Application of asi-kvcbct for Volume Assessment and Dose Estimation: An Offline Adaptive Study for Prostate Radiotherapy

Similar documents
CBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.

CALCULATION OF OPTIMAL MARGINS BETWEEN CLINICAL TARGET VOLUME (CTV) AND PLANNING TARGET VOLUME (PTV)

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Impact of internal variations on the dose distribution during the course of radiotherapy of prostatectomy patients

IGRT Solution for the Living Patient and the Dynamic Treatment Problem

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016

Department of Radiotherapy & Nuclear Medicine, National Cancer Institute, Cairo University, Cairo, Egypt.

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

From position verification and correction to adaptive RT Adaptive RT and dose accumulation

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Image Registration for Radiation Therapy Applications: Part 2: In-room Volumetric Imaging

Leila E. A. Nichol Royal Surrey County Hospital

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Fast cine-magnetic resonance imaging point tracking for prostate cancer radiation therapy planning

Presentation Outline. Patient Setup Imaging in RT: Getting the Most Bang for your Buck. Main Errors in RT. Hypothetical Patient Examples

Utilization of cone-beam CT for offline evaluation of target volume coverage during prostate image-guided radiotherapy based on bony anatomy alignment

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

ART for Cervical Cancer: Dosimetry and Technical Aspects

UvA-DARE (Digital Academic Repository) Towards image-guided radiotherapy of prostate cancer Smitsmans, M.H.P. Link to publication

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies

Image-guided adaptive radiation therapy : retrospective study and assessment of clinical workflow

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

A Thesis. entitled. based on CBCT Data Dose Calculation. Sukhdeep Kaur Gill. Master of Science Degree in Biomedical Science

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

Image Fusion, Contouring, and Margins in SRS

8/3/2016. Outline. Site Specific IGRT Considerations for Clinical Imaging Protocols. Krishni Wijesooriya, PhD University of Virginia

Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm.

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

What do we Know About Adaptive Radiotherapy? Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, California

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population

Image Guided Stereotactic Radiotherapy of the Lung

The effect of region extraction accuracy on the dose calculation with CBCT images combined with MSCT images.

RESEARCH ARTICLE. Shanmugam Thirumalai Swamy 1,2 *, Chandrasekaran Anu Radha 2, Gandhi Arun 1, Murugesan Kathirvel 1, Sai Subramanian 1.

Margins in SBRT. Mischa Hoogeman

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Potential systematic uncertainties in IGRT when FBCT reference images are used for pancreatic tumors

Motion gating and tracking techniques: overview and recent developments

The association of the uterine motion with bladder volume during radiotherapy in gynecological malignancies

Dosimetric impact of different bladder and rectum filling during prostate cancer radiotherapy

Margin Reduction in Prostate. Radiotherapy. Richard Oates, A/Prof Michal Schneider Monash Prof Tomas Kron Peter Mac Dr Michael Lim Joon Peter Mac

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

Ritu Raj Upreti, S. Dayananda, R. L. Bhalawat*, Girish N. Bedre*, D. D. Deshpande

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

Cyberknife Stereotactic Treatment

Which Planning CT Should be Used for Lung SBRT? Ping Xia, Ph.D. Head of Medical Physics in Radiation Oncology Cleveland Clinic

Dosimetric consequences of tumor volume changes after kilovoltage cone-beam computed tomography for non-operative lung cancer during adaptive

Knowledge-Based IMRT Treatment Planning for Prostate Cancer: Experience with 101. Cases from Duke Clinic. Deon Martina Dick

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction

Radiobiological Impact of Planning Techniques for Prostate Cancer in Terms of Tumor Control Probability and Normal Tissue Complication Probability

Implementation of advanced RT Techniques

Vaginal Sparing with Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer. Scott Boulet BSc, RT(T)

Evaluation of deformed image-based dose calculations for adaptive radiotherapy of nasopharyngeal carcinoma

Defining Target Volumes and Organs at Risk: a common language

8/1/2017. Clinical Indications and Applications of Realtime MRI-Guided Radiotherapy

1 : : Medical Physics, Città della Salute e della Scienza, Torino, Italy

biij Initial experience in treating lung cancer with helical tomotherapy

Linac Based SBRT for Low-intermediate Risk Prostate Cancer in 5 Fractions: Preliminary Report of a Phase II Study with FFF Delivery

Collapsed Cone Convolution 2D illustration

Work partially supported by VisionRT

MRI Applications in Radiation Oncology:

Volumetric-based image guidance is superior to marker-based alignments for stereotactic body radiotherapy of prostate cancer

Intensity-Modulated and Image- Guided Radiation Treatment. Outline. Conformal Radiation Treatment

SUPERIORITY OF A REAL TIME PLANNING TECHNIQUE OVER IMAGE GUIDED RADIATION THERAPY FOR THE TREATMENT OF PRIMARY PROSTATE CANCERS

OPTIMIZATION OF COLLIMATOR PARAMETERS TO REDUCE RECTAL DOSE IN INTENSITY-MODULATED PROSTATE TREATMENT PLANNING

Organ Contour Adaptor to create new structures to use for adaptive radiotherapy of cervix cancer using Matlab Bridge and 3DSlicer / SlicerRT

Radiotherapy Advances

Chapters from Clinical Oncology

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

MRI Based treatment planning for with focus on prostate cancer. Xinglei Shen, MD Department of Radiation Oncology KUMC

Pitfalls in SBRT Treatment Planning for a Moving Target

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Tissue Density Mapping of Cone Beam CT Images for Accurate Dose Calculations

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Real-time tumor tracking during VMAT radiotherapy treatments based on 2D/3D registration using CBCT projections

Feasibility study on effect and stability of adaptive radiotherapy on kilovoltage cone beam CT

Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

7/16/2009. Cost-benefit, QALYs and the Value of a Statistical Life

Role of adaptive radiation therapy for pediatric patients with diffuse pontine glioma

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

X-Ray & CT Physics / Clinical CT

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

100 Paluska et al.: Utilization of CBCT for prostate IGRT 100

Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

Transcription:

DOI:10.31557/APJCP.2019.20.1.229 Volume and Dose Assessment on Adapted CT RESEARCH ARTICLE Editorial Process: Submission:08/10/2018 Acceptance:01/05/2019 Application of asi-kvcbct for Volume Assessment and Dose Estimation: An Offline Adaptive Study for Prostate Radiotherapy Arun Gandhi 1,2, Subramani Vellaiyan 2,3 *, Shanmuga Subramanian 1,2, Shanmugam Thirumalai Swamy 1, Kala Subramanian 1, Anantharaman Ayyalusamy 1,2 Abstract Objective: The purpose of this study is to develop a method to estimate the dose using amorphous silicon detector panel cone beam computed tomography (asi-kvcbct) for the OARs and targets in prostate radiotherapy and to compare with the actual planned dose. Methods: The asi-kvcbct is used widely in radiotherapy to verify the patient position before treatment. The advancement in asi-kvcbct combined with adaptive software allows us to verify the dose distribution in daily acquired CBCT images. CBCT images from 10 patients undergoing radical prostate radiotherapy were included in this study. Patients received total dose of 65Gy in 25 fractions using volumetric modulated arc therapy (VMAT). asi-kvcbct scans were acquired before daily treatment and exported to smart adapt software for image adaptation. The planning CT is adapted to daily asi-kvcbct images in terms of HU mapping. The primary VMAT plans were copied on to the adapted planning CT images and dose was calculated using Anisotropic Analytic Algorithm (AAA). The DVH is then used to evaluate the volume changes of organs at risk (OAR), the actual dose received by OARs, CTV and PTV during a single fraction. Results: The normalized volume of the bladder and rectum ranged from 0.70 1.66 and 0.70 1.16 respectively. The cumulative mean Sorensen Dice coefficient values of bladder and rectum were 0.89±0.04 and 0.79±0.06 respectively. The maximum dose differences for CTV and PTV were 2.5% and -4.7% and minimum were 0.1% and 0.1% respectively. Conclusion: The adapted planning CT obtained from daily imaging using asi-kvcbct and SmartAdapt can be used as an effective tool to estimate the volume changes and dose difference in prostate radiotherapy. Keywords: asi-kvcbct- adapted planning CT- prostate radiotherapy- volume change- dose difference Asian Pac J Cancer Prev, 20 (1), 229-234 Introduction Multiple trials have shown that hypo fractionation can shorten treatment schedule for patients with prostate cancer and might also allow us to enhance the therapeutic ratio even further (Dearnaley et al., 2016). Recent technologic improvements have allowed hypo fractionated radiation treatment to be administered with greater precision, improved safety, and an enhanced potential for disease control. The treatment efficacy depends on the patient setup error and interfraction motion throughout the whole treatment. Volume changes in the bladder and rectum can cause prominent variations in the prescribed dose vs the actual dose received in prostate radiotherapy. Inter-fractional prostate motions are corrected through pre-treatment imaging during each fraction. Advancements in treatment delivery techniques like volumetric-modulated arc therapy have improved the sparing of critical structures and reduced the treatment delivery time, thereby limiting intrafraction motion errors (Anantharaman et al., 2016). In this scenario, the development in adaptive algorithms enables us to execute adaptive radiotherapy in an efficient manner using daily CBCT images to account for the intrafraction volume changes of the critical organs in prostate patients. In recent times, asi-kvcbct is a major tool to obtain the 3D image of the patient before treatment. These multidirectional images provide sufficient information to reconstruct patient anatomy in three dimensions, including cross- sectional, sagittal and coronal planes. There is a possibility of changes in the anatomy of bladder, rectum and bowel during the entire course of the treatment for prostate. Thus, position of the PTV may change due to anatomical changes of OARs. Several authors found that the volume change in prostate by an average of +/- 10% during the course of radiation therapy, while bladder and 1 Department of Radiation Oncology, Yashoda Hospital, Hyderabad, 2 Research and Development Centre, Bharathiar University, Coimbatore, 3 Department of Radiation Oncology, All India Institute of Medical Sciences, New Delhi, India. *For Correspondence: vsmani_arun@yahoo.com Asian Pacific Journal of Cancer Prevention, Vol 20 229

Arun Gandhi et al rectal volumes varied by +/- 30% (Roeske et al., 1995). The planning CT data are ideally not representative of the anatomy present during treatment. In principle, dose calculation should be done on asi-kvcbct image but the challenges posed due to the artifacts and non-standard HU values prevent us from using these images for dose calculation. In the current situation, many systems are available to estimate the dose and assess the volume changes on adapted planning CT. Several authors had investigated the dose calculation accuracy on CBCT and compared with planning CT (Richter et al., 2008). CT number mapping is a well-known method to perform adaptive radiotherapy and the dose was calculated on CBCT (Hu et al., 2010). In this study, asi-kvcbct had been used to estimate the dose and asses the volume change during the prostate radiotherapy with the help of commercially available image adaptive system. Materials and Methods A. Patient selection 10 prostate cancer patients were included in this study with low, intermediate risk (stage T1-2N0M0) had radiotherapy (RT) with total prescribed dose of 65 Gy by 2.6 Gy per fraction. Patients were immobilized with vacuum cushion and thermoplastic mask. For planning CT simulation, the patients were asked to drink two glass of water about 20 mins before the acquisition CT and the same were maintained before each fractional treatment. Planning CT acquisition performed with the slice thickness of 3 mm along with Contrast enhanced CT for better delineation of anatomical structures. For all 10 patients, the clinical target volume (CTV) was defined as the prostate gland and seminal vesicle on the CT scan. The planned target volume (PTV) was defined as CTV plus 5 mm in the posterior direction and 8 mm for all other directions. The Bowel, rectum, bladder, femoral heads were also contoured for each patient. The VMAT plan optimization was performed with 6MV photon and dose calculation (Anisotropic Analytic Algorithm) was carried out on the planning CT in the Eclipse v.11 treatment planning system (Varian Medical Systems, Palo Alto, CA) with the calculation grid size of 2.5 mm. B. asi-kvcbct Acquisition The patient position verification consists of daily 2D Kv orthogonal imaging for bony landmark verification and asi-kvcbct for 3D anatomical verification. The following parameters of asi-kvcbct were used for 3D imaging: Kv: 125kVp, high-resolution reconstruction: 512 X 512 pixel, with a slice thickness of 3 mm; Gain mode: Dual gain; Filter: half fan bow tie filter. Detector: Amorphous silicon flat panel imager Active imaging area: 40 x 30 cm2 Resolution: 1,024 x 768 pixels; Max frame acquisition rate: 9.574 frames/second; Energy range: 40-150 kvp; Spatial resolution: 0.19 mm. Daily asi-kvcbct imaging protocol were followed for prostate radiotherapy. 230 Asian Pacific Journal of Cancer Prevention, Vol 20 C. Adaptive image registration To estimate the changes in volume and daily dose to the bladder and rectum, the planning CT is rigidly auto-aligned and deformed to the asi-kvcbct. Image adaptation mainly relies on Deformable Image Registration (DIR) and the same have been discussed in many literatures (Hou et al., 2011). SmartAdapt system (Varian Medical Systems, Palo Alto, CA) system has the ability to perform DIR and map the HU values from planning CT to CBCT for better dose calculation accuracy (Anantharaman et al., 2016) which uses accelerated demons algorithm proposed by Wang et al., (2005). The adaptive image registration consists of two steps: 1) possible bony landmark registration of CBCT with planning CT through a rigid registration algorithm. 2) Deformable image registration of planning CT to asi-kvcbct to deform the contours using the deformation vector field determined by the SmartAdapt system. The deformed structures were checked carefully for its changes in adapted planning CT. The adapted planning CTs along with contours were exported to Eclipse planning system for contour evaluation and dose calculation. D. Volume assessment and Dose estimation The Sorensen Dice coefficient (SDC) is used to compare the similarity of two contoured sets. The volume changes in adapted planning CT were compared with initial planning contours. In this study SDC is used as a volume-based similarity index to evaluate the volume changes quantitatively. The SDC is calculated as follows, where Vp and Va represents initial planned volume and adapted volume respectively. SDC = To estimate the dose of the day all planning parameters were kept the same, including field size, beam arrangement, dose fluence map, planned monitor units and dose calculations were performed. AAA algorithm was used to calculate the dose on adapted planning CT with the calculation grid size of 2.5mm. Dose estimation was computed in dose volume histogram option in Eclipse. The dose-volume parameters were calculated in terms of dose received by 15% of volume (V15%), 25 %( V25%), 35 %( V35%) and 50 %( V50%) (RTOG report no: 0126) for rectum and bladder. DVH analysis of Clinical Target Volume (CTV) and Planning Target Volume (PTV) were performed using doses received by 99% (D99%), 95% (D95%) and 1% (D1%) of volume. Results 2 Vp Va Vp + Va A. Volume assessment Weekly asi-kvcbct images for 10 patients were first analyzed and later adapted to planning CT. For each patient the volumes of bladder and rectum were assessed. Figure 1 shows the volume projection in an axial section of bladder and rectum on the weekly adapted planning CT for a patient. Volumes of bladder and rectum were normalized corresponding to initial planning volumes as shown in

Table 1. The SDC Values for Bladder and Rectum P. No SDC DOI:10.31557/APJCP.2019.20.1.229 Volume and Dose Assessment on Adapted CT 1 2 3 4 5 6 7 8 9 10 B 0.88± 0.023 0.82± 0.071 0.91± 0.021 0.89± 0.015 0.93± 0.010 0.94± 0.020 0.91± 0.035 0.84± 0.063 0.85± 0.021 0.95± 0.035 R 0.76± 0.012 0.80± 0.017 0.64± 0.010 0.77± 0.002 0.85± 0.019 0.88± 0.020 0.78± 0.041 0.86± 0.012 0.76± 0.017 0.79± 0.024 SDC, Sorensen Dice coefficient; P.No, Patient number; B, Bladder; R, Rectum Table 2. The Average Cumulative Dose Differences for CTV CTV Patient No D 95% D 99% D 1% Planned Calculated Variation Planned Calculated Variation Planned Calculated Variation (cgy) (cgy) (%) (cgy) (cgy) (%) (cgy) (cgy) (%) 1 63.6 63.8 0.3 63.1 63.2 0.1 66.7 67.0 0.4 2 65.2 66.4 1.8 64.7 65.7 1.6 67.6 69.3 2.5 3 65.9 66.0 0.2 65.6 65.5-0.2 69.1 69.4 0.4 4 63.9 64.3 0.7 63.4 63.6 0.3 66.6 67.1 0.8 5 64.0 64.4 0.6 63.4 63.1-0.5 66.7 66.3-0.6 6 66.2 66.0-0.3 63.2 63.8 0.9 69.2 69.7 0.7 7 66.5 66.0-0.8 63.4 63.1-0.5 69.9 69.7-0.3 8 65.3 64.4-1.4 64.6 64.0-0.9 67.8 66.9-1.3 9 66.4 66.8 0.6 63 62.5-0.8 66.9 67.4 0.7 10 66.3 66.4 0.2 64.2 63.3-1.4 67.0 66.8-0.3 CTV, Clinical target volume; D95%, D99%, D1%, dose received by 95%, 99%, 1% of volume Figure 1. The Volume Projection in an Axial Section of Bladder and Rectum on the Weekly Adapted Planning CT Figure 2. Normalized Bladder Volume Corresponding to Initial Planning Volumes Asian Pacific Journal of Cancer Prevention, Vol 20 231

Arun Gandhi et al Table 3. The Average Cumulative Dose Differences for PTV PTV Patient No D95% D99% D1% Planned Calculated Variation Planned Calculated Variation Planned Calculated Variation (cgy) (cgy) (%) (cgy) (cgy) (%) (cgy) (cgy) (%) 1 63.1 62.8-0.5 61.8 60.1-2.7 66.9 67.1 0.3 2 63.4 63.1-0.5 61.9 60.5-2.3 67.7 69.5 2.7 3 63.9 63.3-1 59 59.4 0.7 69.3 69.4 0.1 4 62.8 61.3-2.5 61.4 58.5-4.7 66.7 67.4 1 5 62.7 62.9 0.3 61.4 60.2-2 66.5 66.8 0.5 6 64.0 63.5-0.8 62 62.2 0.3 69.9 68.7-1.7 7 63.0 64.0 1.6 60.7 60.3-0.7 68.9 66.7-3.2 8 63.3 64.2 1.4 61.7 60.8-1.5 67.4 67.1-0.4 9 63.0 64.9 3 60.2 59.5-1.2 66.9 69.2 3.4 10 64.2 63.0-1.9 61.7 60.0-2.8 67.1 66.5-0.9 PTV, Planning target volume Table 4. Dose Difference of Bladder Due to the Volume Change Bladder Patient No 1 2 3 4 5 6 7 8 9 10 V 15% Planned (cgy) 42.5 65.1 58.5 52 52.6 56.7 63.2 52.3 58.7 60.6 Calculated (cgy) 47.6 59.5 58 54.3 53.2 52.8 58.9 61.2 60.4 64.2 Dose Difference (%) 12.1-8.6-0.9 4.5 1.1-6.9-6.8 17.0 2.9 5.9 V 25% Planned (cgy) 26.8 59.6 43.0 44.9 44.3 42.5 49.7 49.8 42.6 45.8 Calculated (cgy) 31.9 46.4 41.9 46.3 42.2 44.2 50.3 56.5 47.8 39.4 Dose Difference (%) 19.1-22.1-2.5 3.0-4.7 4.0 1.2 13.5 12.2-14 V 35% Planned (cgy) 16.2 47.6 31.2 39.5 35.1 30.2 40.2 42.1 28.7 30.7 Calculated (cgy) 20.5 39.4 29.5 40.7 38.2 32.3 37.4 40.7 33.5 31.8 Dose Difference (%) 26.3-17.2-5.3 3.0 8.8 7.0-7.0-3.3 16.7 3.6 V 50% Planned (cgy) 7.2 36.6 19.3 32.5 32.6 16.8 32.1 24.1 15.5 19.2 Calculated (cgy) 9.5 32.8 18.2 33.9 30.5 15.6 27.5 33.6 16.2 22.5 Dose Difference (%) 31.4-10.5-5.9 4.3-6.4-7.1-14.3 39.4 4.5 17.2 Figures 2 and 3. The normalized volume of the bladder and rectum ranged from 0.70 1.66 and 0.70 1.16 respectively. From this study we found that the bladder volume changes were comparatively more than the rectum. The SDC between the adapted and primarily delineated contours were determined for bladder and rectum shown in Table Figure 3. Normalized Rectum Volume Corresponding to Initial Planning Volumes 232 Asian Pacific Journal of Cancer Prevention, Vol 20

DOI:10.31557/APJCP.2019.20.1.229 Volume and Dose Assessment on Adapted CT Figure 4. Dose Distribution and DVH Spread on Planning CT and Adapted Planning CT of a Prostate Patient Table 5. Dose Difference of Rectum due to the Volume Change Rectum Patient No 1 2 3 4 5 6 7 8 9 10 V 15% Planned (cgy) 53.3 56.6 58.4 55.9 56 52.5 58.2 50.7 45.2 50.1 Calculated (cgy) 56.7 61.2 60.9 48.7 55.2 58.2 54.5 45.6 49.7 46.5 Dose Difference (%) 6.3 8.1 4.3-12.9-1.4 10.9-6.4-10.1 10-7.2 V 25% Planned (cgy) 38.6 42 40.8 43.7 35.8 41.5 48.7 44.8 38.2 39.8 Calculated (cgy) 40.6 43.8 41.7 37.3 30.9 42.3 46.8 40.2 36.5 36.7 Dose Difference (%) 5.2 4.2 2.3-14.7-13.7 1.9-3.9-10.3-4.5-7.8 V 35% Planned (cgy) 29.2 28.7 29 35.1 27.5 36.6 30.2 34.7 21.6 25.2 Calculated (cgy) 29.9 28.3 28.9 32.5 28.4 33.2 27.5 29.8 25.6 21.8 Dose Difference (%) 2.5-1.5-0.5-7.5 3.3-9.3-8.9-14.1 18.5-13.5 V 50% Planned (cgy) 18.7 16.8 20.3 27.6 25.1 20.8 15.6 20.9 18.7 14.6 Calculated (cgy) 18.4 16.7 21 21.7 25.6 18.6 17.8 16.5 19.6 16.3 Dose Difference (%) -1.4-0.7 3.3-21.5 2-10.6 14.1-21.1 4.8 11.6 1. The cumulative mean SDC values of bladder and rectum were 0.89±0.04 and 0.79±0.06 respectively. B. Dose estimation Figure 4 represents the dose distribution and DVH comparison on planning CT and adapted planning CT of a single patient. The average cumulative dose differences in DVH parameters for CTV and PTV are shown in Tables 2 and 3. The maximum dose differences for CTV and PTV were 2.5% and -4.7% and minimum were 0.1% and 0.1% respectively. Tables 4 and 5 represent the dose difference of bladder and rectum due to the volume change. The average cumulative dose differences were found to be 2.9%±12.9 for bladder and -2.3%±9.5 for rectum. Discussion This study is intended to analyze the application of asi-kvcbct for volume changes of bladder and rectum and dose estimation for the same for prostate radiotherapy. Many literatures showed that the volume, position and shape of the bladder and rectum vary throughout the course of prostate radiotherapy treatment (Li et al., 2011). The Asian Pacific Journal of Cancer Prevention, Vol 20 233

Arun Gandhi et al SmartAdapt system is already validated for deformable image registration by (Ramadaan et al., 2015). Pretreatment bladder and rectum protocol helps to maintain the volume as per planning CT. However, from our results we observed that the changes in volume of bladder are found to be more compared with rectum volume changes. For some patients the bladder filling protocol have been changed during the course of treatment itself after analyzing the volume changes on daily asi-kvcbct to maintain as per planning CT. Patients are being advised to stay in the waiting room for a prescribed time after drinking prescribed amount of water wherein the ambient temperature is maintained every day. This practice is useful to maintain the bladder filling changes due to room temperature setting. When the rectum of patient is filled with gas and feces, it enlarges and encroach the PTV which elevates the dose to rectum and reduces the dose to PTV. The same have been observed by Padhani et al., (1999) in a study and they concluded that the patients should be advised to clear their rectum prior to radiotherapy to reduce rectal distension. If rectum and bladder filling are not sufficient or too filled, the treatment delivery should be aborted. Table 1 show that the SDC fluctuate for each patient and indicates that there is volume changes of OARs. The minimum and maximum SDC for bladder is 0.82± 0.071, 0.95±0.035 and for rectum is 0.64±0.010, 0.88±0.020 respectively. In our study, changes in CTV dose coverage for most patients were not prominent and for PTV it is within ±3.5%. This is because the dose to PTV is more influenced by the volume changes of OARs than CTV. Even though the patient position error can be corrected, deformation of bladder and rectum and its consequences on dose distribution should be considered. The standard deviation of dose difference for bladder and rectum is high because of the dose difference in smaller dose range of V50%. However, this approach is laborious to estimate the volume changes and dose difference every day because of the time taken for image registration, image adaptation, verification, dose calculation and analysis. That is the reason, this study limits itself to check the dose and volume changes weekly once rather than daily. Our results suggest that 3 dimensional imaging is a necessary tool to evaluate the OAR volumes although the filling protocol followed for bladder and rectum. In Conclusion, this study was performed using asi-kvcbt to estimate the volume changes and dose difference in prostate radiotherapy. The adapted planning CT resulted from asi-kvcbct and SmartAdapt system is an effective tool to estimate the volume changes and dose difference. Daily adaptive planning CT can estimate the cumulative dose variation during the entire course of treatment. Faster asi-kvcbct acquisition and enhanced reconstruction algorithm can estimate the dose difference with high accuracy. References Anantharaman A, Subramani V, Subramanian S, et al (2016). Feasibility of offline head and neck adaptive radiotherapy using deformed planning CT electron density mapping on weekly cone beam computed tomography. Br J Radiol, 90, 20160420. Dearnaley D, Syndikus I, Mossop H, et al (2016). Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: 5-year outcomes of the randomised, non-inferiority, phase 3 CHHiP trial. Lancet Oncol, 17, 1047-60. Hou J, Guerrero M, Chen W, D Souza WD (2011). Deformable planning CT to cone-beam CT image registration in head-and-neck cancer. Med Phys, 38, 2088 94. Hu W, Ye J, Wang J, Ma X, Zhang Z (2010). Use of kilovoltage X-ray volume imaging in patient dose calculation for head-and-neck and partial brain radiation therapy. Radiat Oncol, 19, 29. Li T, Thongphiew D, Zhu X,et al (2011) Adaptive prostate IGRT combining online re-optimization and re-positioning: a feasibility study. Phys Med Biol, 56, 1243 58. Nuver TT, Hoogeman MS, Remeijer P, et al (2007) An adaptive offline procedure for radiotherapy of prostate cancer. Int J Radiat Oncol Biol Phys, 67, 1559 67. Piziorska M, Kukołowicz P, Zawadzka A, et al (2012), Adaptive off-line protocol for prostate external radiotherapy with cone beam computer tomography. Strahlentherapie und Onkologie, 11, 1003 9 Ramadaan IS, Peick K, Hamilton DA, et al (2015). Validation of Varian s SmartAdapt deformable image registration algorithm for clinical application. Radiat Oncol, 10, 1. Richter A, Hu Q, Steglich D, et al (2008). Investigation of the usability of conebeam CT data sets for dose calculation. Radiat Oncol, 3, 7215 23. Roeske JC, Forman JD, Mesina CF, et al (1995). Evaluation of changes in the size and location of the prostate, seminal vesicles, bladder, and rectum during a course of external beam radiation therapy. Int J Radiat Oncol, 33, 1321 9. Wang H, Dong L, O Daniel J, et al (2005). Validation of an accelerated demons algorithm for deformable image registration in radiation therapy. Phys Med Biol, 50, 2887 905. This work is licensed under a Creative Commons Attribution- Non Commercial 4.0 International License. Statement conflict of Interest Authors have no conflict of Interest. 234 Asian Pacific Journal of Cancer Prevention, Vol 20