Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Similar documents
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Eric E. Klein, Ph.D. Chair of TG-142

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

EORTC Member Facility Questionnaire

Managing the imaging dose during image-guided radiation therapy

SRS Uncertainty: Linac and CyberKnife Uncertainties

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Leila E. A. Nichol Royal Surrey County Hospital

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Nuclear Associates

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

A quality assurance procedure to evaluate cone-beam CT image center congruence with the radiation isocenter of a linear accelerator

Dosisverifikation mit Delta 4 Discover während der Behandlung

Small field diode dosimetry

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Tolerance levels of CT number to electron density table for photon beam in radiotherapy treatment planning system

Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

A new geometric and mechanical verification device for medical LINACs

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Credentialing for the Use of IGRT in Clinical Trials

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Image Guided Stereotactic Radiotherapy of the Lung

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

IMRT QA: Point Dose Measurements or 2D Array?

Image Registration for Radiation Therapy Applications: Part 2: In-room Volumetric Imaging

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

In-Room Radiographic Imaging for Localization

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

The Journey of Cyberknife Commissioning

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

Quality assurance in external radiotherapy

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Debate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Image quality & dosimetric property of an investigational Imaging Beam Line MV-CBCT

In-Room Radiographic Imaging for Localization

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Work partially supported by VisionRT

Presentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

Collapsed Cone Convolution 2D illustration

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

IMRT Planning Basics AAMD Student Webinar

Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Treatment Efficiency and Optimization of Patient Care with IBA ProteusOne

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Protura Robotic Patient Positioning System. for efficiency + performance

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

Additional Questions for Review 2D & 3D

Case Study. Institution Farrer Park Hospital

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

A method to predict patient specific table coordinates for quality assurance in external beam radiation therapy

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

IGRT Solution for the Living Patient and the Dynamic Treatment Problem

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

Unrivaled, End-to-End

Transcription:

Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Yuting Li 1,2 Tucker Netherton 1,3 Paige L. Nitsch 1 Peter A. Balter 1 Song Gao 1 Ann H. Klopp 1 Laurence E. Court 1 1 Department of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, TX, USA 2 Department of Radiation Oncology, Ohio State University Medical Center, Columbus, OH, USA 3 The University of Texas Graduate School of Biomedical Sciences at Houston, Houston, TX, USA Author to whom correspondence should be addressed. Laurence Court E-mail: lecourt@mdanderson.org; Telephone no: +1 (713) 563 2546 Abstract Purpose: The aim of this study was to measure and compare the mega-voltage imaging dose from the Halcyon medical linear accelerator (Varian Medical Systems) with measured imaging doses with the dose calculated by Eclipse treatment planning system. Methods: An anthropomorphic thorax phantom was imaged using all imaging techniques available with the Halcyon linac MV cone-beam computed tomography (MV-CBCT) and orthogonal anterior-posterior/lateral pairs (MV-MV), both with high-quality and low-dose modes. In total, 54 imaging technique, isocenter position, and field size combinations were evaluated. The imaging doses delivered to 11 points in the phantom (in-target and extra-target) were measured using an ion chamber, and compared with the imaging doses calculated using Eclipse. Results: For high-quality MV-MV mode, the mean extra-target doses delivered to the heart, left lung, right lung and spine were 1.18, 1.64, 0.80, and 1.11 cgy per fraction, respectively. The corresponding mean in-target doses were 3.36, 3.72, 2.61, and 2.69 cgy per fraction, respectively. For MV-MV technique, the extra-target imaging dose had greater variation and dependency on imaging field size than did the in-target dose. Compared to MV-MV technique, the imaging dose from MV- CBCT was less sensitive to the location of the organ relative to the treatment field. For high-quality MV-CBCT mode, the mean imaging doses to the heart, left lung, right lung, and spine were 8.45, 7.16, 7.19, and 6.51 cgy per fraction, respectively. For both MV-MV and MV-CBCT techniques, the low-dose mode resulted in an imaging dose about half of that in high-quality mode. Conclusion: The in-target doses due to MV imaging using the Halcyon ranged from 0.59 to 9.75 cgy, depending on the choice of imaging technique. Extra-target doses from MV-MV technique ranged from 0 to 2.54 cgy. The MV imaging dose was accurately calculated by Eclipse, with maximum differences less than 0.5% of a typical treatment dose (assuming a 60 Gy prescription). Therefore, the cumulative imaging and treatment plan dose distribution can be expected to accurately reflect the actual dose. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------- This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine 52 wileyonlinelibrary.com/journal/jacmp J Appl Clin Med Phys 2018; 19:3:52 57

LI ET AL. 53 PACS 87.55.Qr KEY WORDS halcyon, MV-IGRT, normal tissue imaging dose 1 INTRODUCTION On-line mega-voltage (MV) imaging has been successfully used for daily image-guided patient setup for many years, including orthogonal imaging, helical computed tomography (CT), and cone-beam CT (CBCT). 1 5 Since the early 2000s, however, there has been a significant increase in the use of kilo-voltage imaging for in-room patient setup, including orthogonal imaging, helical CT (CT-on-rails), and CBCT. 6 Although kilo-voltage imaging offers the advantages over MV imaging of lower patient dose and better soft-tissue contrast, MV imaging has the potential advantages of reduced equipment costs and complexity, which can be expected to transfer to gains in reliability. Varian Medical Systems (Palo Alto, CA) recently released Halcyon, a medical linear accelerator which uses daily MV imaging for patient setup. 7 Since there s a concern that MV-IGRT would introduce higher normal tissue dose compared with kv IGRT, 8,9 in this present study we performed extensive measurements of the imaging dose for all available imaging techniques on the Halcyon linac for tissues within and outside the treatment volume. We then compared the measured doses with those calculated by Eclipse, to verify that the Eclipse beam model, which is pre-commissioned for Halcyon, could correctly predict the MV imaging dose and incorporate it into the treatment plan. This report adds practical data specific to the Halcyon, beyond what has been previously reported for other systems. 9 11 imaging on this machine, and the daily imaging dose is incorporated into the treatment plan (in the Eclipse treatment planning system). The maximum field size (28 9 28 cm 2 ) at the isocenter (1 m) is defined by a dual-layer multileaf collimator (MLC) system without physical jaws or light field. Using the internal lasers, the patient is aligned to a virtual isocenter and then shifted into the bore. Subsequent treatment relies entirely on MV-image guidance. The imaging dose is calculated by the treatment planning system (TPS) and incorporated into the final treatment plan. Two imaging modalities are available on the Halcyon system: orthogonal anterior-posterior/lateral pairs (MV-MV) and MV-CBCT, each with Low-Dose and High-Quality modes. MV-MV has fixed gantry angles of 0 and 90 (Fig. 1) with total of two monitor units (MU) for low-dose mode and 4 MU for high-quality mode. The collimator angle is fixed at 0 during imaging. The imaging field size can be adjusted by changing leaf separation along the y-axis to any number less than or equal to 28 cm and separation along the x-axis to any even integer less or equal to 28 cm. MV-CBCT images are acquired via a continuous gantry rotation from 260 to 100, with total of 5 MU for low-dose mode and 10 MU for high-quality mode. The collimator angle is fixed at 0 during imaging. For MV-CBCT, the axial field-of-view (FOV) could not 2 MATERIALS AND METHOD 2.A Phantom selection We used a heterogeneous anthropomorphic thorax phantom (0002LFC; CIRS, Norfolk, VA) with a breast phantom attached to the right chest for both ionization chamber measurements and TPS calculation of the imaging dose to normal tissue structures. This phantom contains ion chamber inserts within tissue equivalent media for the lungs, spine, mediastinum, and breast. The phantom was scanned using CT and the CT images were imported into the Eclipse TPS. 2.B Treatment device and available imaging modalities The Halcyon has an enclosed, ring-mounted gantry with an opposing electronic portal imaging device (EPID) and beam stopper. A 6 MV flattening filter free (FFF) beam is used for both treatment and F IG. 1. Anthropomorphic phantom (CIRS) with dose measurement points and isocenter locations identified. Point 11 is offset by 5.5 cm longitudinally from points 1 10. In this study, isocenter was located at point 1, 2, 6, 8, 10 and 12, which is in the center of the phantom with an 8 cm offset from the axial plane of points 1 10. The target region of MV-MV is illustrated as dark blue (intersection) region. The remaining part is defined as extra-target region.

54 LI ET AL. T ABLE 1 Isocenter locations and field sizes where normal tissue imaging doses were measured using Halcyon and in Eclipse. The location numbers are referred to the points identified in Fig. 1. Measurements Field size (cm 2 ) Isocenter location MV-MV MV-CBCT 1 2 9 2, 6 9 6, 10 9 10, 14 9 14, 20 9 20, 28 9 28 2 9 28, 6 9 28, 10 9 28, 14 9 28, 20 9 28, 28 9 28 1, 2, 6, 8, 10, 12 28 9 28 28 9 28 be adjusted because it is fixed at 28 cm on Halcyon, but the FOV longitudinal length can be changed from 2 cm to 28 cm in 2 cm increments. 2.C Imaging dose calculations Using MV-MV and MV-CBCT in both low-dose and high-quality modes, we created 54 treatment plans in Eclipse (v15.5) with different imaging field sizes (Table 1) and isocenter locations (Fig. 1). The Fourier Transform Dose Calculation (FTDC) Algorithm is used for calculating imaging dose and is built within the Analytic Anisotropic Algorithm (AAA) dose calculation algorithm. 11 The FTDC uses a convolution/superposition algorithm, that is, optimized for speed by simplifying AAA s three-source dose model and uses a 5.0 mm calculation resolution. This dose calculation algorithm is pre-loaded into the Eclipse planning system, and no changes by the user are possible. Imaging doses delivered to 11 points in the heart, lungs, spine, and breast were calculated in Eclipse and later compared with our corresponding measurements described below. 2.D Imaging dose measurements The phantom was placed in the treatment position, images were taken, and imaging doses were measured using a small volume ion chamber (CC04; IBA Dosimetry, Bartlett, TN) after cross-calibration with a Farmer chamber (with Accredited Dosimetry Calibration Laboratory calibration). Imaging dose measurements were taken at 11 points (Fig. 1): Points 1 10 were in a single axial plane, and point 11 was offset in the longitudinal direction by 5.5 cm. All imaging was performed using the Halcyon system in clinical mode. Each measurement was repeated three times, and the average calculated. To examine the effect of collimation during imaging upon normal tissue imaging doses, the isocenter was placed in the center of the phantom (point 1 in Fig. 1), and the imaging doses from various imaging field sizes were measured. To evaluate the effect of isocenter location on imaging dose, we used the maximum field size (28 9 28 cm 2 ) and placed the isocenter at points 2, 6, 8, 10, and 12 (Table 1), where point 12 is in the center of the phantom with an F IG. 2. Normal tissue imaging dose of the extra-target and target regions of MV- MV and of MV-CBCT techniques. Measured data of all measured field sizes and isocenter locations were included in the figure.

LI ET AL. 55 8 cm offset from the axial plane of points 1 10. For MV-MV technique, imaging dose data were designated as target-region if the point being evaluated was within the intersection between the AP and lateral imaging beams, shown as the dark blue region in Fig. 1. Data measured on all other points outside of this region were labeled extra-target region. For MV-CBCT, we measured the imaging dose with various field sizes: 2 9 28, 6 9 28, 10 9 28, 14 9 28, 20 9 28 and 28 9 28 cm 2 with the isocenter at point 1. We also examined the imaging dose with the isocenter at the same six isocenter locations used in the MV-MV study with the MV-CBCT field size fixed at its maximum (Table 1). 2.E Breast imaging dose Since the arrangement of the orthogonal imaging pair (MV-MV) can only be at 0 and 90, we studied the imaging doses to the left and right breasts. Because of the Halcyon linac s bore dimension (1 m diameter), for large patients, placing the isocenter in the treated breast or chestwall was impossible. Thus, for these patients, the isocenter was placed in the ipsilateral lung to avoid collision with the bore. To simulate all possible isocenter locations for breast and chestwall treatments using the Halcyon, we evaluated the imaging doses of ipsilateral and contralateral breasts with the isocenter located in the target breast or ipsilateral lung at the maximum field F IG. 3. Variation in normal tissue imaging dose depending on isocenter location. HQ, high-quality; LD, low-dose.

56 LI ET AL. T ABLE 2 Measured imaging doses per fraction to the breast with daily MV-MV and MV-CBCT imaging using the maximum field size of 28 9 28 cm 2. Breast of interest size (28 9 28 cm 2 ) for both MV-MV and MV-CBCT modalities. All the isocenter locations for breast imaging dose measurements were in the same axial plane. 3 RESULTS Relative position to treatment site 3.A Normal tissue dose MV-MV (cgy) MV-CBCT (cgy) Lowdose Highquality Lowdose Highquality Left Ipsilateral 2.15 4.32 4.82 9.71 Right 1.71 3.40 4.76 9.64 Left Contralateral 1.35 2.76 2.85 5.77 Right 0.41 0.83 2.83 5.72 Fig. 2 shows the imaging doses measured at points in the heart, left lung, right lung, and spine, averaged for all measured field sizes and isocenter locations. For MV-MV, we categorized the tissue doses according to their locations relative to the target volume extratarget and target regions, as shown in Fig. 2. We found that for high-quality MV-MV mode, the mean extra-target doses of heart, left lung, right lung and spine were 1.184, 1.644, 0,802 and 1.105 cgy per fraction, respectively. The corresponding in-target doses were 3.358, 3.719, 2.612 and 2.686 cgy per fraction, respectively. Moreover, the imaging doses in the extra-target regions from MV-MV technique exhibited higher variation and dependency on imaging field size than did those in the in-target regions. The imaging dose from MV-CBCT was less sensitive to the location of the organ relative to the treatment field than MV-MV technique. For highquality mode, the mean imaging doses delivered to the heart, left lung, right lung and spine were 8.448, 7.158, 7.190, and 6.512 cgy per fraction, respectively. For both MV-MV and MV-CBCT techniques, imaging doses to the lung were more sensitive to changes of isocenter location than those of other organs, with a maximum variation of 3.12 cgy in the left lung for high-quality MV-CBCT (Fig. 3). In both MV-MV and MV-CBCT cases, use of the low-dose mode resulted in an imaging dose about half of that using the high-quality mode. 3.B Breast dose Table 2 shows the measured breast imaging doses using MV-MV and MV-CBCT in both low-dose and high-quality modes. Each data point is an average of measured values at two isocenter locations (in target breast and in ipsilateral lung). The first column in the Table 2 indicates the breast of interest where imaging dose measurements were taken, whereas the second column describes the position of the breast of interest relative to the actual treatment site. For MV- MV imaging, the breast dose depended on both the treatment location and laterality. The doses for the high-quality mode ranged from 0.8 cgy to 4.3 cgy. Because the lateral image is always taken with the gantry at 90, the left breast is closest to the x ray source. For this reason, the lowest imaging dose was to the right contralateral breast when the left breast was the target. The highest imaging dose was to the left ipsilateral breast when the left breast was the target. In the cases with same side laterality, the left breast always received higher doses than did then right breast, because the orthogonal imaging fields for Halcyon are fixed at 0 and 90. For high-quality mode, when the breast of interest was the one being treated, the imaging dose to the left breast was about 0.9 cgy higher than the right breast. When the breast of interest was contralateral to the treated site, the left breast received a 1.9 cgy higher imaging dose than the right breast would. The imaging doses received by the breasts from CBCT were less sensitive to the treatment site location than in MV-MV and were only a function of target laterality. In all cases, the imaging doses administered in low-dose mode were approximately half of those from the high-quality mode. 3.C Comparison of measurements and TPS calculations The average differences between the Eclipse calculated and measured imaging doses had were 0.05, 0.35, 0.35, 0.09, and 0.01 cgy for the heart, left lung, right lung, spine, and breast, respectively. As shown in Fig. 4, the TPS tended to overestimate the imaging dose in nearly all cases, with the largest disagreement being in the lung (0.99 cgy). F IG. 4. Comparison of measured and eclipse calculated normal tissue imaging doses.

LI ET AL. 57 4 DISCUSSION MV imaging is fully integrated into the patient treatment workflow for the Halcyon medical linear accelerator. Therefore, it is crucial to understand the (a) impact of imaging doses to normal tissue, and (b) accuracy of MV imaging dose calculation in the treatment planning system. Therefore, we investigated this for all of the imaging techniques offered by Halcyon: MV-MV and MV-CBCT, both with highquality and low-dose mode. The pre-commissioned Eclipse paired with Halcyon exhibited good accuracy in calculating the imaging dose. Based on our imaging dose measurements, for treatment with 30 fractions, the maximum difference between the calculated and actual imaging dose would be about 30 cgy, or approximately 0.5% of the target dose (assuming a 60 Gy prescription). Accurately including the imaging dose in the treatment plan is important for several reasons. It means that it is possible to perform sensible comparisons of plans treated with different treatment devices and different imaging approaches (e.g., comparison with devices with much lower imaging dose). It also means that clinical experience in terms of normal tissue dose constraints can be directly applied to the Halcyon treatment plans. Although other researchers have evaluated MV imaging doses for other systems, we are the first to report imaging dose measurements with the Halcyon treatment delivery system. This information will be useful to users of this new device when introducing it into clinical use. 5 CONCLUSIONS The in-target doses due to MV imaging using the Halcyon ranged from 0.59 to 9.75 cgy, depending on the choice of imaging technique. Extra-target doses from MV-MV technique ranged from 0 to 2.54 cgy. The MV imaging dose was accurately calculated by Eclipse, with maximum differences less than 0.5% of a typical treatment dose (assuming a 60 Gy prescription). Therefore, the cumulative imaging and treatment plan dose distribution can be expected to accurately reflect the actual dose. CONFLICT OF INTEREST This work was partially funded by Varian Medical Systems. REFERENCES 1. Flynn RT, Hartmann J, Bani-Hashemi A, et al. Dosimetric characterization and application of an imaging beam line with a carbon electron target for megavoltage cone beam computed tomography. Med Phys. 2009;36:2181 2192. 2. Court LE, Allen A, Tishler R. Evaluation of the precision of portalimage-guided head-and-neck localization: an intra- and interobserver study. Med Phys. 2007;34:2704 2707. 3. Bissonnette JP, Balter PA, Dong L, et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179. Med Phys. 2012;39:1946 1963. 4. Goyal S, Kataria T. Image guidance in radiation therapy: techniques and applications. Radiol Res Pract. 2014;2014:10. 5. Herman MG. Clinical use of electronic portal imaging. Semin Radiat Oncol. 2005;15:157 167. 6. Yin F-F, Wong J, Balter J, et al. The role of in-room kv X-ray imaging for patient setup and target localization. Report of AAPM Task Group; 2009;104. 7. Halcyon System. (Accessed 6/18/2017, at https://www.varian.com/ oncology/products/treatment-delivery/halcyon.) 8. Murphy MJ, Balter J, Balter S, et al. The management of imaging dose during image-guided radiotherapy: Report of the AAPM Task Group 75. Med Phys. 2007;34:4041 4063. 9. Herman MG, Balter JM, Jaffray DA, et al. Clinical use of electronic portal imaging: Report of AAPM radiation therapy committee Task Group 58. Med Phys. 2001;28:712 737. 10. Morin O, Gillis A, Descovich M, et al. Patient dose considerations for routine megavoltage cone-beam CT imaging. Med Phys. 2007;34:1819 1827. 11. RDS Treatment Planning Instructions for Use Eclipse 15.1.1. Report No.: P1017375-004-D.