Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Similar documents
2D-LC as an Automated Desalting Tool for MSD Analysis

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Chiral Multicolumn Method Development on the Agilent 1260 Infinity II SFC System

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Performance Characteristics of the Agilent 1290 Infinity II Multicolumn Thermostat

Analysis of Counterfeit Antidiabetic Drugs by UHPLC with the Agilent 1220 Infinity Mobile LC

Transferring a Method for Analysis of DNPH-Derivatized Aldehydes and Ketones from HPLC to UHPLC

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Increasing resolution using longer columns while maintaining analysis time Advantages of the wide power range of the Agilent 1290 Infinity LC System

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

Qualitative and quantitative determination of phenolic antioxidant compounds in red wine and fruit juice with the Agilent 1290 Infinity 2D-LC Solution

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Application Note. Abstract. Authors. Pharmaceutical

Agilent 1260 Infinity Analytical SFC System

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

Authors. Abstract. Introduction. Environmental

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

InfinityLab Purification Capillary Kits

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Performance characteristics of the Agilent 1100 Series preparative pump. Technical Note. Abstract

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Online 2D-LC Analysis of Complex N-Glycans in Biopharmaceuticals Using the Agilent 1290 Infinity 2D-LC Solution

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

HPLC to UHPLC Transfer of USP Method for Amlodipine Besylate Using the Agilent 1290 Infinity II LC

SUPPLEMENTARY MATERIAL

Analysis of HMF by HPLC

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Comparison of a UPLC Method across Multiple UHPLC Systems

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Quality Analysis of Extra Virgin Olive Oils Part 7

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

Pesticide Analysis using an Agilent 1290 Infinity LC System with an Agilent 6140 Single Quadrupole LC/MS System

USP purity analysis of pravastatin sodium using the Agilent 1120 Compact LC

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Journal of Chemical and Pharmaceutical Research

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Application Note. Author. Abstract. Introduction. Food Safety

Fast determination of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 1260 Infinity Analytical SFC System

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

Impurity Profiling of Carbamazepine by HPLC/UV

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

Edgar Naegele. Abstract

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Steviol Glycosides from Stevia rebaudiana Bertoni

Extended Application Note

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

Two-Dimensional LC/MS/MS to Reduce Ion Suppression in the Determination of Cannabinoids in Blood Plasma

Ensuring High Sensitivity and Consistent Response in UHPLC-MS Analyses

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Sample Concentration and Analysis of Human Hormones in Drinking Water

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Determination of Amantadine Residues in Chicken by LCMS-8040

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Separation of 15 Underivatized Saccharide and Sialic Acid USP Standards

Agilent 1220 Infinity LC System Performance Specifications

Determination of Opiates and Metabolites in Blood Using Electrospray LC/MS. Application Note

Pelagia Research Library

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Application Note. Preparative Isolation of Piperine from Black Pepper Extracts. Summary. Introduction. Fig. 1

A Novel HILIC Column for High Speed N-linked Glycan Analysis

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Facile Isolation of Carotenoid Antioxidants from Solanum lycopersicum using Flash Chromatography

Dienes Derivatization MaxSpec Kit

PROPOSAL FOR REVISION OF MONOGRAPH PUBLISHED IN The International Pharmacopoeia: REVISION OF ph test ABACAVIR ORAL SOLUTION (JULY 2012)

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Author. Introduction. Abstract

Analysis of bisphenol A leaching from baby feeding bottles

Transcription:

Application Note Food Testing and Agriculture Mass-Based Purification of Natural Product Impurities Using an Agilent 126 Infinity II Preparative LC/MSD System Authors Florian Rieck and Jörg Hippler Agilent Technologies, Inc. Waldbronn, Germany Abstract Organic impurities can be more than a minor inconvenience in the making of high quality products such as food additives or dietary supplements. Preparative scale high performance liquid chromatography (HPLC) is frequently applied to separate and isolate unknown by-products for further characterization. This Application Note demonstrates the workflow from analytical scouting through scale-up to preparative conditions and purification of a natural product extract by mass-based fraction collection.

Introduction Monitoring organic impurities is a critical issue in the pharmaceutical industry, as the presence of unwanted chemicals may, even in small amounts, influence the efficacy and safety of pharmaceutical products. According to regulations, for example the ICH Guidance for Industry Q3A, impurities need to be identified and characterized when exceeding a certain threshold 1. However, it is not only the pharmaceutical industry that needs to be aware of organic impurities. To deliver safe and high-quality products, dietary supplements and foodstuffs based on extracts of natural products need to be monitored in a similar manner. The isolation of unknown impurities is crucial for identification and thorough characterization concerning biological activity and toxicity of a compound. Separation and isolation of impurities can be carried out with preparative-scale HPLC, coupled to an ultraviolet (UV) detector and a mass spectrometer (MS). This Application Note describes the isolation of impurities of a natural product extract using an Agilent 126 Infinity II Preparative LC/MSD System. The extract contains a primary compound of interest associated with various known impurities. Routine quality control analyses, however, revealed two unknown impurities that needed characterizing. With an estimated concentration of less than.5 % of that of the main compound, multiple runs with large injection volumes were necessary to collect sufficient amounts of the impurities for further characterization. A mass selective detector (MSD) was used for highly specific detection and fraction triggering of the unknown compounds. Experimental Instrumentation The Agilent 126 Infinity II LC/MSD System used for analytical scouting and fraction reanalyses consisted of the following modules: Agilent 126 Infinity II Binary Pump (G7112B) Agilent 126 Infinity II Vialsampler (G7129A) Agilent 126 Infinity II Diode Array Detector WR (G7115A) Agilent LC/MSD XT (G6135B) Preparative purification was carried out on an Agilent 126 Infinity II Preparative LC/MSD System comprising the following modules: Agilent 126 Infinity II Preparative Binary Pump (G7161A) Agilent 126 Infinity II Preparative Autosampler (G7157A) Agilent 126 Infinity II Column Organizer (G9328A) Agilent 126 Infinity II Variable Wavelength Detector (G7114A), equipped with a 3-mm preparative flow cell (G1314-624) Agilent 126 Infinity II Isocratic Pump (G711B) Agilent 129 Infinity II MS Flow Modulator (G717B) Agilent 126 Infinity II Delay Coil Organizer (G9324A) Agilent 129 Infinity II Preparative Open-Bed Fraction Collector (G7159B) LC/MSD XT (G6135B) Columns Analytical column: Agilent Prep-C18 Scalar 4.6 1 mm, 5 µm (p/n 44995-92) Preparative column: Agilent Prep-C18 21.2 1 mm, 5 µm (p/n 44995-72) Software Agilent OpenLab CDS ChemStation edition for LC and LC/MS Systems, version C.1.8 [21] Solvents All solvents used were LC grade. Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with a.22-µm membrane point-of-use cartridge (Millipak). 2

Fraction collection Peak detection and fraction collection were performed using the UV and MSD signals with the AND trigger combination. Using the 129 Infinity II Preparative Open-Bed Fraction Collector enabled selection of fraction tubes that suited the flow rate and expected fraction volumes of the method. The connection of the preparative solvent flow with the MSD was implemented with the 129 Infinity II MS Flow Modulator. This module of the InfinityLab LC family integrates seamlessly into the Agilent InfinityLab LC Purification Solutions, both from a software and hardware point of view. Split ratio and dilution, as well as start and stop times, are set and saved directly within the LC method. Depending on the make up and main solvent flow rate, split ratios can be set between 1:1 and 5,:1 in 14 separate modes. To avoid overloading of the MSD, the 129 Infinity II MS Flow Modulator timetable was programmed to stop splitting during the solvent front and around the elution point of the primary ingredient (see Table 1: Split ratio to MSD). Method settings Table 1. Chromatographic conditions of analytical and preparative runs. Parameter Analytical runs Preparative runs Mobile phase A).1 % formic acid in water B).1 % formic acid in methanol Flow rate 1 ml/min 2 ml/min Gradient Stop time Post time minutes 3 %B 15 minutes 45 %B 2 minutes 8 %B 2 minutes 5 minutes Injection volume 2 µl 9 µl Temperature Ambient Ambient UV detection MS detection Split ratio to MSD Fraction collection 297 nm Peak width >.25 minutes (.5 seconds response time) 2 Hz data rate Signal 1: positive scan m/z 5 to 1, Signal 2: positive SIM m/z 823.4 Signal 3: positive SIM m/z 839.4 Full flow Not applicable 1,:1 (mode M1) 3. minutes ON 5. minutes OFF 15. minutes ON 18.6 minutes OFF Peak-based: 3. to 4.2 minutes 16.5 to 18. minutes Table 2. MSD spray chamber and fraction collection settings. Parameter Value Make up solvent.1 % Formic acid in methanol:water (7:3) Make up flow 1.5 ml/min Ionization source Nebulizer pressure Agilent Electrospray 35 psig Drying gas temperature 35 C Drying gas flow Capillary voltage 12. L/min +3, V Target mass 822.4; 838.4 Ion species [M+H] + 3

Results and Discussion Analytical scouting The concentrated extract was diluted tenfold with water, and separated on an analytical LC/MSD system. Under these conditions, the main compound was still overloading the column, whereas all impurities were sufficiently resolved (Figure 1). Using the positive scan signal of the MSD, target ions of m/z 823.4 and m/z 839.4 were identified for unknown impurities 1 and 2, respectively. These two target ions were then constantly monitored as signals 2 and 3 of the MSD, which was set to selected ion monitoring (SIM) mode. Preparative run and fraction collection Based on the analytical results, the separation of the concentrated sample was scaled up to preparative conditions using the same column type as the analytical run, but with a larger inside diameter. This linear scale-up allowed a reproduction of all relevant retention times within a margin of.2 minutes. Injecting 9 µl of the concentrated extract per run, the column was overloaded with the main compound; both targeted impurities, however, were well resolved, enabling peak based fraction collection (Figure 2). With a mass-based trigger using the SIM signals on the two target ions, connected with the UV signal by the AND trigger combination, fraction collection could be triggered with high specificity and sensitivity. Fraction review is carried out in the Purify view in ChemStation Data Analysis, which has been improved in revision C.1.8. Collected fractions are now displayed in a comprehensive view, connecting the tick marks in the chromatogram with a graphical representation of the fraction collector bed, and vice versa, by a single mouse click (Figure 3). 1 8 6 % 4 2 mau 5 6 7 8 9 15 1 5 Impurity 1 Impurity 1 3.616 823.4 921.3 m/z 6 % 4 5 6 7 8 9 5 1 15 1 8 2 Main compound Impurity 2 839.4 m/z Impurity 2 16.99 Figure 1. Diluted sample separated on an analytical system. Target ions for both impurities were extracted from the positive scan signal of the MSD. mau 3, 2,5 2, 1,5 1, 5 Impurity 1 Impurity 2 3.415 Main compound 5 1 15 Figure 2. Preparative purification run with 9 µl injection. Both impurities were separated and collected. After successful collection, fractions were reanalyzed on the analytical system. The typical purity of the collected fractions was found to be greater than 98 %. 17.138 min min 4

Figure 3. Purify view in Agilent OpenLab CDS ChemStation edition C.1.8. Click a collected peak in the chromatogram to display its position in the fraction collector, and vice versa. Conclusion This Application Note demonstrates a workflow from analytical scouting to preparative purification using a 126 Infinity II LC/MSD System and a 126 Infinity II Preparative LC/MSD System. A natural product extract with two unknown impurities was analyzed on an analytical system first. Highly loadable Agilent Prep-C18 stationary phases available in matching dimensions facilitated linear scale-up to preparative conditions with reproducible separation of the target compounds. Fraction triggering by UV and MSD enabled highly specific collection and yielded pure fractions suitable for further analyses to characterize the two unknown impurities of the extract. 5

Reference 1. ICH Guideline Q3A(R2): Impurities in New Drug Substances, 26. www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc. 218 Printed in the USA, October 1, 218 5994-222EN