STANDARD OPERATING PROCEDURE (SOP)

Similar documents
MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH

Procedure for exceptional magnetic resonance scanning in patients with standard (non-mr-conditional) pacemakers

Micra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR)

Advisa MRI and Ensura MRI SureScan pacing systems

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

MRI imaging for patients with cardiac implantable electronic devices (CIEDs)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

Quality Standards for the Implantation of Cardiac Rhythm Management Devices. Pan- London Arrhythmia Project Group. Version 3 (18 th July 2011)

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS

Cardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

MRI SureScan pacemakers & ICDs what you need to know

IMAGEREADY MR CONDITIONAL PACING SYSTEM

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

Call Medtronic at 1 (800) to verify the patient s current implanted system

Peri-operative management of pacemakers and implantable cardiac defibrillators

MRI and CIEDs. Disclosures

Supplemental Material

MRI SURESCAN SYSTEMS. Patient Scanning Process PATIENT PRESCREENING. SureScan Pacing, Defibrillation, and CRT-D Systems Verification

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System.

Title Deactivation of Implantable Cardioverter Defibrillators (ICD) towards the end of life Guidelines

Version überholt. Nicht verwenden. Version obsolète. Ne pas utiliser. Versione obsoleta. Non utilizzare. Verouderde versie. Niet gebruiken.

Cardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers

Perioperative Management of Cardiac Implantable Devices

MRI conditional devices. A luxury or a real clinical need? Sophie Mavrogeni, MD, FESC Onassis Cardiac Surgery Center, Athens, Greece

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

MRI in Patients with Pacemakers and Defibrillators. Paul A. Friedman, MD Win K Shen, MD Robert E. Watson, Jr. MD, Phd October 2016

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE GUIDANCE EXECUTIVE (GE)

Date of Meeting: Ratified Date: 11/03/2010

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

VNS Therapy System Physician s Manual

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Cardiovascular magnetic resonance with an MR compatible pacemaker

8/2/2017. Acknowledgement. Disclaimer. How to manage radiotherapy patients with CIED from initial consult to treatment: TG203 recommendations

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

How to set up a device clinic for heart failure patients Dr C Sridevi

Initial Experience with a Radiologist-Managed MR Imaging Cardiac Pacemaker Program in an Academic-Affiliated Community Hospital

MRI GUIDELINES FOR INSPIRE THERAPY

Management of RT patients with implanted cardiac devices: From recommendation to implementation

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Pediatric pacemakers & ICDs:

Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Dr Randal McRoberts

Lead Management for MRI

Magnetic resonance imaging in patients with cardiac implantable electronic devices

Procedure for the Reprogramming of Pacemakers (PPMs) Prior to Elective and Emergency Surgery

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

MRI Guidelines for the Axonics Sacral Neuromodulation System

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

making a referral for breast imaging Standard Operating Procedure

Protocol. This trial protocol has been provided by the authors to give readers additional information about their work.

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

ImageReady MRI Head Only Guidelines for Spectra WaveWriter Spinal Cord Stimulator System

MRI GUIDELINES FOR INSPIRE THERAPY

Devices and EP- Safety in MRI

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System

Size (mm) (H W D) Mass (g) Volume (cc)

Implant Procedure Concepts

A Closer Look. LATITUDE NXT Alerts SUMMARY. Alerts. Red Alerts

Implantation of an Implantable Cardioverter Defibrillator (ICD)

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

09/05/2018. True/False: All pacemakers have defibrillation capabilities.

28 th September Author Jeremy Gilbert Bariatric Nurse Specialist

Oxford Centre for Clinical Magnetic Resonance Research (OCMR) Adenosine use in stress CMR. Revision: First Issue Version 2.0

CARDIOLOGY. 3 To develop in the trainees the humanistic, moral and ethical aspects of medicine.

ENABLE MRI CLINICAL STUDY SUMMARY

Guidance on competencies for Spinal Cord Stimulation Reviewed

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Division 2, Surgical and Anaesthetics Directorates All surgical and anaesthetics staff Patients with an implanted spinal cord stimulator For:

Pacing Lead Implant Testing. Document Identifier

Safety Issue of Static Magnetic Field (1)

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Open/Upright MRI Criteria Based Access Protocol Supporting people in Dorset to lead healthier lives

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS

Specialised Services Policy Position PP151

TIP CARD. Model Temporary External Pacemaker

BRITISH HEART RHYTHM SOCIETY GUIDELINES FOR THE MANAGEMENT OF PATIENTS WITH CARDIAC IMPLANTABLE ELECTRONIC DEVICES (CIEDs) AROUND THE TIME OF SURGERY

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants

British Society for Dermatological Surgery (BSDS) & British Heart Rhythm Society (BHRS) Guidance on Implanted Devices & Dermatological Surgery

National Cancer Peer Review Programme

Document Control Report

Arrhythmia Care in the DGH What Still Needs to be Done? Dr. Sundeep Puri Consultant Cardiologist

NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)

Implantation of a Cardiac Resynchronisation Therapy-Defibrillator (CRT-D)

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

Model 5392 EPG Temporary Pacer

2009 CPT Codes for Cardiac Device Monitoring

Cardiac Resynchronisation Therapy Defibrillator (CRT-D)

Transcription:

STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0 Date of authorisation Review date April 2019 SOP e-document location Related Trust policies (to be read in conjunction with) Document owner Patricia Feuchter, BHC Head of Radiography

Contents Section 1: Executive summary 2 Section 2: Revision chronology... 2 Section 3: Purpose and scope... 2 Section 4: Background... 3 Section 5: Standard operating procedure... Error! Bookmark not defined. Section 6: SOP review and monitoring compliance/audit requirements... 7 Section 7: SOP dissemination and training... 7 APPENDIX I: Individual Statement of Compliance

Section 1: Executive summary The presence of a cardiac pacemaker or defibrillator has traditionally been regarded as an absolute contraindication to MR scanning. MRI conditional pacemakers now exist however (i.e. safe under certain conditions) which are significantly more resistant to MRI. Similarly there is increasing data suggesting that the risk of scanning non-mri conditional legacy pacemakers is minimal provided specific safety protocols are followed. There are also international guidelines published by British Heart Rhythm Society, European Society of Cardiology and Heart Rhythm Society outlining protocols and best practice for MRI scanning in patients with cardiac implantable electronic devices (CIEDs). This protocol incorporates these recommendations to enable the safe scanning of patients with both MRI conditional and non-mri conditional cardiac devices using MRI. At Barts Heart Centre Cardiac Imaging, we perform approximately 300 scans per annum on patients with CIEDs. This SOP is to protect both staff and patients performing the scans. Section 2: Revision chronology Revision chronology SOP reference 1 Effective date 2 Reason for change 3 Author 4 V1.0 (Draft) First Draft Patricia Feuchter Section 3: Purpose and scope The purpose of this Standard Operating Procedure (SOP) is to specify the processes used to ensure the safe scanning of patients with cardiac devices to be scanned in Cardiac MRI. Section 4: Background Magnetic resonance imaging (MRI) is used increasingly to image patients for diagnostic purposes, and to assess responses to treatment. Formerly, MRI scanning of patients with implantable cardiac devices (loop recorders, pacemakers and implantable cardiac

defibrillators) was felt to be contra-indicated because of the risk of harm to the patient or device. However, the recognition that around 75% of device patients will have a lifetime indication for an MRI scan (17% in the 12 months following device implantation), has led to the development of MRI conditional cardiac devices, to allow these patients to undergo MRI scanning. In addition, selected patients with non-mri conditional devices have undergone MRI scanning safely, without deleterious effects on device function. This has been confirmed in recent clinical studies, and highlighted by inclusion of recommendations for scanning these patients in the 2017 Heart Rhythm Society Guidelines. Section 5. Guidance 5.1 General guidance for doctors referring patients with implantable cardiac devices for MRI scans. Prior to referral for an MRI scan, information should be obtained from the pacing clinic concerning details of the patient s implantable cardiac device (both generator and lead(s) where appropriate). Bookings for MRI will not be accepted until this information has been received. The safety of most medical devices within an MRI scanner can be verified on www.mrisafety.com or via the device manufacturer websites. Currently, the following general principles apply: For a device to be considered MRI conditional, all product labelling must be adhered to: Device and all leads to be labelled as MRI conditional and from the same manufacturer, to form an MRI conditional system. Check on manufacturers webites, or with the Devices Clinic. Pacing / ICD system implanted in either the right or left pectoral region Implanted device in place for more than 6 weeks (there are some exclusions to this). No other implantable devices (including old pacemakers or abandoned leads). If there is doubt then a CXR needs to be performed prior to MRI. Absence of broken leads, lead extenders or adaptors Patient should not be required to be positioned on their side within the scanner. For non-mri conditional devices (i.e. that do not fulfil the above criteria) additional information is required in writing from the referring consultant. Referrals all need to be discussed with the Lead Physician for Device MRI, Dr. Charlotte Manisty (charlotte.manisty@bartshealth.nhs.uk), and we reserve the right to refuse scans if indications or risk are deemed unacceptable. Referrals will not be accepted until this has step has been followed with the correct information provided. Confirmation that the clinical information cannot be obtained any other way (eg by another imaging modality) Confirmation that the clinical information is likely to influence clinical management

Confirmation that the patient is aware that this is an off-label indication, and that they will be asked to sign a consent form making them aware that there is a potential theoretical risk of device malfunction. 5.2 Equipment Requirements MRI scanner Scans currently should only be performed on hydrogen proton magnetic resonance imaging equipment and according to the CIED device conditions for scanning. This is generally - static magnetic field of no greater than 1.5T (Tesla) with RF excitation frequency of 64 MHz. - horizontal closed-bore scanner - maximum gradient slew rate of 200 T/m/s per axis - Whole body SAR 2.0 W/kg and head SAR 3.2 W/kg - Scans should be acquired with Normal operating mode (only some devices allow first level controlled operating mode). There are however some newer CIEDs that are conditional at 3T, and therefore device manufacturer guidance should be followed. Monitoring All patients should be monitored throughout the duration of the MRI scan with pulse oximetry waveform and ECG monitoring. Safety equipment The Resuscitation Trolley must be available in the department and must have an external defibrillator with the potential for transcutaneous pacing (non AECD). Resuscitation drugs must be available. These should be checked regularly and recorded. All resuscitation attempts should be performed only after moving the patient outside of Zone 4 in the MRI department. For the duration of a patient s scan with a non-mri conditional device, the device manufacturer s programmer must be present in the department. Equipment Quantity Cardiac enabled MRI scanner with ECG and pulse oximetry sensors 1 Device Programmer different manufacturers 3 Emergency Resuscitation Trolley (standard) including a manual defibrillator with 1 the capability to deliver external transcutaneous pacing MRI Compatible Evacuation Trolley 1 MRI compatible Cardiac Monitor 1

Image archive and Reporting System CVI42 Report42 Sectra RIS/PACS Trust PC Datix Imaging Database Intranet (Trust Policies) 1 1 5.3 Personnel Requirements for Scanning For patients with an MRI conditional system, there should be A member of staff present nominated the responsible physician Personnel available with the patient for the duration of the scan with the skill to perform advanced cardiac life support, including expertise in the performance of CPR, arrhythmia recognition, defibrillation and transcutaneous pacing. A cardiac device physiologist or cardiologist training in cardiac devices should be present to re-program the device pre and post scan. They need not be in attendance for the duration of the scan but must remain on site. For patients with a non-mri conditional system The cardiac device physiologist should remain in attendance in the MRI department for the duration of the scan (with the appropriate programmer). 5.4 Written Consent for Scanning patients with Non-MRI conditional CIEDs All patients with Non-MRI conditional CIEDs should sign a written consent form on the day of scanning, following a detailed discussion with a cardiology doctor regarding the potential theoretical risks and benefits of scanning. Patients unable to given informed written consent should have a Form 4 signed as per Trust protocol by two doctors. 5.5 Pacemaker re-programming for MRI This should be performed by appropriately trained personnel (cardiac device physiologist, cardiologist or electrophysiology fellow, or specialist device technician. Patients with MRI-conditional CIEDs

All patients should have their devices interrogated immediately prior to scanning, and a record of the battery life (voltage), pacing thresholds, P and R wave amplitudes and lead impedances noted. CIEDs should be re-programmed into MRI-Safe mode as per manufacturer guidance. Post scan, a repeat device interrogation should be performed, and lead and generator parameters recorded as per above. Any significant changes should be noted (lead threshold change >1.0V, P/R wave amplitude reduction of >50%, impedance change of >50ohms) and the patient offered a follow up appointment in the Device Clinic within two weeks. Patients with Non MRI-conditional CIEDs All patients should have their devices interrogated immediately prior to scanning as above. If pacing dependent or high rate of ventricular pacing, program to asynchronous pacing (VOO/ DOO) at a rate higher than their intrinsic rate to avoid competitive pacing If little requirement for pacing, program pacing off (OVO/ ODO) or VVI/ DDI. Deactivate all tachycardia detection and therapies, and switch off advanced magnet mode and noise/ rate response features. Post scan, a repeat device interrogation should be performed, and lead and generator parameters recorded as per above. Any significant changes should be noted (lead threshold change >1.0V, P/R wave amplitude reduction of >50%, impedance change of >50ohms) and the patient offered a follow up appointment in the Device Clinic within two weeks. Section 6 Personnel involved with SOP 6.1 Staff groups SOP applies to This SOP applies to all staff working in the Barts Heart Centre Cardiac Imaging Department, including: Imaging Clinical Director and Consultants (within Radiology and Cardiology) Imaging Fellows and Trainees Cardiac Physiologists

Cardiac MRI Radiographers MRI Physicists 6.2- Key operational relationships Inter-department relationships Radiologists Cardiac Physiologists MRI Physicists Other departments Cardiology Safeguarding Team Resus Team Barts Site DoN/ Deputy DoN External stakeholders Clinicians caring for these patients 6.3 Responsible personnel Clinical Director (Cardiac Imaging) Cardiac Physiologists Operational Lead Radiographer, Cross Sectional General Manager (Specialised, Imaging, Pharmacy, OPD & Network) Cardiac Imaging MDT (all) Section 7: SOP review and monitoring compliance/audit requirements 7.1 SOP Review This policy will be reviewed in 3 years, as set out in the Policy for the Development and Implementation of Procedural Documents, unless changing circumstances necessitate an earlier review. 7.2 Audit requirements Annual compliance audit (Cardiac Imaging Governance Lead) Monitor compliance in protocolling criteria and documentation Non-compliance to be reported and investigated via DATIX

Section 8: SOP dissemination and training 8.1 Dissemination New Staff At local induction and competency assessment Existing Staff - New versions to be disseminated via email for existing staff 8.2 Training There is no mandatory training associated with this policy. Ad hoc training based on an individual s training needs will be defined within their annual appraisal or job plan Read and confirm (Appendix I)

APPENDIX I Individual Statement of Compliance This statement is to be scanned and returned to the stored electronically in your individual HR folder. The original need not be retained. The departmental competency dashboard must be updated. SOP Title.. Version Number.. I have read and understand my duties described in this standard operating procedure Signed Date Name..