Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Similar documents
A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Jose Castro-Perez, Henry Shion, Kate Yu, John Shockcor, Emma Marsden-Edwards, Jeff Goshawk Waters Corporation, Milford, MA, U.S. and Manchester, UK

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Lipid Class Separation Using UPC 2 /MS

Automated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma

Achieve Broad Lipid Quantitation using a High-Throughput Targeted Lipidomics Method

Jody Dunstan, 1 Antonietta Gledhill, 1 Ailsa Hall, 2 Patrick Miller, 2 Christian Ramp 3. Waters Corporation, Manchester, UK 2

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

LC-Based Lipidomics Analysis on QTRAP Instruments

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

A Metabolomics Approach to Profile Novel Chemical Markers for Identification and Authentification of Terminalia Species

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

UPLC/MS/MS Method for the Routine Quantification of Regulated and Non-Regulated Lipophilic Marine Biotoxins in Shellfish

Simultaneous Analysis of Intact Human Insulin and Five Analogs in Human Plasma Using μelution SPE and a CORTECS UPLC Column

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

Quantification of Budesonide Using UPLC and Xevo TQ-S

New Frontiers for MS in Metabolipidomics

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

Latest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Neosolaniol. [Methods listed in the Feed Analysis Standards]

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

Dienes Derivatization MaxSpec Kit

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

The use of mass spectrometry in lipidomics. Outlines

Comparison of a UPLC Method across Multiple UHPLC Systems

Supplementary Information

Phospholipid characterization by a TQ-MS data based identification scheme

SYNAPT G2-S High Definition MS (HDMS) System

Chemical Analysis Business Operations Waters Corporation Milford MA

Increased Identification Coverage and Throughput for Complex Lipidomes

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

Supporting Information

Paul Silcock 1, Anna Karrman 2, and Bert van Bavel 2. Aim. Introduction. Sample preparation

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Profiling and Quantitation of Metabolomic Signatures for Breast Cancer Cell Progression

LC/MS/MS SOLUTIONS FOR LIPIDOMICS. Biomarker and Omics Solutions FOR DISCOVERY AND TARGETED LIPIDOMICS

Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids

Detection of Cotinine and 3- hydroxycotine in Smokers Urine

[ APPLICATION NOTE ] A Generic Kit-Based Approach for LC-MS/MS Quantification of Urinary Albumin for Clinical Research APPLICATION BENEFITS

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Metabolomic and Lipidomic Research in Diabetes and Obesity 08/07/2013

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

[ application note note ] ]

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Mass Spectrometry based metabolomics

CHARACTERIZATION AND DETECTION OF OLIVE OIL ADULTERATIONS USING CHEMOMETRICS

Analysis of drugs of abuse in biological matrix using Time of Flight technology

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

[ APPLICATION NOTE ] The Separation of 8 -THC, 9 -THC, and Their Enantiomers by UPC 2 Using Trefoil Chiral Columns INTRODUCTION APPLICATION BENEFITS

Discovery Metabolomics - Quantitative Profiling of the Metabolome using TripleTOF Technology

Supporting Information (SI)

Key Advantages of Comprehensive Cannabis Analysis

Single Quadrupole Compact Mass Spectrometer for Liquid Sample

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Obtaining Answers to Biological Questions Sample Prep to Data Analysis. Jeremiah D. Tipton, Ph.D. SCIEX Advanced Workflow Specialist in OMICS

Relative Quantitation of Human Polymorphonuclear Leukocyte Cell Membrane GPEtn Lipids

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Modernizing the Forensic Lab with LC-MS/MS Technology

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations

The use of mobile-device data to better estimate dynamic population size for wastewater-based epidemiology

Supplementary Information. Effects of Perfluorooctanoic Acid on Metabolic Profiles in Brain and Liver of Mouse by a

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Transcription:

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS Mark Ritchie and Evelyn Goh Waters Pacific Pte Ltd., Singapore A P P L I C AT ION B E N E F I T S Delivers a rapid 10-min MRM method with Xevo TQ MS for 194 lipids in serum utilizing both positive and negative ESI modes. Achieves a chloroform-free, high performance, reverse phase ACQUITY UPLC separation of complex lipid mixtures. Provides a streamlined workflow for instrument setup, analysis, and results processing using IntelliStart, Quanpedia, and TargetLynx. Waters solutions ACQUITY UPLC System INTRODUCTION The field of lipidomics is gaining momentum at an incredible pace. The biological knowledge and potential applications are becoming very attractive to those seeking to better understand disease, and for those looking to develop diagnostics. From an analytical science perspective, there is not yet a dominant technique or technology platform, and many researchers have adapted workflows from other life science applications to solve their problems. In many laboratories, samples are analyzed directly after sample preparation by ESI-MS without separation, which does provide short acquisition times but it can also create problems. The dynamic range and sensitivity of the experiment is undoubtedly affected by the complexity of the biological sample. Ion suppression from the most abundant analytes and contaminants must be considered and this is compounded when salt and adduct formation are taken into account. These effects combined limit the detection and quantification of isobaric and low abundance lipid species. Specificity and quantitation accuracy may also be affected when dealing with lipids in a non-separated mixture. Lipids are a diverse class of related species that differ through minor variations in structure. Lipids within the same class are typically distinct from each other by 2 Da, and as such are easily separated; however, inter-class interference is possible. Xevo TQ MS IntelliStart Software Quanpedia New method set up IntelliStart Rapid MRM compound tuning using on board fluidics Utilizing existing method Quanpedia Tuning, LC, and MS methods loaded from applications library TargetLynx Application Manager MassLynx Software Xevo TQ MS and ACQUITY UPLC For high sensitivity, high quality separation and speed key words Lipidomics, reversed phase separation, MRM, workflow TargetLynx simplified data analysis method set up, rapid data analysis TargetLynx Data analysis method loaded from Quanpedia Figure 1. Workflow for designing MRM experiments and utilization of Quanpedia for established methods. 1

E X P E R IM E N TA L Sample preparation Monkey plasma was extracted with cold methanol, and the soluble supernatant was dried and re-dissolved in chloroform/methanol (1:1). After adding non-biological lipids for sample normalization (the same lipids used in tuning), the sample was diluted 10x with acetonitrile. Five µl of injected sample was equivalent to 0.14 µl of raw serum and the final solvent composition was injected in acetonitrile/ chloroform/methanol (90:5:5). Triplicate injections were analyzed. LC conditions LC conditions were previously published 1,2 LC system: Waters ACQUITY UPLC Column: ACQUITY HSS T3, 2.1 x 100 mm, 1.8 mm Column temp.: 65 C Flow rate: 500 µl/min Mobile phase A: Acetonitrile/water (40:60) with 10 mm ammonium acetate ph 5.0 Mobile phase B: Acetonitrile/isopropanol (10:90) with 10mM ammonium acetate ph 5.0 Gradient: 40% to 100% B/10 min MS conditions MS system: Xevo TQ MS Ionization mode: Positive/negative switching MS system tuning IntelliStart is used to assist the user in the generation of a new analysis, as shown in Figure 1. Assisted MRM tuning for phospholipids was performed for each class with IntelliStart using mixed standards, as shown below in Table 1. For established methods, all of the required experimental parameters including the data analysis template were automatically loaded from Quanpedia. Compound Formula Mass Parent m/z Cone voltage Daughter m/z Collision Energy Ionization mode DMPA 592.5 591.5 50 153.0 32 ESI- DMPG 666.6 665.5 50 153.0 34 ESI- DMPS 679.6 678.5 42 591.5 22 ESI- DMPC 677.5 678.6 36 184.1 30 ESI+ L-SM 646.5 647.6 36 184.1 24 ESI+ C17-Cer 551.5 552.6 20 264.2 26 ESI+ Glu-Cer 587.5 588.5 20 264.2 20 ESI+ DMPE 635.5 636.5 26 495.5 18 ESI+ Lyso PC 551.5 552.5 42 184.1 26 ESI+ Lyso PE 435.5 452.3 26 313.2 18 ESI+ Table 1. Summary of IntelliStart automated tuning report. ESI+ C8 PI 586.5 585.4 48 241.0 30 ESI- ESI- Lipid Class No. MRMs Lipid Class No. MRMs Lyso PE 10 Lyso PS 4 Lyso PC 8 Lyso PI 4 Ceramides 8 PIs 22 Glucoceramides 13 PAs 11 PEs 38 PGs 11 Sphingomyelins 11 PSs 20 PCs 36 Table 2. Summary of classes of lipid MRMs used. Capillary voltage: 3.2 kv Desolvation temp.: 500 C Source temp.: 150 C Acquisition method: Multiple reaction monitoring (MRM) Collision cell pressure: 3.5 x 10-3 mbar 2

RESULTS AND DISCUSSION Extracted ion chromatograms for target lipids were easily generated using TargetLynx. Mass chromatograms visualized in the MassLynx chromatogram window can be used to generate quantification methods in TargetLynx, with a single mouse click to transfer retention time, channel, and MRM information. Additional information may be entered to normalize or quantify from an internal or external standard. Once generated, the TargetLynx method can be reused for later experiments. Figure 2. Setup for TargetLynx automated data extraction. 1.) Right click and drag across a peak of interest in the chromatogram browser. 2.) MRM transition is automatically added to the compound information. 3.) Observed retention time is automatically added. 4.) Measured area can be normalized, or converted to an absolute value by comparison to standard compound. Targeted phospholipids were measured over approximately five orders of dynamic range. Areas under the extracted mass chromatograms were exported from TargetLynx, and selected lipid intensities are shown in Figure 3. The average %RSD between the three replicates for these intensity measurements was under 10%. Figure 3. Intensities of selected serum lipids. 3

As previously noted 1,, 2 this chromatographic method provides a reliable and reproducible separation of lipids. From this data it is also apparent that lipids in the same class can be separated cleanly and predictably. Figure 4 shows the separation of unsaturated ceramides and glucoceramides as an example. As the aliphatic chain is extended by two carbons, approximately 30 seconds is added to the retention time. Errors in quantification of lipids are often seen in infusion experiments due to overlapping isotopic patterns of lipids that differ by only a single saturated bond. In these cases, the mono-isotopic ion of the higher mass lipid species overlaps with the M+2 isotope of the lower and therefore shows increased abundance for that lipid. In this UPLC method, however, this type of variation in lipids results in significant and predictable chromatographic separation, as shown in Figure 5. In the example shown here, PC 36:1, PC 36:2, PC 36:3, PC 36:4, and PC 36:5 were each separated by approximately 20 s from their closest relative, with the most saturated being the earliest eluting species. In this way, the problem of overlapping isotopic patterns can be overcome. Figure 4. Retention times of taturated ceramides and glucoceramides show predictable elution patterns based on aliphatic carbon chain length when separated by reverse phase UPLC. Figure 5. Retention times of differentially saturated PCs show predictable elution patterns relating to the degree of saturation when separated by reverse phase UPLC. This helps to distinguish isotopic interference brought about by the m/z difference of 2 Da, where the monoisotopic ion of the higher mass (less saturated) lipid species overlaps with the M+2 isotope of the higher mass (less saturated) lipid species, which would affect quantitation in un-separated mixtures. 4

CONCLUSIONS The ACQUITY UPLC/Xevo TQ MS System offers lipidomics researchers the opportunity to achieve high performance reverse phase separation of complex lipid mixtures with highly sensitive and accurate detection. The reverse phase method uses more stable and less toxic solvents than the usual normal phase separation methods applied to lipidomics, and it is able to differentiate between similar and isotopically interfering species. References: 1. Castro-Perez JM, Kamphorst J, DeGroot J, Lafeber F, Goshawk J, Yu K, Shockcor J, Vreeken RJ, Hankemeier T. Comprehensive LC-MS E Lipidomic Analysis using a Shotgun Approach and Its Application to Biomarker Detection and Identification in Osteoarthritis Patients. J. Proteome Res., 2010, 9 (5), pp 2377 2389. 2. Shockcor J, Crowe H, Yu K, and Shion H. Analysis of Intact Lipids from Biological Matrices by UPLC/High Definition MS. Waters Application Note 720003349EN. In addition to the benefits of separation of similar species, UPLC s concentration effect and reduction in ion suppression improve the performance dramatically compared to infusion or HPLC experiments. The sensitivity of the technique is apparent as the sample load required by this approach was 1/10th of the sample load that would typically be used in an infusion experiment. The use of IntelliStart for rapid automated tuning, Quanpedia to generate methods, and TargetLynx for data review provide a simple workflow for efficiently extracting qualitative and quantitative information from large data sets. Waters, UPLC, ACQUITY UPLC, and Xevo are registered trademarks of Waters Corporation. The Science of What s Possible, IntelliStart, Quanpedia, TargetLynx, and MassLynx are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2011 Waters Corporation. Produced in the U.S.A. July 2011 720004035en AG-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com