Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Similar documents
Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Survey of patients CT radiation dose in Jiangsu Province

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Quality Control and Patient Dosimetry on line for Computed Tomography

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Adopting DAP as a dose metric in CT

Computed tomography Acceptance testing and dose measurements

Assessing the standard dose to standard patients for x-ray investigations

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

Assessment of effective dose in paediatric CT examinations

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Why is CT Dose of Interest?

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Accounting for Imaging Dose

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

A multicentric study on patient dose in multislice CT

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Patient dose in routine X-ray examinations in Yazd state

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Medical Physics and Informatics Original Research

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Regional diagnostic reference levels and collective effective doses from CT scanners in India

FDG-18 PET/CT - radiation dose and dose-reduction strategy

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

CT Radiation Risks and Dose Reduction

Introduction and Background

Patient Doses in Chest CT Examinations: Comparison of Various CT Scanners

CT Dosimetry in the Clinical Environment: Methods and Analysis

Chief Radiographer TEI Clinical Associate 2016

Quality Control Status of Radiology Centers of Hospitals Associated with Mashhad University of Medical Sciences

Radiation Protection Dosimetry Advance Access published April 22, 2008

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Patient dose measurement in common medical X-ray examinations in Iran

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

The impact of SID and collimation on backscatter in radiography

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

National Cancer Institute

Managing Radiation Risk in Pediatric CT Imaging

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

IMAGE GENTLY HOW CAN YOU HELP?

A Standardised Approach to Optimisation

International Journal of Radiation Research, January 2017

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

Doses Received by Patients during Thorax X-Ray Examinations

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Managing Patient Dose in Computed Tomography (CT)

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Doses from CT examinations to children suffering from hydrocephalus

Patient Dose Estimates. from CT Examinations. Patient Dose Estimates

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

EFFECTIVE DOSE ESTIMATION AND CANCER RISK ASSESSMENT DUE TO SOME SELECTED CT EXAMINATIONS AT SWEDEN GHANA MEDICAL CENTRE (SGMC)

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Estimating Patient Radiation Dose from Computed Tomography

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Radiation Dose Reduction Strategies in Coronary CT Angiography

Evaluation of Personal Shields Used in Selected Radiology Departments

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

Bureau of Laboratory Quality Standards

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Patient Radiation Doses from Adult and Pediatric CT

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Diagnostic reference levels as a quality assurance tool

National collection of local diagnostic reference levels in Norway and their role in optimization of X-ray examinations

Diagnostic reference levels in intraoral dental radiography in Korea

Local reference dose evaluation in conventional radiography examinations in Iran

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Radiation Protection- Cath lab

Radiation Dosimetry for CT Protocols

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Current status of diagnostic imaging in dental university hospitals in Japan

Radiation Protection Dosimetry Advance Access published September 13, 2006

State of the art and future development for standardized estimation of organ doses in CT

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

An Update of VirtualDose Software Used for Assessing Patient Organ Doses from CT Examinations

CT Quality Control Manual FAQs

Bone Densitometry Radiation dose: what you need to know

ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray

Tremendous progress in computed tomography (CT) technology

Application of international standards to diagnostic radiology dosimetry

Prediction of Organ and Effective dose with known mas and kvp for dose Optimisation Protocol and Recommendations in CT

Overview of Medical Diagnostic and Interventional Imaging: Usage Patterns

Transcription:

Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 2012, 233-238 Received: July 18, 2012; Accepted: November 17, 2012 Original Article Assessment of Patient Dose from CT Examinations in Khorasan, Iran Abstract Mohammad Taghi Bahreyni Toossi 1,2*, Mohammad Bahrami 2 Introduction Computed Tomography scans are a very important tool for diagnosis and assessment of response to treatment in the practice of medicine. Ionizing radiation in medical imaging is undoubtedly one of the most powerful diagnostic tools in medicine. Yet, as with all medical interventions, there are potential risks in addition to the clear potential benefits. Materials and Methods Two reference dose quantities have been defined in order to promote the use of good technique in CT. These are weighted CT dose index (CTDI w ) in (mgy) for a single slice in serial scanning or per rotation in helical scanning, and dose length product (DLP) per complete examination (mgy.cm), All measurements were performed using a pencil shaped ionization chamber introduced into polymethyl methacrylate cylindrical brain and body phantoms. This survey was performed on 7 CT scanners in Khorasan Province-Iran. Results DLP for brain,, abdomen and pelvic examinations had a range of 255-26, 76-1277, 48-737, 69-854 mgy.cm, respectively. Conclusion The results obtained in this study show that the DLP values obtained in this province are lower than European Commission reference dose levels (EC RDL), in other words performance of all the scanners were satisfactory. Keywords: DLP, CT, Ionization Chamber 1- Medical Physics Research Center, School of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran *Corresponding author: Tel: +98 511 8822876; Fax: +98 511 80020; Email: bahreynimt@mums.ac.ir 2- Department of Medical Physics, School of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran Iran J Med Phys, Vol. 9, No. 4, Autumn 2012 233

Mohammad Taghi Bahreyni Toossi et al. 1. Introduction Computed Tomography (CT) was introduced in the early 1970s and quickly became a very important tool in medical diagnosis. Technological developments which have improved the speed and the quality of the resulting images have also encouraged the growth of CT worldwide. The expanding use of CT worldwide has resulted in this modality becoming the major source of the population in exposure to X-rays, contributing to 40% of the resulting collective dose in the UK in 1997 [1]. Data from 1991 to 1996 show that globally CT was responsible for 34% of the annual collective dose from medical exposures [1]. Following new biological information on harm due to, radiation and modified risk estimates, the basic recommendations of the International Commission on Radiological Protection (ICRP) were revised in Publication 60, [2,3] and new quantities and concepts were introduced. In optimization, there is the important new concept of constraint, with the main implication in diagnostic radiology being that of dose constraints for the patient. CT is a radiological imaging modality which is associated with relatively high levels of effective dose exposure of the patient in comparison to conventional radiological techniques [4]. The complex conditions of irradiations in CT, involving highly collimated X-ray beams and at the same time also beam-shaping filters necessitate the use of specially defined dose descriptors, such as a CT dose index (CTDI). This is defined as the integral dose along a line parallel to the axis of rotation (z) of the dose profile (D(z)) for a single slice, divided by the nominal slice thickness T: CTDI= ( ) (mgy) (1) The European Commission (EC) has suggested the use of a normalized dose index, nctdi w, expressed as absorbed dose to air, which takes account of non-uniformities of CTDI values measured at the centre or the periphery of these phantoms: [5] (2) nctdi w = ( CTDI 0.c+ CTDI 0.p) (mgy air (mas) -1 Here C is the radiographic exposure (mas) and CTDI 0,p represents an average of measurements at four different locations around the periphery of the phantom. The weighted CT dose index, CTDI w, which is the first of the two reference dose quantities proposed by the EC, for a single slice in serial scanning or per rotation in helical scanning is then simply calculated using the following equation:[5] CTDI w = n CTDI w. C (mgy air ) (3) The second reference quantity is the dose length product (DLP), which includes the patient, or the phantom volume irradiated during a complete examination: DLP= CTDI w.t.n.c (mgy air.cm) (4) Here i represent each serial scan sequence forming part of an examination, and N is the number of slices, each with a thickness of T (cm). In the case of helical (spiral) scanning: DLP= CTDI w.t.a.t (mgy air.cm) (5) Here, for each of i helical sequences forming part of an examination, T is the nominal irradiated slice thickness (cm), A is the tube current (ma), and t is the total acquisition time (s) for the sequence [6]. Reduction of the dose received by the patient by optimization of CT technical parameters and protocols is of principle concern in patient protection [7]. In this study radiation dose arising from four common CT examinations by conventional and spiral CT machines in Khorasan province was investigated. 2. Materials and Methods This survey was performed on 7 CT scanners operating in Khorasan. Table 1 summarizes the scanners according to their manufacturers. All measurements were performed during 2009-20. Four typical CT examinations, namely routine brain,, abdomen, and, were selected for the study. For each examination, data concerning examination parameters, such as kvp, mas, number of slices, slice thickness and couch increment were taken into consideration. Table 2 presents the different examination protocols used among the 7 CT scanners for the routine brain,,, and abdomen examinations. CTDI 0,air measurements were 234 Iran J Med Phys, Vol. 9, No. 4, Autumn 2012

Patient Dose from CT Examinations made free-in-air using a cm pencil shaped ionization chamber (model RadcalX5-3CT; Radcal Corporation, Monrovia, CA) connected to a Radcal (model 9015; Radcal Corporation, Monrovia, CA) dosimeter. Table 1. Distribution of CT units per manufacturer Manufacturer Scanner Number of units GE GE 9800 1 Siemens Siemens Balance 1 GE GE Sytec 1 Shimadzu ShimadzuSCT 1 Toshiba Toshiba Asteion Multi4 1 Philips Philips 3 1 Philips Philips CX 1 Figure 1. Schematics of phantom and ionization chamber Table 2. Protocol parameters used for the routine examinations (brain,, abdomen, and ) in 7 CT scanners Scanner A B C D E F G Exam Chest kvp mas 290 140 300 300 250 165 340 340 180 144 144 144 300 150 200 200 140 90 1 1 224 280 224 224 T(mm) I(mm) L(cm) 16.1 27 30 15.6 27 30.5 16 25.7 31 31.5 16.7 26 31.5 16.5 26.3 30.5 31.5 17 26.5 30 16 26 31.5 Iran J Med Phys, Vol. 9, No. 4, Autumn 2012 235

Mohammad Taghi Bahreyni Toossi et al. The chamber was aligned parallel to the scanner axis of rotation. Two polymethyl methacrylate (PMMA) cylindrical phantoms according to Food and Drug Administration (FDA) specifications [8, 9] were used for the measurements of (CTDI 0 ) in order to calculate ( n CTDI w ). The brain phantom was 14 cm long and 16 cm in diameter, with five holes drilled parallel to its long axis, one at the centre and four around the periphery, 90 apart and 1 cm from the edge. All holes could be filled with retractable PMMA inserts. The body phantom was cm in diameter 14 cm in length, and was constructed similar to the brain phantom CTDI 0 measurements in this study were performed with the ionization chamber positioned at the four peripheral and central holes after removing the corresponding insert [5]. DLP was calculated by equation (5). 3. Results The DLP values acquired for brain,, abdomen, and examinations were in the range of 255-26, 76-1277, 48-737, and 69-854 mgy.cm, respectively. The average DLP obtained for the 7 studied machines for brain,, abdomen, and were 564, 444, 349 and 467 mgy.cm, respectively (Figure 2). Minimum, maximum and 3 rd quartile the of DLP distribution for brain,, abdomen, and are summarized in Table 3. The values of n CTDI w acquired for different machines were widly different, in some cases up to four folds, the range of figures obtained for brain,, abdomen, and were 5.7-26.2, 2.7-8.8, 2.1-9.9, and 2.1-9.9, mgy (mas -1 ), correspondingly. 1400 0 DLP mgy.cm 00 800 600 400 brain abdomen pelvic 200 0 A B C D E F G Figure 2. Measured DLP of the brain,, abdomen, and Table3. DLP Values as obtained in this study (20) Examination Dosimetry Quantity Min Max DLP 255 26 564 Chest DLP 76 1277 444 DLP 48 737 349 DLP 69 854 467 3 rd 4. Discussion The DLP value obtained for brain CT in this work (564 mgy.cm) is lower than the values reported by other researchers such as Torp et al. (740 mgy.cm) [], Hidajat et al. (587 mgy.cm) [11], Shrimpton et al (787 mgy.cm) [12] and Hatziioannou et al. (677 mgy.cm) [5]. It should be noted, however, that all these reported values are much lower than the European DRL (50 mgy.cm) [6]. The mean DLP (444 mgy.cm) acquired for CT in this study is lower than the IAEA-reported value (455 mgy.cm) [13,14], and the 236 Iran J Med Phys, Vol. 9, No. 4, Autumn 2012

Patient Dose from CT Examinations corresponding values reported by Papadimitriou et al (429 mgy.cm) for some centers in Greece and 483 mgy.cm for some centers in Italy [15]. The recommended DLP value in European guidelines for CT is 650 mgy.cm, which is higher than the corresponding value acquired in this study [6]. The wide range of results reported reveals the differences in the techniques used at different center. Since the DLP is the product of the nctdi w and the length of the patient's body it suggests that a particular examination is preformed differently in different centers, this is due to several reasons such as patient's problem, the operator's experience, and technical parameters applied and/or even the demographics for the given region (i.e., average height of the population). DLP for abdominal CT (349 mgy.cm) is comparable to the values reported by Papadimitriou et al (493 551 mgy.cm) [15]. Hidajat et al [12] reported a lower DLP (247mGy.cm) when helical technique was used which is more than half of the corresponding value obtained in this study, showing once more the significance scan length. DLP for abdominal and pelvic CT was 349 and 467mGy.cm respectively, which is lower than that reported by European guidelines (780 and 570mGy.cm) and Muhogora2009 [16]. These differences can be due to difference in exposure factors, scanner type, and technology. 5. Conclusion Patient dose in general, nctdi w and DLP values in particular are greatly affected by technical parameters applied, technicians experience and CT machine type and technology. Patient's average height and weight are also relevant factors. Acknowledgements This article is based on the data extracted from the M.Sc. desecration code no. A-3 presented to medical physics dept. of Mashhad University of Medical Sciences. The authors would like to thank the office of Vicepresident for research affair for financial support of this work. We are also greatly indebted to the technicians of the studied centers for their cooperation. References 1. Shrimpton PC, Edyvean S. CT scanner dosimetry. Br J Radiol. 1998 Jan;71(841):1-3. 2. ICRP, 1991. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann. ICRP 21 (1-3). 3. Valentin J. Managing patient dose in multi-detector computed tomography(mdct). ICRP Publication 2. Ann ICRP. 2007;37(1):1-79, iii. 4. Schauer DA, Linton OW. NCRP Report No. 160, Ionizing Radiation Exposure of the Population of the United States, medical exposure--are we doing less with more, and is there a role for health physicists? Health Phys. 2009 Jul;97(1):1-5. 5. Hatziioannou K, Papanastassiou E, Delichas M, Bousbouras P. A contribution to the establishment of diagnostic reference levels in CT. Br J Radiol. 2003 Aug;76(908):541-5. 6. Jessen K, Panzer W, Shrimpton P, Bongartzm G, Geleijns J, Golding S, et al. EUR 16262: European guidelines on quality criteria for computed tomography. Luxembourg: Office for Official Publications of the European Communities. 2000. 7. Banghart D, Thomas RH. IAEA Safety Series Report Series: No. 60: Radiation Protection in Newer Medical Imaging Techniques: Cardiac CT IAEA Safety Series Report Series: No. 61: Radiation protection in Newer Medical Imaging Techniques: CT Colonography. Radiation Protection Dosimetry. 20 January 1, 20;138(1):92-6. 8. CFR - Code of Federal Regulations Title 21. Avaliable at: http://www.fda.gov/default.htm. Accessed Feb 16, 2013. 9. Beganović A, Skopljak-Beganović A, Drljević A, Džanić S, Bošnjak J, Gazdić-Šantić M, et al. Doses to patients for computed tomography in Sarajevo. In: Dö ssel O, Schlegel W, editors. World Congress on Medical Iran J Med Phys, Vol. 9, No. 4, Autumn 2012 237

Mohammad Taghi Bahreyni Toossi et al. Physics and Biomedical Engineering, September 7-12, 2009, Munich, Germany: Springer Berlin Heidelberg; 2009. p. 252-4.. Torp C, Olerud H, Einarson G, Gron P, Leitz W, Servomaa A. Use of the EC quality criteria as a common method of inspecting CT laboratories: a pilot study by the Nordic Radiation Protection Authorities radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy: Document IAEA-CN-85-175. Vienna, Austria: International Atomic Energy Agency2001. 11. Hidajat N, Wolf M, Nunnemann A, Liersch P, Gebauer B, Teichgraber U, et al. Survey of conventional and spiral ct doses. Radiology. 2001 Feb;218(2):395-401. 12. Shrimpton PC, Hillier MC, Lewes MA, Dunn M. National Radiological Protection Board (NRPB): doses from computed tomography examinations in the UK 2003 review. Document NRPB-W67. Chilton, England: National Radiological Protection Board, 2005. 13. Agency IAE. Optimization of the radiological protection of patients undergoing radiography, fluoroscopy and computed tomography: final report of a coordinated research project in Africa, Asia and eastern Europe: International Atomic Energy Agency; 2005. 14. Agency IAE, Organization WH, Radiology ISo, Protection ICoR. Radiation Protection in Newer Medical Imaging Techniques: CT Colonography: Safety Report Series: Internat. Atomic Energy Agency; 2008. 15. Papadimitriou D, Perris A, Manetou A, Molfetas M, Panagiotakis N, Lyra-Georgosopoulou M, et al. A survey of 14 computed tomography scanners in Greece and scanners in Italy: Examination frequencies, dose reference values, effective dose and doses to organs. Radiation protection dosimetry. 2003;4(1):47-53. 16. Muhogora W, Ahmed N, Beganovic A, Benider A, Ciraj-Bjelac O, Gershan V, et al. Patient doses in CT examinations in 18 countries: initial results from International Atomic Energy Agency projects. Radiation protection dosimetry. 2009;136(2):118-26. 238 Iran J Med Phys, Vol. 9, No. 4, Autumn 2012