Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Similar documents
Baotian Huang 1, Lili Wu 1, Peixian Lin 2 and Chuangzhen Chen 1*

Dosimetric Comparison: Acuros XB vs AAA

The performance of the progressive resolution optimizer (PRO) for RapidArc planning in targets with low-density media

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Comparison of Anisotropic Analytic Algorithm Plan and Acuros XB Plan for Lung Stereotactic Ablative Radiotherapy Using Flattening Filter-Free Beams

Dosimetric evaluation of the Acuros XB algorithm for a 4 MV photon beam in head and neck intensity-modulated radiation therapy

Accuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams *

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Leila E. A. Nichol Royal Surrey County Hospital

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Diode calibration for dose determination in total body irradiation

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Radiobiological Impact of Planning Techniques for Prostate Cancer in Terms of Tumor Control Probability and Normal Tissue Complication Probability

The effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Quality assurance in external radiotherapy

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

I. Equipments for external beam radiotherapy

Intensity Modulated RadioTherapy

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Abdulhamid Chaikh 1,2, Jacques Balosso 1,2. Introduction

Inhomogeneity correction and the analytic anisotropic algorithm

Limits of Precision and Accuracy of Radiation Delivery Systems

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

A comparative dosimetric analysis of the effect of heterogeneity corrections used in three treatment planning algorithms

Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Tolerance levels of CT number to electron density table for photon beam in radiotherapy treatment planning system

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

Future upcoming technologies and what audit needs to address

Eric E. Klein, Ph.D. Chair of TG-142

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Additional Questions for Review 2D & 3D

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Heterogeneity Corrections in Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Commission, Ganakbari, Savar, Dhaka, Bangladesh

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

The bootstrap method to improve statistical analysis of dosimetric data for radiotherapy outcomes

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

Protection of the contralateral breast during radiation therapy for breast cancer

Activity report from JCOG physics group

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Development of a novel low-radiation-absorbent lok-bar to reduce X-ray scattering and absorption in RapidArc â treatment planning and dose delivery

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

Hampton University Proton Therapy Institute

EORTC Member Facility Questionnaire

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Taeryool Koo 1, Jin-Beom Chung 2*, Keun-Yong Eom 1,2, Jin-Yong Seok 2, In-Ah Kim 1,2 and Jae-Sung Kim 1,2

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Comparative evaluation of modern dosimetry techniques near low- and high-density heterogeneities

Image Guided Stereotactic Radiotherapy of the Lung

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Small field diode dosimetry

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Transcription:

Original Article Research in Oncology 2018; Vol. 14, No. 1: 34-38. DOI: 10.21608/resoncol.2018.3026.1047 Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System Nader Elsherbini 1, Wafaa Ahmed 2, Sherifa O. Elwakeel 1 1 Medical Physics Unit, Kasr Al-Ainy Center of Clinical Oncology and Nuclear Medicine, Kasr Al- Ainy School of Medicine, Cairo University, Cairo, Egypt ; 2 Department of Biophysics, Faculty of Science, Cairo University, Cairo, Egypt. Background: Verification of dose calculation algorithms is important in the radiotherapy process. Aim: The aim of this work is to verify the dose calculation accuracy of Acuros XB version 10 algorithm (AXB10) in homogenous and heterogeneous media. Methods: We compared AXB10 calculated doses with the measured doses using thermo-luminescent dosimeters (TLD) for 6 MV photon beam. Two clinical cases, nasopharynx and lung tumors, were studied to evaluate the ability of the AXB10 to deal with tissue heterogeneity. Selected field sizes were measured in water phantom then compared with that calculated with the Eclipse treatment planning system (TPS) - based on AXB10 algorithm. Rando humanoid phantom was computerized tomography scanned and the images were transferred to the TPS system where a set of similar plans of a single direct field were calculated with AXB10, then the dose was measured within the phantom using the TLD in the brain and lung regions. Lastly the measured and calculated data were compared. Results: There was a good agreement between the measured fields in the water phantom and that calculated with the AXB10 (± 2%). The percentage difference between full Monte Carlo algorithm and AXB10 were 3% for the phantoms with lung. Conclusion: Acuros XB algorithm (version 10) results are in agreement with the International Commission for Units and Measurements recommendations. Keywords: Acuros XB algorithm, Treatment planning system, Thermo-luminescent dosimeters Corresponding Author: Sherifa Elwakeel, Medical Physics Unit, Kasr Al-Ainy Center of Clinical Oncology and Nuclear Medicine, Kasr Al-Ainy School of Medicine, Cairo University, Cairo, Egypt, E-mail: sherifa_osman93@yahoo.com Submitted: 27-February-2018, Revised: 26-March-2018, Accepted: 28-March-2018, Published online: 8-April-2018 INTRODUCTION Radiotherapy is a multi-step process starting with diagnosis and disease staging and ending with the treatment of specific target volume with predetermined radiation energies and beam parameters. The accuracy of the dose calculation and the dose delivery to the target has been increased especially in the last decades with the use of computed tomography (CT) which allowed more anatomical details and material information 1. There are many steps to optimize the process of radiation therapy to deliver the given dose to the target in high accuracy 2. The presence of heterogeneities causes dose calculation uncertainties which should be overcome to maximize the therapeutic benefit of radiation therapy 2. Acuros XB version 10 (AXB10) algorithm is based on Linear Boltzmann Transport Equation (LBTE) which is the linearized form of Boltzmann Transport Equation (BTE), the governing equation which describes the behavior of macroscopic particles (protons, neutrons, electrons, etc.) as they pass through and interact with the matter. Linear Boltzmann Transport Equation considers that the particle interacts only with the matter they are passing through not with the particles each other without the conditions of external magnetic field 1. Eclipse treatment planning system (TPS) is loaded by the AXB10 by Varian Medical Systems, Inc. (VMS) (Palo Alto, CA, USA). The origin of the AXB algorithm found in the Attila LBTE solver was developed by a group based on the founders of the Transpire, Inc. (Gig Harbor, WA, USA) 2. The accuracy of AXB11 was tested for 6 and 18 MV photon in a cubic water phantom which composed of stainless steel and titanium alloy. This resulted in a significant percentage of Gamma Analysis Index values of 96.8% and 91.3% for the comparison between the full Monte Carlo (MC) simulation and the AXB11 with acceptance criteria of 2% or 1 mm 6. Acuros XB version 10 and Analytical Anisotropic Algorithm (AAA) were tested for their accuracy in phantoms which composed of different percentages of solid water and air, for different square and rectangular field sizes (3 3, 5 5, 10 10 and 30 30 cm 2 ) for 6 MV photon beam. Phantom A composed of 2-cm air gap, phantom B composed of 4-cm air gap, and phantom C composed of 6-cm air gap. The three phantoms were imaged by computed tomography (CT) and transferred to the Eclipse Planning System via DICOM for calculation by AXB10 then recalculated by AAA. The deviation between AXB10 and measured data was from 34

3.81% to +0.9%, while the deviation between AAA and measured data was from 3.1% to 10.9% 7. In a study that included lung cancer cases presenting for radiotherapy, the planning target volume (PTV) was calculated by AXB11 and recalculated by AAA. The mean deviation was less than 2% for the PTV and less than 3% for the organs at risk 9. While the cases treated by stereotactic body radiation therapy technique needed more monitor units (MU) due to the presence of heterogeneities 10. In breast cancer cases receiving conformal treatment in the form of two tangential fields of 6 MV, AXB predicted lower doses to the PTV and to the lung compared to AAA. So the AXB may be needed for improving low density materials calculations 11. The aim of this study was to test the accuracy of AXB10 in heterogeneous (nasopharynx and lung) and homogenous (water phantom) regions. METHODS Materials Linear Accelerator: The unique Linac (UNIQUE TM, Varian Palo Alto, California, USA) was used in this study with field size ranging from 4x4 to 40x40 cm 2 at a 100cm Source to Skin Distance (SSD). The linear accelerator was calibrated so that a square symmetric field (10x10 cm 2 ) will deliver an absorbed dose of 1Gy per 100MU at 10 cm depth in water with an SSD of 90 cm. Gantry Rotation angle from the vertical was ±185. Mechanical and Radiation Isocenter Accuracy was < 0.1 cm radius sphere. For arc technique, the gantry was not to deviate >7 degrees from the desired angle which was considered as 0.2 MU from the total dose. Treatment planning system (TPS): The Eclipse TM Treatment Planning Software (version 10.0) with AXB10 and AAA with the same version used in this study, is an integrated and comprehensive system supporting radiation treatment (e.g. photons, flattening filter free beams, protons, electrons, external beams, low-dose-rate brachytherapy and cobalt therapy). Water Phantom: PTW (Freiburg, MP1) water tank is remote controlled and has dimensions of 254 mm x320 mm x370 mm. It is enhanced by holding devices and TBA electronics and MEPHYSTO mc 2 software for software operations. Humanoid Rando: Alderson RANDO phantom (ART) (Radiology Support Devices Inc., Long Beach, CA) was used in this study. The phantom is 175cm tall and 73.5kg weight. It consists of 34 slices each of 2.5 cm thickness. Each slice is perforated for the insertion of thermo-luminescent dosimeter (TLD) chips. The phantom is made of tissue-equivalent material according to International Commission on Radiation Units and Measurements (ICRU) - 44 recommendations. The skeleton is composed of polymer moldings to simulate the shape, mass, density and the lung is molded from synthetic foam. Computed Tomography (CT): The multislice CT machine used in this work was General Electric (UK); light speed RT. This CT machine can record 4 CT 35 images per rotation and the CT unit is provided with a flat carbon fiber table to simulate the Linac couch. Thermo-Luminescent Dosimeter (TLD): System LiF (700) chips were used in this study for measuring dose in the humanoid. The chip material was LiF enriched with Mg and Ti impurities (each chip is 0.45 cm [diameter] x 0.08 cm [thickness]) with an effective atomic number of 7.4. The system also included a Rados TLD reader unit which was used to read the dose measured by TLD chips. Procedures Verification of AXB10 depth dose and beam profile measurements: Phantom of 40 x 40 x 40 cm 3 was created as a 3D structure on Eclipse planning system, which was defined as body structure (CT value=0) and used for computation of percentage depth doses (PDDs) and beam profiles for calculated field sizes for AXB10. This was then compared to the measured data with the water phantom at SSD=100cm for 6 MV photon beam. The PDDs were normalized to 100%. Beam profiles were measured and calculated at depths of 1.5, 5 and 15cm for the 6 MV photon beam. TLD (LiF 700) preparation procedures: A new batch of LiF (700) chips was used. The chips were annealed before use by heating up to 400ºC for one hour and for two hours at 100ºC then left in the oven to cool to room temperature. All chips were then irradiated to 1 mgy using RADOS irradiator. Then sensitivity factor for each TLD (LiF 700) chip was measured by dividing the TLD (LiF 700) reader count by the given dose. The average factor was calculated for all chips and considered as 100%. All chips with sensitivity factors exceeding ±2% were omitted. The absorbed doses were measured by TLD in two CT cuts in RANDO phantom in nasopharnyx and lung regions to test AXB10 in the heterogeneity regions in soft tissue, air bone interface and air-soft tissue interface. Evaluation of AXB10: One of the error sources in dose calculations arises from the ability of TPS to deal with tissue heterogeneity. Humanoid RANDO phantom was CT scanned with slice spacing of 2.5 mm. These series were transferred to the TPS through DICOM network. Organs at risk were delineated in the phantom; including right lung, left lung, heart and spinal cord in the chest region and right optic nerve, left optic nerve, brain stem and optic chiasm in the brain region. A plan of single isocentric direct field size of 20x20 cm 2 was created and doses at certain points (representing the positions of TLD chips) were calculated and recorded. Plan isocenter position was verified with the simulation step in which the anterior and lateral Digital Reconstructed Radiography views were compared to the corresponding 2D simulator images. The TLD chips were inserted in their predefined positions in which two crystals were inserted in each position and irradiated with the previously mentioned field.then crystals were collected and the dose was measured by the TLD reader. The previous step was repeated 3 times for accuracy. Dose assessment: 1) Two plans were created in the TPS for nasopharynx and lung cases and then

calculated by AXB10, 2) The measured and calculated doses were compared. For TLD measurements the percentage difference between measured and calculated doses was calculated as follows: % Dose difference = 100 x ((Dose measured - Dose calculated) / Dose measured) RESULTS Measurements of tissue heterogeneity In the lung case, the differences between measured and calculated dose in the heterogeneous regions ranged from -0.7 to 0.4 (table 1). In the nasopharynx case the differences ranged from 0.04 to 0.1 (table 2). Comparison between standard measurements The PDDs of field sizes 5x5,10x10,15x15 and 25x25 cm 2 were compared between the measured PDDs from the water phantom and the calculated PDDs from the TPS loaded by AXB10 and showed standard deviations of 0.3%,0.2%,0.1% and 0.1% respectively (figure 1). The standard deviations for the dose profiles of field sizes 5x5, 10x10, 15x15 and 25x25 cm 2 between measured and calculated doses in the three depths 1.5, 5, and 15 cm were in the range of 0.1-0.3, 0.2-0.3 and 0.1-0.4 respectively (figure 2). Table 1. The TLD measured doses and the Eclipse TPS calculated relative doses in the lung region Energy Point No. Region TLD Eclipse (Acuros XB 10 Algorithm) Diff.(%) Met criteria 6MV 656 Bone air interface 117.2 118-0.7 Yes 6MV 655 Air soft tissue interface 192.5 193.1-0.6 Yes 6MV 657 Soft lung 157.4 157 0.4 Yes TLD: Thermo-luminescent dosimeters Table 2. The TLD measured doses and the Eclipse TPS calculated relative doses in the Nasopharynx region Energy Point No. Region TLD Eclipse (Acuros XB 10 Algorithm) Diff.(%) Met criteria 6MV 798 Soft tissue 1.3 1.4 0.1 Yes 6MV 797 Bone region 0.99 1.1 0.2 Yes 6MV 796 Tissue-Air interface 1.76 1.8 0.04 Yes TLD: Thermo-luminescent dosimeters Figure 1. Measured and calculated percentage depth dose for field sizes: a) 5x5 cm 2 at SSD = 100 cm, b) 10x10 cm 2 at SSD = 100 cm, c) 15x15 cm 2 at SSD =100 cm and d) 25x25 cm 2 at SSD =100 cm 36

Figure 2. Measured and calculated profiles for field sizes: a) 5x5cm 2 at SSD =100 cm, b) 10x10cm 2 at SSD =100 cm, c) 15x15cm 2 at SSD =100 cm and d) 25x25cm 2 at SSD =100 cm DISCUSSION Tissue heterogeneity could be a source of error which leads to errors in radiotherapy dose delivery resulting from TPS. This highlights the importance of TLD testing to verify the accuracy of algorithms. The American Association of Physicists in Medicine (AAPM) recommended the proportion of each source of uncertainties, determined for each step in the radiotherapy process for achieving the accuracy, to be in the range from 1.0% to 5.0%. This was decreased below 3.0% by the Report 85 of Task Group 65 of AAPM 3. Our results for homogeneity are consistent with these recommendations. Also, ICRU report no. 85 has recommended that the accuracy of the absorbed dose to be 5% and the AAPM has recommended that the uncertainty in computed dose distribution to be less than 2% 4. According to ICRU report no. 83, the deviation on the delivered dose to the PTV should be within +3 % or -3% 5.The results of the current study for heterogeneity are congruent with these recommendations. Our results are also in good agreement with other studies. Bush et al reported that the percentage difference between the AXB10 and full MC simulations was within 3% for the phantoms with lung inserted 11. Also, Fogliata et al tested AXB10 and AXB11 in water phantom for square field sizes ranging from 2 2 to 40 40 cm 2 using flattening filter free 6 MV beam. They found that the percentage difference was within 1% for open beams 12. Conclusion In this study we tested the accuracy of AXB10 in water phantom with the same calculations from TPS which gave a good agreement. The results of the work showed that AXB10 made an improvement in beam modeling in homogenous medium (water phantom) measurement for both PDDs and beam profiles measurement. Also measurements in heterogeneous medium (RANDO phantom) showed good results between measured doses in the lung region by both TPS loaded by AXB10 and doses delivered by LINAC. Acknowledgement The authors would like to express their gratitude to Physicist. Mohamed Sayed, MSc (Kasr Al-Ainy Center of Clinical Oncology and Nuclear Medicine, Cairo University) for his valuable support. Conflict of interest None to declare REFERENCES 1. Ojala J. The accuracy of the Acuros XB algorithm in external beam radiotherapy a comprehensive review. Int J Cancer Ther Oncol. 2014; 2(4): 020417. 37

2. Vassiliev ON, Wareing TA, Davis IM, et al. Feasibility of a multigroup deterministic solution method for threedimensional radiotherapy dose calculations. Int J Radiat Oncol Biol Phys. 2008; 72(1):220-227. 3. Papanikolaou N, Battista JJ, Boyer AL, et al. Report of Task Group No. 65 of the Radiation Therapy Committee of the American Association of Physicists in Medicine: Tissue inhomogeneity corrections for megavoltage photon beams. Madison, WI: Medical Physics Publishing, 2004. 4. International Commission on Radiation Units and Measurements. Report 85: Fundamental quantities and units for ionizing radiation. J ICRU. 2011; 11(1): 1-31. 5. International Commission on Radiation Units and Measurements. Report 83: Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT). J ICRU. 2010; 10(1): 1-106. 6. Lloyd SA, Ansbacher W. Evaluation of an analytic linear Boltzmann transport equation solver for high-density inhomogeneities. Med Phys. 2013; 40(1): 011707. 7. Liu HW, Nugent Z, Clayton R, Dunscombe P, Lau H, Khan R. Clinical impact of using the deterministic patient dose calculation algorithm Acuros XB for lung stereotactic body radiation therapy, Acta Oncol. 2014; 53: 324 329. 8. Rana S, Rogers K, Pokharel S, Cheng C. Evaluation of Acuros XB algorithm based on RTOG 0813 dosimetric criteria for SBRT lung treatment with Rapid Arc. J Appl Clin Med Phys. 2014; 15(1): 4474. 9. Rana S, Rogers K, Lee T, Reed D, Biggs C. Verification and dosimetric impact of Acuros XB algorithm for stereotactic body radiation therapy (SBRT) and RapidArc planning for non-small-cell lung cancer (NSCLC) patients. Int J Med Phys Clin Eng Radiat Oncol. 2013; 2: 6-14. 10. Warren S, Panettieri V, Panakis N, et al. Optimizing collimator margins for isotoxically dose-escalated conformal radiation therapy of non-small cell lung cancer. Int J Radiat Oncol Biol Phys. 2014; 88(5): 1148-1153. 11. Bush K, Gagne IM, Zavgorodni S, Ansbacher W, Beckham W. Dosimetric validation of Acuros XB with Monte Carlo methods for photon dose calculations. Med Phys. 2011; 38(4): 2208-2221. 12. Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. Accuracy of Acuros XB and AAA dose calculation for small fields with reference to RapidArc( ) stereotactic treatments. Med Phys. 2011; 38(11): 6228-6237. 38