PORTO 2 VENT CPAP OS. Operator s Manual. PORTO 2VENT CPAP OS System Operator s Manual Part Number Rev I

Similar documents
I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device

English. Fealite Nasal Pillows System. User Manual

HCS70004P, HCS70004G HCS70004C, HCS70004BL HCS70004DP, HCS70004 AEROMIST COLORS NEBULIZER COMPRESSOR KIT. Instruction Manual

Compressor Nebulizer Instruction Manual Part No.: 5055

PORTABLE COMPRESSOR NEBULIZER Guidebook & Manual Reorder No. 5606

COMPRESSOR NEBULIZER MODEL: JB USER S MANUAL

Compressor Nebulizer. Guidebook

Medescan Nebuliser Med - S600A Instructions Manual

Compressor Nebulizer. Model UN-014

DeVilbiss Pulmo-Aide Compact Compressor/Nebulizer Instruction Guide

Midwest E Motor. Directions For Use. Doc. No

Compressor Nebulizer System

Adult Intubation Skill Sheet

weight.detach/bench ITEM NO: 76310

Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS}

Low Row. User manual E S S E N T I A L S T R E N G T H

VISTA400 REMOVABLE FOOT PLATE. Instructions For Use. Step on yellow pedals to close and open legs to go around chairs & ambulate.

I. Subject: Administration of Virazole (Ribavirin) by aerosol.

SEMINOLE COUNTY EMS PROVISIONAL EMT SKILLS VERIFICATION

STATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS

Single Use Curlew TM Multiple Biopsy Forceps

TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT:

DeltaWave Nasal Pillow Device Rx only. Developing Sleep Solutions. Users Guide (English) Version 1.0. Page 1

Radux StandTall Instructions for Use Sheath Extender and Securement Clasp

DeVilbiss Pulmo-Aide Compressor/Nebulizer Instruction Guide

SelectTech 4.1 Bench Assembly / Owner s Manual

SureFlow. Oxygen Flow Station PROVIDER TECHNICAL MANUAL

KOALA. Adult Bacterial Viral Filters. Medical Pty Ltd

Owner s Manual Cath Ergometer

Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T

Effective Date: August 31, 2006 SUBJECT: CARE AND USE OF NEBULIZER AND INTERMITTENT POSITIVE PRESSURE BREATHING DEVICE

Compressor Nebulizer. Guidebook

St. Jude Medical 8-Channel Adapter. Clinician's Manual

MEDIUM-FLOW PNEUMOTACH TRANSDUCER

TABLE OF CONTENTS 1.1. WARRANTY DAMAGED SHIPMENTS SERVICE INTRODUCTION UNPACKING AND INSTALLATION

SelectTech 4.1 Bench. Assembly Manual

Capnography (ILS/ALS)

HeartCode PALS. PALS Actions Overview > Legend. Contents

WEIGHT BENCH EXERCISER User s Manual

max airless Operating Manual for Dispenser Read this manual thoroughly before operating the Labmax airless!

IMPORTANT INFORMATION - PLEASE READ. Table of Contents. Introduction. General Information

Multipower. User Guide E S S E N T I A L S T R E N G T H

ER75 Electro-Acoustic Ear Simulator. Operating Manual

I. Subject: Medication Delivery by Metered Dose Inhaler (MDI)

WEIGHT BENCH EXERCISER User s Manual

I. Subject: Pressure Support Ventilation (PSV) with BiPAP Device/Nasal CPAP

Comfortmax Thermax Hot/Cold Water Circulation System. Instructions for Use

CHANHASSEN FIRE DEPARTMENT MEDICAL / RESCUE SKILLS

Flowmeters and. Specialty Products

Instruction Manual Aerogen, Inc. Part No. AG-AL1010 Rev. C

COR. Precision Targeting Cartilage Repair System. Arthroscopic Technique for Repair of Osteochondral Defects

User Manual & Repair Manual

CATHETER ACCESS KIT. For use with Prometra Programmable Infusion Systems

Portable Equine Nebuliser System. User Manual

OWNER'S MANUAL OxyTrainer

TIP CARD. Model Temporary External Pacemaker

ZOLL Power Infuser. Operator s Guide. Model M100B-3A

Provider guide. DreamStation CPAP DreamStation CPAP Pro DreamStation Auto CPAP DreamStation BiPAP Pro DreamStation Auto BiPAP

Dip Station G

Blood Glucose Monitoring System. User Guide

Owner s Manual. Model# IMPORTANT: Read all instructions carefully before using this product. Retain this owner s manual for future reference.

ULTR A SERIES G7-S13


ULTR A SERIES G7-S23

Carina. The compact wonder. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

IC031 Magnetic Belt Drive Performance Indoor Cycle Bike

Motorised Treadmill Model No: JI1625

InWave. User s Manual

Device Preparation (all steps to be performed per standard interventional technique)

ULTR A SERIES G7-S75

experiences that matter

OWNER S MANUAL and INSTALLATION INSTRUCTIONS

USER MANUAL Rev. 1.15

PARI INSTRUCTIONS FOR USE COMPRESSOR NEBULIZER SYSTEM. Portable Power Kit (optional)

Adjustable Ab Board G

CROSS TRAINING MAGNETIC TREADMILL

Pain Relief Patch. Operation Manual Read Before Using MODEL # ET

SelectTech 3.1 Bench Assembly / Owner s Manual

OPERATING AND MAINTENANCE MANUAL. - Tube Style Integrated Flowmeter -

Muscle Stimulation. Page 1

Model 9306A Six-Jet Atomizer

Complex Care Hub Manual: Continuous Positive Airway Pressure (CPAP) Ventilation

Surgical Technique Guide

Pocket Diagnostic Instruments. Use, Care, & Maintenance

A82-LIFE-Dog Nebulizer NBB02

Capnography Connections Guide

Power-Pak 445 Adjustable Dumbbell System

Gewässer Umwelt Schutz

Thank you for purchasing the AB COASTER. Our customer service staff is ready to help and to answer any questions.

Instructions for Use Reprocessed Ethicon ENDOPATH XCEL TM Bladeless Trocar with OPTIVIEW Technology

PLEASE READ THIS USER GUIDE BEFORE OPERATING THE SYSTEM

ACCESSORIES AND DISPOSABLES

ULTR A SERIES G7-S40

Stratis INSTRUCTIONS FOR USE. Needle-free Injection System. 0.5mL volume (+/-5%)

Provider guide. DreamStation


Thank you for purchasing the AB COASTER. Our customer service staff is ready to help and to answer any questions.

Transcription:

PORTO 2 VENT CPAP OS Operator s Manual 1

2

TABLE of CONTENTS 1. Introduction 3. Operating Instructions 1a. Definitions 3a. Setting the CPAP Level 1b. General Description 3b. Applying the Breathing Circuit and Mask 1c. Intended Use 3c. Getting Started 1d. Labels and Symbols 3d. Monitoring the Patient 1e. How CPAP OS Works 1f. Breathing on the CPAP OS Unit 4. Care and Maintenance 2. Installation and Assembly 5. Accessories 2a. Functional Test 2b. Oxygen Supply 6. Specifications 2c. Breathing Circuit 3

SAFETY Before you attempt to set up and use your PORTO 2 VENT CPAP OS System ( System ), read this manual in its entirety. The PORTO 2 VENT CPAP OS System is to be used by authorized, trained personnel under the supervision of a physician. If you have any questions regarding the operation of your CPAP OS System an Emergent Respiratory, LLC representative will be able to assist you. You may also contact the Emergent Respiratory, LLC Customer Care and Service Assistance Line, Toll Free, at 866-411-4368 or e-mail: customercare@eresp.com 1. Introduction 1a. Definitions NOTE: A NOTE surrounded by this box indicates a technical or clinical point that may enhance the use of or protect the device. CAUTION: A CAUTION surrounded by this box is an indication to exercise special care for the safe and effective use of the device. WARNING: A WARNING in bold typeface surrounded by this box is an indication of problems that may lead to death or serious injury. 4

1b. General Description The PORTO 2 VENT CPAP OS System is a gas-powered system that delivers Continuous Positive Airway Pressure (CPAP) at operator adjustable levels throughout the breathing cycle, independent of the patient s flow rate. The System conserves the gas supply by terminating gas flow during the expiratory phase of the breath. The System is intended for use on spontaneously breathing patients. CPAP OS Unit DISS Oxygen Inlet Locking Bayonet Main Outlet Port and Breathing Circuit Attachment Airway Pressure Gauge Dust Cover CPAP Adjustment 5

1c. Intended Use To provide Continuous Positive Airway Pressure (CPAP) to spontaneously breathing patients in hospital and prehospital (EMS) environments. CAUTION: The CPAP OS unit and breathing circuit are intended only for use with a positive pressure face mask and should not be connected to endotracheal or oropharyngeal tubes. CAUTION: Always use appropriate monitoring, such as, oxygen monitor, pulse oximeter, EKG and chest auscultation. CAUTION: Federal Law restricts this device to sale by or on the order of a physician. 6

1d. Labels and Symbols DISS Oxygen Inlet CPAP Adjustment Knob Increase clockwise Decrease counterclockwise Airway Pressure Gauge 7

CPAP os Unit Abbreviated Operating Instructions Label: CPAPos Abbreviated Operating Instructions NOTE: for complete detailed operating Instructions please read the CPAPos Operator s Manual in its entirety before use. Service Inspection and Maintenance is recommended every two (2) years. 1. Attach gas source to the GAS inlet located on the side of the device. 2. Connect the CPAPos Breathing Circuit. 3. Instruct Patient and Apply Mask. 4. Set CPAP level: To set Constant Positive Airway Pressure, turn CPAP Adjustment Knob Clockwise and observe the Needle Indicator on the Airway Pressure Gauge. 5. Observe and Monitor Patient Vital Signs. CAUTION: Use only the CPAPos Breathing Circuit. CPAP Model No. 1900-001 Manufactured for Emergent Respiratory, LLC USA 1-866-411-4368 8

1e. How PORTO 2 VENT CPAP OS System Works The System (See figure 1) provides continuous positive airway pressure (CPAP) throughout the breathing cycle (inspiration and expiration), and flow of gas on patient demand during inspiration. The peak inspiratory flow rate from the CPAP OS unit is determined by the patient inspiratory effort. Prior to inspiration the flow rate of gas from the CPAP OS system is zero liters per minute (lpm) (0 lpm). To initiate flow of gas the patient must trigger the CPAP pressure generator. DISS Oxygen Inlet Port The gas inlet port is a 9/16 standard DISS oxygen connector. The specification for inlet pressure is 40 60 psi (2.8 4.1 bar). CPAP Pressure Generator The CPAP Pressure Generator maintains a continuous positive airway pressure and delivers flow from 0 to 100 or greater, liters per minute on demand to a spontaneous breathing patient. CPAP Adjustment Regulator The CPAP adjustment is an operator adjustable pressure control providing pressure to the patient airway between 0 and 15 cmh 2 O. NOTE: Mask and proximal airway pressure will vary. 9

Pressure Balanced Patient Valve and Inspiratory Check Valve The inspiratory valve opens to supply flow to a spontaneous breathing patient during the inspiratory phase of breath delivery. The expiratory valve is pressure balanced to precisely maintain CPAP levels and reduce patient effort to open and exhale through the valve. The patient valve gas collector port rotates 360 O and is a 22 mm I.D. and 30 mm O.D. port to accept filters and spirometry devices. CAUTION: Do not use any other make or manufacturer of the pressure balanced patient valve that is not labeled for use with the CPAP OS device. The patient breathing circuit must be connected as illustrated in section 2c and described in section 3b of this manual. Failure to do so could result in device malfunction. The patient and equipment must be observed constantly. Over Pressure Relief Valve In the event of an occlusion and/or pressure build up in the CPAP OS system, the Over Pressure Relief Valve is preset to relieve pressure at 22 cmh 2 O. Anti-asphyxiation Port In the event of complete gas supply failure the Anti-asphyxiation Port allows the spontaneous breathing patient access to ambient air. Outlet Port Connector The Outlet Port Connector is a locking bayonet connector used to attach the corrugated main tube, and engage the airway pressure line. 10

Airway Pressure Line The Airway Pressure Line is attached from the outlet of the Pressure Balanced Patient Valve through the outlet connector with a standard 1/8 th inch I.D. tube. CAUTION: The airway pressure line is connected internally to the pressure gauge. Disconnection or occlusion of this line may not be detected. The patient and equipment must be observed constantly to check for such problems. Bacteria/Viral Filter The bacteria/viral filter is connected to the Airway Pressure Line and outlet connector. Corrugated Breathing Tube The corrugated main tube directs gas flow from the CPAP os outlet port through the inspiratory valve to the patient and Pressure Balanced Patient Valve. It is six (6) feet in length. Positive Pressure Face Mask The Positive Pressure Face Mask is a standard single-ported mask. It is available in small, medium and large adult sizes. Black Neoprene Head Harness The Black Neoprene Head Harness is used to secure the positive pressure facemask to the patient s face. It is a one size fits all adjustable design. 11

DISS Oxygen Inlet Port Over Pressure Relief Valve Internal Pressure Reducer Anti-asphyxiation Port Airway Pressure Gauge Outlet Port Connector Bacterial/Viral Filter CPAP Adjustment Regulator Airway Pressure Line Figure 1. PORTO 2VENT CPAP OS System 12

2. Installation and Assembly 2a. Suggested functional testing of Control unit prior to initial use. Attach the included oxygen hose to the unit and to an oxygen source. With the CPAP pressure control knob turned completely in the counterclockwise direction (against the stop), apply an inlet pressure of 50 PSI to the Control Module. If the unit free-flows upon initial application of supply pressure, momentarily (3-5 seconds) block the outlet. Unblock the outlet and the flow shall have ceased. Block the outlet. Rotate the CPAP Adjustment Knob clockwise until the Airway Pressure Gauge indicates approximately 5 cmh 2 O. Unblock the outlet. The unit should fully flow. Block the outlet, being careful to not block the Airway Pressure Line (small plastic tube). Slowly rotate the CPAP Adjustment Knob to the fully clockwise position. Observe that the pressure rise indicated on the gauge needle is smooth. An outlet pressure of 15 cmh 2 O or greater on the Airway Pressure Gauge indicates the unit is operating properly. Unblock the outlet. Turn the pressure control knob back to off position. Do not force the knob against the stop. In the majority of cases, the unit will continue to flow. Momentarily (3-5 seconds) block the outlet and then unblock. The unit should shut off. Contact Emergent Respiratory Customer Service if unit fails to pass these tests. 13

2b. Oxygen Supply The CPAP OS device can use oxygen from a cylinder or from a central supply system of 40 to 60 psi. When ready to administer CPAP therapy, connect the oxygen supply to the DISS inlet located on the side of the device. WARNING: To prevent explosion hazard, make sure the oxygen connector is free from oil. NOTE: To turn the CPAP OS unit off, disconnect the O 2 supply hose from the source or turn off at the O 2 cylinder. CAUTION: Certain data on the use of high concentrations of inspired oxygen show adverse effects on cardiac function in patients who have congestive heart failure or other specific conditions. When attached to an oxygen source, the CPAP OS provides an F I O 2 of up to 95% dependent upon outlet flow. NOTE: To monitor F I O 2 accurately the use of an in-line oxygen monitor is recommended. 14

2c. PORTO 2 VENT Breathing Circuit The CPAP OS Breathing Circuit contains the following components: Six (6) foot corrugated main tube Area Ratio/Pressure Balanced Exhalation Valve Bacterial/Viral Filter Positive Pressure Medium Face Mask (Small and Large Adult sizes may be purchased separately) Inspiratory Check Valve Locking Bayonet Outlet Adapter Airway Pressure Line 15

Area Ratio/Pressure Balanced Patient Exhalation Valve and Components Patient Valve Cap Exhalation Port with 360 0 rotation and 22 mm I.D. / 30 mm O.D. Pressure Balanced Exhalation Valve 22 mm Main Tube Connection Patient Mask Connection 22 mm O.D. Inspiratory Check Valve NOTE: The CPAP OS Breathing Circuit and all of its components are sold clean and non-sterile. 16

3. Operating Instructions 3a. Setting the CPAP Level Prior to setting the pressure always observe that the airway pressure gauge needle indicator is at the zero (0) value with the CPAP adjustment knob in the fully counterclockwise position and the breathing circuit connected. To set continuous positive airway pressure, turn the CPAP adjustment clockwise and observe the needle indicator on the airway pressure gauge. 3b. Applying the Breathing Circuit and Mask Assembly Instructions: The CPAP OS Breathing Circuit is pre-assembled at the factory. To attach the breathing circuit to the CPAP OS unit insert and align the locking bayonet outlet adapter to the unit and turn clockwise until securely engaged. CAUTION: Inspect the patient (yellow mushroom) valve to assure that it is properly seated and not cocked. The valve may be re-seated properly by completely unscrewing the patient cap and then re-installing the cap. The valve will automatically re-seat itself without the need to handle the valve. 17

CAUTION: Do not use any other make or manufacturer of the pressure balanced exhalation valve that is not labeled for use with the CPAP OS device. The patient breathing circuit must be connected as illustrated in section 2b of this manual. Failure to do so could result in device malfunction. The patient and equipment must be observed constantly. CAUTION: Do not clean or reuse the single-use, disposable breathing circuit or components. Do not autoclave or submerse the CPAP OS unit or any of its components in any solution. Do not connect a filter directly into the airway path. This could compromise the operation of the control unit. 18

3c. Getting Started 3d. Monitoring the Patient 1. Connect the oxygen supply to the DISS oxygen inlet located on the side of the CPAP OS unit. 2. Connect the Breathing Circuit as illustrated in this manual. (p 15) 3. Instruct the patient and apply the mask to the patient as per your protocol. 4. Set the CPAP level as per your protocol by turning the CPAP adjustment knob clockwise and observe the needle indicator on the airway pressure gauge. During operation, monitor the following periodically. 1. Monitor to assure that there are no leaks at the patient connection. 2. Monitor the Airway Pressure Gauge. 3. Monitor the patient s arterial blood oxygen saturation (S a O 2 ). 4. Monitor the patient s blood pressure and EKG. 5. Monitor the patient s delivered F I O 2. 6. Monitor the patient s breathing with chest auscultation. 7. Monitor the patient for signs of dehydration and discomfort in the upper airways. 19

CAUTION: Check the airway pressure gauge while the patient is breathing to assure that pressure is maintained within specifications (see CPAP OS Specifications page 12). As the patient breathes, larger changes in pressure shows that CPAP is not being effectively delivered. Possible causes are insufficient source gas pressure and flow or possibly a malfunction of the CPAP unit and/or breathing circuit. Excessive pressure variation during CPAP treatment can cause fatigue and respiratory failure. WARNING: Potential side effects of continuous positive airway pressure may include fluid retention, pneumothorax, decreased cardiac output and gastric distension. 4. Care and Maintenance The housing and faceplate of the CPAP OS unit may be cleaned with mild soap and water. A glutaraldehyde solution may be used as a disinfectant on all external surfaces of the CPAP OS housing. CAUTION: Do not clean or reuse the single-use, disposable breathing circuit or components. Do not autoclave or submerse the CPAP OS unit or any of its components in any solution. CAUTION: Do not attempt to disassemble or repair the CPAP OS unit. 20

Service inspection and maintenance is recommended every Two (2) Years. Service and Repair: For all service and repair of the CPAP OS unit return it to an authorized Emergent Respiratory, LLC authorized service and repair center. To return a CPAP OS unit call 1-866-411-4368 to obtain a Return Goods Authorization (RGA) number. Warranty: Emergent Respiratory, LLC warrants the CPAP OS unit and its accessories to be free from defects in materials and workmanship for a period of one year from date of purchase. 5. Accessories The following accessories are compatible with the CPAP OS Unit. Mounting: Any pole or rail system with universal clamp. Oxygen Monitor: Any in-line oxygen monitor. 21

6. Specifications CPAP Treatment Range: Maximum Inspiratory Flow: F IO 2: Adjustable from 2.5 to 15 cmh 2O 100 liters (or greater) per minute (lpm) up to 95% dependent upon outlet flow. Inspiratory Effort Less than 2 cmh 2O Exhalation Resistance Less than 2 cmh 2O* Patient Valve Dead Space: 15 ml Inlet Pressure: 40-60 psi (2.7 to 4.1 bar) Operating Temperatures: 0 0 F to 120 0 F (-18 to 50 0 C) Storage Temperatures: -40 0 F to 160 0 F (-40 to 70 0 C) Over Pressure Relief: 22 cmh 2O (max 35 cmh 2O @ 100 lpm) Anti-asphyxiation Valve: - 8 cmh 2O @ 50 liters per minute (lpm) Positive Pressure Face Masks Small, Medium and Large sizes Dimensions: 3½h x 5d x 8½w in (9 x 12.5 x 21.5 mm) Weight: 3.50 lbs (1.50 kg) *Expiratory Resistance is measured from airway pressure at beginning of expiratory cycle to peak expiratory pressure. 22

This page intentionally left blank PORTO 2VENT CPAP 08 System Operator's Manual 23

2731 Loker Avenue West, Carlsbad, CA 92010 For warranty registration of your CPAP OS unit visit www.eresp.com 24