I - Gel Versus Cuffed Tracheal Tube in Elective Laparoscopic Cholecystectomy A Clinical Comparative Study

Similar documents
I-gel vs cuffed tracheal tube during volume controlled ventilation in elective laparoscopic cholecystectomy

RESPIRATION AND THE AIRWAY Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation

Deposited on: 24 February 2009

Deposited on: 11 February 2010

A prospective randomized study comparing the efficacy of the LMA Classic TM, the

ISSN X (Print) Research Article. *Corresponding author Dr. Souvik Saha

Comparison of the i-gel and the LMA-Unique laryngeal mask airway in patients with mild to moderate obesity during elective short-term surgery

Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.

The Laryngeal Mask and Other Supraglottic Airways: Application to Clinical Airway Management

Comparison of the LMA Supreme vs the i-gel TM in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation*

A Comparative Study of Two Disposable Supraglottic Devices in Diagnostic Laparoscopy in Gynecology

Comparison of clinical performance of i-gel with laryngeal mask airway pro-seal in elective surgery in adults

International Journal of Health Sciences and Research ISSN:

Evaluation of Baska Mask Performance in Laparoscopic Cholecystectomy

Dr. Ranjeet Rana De 1, Dr. Saurav Shekhar 2, Dr. D G Pathak 3, Dr. Harshwardhan 4, Dr. Shashank Dhiraj 5 1,2,4,5

A randomised comparison between Cobra PLA and classic laryngeal mask airway and laryngeal tube during mechanical ventilation for general anaesthesia

Use of the Intubating Laryngeal Mask Airway

DIFFICULT AIRWAY MANAGMENT. Dr.N.SANTHOSH KUMAR MD ANESTHESIA (2 nd Yr)

COMPARATIVE CLINICAL EVALUATION OF PROSEAL LMA WITH I - GEL AIRWAY IN PATIENTS UNDERGOING ELECTIVE SURGICAL PROCEDURES UNDER GENERAL ANAESTHESIA

Citation British journal of anaesthesia, 104. pp ; 2010 is available onlin

A comparative study between i-gel and classical laryngeal mask airway in elective surgery under general anaesthesia

ISPUB.COM. M Roberts, M Mani, A Wilkes, E Flavell, N Goodwin INTRODUCTION

A Comparative Study of Classic LMA and Proseal LMA in Paralyzed Anaesthetized Patients

Comparison of efficacy of the Laryngeal tube with the Laryngeal mask airway in securing the upper airway

Comparison of I-gel with Baska Mask Airway for Controlled Ventilation in Obese Patients Undergoing Ambulatory Surgery: A Prospective Randomized Trial

LMA Supreme Second Seal. Maintain the airway. Manage gastric contents. Meet NAP4 recommendations.

Recent Advances in Airway Management HA Convention 2014

ORIGINAL ARTICLE. Shashank Chitmulwar, MD, Charulata Deshpande, MD, DA ABSTRACT. ANAESTHESIA, PAIN & INTENSIVE CARE

A Clinical Comparative Study Of Evaluation Of Proseal LMA V/S I-GEL For Ease Of Insertion And Hemodynamic Stability; A Study Of 60 Cases

Use of the Aintree Intubation Catheter with the Laryngeal Mask Airway and a Fiberoptic Bronchoscope in a Patient with an Unexpected Difficult Airway

Comparison of the clinical performance of i-gel, LMA Supreme and LMA ProSeal in elective surgery

The LMA CTrach TM, a new laryngeal mask airway for endotracheal intubation under vision: evaluation in 100 patients

Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy

Comparison of Ease of Insertion and Hemodynamic Response to Lma with Propofol and Thiopentone.

University of Groningen

Anesthetic challenges when elective case becomes emergent

CONTINUING EDUCATION IN HONOR OF NORMAN TRIEGER, DMD, MD

Airway Management & Safety Concerns Experience from Bariatric Surgery

Positive pressure ventilation with the laryngeal mask airway in non-paralysed patients: comparison of sevoflurane and propofol maintenance techniques

Comparison of the Baska â mask with the single-use laryngeal mask airway in low-risk female patients undergoing ambulatory surgery

Randomised comparison of the LMA Supreme with the I-Gel in spontaneously breathing anaesthetised adult patients

Received 1 November 2009 Revised 8 December 2009 Accepted 10 December 2009

Clinical Study McGrath Video Laryngoscope May Take a Longer Intubation Time Than Macintosh Laryngoscope

Comparison of the Hemodynamic Responses with. with LMA vs Endotracheal Intubation

A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures

MD (Anaesthesiology) Title (Plan of Thesis) (Session )

Laryngeal mask airway (LMA-ProSeal) malfunction causing acute airway obstruction

The Pro-Seal LMAtm And The Tracheal Tube: A Comparison Of Events At Insertion Of The Airway Device

Similar oropharyngeal leak pressures during anaesthesia with i-gel TM, LMA-ProSeal TM and LMA-Supreme TM Laryngeal Masks

Original Contributions

FAILED ELECTIVE INTUBATION: PLAN A- C

JMSCR Vol 4 Issue 02 Page February 2016

Comparison of the efficacies of I-gel TM and LMA-ProSeal TM for airway management in pediatric patients

EUROANESTHESIA 2007 Munich, Germany, 9-12 June 2007

LEVITAN S FIBREOPTIC STYLET: BEYOND BARRIERS. - Our Perspective.

LMA Unique Airway Portfolio

ASPIRATION DURING ANAESTHESIA

Comparision of Hemodynamic Changes after Insertion of Classic Lma and Proseal Lma

Asai and Shingu Table 1 Size selection and recommended cuff volumes Size Patients Body size Recommended cuff volumes (ml) Connector colour 0 Newborn <

Initial anatomic investigations of the I-gel airway: a novel supraglottic airway without inflatable cuff*

Mouth opening with cervical collars Fig 1 Types of semi-rigid collar used (left to right): Stifneck (Laerdal Medical Corp.); Miami J (Jerome Medical);

Other methods for maintaining the airway (not definitive airway as still unprotected):

Unanticipated difficult tracheal intubation - during routine induction of anaesthesia in an adult patient

In 2011 I received an unrestricted lecture honorarium from Ambu I have been loaned equipment by Intavent Direct (Teleflex) and Aircraft Medical As an

Cricoid pressure: useful or dangerous?

J. Brimacombe, 1 L. Holyoake, 2 C. Keller, 3 J. Barry, 4 D. Mecklem, 4 A. Blinco 5 and K. Weidmann 5

Pharyngolaryngeal Morbidity with the Laryngeal Mask Airway in Spontaneously Breathing Patients

Tibe ProSeal Laryngeal Mask Airway

Materials and Methods

A comparative Study of Tracheal Intubation through i-gel and Intubating Laryngeal Mask Airway ( ILMA ) G. Bhandari, *K.S. Shahi, R.

Clinical Study Randomised Comparison of the AMBU AuraOnce Laryngeal Mask and the LMA Unique Laryngeal Mask Airway in Spontaneously Breathing Adults

CAN T INTUBATE, CAN T VENTILATE: PLAN A- D

Dr.Bharghavi.M 2 nd year post graduate Dept of Anaesthesia

ORIGINAL ARTICLE ABSTRACT

RESPIRATION AND THE AIRWAY. Editor s key points. L. Theiler 1,2 *, M. Kleine-Brueggeney 1,3, N. Urwyler 1,4,T.Graf 1, C. Luyet 1 and R.

J of Evolution of Med and Dent Sci/ eissn , pissn / Vol. 3/ Issue 64/Nov24, 2014 Page 13923

Clinical Comparison of Lightwand-guided versus Conventional Endotracheal Intubation

LMA CTrach TM in patients with anticipated difficult airway: A retrospective study

Original Article INTRODUCTION. Abstract

Veena Patodi*, Maina Singh, Surendra K. Sethi, Vini N. Depal, Neena Jain, Vijay Kumar

Observation of ventilation effects of I-gel TM, Supreme TM and Ambu AuraOnce TM with respiratory dynamics monitoring in small children

AIRWAY MANAGEMENT AND VENTILATION

Airway Management. Teeradej Kuptanon, MD

i-gel TM supraglottic airway in clinical practice: a prospective observational multicentre study

Optimal Size AMBU Laryngeal Mask Airway Among Asian Adult Population

Mao-Kai Chen 1, Hung-Te Hsu 1, I-Cheng Lu 2, Chih-Kai Shih 3, Ya-Chun Shen 1, Kuang-Yi Tseng 1 and Kuang-I Cheng 1,4*

Supra-glottic Airway Devices

Comparative Evaluation of Performance of Videolaryngoscope vs Fastrach Intubating Laryngeal Mask Airway

The Laryngeal Tube. An Evaluation of the Laryngeal Tube During General Anesthesia Using Mechanical Ventilation

Provision of General Anesthesia Out of Hospital: Perspective from The Americas. Anthony Charles Caputo

Z.U.M.J.Vol.19; N.5; September; 2013

The flexible laryngeal mask airway (FLMA) was

Editor s key points. M. Kleine-Brueggeney 1,2, L. Theiler 1,3 *, N. Urwyler 1,4, A. Vogt 1 and R. Greif 1

Randomized Comparison of Actual and Ideal Body Weight for Size Selection of the Laryngeal Mask Airway Classic in Overweight Patients

Buyers guide. Laryngeal masks CEP 08010

Review Article. Summary. Introduction. D. K. Baidya, 1 Chandralekha, 2 V. Darlong, 3 R. Pandey, 3 S. Maitra 4 and P. Khanna 5

Success of Tracheal Intubation with Intubating Laryngeal Mask Airways

Is the Airtraq optical laryngoscope effective in tracheal intubation by novice personnel?

How do you use a bougie as an airway adjunct for endotracheal intubation?

Comparison of Three Disposable Extraglottic Airway Devices in Spontaneously Breathing Adults

Transcription:

Original Research I - Gel Versus Cuffed Tracheal Tube in Elective Laparoscopic Cholecystectomy A Clinical Comparative Study Siddharam Jamagond 1, Anuradha H 2, Ramesh.K 3 1 Senior resident, Department of Anesthesia, Koppal Institute of Medical Sciences, Koppal, Karnataka 2 Assistant Professor, Department of Anesthesia, Koppal Institute of Medical Sciences, Koppal, Karnataka 3 Associate Professor, Department of Community Medicine, VIMS, Ballari, Karnataka Corresponding Author: E-mail: ramspsm@gmail.com ABSTRACT: Background: This study aims to test and compare Endo tracheal tube and i-gel in terms of their: Efficacy: Difference in the leak fraction between two airway devices before and after pneumoperitoneum with different tidal volumes andcomparison of oro pharyngeal leak pressure. Ease of insertion: Number of attempts required for optimal positioning. Methods: Sixty patients, ASA I II, were randomly selected to the study. Standard anaesthetic technique was used for all patients. The i-gel was then inserted. The lungs were ventilated at three different tidal volumes (6, 8 and 10 ml kg1) using volume controlled ventilation (VCV). The leakvolume was calculated as the difference between the inspired and expired tidal volumes. The leakfraction was also calculated as the leak volume divided by the inspired tidal volume. These observations were recorded with every tidal volume before and after pneumoperitoneum with the i-gel and the conventional tracheal tube Results: We found oro pharyngeal leak pressure for i gel as 26cm of H2O and there was no leak in endotracheal tube group even at 40 cm of H2O peak air way pressure. Before and after pneumoperitoneum there was no statistically significant difference in leak fraction or leak volume between i-gel and tracheal tube at tidal volume 6ml kg 1. At 8 and 10 ml kg 1 there was statistically significant difference between i-gel and tracheal tube both before and after pneumoperitoneum Conclusion: In our study we found that i-gel airway can be used safely and effectively during volume controlled ventilation with low and moderate tidal volumes Key words: I-gel; Leak fraction; Pneumoperitoneum; Cholecystectomy Quick Response Code: Access this article online Website: www.innovativepublication.com DOI: 10.5958/2394-4994.2015.00036.0 INTRODUCTION Safe and effective airway management is the foundation of quality anesthetic practice. Supraglottic airway devices have revolutionized airway management since the invention of the LMA Classic (LMA North America Inc., California, USA) by Dr Archie Brain in 1988. They fill a niche between the face mask and the endotracheal tube in terms of both anatomical position and degree of invasiveness 1. The ease of insertion, safety and the global increase in the number of day care surgeries have led to their increased use in routine anesthetic practice. Since the introduction of the LMA Classic, several laryngeal masks have been introduced which differ in shape, stiffness, cuff properties and constituent material 2. The Ambu Aura 40 (Ambu A/S, Copenhagen, Denmark) laryngeal mask and the I-gel (Intersurgical Ltd, Wokingham, U.K.) are two such devices. Apart from being used to maintain the airway routinely during an anaesthetic, laryngeal masks have now come to play an important role in the management of difficult airways and in emergent situations such as cardio-pulmonary resuscitation 1. The i-gel (Intersurgical Ltd., Wokingham, UK) is a new supraglottic airway device (SAD) made of thermoplastic elastomer which is soft, gel-like and transparent. 3 Studies on Cadaver showed that i-gels effectively conformed to the per laryngeal anatomy and consistently achieved proper positioning for supraglottic ventilation 4. Manikins studies and patients have shown that the insertion of the i-gel was significantly easier when compared with insertion of other SADs 5,6. Few studies had been done to evaluate the use of i-gel during controlled ventilation but they did not evaluate its use during procedures with airway pressure more than 25 cm H2O 7. Our study was designed to evaluate the i-gel sealing pressure and as effective airway as cuffed tracheal tube during volume controlled ventilation inelective laparoscopic cholecystectomy.this study aims to test and compare cuffed endotracheal tube and i-gel in terms of their: Efficacy: Difference in the leak fraction between two airway devices before and after pneumoperitoneum with different tidal volumes and comparison of oropharyngeal leak pressure. Ease of insertion: Number of attempts required for optimal positioning. Indian Journal of Clinical Anaesthesia, 2015;2(4):235-239 235

METHODOLOGY The study was conducted in Malabar institute of Medical Sciences, a tertiary care centre during the period of June 2012 to may 2013.This is an institutionbased randomised case-control study. The approval of institutional ethics committee was obtained prior to the commencement of the study. 60 patients of ASA 1 and 2,with BMI <35kg/m2,between 18 and 60 years posted for elective laparoscopic cholecystectomy were enrolled into the study. The sample size was determined by considering a difference in the leak fraction more than 20%for the i-gel when compared to tracheal tube to be significant. They were randomized into two groups of equal number using the chit-in-a-box method for the use of either i-gel or endo tracheal tube for the maintenance of airway during the anaesthesia. After induction of anaesthesia by a suitable intravenous induction agent and after achieving adequate anaesthetic depth, the randomly chosen, appropriately sized airway device was inserted and connected to the breathing circuit. The following parameters were then studied: 1. Number of attempts for correct positioning of the device. 2. Oropharyngeal leak pressure 3. Leak volume and leak fraction The data was analysed using the Statistical Package for Social Sciences software for Windows. The paired t test, t test, fisher exact test, and Pearson s Chi square tests were used for comparing the data. RESULTS Table1: Comparison of age based on group Age I-gel 20-29 9 30.0 7 23.3 30-39 6 20.0 6 20.0 40-49 5 16.7 4 13.3 50-60 10 33.3 13 43.3 Mean ± SD 40.2 ± 13 42.9 ± 12.9 t= 0.80, p= 0.427 (significant if p< 0.05) The mean age of I-gel group was 40.2 years and whereas group was 42.9 years. This difference was not statistically significant. Table2: Distribution according to American society of Anesthesiologists physical status classification American Society of Anesthesiologists physical status classification Grade I 20 66.7 16 53.3 Grade II 10 33.3 14 46.7 2 = 1.11, p = 0.292 (significant if p< 0.05) The airway characteristics of the patients studied i.e. mouth opening (Table 3), thyromental distance (Table 4) and the Mallampati (Table 5) score were also noted and statistically analysed, the results were not statistically significant. Table 3: Distribution according to mouth opening Mouth Opening 5 cm designated 4 13.3 6 20.0 > 5 cm designated 26 86.7 24 80.0 2 = 0.48, p = 0.488 (significant if p< 0.05) Table 4: Distribution according to thyromental distance Thyromental Distance 6 cm designated 4 13.3 3 10.0 > 6 cm designated 26 86.7 27 90.0 p = 0.500 (Fisher Exact test) (significant if p< 0.05) Indian Journal of Clinical Anaesthesia, 2015;2(4):235-239 236

Mallampati grade Table5: Comparison ofmallampati grade 1 3 10.0 2 6.7 2 27 90.0 28 93.3 p = 0.500 (Fisher Exact test) (significant if p< 0.05) The number of attempts at insertion needed to get a proper positioning of each device was noted and analysed. The i- gel could be positioned successfully with a single attempt in 86.2% of the patients in whom the device was used(26 out of the 30 patients studied),whereas successful placement at first attempt could be achieved in 93.3% of the subjects in endotracheal group(28 out of 30 patients). The i- gel and endotracheal tube could be positioned successfully in second attempt in remaining cases. I-gel had to be manipulated in 3 cases. There were no instances of failure to secure an airway with the chosen device. This result does not show a statistical significance (P = 0.335) Table6: Comparison of no. of attempts required to insert the device based on group No. of attempts required to insert the device 1 26 86.7 28 93.3 2 4 13.3 2 6.7 p = 0.335 (Fisher Exact test) (significant if p< 0.05) Table7: Comparison of manipulation of airway to maintain adequate ventilation based on group Manipulation of air way to maintain adequate ventilation Yes 3 10.0 0 0.0 No 27 90.0 30 100.0 p = 0.119 (Fisher Exact test) (significant if p< 0.05) The oropharyngeal leak pressure (OPLP) measured while using each device was measured and the average was calculated. It is as follows median value of OPLP in i-gel group is 26 cm H20 Table8: Distribution according to oropharyngeal leak pressure(cm H2O) Oropharyngeal Leak Pressure 20-29 21 70.0 0 0.0 30-39 9 30.0 0 0.0 > 40 0 0.0 30 100.0 Before and after pneumoperitoneum there was no statistically significant difference in leak fraction between i-gel and tracheal tube at tidal volume 6ml kg 1 (P = 0.620 and 0.956 ).The mean difference in leak fraction was significant between i- gel and endotracheal group before and after pneumoperitoneum at 8 ml kg 1 tidal volume (P = 0.000 ). At 10 ml kg 1 there was statistically significant difference between i-gel and tracheal tube both before and after pneumoperitoneum (P = 0.000). Before and after pneumoperitoneum leak volume at 6 ml kg 1 showed no significant difference (P=0 212and0.972). But we found statistically significant difference before and after pneumoperitoneum at 8ml kg 1 and 10 ml kg 1. Indian Journal of Clinical Anaesthesia, 2015;2(4):235-239 237

Table 9: Comparison before pneumoperitoneum based on group Mean SD N t p **: significant at 0.01 level I- gel 14.8 1.8 30 Endo tracheal tube 15.9 4.6 30 I- gel 27.3 5.2 30 Endo tracheal tube 17.8 4.7 30 I- gel 43.6 7.1 30 Endo tracheal tube 18.9 4.8 30 I- gel 0.0 0.0 30 0.1 0.0 30 I- gel 0.1 0.0 30 1.26 0.212 7.44** 0.000 15.78** 0.000 0.50 0.620 7.35** 0.000 13.55** 0.000 Table10: Comparison after pneumoperitoneum based on group Mean SD N t p **: significant at 0.01 level I gel 16.2 2.5 30 Endo tracheal tube 16.2 4.6 30 I gel 29.7 3.9 30 Endo tracheal tube 17.4 3.4 30 I gel 54.4 10.6 30 Endo tracheal tube 18.5 3.0 30 I gel 0.0 0.0 30 I gel 0.1 0.0 30 I gel 0.1 0.0 30 0.04 0.972 13.07** 0.000 17.8** 0.000 0.06 0.956 8.81** 0.000 13.88** 0.000 DISCUSSION Laryngeal masks have played an important role in airway management since the introduction of the LMA Classic in 1988. Since then, several laryngeal masks varying in their shape, stiffness, cuff properties and clinical applications have come into existence. In addition to their use during routine anesthetics, they have also been recommended for use in difficult airway scenarios 8,9 and in cardio-pulmonary resuscitation. Supra glottic airway devices have several advantages including lower incidence of sore throat 10, less hemodynamic upset during induction and maintenance of anaesthesia 11,12 and better oxygenation during emergence 13. I-gel is a relatively new disposable supraglottic airway device that has no inflatable cuff. It has an integral bite block, wide bore lumen, and an additional distal lumen that allows for the passage of a gastric tube. These features may give the i-gel an advantage over the LMA and even the Pro-Seal LMA. There was too much debate among anesthesiologists about using these devices during procedures requiring positive pressure ventilation. During high tidal volume ventilation and laparoscopic procedures peak airway pressure rises and exceeds airway sealing (leak) pressure leading to increase in leak volume and fraction. These findings explain difficulties in maintaining optimum ventilation We analyzed 60 patients undergoing elective laparoscopic cholecystectomy. They were randomized into two groups of equal numbers using the chit-in-abox method for the use of either i-gel or endotracheal tube for the maintenance of airway during the anesthesia. Both groups were comparable in terms of age, sex and ASA status. Height, weight and BMI were also statistically comparable. The airway characteristics of all patients studied in terms of mouth opening, thyromental distance and the mallampati scores were also comparable Ease of Insertion After induction of anesthesia, the randomly chosen device (i gel or endotracheal tube) of appropriate size was inserted and the number of Indian Journal of Clinical Anaesthesia, 2015;2(4):235-239 238

attempts needed for proper positioning of the device was noted. In our study we did not find any significant difference between two and number of attempts require to secure i -gel in our study is comparable to other international studies. 14 Oropharyngeal Leak Pressure (OPLP) The oropharyngeal leak pressure is the airway pressure at which gases begins to leak around the cuff of the laryngeal mask airway device Uppalet al. 7 found leak pressure for i-gel 28 (20 35) cm H2O by both auscultation and manometer stabilization methods. In our study we concluded that airway leak pressure for i-gel was 26 cm H2O. Ishwar et al. 14 concluded that airway leak pressure for i-gel was 25.27 cm H2O using same methods. Lu et al. 15 compared Pro-Seal laryngeal mask airway (PLMA) with Classic laryngeal mask airway (LMA) for positive pressure ventilation during laparoscopic cholecystectomy. They concluded that PLMA is more effective ventilator device for laparoscopic cholecystectomy than classic LMA. This was attributed to higher leak pressure due to large cuff size (leak pressure was 29± 6 cm H2O). We thought that i-gel could be used during such procedures but unfortunately during our study we found leak pressure for i-gel was 26 cm H2O which is less than peak pressure during pneumoperitoneum especially at moderate and high tidal volumes 16. In our study we found oropharyngeal leak pressure more than 40 cm of H2O for endotracheal tube. We did not correlate anatomical position of i-gel with clinically evident leaks by using fibreoptic bronchoscope 17 Leak fraction and leak volume Before and after pneumoperitoneum there was no statistically significant difference in leak fraction or leak volume between i-gel and tracheal tube at tidal volume 6ml kg 1. At 8 and 10 ml kg 1 there was statistically significant difference between i-gel and tracheal tube both before and after pneumoperitoneum. CONCLUSIONS Our study supports the use of i-gel during VCV in elective laparoscopic cholecystectomy using low to moderate tidal volumes provided that peak airway pressure not more than device leak pressure. Although leak volume was significant, ventilation and oxygenation were optimal in most cases. Tracheal tube should be inserted if failed ventilation and oxygenation. 6. Devitt JH, Wenstone R, Noel AG, O Donnell MP. The laryngeal mask airway and positive-pressure ventilation. Anesthesiology 1994;80:550 5. 7. Levitan RM, Kinkle WC. Initial anatomic investigations of the i-gel_ airway: a novel supraglottic airway without inflatable cuff. Anesthesia 2005:60:1022 6. 8. Jackson KM, Cook TM. Evaluation of four airway training manikins as patient simulators for the insertion of eight types of supraglottic airway devices. Anaesthesia 2007;62:388 93. 9. Wharton NM, Gibbison B, Gabbott DA, Haslam GM, Muchatuta N, Cook TM. i-gel insertion by novices in manikins and patients. Anaesthesia 2008;63:991 5. 10. Uppal V, Fletcher G, Kinsella J. Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation. J Anaesth 2009;102(2):264 8, ISSN: 0007-0912. 11. American Society of Anesthesiologists Task Force on Management of the Difficult Airway- Practice Guidelines for Management of the Difficult Airway. Anesthesiology 2003; 98:1269 77 12. Henderson JJ, Popat MT, Latto IP et al; Difficult Airway Society Guidelines for management of the unanticipated difficult intubation. Anaesthesia 2004:59; 675-694 13. Higgins PP, Chung F, Mezei G. Postoperative sore throat after ambulatory surgery. Br J Anaesth 2002;88:582 4. 14. Dyer RA, Llewellyn RL, James MF. Total i.v.anaesthesia with propofol and the laryngeal mask for orthopaedic surgery. Br J Anaesth 1995;74:123 8. 15. Cork RC, Depa RM, Standen JR. Prospective comparison of use of the laryngeal mask and endotracheal tube for ambulatory surgery. Anesthesia Analgesia 1994;79:719 27 16. Webster AC, Morley-Forster PK, Dain S, et al. Anaesthesia for adenotonsillectomy: a comparison between tracheal intubation and the armoured laryngeal mask airway. Can J Anaesth1993;40:1171 7. 17. Ishwar S, Monika G, Mansi T. Comparison of clinical performance of i-gel_ with LMA ProSeal_ in elective surgeries. Indian J Anaesth 2009;53(3):302 5. 18. Lu PP, Brimacombe J, Yang C, Shyr M. ProSeal versus the Classic laryngeal mask airway for positive pressure ventilation during laparoscopic cholecystectomy. Br J Anaesth 2002;88(6): 824 7 19. Brimacombe J, Keller C, Fullekrug B, et al. A multicenter study comparing the ProSeal and Classic laryngeal mask airway in anesthetized, nonparalyzed patients. Anesthesiology 2002;96: 289 95. 20. Van Zundert A, Brimacombe J, Kamphuis R, Haanschoten M. The anatomical position of three extraglottic airway devices in patients with clear airways. Anaesthesia 2006;61:891 5. REFERENCES: 4. T. M. Cook, The classic laryngeal mask airway: a tried and tested airway. What now? British Journal of Anaesthesia,2006; 96: 149 152, 5. Ng SY, Teoh WH, Lim Y, Cheong VG. Comparison of the AMBU Laryngeal Mask and the LMA Classic in anaesthetised, spontaneously breathing patients. Anaesth Intensive Care. 2007 Feb;35:57-59 Indian Journal of Clinical Anaesthesia, 2015;2(4):235-239 239