Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area. Lightweight single-phase system Adjustable throughout application Rigid construct using three basic components
Mandible External Fixator II The Mandible External Fixator II allows the surgeon to create a rigid construct using three basic components: 4.0 mm rods, 2.0 mm and 2.5 mm Kirschner wires and/or 2.5 mm/4.0 mm Schanz screws, and adjustable snap-on clamps. Features Lightweight single-phase system Adjustable throughout application Adjustable Clamp (03.305.006) Accepts a 2.5 mm/4.0 mm Schanz screw, 2.5 mm Kirschner wire, or 4.0 mm rod on each end of the clamp Snap-on design allows clamp placement between previously applied clamps Connects two rods in any orientation, for greater versatility of Schanz screw/kirschner wire placement and frame construction Maintains rod position during frame assembly and fracture reduction Adjustable Parallel Pin Clamp (03.305.105) Allows two Kirschner wires to be implanted in the same small bone fragment Snap-on design allows clamp placement between previously applied clamps Allows greater versatility of Schanz screw/kirschner wire placement and frame construction Top portion of the clamp accepts a 2.5 mm Kirschner wire and 4.0 mm rod, and the bottom portion of the clamp accepts two 2.0 mm or 2.5 mm Kirschner wires 2.0 mm and 2.5 mm Kirschner Wires (292.79 and 292.75) For fixation of fragments that may be too small for 4.0 mm Schanz screws. Mandible External Fixator II
Anatomic Schanz screws For specific regions of the mandible Shouldered to prevent overinsertion Adjustable throughout application Postoperative adjustability Clamps snap onto rod and pins Self-contained set: implants and instruments in one case Indications The Mandible External Fixator II is intended to stabilize and provide treatment for fractures of the maxillofacial area, including: Severe open mandibular fractures Highly comminuted closed fractures Nonunions and delayed unions (especially associated with infection) Fractures associated with infection Tumor resections Facial deformity corrections Gunshot wounds Panfacial fractures Burn maintenance Bone grafting defects
Alternative Frame Configurations Frame incorporating an Adjustable Parallel Pin Clamp As applied on a mandible with a fragment that is too small to accept two Schanz screws. 03.305.006 Adjustable Clamps 5 03.305.105 Adjustable Parallel Pin Clamp 1 04.305.100 4.0 mm Titanium Pre-Bent Rod, 1 Full Mandible with Ramus 04.305.128 2.5 mm/4.0 mm Titanium 1 04.305.130 Schanz Screws, for Ramus 04.305.140 2.5 mm/4.0 mm Titanium 2 04.305.142 Schanz Screws, for Symphysis 292.79 2.0 mm Kirschner Wires 2 392.179 Protective Caps, for 2.0 mm 2 Kirschner Wires One-half Frame As applied on an infected angle fracture. 03.305.006 Adjustable Clamps 4 04.305.103 4.0 mm Titanium Pre-Bent Rod, 1 One-Half Mandible 04.305.128 2.5 mm/4.0 mm Titanium 2 04.305.130 Schanz Screws, for Ramus Mandible External Fixator II
Modular Carbon Fiber Frame As applied on an infected angle fracture. The modular carbon fiber frame provides improved x-ray imaging. 03.305.006 Adjustable Clamps 6 04.305.128 2.5 mm/4.0 mm Titanium 2 04.305.130 Schanz Screws, for Ramus 395.60 4.0 mm Carbon Fiber Rods, 2 60 mm 395.61 4.0 mm Carbon Fiber Rod, 1 80 mm Double-stacked Frame As applied on a resected mandible. 03.305.006 Adjustable Clamps 7 04.305.100 4.0 mm Titanium Pre-bent Rod, 1 Full Mandible with Ramus 04.305.103 4.0 mm Titanium Pre-Bent Rod, 1 One-Half Mandible 04.305.128 2.5 mm/4.0 mm Titanium 2 04.305.130 Schanz Screws, for Ramus 04.305.140 2.5 mm/4.0 mm Titanium 1 or Schanz Screw, for Symphysis 04.305.142
Alternative Frame Configurations continued Modular Frame As applied on a comminuted fracture. A modular frame can be created to accommodate the fracture location. 03.305.006 Adjustable Clamps 7 04.305.101 4.0 mm Titanium Pre-Bent Rod, 1 Full Mandible 04.305.103 4.0 mm Titanium Pre-Bent Rod, 1 One-Half Mandible 04.305.128 2.5 mm/4.0 mm Titanium 1 or Schanz Screws, for Ramus 04.305.130 04.305.140 2.5 mm/4.0 mm Titanium 2 04.305.142 Schanz Screws, for Symphysis Mandible External Fixator II
Mandible External Fixator II Set (01.305.001) Graphic Case 306.720 Mandible External Fixator II Set Graphic Case Base 306.721 Mandible External Fixator II Set Graphic Case Lid Schanz Screws 2.5 mm/4.0 mm Titanium Schanz Screws, self-drilling, 3 ea. 04.305.128 for Ramus, 8 mm thread length 04.305.130 for Ramus, 10 mm thread length 04.305.134 for Body, 14 mm thread length 04.305.138 for Body, 18 mm thread length 04.305.140 for Symphysis, 20 mm thread length 04.305.142 for Symphysis, 22 mm thread length Wires 292.75 2.5 mm Kirschner Wire with Thread, 150 mm, trocar point, 15 mm thread length (10/pkg.) 292.79 2.0 mm Kirschner Wire with Thread, 150 mm, trocar point, 15 mm thread length (10/pkg.) Rods 4.0 mm Titanium Pre-bent Rods, 2 ea. 04.305.100 Full mandible with ramus 04.305.101 Full mandible 04.305.102 Three-quarter mandible 04.305.103 One-half mandible 4.0 mm Carbon Fiber Rods, 2 ea. 395.60 60 mm 395.61 80 mm 395.62 100 mm 395.63 120 mm Clamps 03.305.006 Adjustable Clamp, 8 ea. 03.305.105 Adjustable Parallel Pin Clamp, 2 ea. Protective Caps 392.179 for 2.0 mm Kirschner wires (10/pkg.) 392.24 for 2.5 mm Kirschner wires (10/pkg.) 394.99 for 4.0 mm bars and rods (10/pkg.)
Mandible External Fixator II Set (01.305.001) continued Instruments 03.305.004 1.8 mm/4.0 mm Stepped Drill Bit, Stryker J-latch, with 24 mm stop, 2 ea. 03.305.005 Measuring Device, for Mandible External Fixator 03.305.007 4.0 mm Rod Bender, for Mandible External Fixator 03.305.009 Rod Template, for 4.0 mm rods, for Mandible External Fixator 03.305.021 Drill Guide for 2.0 mm/2.5 mm Kirschner Wires 03.305.022 2.0 mm/2.5 mm Obturator 03.305.024 Schanz Screw Rapid Driver, 2 ea. 03.305.025 4.0 mm Obturator 03.305.026 Drill Guide/Cannula, extra long, for 4.0 mm Schanz Screws, 2 ea. 03.305.027 Adaptor for Ratcheting Screwdriver Handle 311.023 Ratcheting Screwdriver Handle 395.35 Combination Wrench, 7 mm width across flats 395.36 Socket Wrench, 7 mm width across flats 395.911 Drill Sleeve Handle Also Available* 03.305.500S Mandible External Fixator, 500 mm Rod Kit, sterile (includes one straight 500 mm Titanium Rod and one 500 mm Rod Template) 359.217 Cutter (small) 388.72 Rod Cutter (large) 391.944 Wire Cutters 4.0 mm Carbon Fiber Rods 395.64 140 mm 395.65 160 mm 395.66 180 mm 395.67 200 mm 511.701 Compact Air Drive II * These items will not fit in the Mandible External Fixator II graphic case. Available nonsterile and sterile-packed. Delete S from catalog number for nonsterile product. 1302 Wrights Lane East West Chester, PA 19380 Telephone: (610) 719-5000 To order: (800) 523-0322 Fax: (610) 251-9056 Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185 www.synthes.com 2007 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 11/08 J6841-B