Technique Guide. Titanium Wire with Barb and Needle. Surgical Technique Guide for Canthal Tendon Prodecures.

Similar documents
Titanium Wire with Barb and Needle. Surgical Technique Guide for Canthal Tendon Procedures.

Titanium Wire With Barb and Needle

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

MatrixNEURO. The next generation cranial plating system.

Low Profile Neuro Plating System. Surgical Technique

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

low ProfIle neuro PlaTIng system

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Surgical Technique. Clavicle Locking Plate

LOW PROFILE NEURO. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

Surgical Technique. Olecranon Locking Plate

Surgical Technique. Calcaneal Locking Plate

Technique Guide with Application Forceps. FlapFix. Quick and stable fixation of cranial bone flaps following a craniotomy.

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

Nasal Orbital Ethmoid (NOE) Fractures November, 2014

Technique Guide. Midface Distractor System. For the temporary stabilization and gradual lengthening of the cranial or midfacial bones.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Open reduction; plate fixation 1 Principles

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth.

Surgical Technique. Distal Humerus Locking Plate

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide with Application Instrument. FlapFix. Quick and stable fixation of cranial bone flaps following a craniotomy.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

Dentoalveolar Bone Fixation System. Bone fixation sets for the office.

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system.

RapidSorb Resorbable Tacks. Resorbable Fixation System.

Surgical Technique. Proximal Humerus Locking Plate

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Surgical Technique International Version. Clavicle Locking Plate

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Surgical technique. IMF Screw Set. For temporary, peri opera tive stabilisation of the occlusion in adults.

2.0 mm Mandible Locking Plate System

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. IMF Screw Set. For intermaxillary fixation.

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine.

Surgical Technique. CONQUEST FN Femoral Neck Fracture System

3.5 mm LCP Extra-articular Distal Humerus Plate

Trochanter Stabilization Plate for DHS Implants

The CerviFix System. Including the StarLock Components. CerviFix Clamp and Screw. StarLock Clamp and Screw

3.5 mm LCP Clavicle Hook Plates

The Orthopaedic Cable System TECHNIQUE GUIDE

Tension Band Pin System 2. Surgical Technique

The Titanium Tibial Nail System

Technique Guide. Cranial Clamps with Application Instrument. Quick and stable fixation of cranial bone flaps following a craniotomy.

CRANIAL RECONSTRUCTION SOLUTIONS

MatrixRIB. Stable fixation of normal and osteoporotic ribs.

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

Surgical Technique. Targeter Systems Overview

The Locking Calcaneal Plate Instrument and Implant Sets

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

MatrixNEURO Cranial Plating System

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

3.5 mm LCP Superior and Superior Anterior Clavicle Plates. Part of the Synthes modular clavicle plate system.

MATRIX PLATFORM FEATURES DESIGNED TO COMPLEMENT AO FIXATION PRINCIPLES

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

LCP Distal Humerus Plates

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

Distal Ulnar Locking Plate

Biomet Large Cannulated Screw System

LCP Superior Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

SURGICAL TECHNIQUE THE TENDON ANCHOR SYSTEM

MatrixNEURO. The next generation cranial plating system.

Low Bend Distal Tibia Plates

For Fast and Stable Fixation of the Sternum. Sternal ZIPFIX. System. Surgical Technique

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Extra-articular Distal Humerus Plate.

TELEFIX SURGICAL TECHNIQUE. Implant system for the anterior stabilization of the thoracolumbar spine

AcUMEDr. LoCKING CLAVICLE PLATE SYSTEM

Figure 3 Figure 4 Figure 5

Bio-TransFix ACL Reconstruction. Surgical Technique

VA LOCKING CALCANEAL PLATES 2.7

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

SynPOR HD Facial Shape System. For the augmentation or reconstruction of the craniomaxillofacial skeleton.

A Legacy of Serving the Surgical Community

Distal Radius Plate 2.4/2.7 dorsal and volar

Olecranon Locking Plate II

Cranial System. Neuro Plating and Screw System

TSLP Thoracolumbar Spine Locking Plate

Zimmer Small Fragment Universal Locking System. Surgical Technique

Lag Screw Device Intended for symphyseal fracture fixation of the mandible

30+ MEDPOR biomaterial. years of proven clinical history

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

3.5 mm LCP Olecranon Plates

versatile SOLUTION for COMPLEX fractures

Transcription:

Technique Guide Titanium Wire with Barb and Needle. Surgical Technique Guide for Canthal Tendon Prodecures.

Indications/Features Indications The Synthes Titanium Wire with Barb and straight Needle is indicated for use in soft tissue approximation and/or ligation, for canthoplasty, canthopexy, and/or medial canthal tendon repair. Features Manufactured from titanium and titanium alloy Versatile 28 gauge (0.31 mm diameter) wire size Permanently affixed barb Straight taper-point needle Compatible with titanium bone fixation systems Available in single-use sterile packs Benefits Convenient and easy handling Facilitates capture of the medial canthal tendon Minimizes palpability while providing adequate strength Prevents galvanic corrosion when used with titanium plates and screws Minimal MRI scatter Technique Overview Preoperative Planning Medial Canthal Tendon Repair Postoperative Considerations Product Information Titanium Wire with Barb and Needle Technique Guide Synthes 1

Preoperative Planning Canthal tendon repair using the Synthes Titanium wire with Barb and straight Needle is indicated when the medial canthal tendon is detached from the bone as the result of trauma or a surgical approach. Note: When the medial canthal tendon remains attached to a large bone fragment in the case of trauma, anatomical reduction and stabilization of the bone fragment is sufficient in most cases. The bony skeleton must be properly restored before canthopexy by reduction and osteosynthesis of the fragments. The normal distance between the canthal tendons is approximately half the interpupillary distance. 1 Note: In an adult, the normal intercanthal distance is approximately 32 35 mm.2 It is recommended that the lacrimal duct be intubated prior to the start of the procedure. 2 Synthes Titanium Wire with Barb and Needle Technique Guide

Surgical Technique for Medial Canthal Tendon Repair 1 Surgical approaches In the case of serious injury, a coronal approach is usually necessary to stabilize the bony fragments. 2 Reduce fractures Reduce and stabilize all fractures. Before canthal tendon reattachment, the bony-cartilaginous framework must be precisely repaired. Notes: If the medial canthal tendon is attached to a bone fragment, repositioning and plating the fragment generally leads to the most anatomic appearance. After securing the wire, access to the internal orbit will be limited, therefore orbital wall reconstruction should be completed before canthal resuspension. 3 Locate tendon Locate the traumatized medial canthal tendon. The tendon may be identified from inside the coronal flap, or through a small skin incision, or alternatively through a caruncular incision. These incisions provide direct access to the tendon. The Lacrimal Fossa can be used as a point of reference when locating the medial canthal tendon. Note: The approach to the medial canthal tendon is posterior to the lacrimal duct and should not impinge on the lacrimal system. If using the skin incision, the tendon does not necessarily need to be visualized to complete this procedure. The tendon can be palpated by using the needle to find the area of most resistance. Titanium Wire with Barb and Needle Technique Guide Synthes 3

Surgical Technique for Medial Canthal Tendon Repair 4 Capture cantheal tendon To capture the canthal tendon with the barb on the wire, the needle is guided through a small skin incision below the medial canthus through the site of greatest resistance (approximately 2 mm medial to the canthus) toward the inside of the coronal flap. The titanium wire is guided through this flap until the barb captures the canthal tendon. Instead of a skin incision below the lid margin, an incision can be made in the caruncula. Alternatively: By using the caruncular incision, the barb will become engaged in the substance of the tendon after the needle and wire are passed through it. Notes: If the medial canthal tendon has been severely traumatized, wire fixation may not be possible. Another method may be required. In handling titanium wire, care should be taken to avoid damage from handling, such as kinking or excessive twisting. Avoid crushing or crimping damage due to application of surgical instruments such as forceps or needle holders. 4 Synthes Titanium Wire with Barb and Needle Technique Guide

5 Plan tendon position Proper tendon repair includes positioning the canthal tendon posterior and superior to the lacrimal fossa. 6 Tendon placement To facilitate tendon placement, a titanium adaption plate should be placed on the frontal bone, extending inferiorly and posteriorly toward the medial orbital wall 3. Cut and contour the plate to fit the patient s anatomy. Insert at least three bone screws to affix the plate to the bone. Notes: The most inferior-posterior screw hole in the plate must be located at the planned position of canthal tendon resuspension and must remain empty to allow passage of the titanium wire transnasally. In cases with minimal bone loss, an adaption plate may not be necessary for canthal tendon repair. Other methods used for ensuring the posterior and superior pull of the canthal tendon include the use of medial orbital bone grafts and passage of the titanium wire through the posterior portion of the perpendicular plate of the ethmoid bone. Plate placement may depend on availability of sufficient bone. Titanium Wire with Barb and Needle Technique Guide Synthes 5

Surgical Technique for Medial Canthal Tendon Repair 7 Drill transnasally Using a 2.0 mm to 2.4 mm diameter bit, drill transnasally from the nonaffected orbit to the affected orbit. Notes: Use a drill sleeve to protect the soft tissue and globes when drilling. In cases of severe comminution, drilling may not be required. The use of a transnasal awl may help facilitate wire passing. 8 Transnasal passage of the wire This can be accomplished either with a perforated awl or with the aid of a large cannula serving as a guide for the wire. Alternatively: The wire can be passed through the posterior plate hole, then come forward within the orbit to be fixated to the supraorbital/ frontal bone. After tightening the final screw, the wire may be directed anteriorly to be fixated on the ipsilateral supraorbital or frontal bone. 9 Remove needle Remove the needle directly under the needle crimp. Note: Exercise caution when handling surgical needles to avoid inadvertent needle sticks. Discard used needles in sharps containers. 6 Synthes Titanium Wire with Barb and Needle Technique Guide

10 Apply tension Apply moderate tension and visually check the position of the canthal tendon. For stable fixation, the canthal tendon must be moved into the desired position in a completely relaxed state. 11 Secure wire Secure the titanium wire to the supraorbital rim on nonaffected side. Note: Ensure proper fixation of the wire before closure. 12 Postoperative considerations Frequent examinations of visual acuity during the first 24 hours postoperatively are recommended. Titanium Wire with Barb and Needle Technique Guide Synthes 7

Product Information 493.104.01S Canthal Tendon Wire with Barb and Straight Needle, 28 Gauge (0.31 mm diameter), length 500 mm, sterile References 1 Ellis, Edward III. Sequencing Treatment for Naso-orbitoethmoid Fractures. Journal of Oral and Maxillofacial Surgery, 51 (1993), pp. 543 558. 2 Frodel, John L. Medial Canthal Tendon Repair. In Atlas of Head & Neck Surgery Otolaryngology. Ed. B.J. Bailey, et al. Philadelphia: Lippincott-Raven, 1996, pp. 552 553. 3 Hammer, Beat; et al. Secondary Correction of Posttraumatic Deformities After Naso-Orbito-Ethmoidal Fractures: An Invited Technique. The Journal of Cranio-Maxillofacial Trauma, 7:1 (2001), pp. 44 51. 8 Synthes Titanium Wire with Barb and Needle Technique Guide

02/2011 60100272 Synthes, Inc. or its affiliates Subject to modification Synthes is a trademark of Synthes, Inc. or its affiliates All technique guides are available as PDF files at www.synthes.com/lit Ö036.000.538öABaä 0123 036.000.538 version AB