Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Similar documents
Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

HDR Brachytherapy: Results and Future Studies in Monotherapy

Abstract Purpose: Material and methods: Results: Conclusions: Key words: Purpose Address for correspondence:

SunCHECK Patient Comprehensive Patient QA

Radiobiological evaluation of the influence of dwell time modulation restriction in HIPO optimized HDR prostate brachytherapy implants

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Brachytherapy Planning and Quality Assurance

From position verification and correction to adaptive RT Adaptive RT and dose accumulation

Comprehensive and Practical Brachytherapy March 04-8 March 2018, Ljubljana, Slovenia Day 1 Sunday 4 March 2018

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Leila E. A. Nichol Royal Surrey County Hospital

I. Equipments for external beam radiotherapy

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Content. Acknowledgments. Prostate brachy LDR Prostate brachy HDR. Use of permanent seeds and HDR in prostate: Current practice and advances

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

DOSIMETRIC OPTIONS AND POSSIBILITIES OF PROSTATE LDR BRACHYTHERAPY WITH PERMANENT I-125 IMPLANTS

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

Feasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

Real-time brachytherapy for prostate cancer implant analysis

EORTC Member Facility Questionnaire

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Recent proceedings in Brachytherapy Physics

IMRT Planning Basics AAMD Student Webinar

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

The Physics of Oesophageal Cancer Radiotherapy

HDR vs. LDR Is One Better Than The Other?

CBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.

Image Guided Proton Therapy and Treatment Adaptation

Automated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist

Quality management for Breast Brachytherapy.

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Overview of Clinical and Research Activities at Georgetown University Hospital

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

Radiotherapy Advances

Outline. Contour quality control. Dosimetric impact of contouring errors and variability in Intensity Modulated Radiation Therapy (IMRT)

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

IMAT: intensity-modulated arc therapy

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

SUPERIORITY OF A REAL TIME PLANNING TECHNIQUE OVER IMAGE GUIDED RADIATION THERAPY FOR THE TREATMENT OF PRIMARY PROSTATE CANCERS

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

ART for Cervical Cancer: Dosimetry and Technical Aspects

Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Evaluation of Normal Tissue Complication Probability and Risk of Second Primary Cancer in Prostate Radiotherapy

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Radiotherapy physics & Equipments

Changing Paradigms in Radiotherapy

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy

A review of RTTQA audit. Karen Venables

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

Implementation of advanced RT Techniques

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA

20 Prostate Cancer Dan Ash

Linac Based SBRT for Low-intermediate Risk Prostate Cancer in 5 Fractions: Preliminary Report of a Phase II Study with FFF Delivery

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical

Patient and Linac QA - present and the future

Activity report from JCOG physics group

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Definitions. Brachytherapy in treatment of cancer. Implantation Techniques and Methods of Dose Specifications. Importance of Brachytherapy in GYN

IMRT for Prostate Cancer

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Review of brachytherapy in

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

Ritu Raj Upreti, S. Dayananda, R. L. Bhalawat*, Girish N. Bedre*, D. D. Deshpande

PROSTATE CANCER BRACHYTHERAPY. Kazi S. Manir MD,DNB,PDCR RMO cum Clinical Tutor Department of Radiotherapy R. G. Kar Medical College

Jean Pouliot, PhD Professor and Vice Chair, Department of Radiation Oncology, Director of Physics Division

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

Unrivaled, End-to-End

Trina Lynd, M.S. Medical Physicist Lifefirst Imaging & Oncology Cullman, AL Tri-State Alabama, Louisiana and Mississippi Spring 2016 Meeting April

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

TRANSRECTAL ULTRASOUND-GUIDED PROSTATE BRACHYTHERAPY

Limits of Precision and Accuracy of Radiation Delivery Systems

New research in prostate brachytherapy

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

RPC s Credentialing Programs for Clinical Trials

Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance

DELIVERED PATIENT DOSE

Additional Questions for Review 2D & 3D

Transcription:

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer? Dimos Baltas 1,3, Natasa Milickovic 1, Nikolaos Zamboglou 2 1 Dept. of Medical Physics & Engineering, 2 Strahlenklinik, Klinikum Offenbach GmbH 63069 Offenbach, Germany 3 Nuclear and Particle Physics Section, Physics Department, University of Athens, Panepistimioupolis, Ilisia, 157 71 Athens, Greece E-mail: dimos.baltas@klinikum-offenbach.de Slide 1

List of Content BRT versus ERT from Dosimetry Point of View Treatment Planning and Dose Calculation Treatment Delivery Treatment Delivery Verification Slide 2

Modern Radiation Therapy The Field / Beam: ERT BRT 1 3 1: Catheter / Needle 3D 3D Slide 3

Modern Radiation Therapy Beam Shaping: Plane Field Catheter/Needle/Applicator MLC 2.5 mm or 5.0 mm or 10.0 mm 1.0 mm 2.5 mm 5.0 mm 10.0 mm?? mm MLC ERT BRT Slide 4

Modern Radiation Therapy Beam Shaping: Plane BRT: Catheter/Needle MLC p + / Ions: Raster Scanning 1.0 mm 2.5 mm 5.0 mm 10.0 mm?? mm Raster / Vector Scanning Slide 5

Modern Radiation Therapy Dosimetric Kernel ERT 10 : 1 BRT Depth Dose 1.0 0.9 0.8 18 MV 0.7 0.6 10 MV 0.5 4 MV 0.4 0.3 6 MV 0.2 0.1 0 5 10 15 20 25 30 35 Depth (cm) Dose Rate Normalized to 1.0 cm 3.0 2.8 2.6 2.4 2.2 2.0 1.8 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0.0 20 kev 25 kev 30 kev 40 kev 50 kev 60 kev 70 kev 80 kev 90 kev 100 kev 150 kev 200 kev 300 kev 400 kev 667 kev 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 Radial Distance (cm) Slide 6

Modern Radiation Therapy Dosimetric Kernel ERT BRT Dose Rate Normalized to 1.0 cm 3.0 2.8 2.6 2.4 2.2 2.0 1.8 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0.0 20 kev 25 kev 30 kev 40 kev 50 kev 60 kev 70 kev 80 kev 90 kev 100 kev 150 kev 200 kev 300 kev 400 kev 667 kev 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 Radial Distance (cm) Slide 7

Modern Radiation Therapy Dosimetric Kernel Dose Rate Normalized to 1.0 cm 3.0 2.8 2.6 2.4 2.2 2.0 1.8 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0.0 BRT 20 kev 25 kev 30 kev 40 kev 50 kev 60 kev 70 kev 80 kev 90 kev 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 Radial Distance (cm) 100 kev 150 kev 200 kev 300 kev 400 kev 667 kev Radial Dose Fucntion g(r) BRT r -2 = 0.007 1.6 1.5 1.4 1.3 100 kev 1.2 60 kev 1.1 Yb-169 80 kev 1.0 0.9 Ir-192 0.8 0.7 0.6 50 kev 0.5 0.4 20 kev 40 kev 0.3 30 kev 0.2 0.1 0.0 0 2 4 6 8 10 12 14 16 Radial Distance (cm) Slide 8

Modern Radiation Therapy Dose Shaping: Intensity Modulation (2D) MLC : Step & Shoot ERT BRT Bixel/Beamlet/Segment Dwell Position MUs Dwell Time Slide 9

Modern Radiation Therapy Dose Shaping ERT Energy Dwell Position (3D) BRT Multi-Spots Relative Depth Dose 1.5 1.4 1.3 1.2 159 MeV Protons 192 Ir (400 kev) 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 Depth in Water (cm) Slide 10

Dose Shaping: Intensity Modulation (3D) MSS: Step & Shoot Spot K(r ) F( { r }, { t}) D( r ) = F( r, r { t} ) K( ) Slide 11

Modern Radiation Therapy BRT versus ERT Similarities and Differences Dosimetric Kernel Particles (Spot) Delivery Technology IMRT (X, P) (Modulation, Dose-Volume-Prescriptions ) Dose Distribution SRS / SBRT (Inhomogeneity) Slide 12

Modern Radiation Therapy The Field / Beam: ERT BRT Set-Up of the anatomy (Reconstruction) at the 3D-Reference System of the LINAC. Position of Beams and their configuration (MLC-settings) can then be automatically tracked based on integrated technology into the LINAC (electronics, EPID, kv- or MV-CBCT, ) Anatomy: is reconstructed relative to the Imager Coordinate System (DICOM). Beams: Catheters/needles/Applicators have to be placed and reconstructed by user-driven actions relative to DICOM (Anatomy). ASDPs (MLC) are not tracked automatically but this assumes firstly the reconstruction of the catheters/applicators. There is no Imaging of ASDPs delivery with respect to anatomy. Slide 13

List of Content BRT versus ERT from Dosimetry Point of View Treatment Planning and Dose Calculation Treatment Delivery Treatment Delivery Verification Slide 14

Workflow Imaging and Anatomy Definition Pre-Planning Imaging and Implantation and Treatment Plan Treatment Delivery Slide 15

Current Treatment Planning The Role of Imaging in Anatomy Definition and related uncertainty effects can be assumed similar to ERT with the following exceptions: Image distortion due inserted applicators (e.g. MR and metallic needles, seed artefacts in CTimaging,...) For BRT there exists the following specific issue: Imaging is also used for Identification and Reconstruction of the Catheters/Needles/Applicators (ERT: Beams). Based on their geometry the possible SDPs and required ASDPs are extracted (ERT: MLC settings, beamlets) Slide 16

Sources of Uncertainties Imaging, Fusion/Registration, Contouring Imaging, Fusion/Registration, RTP: Beam - Reconstruction +δvois Treatment Delivery +δvois +δbeams +δbeams +δrtp +δmachine Treatment Delivery: Machine + Anatomy + Beams RTP: Dose Calculation Slide 17

Treatment Planning Imaging, Fusion/Registration, RTP: Beam - Reconstruction +δbeams δdose e.g. Gynaecological Implants: 5% in D 2cm³ per mm offset for bladder & rectum < 4% per mm offset for CTV (D 90, D 100 ) Data available limited to gynaecological and prostate implants! Slide 18

RTP: Dose Calculation +δrtp Beam Data Entry (TG-43) Source Strength S K TG-43 Tables Dose Calculation Clinic measures S K (k=1): 1.3% for LDR low Energy 1.5% for HDR high Energy Total Dose Calculation TG-43 (k=1): 4.4% for low Energy 3.4% for high Energy All values for Best practice Uncertainties Slide 19

List of Content BRT versus ERT from Dosimetry Point of View Treatment Planning and Dose Calculation Treatment Delivery Treatment Delivery Verification Slide 20

Treatment Delivery: I. Machine +δmachine MLC precision of ± 1.0 mm Complex geometries: MLC accuracy 2.5 4.5 mm By courtesy of University of Vienna Attention: Dose distribution Depends not only on the position but also on the orientation (vector) of the sources Slide 21

Treatment Delivery: I. Machine Prostate: 16 catheters & 86 ASDPs ( beamlets/ segments ) Slide 22

Treatment Delivery: I. Machine Prostate: 16 catheters & 86 ASDPs ( beamlets/ segements ) δbeams (Reconstruction) Reconstruction uncertainty of 1.0 mm (...) and 2.0 mm (---) + δmachine MLC / source positioning uncertainty of 1.5 mm along catheter (...) : 2% up to 200%, 5% above 200% (---) : 3% up to 150%, 9% above 150% Even in that case, the impact on DVH calculations is 3% Pantelis E, Papagiannis P, et al. Phys. Med. Biol. 49, 55-67, 2004 Slide 23

Treatment Delivery: I. Machine +δmachine MU - dwell time accuracy of 30 ms Effect of Rounding: less than 1% Effect of MLC travelling finite speed of source drive < 1% Slide 24

Treatment Delivery: II. Anatomy and Catheters/Applicators +δvois + δbeams Between: Plan & Delivery During Delivery Fraction to Fraction (Interfractional: single implant & multiple fractions) + δdose Delivery Slide 25

Treatment Delivery Almost all published data available are based on pretreatment imaging for the purpose of investigating anatomy & implant changes and thus of potential errors for following workflows (mainly gynaecological and prostate implants): Multi-fractions delivery with a single implant Multi-implants with single or multi factions delivery per implant Slide 26

Treatment Delivery Multi-fractions delivery with a single implant Implant U/S Plan U/S 1 st Delivery Imaging N th Delivery Plan CT 1 st Delivery N th Delivery Slide 27

Treatment Delivery Multi-fractions delivery with a single implant CT-based Use of reference: bony structures Implanted markers Slide 28

Treatment Delivery Multi-fractions delivery with a single implant δbeam & δvoi δdose Slide 29

Treatment Delivery Multi-fractions delivery with a single implant! Slide 30

Treatment Delivery Multi-fractions delivery with a single implant Slide 31

Treatment Delivery Multi-fractions delivery with a single implant δbeam Mean shift: f2-f1 = 7.9 mm f3-f1 = 3.9 mm δdose Slide 32

Treatment Delivery Multi-fractions delivery with a single implant Slide 33

Treatment Delivery Multi-fractions delivery with a single implant Slide 34

Treatment Delivery Multi-fractions delivery with a single implant δbeam Median shift: 7.5 mm [2.9, 23.9] Median t between CT & delivery 254 min [81, 367]! Slide 35

List of Content BRT versus ERT from Dosimetry Point of View Treatment Planning and Dose Calculation Treatment Delivery Treatment Delivery Verification Slide 36

Treatment Delivery Verification: Off-Line Procedure CT based 1 x Implant + 4 x Fractions Slide 37

Treatment Delivery Verification: Off-Line Procedure CT based 1 x Implant + 4 x Fractions 5 x cases, 18 x Fractions, 14 x TLDs Example Case # 1 50% 100% 150% 200% 300% PTV TLD catheter, 14 x TLDs Rectum Slide 38

Treatment Delivery Verification: Off-Line Procedure CT based 1 x Implant + 4 x Fractions 5 x cases, 18 x Fractions, 14 x TLDs Example Case # 2 Urethra 50% 100% 150% 200% 300% PTV TLD catheter, 14 x TLDs Slide 39

Treatment Delivery Verification: Off-Line Procedure CT based 1 x Implant + 4 x Fractions 5 x cases, 18 x Fractions, 14 x TLDs Dose (cgy) 1600 1400 1200 1000 800 600 400 200 0 Plato BPS 14.2.2 with D ref = 9,5 Gy TLD 1.Fraktion 04/06/2002 15:30 TLD 2.Fraktion 05/06/2002 08:30 TLD 4.Fraktion 06/06/2002 08.30 0 10 20 30 40 50 60 70 80 90 Distance from catheter tip (mm) Analysis Example Slide 40

Treatment Delivery Verification: Off-Line Procedure CT based 1 x Implant + 4 x Fractions Slide 41

Treatment Delivery Verification: Off-Line Procedure CT based 1 x Implant + 4 x Fractions Slide 42

Treatment Delivery Verification: Plan Currently common/published/realised CT imaging based Verification of anatomy and implant geometry before repeated Treatment Delivery for the same Implant Delivery Imaging From CT-Imaging to Treatment Delivery there exist in the majority of the cases a patient transportation (CT-Imaging room Treatment Room). No verification under Treatment Delivery conditions or no verification of dose delivery (except some efforts for a reliable in-vivo dosimetry). Slide 43

Treatment Delivery Verification: +δvois + δbeams Between: Plan & Delivery During Delivery +δdose Slide 44

The Clinical Procedure at Offenbach Clinic Pre-Plan Needle Implantation Treatment Delivery Live plan Slide 45

The Clinical Procedure at Offenbach Clinic: The QA-Tools 1 1 1 3 2 1 Slide 46

The Clinical Procedure at Offenbach Clinic: The QA-Tools 1 1 1 (1) Intraoperative Real Time Treatment Planning System Oncentra Prostate using 2D- and 3D- Ultrasound Imaging. Ultrasound probe remains in place during treatment enabling 2D and 3D verification at any time during the procedure. 1 Slide 47

The Clinical Procedure at Offenbach Clinic: The QA-Tools (2) Template-Perineum QA-Tool 2 Slide 48

The Clinical Procedure at Offenbach Clinic: The QA-Tools 3 (3) Measurement of Free Length of the needles is utilized for both reconstruction purposes (reconstruction of needle tip) and for quality control purposes for checking needle shifts during the treatment procedure Slide 49

Materials The mean volume of prostate (PTV) for the 25 cases is 35.8cm³ with minimum volume of 17.4cm³ and maximum of 59.6cm³ Average time spent between the clinical (1) and the pre-irradiation (2) 3D acquisition was 51.2 ± 7.1min*, and between the pre- (2) and post-irradiation (3) acquisition 19.3 ± 2.2min**. *as a teaching hospital, we regularly provide the residents training, that increases the time between the clinical set acquisition and irradiation **irradiation time that depends on the source activity and the prostate volume Slide 50

Results: Prostate & OARs Movement δvois Prostate The prostate movement in the Z direction (DICOM) was equal to 0mm for each of our acquisitions BASE plane remained unchanged. The movement in X (horizontal) and Y (vertical) directions was measured compared to the prostate gravity center on the reference plane. Urethra The shift of urethra was measured on the base, reference and apex plane. Rectum and Bladder By rectum and bladder, we have not noticed significant movements. Slide 51

Results: Prostate & OARs Movement δvois VOI Name Type r 21 [mm] r 31 [mm] r 32 [mm] Prostate PTV 0.5±0.6 [0.0 2.1] 0.6±0.7 [0.0 2.1] 0.1±0.3 [0.0 1.1] Urethra OAR Base plane 0.6±0.7 [0.0 2.7] 1.1±1.3 [0.0 5.1] 0.4±0.8 [0.0 3.4] Reference 0.6±0.7 [0.0 2.3] 0.8±0.9 [0.0 2.9] 0.3±0.5 [0.0 1.7] Apex 0.6±0.8 [0.0 3.3] 0.8±0.9 [0.0 2.8] 0.3±0.5 [0.0 1.7] Rectum OAR 0.3±0.4 [0.0 1.4] 0.4±0.4 [0.0 1.4] 0.1±0.2 [0.0 0.6] *The significance test for the observed differences is always done using the student paired t-test, two sided statistical method. For the paired t-tests the significance level was 0.05. Slide 52

Results: Catheter Changes/Movement δbeams The shifts of the catheters on base, reference and apex plane is measured between the clinical and pre-irradiation; clinical and postirradiation and pre- and post- irradiation plan. In all but one case, is the maximum needle movement less than 1.5 mm Slide 53

Results: Dosimetric Impact δdose Prostate D 90 [%] 110 105 100 95 Paired t-test: p 12 <0.001 p 13 <0.001 p 23 =0,411 D 90 of: clinical, pre-treatment and post-treatment plan 1 2 3 Clinical (1), pre (2)- and post (3)-treatment plan There are only three (of 25) pre- irradiation plans with D 90 values of 97.5, 97.7 and 97.8 and thus below 100%. The minimum value for D 90 for the post- irradiation plan was 96.2%. There are five (of 25) post- irradiation plans with D 90 value less than 100% with values within range of [96.2 to 98.3] %. Slide 54

General finding There is much more sense in observing the influence of the relative shift of catheters with reference to prostate on the DVH and also radiobiological parameters than the influence of absolute dislocation of needles and/or OARs separately. The cases that showed large absolute prostate and catheter displacements, showed small relative displacement and fulfilled the clinical protocol. Slide 55

Results: Dosimetric Impact δdose Slide 56

Results: Dosimetric Impact δdose COIN 0,95 0,90 0,85 0,80 0,75 Paired t-test: p 12 <0,001 p 13 <0,001 p 23 =0,409 COIN for: clinical, pre- and post-treatment plan 1 2 3 Clinical (1), pre (2)- and post (3)-treatment plan The COIN values for all preirradiation plans remain above 0.8. For 1 of 25 post- irradiation plans the COIN drops below 0.8 with values 0.78 Slide 57

Radiobiological Evaluation The radiobiological quantities P + and D were employed to evaluate the effectiveness of the examined dose distributions. P + is the probability of complication-free tumor control and can be approximated by: P + = P B - P I where P B and P I are the total probability of tumor control and injury, respectively and D is the biologically effective uniform dose, which is the dose that causes the same tumor control or normal tissue complication probability as the actual dose distribution, D given to the patient and it can be iteratively determined from: PD ( ) PD ( ) Slide 58

Radiobiological Evaluation Slide 59

Radiobiological Evaluation Summary of the model parameter values for the examined prostate cancer cases. D 50 is the 50% response dose, γ is the maximum normalized value of the dose-response gradient and s is the relative seriality, which characterizes the volume dependence of the organ. Radiobiological Model D 50 (Gy) γ s α/β Endpoint PTV 70.0 4.0 3.0 Control Urethra (U) 120.0 3.0 0.03 3.0 Stenosis Bladder (B) 80.0 3.0 0.3 3.0 Rectum (R) 80.0 2.2 0.7 3.0 Symptomatic contracture, volume loss Proctitis, necrosis, fistula, stenosis Slide 60

Results: Dosimetric Impact δdose The average DVHs of the prostate gland (red), urethra (black), bladder (pink) and rectum (blue) are presented for the three HDR treatment plans, namely 1: after implantation (clinical acquisition) (solid), 2: just before starting the treatment execution (preirradiation acquisition) (dashed) and 3: just after finishing the treatment delivery (postirradiation acquisition) (dotted-dashed). Here, the total dose of 34.5 Gy delivered by three fractions of 11.5 Gy is considered to be the total prescription dose (100%). Slide 61

Results: Dosimetric Impact δdose Slide 62

Paper Conclusions The measured mean shift of anatomy and needles ( beams ) is as low as 1.0mm For high modulated plans as those in HDR brachytherapy such small shifts result in dosimetric changes which are in general lower than 5%. These results demonstrate that quality assurance procedures have to be clinically implemented to guarantee anatomy and implant stability of the order of 1mm. This can only be realized without any manipulation of the implant and anatomy as done, for example, in the case of removing the US-probe before treatment delivery (Presentation by Dr. Milickovic on Monday)! Slide 63

The Clinical Procedure at Offenbach Clinic: The QA-Tools 1 1 1 3 2 1 Slide 64

Treatment Delivery Verification: An open Issue... Currently double check (4 x eyes)!!!!? Slide 65

Treatment Delivery Verification: The dose-fluence pattern incident verification, that will enable the real 3Ddose reconstruction (analogue to on-line fluence measurement in ERT), assumes the on-line Beamlets & MU Tracking: source dwell positions & dwell times. ERT Advance in-vivo dosimetry systems could partly support this. Currently we have to build on: Anatomy Reconstruction & Tracking Catheter Reconstruction & Tracking OCTAVIUS III by PTW Freiburg, Germany Under delivery conditions / during delivery! Slide 66

CLINICAL UNCERTAINTY ANALYSIS FOR BRACHYTHERAPY Slide 67

CLINICAL UNCERTAINTY ANALYSIS FOR BRACHYTHERAPY HDR 192 Ir temporary brachytherapy implants for prostate (Monotherapy or Boost) In this example, the relative dosimetric uncertainties are considered with the respect to the D 90 and V 150 dosevolume parameters for the prostate/ptv. Slide 68

Reproducibility of Treatment Delivery in Offenbach Reproducibility of Treatment Delivery: What does it mean for the individual patient or individual implant? 52 consequent patients with a total of 117 plans and treatment deliveries. Only Live Plans (Treatment Plans delivered are considered). Mean PTV of (38.9 ± 16) cm³ Range 15 97 cm³ Slide 69

Offenbach Dosimetric Constraints Case Monotherapy #1 and Boost Case #2 Prostate (PTV = CTV 1): D 90 100% V 100 90% V 150 35% Urethra: D 10 115% V 115% 10% D 0.1cm³ 120 % Rectum & Bladder D 10 75% V 75% 10% D 0.1cm³ 80 % Slide 70

How we contour Since the last 7 years We contour also bladder. Thus urethra stops Cranially where bladder begins. Slide 71

Reproducibility of Treatment Delivery I. PTV (38.9 ± 16) cm³ Volume [%] 100 90 80 70 60 N=117 50 40 30 20 10 0 0 50 100 150 200 250 Dose [%] D 90 = (102.5 ± 2.0)% V 100 = (92.0 ± 1.6)% V 150 = (29.1 ± 3.2)% Prostate (PTV = CTV 1): D 90 100% V 100 90% V 150 35% Slide 72

Reproducibility of Treatment Delivery II. Urethra (17.2 ± 3.2) cm³ Volume [%] 100 90 80 70 60 N=117 50 40 30 20 10 0 0 20 40 60 80 100 120 140 Dose [%] D 10 = (112.9 ± 1.6)% D 0.1cm³ = (114.0 ± 1.3)% D 1.0cm³ = (98.4 ± 13.2)% V 100 = (63.9 ± 9.2)% Urethra: Volume [%] 100 90 80 70 60 50 40 30 20 10 0 0 50 100 150 200 250 Dose [%] D 10 115% V 115 10% D 0.1cm³ 120 % Slide 73

Reproducibility of Treatment Delivery III. Rectum (33.8 ± 9.6) cm³ Volume [%] 100 90 80 70 60 50 40 30 20 10 0 0 50 100 150 200 250 Dose [%] Volume [%] 100 90 80 N=117 70 60 50 40 30 20 10 0 0 10 20 30 40 50 60 70 80 90 100 Dose [%] D 10 = (52.5 ± 4.8)% D 0.1cm³ = (74.9 ± 2.7)% D 2.0cm³ = (58.1 ± 5.3)% Rectum D 10 75% V 75 10% D 0.1cm³ 80 % Slide 74

Reproducibility of Treatment Delivery IV. Bladder (28.6 ± 13.4) cm³ Volume [%] 100 90 80 70 60 50 40 30 20 10 0 0 50 100 150 200 250 Dose [%] Volume [%] 100 90 80 N=117 70 60 50 40 30 20 10 0 0 10 20 30 40 50 60 70 80 90 100 Dose [%] D 10 = (50.5 ± 8.2)% D 0.1cm³ = (73.8 ± 4.9)% D 2.0cm³ = (53.6 ± 5.7)% Bladder D 10 75% V 75 10% D 0.1cm³ 80 % Slide 75

Conclusions Can we deliver the dose distribution we plan? Currently we do not have a close-loop System (Technology) clinically routinely available to answer this for the different localisations (similar to kv-, MV-CBCT in ERT, Industry) We do have data supporting and proving the YES-Answer under specific conditions and scenarios (in the enormous diversity of methods & clinical workflows) We definitively need site-specific guidelines (procedures and action levels) for Treatment Delivery Verification in modern Brachytherapy! Slide 76

Thank you very much for your Attention! Slide 77