Comparison of Mercury and Aneroid Blood Pressure Measurements in Youth

Similar documents
NIH Public Access Author Manuscript JAMA Intern Med. Author manuscript; available in PMC 2015 August 01.

ORIGINAL INVESTIGATION. C-Reactive Protein Concentration and Incident Hypertension in Young Adults

Assessing Blood Pressure Accuracy of an Aneroid Sphygmomanometer in a National Survey Environment

Progress in the Control of Childhood Obesity

ipad Increasing Nickel Exposure in Children

that youth with either type 1 diabetes mellitus (T1DM) or type 2 diabetes mellitus (T2DM) have increased

Trends in Incidence of Type 1 Diabetes Among Non-Hispanic White Youth in the U.S.,

Age Limit of Pediatrics

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

Wellness Coaching for People with Prediabetes

Understanding Confounding in Research Kantahyanee W. Murray and Anne Duggan. DOI: /pir

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

2017 Recommendations for Preventive Pediatric Health Care COMMITTEE ON PRACTICE AND AMBULATORY MEDICINE, BRIGHT FUTURES PERIODICITY SCHEDULE WORKGROUP

change in blood pressure (BP) percentile and development of hypertension (HTN).

THE ACCURATE ASSESSMENT

In recent years, concern has increased markedly

Chapter Two Renal function measures in the adolescent NHANES population

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health

Methods DEVICES AND TECHNOLOGY

A Systematic Review of Vision Screening Tests for the Detection of Amblyopia

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

Chronic kidney disease (CKD) has received

Reduced 10-year Risk of CHD in Patients who Participated in Communitybased DPP: The DEPLOY Pilot Study

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Are Patient-Held Vaccination Records Associated With Improved Vaccination Coverage Rates?

Molluscum BOTE Sign: A Predictor of Imminent Resolution

The Impact of Social Networks on Parents Vaccination Decisions

ORIGINAL INVESTIGATION. Effects of Prehypertension on Admissions and Deaths

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

infant s gestational age, age at phototherapy initiation, and TSB relative to the treatment threshold at phototherapy termination.

In clinical practice, blood pressure (BP) treatment

Family-Based Treatment of Severe Pediatric Obesity: Randomized, Controlled Trial

Analytical Methods: the Kidney Early Evaluation Program (KEEP) The Kidney Early Evaluation program (KEEP) is a free, community based health

Blood Pressure Competency Training

United Recommended Childhood and Adolescent Immunization Schedule States, 2013

Lecture #3 - Blood Pressure Recording Procedure. Equipment

Failure to Confirm High Blood Pressures in Pediatric Care - Quantifying the risks of misclassification

External Oscillatory Blood Pressure - EOBPTM

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Clinical Guidelines. Annals of Internal Medicine. Annals of Internal Medicine

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

Childhood Obesity Predicts Adult Metabolic Syndrome: The Fels Longitudinal Study

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Residency Training in Transition of Youth With Childhood-Onset Chronic Disease Manisha S. Patel and Kitty O'Hare. DOI: /peds.

Screening for High Blood Pressure in Adults During Ambulatory Nonprimary Care Visits: Opportunities to Improve Hypertension Recognition

CHAPTER 3 DIABETES MELLITUS, OBESITY, HYPERTENSION AND DYSLIPIDEMIA IN ADULT CENTRAL KERALA POPULATION

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

Training and Certification of Blood Pressure Observers

Blood Pressure Measurement by Pulse Oxymetric Method and Comparison with Conventional Technique

HHS Public Access Author manuscript Am J Kidney Dis. Author manuscript; available in PMC 2017 July 05.

Data Collection Worksheet

risk of cigarette initiation among adolescents in the transition to adulthood when the sale of cigarettes becomes legal.

Managing Blood Pressure: It Takes a Team

Prevalence of Obesity and Severe Obesity in US Children,

TITLE: Breast Density Assessment by Dual Energy X-ray Absorptiometry in Women and Girls. Rachel Novotny, Ph.D. Honolulu, HI 96822

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Pulmonary Embolism in the Pediatric Emergency Department. abstract ARTICLE

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

A CROSS SECTIONAL STUDY OF RELATIONSHIP OF OBESITY INDICES WITH BLOOD PRESSURE AND BLOOD GLUCOSE LEVEL IN YOUNG ADULT MEDICAL STUDENTS

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

Update on Tuberculosis Skin Testing of Children

The Seventh Report of the Joint National Commission

Importance Of Taking Manual Blood Pressure In Both Arms

doi: /fampra/cmw053

AMERICAN ACADEMY OF PEDIATRICS

9/27/2016. Adherence and Monitoring of Hypertensive Patients. Conflicts of Interest. Brought to you though a partnership:

CLOSING OF THE ANTERIOR FONTANELLE

Bedside Ultrasound in Pediatric Practice

Transitions To and From At-Risk Alcohol Use In Adults In the United States

CLINICAL PRACTICE GUIDELINE Guidance for the Clinician in Rendering Pediatric Care EXECUTIVE SUMMARY FROM THE AMERICAN ACADEMY OF PEDIATRICS

Reflections of Dietary Studies With Children in the Ten- State Nutrition Survey of

Hypertension and the SPRINT Trial: Is Lower Better

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

Obesity prevalence, disparities, trends and persistence among US children <5 y

Asthma and Tobacco: Double Trouble for Wisconsin Adolescents

AMERICAN ACADEMY OF PEDIATRICS

Strength Training, Weight and Power Lifting, and Body Building by Children and. Adolescent. 0 Committee on Sports Medicine

Home Blood Pressure Log

Chapter-III METHODOLOGY. In this chapter, the procedure adopted for the selection of the

CLINICAL PRACTICE GUIDELINE Guidance for the Clinician in Rendering Pediatric Care EXECUTIVE SUMMARY FROM THE AMERICAN ACADEMY OF PEDIATRICS

APPENDIX AVAILABLE ON THE HEI WEB SITE

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

Cost-effectiveness of Metered-Dose Inhalers for Asthma Exacerbations in the Pediatric Emergency Department

HYPERTENSION IS A MAJOR RISK

Society for Behavioral Medicine 33 rd Annual Meeting New Orleans, LA

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

Prevalence, awareness of hypertension in rural areas of Kurnool

Bronchiolitis Management Before and After the AAP Guidelines

Prevalence of Cardiovascular Disease Risk Factors Among US Adolescents,

ARTICLE. Dietary Nutrients and Blood Pressure in Urban Minority Adolescents at Risk for Hypertension

NIH Public Access Author Manuscript Sex Transm Infect. Author manuscript; available in PMC 2013 June 21.

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Transcription:

ARTICLE Comparison of Mercury and Aneroid Blood Pressure Measurements in Youth AUTHORS: Amy S. Shah, MD, MS, a Lawrence M. Dolan, MD, a Ralph B. D Agostino Jr, PhD, b Debra Standiford, CNP, a Cralen Davis, BS, MS, b Lisa Testaverde, MS, c Catherine Pihoker, MD, d Stephen R. Daniels, MD, PhD, e and Elaine M. Urbina, MD, MS, a for the SEARCH for Diabetes in Youth Study Group a Department of Pediatrics, Cincinnati Children s Hospital, University of Cincinnati College of Medicine, Cincinnati, Ohio; b Department of Biostatistical Sciences, Wake Forest University School of Medicine, Winston-Salem, North Carolina; c Department of Epidemiology, Colorado School of Public Health, University of Colorado Denver, Aurora, Colorado; d Department of Pediatrics, University of Washington, Seattle, Washington; and e Department of Pediatrics, Children s Hospital Colorado, Aurora, Colorado KEY WORDS pediatrics, blood pressure ABBREVIATIONS BP blood pressure DPPOS Diabetes Prevention Program Outcomes Study SEARCH SEARCH for Diabetes in Youth Study Dr Shah aided in the concept and study design, interpreted the data analysis, and wrote the manuscript; Dr Dolan aided in the concept and study design, interpreted the data analysis, and revised the manuscript for intellectual content; Dr D Agostino aided in the concept and study design, conducted the analysis, and helped in the interpretation of the data; Ms Standiford aided in the concept and study design and revised the manuscript for intellectual content; Mr Davis aided in the concept and study design and conducted the analysis; Ms Testaverde revised the manuscript for intellectual content; Dr Pihoker revised the manuscript for intellectual content; Dr Daniels revised the manuscript for intellectual content; and Dr Urbina aided in the concept and study design, interpreted the data analysis, and revised the manuscript for intellectual content. The contents of this article are solely the responsibility of the authors and do not necessarily represent the official position of the Centers for Disease Control and Prevention and the National Institute of Diabetes and Digestive and Kidney Diseases. www.pediatrics.org/cgi/doi/10.1542/peds.2011-3087 doi:10.1542/peds.2011-3087 Accepted for publication Jan 12, 2012 Address correspondence to Amy S. Shah MD, MS, Division of Endocrinology, ML 7012, Cincinnati Children s Hospital, 3333 Burnet Ave, Cincinnati, OH 45229. E-mail: amy.shah@cchmc.org (Continued on last page) WHAT S KNOWN ON THIS SUBJECT: As a result of safety and environmental concerns about mercury, aneroid sphygmomanometers have replaced mercury-filled devices for blood pressure measurements. Despite this change, few studies have compared the 2 devices. WHAT THIS STUDY ADDS: Little clinical variation exists between blood pressure measurements obtained from an aneroid or mercury device, suggesting that either device could be used in a research or clinical setting. abstract OBJECTIVE: Because of concerns about the safety and environmental impact of mercury, aneroid sphygmomanometers have replaced mercury-filled devices for blood pressure (BP) measurements. Despite this change, few studies have compared BP measurements between the 2 devices. METHODS: The SEARCH for Diabetes in Youth Study conducted a comparison of aneroid and mercury devices among 193 youth with diabetes (48% boys, aged 12.9 6 3.7 years; 89% type 1). Statistical analyses included estimating Pearson correlation coefficients, Bland-Altman plots, paired t tests, and fitting regression models, both overall and stratified by age (,10 vs $10 18 years). RESULTS: Mean mercury and aneroid systolic and diastolic BPs were highly correlated. For the entire group, there was no significant difference in mean systolic BP using the aneroid device, but there was a 2 1.53 6 5.06 mm Hg difference in mean diastolic BP. When stratified by age, a lower diastolic BP (21.78 6 5.2 mm Hg) was seen in those $10 to 18 years using the aneroid device. No differences in systolic BP were observed, and there were no differences in BP by device in individuals,10 years. Regression analyses did not identify any explanatory variables. CONCLUSIONS: Although a small discrepancy between diastolic BP measurements from aneroid versus mercury devices exists, this variation is unlikely to be clinically significant, suggesting that either device could be used in research or clinical settings. Pediatrics 2012;129:e1205 e1210 PEDIATRICS Volume 129, Number 5, May 2012 e1205

Auscultation measurements using a mercury sphygmomanometer are the gold standard for documenting blood pressure (BP). 1,2 However, because of concerns about the environment and personal health, regulatory agencies have recommended removal of these devices. 3 Thus, despite a lack of studies documenting the comparability of other devices, clinical programs and research studies have been required to use electronic and aneroid devices. Only 1 study has previously compared BP measurements obtained by mercury and aneroid devices in pediatrics. NHANES reported a small mean difference in systolic BP (+1.1 mm Hg) in youth aged 8 to 17 years with the aneroid sphygmomanometer. 4 Therefore, the SEARCH for Diabetes in Youth Study (SEARCH) sought to validate these results and to determine whether differences between the 2 devices were present by age. METHODS SEARCH SEARCH is a multicenter, populationbased, observational study of diabetes mellitus in youth. Detailed study methods have been described previously. 5 In brief, SEARCH began conducting population-based ascertainment of cases of diabetes in youth,20 years at diagnosis in 2001 and continues through the present. Cases are identified in geographically defined populations in Colorado, Ohio, South Carolina, and Washington; among health plan enrollees in California and Hawaii; and Indian Health Service beneficiaries from 4 American Indian populations. Cases are considered valid if they were diagnosed with diabetes by a health care provider. Youth with diabetes were identified and asked to complete a survey that collected information on date of birth, race/ethnicity, type of diabetes, date of diagnosis, and current diabetes treatment. All participants who completed this survey (except those with known secondary diabetes) were invited to participate in an in-person study visit while metabolically stable (no episode of diabetic ketoacidosis during the previous month). At the beginning of the visit, consent was obtained from participants or their parents (if,18 years of age), and assent was obtained from younger subjects as directed by local institutional review boards. Information collected at the visit included questionnaires for demographics, blood and urine samples, and a brief physical examination that included height, weight, and BP data. Study Population The study population for this analysis consisted of all individuals who participatedinanin-personresearchvisitbetween September 1, 2007, and June 30, 2008, at 3 of the 6 SEARCH clinical sites: Hawaii, Ohio, and Washington. Approval for this study was obtained from local institutional review boards. Research personnel were trained in BP measurements by using a standard procedure derived from the National High Blood Pressure Education Program Working Group on High Blood Pressure in Children and Adolescents (Fourth Report). 1 BP Equipment and Measurements Mercury measurements were performed by using either a portable or wallmounted Baumometer sphygmomanometer Kompak Model-260 mm Hg (WA Baum, Copiague, NY). Aneroid measurements were performed by using the Welch Allyn Tycos 767-Series Mobile Aneroid with a weighted base (model 7670-04; Welch Allyn, Skaneateles Falls, NY). Before initial use, calibration checks of the aneroid sphygmomanometers were performed by using the Netech DigiMano 1000 digital pressure-vacuum meter (part number 200-2000 IN). Calibration checks were repeated every 6 months and whenever the device was transported in a motor vehicle from 1 location to another. Before the first BP measurement, all participants were seated with both feet on the floor after resting quietly for a minimum of 5 minutes. The right arm wasusedforallbpmeasurementsunless there was a contraindication. The subject s arm was supported on a table at heart level. The examiner selected an appropriate-sized cuff based on the circumference of the upper arm. 1 The pulse pressure was determined by locating the radial pulse and inflating the cuff until the radial pulse disappeared. The maximum inflation level was then recorded as the pulse pressure plus 30 mm Hg. All participants then had 4 BP measurements performed with the same sized cuff. The first and fifth Korotkoff sounds were recorded for each of the 4 measurements. Subjects rested quietly for a minimum of 30 seconds between measurements. The first 2 measurements were always taken with an aneroid sphygmomanometer. The third and fourth measurements were measured by using either aneroid then mercury, or mercury then aneroid. The order of these third and fourth measurements was randomly assigned by the SEARCH Coordinating Center. Each of the 3 clinical sites were sent a list specifying the order(mercury then aneroid oraneroid then mercury) for performing the third and fourth BP measurements for each participant. This list was generated by using Proc Plan, a statistical procedure in SAS (version 9.2; SAS Institute, Cary, NC) for generating randomization plans. Only the third and fourth BP measurements were used in these analyses. Statistical Analysis Statistical analyses were performed by using SAS software (version 9.2; SAS Institute, Cary NC). All data were analyzed for the entire cohort and then by age group,,10 years and $10 to 18 e1206 SHAH et al

ARTICLE years. Pearson correlation coefficients for systolic and diastolic BP measurements from the 2 BP devices (aneroid and mercury) were calculated. In addition, we fit Bland-Altman plots to provide a visual comparison between the 2 methods. Next, mean differences in systolic and diastolic BP obtained from each device were compared by using paired t tests to determine if they were different from zero. This was used to estimate the difference in millimeters of mercury observed between devices and to generate a correction factor between devices. Thus, if we assume that the observed average difference between devices represents the shift in blood pressure measurements due to the different devices, then adding this difference as a correction factor to the aneroid measures would allow for the new (aneroid) measures to have the same observed BP values as the mercury measures. Finally, general linear models were fit to determine if the reason for the observed differences in BP measures between devices could be identified. Models were adjusted for age, gender, race/ethnicity, BMI, clinic site, or type of diabetes. P values of,.05 were deemed significant in these analyses. RESULTS One hundred ninety-three youth (48% male youth) with a mean age of 12.9 6 3.7 years (age range 3.9 18.9 years) who had a mean duration of diabetes of 9.2 6 5.6 months participated in the comparison of the 2 BP devices (Table 1). Of these participants, 171 had a clinical diagnosis of type 1 diabetes, and 22 had type 2 diabetes. A majority of the participants were non-hispanic white. When stratified by age (aged,10 and $10 18 years), there were no significant differences between the groups except for diabetes type and BMI because the youth with type 2 diabetes were exclusively in the $10- to 18-year age category. TABLE 1 Characteristics of the Study Population The randomization of the sequence of BP measurements (aneroid then mercury or mercury then aneroid) resulted in 92 subjects randomized to receive aneroid then mercury and 101 individuals to mercury then aneroid for the third and fourth BP measurements. Mercury and aneroid systolic and diastolic BP measures were significantly correlated for the total sample and by age group (Table 2). For systolic BP, the overall correlation was r =.94, P,.0001. By age, r =.90, P,.0001 for,10 years and r =.94, P,.0001 for $10 to 18 years. Likewise, for diastolic BP, the overall correlation was very high: r =.82, P,.0001 and r =.83, P,.0001 and r =.81, P,.0001 for,10 years and $10 to 18 years, respectively. We examined Bland-Altman plots to further assess agreement between the 2 methods. These plots are presented in Fig 1 with 4 panels, 2 for each measure (systolic and diastolic BP) and 2 for each age group (,10 years and $10 18 years). For the older group (Fig 1d) larger differences in diastolic BP seemed to occur for diastolic blood pressures.65 mm Hg; however, there is no clear trend in these data to suggest a specific cutpoint where the 2 methods become discordant. We then tested whether the mean difference in systolic and diastolic BPs Total,10 y $10 to 18 y No. of participants 193 45 148 Male, n (%) 93(48) 21 (47) 72 (49) Female, n (%) 100 (52) 24 (53) 76 (51) Age, y, mean 6 SD 12.85 6 3.7 7.8 6 1.8 14.4 6 2.6 BMI, kg/m 2 21.7 6 7.4 16.7 6 2.3 23.1 6 7.7 Diabetes duration, mo, mean 6 SD a 9.2 6 5.6 10.1 6 6.1 9.0 6 5.4 Type 1 diabetes, n (%) 171 (89) 45 (100) 126 (85) Non-Hispanic white, % 75 69 77 African American, % 9 9 9 Hispanic/Asian Pacific Islander/Other, % 16 22 14 Type 2 diabetes, n (%) 22 (11) 0 (0) 22 (15) Non-Hispanic white, % 23 0 23 African American, % 27 0 27 Hispanic/Asian Pacific Islander/Other, % 50 0 50 a Duration of diabetes indicates duration of diabetes at time of study visit. obtained by the mercury and aneroid devices was significantly different from zero. This was used to estimate a correction factor between devices. Among individuals,10 years, the mean difference in systolic and diastolic BPs was not different from zero (P =.85 and P =.29, respectively). In individuals $10 to 18 years of age, the mean difference in systolic BP was not significant (2.67 6 4.33, P =.06), but the mean difference for diastolic BP was significantly different from zero (21.78 6 5.22, P,.0001). Thus, to equate BPs between devices a correction factor of +1.8 could be added to an aneroid BP measurement in individuals $10 years. Regression models were used to seek for the reason for the differences in diastolic BP by device. Models were only created for diastolic BP as systolic BP was not statistically different by device. Univariate analysis revealed race/ethnicity and BMI z scoretobesignificantly associated with the mean difference in diastolic BP. Nonwhites were found to have a larger difference in diastolic BP (1.71 mm Hg, P =.03),anda1-unitchangeinBMI z score corresponded to a 0.85 mm Hg (P =.014) difference in diastolic BP. In the fully adjusted models, race/ethnicity and BMI z score were no longer significant, and no other predictors to explain the difference in the diastolic BP measures were identified. PEDIATRICS Volume 129, Number 5, May 2012 e1207

TABLE 2 Comparison of Mercury versus Aneroid BP Measurements Systolic BP, (mm Hg) Pearson Correlation Diastolic BP, mm Hg Pearson correlation Systolic BP Different From zero, Diastolic BP Different From Zero, mm Hg a mm Hg a Mercury Aneroid Mercury Aneroid All 105.0 6 13.2 104.5 6 13.3 r =.94, P,.0001 65.6 6 11.1 64.4 6 9.1 r =.82, P,.0001 20.48 6 4.41, P =.13 21.25 6 6.35, P =.007 #10 y 96.0 6 10.4 96.1 6 10.4 r =.90, P,.0001 60.6 6 7.8 59.5 6 7.2 r =.83, P,.0001 0.13 6 4.66, P =.85 20.71 6 4.45, P =.29 $10 18 y 107.7 6 12.8 107.0 6 13.0 r =.94, P,.0001 67.1 6 11.6 65.7 6 9.3 r =.81, P,.0001 20.67 6 4.33, P =.062 21.78 6 5.22, P,.0001 Data are mean 6 SD. a Reference device is the mercury sphygmomanometer. DISCUSSION This study demonstrates that systolic and diastolic BPs measured by mercury and aneroid devices are highly correlated for the entire sample and by age group. In individuals $10 to 18 years, a significantly lower mean diastolic BP was found when using the aneroid sphygmomanometer. On the basis of our analyses, this corresponds to a mean correction of +1.8 mm Hg that could be added to an aneroid BP measurement to equate the 2 devices if both are used in asinglestudy.however,giventhegood agreement between devices, this correction is unlikely to result in clinically significant differences in BP, suggesting that either device could be used in a clinical or research setting. Two previous studies have compared aneroid to mercury devices with only 1 performed in a pediatric population. In middle-aged adults, the Diabetes Prevention Program Outcomes Study(DPPOS) found no difference in mean systolic BP and a small but statistically significant lower diastolic BP when comparing aneroid and mercury devices. 6 This pattern is identical to the findings in our study. These results are in contrast to those reported by NHANES in which a statistically higher systolic BP among 8- to 17- year-olds was found using an aneroid sphygmomanometer. 4 Discrepant results between NHANES and our study may be due to several factors, including participant differences, different analysis techniques, and the presence of diabetes. First, it is possible the study subjects in the 2 cohorts were of different age, gender, race/ethnicity, and BMI distribution. Although the study population was not described in the NHANES article, the DPPOS found participant differences explained 8% to 10% of the variation in BPs between the 2 devices. 6 Second, we stratified our subjects by age to separate those prepubertal (,10 years) from those who were pubertal or postpubertal ($10 18 years). Subjects in the NHANES cohort were analyzed as a single group (8 17 years). 4 Thus, it is possible changes occur in BP measurements during or after puberty that were not accounted for in NHANES report. Finally, there may be differences in BP related to the presence of diabetes, although to our knowledge, this has not been previously described. Thus, it is not clear why there are differences between our study and NHANES. Multivariate analyses did not identify any factors that contributed to the differences in diastolic BP by device. Thus, the reasons for the differences in blood pressure by device can only be speculated. As mentioned earlier, a correction factor of 1.8 could be applied to diastolic BP measurement obtained by using an aneroiddeviceinadolescentsaged$10 to 18 years. This correction factor is only applicable to equate BPs in clinical or research settings where both devices have been used or there has been a change of devices due to more recent regulations preventing mercury BP measurements. In settings where aneroid only technology has been used, we cannot conclude from this study that this correction factor should apply. A potential limitation of our report is that this study was performed in youth who have diabetes. As previously mentioned, the presence of diabetes should not affect the measurements of BP obtained by using the mercury or aneroid devices. Given that our results coincide with those published by the DPPOS, it appears reasonable to conclude that the data presented here are applicable to all youth. It should also be noted that youth with hypertension were not studied, and thus it is unclear whether larger discrepancies between devices exist at higher BPs. CONCLUSIONS This study found no significant difference in mean systolic BP but a slightly e1208 SHAH et al

ARTICLE FIGURE 1 Agreement between aneroid and mercury devices. The red line indicates the mean difference between readings, and the green line is the 95% confidence intervals of agreement. Bland-Altman Plots: A, systolic BP, age 0 9. B, diastolic BP, age 0 9. C, systolic BP, age 10 18. D, diastolic BP, age 10 18. significantly lower mean diastolic BP in youth using the aneroid device. Although the differences described are statistically significant, as suggested by others, 4,6 the differences in BP measurements between aneroid and mercury sphygmomanometers are small and are unlikely to be clinically significant. In conclusion, either device could be used. ACKNOWLEDGMENTS The SEARCH for Diabetes inyouth Studyis indebted to the many youth and their families, and their health care providers, whose participation made this study possible. SiteContractNumbers:KaiserPermanente Southern California (U48/CCU919219, U01 DP000246, and U18DP002714), University of Colorado Denver (U48/CCU819241-3, U01 DP000247, and U18DP000247-06A1), Kuakini Medical Center (U58CCU919256 and U01 DP000245), Children s Hospital Medical Center (Cincinnati) (U48/CCU519239, U01 DP000248, and 1U18DP002709), University of North Carolina at Chapel Hill (U48/CCU419249, U01 DP000254, and U18DP002708-01), University of Washington School of Medicine (U58/ CCU019235-4, U01 DP000244, and U18DP002710-01), Wake Forest University School of Medicine (U48/CCU919219, U01 DP000250, and 200-2010-35171). We acknowledge the involvement of General Clinical Research Centers at the South Carolina Clinical & Translational Research (SCTR) Institute, at the Medical University of South Carolina (National Institute of Health/National Center for Research Resources [NIH/NCRR] grant UL1RR029882); Children shospital and Regional Medical Center (grant M01RR00037); Colorado Pediatric General Clinical Research Center (grant M01 RR00069), and the Barbara Davis Center at the University of Colorado at Denver (Diabetes and Endocrinology Research Center/National Institutes of Health grant P30 DK57516); and the Institutional Clinical and Translational Science Award, NIH/NCRR at the University of Cincinnati (grant 1UL1RR026314-01). PEDIATRICS Volume 129, Number 5, May 2012 e1209

REFERENCES 1. National High Blood Pressure Education Program Working Group on High Blood Pressure in Children and Adolescents. The fourth report on the diagnosis, evaluation, and treatment of high blood pressure in children and adolescents. Pediatrics. 2004(suppl 2, 4th report);114: 555 576. 2. Chobanian AV, Bakris GL, Black HR. The seventh report of the Joint National Committee on the Prevention, Detection, Evaluation, and Treatment of High Blood Pressure. Hypertension. 2003;42(6)1206 1252. 3. Messelbeck J, Sutherland L. Applying environmental product design to biomedical products research. Environ Health Perspect. 2000;108(suppl 6):997 1002 4. Ostchega Y, Prineas RJ, Nwankwo T, Zipf G. Assessing blood pressure accuracy of an aneroid sphygmomanometer in a national survey environment. Am J Hypertens. 2011; 24(3):322 327 5. SEARCH Study Group. SEARCH for Diabetes in Youth: a multicenter study of the prevalence, incidence and classification of diabetes mellitus in youth. Control Clin Trials. 2004;25 (5):458 471 6. Ma Y, Temprosa M, Fowler S, et al; Diabetes Prevention Program Research Group. Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting. Am J Hypertens. 2009;22(3): 263 266 (Continued from first page) PEDIATRICS (ISSN Numbers: Print, 0031-4005; Online, 1098-4275). Copyright 2012 by the American Academy of Pediatrics FINANCIAL DISCLOSURE: The authors have indicated they have no financial relationships relevant to this article to disclose. FUNDING: SEARCH for Diabetes in Youth is funded by the Centers for Disease Control and Prevention (PA numbers 00097, DP-05-069, and DP-10-001) and supported by the National Institute of Diabetes and Digestive and Kidney Diseases. Funded by the National Institutes of Health (NIH). e1210 SHAH et al

Comparison of Mercury and Aneroid Blood Pressure Measurements in Youth Amy S. Shah, Lawrence M. Dolan, Ralph B. D'Agostino Jr, Debra Standiford, Cralen Davis, Lisa Testaverde, Catherine Pihoker, Stephen R. Daniels, Elaine M. Urbina and for the SEARCH for Diabetes in Youth Study Group Pediatrics 2012;129;e1205; originally published online April 16, 2012; DOI: 10.1542/peds.2011-3087 Updated Information & Services References Subspecialty Collections Permissions & Licensing Reprints including high resolution figures, can be found at: /content/129/5/e1205.full.html This article cites 4 articles, 2 of which can be accessed free at: /content/129/5/e1205.full.html#ref-list-1 This article, along with others on similar topics, appears in the following collection(s): Emergency Medicine /cgi/collection/emergency_medicine_sub Information about reproducing this article in parts (figures, tables) or in its entirety can be found online at: /site/misc/permissions.xhtml Information about ordering reprints can be found online: /site/misc/reprints.xhtml PEDIATRICS is the official journal of the American Academy of Pediatrics. A monthly publication, it has been published continuously since 1948. PEDIATRICS is owned, published, and trademarked by the American Academy of Pediatrics, 141 Northwest Point Boulevard, Elk Grove Village, Illinois, 60007. Copyright 2012 by the American Academy of Pediatrics. All rights reserved. Print ISSN: 0031-4005. Online ISSN: 1098-4275.

Comparison of Mercury and Aneroid Blood Pressure Measurements in Youth Amy S. Shah, Lawrence M. Dolan, Ralph B. D'Agostino Jr, Debra Standiford, Cralen Davis, Lisa Testaverde, Catherine Pihoker, Stephen R. Daniels, Elaine M. Urbina and for the SEARCH for Diabetes in Youth Study Group Pediatrics 2012;129;e1205; originally published online April 16, 2012; DOI: 10.1542/peds.2011-3087 The online version of this article, along with updated information and services, is located on the World Wide Web at: /content/129/5/e1205.full.html PEDIATRICS is the official journal of the American Academy of Pediatrics. A monthly publication, it has been published continuously since 1948. PEDIATRICS is owned, published, and trademarked by the American Academy of Pediatrics, 141 Northwest Point Boulevard, Elk Grove Village, Illinois, 60007. Copyright 2012 by the American Academy of Pediatrics. All rights reserved. Print ISSN: 0031-4005. Online ISSN: 1098-4275.