Pankti M. Shah et al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; 07-16

Similar documents
J Pharm Sci Bioscientific Res (4): ISSN NO

The present manuscript describes simple, sensitive, rapid, accurate, precise and cost effective First derivative

World Journal of Pharmaceutical Research

Available online Research Article

ISSN (Print)

Journal of Global Trends in Pharmaceutical Sciences

Dr. Dilip G. Maheshwariet al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; Research Article

DETERMINATION OF LOTEPREDNOL ETABONATE AND TOBRAMYCIN IN COMBINED DOSAGE FORM USING RP-HPLC METHOD

Journal of Global Trends in Pharmaceutical Sciences

Journal of Chemical and Pharmaceutical Research

Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin in Pharmaceutical Dosage Form using RP-HPLC Method

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

Int. J. Pharm. Sci. Rev. Res., 31(1), March April 2015; Article No. 46, Pages:

Scholars Research Library

Pelagia Research Library

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET

Development and Validation of RP-HPLC Method for the Estimation of Gemigliptin

Pelagia Research Library

for Methotrexate. The result of analysis has been validated statistically and recovery studies confirmed the accuracy of

JPSBR: Volume 4, Issue 6: 2014 ( ) ISSN NO

MEDAK DIST. ANDHRA PRADESH STATE, INDIA. Research Article RECEIVED ON ACCEPTED ON

pharmaceutical formulations. Anagliptin shows absorption maximum at 246 nm and obeys beer s law in the

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms

International Journal of Pharma and Bio Sciences

Tentu Nageswara Rao et al. / Int. Res J Pharm. App Sci., 2012; 2(4): 35-40

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD

Pelagia Research Library

Stability indicating RP-HPLC method development and validation of Etizolam and Propranolol hydrochloride in pharmaceutical dosage form

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

Absorption Correction Method for Simultaneous Estimation of Phenazopyridine HCl and Ciprofloxacin HCl in Combined Tablet Dosage Form

Development and validation of RP-HPLC method for simultaneous estimation of gliclazide and metformin in pure and tablet dosage form

Vol-3, Issue-4, Suppl-2, Nov 2012 ISSN: Sheth et al PHARMA SCIENCE MONITOR

Hyderabad, India. Department of Pharmaceutical Chemistry, Glocal University, Saharanpur, India.

Corresponding Author:

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

36 J App Pharm Vol. 6; Issue 1: 36-42; January, 2014 Rao et al., 2014

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ASSAY AND DISSOLUTION OF METOPROLOL SUCCINATE EXTENDED RELEASE TABLETS

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (

AMERICAN JOURNAL OF BIOLOGICAL AND PHARMACEUTICAL RESEARCH

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

Available online at Scholars Research Library

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD ESTIMATION OF TOLVAPTAN IN BULK PHARMACEUTICAL FORMULATION

Development and Validation of Stability Indicating HPTLC Method for Estimation of Seratrodast

Impact factor: 3.958/ICV: 4.10 ISSN:

Development and Validation of Area Under Curve Method for Simultaneous Estimation of Thiocolchicoside and Lornoxicam in Tablet Dosage Form

Volume 2 (6), 2014, Page CODEN (USA)-IJPRUR, e-issn: International Journal of Pharma Research and Health Sciences

International Journal of Drug Research and Technology

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

IJRPC 2011, 1(4) Rohan et al. ISSN: INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND CHEMISTRY

New RP-HPLC Method for the Determination of Fludarabine in Pharmaceutical Dosage Forms

DEVELOPMENT AND STABILITY INDICATING HPLC METHOD FOR DAPAGLIFLOZIN IN API AND PHARMACEUTICAL DOSAGE FORM

Simultaneous estimation of Metformin HCl and Sitagliptin in drug substance and drug products by RP-HPLC method

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR DETERMINATION OF ROSUVASTATIN CALCIUM IN BULK AND PHARMACEUTICAL DOSAGE FORM

Journal of Chemical and Pharmaceutical Research

Research Article Simultaneous Estimation of DL-Methionine and Pyridoxine Hydrochloride in Tablet Dosage Form by RP-HPLC

UV SPECTROPHOTOMETRIC METHOD DEVELOPMENT AND VALIDATION FOR THE DETERMINATION OF ATENOLOL AND LOSARTAN POTASSIUM BY Q-ANALYSIS

Development and validation of UV spectrophotometric estimation of lisinopril dihydrate in bulk and tablet dosage form using area under curve method

Development and validation of UV and RP-HPLC method for estimation of Nepafenac in bulk drug and opthalmic formulation

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF ETORICOXIB AND THIOCOLCHICOSIDE IN PHARMACEUTICAL DOSAGE FORMS

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method

CHAPTER 3: ANALYTICAL METHOD DEVELOPMENT AND VALIDATION

Development of a Validated RP-HPLC Method for the Analysis of Citicoline Sodium in Pharmaceutical Dosage Form using Internal Standard Method

Estimation of zolmitriptan by a new RP-HPLC method

Simple and stability indicating RP-HPLC assay method development and validation lisinopril dihydrate by RP-HPLC in bulk and dosage form

Simultaneous Estimation of Gemcitabine Hydrochloride and Capecitabine Hydrochloride in Combined Tablet Dosage Form by RP-HPLC Method

RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR CILOSTAZOL IN TABLET DOSAGE FORM

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

S. G. Talele, D. V. Derle. Department of Pharmaceutics, N.D.M.V.P. College of Pharmacy, Nashik, Maharashtra, India


Research Article. ISSN Available online at 746

A stability indicating RP-HPLC method for simultaneous estimation of darunavir and cobicistat in bulk and tablet dosage form

Method Development and Validation for the Estimation of Saroglitazar in Bulk and Pharmaceutical Dosage Form by RP-HPLC

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

Method Validation for Simultaneous Estimation of Prednisolone and Abiraterone Acetate by RP-HPLC

Analytical Method Development and Validation for the Estimation of Guaifenesin and Dextromethorphan by RP-HPLC

NOVEL RP-HPLC METHOD. B.Lakshmi et a. concentration range KEY INTRODUCTION. Diltiazem is used to. . It works by of contractionn of the. (dilate).

Research and Reviews: Journal of Pharmacy and Pharmaceutical Sciences

438 Full Text Available On

Scholars Research Library

International Journal of Pharmacy and Pharmaceutical Sciences. Research Article

Available online Research Article

Development of RP-HPLC Method for Estimation of Valsartan and Hydrochlorothiazide in Tablets

Journal of Chemical and Pharmaceutical Research, 2018, 10(2): Research Article

Scholars Research Library. Der Pharmacia Lettre, 2015, 7 (5):44-49 (

Simultaneous Estimation of Lamivudine and Stavudine by using RP-HPLC and Method Development as per ICH Guidelines

A HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC ASSAY FOR LERCANIDIPINE HYDROCHLORIDE

Airo International Research Journal ISSN : Volume : 7 October 2015

Analytical method validation for tablet of phenoxymethyl penicillin potassium by RP-HPLC method

Development and Validation of Improved RP-HPLC method for Identification and Estimation of Ellagic and Gallic acid in Triphala churna

Analytical Method Development and Validation for the Estimation of Imatinib Mesylate and its Impurity in Pharmaceutical Formulation by RP-HPLC

Transcription:

Asian Journal of Pharmaceutical Technology & Innovation ISSN: 2347-8810 Research Article Received on: 30-03-2016 Accepted on: 01-04-2016 Published on: 15-04-2016 Corresponding Author: *Pankti M. Shah, Department of Quality Assurance & Pharma Regulatory Affairs, L. J. Institute of Pharmacy, Ahmedabad, Gujarat, India Development and Validation RP-HPLC Method for Simultaneous Estimation of Valsartan and Nifedipine in Synthetic Mixture ABSTRACT Pankti M. Shah*, Jignesh S. Shah 1, Dilip G. Maheshwari 2 A simple, specific and accurate Reverse Phase High Performance Liquid Chromatography Method was developed for the simultaneous determination of Valsartan and Nifedipine in Synthetic Mixture. The using Phenomenex Luna C 18 (250 mm x 4.6 mm, 5 μm) column in Isocratic mode, with Mobile Phase containing Methanol: Water: ACN (ph 3.7 adjusted with 10% Ortho Phosphoric Acid) (70:25:05 %v/v/v). The Flow Rate was 1 ml/min and effluents were monitored at 233 nm. The Retention Time of Valsartan and Nifedipine were found to be 8.003 min and 5.290 min respectively. The Linearity for Valsartan and Nifedipine were found to be 4-20 µg/ml and 1.5-7.5 µg/ml respectively. The Recoveries of Valsartan and Nifedipine were found to be 99.50% 100.15% and 99.33% 100.40% respectively. The proposed method was validated and successfully applied for the estimation of Valsartan and Nifedipine in Synthetic Mixture. e-mail: pshah3089@gmail.com Key-words: Valsartan, Nifedipine, RP-HPLC. Cite this article as: Pankti M. Shah, Jignesh S. Shah, Dilip G. Maheshwari, Development and Validation RP-HPLC Method for Simultaneous Estimation of Valsartan and Nifedipine in Synthetic Mixture,Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016, 07-16. 1 Department of Quality Assurance & Pharma Regulatory Affairs, L. J. Institute of Pharmacy, Ahmedabad. 2 Head of Department of Quality Assurance and Pharm Regulatory Affairs, L. J. Institute of Pharmacy, Ahmedabad. 7

INTRODUCTION: [1] Valsartan is an Angiotensin II Receptor Antagonist (commonly called an ARB, or angiotensin receptor blocker), that is selective for the type I (AT1) angiotensin receptor. It is used to treat high blood pressure, congestive heart failure, and to reduce death for people with left ventricular dysfunction after having had a heart attack. Nifedipine is a dihydropyridine calcium channel blocker that primarily blocks L-type calcium channels. Its main uses are as an antianginal and antihypertensive, although a large number of other indications have recently been found for this agent, such as Raynaud's phenomenon, premature labour, and painful spasms of the esophagus such as in cancer and tetanus patients. Figure 1: Valsartan INSTRUMENTAL: HPLC (Shimadzu) Model: SPD-20A, LC-20AD Column:Phenomenex Luna C 18 (250 mm x 4.6 mm, 5 μm) Detector: U.V Detector Software: Spinchrome Hamilton Syringe Analytical Weighing Balance (Wensar DAB-220) Sonicator (Equitron) Digital ph Meter (Systronic) High Vacuum Pump (Parag engineering) Figure 2: Nifedipine CHEMICALS AND MATERIALS: Acetonitrile Avantor Performance Material India Ltd. (HPLC grade) Methanol Finar Ahmedabad. (HPLC grade) Water Astron Chemical India. (HPLC grade) OPA (75% Ortho Phosphoric Acid) - AR Grade, Astron Chemical India. Valsartan and Nifedipine were supplied by Torrent Pharmaceuticals, Ahmedabad, India and Mediwin Pharmaceutical, Ahmedabad, India respectively. SELECTION OF DETECTION WAVELENGTH: The sensitivity of HPLC method that uses UV detection depends upon proper selection of detection wavelength. At 233 nm both the drug give good peak height and shape. So, 233 nm was selected for detection of Valsartan and Nifedipine in RP-HPLC system. (Figure 3) 8

MOBILE PHASE SELECTION: The composition and flow rate of mobile phase were changed to optimize the separation condition using combined solution. The pka value for Valsartan and Nifedipine is 4.37 and 5.33 respectively. After number of trial experiments, it was established that the mobile phase Methanol: Water: ACN (ph 3.7 adjusted with 10% Ortho Phosphoric Acid) (70:25:05 %v/v/v) shows good peak shape and resolution. CHROMATOGRAPHIC CONDITION: Column: Phenomenex LunaC 18 (240 mm 4.6 mm, 5 μm) Mobile Phase: Methanol: Water: ACN (ph 3.7 adjusted with 10% Ortho Phosphoric Acid) (70:25:05 %v/v/v) Flow Rate: 1 ml/min Detection Wavelength: 233 nm Run Time: 15 min Detector: U.V Detector Injection Volume: 20 μl PREPARATION OF STANDARD STOCK SOLUTION: Valsartan (100 μg/ml): Accurately weighed VAL (10 mg) was transferred to a 100 ml volumetric flask, and diluted to the mark with Methanol to obtain a standard stock solution (100 μg/ml). Nifedipine (100 μg/ml): Accurately weighed NIFE (10 mg) was transferred to a 100 ml volumetric flask, and diluted to the mark with Methanol to obtain a standard stock solution (100 μg/ml). PREPARATION AND ANALYSIS OF SYNTHETIC MIXTURE [2] : The Synthetic Mixture of Valsartan and Nifedipine was prepared in ratio of 80:30. Common excipients, Lactose (95 mg), MCC [Micro Crystalline Cellulose] (129 mg), Magnesium Stearate (9 mg), Talc (7 mg) along with the drug Valsartan 80 mg and Nifedipine 30 mg. Accurately weighed equivalently weight of Valsartan 80 mg which contain Valsartan (80 mg) and Nifedipine (30 mg) which transferred in 100 ml volumetric flask and make up half mark with Methanol. This solution was sonicated till the drug dissolves and was made upto mark with Methanol. Then this solution was filtered through Whatmann filter paper. So, obtained concentration of Valsartan is 800 μg/ml and Nifedipine is 300 μg/ml. PREPARATION OF WORKING SAMPLE SOLUTION: From above synthetic mixture solutions take 0.1 ml and transferred in to a 10 ml volumetric flask and the volume was adjusted up to the mark with Mobile Phase to make final concentration of VALSARTAN 8 μg/ml and NIFEDIPINE 3 μg/ml. (Table 6) METHOD VALIDATION [3] : The developed method was validated with respect to specificity, linearity, range, accuracy, and precision, limit of detection and limit of quantification in accordance with the ICH guideline. Specificity: Specificity is the ability to assess unequivocally the analyte in the presence of components which may be expected to be present. Typically these might include impurities, degradants, matrix, etc. 9

Linearity & Range: The linearity of VALSARTAN and NIFEDIPINE was found to be in the range of 4-20 μg/ml and 1.5-7.5 μg/ml respectively. Plot the calibration curve of Area (μv.s) vs. Concentration (μg/ml). Linearity of both the drugs was checked in term of Slope, Intercept and Correlation Coefficient. (Table 1) Precision: The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions. Precision may be considered at three levels: Intermediate (Intraday) Precision, Reproducibility (Interday) Precision, and Repeatability.(Table 2, 3) 1) Intraday Precision: Solutions containing 8, 12, 16 μg/ml of VAL and 3, 4.5, 6 μg/ml of NIFE were analyzed three times on the same day and %R.S.D was calculated. 2) Interday Precision: Solutions containing 8, 12, 16 μg/ml of VAL and 3, 4.5, 6 μg/ml of NIFE were analyzed on three different successive days and %R.S.D was calculated. 3) Repeatability: Solutions containing 12 μg/ml of VAL and 4.5 μg/ml of NIFE were analyzed for six times and %R.S.D. was calculated. Accuracy: The accuracy of an analytical procedure expresses the closeness of agreement between the value which is accepted either as a conventional true value or an accepted reference value and the value found. Recovery studies were carried out by addition of standard drug to the sample at three different concentration levels 50%, 100%, 150%. This performance was done in triplicate. The amount of VALSARTAN and NIFEDIPINE were calculated at each level and % recoveries were calculated by measuring the Peak Area and fitting the values in equation. (Table 4) Limit of Detection (LOD): Limit of detection can be calculated using following equation as per ICH guidelines. (Table 5) LOD = 3.3 x (σ/ S) Where, σ = standard deviation of the Y intercept of calibration curve S = Mean slope of the corresponding calibration curve. Limit of Quantification (LOQ): Limit of quantification can be calculated using following equation as per ICH guidelines. (Table 5) LOQ = 10 x (σ/ S) Where, σ = standard deviation of the Y intercept of calibration curve S = Mean slope of the corresponding calibration curve. Robustness: The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage. (Table 7) It should show the reliability of an analysis with respect to deliberate variation in method parameter. In case of liquid chromatography, examples of typical variations are: 10

- Influence of variations of ph in mobile phase; - Influence of variations in mobile phase composition; - Different columns (different lots and/or suppliers) - Temperature - Flow rate - System Suitability Tests: A system suitability test is an integral part of liquid chromatography. They are used to verify that resolution and reproducibility of chromatography system are adequate for the analysis to be done. The Test includes the Resolution, Column Efficiency, Tailing Factor and Theoretical Plates. Valsartan 233 nm Nifedipine Figure 3: Selection of Wavelength (Overlain Spectra of Valsartan (10 μg/ml) and Nifedipine (8 μg/ml)) Figure 4: Chromatogram of VAL (8 μg/ml) and NIFE (3 μg/ml) in Methanol: Water: ACN (ph 3.7) (70:25:05 %v/v/v) 11

Figure 5: Overlay Chromatogram of Nifedipine (1.5-7.5 µg/ml) and Valsartan (4-20 µg/ml) 800000 700000 600000 500000 400000 300000 200000 100000 0 VALSARTAN y = 29994x + 68009 R² = 0.9995 0 5 10 15 20 25 Figure 6: Calibration Curve of VALSARTAN (4-20 µg/ml) 600000 500000 400000 300000 200000 100000 0 NIFEDIPINE y = 76562x - 18115 R² = 0.9984 0 1 2 3 4 5 6 7 8 Figure 7: Calibration Curve of NIFEDIPINE (1.5-7.5 µg/ml) 12

Table 1: Linearity Data for VAL (4-20 µg/ml) and NIFE (1.5-7.5 µg/ml) VALSARTAN NIFEDIPINE Conc. Mean Peak Area Conc. Mean Peak Area (µv.s) % RSD (µg/ml) (µv.s) ± S.D. (n=6) (µg/ml) ± S.D. (n=6) % RSD 4 192308 ± 1294.68 0.6732 1.5 102180 ± 691.87 0.6771 8 305672 ± 1588.84 0.5197 3 206646 ± 1031.32 0.4990 12 424056 ± 1679.95 0.3961 4.5 318262 ± 1189.62 0.3737 16 545254 ± 1860.66 0.3412 6 450535 ± 1388.88 0.3082 20 672398 ± 1930.99 0.2871 7.5 554453 ± 1636.37 0.2951 Table 2: Precision Study for VALSARTAN VALSARTAN INTRADAY PRECISION Conc. (µg/ml) Mean Peak Area (µv.s) ± S. D. (n=3) % R.S.D 8 308157 ± 1551.18 0.5033 12 424210 ± 1571.61 0.3704 16 546753 ± 1959.70 0.3584 INTERDAY PRECISION Conc. (µg/ml) Mean Peak Area (µv.s) ± S. D. (n=3) % R.S.D 8 307065 ± 1847.81 0.6017 12 427302 ± 2180.14 0.5102 16 548962 ± 2313.50 0.4214 REPEATABILITY Conc. (µg/ml) Mean Peak Area (µv.s) ± S. D. (n=6) % R.S.D 12 425756 ± 2399.31 0.5635 Table 3: Precision Study for NIFEDIPINE NIFEDIPINE INTRADAY PRECISION Conc. (µg/ml) Mean Peak Area (µv.s) ± S. D. (n=3) % R.S.D 3 208491 ± 1001.66 0.4804 4.5 318766 ± 1213.05 0.3805 6 452956 ± 1351.31 0.2983 INTERDAY PRECISION Conc. (µg/ml) Mean Peak Area (µv.s) ± S. D. (n=3) % R.S.D 3 210878 ± 1158.16 0.5492 4.5 321398 ± 1543.43 0.4802 6 455000 ± 1992.11 0.4378 REPEATABILITY Conc. (µg/ml) Mean Peak Area (µv.s) ± S. D. (n=6) % R.S.D 4.5 319475 ± 1478.64 0.4628 13

Table 4: Recovery Study Data Name of Drug Valsartan Nifedipine % Level of Recovery Test Amt. (µg/ml) Amt. of Drug Spiked (µg/ml) Total Std Amt. (µg/ml) Total Amount Recovered (µg/ml) % Recovery ± R.S.D. (n=3) 50 8 4 12 11.94 99.50 ± 0.3615 100 8 8 16 15.98 99.87 ± 0.3587 150 8 12 20 20.03 100.15 ± 0.2911 50 3 1.5 4.5 4.47 99.33 ± 0.4431 100 3 3 6 5.98 99.67 ± 0.3571 150 3 4.5 7.5 7.53 100.40 ± 0.3347 Table 5: LOD and LOQ Data Drug Name VALSARTAN NIFEDIPINE Standard Deviation 1294 723 Slope 39992 11483 LOD (μg/ml) 0.1067 0.2078 LOQ (μg/ml) 0.3236 0.6298 Drug Name Table 6: Application of HPLC Method to Synthetic Mixture Amount in Synthetic Mixture (μg/ml) Amount Found (μg/ml) ± S.D. (n=3) % Assay ± R.S.D. (n=3) VALSARTAN 8 8.03 ± 1688.82 100.37 ± 0.3848 NIFEDIPINE 3 2.98 ± 1404.87 99.33 ± 0.3091 Table 7: Robustness Data of VALSARTAN and NIFEDIPINE Condition Flow Rate (1 ml ± 0.1 ml/ min) Detection Wavelength (257 nm ± 1 nm) Change in Mobile Phase Composition (%v/v/v) Variation VALSARTAN NIFEDIPINE %Assay ± R.S.D (n=3) %Assay ± R.S.D (n=3) 0.9 ml/min 99.50 ± 0.5461 98.50 ± 0.3744 1.0 ml/min 100.50 ± 0.5645 100.50 ± 0.4659 1.1 ml/min 100.83 ± 0.6026 99.00 ± 0.4884 232 nm 99.67 ± 0.4705 99.67 ± 0.3877 233 nm 100.50 ± 0.5645 100.50 ± 0.4659 234 nm 101.00 ± 0.5910 99.50 ± 0.5735 69: 26: 05 99.33 ± 0.4301 98.00 ± 0.4093 70: 25: 05 100.50 ± 0.5645 100.50 ± 0.4659 70: 24: 06 100.66 ± 0.6495 98.50 ± 0.5310 14

Sr. No. Table 8: Summary of Validation Parameters Parameters VALSARTAN NIFEDIPINE 1 Linearity Range (μg/ml) 4-20 1.5-7.5 2 Regression Line Equation (y = mx + c) y = 39992x+68009 y = 11484x-18115 3 Correlation Coefficient (R²) 0.999 0.998 4 Intraday Precision (%RSD, n=3) 0.3584 0.5033 0.2983 0.4804 5 Interday Precision (% RSD, n=3) 0.4214 0.6017 0.4378 0.5492 6 Repeatability (% RSD, n=6) 0.5635 0.4628 7 LOD (μg/ml) 0.1067 0.2078 8 LOQ (μg/ml) 0.3236 0.6298 9 % Recovery Study (n=3) 99.50 100.15 99.33 100.40 RESULTS: A Reverse Phase Column proposed as a suitable method for the determination of Valsartan and Nifedipine in Synthetic Mixture. The Chromatographic condition were optimized by changing the Mobile Phase Composition. Different ratios were experimented to optimize the Mobile Phase. Finally, : Methanol: Water: ACN (ph 3.7 adjusted with 10% Ortho Phosphoric Acid) (70:25:05 %v/v/v) as Mobile Phase which shows good resolution of Valsartan and Nifedipine peak. The wavelength of detection selected was 233 nm, as the drug shows optimized absorbance at this wavelength. By our proposed method Retention Time of Valsartan and Nifedipine were about 8.003 min and 5.290 min respectively and none of the Impurities were interfering in its assay. DISCUSSIONS: The statistical analysis of data and the drug recovery data showed that the method was simple, rapid, economical, sensitive, precise and accurate. It can thereby easily adopt for routine quality control analysis. The results of this analysis confirmed that the proposed method was suitable for determination of drug in Synthetic Mixture with virtually no interference of additives. Hence the proposed method can be successfully applied in estimation of Valsartan and Nifedipine in Synthetic Mixture. CONCLUSION: A simple, rapid, sensitive, accurate and precise RP-HPLC Method has been developed and validated for routine analysis of VALSARTAN and NIFEDIPINE in Synthetic Mixture. The RP-HPLC method is suitable for simultaneous estimation of VALSARTAN and NIFEDIPINE in Synthetic Mixture without interference of each other. The developed method was successfully applied in Synthetic Mixture. The proposed Method can be utilized for the routine analysis of VALSARTAN and NIFEDIPINE in Synthetic Mixture. ACKNOWLEDGEMENT: We are heartly thankful to Dr. K. Pundarikakshudu, Director of L.J Institute of Pharmacy, Ahmedabad for providing all the facilities and the valuable Guidance during the Research work. 15

REFERENCES: 1. Indian Pharmacopeia 2010, Ghaziabad: Govt. of India Ministry of Health and Family Welfare, The Controller of Publication Indian Pharmacopoeia Commission, 2010vol- 1, 3, pp 382, 456, 1779-1780, 2287. 2. JB Park, JW Ha, HO Jung and MY Rhee, Randomized Trial comparing the effects of a low-dose combination of Nifedipine GITS and Valsartan versus High dose Monotherapy on Central Hemodynamic in patient with inadequately controlled Hypertension: FOCUS Study NCBI.2014, 19(5), 294-301. 3. ICH, Q2 (R1), 2005, Validation of Analytical Procedure: Text and Methodology, International Conference on Harmonization, IFPMA, Geneva, Switzerland. 16