TECHNICAL BULLETIN GENERAL INFORMATION

Similar documents
Energy-Efficient Biological Safety Cabinet Reduces Energy Costs While Ensuring Safety

THE BAKER COMPANY 161 Gatehouse Rd., Sanford, Maine USA Creating Immaculate Atmospheres

Increased Microbiological Challenge Test with Bunsen Burner Inside the Cabinet

Evaluation of Laminar Flow Microbiological Safety Cabinets

The Whitley Internal HEPA filtration system bacteriological testing

Validation of Respirator Filter Efficacy

Biohazard Hood to Prevent Infection During Microbiological Procedures

EVALUATION OF THE CYCLEX IMPACTOR PHASE 2: REPORT 3

Laboratory Ventilation Policy

Biosafety in the Animal Setting. Tina Bogac Institutional Biosafety Officer

Biosafety Level 2 Standard Practices and Containment Office of Research Safety Affairs

THE UNIVERSITY OF ALABAMA FUME HOOD MANAGEMENT PLAN

Setting up Your Home Office

SurgiCube Clinical Tests

Laboratory Orientation and Safety

Assessing the Risk of Laboratory Acquired Allergies

Healthy Buildings 2017 Europe July 2-5, 2017, Lublin, Poland

CLEAN ROOMS CIEA TESTING DATA

Unidirectional Airflow Smoke Test Protocol

Uni-directional Airflow Systems Working Height Velocity (WHV) Gordon Farquharson July 2017

Guidance for the Control of Animal Allergens

Environmental Monitoring in the Cell Processing Laboratory What to do with the data?

Biosafety Level 3 (BL3)

AIR PURIFICATION AND AROMATHERAPY SYSTEM

FINAL REPORT. EMSL Analytical, Inc. Airfree Products RUA Mouzinho Da Silveira Lisboa, Portugal PT. Efficacy of an Air Sanitizer

STANDARD OPERATING PROCEDURES (SOP) FOR CARCINOGENS

The following guidelines are applicable to office workers who sit at a desk and/or work on computers.

Interpretation Guide. Enterobacteriaceae Count Plate

Count. Enterobacteriaceae. Interpretation Guide. 3M Food Safety 3M Petrifilm Enterobacteriaceae Count Plate

Laboratory Animal Allergen (LAA) Exposure Control Program

Sterile U Network. Quality Control for Table-top Steam Sterilizers. Introduction: T U T O R I A L S

All of our Radioiodine Cartridges

The integrity of pharmaceutical and nutritional packaging

CONCENTRATIONS OF VIABLE FUNGAL SPORES ON PAPER DOCUMENTS

RiGO801 Laboratory Requirements

Infection Control Standard Precautions and Isolation

Experimental study of the effect of operation lamps on downward airflow distribution in an operating theatre in St. Olavs Hospital in Norway

How Do You Know If Your Sterilized Instruments Remain Sterile?

Adjust the chair height so that your feet rest comfortably on the floor, footrest or foot ring.

Chief Executive Office Risk Management Division P.O. Box 1723, Modesto, CA Phone (209) Fax (209)

How Do You Know If a Sterilized Package Remains Sterile?

Biosafety Basics. Patrick Stockton, Director, Office of Biosafety Biosafety Community Liaison Committee June 26, 2017

Managing Aerial Spray Drift. by Paul E. Sumner Extension Engineer

MEASUREMENT OF THE DEGREE OF PROTECTION AFFORDED BY RESPIRATORY PROTECTIVE EQUIPMENT AGAINST MICROBIOLOGICAL AEROSOLS

Indoor Air Quality: Mold, Patients

Procine sphingomyelin ELISA Kit

EUNETIPS Survey on guidelines or recommendations or official rules about the air quality in operating theatres in Europe

PDA Points to Consider relating to EU GMP Annex 1 changes. Presented by Ashley Isbel 30 November, 2016

Reducing Waste Material and Increasing Productivity by Butt Splicing - R. Joseph Gotshall,

Name Date: Pharmacy Education Resources National Intravenous Training Manual for Pharmacists TSHP Post-Test

4 Pull with different grips

Texas Association of Healthcare Facilities Management December 12, 2014

PATIENT INSTRUCTIONS Jaw Dynasplint System

Petrifilm Enterobacteriaceae Count Plate

Respiratory Protection and Swine Influenza

(*Judges notes: We will have 4 working platforms, and 2 warm-up platforms. Warm-Up platforms will be outside, weather permitting)

Protection for the Human Eye The Lacrimal System

Petrifilm. Staph Express Count Plate. Interpretation Guide

Scoring: Score is the total reps completed in both portions of the workout.

By: Stephanie Bendtsen, Joseph Calderan, and Celeste Dupont Team 17 Client: Dr. Sun

Orthotics Measurement Board for Tibial Torsion and Toe-Out

FAIRFIELD SCHOOL HEALTH PROGRAM. PROCEDURE: Administration of Oral, Topical or Inhalant Medications by Principals and Teachers

Quads (machines) Cable Lunge

PH-04A: Clinical Photography Production Checklist With A Small Camera

Aseptic Technique and Sterile Compounding

Rhino Clear Sprint Atomizer

Supplementary Figure S1 Black silicon and dragonfly wing nanotopography.

Office Ergonomics and Workstation Analysis

New Test Battery / Sequencing of Physical Fitness Assessments

Osteoporosis Exercise:

Misty Max 10 nebulizer

UMD LAA ECP 20 June University of Maryland Laboratory Animal Allergen Exposure Control Plan 20 June 2014

Ergonomics Checklist - Computer and General Workstations 1

M. A. CECIL & ASSOCIATES, INC Shannon Way, Port Republic, Maryland (301) INDUSTRIAL HYGIENE AND ENVIRONMENTAL HEALTH

Novel Techniques for Excellent Air Cleanliness in an Operating Theatre and a Cardiac Catheterization Laboratory

Mouse Hepatic Progenitor Organoid Culture: Supplementary Protocols

PRESENTER: DENNIS NYACHAE MOSE KENYATTA UNIVERSITY

CUB GAMES WEEK 2 BROUGHT TO YOU BY 2115NINE 17:00 PM AEST, THURSDAY, MARCH 1st THROUGH TO TUESDAY, MARCH 6th 10 AM AEST

Mycotech Biological, Inc.

8 Performing Medical Procedures

Influence of Environmental Parameters on Mold Sampling Results

Product Specifications and Usage Information P4405, P4410, & P4420 Genesys 30 Dosimeter Holder System

"Understanding USP 71 Sterility Tests and Extending BUD"

Understanding cascade impaction and its importance for inhaler testing

Living with Newton's Laws

Eyewash and Safety Showers HM 20 Page 1

Report on the Performance of the. VidaShield System

Cytotoxic Drugs. Table of Contents

Stairways and Ladders. OSHA Office of Training & Education 1

8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy

3. Dimensions, weights Placement and Assembly Pressure drops Noise data Material... 14

Index. Immunol Allergy Clin N Am 23 (2003) Note: Page numbers of article titles are in boldface type.

Distributed exclusively by.

Training For The Triple Jump. The Aston Moore Way

NORCAL MASTERS 2017 WORKOUTS WITH EXPLANATIONS AND STANDARDS

Selection and Optimization of the Sample Introduction System

Name Date: Pharmacy Education Resources National Intravenous Sterile Training Manual for Pharmacy Technicians TSHP Post-Test

Laboratory animal allergy (LAA) is a serious occupational hazard

FINAL REPORT- Part 1 & Part 2

Transcription:

TECHNICAL BULLETIN GENERAL INFORMATION NU-612 SMALL ANIMAL CAGE CHANGING STATION PERFORMANCE EVALUATION 2100 Fernbrook Lane Plymouth, MN 55447-4722 U.S.A. Phone: 763.553.1270 Fax: 763.553.0459 Toll Free: 1.800.328.3352 Background The NuAire Small Animal Cage Changing Station Model NU-612 is a station whose unique configuration offers the benefits of vertical HEPA filtered airflow for product protection on the worksurface. In addition, the NU-612 provides improved allergen personnel protection by incorporating a HEPA filtered down draft vacuum on the worktray to assure minimal exposure to animal hair/dander as well as cage bedding material. However, the NuAire Model NU-612 should not be used as a substitute for a Class II, Type A2 Biological Safety Cabinet. The NU-612 is not a containment device for airborne particulate in low to moderate risk-hazard research as prescribed by the CDC/NIH Biosafety in Microbiological and Biomedical Laboratories. As with all airflow devises, performance of the NU-612 is dependent upon the environment in which it is used. Activity in the room itself should be held to a minimum. Unnecessary activity may create disruptive air currents. A person walking past the station can cause draft velocities up to 250 fpm (1.27 m/s), which can interfere with the stations proper function. The NU-612's design and airflow pattern is illustrated on Drawing BCD-08610. The NU-612's design incorporates two impeller/blowers, both a supply and exhaust, each with its own associated HEPA filter providing excellent system function and maximum filter loading capacity. The designed airflow pattern, being a single pass system, also eliminates any heat build up within the workzone. The airflow pattern starts with the supply prefilter located on top of the station. The airflow continues through the supply impeller/blower and forced into the supply HEPA filter providing 99.99% particle removal at 0.3 micron. The airflow continues to move in a uniform vertical downflow pattern providing particulate-free air in the workzone. As the airflow continues to move toward the worktray, some particulate free downflow is pushed out the top two-thirds of the work access opening. The worktray configuration provides a unique downdraft airflow pattern that offers improved allergen personnel protection. Downdraft vacuum slots are provided around the worktray perimeter for improved particulate containment of animal hair/dander and bedding materials. Additional downdraft vacuum slots are provided dividing the worktray into quadrants for improved cross-contamination protection from cage to cage during the transfer process. The downdraft HEPA filter provides 99.99% particle removal at 0.3 micron and is located under the worktray in its own impeller/blower/filter cabinet. As the air is pulled from the worktray, through the prefilter and into the blower, the air then passes through the downdraft HEPA filter and is exhausted back to the room providing improved allergen personnel protection. PG 1-10

The exhaust volume is sufficient for proper station function, yet generates low velocities of less than 100 fpm (.51 m/s) at the base of the station minimizing air currents in the room. In comparison, room supply air ducts and personnel traffic can cause draft velocities exceeding 250 fpm (1.27 m/s). The NU-612 performance evaluation was conducted by NuAire to quantify the NU-612's ability to provide product and cross contamination protection as well as improved personnel protection. Product protection was evaluated using two methods: biological testing and particle measurement. Personnel and cross contamination protection was evaluated using biological testing only. 2

3

PRODUCT PROTECTION BIOLOGICAL TESTING Procedure The product protection biological testing was performed using a modified NSF Standard 49 testing method for Biological Safety Cabinets (BSC's). A biological challenge of Bacillus subtilis var. niger (B. subtilis) spore suspension at a concentration of no less than 5.0 x 10 6 spores/ml was nebulized outside the station. Two different test setups were used to evaluate product protection in different areas of the station (see Figures 1 & 2). The nebulizer was positioned at 4 inches (102mm) outside the window opening and at a height of 12 inches (305mm) above the worktray. A 2-1/2 inch (63mm) outside diameter cylinder is placed within the center of the test set-up. Seven agar plates were placed on the worksurface in a pyramid shape in the center of the test set-up. A control plate was placed directly under the cylinder and worktray grill. The station was set to nominal airflows for the test. Figure 1 Product Protection Worktray Test Setup #1 Centered in Left Quadrant Figure 2 Product Protection Worktray Test Setup #2 Centered in Workstation 4

The test procedure was to the setup. Turn on the station for 10 minutes. Remove the lids on the agar plates and operate the nebulizer for five minutes. Five minutes after nebulization is terminated, place the lids back on the agar plates and incubate for a minimum of 24 hours at 37 C. Each test setup was run in triplicate. Results A) Test Setup #1 Run #1 Control - 50 B) Test Setup #2 Run #1 Control - 43 Run #2 Control - 23 Run #2 Control - 66 Run #3 Control - 46 Run #3 Control - 47 Conclusion The test results indicate that no contamination is migrating onto the worktray. No Colony Forming Units (CFU's) were recovered on the agar plates. One point to note, normally when this test is conducted, a minimum of 300 CFU's present on the control plates. This would validate that the challenge was sufficient for the test. However, since the nebulizer is positioned at 12 inches above the worktray, most of the airflow at that level is pushing out of the station. Even with this airflow pattern, the challenge was sufficient to pick up countable CFU's. In addition, the nebulizer was weighed before and after the test to validate the correct quantity of spore suspension was nebulized. 5

PRODUCT PROTECTION PARTICLE MEASUREMENT Procedure The product protection particle measurement was performed using a laser particle counter manufactured by Met One, Model Number A2408. The particle counter sample rate was 1 CFM for a particle size of 0.3 micron. Five sampling locations at 2 different height levels were taken within the workzone (see Figure 3). Sampling locations 1 through 4 were positioned 6 inches from the worktray side edge and 4 inches from the worktray front edge. Sampling location 5 was centered in the worktray. The sampling locations were taken in a horizontal plane at window height 14 inches (356mm) and a typical level for a mouse cage at 6-1/2 inches (165mm). Each sampling location was zeroed and particles counted for 1 minute. The total number of particles were then recorded. Figure 3 Particle Measurement Sampling Locations Results A) Test level 14 inches (356mm) above worktray sampling location - particles per cubic foot. B) Test level 6-1/2 inches (165mm) above worktray sampling location - particles per cubic foot. 1-14 3-6 4-12 Conclusion The test results indicate that the workzone cleanliness level is maintained at better than a Class 100 environment throughout the workzone. 6

PERSONNEL PROTECTION BIOLOGICAL TESTING Procedure The personnel protection biological testing was performed using a modified NSF Standard 49 testing method for BSC's. A biological challenge of B. subtilis spore suspension at a concentration of no less than 5.0 x 10 8 spores/ml was nebulized inside the station. Three different test setups were used to evaluate the personnel protection level. The first two test setups used a typical mouse cage centered within the front left worktray quadrant (see Figure 4). The only difference was the first test had the station's hinged windows in their normal down position. The second test had the station's hinged windows in a raised position. The nebulizer was placed inside the mouse cage along with control plate. Several more agar plates were placed around the outside of the mouse cage and in the surrounding quadrants of the worktray. A 2-1/2 inch (63mm) outside diameter cylinder was placed in the center of the worktray extending out of the station. Around the cylinder on the outside of the station, 6 impingers are placed around the cylinder, 2-1/2 inches (63mm) from the station front. Outside the station, two slit samplers are placed 8 inches (203mm) from each end and 6 inches (152mm) from the station front. The test duration is 30 minutes, performed in triplicate, using the following sequence: Time Remaining (min.) Activity 30 start slit samplers 25 start nebulizer 24 start impingers 19 stop impingers 18.5 stop nebulizer 0 stop slit samplers The third test set-up was identical to the first two, except the mouse cage was removed (see Figure 5) Figure 4 Personnel Protection Worktray Test Setup #1 & #2 for both Hinged Windows Closed and Open Test Runs 7

Figure 5 Personnel Protection Worktray Test Setup #3 Results A) Test setup #1 w/hinged windows in normal closed position. Run #1 Run #2 Slits Left-39 Right - 0 Slits Left - 50 Right - 0 Impingers - 20 Impingers - 21 1 > 100 8-0 1 > 100 8-0 2 > 100 9-0 2 > 100 9-0 3 > 100 10-1 3 > 100 10-0 4-4 1 4-4 1 1 1 1 1 1 1 Run #3 Slits Left - 50 Right - 0 Impingers - 33 1 > 100 8-0 2 > 100 9-0 3 > 50 10-0 4-13 1 1 1 1 8

B) Test setup #2 w/hinged window in open position Run #1 Run #2 Slits Left-12 Right - 0 Slits Left -11 Right - 0 Impingers - 1 Impingers - 0 1 - TNTC 8-0 1 - TNTC 8-0 2 - TNTC 9-0 2 - TNTC 9-0 3-2 10-0 3 - TNTC 10-0 4-11 1 4-7 1 1 1 1 1 1 1 Run #3 Slits Left - 2 Right - 0 Impingers - 1 1 - TNTC 8-0 2 - TNTC 9-0 3 - TNTC 10-1 4-8 1 1 1 1 9

C) Test setup #3 w/hinged window in normal closed position. Run #1 Run #2 Slits Left-21 Right - 3 Slits Left - 39 Right - 4 Impingers - 22 Impingers - 27 1 - TNTC 8-0 1 - TNTC 8-0 2-34 9-0 2 - TNTC 9-0 10-0 10-0 1 1 1 1 1 1 1 1 Run #3 Slits Left - 38 Right - 2 Impingers - 43 1 - TNTC 8-0 2 - TNTC 9-0 10-0 1 1 1 1 Conclusion The test results indicate that the downdraft vacuum on the worktray provides both excellent improved allergen protection and cross contamination from quadrant to quadrant. In review of the above data regarding personnel protection, the typical results from the above test runs recorded 20 to 50 CFU's on the Impingers and 25 to 50 CFU's on the slit samplers. In a comparison to the NSF Standard 49 requirements, NSF allows no more than 10 total CFU's on the impingers and no more than 5 total CFU's for the slit samplers. Achieving countable colonies close to the NSF allowable limits with such a large challenge 5.0 x 10 8 B. subtilis spores/ml, indicates that under normal usage, particulates generated in the workzone both in and around the animal cage will most likely be contained. Based upon the above test methods and results, it is our opinion that the NU-612 will provide greater than a 95% allergen personnel protection level. However, again please note, the NU-612 should not be used as a substitute for a Class II, Type A2 BSC. The NU-612 is not a containment device for airborne particulate in low to moderate risk-hazard research. In review of the above data regarding cross-contamination from quadrant to quadrant, only 1 CFU was recorded outside the challenge quadrant on test setup #1, agar plate #10. Based upon the above test methods and results, the NU-612 provides excellent cross-contamination protection from quadrant to quadrant. 10