Dosimetric Benefit of a New Ophthalmic Radiation Plaque

Similar documents
A REANALYSIS OF THE COLLABORATIVE OCULAR MELANOMA STUDY MEDIUM TUMOR TRIAL EYE PLAQUE DOSIMETRY

Factory loaded, sterilized, ready to implant plaques:!

Seed coordinates of a new COMS-like 24 mm plaque verified using the FARO Edge

Research Article Outcomes and Control Rates for I-125 Plaque Brachytherapy for Uveal Melanoma: A Community-Based Institutional Experience

Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma

Inherent precision. Ru-106 Eye Applicators and I-125 Ophthalmic Seeds.

A practical approach to estimating optic disc dose and macula dose without treatment planning in ocular brachytherapy using 125 I COMS plaques

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

An Interactive TreatmentPlanning System For Ophthalmic Plaque Radiotherapy

CLINICAL SCIENCES. Pretreatment Characteristics and Response to Plaque Radiation Therapy

Quantitative OCT Angiography Evaluation of Peripapillary Retinal Circulation after Plaque Brachytherapy

Can Protons replace Eye Brachytherapy? 1 Department of Radiation Oncology

CLINICAL SCIENCES. Conclusions: A distinctive postbrachytherapy regression

Episcleral plaque brachytherapy has been proposed as an

Visual Acuity, Contrast Sensitivity and Color Vision Three Years After Iodine-125 Brachytherapy for Choroidal and Ciliary Body Melanoma

LABORATORY SCIENCES. Attenuation of Iodine 125 Radiation With Vitreous Substitutes in the Treatment of Uveal Melanoma

Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment

Tall, dark and.. Uh oh

Gender Differences in Clinical Presentation and Prognosis of Uveal Melanoma METHODS. Patients

Systemic and ocular follow-up after conservative management of an intraocular tumor

Carlo Mosci. Ocular Oncology Service Galliera Hospital Genova Italy (

Ophthalmic tumors. Wolfgang Sauerwein

National Guidelines for the management of uveal melanoma. Sponsored by Melanoma Focus

Retina Center of Oklahoma Sam S. Dahr, M.D. Adult Intraocular Tumors

NOVEL SOFTWARE MODULES FOR TREATMENT PLANNING OF 106 RU EYE PLAQUE BRACHYTHERAPY

MEDICAL POLICY SUBJECT: TRANSPUPILLARY THERMOTHERAPY. POLICY NUMBER: CATEGORY: Technology Assessment

FEP Medical Policy Manual

Episcleral eye plaque dosimetry comparison for the Eye Physics EP917 using Plaque Simulator and Monte Carlo simulation

Proton Radiation Therapy of Ocular Melanoma at PSI

Vitreoretinal surgical management In ocular oncology

Ruthenium-106 Brachytherapy with or without Additional Local Therapy Shows Favorable Outcome for Variable-Sized Choroidal Melanomas in Korean Patients

Case Rep Oncol 2012;5: DOI: /

A Little Physics. How Does It Work? Radiation Therapy for Choroidal Neovascularisation in AMD A Review. => cell death

Cyberknife Radiosurgery for Uveal Melanoma

Update on Intraocular Oncology

Intraocular Radiation Therapy for Age-Related Macular Degeneration

Gamma Knife radiosurgery for locally recurrent choroidal melanoma following plaque radiotherapy

Dose to professionals in epiescleral brachytherapy

Ruthenium-106 Plaque Radiotherapy for Retinal Vasoproliferative Tumors

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Optimisation of eye plaque dosimetry using Monte Carlo method

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Optimisation of eye plaque dosimetry using Monte Carlo method. J. Green, D. Cutajar, S. Guatelli, & Rosenfeld, A.B.

Continuing Medical Education

Plaque Radiotherapy in Recurrent or Incomplete- Excised Conjunctival Squamous Cell Carcinoma and Melanoma

EVIDENCE BASED MANAGEMENT FOR Retinoblastoma

Applications in brachytherapy

Intraocular Radiotherapy for Age- Related Macular Degeneration

Michael P. Blair, MD Retina Consultants, Ltd Libertyville/Des Plaines, Illinois Clinical Associate University of Chicago 17 October 2015

Gold nanoparticles-based brachytherapy enhancement in choroidal melanoma using a full Monte Carlo modelling of human eye

CLINICAL SCIENCES. Three-Dimensional Ultrasound for the Measurement of Choroidal Melanomas

Management and Outcome of Uveal Melanoma in a Single Tertiary Cancer Center in Jordan

Intraocular Radiation Therapy for Age-Related Macular Degeneration

Development and validation of a patient based measure of outcome in ocular melanoma

Financial Disclosures

T.O. Romanova, a Post Graduate student

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

UCSF Uveal Melanoma Program: Outcomes with Proton Beam Radiation Therapy Kavita K. Mishra, M.D., M.P.H. UCSF Comprehensive Cancer Center

University of Wollongong. Research Online

Acute Retinal Necrosis Secondary to Varicella Zoster Virus in an Immunosuppressed Post-Kidney Transplant Patient

LONG-TERM VISUAL ACUITY RESULTS OF TREATED AND UNTREATED RADIATION RETINOPATHY (AN AOS THESIS)

Case Report: Indocyanine Green Dye Leakage from Retinal Artery in Branch Retinal Vein Occlusion

Intravascular Brachytherapy Dosimetry Techniques for Gamma systems

Eye plaque dosimetry verification using novel solid state devices

Royal Berkshire Hospital Dunedin Hospital. Prince Charles Eye Unit Pi Princess Margaret Hospital

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

Retinoblastoma: A Review of Current Treatment Strategies

ROLE OF LASER PHOTOCOAGULATION VERSUS INTRAVITREAL TRIAMCINOLONE ACETONIDE IN ANGIOGRAPHIC MACULAR EDEMA IN DIABETES MELLITUS

Adaptive replanning intensity-modulated radiotherapy for choroidal metastasis of breast cancer using optical coherence tomography

M alignant melanoma of the uvea causes clinical metastases

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER

Local Recurrence Significantly Increases the Risk of Metastatic Uveal Melanoma

OUR EYES & HOW WE SEE

Clinically Significant Macular Edema (CSME)

Ruthenium-106 plaque brachytherapy in the primary management of ocular medulloepithelioma

Andrew J. Barkmeier, MD; Benjamin P. Nicholson, MA; Levent Akduman, MD

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

CLINICAL SCIENCES. Low-Dose Proton Beam Therapy for Circumscribed Choroidal Hemangiomas

Detecting ultrasonographic hollowness in small choroidal melanocytic tumors using 10 MHz and 20 MHz ultrasonography: a comparative study

Tiffany L. Kruger, D.O. Children s Hospital of Michigan Wayne State University/Kresge Eye Institute

optic disc neovascularisation

Study of clinical significance of optical coherence tomography in diagnosis & management of diabetic macular edema

Fluorescein and Indocyanine Green Videoangiography of Choroidal Melanomas

MEDICAL POLICY. SUBJECT: BRACHYTHERAPY OR RADIOACTIVE SEED IMPLANTATION FOR PROSTATE CANCER POLICY NUMBER: CATEGORY: Technology Assessment

Osteoradionecrosis and Radiation Dose to the Mandible in Patients With Oropharyngeal Cancer

Although photocoagulation and photodynamic PROCEEDINGS PEGAPTANIB SODIUM FOR THE TREATMENT OF AGE-RELATED MACULAR DEGENERATION *

PEDIATRIC ORBITAL TUMORS RADIOTHERAPY PLANNING

Retina Conference. Janelle Fassbender, MD, PhD University of Louisville Department of Ophthalmology and Visual Sciences 09/04/2014

Clinical Trials in Diabetic Retinopathy. Harry W. Flynn Jr., M.D. Nidhi Relhan Batra, M.D.

Gold nanoparticle-based brachytherapy enhancement in choroidal melanoma using a full Monte Carlo model of the human eye

Long-Term Follow-Up of Patient with Diabetic Macular Edema Receiving Fluocinolone Acetonide Intravitreal Implant

A comparison of community-based versus university-based centers in clinical trial performance

The Impact of Image Guided Radiotherapy in Breast Boost Radiotherapy

J of Evolution of Med and Dent Sci/ eissn , pissn / Vol. 4/ Issue 55/ July 09, 2015 Page 9665

Subgroup Analysis of the MARINA Study of Ranibizumab in Neovascular Age-Related Macular Degeneration

I t is possible to apply large doses of

International Journal of Health Sciences and Research ISSN:

SUMMARY. Heather Casparis, MD,* and Neil M. Bressler, MD MARINA AND ANCHOR

Transcription:

International Journal of Radiation Oncology biology physics www.redjournal.org Physics Contribution Dosimetric Benefit of a New Ophthalmic Radiation Plaque Gaurav Marwaha, MD,*,x Allan Wilkinson, PhD,*,x James Bena, MS, y,x Roger Macklis, MD,*,x and Arun D. Singh, MD*,z,x Departments of *Radiation Oncology, Taussig Cancer Center, y Quantitative Health Sciences, and z Ophthalmic Oncology, Cole Eye Institute, x Cleveland Clinic Foundation, Cleveland, Ohio Received Nov 10, 2011, and in revised form Jan 20, 2012. Accepted for publication Jan 29, 2012 Summary Uveal melanoma is most often treated with Collaborative Ocular Melanoma Study (COMS) I-125 brachytherapy plaques with excellent disease control, albeit with radiation-induced visual loss. For 100 uveal melanoma patients, a dosimetric study between COMS plaques and a new ophthalmic plaque, EP917, compared radiation dosage to vision critical structures for each plaque. EP917 delivered significantly less radiation to certain visual structures while maintaining adequate tumor coverage. Translation to clinical visual improvement with the new plaque remains to be proved. Purpose: To determine whether the computed dosimetry of a new ophthalmic plaque, EP917, when compared with the standard Collaborative Ocular Melanoma Study (COMS) plaques, could reduce radiation exposure to vision critical structures of the eye. Methods and Materials: One hundred consecutive patients with uveal melanoma treated with COMS radiation plaques between 2007 and 2010 were included in this study. These treatment plans were generated with the use of Bebig Plaque Simulator treatment-planning software, both for COMS plaques and for EP917 plaques using I-125. Dose distributions were calculated for a prescription of 85 Gy to the tumor apex. Doses to the optic disc, opposite retina, lens, and macula were obtained, and differences between the 2 groups were analyzed by standard parametric methods. Results: When compared with the COMS plaques, the EP917 plaques used fewer radiation seeds by an average difference of 1.94 (P<.001; 95% confidence interval [CI], 2.8 to 1.06) and required less total strength of radiation sources by an average of 17.74 U (air kerma units) (P<.001; 95% CI, 20.16 to 15.32). The total radiation doses delivered to the optic disc, opposite retina, and macula were significantly less by 4.57 Gy, 0.50 Gy, and 11.18 Gy, respectively, with the EP917 plaques vs the COMS plaques. Conclusion: EP917 plaques deliver less overall radiation exposure to critical vision structures than COMS treatment plaques while still delivering the same total therapeutic dose to the tumor. Ó 2012 Elsevier Inc. Keywords: Uveal melanoma, Episcleral plaque, Dosimetry, Ophthalmic, Brachytherapy Introduction Uveal melanoma is most frequently treated with brachytherapy by the use of custom-designed episcleral plaques (1). Since the Reprint requests to: Gaurav Marwaha, MD, Department of Radiation Oncology, Taussig Cancer Center, T-28, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. Tel: (216) 392-9651; Fax: (216) 445-1217; E-mail: marwahg2@ccf.org Collaborative Ocular Melanoma Study (COMS) was initiated in 1985 (2-4), I-125 has been the most frequently used radionuclide in the United States, whereas ruthenium-106 is preferred in Europe (5). Excellent primary tumor control rates ranging from 90%-95% have been reported with I-125 episcleral plaque Presented in poster format at the 53rd annual meeting of the American Society for Radiation Oncology, Miami Beach, FL, October 2-6, 2011. Conflict of interest: none. Int J Radiation Oncol Biol Phys, Vol. -, No. -, pp. 1e5, 2012 0360-3016/$ - see front matter Ó 2012 Elsevier Inc. All rights reserved. doi:10.1016/j.ijrobp.2012.01.084

2 Marwaha et al. International Journal of Radiation Oncology Biology Physics radiotherapy. However, moderate loss of vision secondary to radiation-related complications like retinopathy and optic neuropathy is observed in a large proportion of cases (6). In the COMS randomized trial of I-125 brachytherapy for medium-sized choroidal melanoma, loss of 6 or more lines of visual acuity from the pretreatment level occurred in 49% of eyes after 3 years (7). The complications of radiation retinopathy and optic neuropathy are essentially untreatable over the long term, although shortterm benefit with laser photocoagulation, the use of intravitreal triamcinolone acetonide, or anti-vascular endothelial growth factor treatment in nonrandomized case series has been reported (8). Therefore, it has become increasingly more important to find ways to avoid radiation retinopathy and optic neuropathy. The total dose of radiation to critical ocular structures such as the macula and optic disc is the major modifiable risk factor for radiation retinopathy and optic neuropathy after I-125 plaque radiation therapy (7, 9-12). The approaches available to reduce collateral radiation dose include prescription of less than a conventional therapeutic dose of 85 Gy to the tumor apex (13), the use of alternative radionuclides such as strontium-90, ruthenium- 106, and palladium-103 (14), or attenuation of radiation using silicon endotamponade (15). Another approach would be to improve on existing COMS plaque designs (16, 17). In the COMS plaque, I-125 seeds are loaded in a Silastic carrier that is placed within gold casing in such a manner that the seeds are 1 mm away from the sclera, allowing for a more homogeneous dose distribution (Fig 1A) (3). By designing plaques wherein the seeds are glued into the shallow grooves of the gold, the need for a carrier disappears, and the I-125 seeds can now be placed in close proximity to the sclera (Eye Physics, LLC, Los Alamitos, CA) (Fig 1B) (18). The collimated field from each seed is designed to overlap just below the base of the tumor, which also provides a homogenous distribution of the radiation (16). Such a plaque has a theoretical advantage of reduced lateral dose as confirmed by thermoluminescent dosimeter and radiochromic film studies (16). We performed a study of 100 consecutive cases of posterior uveal melanoma to determine whether the dosimetry of a new ophthalmic plaque (EP917) when compared with the COMS plaques would reduce radiation exposure to vision critical structures of the eye. Methods and Materials A series of 100 consecutive patients treated at the Department of Ophthalmic Oncology, Cole Eye Institute, Cleveland Clinic, between the years 2007 and 2010, with a clinical diagnosis of posterior uveal melanoma based on ophthalmoscopy, ultrasonographic features, and angiographic studies, was used in this study. The tumors ranged in height from 2.42 mm-14.00 mm, representing the full spectrum of small, medium, and large melanomas according to COMS criteria. Each patient was planned with Plaque Simulator (Eckert & Ziegler BEBIG GmbH, Berlin, Germany) and treated with COMS radiation plaques (Trachsel Dental Studio, Rochester, MN). Each plaque was chosen to be the smallest one that provided adequate target coverage (V100 >98%). Later, these same cases were replanned with the EP917 plaque (Eye Physics, LLC, Los Alamitos, CA). Each new plan involved treatment planning and dosimetry based on retinal diagrams from the original COMS plans. No patient was treated with the EP917 plaque. The location of each tumor was expressed as the distance in millimeters of the nearest tumor margin to the center of the optic disc and center of the macula. In all cases, each type of plaque was centered on the tumor base. The COMS plaques were fully loaded with I-125 seeds, whereas the EP917 plaques were loaded with only enough seeds to ensure that the V100 exceeded 98%. In each plaque, the seeds were of equal source strength adjusted to deliver a total therapeutic dose of 85 Gy to the tumor apex over the course of 72 hours. Again, the treatment planning system used was Plaque Simulator. Dosimetric verification of this system has been Figure 1. Inner views of the Collaborative Ocular Melanoma Study (COMS) plaque (A) and the Eye Physics (EP)917 plaque (B).

Volume - Number - 2012 Dosimetric benefit of a new ophthalmic radiation plaque 3 Table 1 Summary of radiation data: COMS and EP917 plaques Plaque Factor Mean SD Minimum Median Maximum COMS Number of seeds 13.96 5.43 7.00 13.00 24.00 Total activity (U) 79.54 57.38 32.40 59.56 373.00 Dose to optic disc (Gy) 31.85 25.69 5.05 27.56 206.10 Dose to opposite retina (Gy) 6.19 4.69 2.20 4.66 28.10 Dose to lens (Gy) 17.29 19.06 3.40 9.71 98.48 Dose to macula (Gy) 57.54 44.52 5.63 50.56 238.20 Dose to sclera (Gy) 229.35 117.90 118.20 189.75 820.20 EP917 Number of seeds 12.03 3.96 2.00 13.00 17.00 Total activity (U) 61.80 46.93 20.83 44.75 298.00 Dose to optic disc (Gy) 27.28 22.50 4.72 22.20 162.50 Dose to opposite retina (Gy) 5.69 4.35 1.80 4.10 27.27 Dose to lens (Gy) 17.07 19.75 2.70 9.48 110.70 Dose to macula (Gy) 46.36 41.06 5.13 37.05 233.00 Dose to sclera (Gy) 220.01 148.76 104.00 163.65 931.90 Abbreviations: COMS Z Collaborative Ocular Melanoma Study; EP Z Eye Physics; SD Z standard deviation. published by Knutsen et al (19). The computations for this study incorporated the nonhomogeneity corrections for the Silastic carrier and the gold alloy backing. Next, the number of seeds, total required radiation source strength, and total doses to critical structures of the eyes were compared head to head between the original COMS plaque and the EP917 plaque. The differences were described by means, standard deviations (SD), and 95% confidence intervals (CI) and were compared by dependent t tests. Differences in radiation doses to the critical structure between COMS plaque and EP917 plaque were calculated as EP917-COMS plaque, so that a negative measure reflects less radiation dose with EP917 plaque. Also, the relationship between the dose differences with COMS vs EP917 and the distance of the tumors from critical vision structures was assessed by a Pearson correlation test. Normality of the differences between the COMS and EP917 data was assessed empirically before testing, and nonparametric methods (Wilcoxon signed ranked tests and Spearman correlations) were used as a sensitivity analysis. Additionally, a McNemar test was used to evaluate the ability to keep macula doses under thresholds of 30 Gy, 40 Gy, or 50 Gy. Statistical analysis was performed with R software (version 2.8; Vienna, Austria), and all statistical tests used a significance level of.05. Results The 100 consecutive patients had a median age of 63 years (range, 26-89 years) with uveal melanoma treated with COMS plaques and then replanned for treatment with the EP917 plaques (Table 1). When compared with the COMS plaque, the mean number of radiation seeds used in the EP917 plaque was fewer by 1.94 (P<.001; 95% CI; 2.8 to 1.06), and the total strength of radiation sources was also reduced by 17.74 U (P<.001; 95% CI, 20.16 to 15.32). The mean radiation dose (Gy) delivered to the optic disc, opposite retina, and macula was less by 4.57 (P<.001; 95% CI, 6.91 to 2.23), 0.50 (P<.001; 95% CI, 0.72 to 0.28), and 11.18 (P<.001; 95% CI, 14.58 to 7.78), respectively, for the EP917 plaque vs the COMS (Table 2). The mean of total Gy delivered to the lens was less by 0.23 Gy, although this was not statistically significant (PZ.48). The correlation between distance of tumor from center of macula, and difference in dose delivered between EP917 and COMS, was 0.44 (P<.001; 95% CI, 0.26-0.58). Statistically significant positive associations were observed between the distance and the difference in the total radiation doses to the macula. For tumors near the macula, the reduction in dose with the EP917 plaque was greater. For tumors farther away from the macula (8 mm or more), the difference in the dose was very close to zero. Correlation between distance of tumor from optic disc, and difference in dose delivered to optic disc between EP917 and COMS, was 0.06 not statistically significant (PZ.54). For threshold doses to the macula, both methods delivered 40 Gy in 44 cases (44%) and >40 Gy for 46 cases (46%); for the remaining 10 cases, the COMS delivered a dose >40 Gy, and EP917 delivered a dose 40 Gy (PZ.004). There were no cases wherein a COMS plaque delivered 40 Gy and the EP917 plaque Table 2 Difference in radiation dose (Gy) delivered to various critical structures (calculated as EP917 plaque dose minus COMS plaque dose) Structure n Mean SD Lower 95% CI Upper 95% CI P value Optic disc 100 4.57 11.81 6.91 2.23 <.001 Opposite retina 100 0.50 1.13 0.72 0.28 <.001 Lens 100 0.23 3.23 0.87 0.41.48 Macula 100 11.18 17.16 14.58 7.78 <.001 Sclera 100 9.34 44.67 18.20 0.48.039 Abbreviations: COMS Z Collaborative Ocular Melanoma Study; EP Z Eye Physics; SD Z standard deviation; CI Z confidence interval; ( ) Z COMS plaque dose is greater than EP917 plaque dose; (þ) Z EP917 plaque dose is greater than COMS 125 Iodine plaque dose.

4 Marwaha et al. International Journal of Radiation Oncology Biology Physics Table 3 Macular radiation dose: COMS vs EP917 Dose (Gy) EP917: No EP917: Yes P value >30 COMS 125 I: No 35 1.027 COMS 125 I: Yes 9 55 >40 COMS 125 I: No 44 0.004 COMS 125 I: Yes 10 46 >50 COMS 125 I: No 50 0 <.001 COMS 125 I: Yes 13 37 Abbreviations: COMS Z Collaborative Ocular Melanoma Study; EP Z Eye Physics. delivered >40 Gy. Similar observations favoring the EP 917 plaque were made for threshold values >30 Gy and >50 Gy (Table 3). Analysis of radiation dose to the optic disc in this study revealed that in 17 cases the COMS delivered a dose >30 Gy, and EP917 delivered a dose 30 Gy (PZ.001). Both plaques delivered 30 Gy in 57 cases (57%) and >30 Gy for 26 cases (26%). There were no cases wherein COMS delivered 30 Gy and EP917 plaque >30 Gy. Similar observations favoring the EP 917 plaque were made for threshold values >30 Gy. Discussion The total dose of radiation to critical ocular structures such as the macula and the optic disc is the major modifiable risk factor for radiation retinopathy and optic neuropathy after I-125 plaque radiation therapy (7, 9e12). Collimated I-125 plaques have been designed to offer conformal therapy by providing greater dose homogeneity within the tumor and reducing the radiation to uninvolved adjacent structures (16). Our data suggest that in a head-to-head dosimetric comparison of the same radioisotope, I-125, EP917 has superior dose sparing to critical vision structures of the eye vs the COMS. Specifically, 2 of the most vital vision structures, the macula and the optic disc, received significantly less radiation dosage with the EP917 vs the COMS (Fig 2). Our findings are similar to a previous study wherein collimating plaques were simulated for large melanomas (17). Studies based on treatment with proton beam radiation have indicated that the risk of maculopathy increases linearly up to 40 Gy before plateauing (20). Analysis of radiation dose to the macula in our study indicates that in several cases, the COMS plaque delivered a dose >40 Gy and EP917 delivered a dose of 40 Gy, but not vice versa. These findings indicate the possibility for a lower likelihood of radiation maculopathy with the use of EP917 plaques. Regarding radiation optic neuropathy, the risk is negligible below 30 Gy, with a gradual rise to 100% risk at 70 Gy (20). Analysis of radiation dose to the optic disc in our study indicates that in a large number of cases, the COMS plaque delivered a dose >30 Gy and EP917 delivered a dose 30 Gy. These significant findings indicate the possibility for lower likelihood of radiation optic neuropathy with EP917 plaques. Given that the risks of radiation retinopathy and optic neuropathy are dose dependent and both entities are essentially untreatable, one way to achieve superior visual outcomes after brachytherapy for uveal melanoma would be to reduce radiation dose to the vision critical structures. Our data, based on analysis of 100 consecutive cases, suggest that the EP917 plaque offers theoretical advantage over the COMS plaque. However, benefit from a clinical outcome standpoint has not yet been demonstrated. One of the limitations of our study is that we used only one size of EP917 plaque for simulation because it is the only collimated Eye Physics plaque with verified dosimetry. We compared one size of EP917 plaque with various sized, notched/not-notched COMS plaques that had all been dosimetrically verified. The COMS plaque may still be superior for tumors that are in proximity to the optic disc, because of the availability of notched COMS plaques. However, in our study, there was no selection bias to the shape, size, or location of each tumor because we randomly selected 100 consecutive cases to estimate benefit in the clinical setting. Regardless of these factors, less dose was delivered to vision critical structures with the EP917 plaque. Additionally, our data suggest that, on average, fewer sources and reduced total source strength are associated with use of the Figure 2. Two-dimensional volumetric diagram with prescription isodose lines. The Collaborative Ocular Melanoma Study (COMS) 18-mm plaque (A) has 21 seeds. The EyePhysics 16-14-mm plaque (B) has 17 slots (each slot measures 1.0 5.0 0.8 mm), but only 14 are loaded, wherein all seeds are of the same strength in units of air kerma (U). For both plaque diagrams, isodose lines are in Gray units. Treatment dose is 85 Gy to tumor apex (light blue) delivered in 72 hours.

Volume - Number - 2012 Dosimetric benefit of a new ophthalmic radiation plaque 5 EP917 plaque. This has the potential of reducing costs because individual seed pricing depends somewhat on source strength. In summary, our data indicate that the commercially available EP917 plaque (which has still not received approval from the US Food and Drug Administration) has distinct dosimetric advantages over COMS plaques. For the full range of tumor sizes treated with plaque radiation therapy, the use of EP917 plaques is expected to reduce costs. The EP917 plaques provide for a conformal distribution of radiation with lower exposure to critical ocular structures. Whether these dosimetric benefits translate into clinically significant improved visual outcomes after brachytherapy of posterior uveal melanoma remains to be proved. References 1. Singh AD, Turell ME, Topham AK. Uveal melanoma: trends in incidence, treatment, and survival. Ophthalmology 2011;118:1881-1885. 2. Packer S, Rotman M, Fairchild RG, et al. Irradiation of choroidal melanoma with iodine 125 ophthalmic plaque. Arch Ophthalmol 1980; 98:1453-1457. 3. Earle J, Kline RW, Robertson DM. Selection of iodine 125 for the Collaborative Ocular Melanoma Study. Arch Ophthalmol 1987;105: 763-764. 4. Garretson BR, Robertson DM, Earle JD. Choroidal melanoma treatment with iodine 125 brachytherapy. Arch Ophthalmol 1987;105: 1394-1397. 5. Lommatzsch PK, Werschnik C, Schuster E. Long-term follow-up of Ru-106/Rh-106 brachytherapy for posterior uveal melanoma. Graefes Arch Clin Exp Ophthalmol 2000;238:129-137. 6. Wen JC, Oliver SC, McCannel TA. Ocular complications following I- 125 brachytherapy for choroidal melanoma. Eye 2009;23:1254-1268. 7. Melia BM, Abramson DH, Albert DM, et al. Collaborative ocular melanoma study (COMS) randomized trial of I-125 brachytherapy for medium choroidal melanoma. I. Visual acuity after 3 years COMS report no. 16. Ophthalmology 2001;108:348-366. 8. Finger PT. Radiation retinopathy is treatable with anti-vascular endothelial growth factor bevacizumab (Avastin). Int J Radiat Oncol Biol Phys 2008;70:974-977. 9. Jones R, Gore E, Mieler W, et al. Posttreatment visual acuity in patients treated with episcleral plaque therapy for choroidal melanomas: dose and dose rate effects. Int J Radiat Oncol Biol Phys 2002; 52:989-995. 10. Jensen AW, Petersen IA, Kline RW, et al. Radiation complications and tumor control after 125I plaque brachytherapy for ocular melanoma. Int J Radiat Oncol Biol Phys 2005;63:101-108. 11. Stack R, Elder M, Abdelaal A, et al. New Zealand experience of I125 brachytherapy for choroidal melanoma. Clin Exp Ophthalmol 2005; 33:490-494. 12. Krohn J, Monge OR, Skorpen TN, et al. Posterior uveal melanoma treated with I-125 brachytherapy or primary enucleation. Eye 2008;22: 1398-1403. 13. Saconn PA, Gee CJ, Greven CM, et al. Alternative dose for choroidal melanoma treated with an iodine-125 radioactive plaque: a singleinstitution retrospective study. Int J Radiat Oncol Biol Phys 2010; 78:844-848. 14. Finger PT. Radiation therapy for choroidal melanoma. Surv Ophthalmol 1997;42:215-232. 15. Oliver SC, Leu MY, DeMarco JJ, et al. Attenuation of iodine 125 radiation with vitreous substitutes in the treatment of uveal melanoma. Arch Ophthalmol 2010;128:888-893. 16. Astrahan MA, Luxton G, Pu Q, et al. Conformal episcleral plaque therapy. Int J Radiat Oncol Biol Phys 1997;39:505-519. 17. Puusaari I, Heikkonen J, Kivela T. Effect of radiation dose on ocular complications after iodine brachytherapy for large uveal melanoma: empirical data and simulation of collimating plaques. Invest Ophthalmol Vis Sci 2004;45:3425-3434. 18. http://www.eyephysics.com/ep/epplaques.html. 19. Knutsen S, Hafslund R, Monge OR, et al. Dosimetric verification of a dedicated 3D treatment planning system for episcleral plaque therapy. Int J Radiat Oncol Biol Phys 2001;51:1159-1166. 20. Gragoudas ES, Li W, Lane AM, et al. Risk factors for radiation maculopathy and papillopathy after intraocular irradiation. Ophthalmology 1999;106:1571-1577. discussion 1577-1578.