Radiation Protection Dosimetry Advance Access published April 22, 2008

Similar documents
EVALUATION OF EXPOSURE INDEX (lgm) IN ORTHOPAEDIC RADIOGRAPHY L. Lança 1, * and A. Silva 2

THE LEVEL OF COMPLIANCE OF SELECTED NIGERIAN X-RAY DEPARTMENTS TO EUROPEAN GUIDELINES ON GOOD RADIOGRAPHIC TECHNIQUES

Radiation exposure of the Yazd population from medical conventional X-ray examinations

A Standardised Approach to Optimisation

Radiation Protection Dosimetry Advance Access published September 13, 2006

Assessing the standard dose to standard patients for x-ray investigations

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

Trends in examination frequency and population doses in Norway,

Diagnostic reference levels as a quality assurance tool

Quality Control and Patient Dosimetry on line for Computed Tomography

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Research Article. Corresponding author H. Osman

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

JMSCR Vol 06 Issue 12 Page December 2018

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

European Guidelines on DRLs for Paediatric Imaging

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

Patient dose in routine X-ray examinations in Yazd state

Introduction to RP154 methology for population dose estimates

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

ABSTRACT I. INTRODUCTION

International Journal of Radiation Research, January 2017

Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray

Survey of patients CT radiation dose in Jiangsu Province

Evaluation of radiation dose and image quality following changes to tube potential (kvp) in conventional paediatric chest radiography

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Doses Received by Patients during Thorax X-Ray Examinations

Patient doses in general radiography examinations

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Receptor dose and Patient Dose in Radiographic exposures - a 15 year review

Iranian Journal of Medical Physics

The use of gonadal shielding in singular common diagnostic radiographic procedures

Radiation Protection education of Radiology trainees

Limited information related to: - Type and frequency of procedures; - Exposure levels involved.

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Radiation Protection- Cath lab

The impact of SID and collimation on backscatter in radiography

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Local reference dose evaluation in conventional radiography examinations in Iran

The effect of breast shielding during lumbar spine radiography

Shielding Calculation: Radiographic Room. Jerry Williams

Trends in Patients Exposure Doses during Radiographic Examination in Japan

Current status of diagnostic imaging in dental university hospitals in Japan

Radiation Safety. - Orientation Session - Environmental Health, Safety & Risk Management Health Physics Services Unit

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

Radiologic Technology Training Program

Dosimetric Consideration in Diagnostic Radiology

Radiation Doses to Patients Undergoing Selected Radiographic Examinations in Three Hospitals South-West, Nigeria

RADIATION DOSES TO PATIENTS UNDERGOING BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS

Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp Advance Access publication 23 August 2008

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED *

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

1 DECISION 1 (6) REFERENCE LEVELS FOR THE PATIENT S RADIATION EXPOSURE FOR PAEDIATRIC X-RAY EXAMINATIONS

The Fluoroscopic Technique for Monitoring Distraction of a Non-Invasive Lengthening Device in Early Onset Scoliosis

Early in Life. Noboru Takamura, M.D., Ph.D.

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Radiation Safety in the Modern Radiology Department: A Growing Concern

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Pilot study of patient and phantom breast dose measurements in Bulgaria

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Scientific Exhibit Authors:

The print quality of this copy is not an accurate representation of the original.

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

A STUDY ON SOME PHYSICAL PARAMETERS RELATED TO IMAGE QUALITY AND RADIATION SAFETY IN DIAGNOSTIC RADIOLOGY

Catalog Addendum

Radiation Protection Dosimetry Advance Access published February 17, Radiation Protection Dosimetry (2011), pp. 1 9

Entrance Surface Dose from Pediatric Diagnostic X-ray Examinations in a Developing World Setting: Are We ALARA Principle Compliant?

DIAGNOSTIC IMAGING (DMI)

"Optimal Dose Techniques and Image Quality: Can We Have Both?" Lorusso, J. R., Fitzgeorge, L., Lorusso, D., & Lorusso, E. 2014

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

Radiation Quantities and Units

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

HEALTH SCIENCES AND ATHLETICS Institutional (ILO), Program (PLO), and Course (SLO) Alignment

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Influence of the exposure time in the area monitors at radiodiagnostic

Attention. Therefore, all the data and statements made in this presentation are preliminary and might change in the future.

Highlights and Conclusions Thematic Area 4: Radiation of Patients

Diagnostic Reference Levels (DRLs)

Coaching Reduced the Radiation Dose of Pain Physicians by Half during Interventional Procedures

RADIATION PROTECTION N 180

Diagnostic reference levels in intraoral dental radiography in Korea

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety

Exploring the use of the Swiss medical tariffication codes (TARMED) in the establishment of the frequency of radiodiagnostic examinations

Patient dose measurement in common medical X-ray examinations in Iran

Managing Patient Dose in Computed Tomography (CT)

ROMANIA MEDICAL EXPOSURE IN 2016 IN ROMANIAN RADIOLOGICAL DEPARTMENTS. Olga GIRJOABA and Alexandra CUCU

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

Ionizing Radiation Exposure of the Population of the United States. David A. Schauer Executive Director

CT Radiation Risks and Dose Reduction

Application of international standards to diagnostic radiology dosimetry

Transcription:

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE STUDY WITH EUROPEAN GUIDELINES L. Lança 1, *, A. Silva 2, E. Alves 3, F. Serranheira 1 and M. Correia 1 1 School of Health Technology, Lisbon Polytechnics, Lisbon, Portugal 2 Department of Electronic, Telecommunications and Informatics, Aveiro University, Aveiro, Portugal 3 Nuclear and Technological Institute, Physics Sector, Lisbon, Portugal Typical distribution of exposure parameters in plain radiography is unknown in Portugal. This study aims to identify exposure parameters that are being used in plain radiography in the Lisbon area and to compare the collected data with European references [Commission of European Communities (CEC) guidelines]. The results show that in four examinations (skull, chest, lumbar spine and pelvis), there is a strong tendency of using exposure times above the European recommendation. The X-ray tube potential s (in kv) are below the recommended s from CEC guidelines. This study shows that at a local level (Lisbon region), radiographic practice does not comply with CEC guidelines concerning exposure techniques. Further national/local studies are recommended with the objective to improve exposure optimisation and technical procedures in plain radiography. This study also suggests the need to establish national/local diagnostic reference levels and to proceed to effective measurements for exposure optimisation. INTRODUCTION Patient exposures in diagnostic radiology are increasing at an alarming rate for certain radiographic, fluoroscopic and computed tomography (CT) examinations (1). According to the UNSCEAR 2000 report, the overall mean effective dose per examination has increased by about 20% and the annual collective effective dose by nearly 50%, from 1991 to 1996 (2). Significant differences in national practices with medical radiation exposure with real impact on population mean annual effective dose are described by several studies (2,3). Reports of wide variations in patient dose for the same radiographic examinations within and among hospitals in the UK and Europe are also described previously (4,5). The following question is addressed in a Special Report (6) : Is it really justified for one facility to use an exposure that is 10, 20 or 126 times greater than that used by another facility to produce a radiographic image?. In addition, a recent study estimates that the diagnostic use of X ray causes an increase of cumulative risk of cancer at an age of 75 y: UK (0.6%), The Netherlands (0.7%), Switzerland (1%) and Germany (1.5%) (7). For diagnostic purposes, the optimisation of exposure involves the relationship of three core aspects of the imaging process (8) : (i) choice of radiographic technique; (ii) radiation dose to the patient and (iii) diagnostic quality of the radiographic image. These three aspects are critical for the diagnostic quality of the radiographic image. *Corresponding author: luis.lanca@estesl.ipl.pt Exposure parameters influence and determine the quantity and quality of the X-ray beam. The choice of the most adequate radiographic technique to each clinical situation involves the selection of the correct exposure parameters. It is the radiographer s responsibility to adequately select exposure technical parameters except those related to particular equipment characteristics (9). The four main exposure parameters are tube potential (in kv), tube intensity (in ma), exposure time (in s) and focus-to-detector distance (in cm). Exposure time and tube intensity could be a unique exposure factor (in mas). In 1996, the Commission of European Communities (CEC) published recommendations concerning Quality Criteria for Diagnostic Radiographic Images (10).These recommendations or Guidelines define diagnostic requirements for a normal, basic radiograph, specifying anatomical image criteria and important image details; indicate criteria for the radiation dose delivered to the patient and give an example for good radiographic technique by which the diagnostic requirements and the dose criteria can be achieved (10). Several European countries have already incorporated in their national legislation these recommendations and the European directives on radiation protection. In Portugal, the European Directive 97/43/Euratom is already incorporated in Portuguese legislation as well as the European Guidelines recommendations (11). Although regulation on radiation protection is already incorporated in national legislative documents, in Portugal there is no real implementation of these recommendations in daily clinical practice # The Author 2008. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

or real established diagnostic reference levels (DRLs). As an example, typical distributions of exposure parameters in plain radiography are unknown in Portugal. This exploratory study could constitute a contribution to the identification of these distributions. We are investigating the typical exposure parameters in plain radiography for skull, chest, lumbar spine and pelvis. This study aims to identify exposure parameters used in plain radiography in the Lisbon area and to compare collected data with European references (10). MATERIALS AND METHODS A questionnaire was developed and delivered to 85 radiographers working in the Lisbon area, and 35 (41.1%) valid responses were returned. They were asked to identify exposure parameters that they are using in their clinical practice for a reference adult patient of 70 kg (10). Four anatomic regions were considered in seven radiological projections: skull (PA; ), chest (PA; ), lumbar spine (AP; ) and pelvis (AP). Collected data indicates kv, ma, exposure time (ms), mas and focus-to-detector distance (cm) that are being used in daily practice for each radiological projection. Data were analysed using the SPSS w version 14.0 statistical software. L. LANÇA ET AL. Table 1. Tube potential (in kv) one-sample t-test at 95% CI. Projection Mean (kv) Standard deviation Test (kv) t- statistic P- Skull PA 63.03 6.386 70 6.732,0.001 85 21.211,0.001 Skull 60.11 7.129 70 8.440,0.001 85 21.239,0.001 Chest PA 106.08 18.850 125 6.188,0.001 Chest 118.13 14.114 3.000 0.005 Lumbar 66.58 6.942 75 7.274,0.001 spine AP 90 20.239,0.001 Lumbar 82.21 7.253 75 6.128,0.001 spine 90 6.621,0.001 Pelvis AP 66.82 6.900 75 7.311,0.001 90 20.712,0.001 Table 2. Exposure time (in ms) one-sample t-test at 95% CI. Projection Mean (ms) Standard deviation Test (ms) t- statistic P- RESULTS The results shown in this study allow us to identify the main exposure parameters in seven radiological projections in four anatomic regions. One sample t-test at a 95% confidence interval (CI) of the difference shows a significant deviation (P, 0.001) from the test (Table 1), except for chest. kv s (mean) for skull PA (63.03), skull (60.11), chest PA (106.08), chest (118.13), lumbar spine AP (66.58) and pelvis AP (66.82) are below the CEC kv reference s (Test column). Lumbar spine projection mean kv (82.21) is within the recommended interval, although the results show a significant difference (P, 0.001) from the lowest (70 kv) and highest (85 kv) reference s. Table 2 shows that there is no significant deviation from test for most projections. A significant deviation (P, 0.001) is shown for lumbar spine. Mean exposure time s for skull PA (157.01), skull (195.32) and chest PA (21.84) are above the CEC exposure time reference s (Test column). Chest (33.45), lumbar spine AP (320.14), lumbar spine (466.11) and pelvis AP (287.68) are below the CEC reference level. Exposure parameters for all projections (tube potential and exposure time) are shown in Figure 1(a g). Skull PA 157.01 116.112 100 1.837 0.089 Skull 195.32 133.106 3.581 0.002 Chest PA 21.84 23.196 20 0.286 0.780 Chest 33.45 27.751 40 0.851 0.412 Lumbar 320.14 312.528 400 1.225 0.233 spine AP Lumbar 466.11 428.648 1000 5.842,0.001 spine Pelvis AP 287.68 268.652 400 1.961 0.063 Skull observations (Figure 1a and b) are below the CEC recommendation interval for tube potential (76% PA; 89% ) and above the recommendation for exposure time of 100 ms (85% PA; 80% ). Exposure parameters for chest radiography are shown in Figure 1c and d. Tube potential results are below the reference of 125 kv (90% PA; 62% ). Only a few observations (23%) are set on 125 kv or above in projection. Exposure time is below the 20 ms (70% PA) and 40 ms (62% ) recommendation. Some results are observed far above the recommended exposure time for both projections. Page 2 of 5

COMPARING EXPOSURE PARAMETERS WITH CEC GUIDELINES Figure 1. (a g) Scatter plot showing exposure parameters (tube potential and exposure time). Page 3 of 5

Lumbar spine AP projection (Figure 1e) shows 83% of observations below the reference interval (75 90 kv). In this projection, 30% of exposure time observations are above the reference of 400 ms. Only 26% of cases in lumbar spine projection (Figure 1f) are below the reference interval for tube potential and 22% above the limit of 1000 ms. Figure 1g shows that in 73% of cases tube potential is below the recommended range of 75 90 kv for pelvis AP projection. Exposure time is.400 ms reference in 27% of cases. DISCUSSION Scatter plots (Figure 1a g) show a dispersion of exposure parameters without a consistent pattern and a wide variation in exposure parameters for the same radiological projection. A wide dispersion is observed with a predominance of results below the recommended kv interval and above the recommended exposure time. Compared with CEC guidelines observations show an inadequate selection of exposure parameters. Other studies also demonstrate a wide variation in exposure parameters (12). This could configure a practice of overexposure and collected data suggest that entrance skin dose (ESD) could be above the reference in several projections since long exposure times are observed in most projections. The wide variability found in exposure parameters could be explained by differences in equipment characteristics (e.g. screen-film, computed radiography or digital radiography) that is used in the clinical practice by the respondents. Other possible explanations could be the fact that there is a lack of implemented local/national recommendations and lack of criteria for good radiographic techniques in Portugal. Further actions should be taken to optimise exposure in plain radiography. This study shows tube potential s (kv) systematically below the CEC reference. Literature shows that higher potential s allows considerable dose reduction without loosing image quality. A reduction of up to 29.9% in patient effective dose was demonstrated in Doherty (13). It must be mentioned as a possible restriction that collected data are based on personal experience of radiographers. Data were not collected or measured at the time the radiological procedures being carried out. This is important because ESD was not measured or calculated. In some projections, the results show exposure times above the CEC reference level and this could configure a practice of overexposure to the patients. Implementation of local references could achieve an ESD reduction between 30 and 60% below the L. LANÇA ET AL. Page 4 of 5 CEC recommendation (14). Several studies show that it is possible to achieve a dose reduction of 50% without loosing image quality when CEC guidelines are well established (13,15,16). CONCLUSIONS This study shows that at a local level (Lisbon region) radiographic practice does not comply with CEC guidelines concerning exposure techniques. Further national/local studies are recommended with the objective to improve exposure optimisation and technical procedures in plain radiography. This study also suggests the need to establish national/local DRLs and to proceed to effective measurements for exposure optimisation. REFERENCES 1. Seeram, E and Brennan, P. Diagnostic reference levels in radiology. Radiol. Technol. 77, 373 384 (2006). 2. United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR). Sources and effects of ionizing radiation. Annex D medical radiation exposures (New York: United Nations) (2000). 3. Regulla, D. F. and Eder, H. Patient exposures in medical X-ray imaging in Europe. Radiat. Prot. Dosim. 14, 11 25C (2005). 4. Johnston, D. A. and Brennan, P. C. Reference dose levels for patients undergoing common diagnostic X-ray examinations in Irish hospitals. Br. J. Radiol. 73, 396 402 (2000). 5. Carroll, E. M. and Brennan, P. C. Radiation doses for barium enema and barium meal examinations in Ireland: potential diagnostic reference levels. Br. J. Radiol. 76, 393 397 (2003). 6. Gray, J. E. et al. Reference s for diagnostic radiology: application and impact. Radiology 235, 354 358 (2005). 7. Berrington de González, A and Darby, S. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet 363, 345 351 (2004). 8. European Commission (EC). Optimisation of protection in the medical uses of radiation. EUR 19793. (Luxembourg: Office for Official Publications of the European Communities) (2002). 9. Bushong, S. C. Radiologic Science for Technologists (St. Louis: Mosby) (2001) ISBN 0-323-01337-6. 10. European Commission (EC). European guidelines on quality criteria for diagnostic radiographic images. EUR 16260. (Luxembourg: Office for Official Publications of the European Communities) (1996). 11. Portugal. Health Ministry Law no. 180/2002. Establish the rules concerning ionizing radiation protection in medical exposures (8 August 2002). 12. Papadimitriou, D, Perris, A, Molfetas, M. G., Panagiotakis, N, Manetou, A, Tsourouflis, G, Vassileva, J, Chronopoulos, P, Karapanagiotou, O and Kottou, S. Patient dose, image quality and radiographic techniques for common x ray examinations in two Greek

COMPARING EXPOSURE PARAMETERS WITH CEC GUIDELINES hospitals and comparison with European guidelines. Radiat. Prot. Dosim. 95, 43 48 (2001). 13. Doherty, P, O Leary, D and Brennan, P. C. Do CEC guidelines under-utilise the full potential of increasing kvp as a dose-reducing tool? Eur. Radiol. 13, 1992 1999 (2003). 14. Vano, E, Fernandez, J. M., Ignacio Tem, J. I., Guibelalde, E, Gonzalez, L and Pedrosa, C. S. A. Real-time measurement and audit of radiation dose to patients undergoing computed radiography. Radiology 225, 283 288 (2002). 15. Mooney, R. and Thomas, P. Dose reduction in a paediatric X-ray department following optimization of radiographic technique. Br. J. Radiol. 71, 852 860 (1998). 16. Saure, D., Hagemann, G, and Stender, H.S. Image quality and patient dose in diagnostic radiology. Radiat. Prot. Dosim. 57, 167 170 (1995). Page 5 of 5