Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis

Similar documents
In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

Dental Research Journal

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Australian Dental Journal

Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA

EVALUATION OF STRESS DISTRIBUTION AND MICROMOTION OF DENTAL IMPLANT: IN VIVO CASE STUDY

Finite Element Analysis of Dental Implant as Orthodontic Anchorage

ABSTRACT INTRODUCTION

Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring

Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element analysis

Effects of Mesiodistal Inclination of Implants on Stress Distribution in Implant-Supported Fixed Prostheses

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study

MECHANISM OF LOAD TRANSFER ALONG THE BONE- DENTAL IMPLANT INTERFACE

INFLUENCE OF CANTILEVER ON STRESS DISTRIBUITION IN BONE AROUND DENTAL IMPLANTS: A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS

BIOMECHANICS AND OVERDENTURES

A mimic osseointegrated implant model for three-dimensional finite element analysis

Objective: The occlusal patterns are key requirements for the clinical success of oral

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

IJPCDR INTRODUCTION ABSTRACT

Comparative study of single-thread, double-thread, and triple-thread dental implant: a three-dimensional finite element analysis

Components Of Implant Protective Occlusion A Review

Stress in the Mandible with Splinted Dental Implants caused by Limited Flexure on Mouth Opening: An in vitro Study

Objective: This paper focused on optimal stress distribution in the mandibular bone

Peri-Implant Stress Analysis of Immediate Loading and Progressive Loading Implants in Different /jp-journals

INTRODUCTION. Graduate student, Department of Prosthodontics and Dental Research Institute, Seoul National University 2

The fundamental biomechanics of oral implants. Dr. Szűcs Attila

Bicortically Stabilized Implant Load Transfer

A review: Mechanical Design of Dental Implants to Reduce Stresses around Implant-Bone Interface

Evaluation of Effect of Connector Designs in Implant Tooth-supported Fixed Partial Denture: A Two-dimensional Finite Element Analysis

JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH

Since the introduction of the Branemark

Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw Fracture in Dental Implant System

BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA

STUDYING THE EFFECT OF THE PRELOAD INDUCED BY SCREW TIGHTENING ON STRESS DISTRIBUTION OF A DENTAL IMPLANT

Shrikar R. Desai, I. Karthikeyan, and Rika Singh. 1. Introduction

Performance Assessment of Various Molar Crown Designs

Three dimensional finite element analysis of a novel osteointegrated dental implant designed to reduce stress peak of cortical bone

Available online at ScienceDirect

FINITE ELEMENT ANALYSIS OF PERIODONTAL STRESSES IN FIXED PROSTHODONTICS

Mechanical and technical risks in implant therapy.

Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis.

International Journal of Engineering Research and General Science Volume 3, Issue 1, January-February, 2015 ISSN

IS THE PALATE AN OPTIMAL SITE FOR MINISCREW PLACEMENT?

Anuradha Borse 1* Rubina Tabassum 2 Omkar Shetty 3

Biomechanical factors play an important role in

Recently, promising results have been observed when

Influence of Implant Position on Stress Distribution in Implant-Assisted Distal Extension Removable Partial Dentures: A 3D Finite Element Analysis

Preeti Satheesh Kumar, 1 Kumar K. S. Satheesh, 1 Jins John, 1 Geetha Patil, 2 and Ruchi Patel Introduction

Dental implants have been

SHAPE OPTIMIZATION AND TOLERANCE ANALYSIS OF DENTAL IMPLANTS BY MEANS OF VIRTUAL MODELS

Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect

High Crown to Implant Ratio as Stress Factor in Short Implants Therapy

Mechanical and technical risks in implant therapy.

Sho Kayumi 1,2, Yoshiyuki Takayama 1*, Atsuro Yokoyama 2 and Nana Ueda 1

Rehabilitating a Compromised Site for Restoring Form, Function and Esthetics- A Case Report

Implant Occlusion. Dora Kompotiati Boris Stein David Wilson

Comparison of Platform Switched and Sloping Shoulder Implants on Stress Reduction in various Bone Densities: Finite Element Analysis

Comparison of Short and Standard Implants in the Posterior Mandible: A 3D Analysis Using Finite Element Method

Modelling of temporomandibular joint and FEM analysis

The Use of Alpha-Bio Tec's Narrow NeO Implants with Cone Connection for Restoration of Limited Width Ridges

Dental Research Journal

MULTIDIRECTIONAL APPROACH OF ORAL REHABILITATION WITH IMPLANTS IN A PATIENT WITH LIMITED MOUTH OPENING: A CASE REPORT

Tilted or Parallel Implant Placement in the Completely Edentulous Mandible? A Three-Dimensional Finite Element Analysis

Department of Prosthodontics and Periodontics, Piracicaba Dental School, Campinas State University, Piracicaba, SP, Brazil

Journal of Dental Research, Dental Clinics, Dental Prospects. Original Article

Osseointegrated implant-supported

Occlusal Principles and Considerations for Implants: An Overview

FEM analysis of the mandibular first premolar with different post diameters

synocta Meso abutment for cement-retained restorations PROSTHETICS Step-by-step instructions

A retrospective study on separate single-tooth implant restorations to replace two or more consecutive. maxillary posterior teeth up to 6 years.

Journal of Dental Research, Dental Clinics, Dental Prospects. Original Article

Fixed Partial Dentures /FPDs/, Implant Supported. in implant prosthodontics

Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone

Objectives: In some clinical situations, dentists come across partially edentulous patients,

Rehabilitation of atrophic partially edentulous mandible using ridge split technique and implant supported removable prosthesis

Effect of diameter and length on stress distribution of the alveolar crest around immediate loading implants.

Bone Reduction Surgical Guide for the Novum Implant Procedure: Technical Note

Finite Element Study of Three Different Treatment Designs of a Mandibular Three Implant-Retained Overdenture

Stress Analysis of an Endosseus Dental Implant by BEM and FEM

The first molars are the first permanent teeth to erupt

Design and Analysis of Dental Surgical Fixture

EFFECTIVE DATE: 04/24/14 REVISED DATE: 04/23/15, 04/28/16, 06/22/17, 06/28/18 POLICY NUMBER: CATEGORY: Dental

Compressive Stress, Shear Stress, and Displacement Study on Different Structured Dental Implant: 3-Dimensional Finite Element Analysis

Three-dimensional optimization and sensitivity analysis of dental implant thread parameters using finite element analysis

Samantha W. Chou, D.M.D N. Southport Ave. Chicago, Illinois Phone: Fax:

Influence of Bone Implant Contact on Biomechanical performance of Short Implant Placed in Atrophic Posterior Mandible

Use of Implant Retained Overdenture in Atrophic Mandible - A Case Report

Occlusion & Prosthodontics

Surveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1

Immediate Loading of a Multi-Threaded Implant

Summer Workshop of Applied Mechanics. Influence of implant length and diameter on stress distribution

Influence of prosthesis type and material on the stress distribution in bone around implants: A 3-dimensional finite element analysis

Influence of Bone Implant Contact on Biomechanical Performance of Short Implant Placed in Atrophic Posterior Mandible

Student Presentation # 3 Biomechanics

Osseointegrated dental implant treatment generally

Retreatment: Fractured Implants Due To Biomechanical Overload

Finite Element Analysis of the Effect of Superstructure Materials and Loading Angle on Stress Distribution around the Implant

THE BIOMECHANICAL EFFECT OF SUPRASTRUCTURE LOADING AND MATERIALS ON STRESS STATE OF IMPLANT SUPPORTED SINGLE RESTORATIONS 3D FINITE ELEMENT ANALYSIS

BIOMECHANICAL ANALYSIS OF THE STRESSES GENERATED BY DIFFERENT DISOCCLUSION PATTERNS IN AN IMPLANT-SUPPORTED MANDIBULAR COMPLETE DENTURE

Transcription:

J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 DOI 10.1007/s13191-012-0189-7 ORIGINAL ARTICLE Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis K. Premnath J. Sridevi N. Kalavathy Prakash Nagaranjani M. Ranganath Sharmila Received: 16 July 2012 / Accepted: 8 October 2012 / Published online: 26 October 2012 Ó Indian Prosthodontic Society 2012 Abstract A key factor for the success or failure of a dental implant is the manner in which stresses are transferred to the surrounding bone. This depends on the type of loading, bone implant interface, the shape and characteristics of the implant surface and the quality and quantity of the surrounding bone. This study was done to evaluate the pattern of stress distribution with two different implant designs in four different densities of bone using 3D finite element analysis. Graphic pre-processing software Ansys version 10 was used for creating the geometric configuration of a section of the mandible with a missing first molar. Eight 3D models of this section restored with implantsupported all ceramic crowns were created. Four of these models were created to simulate a single threaded implant placed in four different densities of bone (D1, D2, D3 and D4). The other four models were created to simulate a single cylindrical implant placed in four different densities of bone (D1, D2, D3, and D4). The Poisson s ratio (l) and Young s modulus (E) of elasticity of the material were incorporated into the model. An average vertical load of 400 N was applied on the occlusal surface of the first molar between the buccal cusp, central fossa and the marginal ridge. Maximum Von Mises stresses in all the eight models were observed at the crestal region or neck of the implant. K. Premnath J. Sridevi N. Kalavathy M. R. Sharmila Department of Prosthodontics, D.A.P.M.R.V. Dental College, Bangalore, India K. Premnath (&) 128, Reservoir Street, K.R. Road, Basavangudi, Bangalore 560004, India e-mail: malarprem@yahoo.com P. Nagaranjani GDCRI, Bangalore, India The stresses observed were more for the threaded implants in all the four densities of bone when compared to that of the cylindrical implants. The study concluded that the cylindrical implant design was more favorable in softer bone than the threaded implant design. Keywords Finite element analysis Finite element models Bone density Threaded implant Cylindrical implant Introduction The application of dental implants for Prosthodontics reconstruction can be traced back to ancient Egypt where sea shells were hammered into human jaw bone to replace missing teeth [1]. Since the late 1960s when the dental implants were introduced for the rehabilitation of completely edentulous patients, an awareness and subsequent demand for this form of therapy has increased [2]. A key factor for the success or failure of a dental implant is the manner in which stresses are transferred to the surrounding bone. This depends on the type of loading, bone implant interface, the shape and characteristics of the implant surface and the quality and quantity of the surrounding bone [3]. The interrelationship between the bone quality, quantity and the design of the implant play a vital role for clinical success. A compromise in any of these factors will often lead to implant failure [4 6]. The density of available bone in an edentulous site is the determining factor in treatment planning, and will determine implant design, surgical approach, healing time and if initial progressive bone loading is feasible during prosthetic reconstruction [2]. The various implant designs like the threaded, cylindrical or the tapered design have been shown to have a

556 J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 profound influence on implant biomechanics and stress distribution in the surrounding bone. Finite element analysis has been used in the field of medicine and dentistry since many decades. It can simulate the interaction phenomena between implants and the surrounding tissues [7, 8]. The effect of different implant designs on stress distribution in different bone densities have not been studied extensively using finite element analysis. Hence this study was done to evaluate the pattern of stress distribution with two different implant designs in four different densities of bone using 3D finite element analysis. The objectives of the study were: to analyze the pattern of stress distribution around a threaded implant and a cylindrical implant when placed in four different densities of bone and also to determine the most favorable implant design for the different bone densities. Methodology This study was conducted using finite element models to evaluate the pattern of stress distribution in a mandibular section of bone with a missing first molar restored with an implant-supported all ceramic crown. Fig. 1 Graphic representation of model (T) Finite Element Model Graphic pre-processing software Ansys version 10 was used for creating the geometric configuration of a section of the mandible with a missing first molar. Eight 3D models of this section restored with implant-supported all ceramic crowns were created. Four of these models were created to simulate a single threaded implant placed in four different densities of bone (D1, D2, D3 and D4) [Model-T, Figs. 1, 2, with elements 150,802 and nodes 28,298]. The other four models were created to simulate a single cylindrical implant placed in four different densities of bone (D1, D2, D3, and D4) [Model-C, Figs 3, 4, with elements 247,975 and nodes 46,527], thereby totally eight 3D models were created. The average height of the mandibular section in the first molar region was 28 mm and width was 12 mm. The dimension of the standard threaded implant was 4 9 10 mm (Innova implant system) and that of the standard cylindrical implant was 4.1 9 9 mm (Innova implant system). Material Properties All materials used in the model were considered to be homogenous, isotropic and linear elastic. The Poisson s ratio (l) and Young s modulus (E) of elasticity of the material were incorporated into the model as shown in Table 1 [2, 3, 9]. Fig. 2 Meshed model (T) The finite element model was divided into small elements. Each element was considered to be interconnected at a number of discrete points called nodes. Each model

J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 557 Table 1 Material properties of finite element model [2, 3, 9] Material Modulus of elasticity (GPa) Poisson s ratio (l) Porcelain 82.8 0.35 Cancellous bone D1 9.5 0.3 D2 5.5 D3 1.6 D4 0.69 Cortical bone 14.8 0.3 Titanium (implant, 110 0.35 abutment) Mucosa 10 0.40 Loading and Boundary Conditions Fig. 3 Graphic representation of model (C) Based on previous studies pertaining to masticatory loads, an average vertical load of 400 N was applied [10] on the occlusal surface of the first molar between the buccal cusp, central fossa and the marginal ridge [11]. A support was provided at the inferior surface of the model and also at the distal ends of the mandibular section to simulate the action of the muscles and ligaments [12]. The Von Mises stresses were analyzed in all the eight models at the following sites: (a) (b) (c) (d) Cancellous bone Cortical bone Crown Implant bone interface Results Maxmium Von Mises Stress in the Different Bone Densities with the Threaded Implant (Fig. 5) The height of each bar depicts the maximum stress level. The color of the bar denoted the different models or bone densities. Fig. 4 Meshed model (C) was meshed by elements defined by ten nodes and three degree of freedom in tetrahedral nodes. The displacement of each of these nodes was calculated to determine the maximum Von Mises stresses throughout the structure. 1. The stress level was observed least in cancellous bone and maximum in the implant and considerably higher in the crown. 2. The heights of the bars in the crown region were almost equal indicating that the decrease in bone density does not affect the stress level, similar interpretation holds good for cancellous bone. 3. The heights of the bars in cortical bone and in the implant increases corresponding to the bone density

558 J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 Fig. 5 Maximum Von Mises stress of threaded implant design Fig. 6 Maximum Von Mises stress of cylindrical implant design Table 2 Maximum Von Mises stress (MPa) on and around threaded implant in different densities of bone Type of bone Crown Cancellous bone Cortical bone Implant T1 73.092 7.353 60.0 102.55 T2 73.003 5.875 63.33 106.41 T3 76.709 8.939 74.36 120.29 T4 69.191 10.492 84.79 134.00 (T1, T2, T3 and T4). Hence the decrease in bone density increased in stress level (Table 2). Maxmium Von Mises Stress in the Different Bone Densities with the Cylindrical Implant (Fig. 6) The height of each bar depicts the maximum stress level. The color of the bar denoted the different models or bone densities. 1. The stress level was least in cancellous bone and maximum in implant and considerably higher in the crown. 2. The heights of the bars in the crown region where almost equal indicating that the decrease in bone density did not affect the stress level; similar interpretation holds good for cancellous bone, cortical bone and the implant (Table 3). Discussion Osseointegrated dental implants have been a well accepted and predictable treatment modality for the rehabilitation of partially and completely edentulous patients. Long-term success rates as high as 95 % for mandibular implants and 90 % for maxillary implants have been reported. The success of dental implants depends on various factors like Table 3 Maximum Von Mises stress (MPa) on and around cylindrical implant in different densities of bone Type of bone Crown Cancellous bone Cortical bone Implant C1 122 5.015 26.66 122 C2 122 4.04 28.29 122 C3 121.76 5.23 33.66 121.93 C4 121.55 4.81 38.58 121.67 implant design, bone density, type of prosthesis and loading of implants, oral microflora and parafunctional forces. Misch [13] has classified bone density into four types: D1 is dense cortical bone, D2 is porous cortical and coarse trabecular bone, D3 is porous cortical bone (thin) and fine trabecular bone, and D4 is fine trabecular bone. Different implant designs have been suggested for different bone densities because bone has a ten-fold difference in strength and flexibility between D1 and D4 bone qualities. Finite element analysis has been used for many years in industry to provide analytical solutions to problems involving complex geometrical forms [8]. In 1976, Weinstein et al. were the first to use FEA in implant dentistry; subsequently, FEA was applied rapidly in the field of implant dentistry. [3] The results of this study showed that the maximum Von Mises stresses were observed at the crestal region or the neck of the implant for both the threaded and cylindrical types in all the four densities of bone. This is in similar to the results obtained by various other studies that demonstrated that bone loss begins around the implant neck due to higher bone stresses at the crestal region [9]. The results also showed that a decrease in bone density causes an increase in stress level around the neck of the implants, especially for the threaded implants; whereas the stress level was not influenced to a great extent by bone density in the case of cylindrical implants. This is consistent with the results obtained in a study conducted by Sevimay and Turhan [2] which utilized the 3D finite

J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 559 element analysis to analyze the effect of different bone densities on stress distribution. They concluded that higher stress magnitudes were seen in D3 and D4 bone as the trabecular bone is weaker and there is less resistance to deformation than in the other bone qualities modeled. When the magnitude of stresses are compared between the threaded and the cylindrical implants, it was seen that maximum Von Mises stresses observed for the threaded implants in D4 bone was more compared to that of cylindrical implants. In a study conducted by Siegele and Soltesz, the results showed that different implant shapes lead to significant variations in stress distribution in bone. [3] The stepped or the threaded implants induced greater stresses than the cylindrical shaped implants. The reason for this could be that the areas of threads would form the frontline of stress concentration due to the sharp line angles located at the sides. The success rate of cylindrical implants in D4 bone was higher than that of the threaded implants. This was attributed to the fact that cylindrical implants generate less lateral force in spongy D4 bone than the threaded implants [4]. The present study thereby suggests that in case of D4 density of bone a cylindrical implant design may be considered. This result is also supported by the fact that use of a threaded implant in D4 density bone can be difficult as the threads cannot be engaged by the spongy bone, where as a friction fit can be obtained with the cylindrical implant design. Hence from biomechanical and surgical points of views, cylindrical implant design may be considered for placement in D4 type bone. Conclusion 1. Maximum Von Mises stress was observed at the crestal region of the bone in all the models. 2. Maximum Von Mises stress around the threaded implant was observed in the D4 type of bone when compared to the stress distribution in D1, D2 and D3 types of bone. 3. Maximum Von Mises stress around the cylindrical implant was observed in the D3 type of bone when compared to the stress distribution in D1, D2 and D4 types of bone. 4. When the stress distribution was compared between the threaded and cylindrical implant in D1, D2, D3, and D4 bone, maximum Von Mises stresses were observed around the threaded implant. 5. In view of the above conclusions it may be inferred that the cylindrical type of implant design may be more suitable in softer bone. References 1. Lee JH, Fria V, Lee KW, Wright RF (2005) Effect of implant size and shape on implant success rate: a literature review. J Prosthet Dent 94:377 381 2. Sevimay M, Turhan F (2005) Three-dimensional finite element analysis on the effect of different bone quality on stress distribution in implant-supported crown. J Prosthet Dent 93:227 233 3. Geng JP, Tan KBC, Liu GR (2001) Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent 85:585 598 4. Paul AF, Stpehen L, Wheeler, Lindsay JA (1993) Success and failure rates of cylinder implants in type IV bone. J Periodontol 64:1085 1087 5. Meijer HJA, Kupier JH, Starmans FJM, Bosma F (1992) Stress distribution around dental implants: influence of superstructure, length of implants, and height of mandible. J Prosthet Dent 68:96 102 6. Papavasilion G, Kamposiors P, Bayne SC, Felton DA (1996) Three-dimensional finite element analysis of stress-distribution around single tooth implants as a function of bony support, prosthesis type and loading during function. J Prosthet Dent 76:633 640 7. Vidyasagar L, Apse P (2004) Dental implant design and biological effects on Bone implant interface. Stomato Balt Dent Maxil-facial 6:51 54 8. Germay A, Morgano SM (2004) Finite element analysis of three designs of an implant-supported molar crown. J Prosthet Dent 92:434 440 9. Lin CL, Kuo YC, Lin TS (2005) Effect of dental implant length and bone quality on biomechanical responses in bone around implants: a three-dimensional linear finite element analysis. Biomed Eng Appl Basis Commun 17:44 49 10. Anusavice KJ (1996) Mechanical properties of dental materials. In: Anusavice KJ (ed) Phillip s Science of dental materials, 10th edn. Elsevier, Philadelphia, p 66 11. Misch CE, Bidez MW (2005) Occlusal considerations for implant supported prostheses. In: Misch CE (ed) Dental implant prosthetics. Mosby, Maryland Heights, pp 472 510 12. Yokoyama S, Wakabayashi N, Shiota M, Ohyama T (2004) The influence of implant location and length on stress distribution for three unit implant supported posterior cantilever fixed partial dentures. J Prosthet Dent 91:234 240 13. Misch CE (2005) Bone density: a key determinant for clinical success. In: Misch CE (ed) Dental implant prosthetics. Mosby, Maryland Heights, pp 130 141