The change in energy absorbed post removal of metalwork in a simulated paediatric long bone fracture

Similar documents
Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4

Fracture fixation. Types. Mechanical considerations. Biomechanics of fracture fixation. External fixation. Internal fixation

Study of Ender s Nailing in Paediatric Tibial Shaft Fractures

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Biomechanics of Fractures and Fixation

Development of a Finite Element Model to Study the Torsional Fracture Strength of an Analogue Tibia with Bicortical Holes

Fractures of the Radial and Ulnar Shafts In the Pediatric Patient

TENS in paediatrics both bone forearm fractures

Fractures of the tibia shaft treated with locked intramedullary nail Retrospective clinical and radiographic assesment

Plate and Bone Stresses for Single- and Double-Plated Femoral Fractures. D.R. Carter and R. Vasu

3.5 mm Locking Attachment Plate

LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL

THE NANCY NAIL. The End Caps ADVANTAGES OF NANCY NAIL

AcUMEDr. Anatomic Midshaft Forearm Plates

Treatment Alternatives for Pediatric Femoral Fractures

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Design Team Vipul Nanavati, MD James M. Paci, MD Frederick W. Werner, MME Levi G. Sutton, MS

Comparison of Flexible Intramedullary Nailing with External Fixation in Pediatric Femoral Shaft Fractures

Percutaneous Transphyseal Intramedullary Pinning for Displaced Diaphyseal Forearm Fractures in Children

Improving Fixation of a Previously Designed Pediatric Tibial Stent

Results of tibia nailing with Angular Stable Locking Screws (ASLS); A retrospective study of 107 patients with distal tibia fracture.

LCP Distal Tibia Plate

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

EXTERNAL FIXATION SYSTEM

Revision. Hip Stem. Surgical Protocol

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system.

We have assessed the influence of isolated and

PLR. Proximal Loading Revision Hip System

Case Report. Antegrade Femur Lengthening with the PRECICE Limb Lengthening Technology

Introduction. Mohammad Ruhullah 1*, Sanjay Shah 1, Hare Ram Singh 1, Dipak Shrestha 2. Bagmati, Nepal. Abstract

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Intramedullary Nailing: History & Rationale

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

Femur Shaft Fractures Under 10 years old

COMPARATIVE STUDY OF MANAGEMENT OF DIAPHYSEAL FEMUR FRACTURE WITH INTRAMEDULLARY INTERLOCKING NAIL AND K. NAIL

TIPMED EXTERNAL FIXATION SYSTEMS

EVOS MINI with IM Nailing

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

AcUMEDr. LoCKING CLAVICLE PLATE SYSTEM

The Mechanical Properties Of Fixing Greater Trochanter Or Lesser Trochanter In Complex Four Part Intertrochanteric Fractures

Distal Ulnar Locking Plate

Techique. Results. Discussion. Materials & Methods. Vol. 2 - Year 1 - December 2005

GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE

NEW INSTRUMENTS FOR INTERNAL FIXATION OF FRACTURES USING MINIMALLY INVASIVE TECHNIQUES

Pediatric LCP Plate System. For osteotomies and fracture fixation of the proximal and distal femur.

Femur Condylar Plate System Procedural Steps.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION

TRAUMATOLOGY. Trochanter

Pediatric Femoral Shaft Fracture Management By Titanium Elastic Nailing; A Prospective Study Of 112 Patients

PROXIMAL TIBIAL PLATE

Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study

Surgical Technique. Targeter Systems Overview

Journal of Foot and Ankle Research

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates

Locked Plating: Biomechanics and Biology

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Designing a Novel Fixation Device for Pediatric Orthopaedic Tibia Fractures

This is a repository copy of Mechanics of musculoskeletal repair devices.

Titanium elastic intramedullary nailing : closed or mini-open reduction?

DHS Plate 135º. DHS Platte 135º

PAEDIATRIC FEMORAL SHAFT FRACTURES MANAGED WITH TENS

ZMR Crossover Instruments. Abbreviated Surgical Technique

Practical Reduction Techniques: Diaphyseal Reduction. Philip Wolinsky University of California at Davis Medical Center

Chapter 4: Forearm 4.3 Forearm shaft fractures, transverse (12-D/4)

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

Stress distribution within circumferential clasp arms

Complications Cast Treatment Psychosocial effect on children Economoic effects on their families [Hughes 1995] ORIF with Plate Ex Fix Antegrade Nail L

Knee spanning solutions

The Lateral Trochanteric Wall A Key Element in the Reconstruction of Unstable Pertrochanteric Hip Fractures

Technique Guide. DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP).

InternationalJournalofAgricultural

Initial Fixation Strength of Bio-absorbable Magnesium Screw

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

PFNA-II. Proximal Femoral Nail Antirotation.

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

11 5 New Cavendish Street, The Impact Properties of Bane. R.P. Oickenson, J.W. Wall and w.c. Hutton

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Journal of American Science 2017;13(11) Submuscular Plating in Spiral Unstable Pediatric Femur Fractures

Comparison of Below The Elbow Cast with Above The Elbow Cast in Treating Distal Third Forearm Fractures in Children

The Surgical Management of Rickets & Osteogenesis Imperfecta

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

The Vilex FUZETM. Dual Thread Screw & Intramedullary Nail in One Implant. The Ultimate TTC Arthrodesis Internal Fixator

Variable Angle LCP Locking Technology

Surgical technique. IMF Screw Set. For temporary, peri opera tive stabilisation of the occlusion in adults.

Small Fragment Plating System. Securing optimal fixation through locked and compression plating technology

Pediatric LCP Hip Plate. For osteotomy and trauma applications in the proximal femur.

Small Fragment Plating System

Principles of intramedullary nailing. Management for ORP

Flexible Intramedullary Nailing of Forearm Fractures at the Distal Metadiaphyseal Junction in Adolescents

Pediatric Tibia Fractures Key Points. Christopher Iobst, MD

PFNA. With Augmentation Option.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

An effort is made to analyse the stresses experienced by the human femur. In order

The Birmingham Hook Plate Treatment Of Irreduceable Displaced Mallet Avulsion Fractures: A Technical Note

CASE REPORT. Bone transport utilizing the PRECICE Intramedullary Nail for an infected nonunion in the distal femur

Management of Fractures. Traumatology RHS 231 Dr. Einas Al-Eisa Lecture 5

Transcription:

J Child Orthop (2014) 8:443 447 DOI 10.1007/s11832-014-0606-z BASIC SCIENCE The change in energy absorbed post removal of metalwork in a simulated paediatric long bone fracture Alan J. Howieson Michael D. Jones Peter S. Theobald Received: 1 May 2014 / Accepted: 7 August 2014 / Published online: 27 August 2014 The Author(s) 2014. This article is published with open access at Springerlink.com Abstract Purpose The surgical treatment of paediatric fractures is increasing. Open reduction and internal fixation (ORIF) with plates and screws is long established, whilst the use of elastic stable intramedullary nailing (ESIN) has become increasingly popular. This study quantifies, in terms of the energy required to produce a fracture, the biomechanical sequelae of both techniques post removal of metalwork, to provide clinicians with evidence to guide post-operative advice. Methods An immature bovine model was adopted to ascertain whether these techniques exposed the bone to a greater re-fracture risk following removal of the device. Bones were prepared to reflect ORIF or ESIN techniques, or prepared intact for the acquisition of control data. Each bone was tested to failure at 90 /s, with the absorbed energy then being calculated to determine the relative difference between each technique and versus control data. Data describing peak shear stress and torque were recorded. Results Absorbed energy was reduced by 47 % in the ORIF group compared to both the control (p = 0.011) and ESIN (p = 0.018) groups. The peak shear stress and torque were also significantly different. All ORIF bones failed through drill holes, suggesting stress localisation around the defects. A. J. Howieson M. D. Jones P. S. Theobald (&) Trauma Biomechanics Research Group, Institute of Medical Engineering and Medical Physics, Cardiff University, Queen s Building, The Parade, Cardiff CF24 3AA, UK e-mail: theobaldps@cardiff.ac.uk A. J. Howieson e-mail: alanjhowieson@hotmail.com Conclusion This study suggests that there is a significantly higher re-fracture risk following the removal of ORIF plates when compared to both ESIN and the control environment. Whilst this may reflect the intuitive view of many clinicians, this study provides a quantitative value of the reduction in strength and should help clinicians to appropriately caution patients and parents prior to surgery. Keywords Diaphyseal fracture Paediatric Open reduction internal fixation Elastic stable intramedullary nailing Introduction Historically, the vast majority of paediatric fractures have been managed non operatively, but a ten-fold increase in operative treatment rates for paediatric forearm fractures was reported over the last two decades of the 20th century [1, 2]. Open reduction and internal fixation (ORIF) with plates and screws was the traditional surgical technique, although elastic stable intramedullary nailing (ESIN) has become increasingly popular [3 9]. The need to remove both ORIF and ESIN implants from children following bone healing is, however, associated with a relatively high risk of re-fracture [10 12]. The removal of ORIF implants exposes a series of cortical screw holes, which gradually heal over time. Intuitively, ESIN would, therefore, be associated with a lower risk of re-fracture following implant removal, but this risk has never been quantified and is extremely hard to define experimentally. A reduction in energy absorbed before fracture is one quantifiable variable that would give some indication as to whether there is a difference between

444 J Child Orthop (2014) 8:443 447 the techniques and, to date, there are no published studies examining this factor. There are many reasons for the removal of metalwork in children: all metal implants, with the exception of titanium, will lead to corrosion over time and the release of metal ions into the body. Whilst tumour development next to metal implants has been reported, reports are limited to case reports and large-scale studies are difficult to implement. There is also the risk of foreign body reactions and infection. Pain from prominent implants, implant migration, implant failure, wound problems and allergy are other problems [10]. The potential short- and long-term problems are illustrated by two case reports both involving children who had their femur plated at the age of 6 years: one child sustained a re-fracture just distal to the plate at the age of 12 years and the other sustained a re-fracture at the age of 67 years radiographs taken at the time of re-fracture showed almost complete loss of bone under the plate due to stress shielding [10]. Experimental evidence regarding the force required to fracture a paediatric long bone is scare and no relevant published studies were identified which quantified its magnitude. Forearm fractures commonly result in a spiral pattern or with the fracture lines at different levels on the radius and ulna, which suggests a torsional force [13]. The energy involved in falls from a height can be calculated using Newton s equations of motion. Calculations necessarily include complex equations which include a tip mechanism, an estimation of the body s centre of gravity and joint movements during the fall, and changes in linear momentum: joint movement in particular makes calculating ground reaction forces extremely difficult [14]. This study focuses on comparing the biomechanical consequences of the operative treatment of paediatric long bone fractures with either ORIF or ESIN, with the aim of providing clinicians with a quantitative difference between the two techniques and between intact normal bone. Immature bovine metacarpals were chosen as a representative model, having similar dimensions to paediatric femora. The primary outcome measure was the reduction in energy absorbed between the two techniques and versus normal bone. Secondary outcome measures were the maximum torque to failure, peak shear stress, shear modulus and fracture pattern. slaughter of commercial cows (\1 year old), before being cleared of all soft tissues and inspected for any previous fractures or obvious defects that could potentially affect the results. The diaphyseal lengths and central diameters were recorded (mean value of three measurements), before identifying the centre of rotation through the longest axis. The bones were then randomly sub-divided to form equal groups of control, ORIF and ESIN specimens. ORIF specimens were prepared by using a broad dynamic compression plate as a template. Three bi-cortical holes were drilled (2.5 mm diameter) and tapped (3.5 mm diameter), using the template, on either side of the bone mid-point. This represented the recent removal of a plate fixed with six, 3.5-mm bicortical screws. The cortical defect ratio, a measure of the extent of the defect caused by the screw-hole, was calculated for all samples. The ESIN specimens were prepared by identifying the centre of the proximal joint and by passing a wire (5.6 mm diameter) from this point along the length of the intramedullary canal, before removing it and, thus, leaving an intramedullary defect. The wire diameter was specifically calculated and chosen to represented the insertion of two flexible intramedullary nails, each with a diameter 30 40 % of the Materials and methods Bovine metacarpals, which have similar dimensions to human paediatric femora, were used as a model to explore the reduction in energy absorption and, thus, identify the fixation technique most likely to cause re-fracture. Metacarpal bones (n = 18) were harvested following normal Fig. 1 The alignment phase of the experimental preparations necessitated the use of two right-angled supports to achieve alignment of the distal and proximal bone ends prior to securing with Wood s metal. Aligning the centroidal axis of the bone to the vertical ensured a consistent application of load across all samples

J Child Orthop (2014) 8:443 447 445 internal diameter of the bone, as per standard practice. Each bone was then embedded to a depth of 50 mm in Wood s alloy (Flinn Scientific, Inc., USA), before being clamped into a tension torsion testing machine (MTS 858 Mini Bionix II) (Fig. 1). Each bone was tested in torsion to failure, at a rotational rate equal to 90 /s. This rate was adopted from a previous study, in view of the absence of literature quantifying the rotational rate experienced during equivalent, real-life fracture scenarios [15]. Previous biomechanical studies have shown that testing in torsion is the most valid method for assessing the energy absorbed to failure [16]. Data (time, torsional load) were automatically acquired at 100 Hz, via a wired personal computer, whilst the cortical thickness was retrospectively measured at the previously identified mid-shaft position. All data were then analysed to calculate the shear stress (Eq. 1), shear modulus (Eq. 2) and energy absorption (Eq. 3): Shear stress ¼ Shear modulus ¼ bone radius torque polar moment of inertia ð1þ bone length torque polar moment of inertia rotated angle ð2þ rotated angle torque Energy absorbed ¼ ð3þ 2 MS Excel 2010 (Microsoft Corporation, USA) and GraphPad InStat version 3 (Graphpad Software, Inc., USA) were used to compute analysis of variance (ANOVA) and Student s t-test, enabling the identification of any statistical significance between the datasets. Data were considered statistically significant when p \ 0.05. Results Data describing the dimensions of the immature bovine bones and the performance of the bones, when subject to a torsional load-to-failure, are presented in Table 1 and Fig. 2, respectively. There were no significant differences between the dimensions of the three groups. A 47 % reduction in energy absorption was observed by the comparison of ORIF to both the ESIN (p = 0.018) and control (p = 0.011) datasets; however, there was no significant difference between the ESIN and control datasets (Fig. 2). All ORIF bones had a cortical defect ratio 0.12 0.16, with all failing through at least one drill hole. Table 2 describes the secondary outcome measures. ORIF specimens required significantly less torque to produce fracture than the control dataset (p = 0.049); ESIN bones also required less mean torque to produce fracture than the control, but the results were not significant Table 1 Data describing the mean dimensions of each group (p = 0.507). Direct comparison of the ESIN and ORIF specimens revealed that the former required a higher torque to produce fracture, although the difference between the two did not reach statistical significance (p = 0.247). The maximum shear stress differed significantly when comparing the control and ORIF data (p = 0.043). The shear modulus for immature bovine bone was found to be in the range 15.5 18.8 GPa and provides a reference value for further studies. Discussion Mean length (mm) ± SD Mean polar moment of inertia (910-9 m 4 ) ± SD Mean cortical thickness (mm) ± SD Control 204 ± 8 18.9 ± 1.99 4.4 ± 0.7 ORIF 207 ± 12 18.5 ± 2.14 5.0 ± 0.6 ESIN 194 ± 9 16.7 ± 5.23 4.9 ± 0.6 Fig. 2 Data describing the energy absorption across all tests This study helps to quantify the reduction in bone strength following the removal of paediatric internal fixation. There are numerous clinical reports of long bone re-fractures: Kanlic et al. [17] reported a case after plate removal from the femur and, in a different multicentre study, Flynn et al. [18] reported re-fracture after nail removal. Re-fracture after paediatric forearm fracture occurs more commonly than in any other bone [12]. Multiple cases have been published: Deluca et al. [19] reported seven re-fractures from a set of 37 adults who had plates removed from the forearm: four of these patients required further open surgical treatment and three were treated with a cast. All arms

446 J Child Orthop (2014) 8:443 447 Table 2 Data describing the mean biomechanical values of each group Mean angular deformation to failure ( ) Mean peak torque (Nm) Mean energy absorbed (joules) Mean polar moment of inertia (910-9 m 4 ) Mean maximum shear stress (MPa) Control 23.2 70.6 835 18.93 40.3 ORIF 15.9 54.5 443 18.50 31.3 ESIN 26 64.1 843 16.59 39.8 were protected for a period of 4 8 weeks post removal of metalwork, although the metalwork was removed at an average of 87 days. Nielsen and Simonsen [20] removed plates at a mean period of 26 weeks post surgery in 23 children with fractures of the forearm, and one child sustained a re-fracture 3 months after the plate was removed. Vainionpää et al. [21], as part of a large series of 192 children with fractured forearms, reported one re-fracture amongst a series of ten patients treated with ORIF; the plates were removed at a range of 6 14 months post injury. Kim et al. [11] examined specific complications related to the removal of metalwork in paediatric forearm fractures and found, from a total of 43 children with a mean age of 10.6 years, that there were three re-fractures, two of which occurred in the same patient. The clinical relevance of the bone model adopted in this study is confirmed by considering the favourable comparison with pre-pubertal female and male femora. The average overall diameter for all the bovine bones tested was 21.2 mm, with an average cortical thickness of 4.8 mm. Previously published femoral dimensions from children with an age range of 8.2 9.4 years found an average femora diameter in pre-pubertal females and males of 19.9 and 21.0 mm, respectively, with cortical thicknesses of 4.6 and 5.0 mm [22 24]. It has also been shown by Vashishth et al. [25] that, whilst the elastic modulus of bovine bone is different from human bone, the ratio of yield stress in torsion, compression and tension correlates well with human bone and they have been used in previous studies to model human bone [26]. We calculated the mean shear modulus for our immature bovine bone specimens to be in the range 15.5 18.8 GPa this provides a useful reference value for further work. The diameter of the ESIN wire used corresponds to the insertion of two flexible intramedullary nails measuring 30 40 % of the internal diameter of the bone, as recommended for standard practice. Unfortunately, radiographic facilities were not available to examine the relative position of the wire within the canal; however, this potential variation was deemed unlikely to have any significant influence on the data. It was also important that the size of instruments used in this study reflected clinical practice to allow some extrapolation of results. A significant difference in energy absorbed to failure was observed between the ORIF and control groups, data which correlate to that of Edgerton et al. [27], who report a 20 % reduction in strength for a single drill hole using sheep femora. Our data, which describe multiple drill holes, found a greater, 47 %, reduction in absorption; however, compared to the data of Brooks et al. [28], our reduction in strength with multiple drill holes is less than the 55 % reduction they noted for a single hole in canine bone. It should be noted, however, that our specimens had a cortical defect ratio of 012 0.16, whereas Brooks et al. report a ratio of 0.14 0.18 [28]. This correlates with finite element simulations, as comparing the cortical defect ratio reported here with other studies, simulations have predicted little decline in torsional strength up to a defect ratio equal to 0.1, then a rapid decline in torsional strength between ratios of 0.1 0.2, before declining less steeply above 0.2 [27, 29]. Furthermore, the qualitative data acquired during this investigation clearly indicates that drill holes do localise stress, with all ORIF bones failing through at least one such cortical defect. The effect of occluding a drill hole was reported by Ho et al. [30]. They found that a 4-mm-diameter bicortical defect led to a 38 % reduction in energy absorbed by healthy porcine femora but that, by occluding the hole with plaster of Paris, there was a negligible difference between the specimen and intact bone. In conclusion, this study has adopted an appropriately dimensioned, accepted animal model to evaluate the change in energy absorbed post simulated removal of a fixation device in pre-pubertal children. The data presented suggest that children are at a higher risk of re-fracture immediately following ORIF removal than at an equivalent phase post ESIN. It would be interesting to know how the strength of the bone recovered over time as the defect consolidated, but this study would be impossible to perform in the paediatric population and very difficult even using animal models. For the purpose of this study, we have isolated one important variable, namely the cortical defect: future areas of interest include the effect of disuse

J Child Orthop (2014) 8:443 447 447 osteopaenia and the relevance of the entry point of the nail in the metaphyseal region of the bone. Conflict of interest None of the authors have any conflicts of interest. This study did not receive any external funding. Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. References 1. Wilkins KE, Aroojis AJ (2006) Incidence of fractures in children. In: Beaty JH, Kasser JR (eds) Rockwood and Wilkins fractures in children, 6th edn. Lippincott Williams & Wilkins, Philadelphia, pp 3 20 2. Cheng JC, Ng BK, Ying SY et al (1999) A 10-year study of the changes in the pattern and treatment of 6,493 fractures. J Pediatr Orthop 19:344 350 3. Van der Reis WL, Otsuka NY, Moroz P et al (1998) Intramedullary nailing versus plate fixation for unstable forearm fractures in children. J Pediatr Orthop 18:9 13 4. Ortega R, Loder RT, Louis DS (1996) Open reduction and internal fixation of forearm fractures in children. J Pediatr Orthop 16:651 654 5. Lascombes P, Prevot J, Ligier JN et al (1990) Elastic stable intramedullary nailing in forearm shaft fractures in children: 85 cases. J Pediatr Orthop 10:167 171 6. Ligier JN, Metaizeau JP, Prévot J et al (1988) Elastic stable intramedullary nailing of femoral shaft fractures in children. J Bone Joint Surg Br 70:74 77 7. Myers GJ, Gibbons PJ, Glithero PR (2004) Nancy nailing of diaphyseal forearm fractures. Single bone fixation for fractures of both bones. J Bone Joint Surg Br 86:581 584 8. Till H, Hüttl B, Knorr P et al (2000) Elastic stable intramedullary nailing (ESIN) provides good long-term results in pediatric longbone fractures. Eur J Pediatr Surg 10:319 322 9. Richter D, Ostermann PA, Ekkernkamp A et al (1998) Elastic intramedullary nailing: a minimally invasive concept in the treatment of unstable forearm fractures in children. J Pediatr Orthop 18:457 461 10. Peterson HA (2005) Metallic implant removal in children. J Pediatr Orthop 25:107 115 11. Kim WY, Zenios M, Kumar A et al (2005) The removal of forearm plates in children. Injury 36:1427 1430 12. Landin LA (1997) Epidemiology of children s fractures. J Pediatr Orthop B 6:79 83 13. Mehlman CT, Wall EJ (2006) Injuries to the shafts of the radius and ulna. In: Beaty JH, Kasser JR (eds) Rockwood and Wilkins fractures in children, 6th edn. Lippincott Williams & Wilkins, Philadelphia, pp 399 441 14. Hannon P (2006) Biomechanics of bone tissue. In: Hannon P, Knapp K (eds) Forensic biomechanics. Lawyers & Judges Publishing Company, Inc., Tuscon, pp 73 92 15. Theobald PS, Qureshi A, Jones MD (2012) Biomechanical investigation into the torsional failure of immature long bone. J Clin Orthop Trauma 3:24 27 16. Burstein AH, Frankel VH (1971) A standard test for laboratory animal bone. J Biomech 4:155 158 17. Kanlic EM, Anglen JO, Smith DG et al (2004) Advantages of submuscular bridge plating for complex pediatric femur fractures. Clin Orthop Relat Res 426:244 251 18. Flynn JM, Hresko T, Reynolds RA et al (2001) Titanium elastic nails for pediatric femur fractures: a multicenter study of early results with analysis of complications. J Pediatr Orthop 21(1):4 8 19. Deluca PA, Lindsey RW, Ruwe PA (1988) Refracture of bones of the forearm after the removal of compression plates. J Bone Joint Surg Am 70(9):1372 1376 20. Nielsen AB, Simonsen O (1984) Displaced forearm fractures in children treated with AO plates. Injury 15(6):393 396 21. Vainionpää S, Böstman O, Patiala H et al (1987) Internal fixation of forearm fractures in children. Acta Orthop Scand 58(2):121 22. Högler W, Blimkie CJ, Cowell CT et al (2008) Sex-specific developmental changes in muscle size and bone geometry at the femoral shaft. Bone 42:982 989 23. Quick JL, Ward KA, Adams JE et al (2006) Cortical bone geometry in asthmatic children. Arch Dis Child 91:346 348 24. Anderson M, Messner MB, Green WT (1964) Distribution of lengths of the normal femur and tibia in children from one to eighteen years of age. J Bone Joint Surg Am 46:1197 1202 25. Vashishth D, Tanner KE, Bonfield W (2001) Fatigue of cortical bone under combined axial-torsional loading. J Orthop Res 19:414 420 26. Reilly DT, Burstein AH (1975) The elastic and ultimate properties of compact bone tissue. J Biomech 8:393 405 27. Edgerton BC, An KN, Morrey BF (1990) Torsional strength reduction due to cortical defects in bone. J Orthop Res 8:851 855 28. Brooks DB, Burstein AH, Frankel VH (1970) The biomechanics of torsional fractures. The stress concentration effect of a drill hole. J Bone Joint Surg Am 52:507 514 29. Hipp JA, Edgerton BC, An KN et al (1990) Structural consequences of transcortical holes in long bones loaded in torsion. J Biomech 23:1261 1268 30. Ho KW, Gilbody J, Jameson T et al (2010) The effect of 4 mm bicortical drill hole defect on bone strength in a pig femur model. Arch Orthop Trauma Surg 130:797 802