they are moved from one imaging system to another. MRI Signal Generator REF 70063 The signal generator fits into the large cavity of the Lucy 3D QA Phantom and contains a manganese chloride solution. The volume of manganese is required to produce enough MRI signal strength to image the MRI Marker Cylinders. Evaluate the accuracy of CT and MRI image fusion

Mechanical Alignment Pointers Perform radiation alignment machine QA Radiation Alignment Pointers REF 70079 The Radiation Alignment Pointers attach to the Precision Leveling and Rotational Alignment Base for secure positioning. Two alignment pointers are provided, one with a 5 mm tungsten sphere tip and one with a 3 mm aluminum sphere tip. The spheres are used for optical alignment and also for radiation isocenter alignment when used with radiographic film. Assess optical and geometric isocenter, laser alignment, indexed table positioning alignment, and positioning coordinates for a comprehensive evaluation of geometric accuracy. Winston-Lutz Test for Radiation Isocenter The tungsten or aluminum sphere tips of the Radiation Alignment Pointers can be imaged for the Winston-Lutz test. The sphere is placed at the mechanical isocenter of the treatment room as determined by the room lasers. The accelerator is then rotated to each required gantry angle, and the sphere is imaged with film or an EPID. Utilize PIPSpro Comprehensive Software to perform an automatic analysis of these images to determine the absolute size and position of the radiation isocenter. Contact your Standard Imaging representative for more information or a demonstration. Additional Accessories MRI Isocentric Volume Insert REF 70107 The MRI Isocentric Volume Insert features an isocentrically placed target of known volume which allows for integrated testing of CT and MR imaging, image fusion, and treatment planning. Used in combination with available dosimetric inserts for treatment delivery verification, the Lucy Phantom is a complete end-toend QA system for your SRS program. This insert was developed in cooperation with Sam Hancock, Ph.D. of Southeast Missouri Hospital5. 3D Volumetric Target Dosimetry Kit REF 76013 The Multiple Target Shapes Insert offers you complex geometric targets of known dimensions and volumes to challenge and assess the treatment planning system s re-construction capabilities. Evaluate the ability of your TPS to include and avoid critical structures and then utilize the included film cassette to verify dose delivery to the target. Tools for radiation isocenter and optical alignment 5 Hancock, S., End-to-End QC of Stereotactic Radiosurgery with the Lucy Phantom, Presentation, BrainLAB User s Meeting, Chicago, IL June 2008.

Additional Accessories continued IGRT kv and X-Ray Angiography Marker Cylinders REF 70073 Dosimetry Film Cassette for Three 3 x 3 Inch Films REF 70091 Dosimetry Cassette for Thirteen 2.5 x 2.5 Inch Films REF 70089 The IGRT kv and X-Ray Angiography Marker Cylinders are used with the CT Marker Cylinders and the MRI Marker Cylinders to evaluate the fusion function of treatment planning programs. This cassette positions three films at the exact center of the Lucy 3D QA Phantom for film dosimetry measurements. The two films on either side of the central film are separated by 2.25 mm acrylic spacers. The cassette and spacers are made of clear acrylic. This cassette positions thirteen films in the large void in one hemisphere of the Lucy 3D QA Phantom for film dosimetry measurements. The cassette and spacers are made of clear acrylic. TLD Dosimetry Inserts for TLD Detectors Dosimetry Insert for 3mm TLD Detectors REF 70059 Dosimetry Insert for 1mm TLD Detectors REF 70106 This dosimetry insert allows the positioning of up to [49 TLDs - 3mm] or [85 TLDs - 1mm] in the central plane of the Lucy 3D QA Phantom. One TLD is positioned at the exact center of the phantom. The size of the cavities are [3.4 x 3.4 x 1 mm] or [1.5 x 1.5 x 1 mm]. MOSFET Dosimetry Insert for MOSFET Detectors REF 70058 This dosimetry insert positions 15 MOSFET detectors in the central plane of the Lucy 3D QA Phantom. One detector is positioned at the exact center of the phantom. The size of the cavities are 2.5 x 8 x 1 mm. Inserts for a wide variety of applications

Lucy 3D QA Phantom Components Lucy 3D QA Phantom Lucy MRI Imaging QA Accessory Package Spherical Lucite Phantom MRI Marker Cylinders, set of four Precision Leveling and Rotational Alignment Base MRI Volume Insert with three known geometries Pelican Carrying Case MRI Signal Generator Stereotactic Frame Interfaces Radiation Alignment Pointers BrainLAB Includes a 5 mm tungsten sphere tip and a 3 mm aluminum sphere tip CyberKnife Leksell Gamma Knife Additional Accessories Sandstrom Stereo-Adapter IGRT kv and X-Ray Angiography Marker Cylinders, set of four TomoTherapy Hi-Art System Dosimetry Film Cassette for three 3 x 3 inch films Varian / Radionics / CRW/BRW Dosimetry Film Cassette for thirteen 2.5 x 2.5 inch films Precision Leveling and Rotational Alignment Base for frameless systems and extracranial work MRI Isocentric Volume Insert Lucy Dosimetry QA Accessory Package Other accessories are available for many ion chambers, TLD dosimetry, diode detectors, MOSFET detectors and MRI compatible bases for phantom positioning Dosimetry Insert for ion chamber of your choice Dosimetry Film Cassette for three 2.5 x 2.5 inch films Target / Treatment Verification Film Cassette 3D Volumetric Target Dosimetry Kit Dosimetry Instruments SuperMAX Electrometer Lucy CT Imaging QA Accessory Package MAX 4000 Electrometer CT Marker Cylinders, set of four Exradin A16 MicroChamber CT Volume Insert with three known geometries Exradin A14SL MicroChamber CT Grid Insert for spatial distortion Exradin A1 Thimble Ion Chamber Electron Density QA Insert Exradin A18 Thimble Ion Chamber Visit www.standardimaging.com or call 800-261-4446 for more details. p e r f o r m a n c e validati o n R. Ramasesham, M. Heydarian, Comprehensive quality assurance for stereotactic radiosurgery treatments, Phys. Med. Biol. 48(2003) N119-N205. R. Ramani, et.al., A QA phantom for dynamic stereotactic radiosurgery: Quantitative measurements, Med. Phys. 22(8), August 1995. R. Ramani, et.al., The use of Radiochromic film in treatment verification of dynamic stereotactic radiosurgery, Med. Phys. 21(3), 1994. AAPM Radiation Therapy Committee TG 53: Quality assurance for clinical radiotherapy treatment planning, Med. Phys. 25(10), October 1998. AAPM Radiation Therapy Committee TG 66: Quality assurance for computed-tomography simulators and the computed-tomography-simulation process, Med. Phys. 30(10), October 2003 This product is available through: JRT Associates 5 Nepperhan Avenue, Suite 2B Elmsford, NY 10523 800-221-0111 Versatile and comprehensive QA

l u c y 3 d qa p h a n t o m ( r e f 9 1 2 1 0 ) s p e c i f i c ati o n s Lucy 3D QA Phantom Acrylic Sphere 140 mm (5.51 in) diameter Blank Filler Plug (HEMI-C) 81 x 81 x 35 mm (3.19 x 3.19 x 1.38 in) Blank Filler Plug (HEMI-A) 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) Accessory Port for Ionization Chambers 8 mm (.31 in) diameter Precision Leveling and Rotational Alignment Base Height 30 cm (11.81 in) Depth 23 cm (9.06 in) Width 20 cm (7.87 in) Weight 3.4 kg (7.5 lbs) Dosimetry QA Accessories Dosimetry Insert for Ion Chamber 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) Dosimetry Film Cassette for Three 2.5" x 2.5" films 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) Target/Treatment Verification 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) - Film Cassette - Four Titanium Fiducial Markers 60 mm (2.36 in) square pattern - One Titanium Fiducial Marker for Orientation Dosimetry Film Cassette for Three 3.0" x 3.0" films 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) Dosimetry Film Cassette for Thirteen 2.5" x 2.5" films 85 x 85 x 35 mm (3.35 x 3.35 x 1.38 in) CT QA Accessories CT Marker Cylinders, set of four 10 mm (.39 in) length, 25 mm (.98 in) diameter - Five 2.0 mm Aluminum Spheres per cylinder 5 mm (.197 in) center-to-center CT Volume Insert with 3 irregular known volumes 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) - Geometry One Area 250 mm, Volume 2500 mm3 - Geometry Two Area 750 mm, Volume 7500 mm3 - Geometry Three Area 1750 mm, Volume 17500 mm3 CT Grid Insert for spatial distortion 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) 0.5 mm (.02 in) aluminum wire spaced 5 mm (.197 in) apart MRI QA Accessories MRI Marker Cylinders, set of four 10 mm (.39 in) length, 25 mm (.98 in) diameter - Five 2.0 mm mineral oil spheres per cylinder 5 mm (.197 in) center-to-center MRI Volume Insert with three known Geometries 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) - Geometry One Area 250 mm, Volume 1700 mm3 - Geometry Two Area 750 mm, Volume 5250 mm3 - Geometry Three Area 1750 mm, Volume 12250 mm3 MRI Signal Generator 85 x 85 x 35 mm (3.35 x 3.35 x 1.38 in) - Cavity Filled with Manganese Chloride Imaging QA Accessories IGRT Localization & Angiography Marker Cylinders, set of four 10 mm (.39 in) length, 25 mm (.98 in) diameter - One 2.0 mm Lead Sphere in each cylinder Radiation Alignment Pointer Includes 3 mm Optical Alignment Tip and Radiation 5 mm Alignment Tip MOSFET Dosimetry Cassette Cassette with 15 cavities 2.5 x 8 x 1 mm to accommodate MOSFET Diodes 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) TLD Dosimetry Cassette Cassette with 49 cavities 3.4 x 3.4 x 1 mm to accommodate TLDs 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) Cassette with 85 cavities 1.5 x 1.5 x 1 mm to accommodate TLDs 85 x 85 x 10 mm (3.35 x 3.35 x.39 in) Pelican is a trademark of Pelican Products, Inc. Leksell Gamma Knife is a registered trademark of Elekta Instruments, Inc. CyberKnife is a registered trademark of Accuray Incorporated. TomoTherapy Hi-Art System is a trademark of TomoTherapy, Inc. Specifications subject to change without notice. All critical dimensions have a tolerance of 0.1 mm or less