ABSTRACT. Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray

Similar documents
Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Assessment of effective dose in paediatric CT examinations

Skyscan 1076 in vivo scanning: X-ray dosimetry

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

CT Radiation Risks and Dose Reduction

Dosimetric Consideration in Diagnostic Radiology

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

CT Dosimetry in the Clinical Environment: Methods and Analysis

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Accounting for Imaging Dose

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Bureau of Laboratory Quality Standards

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

RADIATION SAFETY. Junior Radiology Course

Why is CT Dose of Interest?

AUA Guidelines for Imaging Known or Suspected Ureteral Calculi. Michael Ferrandino, MD Assoc Professor of Urology Duke University Medical Center

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Radiologic Units: What You Need to Know

Managing Patient Dose in Computed Tomography (CT)

Survey of patients CT radiation dose in Jiangsu Province

Radiation Protection Services in R.Macedonia

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Prepublication Requirements

Managing the imaging dose during image-guided radiation therapy

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

CT Quality Control Manual FAQs

Managing Radiation Risk in Pediatric CT Imaging

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Course Syllabus + Study Guide for Lecture and Laboratory

Variation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito

1 Course Syllabus + Study Guide for Lecture and Laboratory

Prediction of Organ and Effective dose with known mas and kvp for dose Optimisation Protocol and Recommendations in CT

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

Atoms for Health. Atoms for Health The. Atoms for Health - Division of Nuclear Health - Dept of Nuclear Aplications P Andreo DIR-NAHU 1

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Patti Edwards, Senior Radiographer, West Herts Hospitals, UK. February Radiation Safety

Radiation Protection- Cath lab

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Ask EuroSafe Imaging Tips & Tricks. CT Working Group

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

45 Hr PET Registry Review Course

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Research Article. Corresponding author H. Osman

How do the Parameters affect Image Quality and Dose for Abdominal CT? Image Review

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Computed tomography Acceptance testing and dose measurements

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Radiation Safety for New Medical Physics Graduate Students

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters


ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Adopting DAP as a dose metric in CT

Dosimetry in digital mammography

Radiologic Imaging Magnetic Resonance Imaging (MRI)

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Survey of Radiation Dose Levels in Patients in X-Ray Units of Some Selected Hospitals in Jos Metropolis

Application of international standards to diagnostic radiology dosimetry

Radiation Dose Reduction Strategies in Coronary CT Angiography

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Quanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Diode calibration for dose determination in total body irradiation

Radiation related cancer risk & benefit/risk assessment for screening procedures

To Shield or Not to Shield? Lincoln L. Berland, M.D.

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Patient Dosimetry in Mammography and Tomosynthesis:

Radiation Dose Assessment and Risk Estimation During Extracorporeal Shock Wave Lithotripsy

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Recent Progress in Radiation Dosimetry for Epidemiology and Radiological Protection. John Harrison ICRP Committee 2

Case Report Three-Dimensional Dual-Energy Computed Tomography for Enhancing Stone/Stent Contrasting and Stone Visualization in Urolithiasis

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Transcription:

Radiation Dose Distribution for Patients Undergoing Routine Radiological Scans for Kidney Stone Diagnosis at the University Hospital of the West Indies DC Walker 1, WD Aiken 2, S Shah 3, MK Voutchkov 1, L-GM Burnett 4, A Gordon 1, CA McKenzie 5 ABSTRACT Objectives: To investigate the radiation dose received by patients undergoing routine plain x-ray and computed tomography scans for kidney stone studies. Methods: For x-ray scans two sets of experiments were performed. Thermo-Luminescence Dosimeters were used to determine the distribution of radiation dose across the abdominal surface and an acrylic water phantom equipped with a Farmer ionization chamber was used to measure the radiation dose as x-ray photons travel through the abdomen. For CT scans, the Dose Length Product (DLP) over 3 months for abdominal studies was recorded. Calculations were done to convert the DLP to effective dose using k factor values from the International Commission on Radiological Protection (ICRP) report 103. The results were then mapped and analysed using Microsoft excel and Wolfram Mathematical software. Results: Skin dose was greatest at the centre of the abdomen. Radiation equivalent dose to the skin for a plain abdominal x-ray ranged from 2.38mSv for patients with a sthenic body habitus to 5.45mSv for hypersthenic patients. For CT scans, effective dose was found to range between 6.72mSv and 24.27mSv. Conclusion: All radiation dose measured were found to be within recommended ICRP operational limits. The developed procedure could be implemented in other radiology departments in Jamaica as a part of their quality assurance program. Keywords: Computed tomography, kidney stone, radiation dose, x-ray scan From: 1 Department of Physics, Faculty of Science and Technology, 2 Section of Urology, Department of Surgery, UWI, Faculty of Medical Sciences, 3 Section of Radiology, Department of Surgery, 4 X-Ray Department, University Hospital of the West Indies Mona, and 5 School of Medical Radiation Technology, Faculty of Medical Science, The University of the West Indies Mona, Kingston, Jamaica. Correspondence: Mr D Walker, Department of Physics, Faculty of Science and Technology, The University of the West Indies, Mona. Kingston, Jamaica. Email: darrion.walker02@uwimona.edu.jm West Indian Med J DOI: 10.7727/wimj.2016.589

Radiation Dose and Kidney Stone INTRODUCTION Kidney Stones are developed from the urine when chemicals (usually calcium oxalate and phosphate) that are normally excreted by the kidney become concentrated and are unable to dissolve in solution (1, 2). They have become increasingly prevalent over the past two decades with the prevalence rate worldwide being approximately 1.7% to 8.8%. 5% of kidney stones are non-calcified with the other 95% being calcified to varying degrees (3). Computed Tomography (CT) is the best imaging study to confirm the diagnosis of a urinary stone, and is now the first line imaging study for suspected kidney stone in the United States, however the disadvantages to Computed Tomography involves it being costly and that it uses ionizing radiation (4). In Jamaica, the frontline modalities in patient diagnosis are plain x-ray, computed tomography and ultrasonography. In 2012, there were 28 radiology departments in the island. Plain x-ray was the most widely available modality followed by Computed Tomography and then ultrasonography (5). Non-calcified stones however are usually more difficult to diagnose on radiographs (6). Except for ultrasonography, radiological scans most commonly used for the diagnosis of kidney stones (these include plain x-ray, CT and fluoroscopy) employs the use of ionizing radiation. This therefore means that patients undergoing these examinations are at a risk for possible radiation health effects which could be either of a Stochastic or Non-Stochastic nature (7). To protect patients undergoing radiological studies, the International Commission on Radiological Protection (ICRP) has established recommended dose limits for different radiological studies including for those which are used in kidney stone diagnosis such as x-ray of the abdomen, CT scan of the abdomen and intravenous pyelogram. 2

Walker et al This study was therefore aimed at investigating the radiation dose received by patients that undergo these examinations at the University Hospital of the West Indies and compared it with the recommended dose levels of the ICRP as a part of a quality control exercise (8, 9). METHOD This study included the use of Thermo-Luminescence Dosimeters (TLD), a PTW Farmer ionization chamber and a 1D water phantom tank. The field of view (FOV) for plain x-ray scan for abdominal studies was determined as shown in fig 1 and the radiation dose within this FOV was measured. The study included the mapping of radiation dose distribution across the surface of the abdomen using TLD films and the measurement of radiation dose fall off with depth as x-ray traverse the abdomen using a water phantom. The mapping of radiation dose distribution across the surface of the abdomen for abdominal studies involved the use of sixteen TLD films placed across a horizontal surface within the dimensions of 35.56x43.15cm. This configuration resembles the widest radiation field used by Radiographers to carry out an abdominal x-ray scan. The human abdomen is curved with the greatest distance between the abdomen and back located in the mid sagittal plane. The study however was carried out on a flat surface as a quality control test to investigate the homogeneity of the radiation field as it is distributed across the anatomical surface. The TLDs were paired and separated by a 2cm distance at each selected point for measurement within the exposed field (figure 1). 3

Radiation Dose and Kidney Stone Fig. 1: TLD placement in FOV. The source - image receptor distance (SID) was set at 101.6cm as is used for supine abdominal studies. The source - object distance (SOD) was established at 63.5cm, 50.8cm and 25.4cm for simulating the SOD of the different body habitus of hypo-sthenic, sthenic and hyper-sthenic respectively. Three sets of exposures were delivered to three different sets of TLD films. The three different set of exposures were set to simulate low, medium and high exposure factors in an effort to analyse the dose distribution across the skin for varying exposure settings, as well as to analyse the tube performance. The TLDs were sent to the International Centre for Environmental and Nuclear Sciences (ICENS) at the University of the West Indies where they were analysed using an automated HARSHAW Model 6600+ reader. This model comes with a built in 90 Sr irradiator for calibration purposes. Fig 2. Water Phantom The measurement of radiation dose fall off with depth as x-ray traverse the abdomen involved the use of a 1D water phantom tank, a 0.6cm 3 PTW Farmer ionization chamber and a PTW 4

Walker et al Unidos E universal dosimeter. The phantom was filled with water to the dimensions of 42cm x 40cm x 39cm (length, width and height respectively) and placed on the radiographic table and left to rest for 10 minutes. The x-ray tube was then centred to the ion chamber and an SID of 101.6cm was used in relation to the table bucky. Two sets of eight exposures were made. Each set of exposure maintained the exposure factors used (ma, kv, mas) and varied the depth of submersion of the ionization chamber. The two set of exposures used simulated the exposure factors normally given to patients of the Sthenic and Hyper-Sthenic body habitus (70kV@20mAs (Parameter 1) and 74kV@40mAs (Parameter 2) respectively). The depth of submersion of the ionization chamber was decreased by 2cm for each exposure and ranged from the surface (0) to 14cm deep in the water. The data was recorded and analysed. In assessing the dose received by patients that undergo abdominal CT scans for the diagnosis of kidney stones, the Dose Length Product (DLP) over 3 months for such studies was recorded. Calculations were done to convert the DLP to effective dose using k factor values from the International Commission on Radiological Protection (ICRP) report 103 (10). RESULTS Figure 3 shows the x-ray surface dose distribution maps for low, medium and high radiographic exposures. The maps were developed using Wolfram Mathematica. A B C Fig. 3: X-ray surface dose distribution for low (A), medium (B) and High (C) exposure. 5

Radiation Dose and Kidney Stone For surface dose distribution, the maximum dose was recorded at the centre of the radiation field for all exposures. The dose fell off significantly towards the left of the radiation field in comparison to the right as well as toward the anode side (top) more significantly than towards the cathode side (bottom) for images A and B but relatively even for image C. For dose distribution with increased depth, the dose decreased linearly, starting with surface dose readings of 5.45mGy for parameter 2 (74kV@40mAs) and 2.38mGy for parameter 1 (70kV@20mAs). The Dose Length Product (DLP) recorded by the CT unit for abdominal scans ranged between 448.1 mgy*cm and 1618.1mGy*cm with a mean DLP 940.9mGy*cm. DISCUSSION The non-homogenous radiation dose distribution (Figure 3) often times leads to a decrease in the diagnostic quality of radiographs. The persistent sharp fall-off in radiation dose towards the left indicates that there was a misalignment with the collimator light field and the x-ray field (7). There was also evidence of anode heel effect (Figures 3 images A&B). The anode heel effect stipulates that the x-rays emitted toward the anode side of the tube will be of less energy and intensity than those emitted towards the cathode side of the tube as some of the x-ray photons emitted towards the anode side will pass through the heel of the anode and as such are attenuated to varying extent (7, 11). Over time, depending on the work load of the tube, Pitting may occur where the anode becomes damaged due to constant bombardment from electrons. This increases the heat at the anode surface and results in melting and cracking in the anode (7).The highest equivalent dose for each of the test carried out for plain x-ray scans was recorded at the centre of the radiation field. This is because the distance from the x-ray tube to the dose receptors was shortest at the 6

Dose (mgy) Walker et al centre as the tube was at a 90 angle to the centre of the radiation field. Photons therefore not impacting at the centre of the field will therefore travel a longer distance to the dose receptors. The intensity of radiation photon is inversely proportional to the square of the distance (7, 11). Increase distance also means more energy loss due to air kerma (7). Water was used as the tissue simulating material as it has the same effective atomic number as soft tissue (7.4), similar density (1g/cm 3 for water and 1.05g/cm 3 for soft tissue) and similar electron density as soft tissue (3.34x10 26 electrons/kg for water and 3.36x10 26 electrons/kg for soft tissue) (12). 6 Dose Distribution With Depth 4 2 0 0 2 4 6 8 10 12 14 16 Depth (cm) Parameter 1 Parameter 2 Figure 4: Absorbed dose with depth. The data in figure 4 shows that for plain x-ray scans a hypersthenic patient will receive a higher skin dose than a sthenic patient; 5.45mGy and 2.38mGy respectively. With the weighting factor of x-ray being 1, this results in a dose equivalent of 5.45mSv and 2.38mSv respectively. 7

Dose % Radiation Dose and Kidney Stone Dose % with Depth 150 100 50 0 y = -0.0075x 4 + 0.2239x 3-1.7394x 2-5.5497x + 100.5 R² = 0.9988 y = -0.0071x 4 + 0.2098x 3-1.5555x 2-6.5696x + 100.5 R² = 0.9988 0 2 4 6 8 10 12 14 16 Depth (cm) Parameter 1 Parameter 2 Fig. 5: Percentage depth dose. Figure 5 shows the dose distribution with depth when both readings have been normalised to 100%. The data has been fitted with the 4 th order of polynomial. With R 2 values of 0.99, the data has a high correlation coefficient. The data trajectory suggests a similar behaviour of x- ray photon interaction with soft tissue for both parameters with parameter 2 having a slightly higher percentage depth dose. This is because as the energy of the photon increases, so too does the percentage depth dose (13). The table shows the effective dose received by patients that underwent CT abdominal scans over a three-month period. It shows the range of dose length (DLP) product recorded by the unit as well as the mean DLP over this time period. Table: Effective dose received by patients that underwent CT Abdominal scan. Effective dose (DLP x 0.015mSv mgy-1 cm-1) DLP Range for Abdominal scans (without contrast) (mgy*cm) Mean DLP (mgy*cm) 448.1 1618.1 940.9 6.72mSv 24.27mSv 14.11mSv 8

Walker et al Calculations were done to convert the DLP to effective dose using k factor values from ICRP 103 (DLP x k factor for abdomen) (10). The dose readings for CT scans varied depending on the slice thickness and patient habitus. A decrease slice thickness results in an increase in image noise and as such an increase in mas is necessary to combat the increase in image noise (14). For scans of the abdomen for renal calculi, the entire abdomen is exposed to radiation. This leads to irradiation of all organs in the abdominal and pelvic cavity. Some organs are more radiosensitive than others and are given weighting factor numbers (WT). The higher the tissue weighting factor number, the more radiosensitive the organ is (15). There is no specific value for the tissue weighting factor for kidney, however other organs that are often exposed during an x-ray or CT scan for kidney stones and are more radiosensitive includes the Bone Marrow, Stomach and Colon with a WT of 0.12 and the Gonads with a WT of 0.08. Moderately radiosensitive tissues such as the Bladder and Liver (0.04WT) are also irradiated (15). The irradiation of these tissues increases the risk of stochastic effects to these organs more than to the kidneys and as such the use of ionizing radiation scans should be employed with caution. Despite the non-homogenous distribution of radiation dose in the irradiated field, all recorded dose conformed to those estimated by the ICRP for abdominal and urinary studies (8, 9). CONCLUSION Skin dose is greatest at the centre of the abdomen for plain x-ray abdominal scans. Radiation equivalent dose to the skin ranged from 2.38mSv for a sthenic body habitus patient to 5.45mSv for a hypersthenic patient. For CT scans, effective dose was found to range between 6.72mSv and 24.27mSv. All radiation dose measured were found to be within recommended ICRP 9

Radiation Dose and Kidney Stone operational limits (8, 9). The developed procedure could be implemented in other radiology departments in Jamaica as a part of their quality assurance program. ACKNOWLEDGEMENT We thank the International Atomic Energy Agency (IAEA) for fellowship training and equipment supplied through technical corporation programme. We also thank Mr Clifton McCook for his assistance in coding the Wolfram Mathematica software. 10

Walker et al REFERENCES 1. Morrison B. Hot days, increased risk of kidney stones Jamaica: The Jamaica Observer; 2014 [cited 2016 May 9]. Available from: http://www.jamaicaobserver.com/news/hotdays--increased-risk-of-kidney-stones_17195585 2. Jamaica: Jamaica Urological Society [cited 2016 May 9, 2016]. Available from: http://www.jus.org.jm/faq.php 3. Chang-Xu CZ, Wang XL, Liu TZ, Zeng X-T, Li S, Duan X-W. Self-fluid management in Prevention of Kidney Stones: A PRISMA-Compliant Systematic Review and Dose Response Meta-Analysis of Observational Studies. Medicine 2015 94. 4. Moore C, Daniels B, Ghita M, Gunabushanam G, Luty S, Molinaro A et al. Accuracy of reduced-dose computed tomography for ureteral stones in emergency department patients, 2014. 5. List of Health Facilities Non-Governmental organizations/community based organization and International Agencies In: Health Mo, editor. Jamaica 2012. 6. Kowalczyk N, Mace J. Radiographic pathology for technologists: Mosby Elsevier; 2009. 7. Bushong SC. Radiologic science for technologists: physics, biology, and protection. 10 ed. United States of America Elsevier Mosby; 2013. 8. What are the Radiation Risks from CT? United States of America United States Food and Drug Administration [cited 2016 May 9]. Available from: http://www.fda.gov/radiation- EmittingProducts/RadiationEmittingProductsandProcedures/MedicalImaging/Medical X-Rays/ucm115329.htm 9. Radiation Dose in X-Ray and CT Exams [cited 2016 May 9]. Available from: http://www.radiologyinfo.org/en/info.cfm?pg=safety-xray 11

Radiation Dose and Kidney Stone 10. AAPM. The measurement, reporting, and management of radiation dose in CT. United States of America American Association of Physicists in Medicine 2008 96. 11. Fauber T. Radiographic imaging and exposure 3ed. United States of America: Mosby Elsevier; 2009. 12. Hasegawa B. Physics of Medical X-ray Imaging. 2 ed. United States of America Medical Physics Publishing Corporation 01 Sep 1991 13. Gazda M, Coia L. Principles of Radiation Therapy. In: Cancer Management: A Multidisciplinary Approach [Internet]. Available from: http://www.thymic.org/uploads/reference_sub/02radtherapy.pdf 14. Goldman L. Principles of CT: radiation dose and image quality. Society of Nuclear Medicine 2007. 15. Health effects of security scanners for passenger screening (based on X-ray technology). Europe: Scientific Committee on Emerging and Newly Identified Health Risks (SCENIHR) 2012 April 26. Report No. 12