Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Similar documents
MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Feasibility of portal dosimetry for flattening filter-free radiotherapy

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Optimized Varian asi portal dosimetry: development of datasets for collective use

Leila E. A. Nichol Royal Surrey County Hospital

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

EPID dosimetry for pretreatment quality assurance with two commercial systems

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

iviewdose Confidence and assurance in dose delivery

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

Dosisverifikation mit Delta 4 Discover während der Behandlung

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

Small field diode dosimetry

The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

SunCHECK Patient Comprehensive Patient QA

Quality assurance in external radiotherapy

Portal dosimetry in wedged beams

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

The impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

IMRT QA: Point Dose Measurements or 2D Array?

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Tuning of AcurosXB source size setting for small intracranial targets

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

University, Vellore, Tamilnadu, India, ABSTRACT

Flattening Filter Free beam

Gamma analysis dependence on specified low-dose thresholds for VMAT QA

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Statistical process control analysis for patient-specific IMRT and VMAT QA

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Elekta Synergy Digital accelerator for advanced IGRT

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Technique For Plan Quality and Efficiency Using VMAT Radiosurgery For Patients with Multiple Brain Metastases

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

REPLACING PRETREATMENT VERIFICATION WITH IN VIVO EPID DOSIMETRY FOR PROSTATE IMRT

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Results of a 10-year survey of workload for 10 treatment vaults at a high-throughput comprehensive cancer center

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

A new geometric and mechanical verification device for medical LINACs

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

EORTC Member Facility Questionnaire

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Follow this and additional works at:

PMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.

Case Study. Institution Farrer Park Hospital

biij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *

Volumetric Modulated Arc Therapy - Patient Specific QA

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

The effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Baotian Huang 1, Lili Wu 1, Peixian Lin 2 and Chuangzhen Chen 1*

Reliability of the gamma index analysis as a verification method of volumetric modulated arc therapy plans

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

IMRT Planning Basics AAMD Student Webinar

Learning objective. Outline. Acknowledgements. KV CBCT Imaging Part I. R Hammoud AAPM 2008 CE-Therapy (SAM) 1

Clinical Precision for Best Cancer Care. Dee Mathieson Senior Vice President Oncology Business Line Management

Radiotherapy In-vivo Treatment Verification: Multi-center analysis for more than 6 million fractions delivered

The Journey of Cyberknife Commissioning

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Patient and Linac QA - present and the future

Limits of Precision and Accuracy of Radiation Delivery Systems

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

DELIVERED PATIENT DOSE

Realtime dose monitoring technologies: Luxury or Necessity

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

Eric E. Klein, Ph.D. Chair of TG-142

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

The performance of the progressive resolution optimizer (PRO) for RapidArc planning in targets with low-density media

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Performance Evaluation of Calypso (R) 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

Diode calibration for dose determination in total body irradiation

Application of asi-kvcbct for Volume Assessment and Dose Estimation: An Offline Adaptive Study for Prostate Radiotherapy

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

On the use of bolus for pacemaker dose measurement and reduction in radiation therapy

Transcription:

Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Zhigang Xu Jinkoo Kim James Han An Ting Hsia Samuel Ryu Department of Radiation Oncology, Stony Brook University Hospital, Stony Brook, NY, USA Author to whom correspondence should be addressed. Zhigang Xu E-mail: Zhigang.xu@sbumed.org; Telephone: (631) 444-3617. Abstract In this study we investigated the dose rate response characteristics of the Digital Megavolt Imager (DMI) detector, including panel saturation, linearity, and imager ghosting effects for flattening filter-free (FFF) beams. The DMI detector dose rate response characteristics were measured as a function of dose rate on a Varian True- Beam machine. Images were acquired at dose rates ranging from 400 to 1400 MU/ min for 6XFFF and 400 to 2400 MU/min for 10XFFF. Line profiles and central portal doses derived from the images were analyzed and compared. The linearity was verified by acquiring images with incremental Monitor Unit (MU) ranging from 5 to 500 MU. Ghosting effects were studied at different dose rates. Finally, for validation, test plans with optimal fluence were created and measured with different dose rates. All test plans were analyzed with a Gamma criteria of 3%-3 mm and 10% dose threshold. Our study showed that there was no panel saturation observed from the profile comparison even at the maximum dose rate of 2400 MU/min. The central portal doses showed a slight decrease (1.013 1.008 cgy/mu for 6XFFF, and 1.020 1.009 cgy/mu for 10XFFF) when dose rate increased (400 1400 MU/min for 6XFFF, and 400 2400 MU/min for 10XFFF). The linearity of the DMI detector response was better than 0.5% in the range of 20 500 MU for all energies. The residual image was extremely small and statistically undetectable. The Gamma index measured with the test plans decreased from 100% to 97.8% for 6XFFF when dose rate increased from 400 to 1400 MU/min. For 10XFFF, the Gamma index decreased from 99.9% to 91.5% when dose rate increased from 400 to 2400 MU/min. We concluded that the Portal Dosimetry system for the TrueBeam using DMI detector can be reliably used for IMRT and VMAT QA for FFF energies. PACS 87.53.Bn, 87.53 Jw, 87.55.Qr, 87.56.Fc KEY WORDS FFF, PDC, IMRT, VMAT, EPID, QA ---------------------------------------------------------------------------------------------------------------------------------------------------------------------- This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. J Appl Clin Med Phys 2018; 19:4:141 147 wileyonlinelibrary.com/journal/jacmp 141

142 XU ET AL. 1 INTRODUCTION The Electronic Portal Imaging Devices (EPID) is a very useful device for routine clinical use because of its prompt setup, easy data acquisition, and high resolution. The Portal Dosimetry is an application integrated on the ARIA electronic medical record system (Varian Medical Systems VMS, Palo Alto, CA, USA) that allows IMRT or VMAT field verification. The dosimetry process consists of three steps: (a) Fluence prediction using Eclipse Portal Dose Calculation (PDC) algorithm, (b) Fluence acquisition using dosimetry mode on the linac, and (c) Fluence comparison using the Portal Dosimetry module in Aria. The Portal Dosimetry adds a fast and efficient workflow for the verification of IMRT and VMAT plans. 1 Since the option of removing the flattening filter (FF) in the linacs for IMRT and VMAT treatments was introduced in 2010, there has been a lot of interest generated in using flattening filter-free (FFF) beams which give the benefit of reduced headscatter and hence, reduced dose outside the field. 2 These beams also deliver dose faster than flattened beams, which would be beneficial for hypofractionated treatments by reducing treatment time and potential intrafractional organ motion. 3 The Digital Megavolt Imager (DMI) detector is now a standard MV imaging detector installed on all Varian TrueBeam machines. The DMI detector offers not only the possibility to image large field size (43 9 43 cm 2 ) but also the images with a higher pixel resolution (1280 9 1280) than the older detectors (i.e., IDU20, as1000). The detector area used for dosimetry measurements (integrated images) is a little smaller than the complete imaging size (40 9 40 cm 2 with 1190 9 1190 pixel). 4 With faster readout electronics and a higher pixel capacitance, the DMI detector allows for much higher dose rate than the older detectors (i.e., IDU20, as1000). It has been adapted by Varian for use in FFF beams at any source-to-detector distance. 5 Since the dose rates for FFF beams are up to six times higher than for conventional flattened beams, portal images taken at maximum FFF dose rate may saturate the images. The EPID saturation occurs when no additional photocurrent outputs from the photodiode as the incident optical power increases. There are a number of parameters that can affect the saturation limit. However, the detailed information of the hardware and software components is proprietary and unknown to the public. Several studies investigated the feasibility of using Portal Dosimetry to FFF beams for IMRT and VMAT plan verification. 6 10 Chuter et al. studied the Portal Dosimetry of Elekta Synergy linear accelerator (Elekta, AB, Stockholm, Sweden) with an Agility collimator capable of delivering FFF beams. They found that, with iviewgt amorphous silicon (asi) panel (RID 1640 AL5P) at a fixed SSD of 160 cm, images taken at maximum FFF dose rate could saturate the EPID. A dose rate of 800 MU/min was found not to saturate the EPID for open fields. 6 Pardo et al. presented a method for FFF Portal Dosimetry by placing the EPID at a greater distance to avoid panel saturation with Varian TrueBeam STX 1.6 and an as1000 model. 7 Miri et al. studied the Portal Dosimetry of a TrueBeam with an as1200 panel for flattened filter (FF) and flattening filter-free (FFF) beams. They studied the linearity of dose response with MU, the imager lag, and the effectiveness of backscatter shielding. They concluded that significant improvements were observed in the dosimetric response of the as1200 imager compared to previous imaging detectors (IDU20, as1000). 5 To the best of our knowledge, the study of the dose rate response characteristics for the DMI detector has not been reported in the literature. The purpose of this work was to study the panel saturation as a function of the dose rate, including dose linearity and imager ghosting effects of the DMI detector for FFF beams. 2 MATERIALS AND METHOD 2.A Portal dosimetry commissioning The Portal Dosimetry was used to acquire delivered dose images, while Eclipse was used to compute a corresponding dose distribution. Both images were viewed and quantitatively compared in the Portal Dosimetry. The PDC was used to calculate portal dose images for fields containing fluences as part of a pre-treatment verification for IMRT and VMAT planning. 11 The portal imager was calibrated in dosimetric acquisition mode at isocenter according to the official Portal Dosimetry calibration procedure. 12 The procedure includes applying a dark and flood-field correction, an absolute calibration, and a beam profile correction. The absolute calibration is defined using 100 MU to correspond to 1 calibrated unit (CU) with a 10 9 10 cm 2 field. The beam profile correction was made using diagonal profiles measured in a water phantom at a depth of maximum dose (dmax). During the commissioning of Portal Dosimetry, all calibration datasets were measured at 400 MU/min to eliminate potential panel saturation. The configuration of the Eclipse PDC algorithm requires output factor measurements measured with the DMI panel, the beam intensity profile and a set of pyramid-shaped test images from which the algorithm configuration derives the DMI pencil beam kernel. 12 The pyramid-shaped fluence provided by Varian is an optimal fluence specially designed for the configuration and verification of PDC. The fluence consists of a port of 120 9 250 mm with five rectangular slabs of the intensity of 1.0 and fluence of 0.0 outside the slabs. Output factors were acquired with the imager panel at isocenter for field sizes ranging from 1 9 1to38938 cm 2. Figure 1 displays the output factors as a function of square field size for all the energies on our TrueBeam machine. The beam intensity profiles were derived from the previously acquired diagonal beam profiles at dmax in a water phantom for the configuration of the Eclipse AAA photon dose calculation algorithm. After the PDC was configured as described above, the Portal Dosimetry was tested with fields that have pyramid-shaped fluence imported. The actual fluences were then recalculated using the Eclipse Smart Leaf Motion Calculation (LMC 13.6.23), and the portal dose images were subsequently predicted, measured, and analyzed for all energies. All of the measurements met the Varian acceptance criteria: Dose Difference: 4%, Distance-To-Agreement: 4 mm,

XU ET AL. 143 rates from 400 to the maximum of 1400 and 2400 MU/min for 6XFFF and 10XFFF, respectively, in six steps, delivering a 100 MU 20 9 20 cm 2 field. Line profiles and central portal doses derived from the images were compared and analyzed. The central portal doses were measured as the average of pixel value in the center of the field, using the output factor tool of the Portal Dosimetry. 2.C The linearity of the DMI detector response F IG. 1. size. Output factors are plotted as a function of square field The linearity of the DMI detector response as a function of MU was investigated by delivering a 10 9 10 cm 2 field with MU ranging from 5 to 500 for both of the unflattened modes (6XFFF and 10XFFF) and the conventional 6X mode, for clinical dose rates at 1400, 2400, and 400 MU/min, respectively. Gamma pass rate: >99%. The Portal Dosimetry user interface is shown in Fig. 2. 2.B Profile and central point dose measurement as function of dose rate The DMI detector dose rate response characteristics were studied as a function of dose rate. Images were acquired at dose 2.D The dose rate response of ghosting effects The dose rate response of ghosting effects was studied by measuring the imager lag after delivering 500 MU at different dose rates for all energies. A 10 9 10 cm 2 treatment field was acquired immediately after exposing the DMI detector using a 20 9 20 cm 2 field. The time delay between two consecutive images was approximately 30 s which is the time delay required by the TrueBeam machine. The residual image (ghost) in the F IG. 2. Portal Dosimetry user interface: Predicted Dose (up-left), Evaluated Dose (up-center), Portal Dose (up-right), Profiles (bottom-left), and Histogram (bottom-right).

144 XU ET AL. 20 cmx 20 cm 10 cmx 10 cm 10 cmx 10 cm 100 cm SID, 500 MU F IG. 3. The residual image (ghost) was studied by measuring imager lag after 500 MU delivery in the periphery region (four points in red). periphery region outside the central 10 9 10 cm 2, but within the 20 9 20 cm 2 treatment field, was carefully checked by comparing the portal doses in the same region that was acquired after a long time interval without image acquisitions (see Fig. 3 for illustration). The residual image (RI) is then RI ¼ PD After PD Before PD Center 100% (1) where PD Before and PD After are the portal doses before and after 20 9 20 cm 2 treatment field is delivered, respectively. PD Center is the portal dose at filed center for 10 9 10 cm 2 field size. An average of F IG. 4. Portal dose image (a) and line profiles (b) at different dose rates for 6XFFF.

XU ET AL. 145 RI calculated based on four points at 8 cm away from the center was used to determine the residual signals. Finally, plans with pyramid-shaped fluence were created. Portal images of the field were acquired at 6XFFF and 10XFFF energies delivered at different dose rates. All test plans were analyzed with a gamma criteria of 3%-3 mm and 10% dose threshold, which is a typical gamma criteria for IMRT QA in our institution. 3 RESULTS AND DISCUSSION The portal dose image and their line profiles for 100 MU and 20 9 20 cm 2 field size for 6XFFF at different dose rates are shown in Figs. 4(a) and 4(b), respectively. The corresponding image and line profiles of 10XFFF are shown in Figs. 5(a) and 5(b). No panel saturation was detected even with a maximum dose rate of 2400 MU/min for 10XFFF by comparing the profiles with different dose rates. Figure 6 displays the central portal dose as the function of dose rate for 100 MU and 20 9 20 cm 2 field size for energies 6XFFF (in red) and 10XFFF (in blue). As mentioned above, the central portal dose was measured as the average of pixel value in the center of the field. This was done using the output factor tool of the Portal Dosimetry. A slight decrease in central portal dose, from 1.013 to 1.008 cgy/ F IG. 6. Central portal doses as a function of dose rate for 100 MU and 20 9 20 cm 2 field size. MU, was observed when the dose rate increased from 400 to 1400 MU/min for 6XFFF. A similar decrease, from 1.020 to 1.009 cgy/mu, was measured for 10XFFF when the dose rate increased from 400 to 2400 MU/min. However, both dose reductions are less than 1.09%, meaning the dose rate effect is clinically insignificant. Results of the linearity tests are shown in Fig. 7 (a) for the conventional flattened beam (6 MV) and unflatten beams (6XFFF F IG. 5. Portal dose image (a) and line profiles (b) at different dose rates for 10XFFF.

146 XU ET AL. F IG. 7. (a, b) Image response is linear in wide range of MU (20 500MU). T ABLE 1 Gamma analysis for test plans with different dose rates (3%-3 mm). Dose rate (MU/min) 6XFFF Dose rate (MU/min) 10XFFF 400 100 400 99.9 600 99.9 800 99.8 800 99.8 1200 99.2 1000 99.7 1600 97.5 1200 98.8 2000 95.5 1400 97.8 2400 91.5 and 10XFFF). All dose rates were calibrated to yield 1 CU per 100 MU. The linearity was better than 0.5% in wide range of MU (20 500 MU) for all energies as shown in Fig. 7 (b). The measured residual image signals at different dose rates for 6XFFF and 10XFFF were extremely small (<0.1%) and statistically undetectable after delivering 500 MU. Finally, the results of the Gamma analysis (3%-3 mm) for test plans with different dose rates are summarized in Table 1. The Gamma agreement index was decreased from 100% to 97.8% when dose rate increased from 400 to 1400 MU/ min for 6XFFF, and from 99.9% to 91.5% when dose rate increased from 400 to 2400 MU/min for 10XFFF. As the data shown in Fig. 6 for 10XFFF, only a slight decrease (about 1%) of central portal dose was observed for an open field when the dose rate increased from 400 to 2400 MU/min. The decrease in Gamma agreement index here is mainly due to the dose rate response of the Multileaf Collimator (MLC) motion. Increasing the dose rate increases the number of control points per unit time and, thus, it increases the complexity of the MLC delivery. 0.12 CU 0.10 CU 0.05 CU 0.00 CU -0.05 CU -0.10 CU -0.15 CU -0.20 CU -0.25 CU -0.30 CU 0.08 CU 0.05 CU 0.00 CU -0.05 CU -0.10 CU -0.15 CU -0.20 CU -0.25 CU -0.30 CU -0.32 CU F IG. 8. Portal dose difference images for 10XFFF: 400 MU/min dose rate (top) and 2400 MU/min dose rate (bottom).

XU ET AL. 147 Lowering the dose rate helps to improve the gamma pass rate. This is in agreement with the work done by Kaviarasu et al. 13 Figure 8 shows the portal dose difference images (measured image predicated image) for 10XFFF with dose rate of 400 MU/min (top) and 2400 MU/min (bottom). The highest dose difference was appeared near the 1 2 cm bands with a bigger difference for 2400 MU/min dose rate beam. With the consideration of the extreme dose gradients of the high-dose rate beams, the low passing rate near the short band width is clinically acceptable. In general, the fluence patterns of clinical plans will have smoother profiles and the machine is unlikely to be running at a constant maximum dose rate of 2400 MU/min. 4 CONCLUSION The Portal Dosimetry utilizing EPID is an existing technique that has been shown to work well with standard, flattened radiotherapy beams. With the emerging use of high-dose-rate FFF radiotherapy there is an associated need to verify these treatments efficiently. This study has shown that the DMI panel saturation was clinically insignificant even at the maximum dose rate of 2400 MU/min. The linearity of the DMI detector response was better than 0.5% in the range of 20 500 MU for all energies. The residual image (ghost) was extremely small and statistically undetectable. With faster readout electronics and a higher pixel capacitance, the DMI detector allows for a much higher dose rate than the older detectors (i.e., IDU20, as1000) without saturation. Therefore, we conclude that the Portal Dosimetry system for the TrueBeam with the DMI detector can be reliably used for IMRT and VMAT pretreatment QA verification for FFF energies. CONFLICTS OF INTEREST The authors do not have any conflicts of interest to declare. REFERENCES 1. van Elmpt W, McDermott L, Nijsten S, et al. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol. 2008;88:289 309. 2. Fogliata A, Garcia R, Kn o os T, et al. Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy. Med Phys. 2012;39:6455 6464. 3. Vassiliev ON, Titt U, P onisch F, et al. Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol. 2006;51:1907 1917. 4. Varian manual. TrueBeam Technical Reference Guide Volume 2: Imaging, P1005924-001-A. 5. Miri N, Keller P, Zwan BJ, et al. EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams. J Appl Clin Med Phys. 2016;17:292. 6. Chuter RW, Rixham PA, Weston SJ, et al. Feasibility of portal dosimetry for flattening filter-free radiotherapy. J Appl Clin Med Phys. 2016;17:112. 7. Pardo E, Novais JC, Lopez MYM, et al. On flattening filter-free portal dosimetry. J Appl Clin Med Phys. 2016;17:132. 8. Tyner E, McClean B, McCavana P, et al. Experimental investigation of the response of an a-si EPID to an unflattened photon beam from an Elekta Precise linear accelerator. Med Phys. 2009;36: 1318 1329. 9. Nicolini G, Clivio A, Vanetti E, et al. Evaluation of an asi-epid with flattening filter free beams: applicability to the GLAaS algorithm for portal dosimetry and first experience for pretreatment QA of RapidArc. Med Phys. 2013;40:111719. 10. Nicolini G, Clivio A, Vanetti E, et al. Dosimetric testing of the new as1200 MV imager with FF and FFF beams [poster]. Radiother Oncol. 2015;115:S439. 11. van Esch A, Depuydt T, Huyskens D. The use of an asi-based EPID for routine absolute dosimtric pre-treatment verification of dynamic IMRT fields. Radiother Oncol. 2004;71:223 234. 12. Varian manual. TrueBeam 2.0 Administrators Guide, ID 100058502. 13. Kaviarasu K, Raj NA, Murthy KK, et al. Impact of dose rate on accuracy of intensity modulated radiation therapy plan delivery using the pretreatment portal dosimetry quality assurance and setting up the workflow at hospital levels. J Med Phys. 2015;40:226 232.