Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Similar documents
Chapters from Clinical Oncology

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Comparing Study between Conventional and Conformal Planning Radiotherapy

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

SUPERIORITY OF A REAL TIME PLANNING TECHNIQUE OVER IMAGE GUIDED RADIATION THERAPY FOR THE TREATMENT OF PRIMARY PROSTATE CANCERS

PRINCIPLES and PRACTICE of RADIATION ONCOLOGY. Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology

REVISITING ICRU VOLUME DEFINITIONS. Eduardo Rosenblatt Vienna, Austria

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

New Technologies for the Radiotherapy of Prostate Cancer

First, how does radiation work?

Defining Target Volumes and Organs at Risk: a common language

Role of Belly Board Device in the Age of Intensity Modulated Radiotherapy for Pelvic Irradiation

CBCT of the patient in the treatment position has gained wider applications for setup verification during radiotherapy.

TomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

Vaginal Sparing with Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer. Scott Boulet BSc, RT(T)

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

Radiotherapy physics & Equipments

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

The Physics of Oesophageal Cancer Radiotherapy

Managing the imaging dose during image-guided radiation therapy

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Knowledge-Based IMRT Treatment Planning for Prostate Cancer: Experience with 101. Cases from Duke Clinic. Deon Martina Dick

Grayden MacLennan Case Study 1 February 28, 2015 DCIS Treatment with Conformal Breast Tangents and Field in Field Dose Shaping History of Present

Trends in radiation oncology: a review for the nononcologist

Radiation Damage Comparison between Intensity Modulated Radiotherapy (IMRT) and Field-in-field (FIF) Technique In Breast Cancer Treatments

Subject: Image-Guided Radiation Therapy

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Clinical Implementation of a New Ultrasound Guidance System. Vikren Sarkar Bill Salter Martin Szegedi

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

Additional Questions for Review 2D & 3D

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Radiation Therapy for Metastatic Non-Small Cell Lung Carcinoma of the Right Hip

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Herlev radiation oncology team explains what MRI can bring

The technique of craniospinal irradiation of paediatric patients in supine position

Radiation therapy treatment plan optimization accounting for random and systematic patient setup uncertainties

ART for Cervical Cancer: Dosimetry and Technical Aspects

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Advances in Radiotherapy

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Left Chest Wall and Supraclavicular Irradiation Using Photon and Electron Energies

Dose-Volume Histogram Analysis in Point A-based Dose Prescription of High-dose-rate Brachytherapy for Cervical Carcinoma

IGRT Solution for the Living Patient and the Dynamic Treatment Problem

LA GESTIONE DELLE NUOVE TECNOLOGIE Cinzia Iotti. Azienda Arcispedale S. Maria Nuova IRCCS Reggio Emilia

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer

Treatment of exceptionally large prostate cancer patients with low-energy intensity-modulated photons

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

IMRT - Intensity Modulated Radiotherapy

Institute of Oncology & Radiobiology. Havana, Cuba. INOR

Automated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist

Statistical Analysis and Volumetric Dose for Organ at Risk of Prostate Cancer

TRANS-TASMAN RADIATION ONCOLOGY GROUP INC. Quality Assurance. Treatment Planning Benchmark

RTOG DOSIMETRY DATA SUBMISSION

Silvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

3D ANATOMY-BASED PLANNING OPTIMIZATION FOR HDR BRACHYTHERAPY OF CERVIX CANCER

Page 1. Helical (Spiral) Tomotherapy. UW Helical Tomotherapy Unit. Helical (Spiral) Tomotherapy. MVCT of an Anesthetized Dog with a Sinus Tumor

MR-Guided Brachytherapy

Computerized Treatment Planning in Radiation Therapy of Intact Breast: Influence of Number of CT-Cuts

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

From position verification and correction to adaptive RT Adaptive RT and dose accumulation

ENDPOINTS IN ONCOLOGY- HOW LONG WILL A CANCER PATIENT SURVIVE? DR GUNJAN BAIJAL CONSULTANT RADIATION ONCOLOGY MANIPAL GOA

Understanding Radiation Therapy. For Patients and the Public

Medical Errors in Radiation Therapy 2014

CALCULATION OF OPTIMAL MARGINS BETWEEN CLINICAL TARGET VOLUME (CTV) AND PLANNING TARGET VOLUME (PTV)

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Does conformal therapy improve dose distribution in comparison to old techniques in teleradiotherapy of cervical cancer patients?

3-Dimensional conformal radiotherapy versus intensity modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic analysis

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Implementation of advanced RT Techniques

Evaluation of Normal Tissue Complication Probability and Risk of Second Primary Cancer in Prostate Radiotherapy

Changing Paradigms in Radiotherapy

Clinical Quality Assurance for Particle Therapy Plus ça même Plus

Transcription:

Turkish Journal of Cancer Volume 37, No. 2, 2007 59 Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy MUHAMMAD BASIM KAKAKHAIL 1,2 1 Pakistan Institute of Engineering and Applied Sciences, Department of Physics and Applied Mathematics, Islamabad and 2 Karachi Institute of Radiotherapy and Nuclear Medicine, Department of Radiotherapy, Karachi-Pakistan ABSTRACT The aim of this study was to quantitatively compare treatment plans of high (15 MV) and low energy photon (6 MV) beams in pelvic sites (Bladder, Prostate, Cervix and Femur heads) for conventional radiation therapy. A total of 13 patients were studied. X-ray simulation was performed on Shimadzu treatment simulator. Patient contours were acquired using a conformator type contour plotter. Optimal treatment plans were prepared with the help of Decision Support System (DSS by Multidata). In all cases three field (3-F) technique was compared for high and low energy photons, except for cervix and femur heads where in addition to 3-F technique box technique was compared as well. The parameters analyzed for the effectiveness of a treatment plan were; Quality Index, Global maximum dose, Doses to Organs at Risk (OAR), dose uniformity with in the tumor and entrance doses. The entrance doses from anterior side for 15 MV photons in case of bladder and prostate were reduced by 20% and 19.2% respectively. Where as the dose to rectum (OAR) for 15 MV photons, was reduced by 300 cgy and 210 cgy for bladder and prostate patients respectively. A 26% reduction in entrance dose from the anterior aspect was observed for 15 MV photons in Cervix patients (3-F technique). The dose to bladder and rectum were reduced by 810 cgy and 255 cgy respectively for 15 MV photons in cervix (3-F). Dose to left femur head for 6 MV increased by 16% as compared to 15 MV photons. For right femur head the dose for 6 MV photons was 16.5% higher as compared to 15 MV photons. The major conclusion was that patient replanning was necessary in case of pelvic sites; otherwise very high doses were received by OAR. [Turk J Cancer 2007;37(2):59-65] KEY WORDS: High and low energy photons, pelvic sites, OAR, three field technique INTRODUCTION Radiation therapy has progressed significantly over the last few decades in the treatment of cancer. The arrival of sophisticated treatment modalities like IMRT, IGRT and Tomo-therapy has enhanced the effectiveness of Tele therapy in the cure of cancer. Nevertheless many centers hospitals around the world especially in developing countries are still using conventional radiation therapy. The aim of this study which was carried out at Radiation Therapy Department of Karachi Institute of Radiotherapy and Nuclear Medicine (KIRAN), was to assess the impact of patient transition from 15 MV to 6 MV in body sites where the use of high-energy beams has been preferred (like pelvic region). The important consideration in any such transition was to keep in mind

60 Comparison of High and Low Energy Treatment Plans the tumor control probability (TCP) and the dose to the organs at risk (OAR). Prior to site selection a literature survey was carried (1-4). MATERIALS AND METHODS A total of 25 patients were studied. Two types of patients were included in this study: 1) those whose body contours were already obtained for planning purposes and were maintained in the hospital record and, 2) those who arrived at KIRAN during the course of this study (two and half months). Patient s consent was not obtained for this study as only their data was used to compare the treatment plans and no change in the treatment was made for the sake of this study. During the course of study at every stage patient s privacy was respected fully. X-ray simulation was performed on Shimadzu simulator. The three main areas of interest that were addressed during the simulation process were: Alignment and positioning of the patient, tumor localization and field marking. During the simulation process the physicist was accompanied by the radiation oncologist, so as to discuss various issues regarding target localization, beam parameters and patient setup (5,6). Single slice body contours were obtained for each patient and the oncologist drew the target area and OAR. These marked contours were digitized. One better alternative to the above step was to use CT scans instead of manual contour taking; manual contours were not taken for those patients who had already undergone a CT scan. Vidar VXR-12 Plus film Digitizer was used to scan these CT slices to treatment planning system. Target localization with the help of CT scan was more accurate. But there were very few patients available with CT scans. Optimal treatment plans were prepared for 6 MV and 15 MV photon beams for each patient. The treatment planning software used was DSS (Decision Support System) by Multidata. This software closely resembled the properties of a 2.5D modern treatment planning system. Finally both the plans were reviewed by the radiation oncologist. RESULTS AND DISCUSSION High and low energy treatment plans comparison results have been provided, also a hypothetical situation was considered where patient originally planned for 15 MV has been mistakenly treated on 6 MV or directly shifted from high to low energy without re-planning, and this case was called the direct transition case (DT). Treatment plans have been provided for one patient from each pelvic site (because of restriction of space). In addition the mean data in the form of tables for each pelvic site has been furnished as well. Unless and otherwise mentioned, three field technique was used. Bladder Seven bladder patients were studied. The total dose prescribed to the tumor was 6000 cgy delivered in daily fraction of 200 cgy (30 fractions). In all cases 15 MV photon beams were superior to 6 MV in terms of quality index, global maximum doses, doses to the organs at risk Fig 1. Bladder 15 MV plan (OAR), target uniformity and the entrance doses (table 1). Comparison of entrance doses amongst 15 MV and 6 MV revealed that from the anterior field an additional 20% dose was delivered in case of 6 MV photons. Direct Fig 2. Bladder 6 MV plan transition (DT) resulted in 10% extra dose from the anterior field, and also delivered high doses from lateral aspects. However no such dose concentration was seen for 15 MV and 6 MV optimum plans (Figures 1 & 2). In case of DT the entrance dose from left lateral side was higher by 13.3% and 10.9% as compared to 15 MV and 6

Kakakhail et al. 61 Fig 3. Bladder DT plan MV respectively, where as from right lateral side in case of DT the entrance dose was up by 13.6% and 10.4% from 15 MV and 6 MV respectively (Figures 2 & 3). The above trends indicated that careful patient planning was required before making any transition from 15 MV to 6 MV incase of bladder, other wise there were chances of skin reaction from the lateral sides, normally above 3000 cgy skin reactions are very common. Mean quality index for tumor (Table 1) indicated that in case of 15 MV the area under DVH was more and target coverage was better, because quality index is the ratio of area under the DVH and the total area. Mean dose to rectum for 15 MV was less by 10 cgy that implied that in 30 fractions a total reduction of 300 cgy was made to bladder dose, which was clinically significant. The TD5/5 dose for bladder is 6000 cgy. For bladder and other pelvic sites to be discussed in the following sections, superiority of high-energy beams in various parameters of interest was because of the reason that lateral body separation in pelvic region was greater as compared to any other body site, which favored the use of higher energy beams. Prostate Three prostate patients were studied. 7000 cgy was delivered in 35 fractions (200 cgy per fraction). In all cases 15 MV photon beams were superior to 6 MV in terms of Fig 5. Prostate 6 MV plan quality index, global maximum doses, doses to the organs at risk (OAR), target uniformity and entrance doses (Table 2). The 50% isodose curve for 15 MV photon beam was closer to the tumor both from anterior and posterior ends Fig 6. Prostate DT plan as compared to 6 MV and DT cases (Figures 4, 5 & 6). Comparison of entrance doses between 15 MV and 6 MV showed that from the anterior field an additional 19.2% dose was delivered in case of 6 MV photons. DT resulted in an additional 6.8% dose from the anterior field. From left lateral side in case of direct transition the entrance dose was up by 13.2% and 9.7% as compared to 15 MV and 6 MV respectively. For right lateral side in case of direct transition the entrance dose was up by 13.3% and 10.2% as compared to 15 MV and 6 MV respectively. As high as 85% isodose curve was visible from the lateral sides (Figure 6), where no such dose concentration was visible for 15 MV and 6 MV (Figures 4 & 5). Mean dose to rectum for 15 MV beam was lower by 7 cgy as compared to 6 MV (Table 2). This suggested a reduction of 245 cgy in bladder dose during 35 fractions. Quality index (Table 2), global maximum doses, and doses to rectum and target uniformity followed similar trends as explained in section (7,8). Uterine cervix Fig 4. Prostate 15 MV plan Two uterine cervix cases were studied (high and low

62 Comparison of High and Low Energy Treatment Plans 30 fractions. Fig 7. Uterine cervix 15 MV 3-F plan energy for both 3-F and box techniques, Table 3). The prescribed dose was 6000 cgy with 200 cgy per fraction and a total of 30 fractions, the organs at risk were bladder and rectum and the TD5/5 for both these organs was 6000 cgy. For 3-F technique entrance doses from anterior side for 15 MV photons was reduced by 26% and 10% as For box technique entrance doses from anterior side was up by 8.25% and 5.75% for 6 MV and DT cases respectively as compared to 15 MV. From posterior side dose was up by 14.5% and 6.5% for 6 MV and DT cases respectively In case of DT the doses from right lateral side were up by 14.5% and 7.75% in comparison with 15 and 6 MV respectively. For left lateral side the entrance dose was up by 9.75% and 6.75% as compared to 15 MV and 6 MV respectively. Mean doses to rectum and bladder were lower by 10 cgy and 13 cgy respectively for 15 MV beams in case of 3-F technique, resulting in a reduction of 260 cgy and 200 cgy to rectum and bladder doses respectively in 30 fractions. Directly shifting the patients from 15 MV to 6 MV was not feasible as there were high doses from lateral aspect and the doses to femur head was quite significant. Fig 8. Uterine cervix 6 MV 3-F plan compared to 6 MV and DT cases respectively (Figures 7, 8 & 9). For DT the doses from right lateral sides were higher by 15.5% and 12% as compared to 15 MV and 6 MV respectively. For left lateral side the entrance dose was up by 9.75% and 6.75% as compared to 15 MV and 6 MV. Mean dose to rectum and bladder were reduced by 8.5 cgy and 27 cgy respectively for 15 MV beams in case of 3-F technique, resulting in a reduction of 255 cgy and 810 cgy to rectum and bladder doses respectively during Fig 10. Femur head 15 MV plan Data in table 3 also suggested that dose to bladder and rectum in box as well as 3-F techniques was quite reduced in case of 15 MV beams and especially in case of 3-F technique the dose to bladder was significantly reduced; therefore DT would result in unacceptable doses to bladder and there fore should be avoided. The quality index in case of high- energy beams was improved. (Table 3). Fig 8. Uterine cervix DT 3-F plan Fig 11. Femur head 6 MV plan

Kakakhail et al. 63 Comparison of doses on femur head While irradiating the pelvis, femur head doses are important; it is desired to keep these doses as low as possible. The doses for femur head were calculated on a single patient, where the femur heads were marked on the CT scan so as to determine their exact location. Doses on left and right femur heads for 15 MV photons were 42% & 43% of the prescribed dose respectively. For 6 MV left and right femur heads received 58% & 59.5% of the prescribed dose. This trend could be verified from the isodose distributions (Table 4, Figures 10, 11 & 12). Therefore before shifting a patient directly from 15 MV to 6 MV proper planning for femur head doses was required. Despite the fact that quality index for all cases were the, still the lateral doses were high and therefore re-planning was necessary. Fig 12. Femur head DT plan Table 1 Bladder patient mean data 15 MV 6 MV *DT Quality index 98 95 97 Global maximum dose (cgy) 200.6 213 208.8 Relative target uniformity (%) 5 10 7 Dose to rectum (cgy) 197 204 203 Anterior beam 48.5 68.6 58.7 Right lateral beam 57.3 56.6 67.3 Left lateral beam 53.3 55.7 66.7 Table 2 Prostate patient mean data 15 MV 6 MV *DT Quality index 97 95 97 Global maximum dose (cgy) 205.4 214.5 210.1 Relative target uniformity (%) 5 11 7.4 Dose to rectum (cgy) 75 82 71 Anterior beam 47.8 67 54.6 Right lateral beam 52.5 55.8 65.8 Left lateral beam 52.1 55.6 65.3

64 Comparison of High and Low Energy Treatment Plans Table 3 Cervix patient mean data Box Technique 15 MV 6 MV DT Quality index 99 99 99 Global maximum dose (cgy) 200.5 200.2 201.2 Relative target uniformity (%) 5 3 4 Dose to rectum (cgy) 93 103 92 Dose to bladder (cgy) 85 98 87 Anterior beam 26.7 37.5 32.2 Right lateral beam 54.5 62.2 69 Left lateral beam 49.5 54.5 60.7 Posterior beam 22.2 29.5 25.5 Three Field Technique Quality index 99 98 99 Global maximum dose (cgy) 200.4 203 200.7 Relative target uniformity (%) 5 6 4 Dose to rectum (cgy) 87.6 96 85 Dose to bladder (cgy) 97 124 101 Anterior beam 51.5 77.5 61.5 Right lateral beam 53 56.5 68.5 Left lateral beam 52 55 61.7 Table 4 Femur head patient mean data 15 MV 6 MV *DT Quality index 99 99 99 Global maximum dose (cgy) 200.4 206.4 204.6 Relative target uniformity (%) 5 4 3 Dose to rectum (cgy) 197 204 203 Dose to left femur head (cgy) 84 116 88 Dose to right femur head (cgy) 86 119 90 Anterior beam 42.5 40 32.2 Right lateral beam 34.5 60.5 45 Left lateral beam 34 59 44 Posterior beam 40 37 46.5

Kakakhail et al. 65 CONCLUSIONS The major conclusion of this study was that patient re-planning was necessary in case of bladder, cervix and prostate, because if the patient was not re-planned very high entrance doses from lateral aspects were received and therefore direct transition was not recommended. Similarly the doses received by organs at risk were reduced significantly for high-energy photons. Quantitatively 15 MV beams demonstrated better results for normal tissue sparing and the doses to the organs at risk. Pakistan being a developing nation is still in process of installing linear accelerators at various medical centers where already Co- 60 units are used for external beam therapy; therefore if high-energy beams are available they should be used in pelvic cancers. FUTURE RECOMMENDATIONS In the future the findings of this study should be verified from in vivo dosimetry, so as to compare the values of doses calculated on the planning software with that of the actual doses received by the patient. Similarly a second comparison could be made to compare these results with that obtained from true 3D planning software and to assess that how much extra dose was delivered by a 2D planning system. The difference between dose calculated and actual dose received by the patient would serve as an important parameter for the integrity of the whole radiation treatment planning process. References 1. Khan FM, Potish RA. Treatment Planning in Radiation Oncology. Williams & Wilkins, USA, 1998. 2. Bentel GC. Radiation Therapy Treatment Planning. 2nd Edition. McGraw-Hill Health Professions Division, 1996. 3. Chao KSC, Perez CA, Brady LW. Radiation Oncology: Management Decisions. Philadelphia: Lippincott Williams & Wilkins, 1999,2002. 4. Dobbs J, Barrett A, Ash D. Practical Radiotherapy Planning. 3rd edition. London, 1999. 5. ICRU Report 50, Prescribing, Recording, and Reporting Photon Beam Therapy. International commission on radiation units and measurements, Maryland 20614 U.S.A, 1993. 6. Khan FM. The Physics of Radiation Therapy. 3rd Edition, Philadelphia: Lippincott Williams & Wilkins, 2003. 7. Review of Radiation Oncology Physics: A Handbook for Teachers and Students. Podgorsak EB, editor. International Atomic Energy Agency, Austria 2003. 8. ICRU Report 62, Prescribing, Recording and Reporting Photon Beam Therapy (Supplement to ICRU Report 50). International commission on radiation units and measurements, Maryland 20614 U.S.A, 1999.