Automatic buffy coat extraction: methodology, feasibility, optimization and validation study

Similar documents
CyBio Felix Your Automated Application Starts Here

CEPHIA Consortium for the Evaluation and Performance of HIV Incidence Assays STANDARD OPERATING PROCEDURE

Fast and Easy Isolation of T Cells

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

For Research Use Only Ver

Product Contents... 1 Specifications... 1 Product Description... 2 Buffer Preparation... 3 Protocol Appendix... 5 Ordering Information...

In vitro human regulatory T cell expansion

ABIOpure TM Viral (version 2.0)

HbA1c (Human) ELISA Kit

For Research Use Only Ver

In vitro human regulatory T cell expansion

Chapter 19(1) An Introduction to the Circulatory System and Blood

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell Culture (CD63 / CD81)

Quality and standardization in blood component preparation with an automated blood processing technique

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents

Chapter 19(1) An Introduction to the Circulatory System and Blood

XTRAKT FFPE Kit / Manual

Establishing a Pure Lymphocyte Gate for Subset Analysis by Flow Cytometry

Primary Adult Naïve CD4+ CD45RA+ Cells. Prepared by: David Randolph at University of Alabama, Birmingham

For purification of viral DNA and RNA from a wide range of sample materials

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies

Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit

BIOCHEMISTRY OF BLOOD

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Concentration Estimation from Flow Cytometry Exosome Data Protocol

CoQ10(Coenzyme Q10) ELISA Kit

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood

Auto Validation and Management Reports Using Sysmex WAM. Alfonso Ziccardi Laboratory Operation Manager for AP/CP/Safety Officer

Performance evaluation of Celltac G: a new automated hematology analyzer

Rapid antigen-specific T cell enrichment (Rapid ARTE)

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Start your T cell research right

PRODUCT: RNAzol BD for Blood May 2014 Catalog No: RB 192 Storage: Store at room temperature

90 min 18 min. 45 min. 14 d

Sponsored and reviewed by ICCS Quality and Standards Committee Title: Verification of PNH assay sensitivity through spiking experiment Written by:

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell culture

IMMUNOLOGY. Source, Isolate, Culture, And Analyze Immune Cells. Scientists Helping Scientists

Activity Overview. P.L.E.P: Parts of Blood. Cast Your Net: Adventures With Blood. Activity 1A. Activity Objectives: Activity Description:

CHAPTER 3 LABORATORY PROCEDURES

Technical Bulletin No. 157

Analysis of Micronutrients in Milk by Flame Atomic Absorption Using FAST Flame Sample Automation for Increased Sample Throughput

AquaPreserve DNA/RNA/Protein Order # Preservation and Extraction Kit 8001MT, 8060MT

BD Multitest IMK Kit

PRODUCT INFORMATION & MANUAL

PRODUCT INFORMATION & MANUAL

Consumables PERFORMANCE ASSURED. QUALITY ASSURED. APPLICATION ASSURED.

EXO-DNA Circulating and EV-associated DNA extraction kit

Direct ex vivo characterization of human antigen-specific CD154 + CD4 + T cells Rapid antigen-reactive T cell enrichment (Rapid ARTE)

Supplementary material page 1/10

MagniSort Human CD4 Memory T cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

Mouse C-Peptide ELISA Kit

Amendment No. 2. Item No. 2 (Rfx/ Event number )

ExiPrep TM Viral DNA/ RNA Kit

DuraClone IM. Standardized phenotyping panels for studies of the human immune system

PointCare NOW TM Technical Note

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

ab Nuclear Extract Kit

BLOOD. EEI n: t.ee# Required Name : Due : April 12,2018 COMPOSITION AND FUNCTIONS OF BLOOD. Beginningof Class

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx

Leukemia BCR-ABL Fusion Gene Real Time RT-PCR Kit

Human HIV (1+2) antigen&antibody ELISA Kit

MagniSort Human CD4 T cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

ab Lipid Extraction Kit (Chloroform-Free)

VEGFR2-Mediated Vascular Dilation as a Mechanism of VEGF-Induced Anemia and Bone Marrow Cell Mobilization

Viral RNA / DNA purification products from MACHEREY-NAGEL. MN guide for viral RNA / DNA purification Multiple solutions for many needs

Blocking antibodies and peptides. Rat anti-mouse PD-1 (29F.1A12, rat IgG2a, k), PD-

Copyright 2010 Pearson Education, Inc. Blood Vessel Structure

For Research Use Only Ver

EliKine Free Thyroxine (ft4) ELISA Kit

CELL-DYN 3700 Strength in Technology, Proven Reliability

Abstract for Blood Bag Production. This abstract refers to blood bag production considering general and process information

HEAMATOLOGICAL INDICES AND BONE MARROW BIOPSY

Gladstone Institutes, University of California (UCSF), San Francisco, USA

Human Alpha 1 microglobulin ELISA Kit

Fluorochrome Panel 1 Panel 2 Panel 3 Panel 4 Panel 5 CTLA-4 CTLA-4 CD15 CD3 FITC. Bio) PD-1 (MIH4, BD) ICOS (C398.4A, Biolegend) PD-L1 (MIH1, BD)

Flow Cytometric Analysis of Cerebral Spinal Fluid Involvement by Leukemia or Lymphoma

Update on CD4+ T-Cell Counting Technologies

FinTest IgG4 Screen 20 ELISA KIT

IV2-113E Use by. Invitron Glargine ELISA Kit REF LOT IVD. Definitions. English. For in-vitro diagnostic use. Instructions for use.

Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay

Procine sphingomyelin ELISA Kit

MACSxpress Technology Untouched cell isolation directly from blood products

BLOOD RUNS THROUGH YOUR BODY

Incorporating Differentials Into Every Complete Blood Count. Paige Flowers, LVT Dogwood Veterinary Internal Medicine

Sputum DNA Collection, Preservation and Isolation Kit 50 Individual Devices

idc 20 AUTOMATED LIQUID-BASED CYTOLOGY SLIDE PREPARATION SYSTEM

EXO-DNAc Circulating and EV-associated DNA extraction kit

STANDARD OPERATING PROCEDURE TITLE: TOTAL IGE EIA (HYTEC 288)

155 S. Milpitas Blvd., Milpitas, CA USA T: (408) F: (408)

Bead Based Assays for Cytokine Detection

Standard Operation Procedure (SOP) for Biobanking Sampling Procedure Manual Use

Cotinine (Mouse/Rat) ELISA Kit

attomol HLA-B*27-Realtime LT 2 Assay for the detection of the human HLA-B*27-locus using LightCycler (Do not use for tissue typing!

ab Mammalian Cell Lysis Buffer 5X

Supplemental Figure 1. Activated splenocytes upregulate Serpina3g and Serpina3f expression.

Clinico-cytometric classification of PNH

Transcription:

Automatic buffy coat extraction: methodology, feasibility, optimization and validation study Conny Mathay 1, Wim Ammerlaan 1 and Fay Betsou 1 1 Integrated Biobank of Luxembourg, 6 rue Nicolas Ernest Barblé, 1210 Luxembourg, Luxembourg Buffy coat isolation from has typically been a long, manual process. We tested and evaluated the feasibility, efficiency and reproducibility of extracting buffy coat by an automated process with the Tecan pipetting robot Freedom Evo200. The Buffy coat (BC) is defined as the layer of leukocytes and platelets that forms between the red blood cells and the plasma when unclotted blood is centrifuged or allowed to stand. The automation of BC extraction is challenging because not only do the blood volumes in the individual blood collection tubes vary but also the quantity, viscosity and texture of the BC layer itself can present important inter-individual variations (e.g. leukocyte numbers increase upon infection). Compared to manual BC extraction, automation guarantees high sample throughput, electronically controlled speed and volume of BC aspiration, no operator-due variability, tracking of sample identifications and recording of all instrument-specific parameters of BC extraction. Methodology and Feasibility Essential platform components for the automation of the BC extraction are the software, the pipetting arm and the Tube Inspection Unit (TIU), which detects differently colored layers of centrifuged blood by laser. For platform specifications please refer to Figure 1. K 2 -EDTA, 10 ml blood collection tubes were centrifuged at 2000g for 10 minutes at room temperature so that the blood separated into plasma, red blood cells and the intermediate thin, white colored layer: the buffy coat (Figure 2). The pipetting robot Freedom Evo200 used a customized human BC script where the blood tubes are first barcode-identified then the software converts the location of the BC location into X, Y, Z vectors, which guide the tips of the pipetting arm to dip precisely into the BC layer and to slowly aspirate the BC. Finally the extracted BC (approximately 1ml) is ejected into a separate tube. One BC extraction run for 1-8 samples takes 5 min 20 sec, our configuration of the platform permits to process 24 samples in one run. Leukocyte content of automatically extracted BCs: Verification The leukocyte subpopulation content of automatically extracted buffy coats from three adult donors was analysed by flow cytometry (Table 1). Compared to reference ranges of leukocyte subpopulations in blood of healthy adults (1, 2), normal values for leukocyte subpopulations (lymphocytes (B-cells and T-cells), monocytes and granulocytes) were detected when BCs were extracted on the Freedom Evo200 platform (natural killer cells were not investigated). Optimization and efficiency 1

BC is often used for DNA extraction. Quantification of the extracted DNA, from either automatically or manually prepared BCs, permits to assess the efficiency of the BC automation process. Data presented in Table 2 were obtained from 364 healthy volunteer donors over a 6-month timespan during which the BC automation was optimized. The same DNA extraction and quantification protocols were used for all specimens. BCs were prepared from 8ml blood either automatically or manually, and half of the BC volume was used for DNA extraction, corresponding to an initial 4ml blood volume. The automated BC extraction resulted in a significantly lower DNA yield (47.8 µg) compared to manual BC preparation (61.3 µg). When the whole BC prepared by the Freedom Evo200 was used for DNA extraction, corresponding to 8ml of, a median DNA yield of 147.9 µg was obtained. Manual BC extraction from 8ml blood resulted in a similar median DNA yield (145.2 µg). Optimization of the BC script for automation resulted in a significantly higher DNA yield (199 µg) when compared to the yields obtained with the previous script. The optimization took into account some specific mechanical parameters of the used tube type. Additionally the settings were optimized to allow a complete aspiration of the BC. As a result, the efficiency of optimized automated BC preparation in terms of DNA yield is higher than in manually prepared BCs. Validation To compare the physical content of extracted BCs to, cell counting was performed on blood before BC extraction and on automatically (optimized script) and manually extracted BCs from the same three donors (Table 3). The automatically extracted BCs contained significantly more leukocytes and platelets than the manually prepared BCs. Compared to, 95% of the blood leukocytes could be recovered in the automatically prepared BCs, whereas in the manually extracted BCs the leukocyte recovery percentage was 56%. Confirmation of these data came from a visual inspection of the collection tubes 30 minutes after the buffy coat extraction, where less buffy coat left-over in the automated tubes was clearly observed (Figure 3). Red blood cell (RBC) contamination in manually or automatically prepared BCs was not significantly different, however in two of the three examined cases, RBC contamination was higher in the manually extracted buffy coats. Haemoglobin values were low both in manually and automatically prepared BCs showing that in general no RBC lysis happened during the BC extraction. The median haemoglobin value was lower in the automatically extracted BCs (in two of the three cases). In conclusion, the automatically extracted BCs were of high quality presenting high leukocyte and platelets numbers and low red blood cell contamination compared to manually prepared BCs. In a second part of this experiment the DNA content of the automatically or manually prepared BCs was analysed from the same 3 donors (Table 4). The mean DNA yield of 204 g obtained from BC from 8 ml blood (very similar to yields shown in Table 2), was 1.5 times higher in the automatically extracted BCs compared to the manually extracted BCs, however the difference was not significantly different due to low numbers of donors. The ratios A260/A280 were high and very stable for both manually and automatically extracted BCs, thus indicating highly pure DNA samples. 2

Discussion We demonstrate the feasibility of the BC extraction automation on a Tecan Freedom Evo200 platform. Optimization of the script and validation of the method were based on BC cell content analysis (Blood cell counting and FACS) and on DNA extraction yields. Median DNA yield obtained with the optimized BC script (199 ug DNA from 8ml blood) is significantly higher than the one obtained with the first BC script, highlighting that script optimization was a critical factor for best DNA yield outcomes. The optimized script generated DNA extraction yields which were even significantly higher than the ones from manually prepared BCs (Table 2 and 4). In conclusion, automation of buffy coat extraction permits to isolate all leukocyte subpopulations and to extract high DNA quantities, it eliminates inter-operator variability and is time-saving, since up to 24 samples can be processed at the same time. Besides the Tecan system, Hamilton platforms have also been reported by the manufacturer as being capable of offering automation of buffy coat extraction. In our system, the automated BC extraction can be combined, on one hand, with upstream automated plasma aliquoting in the same run, and downstream automated DNA extraction from BCs on the other hand. Acknowledgements We are grateful to Hans-Peter Sattler and Matthias Greuter from Tecan for continuous assistance with the platform and Katy Beaumont, Estelle Henry and Audur Thorlaksdottir for excellent technical help. Figure legends Figure 1: Illustration of the Tecan Freedom Evo200 platform at IBBL. The platform is composed of a pick and place arm (PnP), a liquid handling arm (LiHa) with eight adapters for disposable tips (DiTi) and a robotic arm with centric gripper. A Tecan Tube Inspection Unit (TIU) is installed to detect blood separation levels. Tube barcodes are identified by a Datalogic Matrix200 scanner (scanner), 2D barcodes of individual destination tubes are identified by a FluidX flatbed barcode scanner. FluidX decappers (XSD-96Pro and XSD-48Pro for 96 respectively 48 individual tubes in racks), a laminar flow hood and a reagent-cooling unit complete the instruments of the platform. The system is operated by the software EVOware Standard V2.4. Figure 2: Typical appearance of a centrifuged blood collection tube after plasma withdrawal. The buffy coat (arrow) is located between the plasma and the dark red erythrocytes. Figure 3: Six blood collection tubes from 3 donors after withdrawal of equivalent volume of buffy coat by manual process (Man, donors 1-3) or automated process (Aut, donors 1-3). References 1 Herklotz R, Luthi U, Ottiger C, Huber AR. [Metaanalysis of reference values in hematology]. Therapeutische Umschau Revue therapeutique. 2006 Jan;63(1):5-24. 2 Peter HH. Labor und Diagnose: TH-Books Verlagsgesellschaft Frankfurt/Main (2000). 3

Population percentages Reference ranges (healthy adults) automated BC#1 automated BC#2 automated BC#3 Lymphocytes (% leukocytes) 20-44 % (1) 33.9 % 39.4 % 37.1 % Monocytes (% leukocytes) 2-9.5 % (1) 6.5 % 6.7 % 9.8 % Granulocytes (% leukocytes) 42-77 % (1) 59.6 % 53.9 % 53.1 % CD3 T cells (% lymphocytes) 69-93 % (2) 66.7 % 72.9 % 67.1 % CD4 T cells (% lymphocytes) 35-72 % (2) 29.7 % 55.3 % 41.8 % CD19 B cells (% lymphocytes) 3-13 % (2) 11.1 % 10.9 % 14.3 % Table 1: Flow cytometry analysis of leukocyte subpopulations present in automatically prepared buffy coats from 3 different healthy adult donors. Flow cytometry experiments were performed on a BD Influx cytometer using the following antibodies: anti-human CD3 APC-conjugated antibody (Immunotools, ref 21620036) for Pan T cells detection, anti-human CD4 PE-conjugated antibody (Immunotools, ref 21270044) for T cell subset detection, anti-human CD19 PE-conjugated antibody (Immunotools, ref. 21270194) for B cell detection, anti-human CD14 FITC-conjugated antibody (Immunotools, ref. 21279143) for monocyte detection, anti-human CD15 PE-conjugated antibody (Immunotools, ref. 21270154) for granulocyte detection, anti-human CD45 FITC-conjugated antibody (Immunotools, ref. 21270453) for total leukocyte detection. Isotype controls for IgG1, IgG2a and IgG2b were performed.

A DNA yield (µg) obtained from automatically extracted BC DNA yield (µg) obtained from manually extracted BC Buffy coat (BC) prepared from Median 25% 75% n Median 25% 75% n 4 ml blood (1 st script) 47.803 22.816 79.482 80 61.258 46.39 101.84 26 8 ml blood (1 st script) 147.891 92.204 200.02 156 145.242 111.51 184.2 53 8 ml blood (optimized script) 199.086 181.62 239.27 49 NP B p-value (Mann- Whitney test) significance at =0.05 Comparison Automatically extracted BC from 4ml blood versus manually extracted BC from 4ml blood p=0.029 significant Automatically extracted BC from 8ml blood versus manually extracted BC from 8ml blood p=0.967 not significant Automatically extracted BC from 8ml blood: optimized script versus previous script p<0.001 highly significant Table 2: A. 5 different populations were studied (n=80, n=26, n=156, n=53, n=49). DNA extractions were performed with an automated process based on magnetic bead extraction (Chemagic DNA Blood 4ml kit special) on a MSM I instrument (Chemagen). DNA eluates were quantified using spectrophotometry on a Synergy Mx monochromator instrument. B: DNA yield data were compared using a Mann-Whitney Rank Sum Test.NP: not performed.

donor 1 (male, 25 years) automated BC manual BC measured parameter WBC (10⁶) 60.00 61.32 102% 43.75 73% RBC (10⁶) 39760 8852 22% 11048 28% HGB (g) 1.168 0.260 22% 0.323 28% PLT (10⁶) 1576 1157 73% 736 47% donor 2 (male, 60 years) automated BC manual BC measured parameter WBC (10⁶) 40.00 38.71 97% 17.12 43% RBC (10⁶) 36400 6800 19% 9903 27% HGB (g) 1.096 0.211 19% 0.301 27% PLT (10⁶) 1928 1500 78% 588 30% donor 3 (female, 44 years) automated BC manual BC measured parameter WBC (10⁶) 41.60 35.02 84% 18.80 45% RBC (10⁶) 32160 6562 20% 6110 19% HGB (g) 1.000 0.204 20% 0.190 19% PLT (10⁶) 2288 1358 59% 705 31% measured parameter automated BC manual BC Comparison tot quantity recovery (%) vs whole blood tot quantity recovery (%) vs whole blood automated BC vs manual BC mean StDev mean StDev mean StDev WBC (10⁶) 47.20 11.11 45.02 14.24 95% 26.56 14.91 56% p=0.008 RBC (10⁶) 36107 3808 7405 1259 21% 9020 2584 25% NS HGB (g) 1.088 0.084 0.225 0.030 21% 0.271 0.071 25% NS PLT (10⁶) 1931 356 1339 172 69% 676 78 35% p=0.043 Table 3: Complete blood cell count data for, automated buffy coat (optimized script) and manual buffy coat prepared from the 3 same healthy donors (WBC: white blood cells, RBC: red blood cells, HGB: haemoglobin, PLT: platelets). Blood cell counting was performed on a Horiba ABX Micros CRP 200 instrument. Statistical t-test analysis on mean values of the three donors for WBC, RBC, HGB and PLT (NS=non significant).

donor 1 (male, 25 years) measured parameter automated BC manual BC Recovery: automated vs manual (%) DNA Yield (ug) 228.4 213.2 107% ratio A260/A280 1.925 1.93 100% donor 2 (male, 60 years) measured parameter automated BC manual BC Recovery: automated vs manual (%) DNA Yield (ug) 205.0 82.3 249% ratio A260/A280 1.917 1.968 97% donor 3 (female, 44 years) measured parameter automated BC manual BC Recovery: automated vs manual (%) DNA Yield (ug) 178.6 105.7 169% ratio A260/A280 1.931 1.947 99% measured parameter automated BC manual BC Comparison: automated vs manual mean StDev mean StDev Average DNA yield (ug) 204.0 24.9 133.7 69.8 NS Average A260/A280 ratio 1.924 0.007 1.948 0.019 NS Table 4: DNA extraction yields and ratio from individual donors and average values, statistical analysis t-test, NS=non significant