Assessment of penetration of dorsal screws after fixation of the distal radius using ultrasound: cadaveric study

Similar documents
The Utility of the Fluoroscopic Skyline View During Volar Locking Plate Fixation of Distal Radius Fractures

Radiographic Evaluation of Dorsal Screw Penetration After Volar Fixed-Angle Plating of the Distal Radius: A Cadaveric Study

Failed Extensor Indicis Proprius Tendon Transfer for Extensor Pollicis Longus Tendon Rupture after Distal Radial Fracture

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

WINSTA-R. Distal Radius System

Distal Radius and Distal Ulna Plates System Self-Tapping Spherical Locking Screw Self-Tapping Conical Locking Screw Cortex Screw

Disclosures. Distal Radius Fractures 5/16/2017. Distal Radius Fractures: Complications & Limitations of the Volar Approach

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Complications of Distal Radius Fractures. How to Treat a Distal Radius Fx 11/13/2017. Michael S. Bednar, M.D. Loyola University Chicago

SURGERY OF THE HAND INTRODUCTION CASE REPORT. Hee-June Kim 1, Hyun-Joo Lee 1, Dong-Hyun Kim 1, Joon-Woo Kim 1, Ji Won Oh 2

Integra. Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

Disclosure. 85% Dorsal. 15% Volar. Distal Radius Fractures- Volar Plating for All. Skeletal Dynamics- Consultant

De Maeseneer et al. Sonography of the Extensor Pollicis Longus Tendon

6/5/2018. DISCLOSURES Hassan R. Mir, MD, MBA, FACS. Evolution of Distal Radius Fracture Treatment [Chung Hand Clinics 2012]

Instrument and Implant for wrist fracture

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y.

Acu-Loc Wrist Plating System. Surgical Technique

Distal Radius Plate 2.4/2.7 dorsal and volar

Treatment of unstable distal radius fractures with the volar locking plate

VariAx TM Distal Radius Locking Plate System

Emile N. Brown, MD, and Scott D. Lifchez, MD

From the Department of Orthopedics, Tongji Hospital, Tongji University, Shanghai, China

Locking Radial Head Plates

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures.

FUNCTIONAL AND RADIOLOGICAL OUTCOMES AFTER VOLAR LOCKING PLATE FIXATION IN TYPE C DISTAL RADIUS FRACTURES A RETROSPECTIVE STUDY

2.4 mm LCP Distal Radius System

AcUMEDr. FoREARM ROD SYSTEM

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

Small External Fixator Wrist Spanning Frame. For the treatment of wrist fractures.

Acu-Loc 2 Wrist Plating System. Surgical Technique

Acu-Loc 2 Wrist Plating System

Surgical Technique. Wrist Plating System

Volar Distal Radius Plating System Surgical Technique

VariAx Distal Radius Locking Plate System

Surgical Technique. Forearm Fracture Solutions

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬

Volar Distal Radius Plating System Surgical Technique

Surgical Technique Guide

The Wrist Fusion Set. Stainless Steel and Titanium TECHNIQUE GUIDE. Instruments and implants approved by the AO Foundation

epak Single-Use Delivery System

Reproducibility of Three Dimensional Digital Preoperative Planning for the Osteosynthesis of Distal Radius Fractures

Wrist Fusion Instrument and Implant Set.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES

Aesculap Orthopaedics Targon DR

QUICK REFERENCE GUIDE. The Pennig Dynamic Wrist Fixator. Part A: Trans-articular application

Index. Springer International Publishing Switzerland 2016 J.N. Lawton (ed.), Distal Radius Fractures, DOI /

CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Comparison of volar locking plate and external fixation with K wire augmentation in unstable distal radius fractures

Introduction to Ultrasound Examination of the Hand and upper

Evaluation of volar locking plate fixation for management of intraarticular fractures of distal end of radius

Angle Stable Distal Radial Plate System WINSTA-R

Acute Rupture of Flexor Tendons as a Complication of Distal Radius Fracture

Wrist Fixation System

Mark VanDer Kaag 1, Ajmal Ikram 2. Hand Unit, Tygerberg Hospital University of Stellenbosch

)372( COPYRIGHT 2018 BY THE ARCHIVES OF BONE AND JOINT SURGERY

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Long-term Results of Dorsally Displaced Distal Radius Fractures Treated With the Pi-Plate: Is Hardware Removal Necessary?

A N D R E W I R W I N, F R C S E D ( O R T H ) C O N S U L T A N T O R T H O P A E D I C S U R G E O N W E S T H E R T S N H S T R U S T, U K

Int J Clin Exp Med 2017;10(1): /ISSN: /IJCEM

Fractures of the distal radius have occupied a storied history

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Disclosures. Epidemiology 10/12/2015. Distal Radius Fractures: Staying Out of Trouble

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

The Flower Forefoot PROCEDURE GUIDE.

The Flower Medial Column Fusion Plate

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Distal radius fractures raises considerable interest and

The Flower Four Corner Fusion Plate

Surgical Technique Carpal Fusion

Acomparison of percutaneous and pin-and-plaster techniques in distal radius fracture

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

DVR Anatomic Plate. Surgical Technique

Ultrasound-guided reduction of distal radius fractures

Forearm Fracture Solutions. Product Overview

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

ANATOMIC LOCKED PLATING SYSTEM

Factors Predicting Late Collapse of Distal Radius Fractures

Ultrasonography of the wrist - a step-by-step approach to study protocols and normal findings

VariAx. Hand & Wrist. Distal Radius Locking Plate System

Visualize, stabilize, mobilize. Wristore * Distal Radius Fracture Fixator Abbreviated Surgical Technique

Breakage of a volar locking plate after delayed union of a distal radius fracture

Bridging external fixation and supplementary Kirschner-wire fixation versus volar locked plating for unstable fractures of the distal radius

Distal Ulnar Locking Plate

of intramedullary nailing for correction of posttraumatic

Acu-Loc Wrist Spanning Plate System. Surgical Technique

Fractures of the distal end of the radius should be

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

Hand and wrist emergencies

Kousuke Iba 1*, Yasuhiro Ozasa 1, Takuro Wada 1, Tomoaki Kamiya 1, Toshihiko Yamashita 1, Mitsuhiro Aoki 2. Abstract

Abd Ali Muhsin FICMS.

The effect of an associated ulnar styloid fracture on the outcome after fixation of a fracture of the distal radius

ORIGINAL ARTICLE TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY

Surgical Technique. Olecranon Locking Plate

SCAHPO-LUNATE DISSOCIATION

Transcription:

UPPER LIMB Ann R Coll Surg Engl 2016; 98: 138 142 doi 10.1308/rcsann.2016.0045 Assessment of penetration of dorsal screws after fixation of the distal radius using ultrasound: cadaveric study D Williams, J Singh, N Heidari, M Ahmad, A Noorani, L Di Mascio The Royal London Hospital, Barts Health NHS Trust & Queen Mary University of London, UK ABSTRACT INTRODUCTION Volar locking plates are used to treat unstable and displaced fractures of the distal radius. Potential advantages of stable anatomical reduction (eg early mobilisation) can be limited by penetration of dorsal screws, leading to synovitis and potential rupture of extensor tendons. Despite intraoperative imaging, penetration of dorsal screws continues to be a problem in volar plating of the distal radius. Ultrasound is a well recognised, readily available, diagnostic tool used to assess soft-tissue impingement by orthopaedic hardware. In this cadaveric study, we wished to ascertain the sensitivity and specificity of ultrasound for identification of protrusion of dorsal screws after volar plating of the distal radius. METHODS Four adult, unpaired phenol-embalmed cadaveric distal radii were used. A VariAx Distal Radius Volar Locking Plate system (Stryker, Kalamazoo, MI, USA) was employed for instrumented fixation. A portable SIUI CTS 900 ultrasound machine (Providian Medical, Eastlake, OH, USA) was used to image the dorsal cortex to ascertain screw penetration. RESULTS Specificity and sensitivity of ultrasound for detection of screw protrusion through the dorsal cortex was 100%. CONCLUSIONS Ultrasound was found to be a safe and accurate method for assessment of dorsal-screw penetration through the dorsal cortex of the radius after volar plating of the distal radius. It also aids diagnosis of associated tendon disorders (eg tenosynovitis) that might cause pain and limit wrist function. KEYWORDS Ultrasound Radius Dorsal screws Volar plates Accepted 25 June 2015 CORRESPONDENCE TO Jagwant Singh, E: drjagwant@gmail.com Percutaneous pinning and plastering are established treatments for dorsally displaced fractures of the distal radius. 1 3 Fixation using volar plates is first-line treatment for unstable intra-articular or extra-articular distal fractures of the radius. 4 10 Penetration of the dorsal cortex leading to irritation or attrition of extensor tendons is a complication of fixation using volar plates. 12 21 Screws can compromise the volume of the dorsal compartments. Chronic conflict between tendons and screws can result in local hyperaemia, tenosynovitis, and persistent impingement, eventually leading to tendon rupture. 21 23 Prevalence of tendon-related problems (irritation and tears) is 2 14%. 12,13,15 In a meta-analysis, Margaliot et al 22 reported the prevalence of tendon rupture to be 2.6% and that of tenosynovitis to be 5.2%. These values can lead to poor outcomes despite satisfactory reduction and fixation of fractures, as well as a delay to return to work in young, active patients. 16 Conflict between orthopaedic hardware and soft tissue is difficult to visualise on standard radiographs owing to: overlying bone; 21,24 static nature of images; insufficient contrast for soft tissues; absence of real-time correlation between pain and the structures being observed. Ultrasound eliminates all of these problems and also offers excellent spatial resolution. 21 We use ultrasound in the outpatient clinic to identify screw penetration of the dorsal cortex after volar plating of fractures of the distal radius. This simple test can be done with minimal training, and could be used to reduce the need for computed tomography (CT) and, as such, radiation exposure and cost. With the ready availability of sterile ultrasound units in theatres, this test could be done intraoperatively for difficult cases. In this cadaveric study, we wished to ascertain the sensitivity and specificity of ultrasound for identification of protrusion of dorsal screws after volar plating of the distal radius. 138 Ann R Coll Surg Engl 2016; 98: 138 142

Methods Four adult, unpaired phenol-embalmed cadaveric distal radii were used. A VariAx Distal Radius Volar Locking Plate system (Stryker, Kalamazoo, MI, USA) was employed for instrumented fixation. Plates are contoured anatomically and have four options for variable-angle distal locking screws that allow for 15 of freedom. A portable SIUI CTS 900 ultrasound machine (Providian Medical, Eastlake, OH, USA) was used to image the dorsal cortex for screw penetration. The volar aspect of four cadaveric distal radii was exposed using a modified version of Henry's approach. The pronator quadratus was exposed through the bed of the flexor carpi radialis tendon. Then, the pronator quadratus was reflected from its radial insertion. Plates were applied to the volar cortex of the distal radius (just proximal to the watershed line) and fixed with a single cortical non-locked screw in the distal diaphysis. Distal screw holes were numbered from one to four starting from the ulna side (Fig 1). Bi-cortical holes were drilled sequentially through the locking holes in the distal row of the plate without fluoroscopic guidance. The optimum length of each screw was measured using a standard depth gauge from Stryker VariAx instrumentation. Then, locking screws (2.3mm) were inserted at different lengths into all four distal locking holes. Screws were inserted at 2mm to the measured depth in cadaver specimen number (CSN)1, at 0mm in CSN2, at +2mm in CSN3, and at +4mm in CSN4. Using the SIUI CTS 900 ultrasound machine, the dorsal surface of the wrist was imaged to detect the site (extensor compartment) and length of screw penetration. Figure 2 shows a typical appearance of a screw penetrating the dorsal cortex. Ultrasound findings for each specimen were recorded. Figure 2 Ultrasound image showing screw penetration just radial to Lister s tubercle as a hyperechoic structure Cadaveric specimens were dissected in the dorsal plane to identify and quantify screw penetration. Penetration of the extensor compartment and screw length were recorded. Results In CSN1, the depth was measured and screws were inserted 2mm short of the measured depth. No screws were found to be prominent of the dorsal cortex upon ultrasound or dissection. Upon dissection, the most ulnar-directed screw was found to have penetrated the distal radioulnar joint (DRUJ) but this phenomenon was not detected on ultrasound. In CSN2, the screw lengths selected were the same as the measured depth. Ultrasound detected one screw in the second extensor compartment that was prominent by 1mm, and this phenomenon was confirmed upon dissection. Dissection Figure 1 The volar cortex of the distal radius was divided into four zones from medial (ulnar) to lateral Figure 3 Axial CT of same patient in Figure 2 demonstrating screw penetration through the dorsal cortex of the distal radius just radial to Lister s tubercle Ann R Coll Surg Engl 2016; 98: 138 142 139

Figure 4 Screw protrusion in the radiocarpal joint upon dissection also revealed that one screw had penetrated the radiocarpal joint (RCJ) (Fig 3): this phenomenon was not detected on ultrasound. In CSN3, 2mm-longer screws were inserted. Three of the four screws were prominent upon ultrasound and the fourth screw was at the level of dorsal cortex in extensor compartment four. In CSN4, screws were inserted 4mm-longer than the measured depth. Three of the four screws were found to be long on ultrasound and dissection (Fig 5). All results are summarised in Table 1. Specificity of ultrasound at detection of prominent screws was 100%. No screws were classified as prominent upon ultrasound and found subsequently not to be prominent upon dissection. Sensitivity of ultrasound was 100% for dorsal penetration of screws. One screw was found in the RCJ and another screw penetrated into the DRUJ, but this was not prominent beyond the dorsal cortex. We expect these to be picked up by fluoroscopy intra operatively. We did not use fluoroscopy for our cadaveric study. We wished only to investigate prominence of dorsal screws on ultrasound and correlate findings by dorsal dissection. Discussion The complex geometry of the dorsal cortex of the distal radius makes confirmation of screw length with fluoroscopic or standard radiographic imaging methods quite difficult. 24 26 Protrusion of distal locking screws on the dorsum of the distal radius can lead to attrition and rupture of extensor tendons. 15,16,27 30 and hence reduced function, pain and disability The dorsal tangential view 31,32 has been adopted in many UK centres (including our unit) to assess screw length intraoperatively, and often results in a change of screw. Nevertheless, whether routine use can reduce the prevalence of penetration of dorsal screws is not known. Figure 5 Screw prominence revealed upon dissection The gold standard for assessment of long screws is CT but is not practical intraoperatively as well as being timeconsuming, expensive and exposes the patient to radiation. Ultrasound has two advantages for assessment of penetration of dorsal screws: (i) allows for dynamic examination; (ii) can be used if hardware limits use of magnetic resonance imaging. 17,31,32 The screw appears as a hyperechoic structure created by multiple, parallel, oblique echogenic lines that correspond to the screw thread (Figure 2). Figure 3 shows similar pathology on axial CT. Bianchi et al 17 used ultrasound to detect penetration of dorsal screws after fixation of the distal radius using a volar plate. The diagnosis was confirmed surgically in all nine patients who had fixation using a volar plate. They correlated the ultrasound appearance in vitro with appearance of a protruding screw tip as multiple hyperechoic lines in vivo, similar to our findings. They concluded that, apart from detection of dorsal-screw penetration, ultrasound offers the added advantage of the diagnosis of tears to the extensor tendon and tendon disorders such as tenosynovitis. Tenosynovitis can also affect outcomes after volar plating. Ultrasound has the advantage of detection of this disorder 14,17 (though this is operator-dependent). Thus, ultrasound can be helpful for the detection of associated tendon disorders, which can limit functional recovery of distal-radius fractures. In their clinical and cadaveric study, De Maeseneer et al 33 showed that ultrasound can be used to diagnose tears of the extensor pollicis longus, and similar findings were noted by Santiago et al in their comparative study. 34 We found that the sensitivity and specificity of ultrasound was 100% for the detection of penetration of dorsal 140 Ann R Coll Surg Engl 2016; 98: 138 142

Table 1 Ultrasound and dissection findings in cadavers using different screw lengths Cadaver specimen number Screw length* (mm) Ultrasound protrusion (mm) Dissection prominence Extensor zone 1 (2mm shorter screw) DRUJ 2 (normal length screw) 0 0 0 RCJ 0 0 0 0 0 0 0 1 1 3 (2mm longer screw) +2 1 1 4 +2 0 0 +2 3 3 3 (in EPL groove) +2 0 0 4 (4mm longer screw) +4 4 3 4 +4 0 0 +4 4 4 2 +4 4 5 1 * Screws inserted in order 1 4 as shown in Fig1. DRUJ, distal radioulnar joint; EPL, extensor pollicis longus; RCJ, radiocarpal joint screws after volar plating in fractures of the distal radius. Importantly, no screws were identified as being long when they were in fact not prominent. These findings suggest that, if screws are found to be prominent on ultrasound, then undertaking CT before elective removal of hardware is not necessary. Intra-articular penetration of screws occurred in our study because screws were placed without fluoroscopic guidance. Ultrasound could not be used to identify these screws as being intra-articular. We did not attempt to identify screws within the joint, but one would expect these screws to be readily identifiable using standard fluoroscopic views intraoperatively. Ultrasound can complement fluoroscopic views obtained intraoperatively, and its use may be helpful in difficult cases. The method described here is an accurate and costeffective way of identification of penetration of dorsal screws after volar plating of the distal radius. It could be used as a simple screening method to complement clinical examination in the outpatient setting. Conclusion Ultrasound is a quick, inexpensive and a non-traumatic imaging method that allows assessment of tendons and measurement of protruding dorsal screws. This method is simple, safe, can be undertaken with minimal training, and could be used for the early diagnosis of dorsal-screw prominence, thereby reducing the need for CT. Acknowledgement We thank Dr Catherine Molyneux, Director of Anatomical Studies, for providing the cadavers and the facilities within the department of Anatomy. References 1. Karantana A, Davis TRC. Extra-articular fractures of the distal radius a European view point. Hand Clin 2012; 28: 145 150. 2. Costa ML, Achten J, Parsons NR et al. UK DRAFFT a randomised controlled trial of percutaneous fixation with kirschner wires versus volar locking-plate fixation in the treatment of adult patients with a dorsally displaced fracture of the distal radius. BMC Musculoskelet Disord 201; 12: 201. 3. Rozental TD, Blazar PE, Franko OI, Chacko AT, Earp BE, Day CS. Functional outcomes for unstable distal radial fractures treated with open reduction and internal fixation or closed reduction and percutaneous fixation. A prospective randomized trial. J Bone Joint Surg Am 2009; 91:1,837 1,846. 4. Esposito J, Schemitsch EH, Saccone M, Sternheim A. External fixation versus open reduction with plate fixation for distal radius fractures: a meta-analysis of randomised controlled trials. Injury 2013; 44: 409 416. 5. Jeudy J, Steiger V, Boyer P, Cronier P, Bizot P. Treatment of complex fractures of the distal radius: a prospective randomised comparison of external fixation "versus" locked volar plating. Injury 2012; 43: 174 179. 6. Karantana A, Downing ND, Forward DP. Surgical treatment of distal radial fractures with a volar locking plate versus conventional percutaneous methods a randomized controlled trial. J Bone Joint Surg Am 2013; 95: 1,737 1,744. 7. Gomez Rice A, Villarrubia E, Bau A. Methodological issues of: should unstable extra-articular distal radial fractures be treated with fixed-angle volar-locked plates or percutaneous Kirschner wires? A prospective randomised controlled trial. Injury 2012; 43: 1,219 1,222. 8. Chung KC, Watt AJ, Kotsis SV, Margaliot Z, Haase SC, Kim HM. Treatment of unstable distal radial fractures with the volar locking plating system. J Bone Joint Surg 2006; 88: 2,687 2,694. Ann R Coll Surg Engl 2016; 98: 138 142 141

9. Twigt B, Bemelman M, Lansink K, Leenen L. Type C distal radial fractures treated with conventional AO plates: an easy and cost-saving solution in a locking plate era. Int Orthop 2013; 37: 483 488. 10. Murakami K, Abe Y, Takahashi K. Surgical treatment of unstable distal radius fractures with volar locking plates. J Orthop Sci 2007; 12: 134 140. 11. Hull P, Baraza N, Gohil M et al. Volar locking plates versus K-wire fixation of dorsally displaced distal radius fractures a functional outcome study. J Trauma 2011; 70: E125 E128. 12. Obert L, Rey P-B, Uhring J et al. Fixation of distal radius fractures in adults: a review. Orthop Traumatol Surg Res 2013; 99: 216 234. 13. Wang J, Yang Y, Ma J et al. Open reduction and internal fixation versus external fixation for unstable distal radial fractures: a meta-analysis. Orthop Traumatol Surg Res 2013; 99: 321 331. 14. De Maeseneer M, Marcelis S, Osteaux M, Jager T, Machiels F, Van Roy P. Sonography of a rupture of the tendon of the extensor pollicis longus muscle: initial clinical experience and correlation with findings at cadaveric dissection. Am J Roentgenol 2005; 184: 175 179. 15. Hattori Y, Doi K, Sakamoto S, Yukata K. Delayed rupture of extensor digitorum communis tendon following volar plating of distal radius fracture. Hand Surg 2008; 13: 183 185. 16. Maschke SD, Evans PJ, Schub D, Drake R, Lawton JN. Radiographic evaluation of dorsal screw penetration after volar fixed-angle plating of the distal radius: a cadaveric study. Hand (NY) 2007; 2: 144 150. 17. Bianchi S, van Aaken J, Glauser T, Martinoli C, Beaulieu J-Y, Santa DD. Screw impingement on the extensor tendons in distal radius fractures treated by volar plating: sonographic appearance. Am J Roentgenol 2008; 191: W199 W203. 18. Cho N-Y, Seo C-Y, Kim M-S, Kim H-S, Lee K-B. Extensor pollicis longus rupture after distal radius fracture. J Korean Fracture Soc 2012; 25: 52 57. 19. Bianchi S, van Aaken J, Glauser T, Martinoli C, Beaulieu J-Y, Santa Della D. Screw Impingement on the Extensor Tendons in Distal Radius Fractures Treated by Volar Plating: Sonographic Appearance. http://dxdoiorg/102214/ajr081057. American Roentgen Ray Society; 2012 Nov 23;191(5):W199 W203. 20. Sügün TS, Karabay N, Gürbüz Y, Ozaksar K, Toros T, Kayalar M. Screw prominences related to palmar locking plating of distal radius. J Hand Surg Eur Vol 2011; 36: 320 324. 21. Guillin R, Bianchi S. Sonographic assessment of orthopedic hardware impingement on soft tissues of the limbs. J Ultrasound 2012; 15: 50 55. 22. Margaliot Z, Haase SC, Kotsis SV, Kim HM, Chung KC. A meta-analysis of outcomes of external fixation versus plate osteosynthesis for unstable distal radius fractures. J Hand Surg 2005; 30: 1,185.e1 1,185.e17. 23. Pichler W, Windisch G, Schaffler G, Rienmüller R, Grechenig W. Computer tomography aided 3D analysis of the distal dorsal radius surface and the effects on volar plate osteosynthesis. J Hand Surg Eur Vol 2009; 34: 598 602 24. Thomas AD, Greenberg JA. Use of fluoroscopy in determining screw overshoot in the dorsal distal radius: a cadaveric study. J Hand Surg Am 2009; 34: 258 261. 25. Sügün TS, Karabay N, Gürbüz Y, Ozaksar K, Toros T, Kayalar M. Screw prominences related to palmar locking plating of distal radius. J Hand Surg Eur Vol. 2011 May;36(4):320 4. 26. Park DH, Goldie BS.. Volar plating for distal radius fractures do not trust the image intensifier when judging distal subchondral screw length. Tech Hand Up Extrem Surg 2012; 16: 169 172. 27. Failla JM, Koniuch MP, Moed BR. Extensor pollicis longus rupture at the tip of a prominent fixation screw: report of three cases. J Hand Surg Am 1993; 18: 648 651. 28. Benson EC, DeCarvalho A, Mikola EA et al. Two potential causes of EPL rupture after distal radius volar plate fixation. Clin Orthop Relat Res 2006; 451: 218 222. 29. Al-Rashid M, Theivendran K, Craigen MA. Delayed ruptures of the extensor tendon secondary to the use of volar locking compression plates for distal radial fractures. J Bone Joint Surg Br 2006; 88: 1,610 1,612. 30. Wong-Chung J, Quinlan W. Rupture of extensor pollicis longus following fixation of a distal radius fracture. Injury 1989; 20: 375 376. 31. Ozer K, Toker S. Dorsal tangential view of the wrist to detect screw penetration to the dorsal cortex of the distal radius after volar fixed-angle plating. Hand (NY) 2011; 6: 190 193. 32. Ozer K, Wolf JM, Watkins B, Hak DJ. Comparison of 4 fluoroscopic views for dorsal cortex screw penetration after volar plating of the distal radius. J Hand Surg Am 2012; 37: 963 967. 33. De Maeseneer M, Marcelis S, Osteaux M, Jager T, Machiels F, Van Roy P.. Sonography of a rupture of the tendon of the extensor pollicis longus muscle: initial clinical experience and correlation with findings at cadaveric dissection. AJR Am J Roentgenol. 2005 Jan; 184(1):175 9 34. Santiago FR, Plazas PG, Fernández JM.. Sonography findings in tears of the extensor pollicis longus tendon and correlation with CT, MRI and surgical findings. Eur J Radiol 2008; 66: 112 126. 142 Ann R Coll Surg Engl 2016; 98: 138 142