Sensitivity and Specificity in Detection of Labral Tears with 3.0-T MRI of the Shoulder

Similar documents
Usefulness of Unenhanced MRI and MR Arthrography of the Shoulder in Detection of Unstable Labral Tears

Prevalence of Meniscal Radial Tears of the Knee Revealed by MRI After Surgery

Evaluation of the Glenoid Labrum With 3-T MRI: Is Intraarticular Contrast Necessary?

MRI Shoulder - Slap tear

Modified Oblique Sagittal Magnetic Resonance Imaging of Rotator Cuff Tears: Comparison with Standard Oblique Sagittal Images

MR Arthrography in the Differential Diagnosis of Type II Superior Labral Anteroposterior Lesion and Sublabral Recess

Hugo C. van der Veen James P. M. Collins Paul C. Rijk

Meniscal Tears: Role of Axial MRI Alone and in Combination with Other Imaging Planes

MRI of the Shoulder What to look for and how to find it? Dr. Eric Handley Musculoskeletal Radiologist Cherry Creek Imaging

Articular cartilage and labral lesions of the glenohumeral joint: diagnostic performance of 3D water-excitation true FISP MR arthrography

Intramuscular Rotator Cuff Cysts: Association with Tendon Tears on MRI and Arthroscopy

Musculoskeletal Imaging Clinical Observations

Rotator Cable: MRI Study of Its Appearance in the Intact Rotator Cuff With Anatomic and Histologic Correlation

Distinguishing Superior Labral Tears from Normal Meniscoid Insertions with Magnetic Resonance Imaging

Department of Orthopaedic Surgery University of California, Davis

MRI Findings of 26 Patients with

Sensitivity of MR Arthrography in the Evaluation of Acetabular Labral Tears

Added value of MR myelography using 3D COSMIC sequence in the diagnosis of lumbar canal stenosis: comparison with routine MR imaging

Delaminated Tears of the Rotator Cuff: Prevalence, Characteristics, and Diagnostic Accuracy Using Indirect MR Arthrography

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

Sports Medicine: Shoulder Arthrography. Christine B. Chung, M.D. Professor of Radiology Musculoskeletal Division UCSD and VA Healthcare System

Lesions of the Menisci of the Knee: Value of MR Imaging Criteria for Recognition of Unstable Lesions

Musculoskeletal Imaging Original Research

Diagnosis of SLAP lesions with Grashey-view arthrography

MRI SHOULDER WHAT TO SEE

FieldStrength. Achieva 3.0T enables cutting-edge applications, best-in-class MSK images

MRI of Bucket-Handle Te a rs of the Meniscus of the Knee 1

Superior Labral Anteroposterior Lesions of the Shoulder: Part 1, Anatomy and Anatomic Variants

Musculoskeletal Imaging What to order? Brian Cole, MD

Comparative Study of MRI and Operative Findings in Rotator Cuff Tears. University of Medical Sciences, Iran. Medical Sciences, Iran.

COPYRIGHT 2005 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED

Introduction & Question 1

Original Report. Sonography of Tears of the Distal Biceps Tendon. Theodore T. Miller 1,2 Ronald S. Adler 3

Lawrence Gulotta Gillian Lieberman, MD October Gillian Lieberman, MD. Shoulder Imaging. Lawrence V. Gulotta, HMS IV 10/16/02

Viviane Khoury, MD. Assistant Professor Department of Radiology University of Pennsylvania

The magic angle phenomenon in tendons: effect of varying the MR echo time

MR measurements of subcoracoid impingement using a new method and its relationship to rotator cuff pathology at MR arthrography

Arthroscopy / MRI Correlation Conference. Department of Radiology, Section of MSK Imaging Department of Orthopedic Surgery 7/19/16

The Low Sensitivity of Fluid-Attenuated Inversion-Recovery MR in the Detection of Multiple Sclerosis of the Spinal Cord

Comparison of Three-dimensional Isotropic T1-weighted Fast Spin-Echo MR Arthrography with Two-dimensional MR Arthrography of the Shoulder 1

Generated by Foxit PDF Creator Foxit Software For evaluation only. Tishreen University Journal. Med.

This presentation is the intellectual property of the author. Contact them for permission to reprint and/or distribute.

Fosbury Flop Tear of the Rotator Cuff: Diagnostic Assessment with Magnetic Resonance Arthrography. Running Title: Fosbury Flop Tears

This presentation is the intellectual property of the author. Contact them at for permission to reprint and/or distribute.

Demystifying ABER (ABduction and External Rotation) sequence in shoulder MR arthrography

imri 2018;22:

The mandibular condyle fracture is a common mandibular

Elbow Effusions in Trauma in Adults and Children: Is There an Occult Fracture?

dgemric Effectively Predicts Cartilage Damage Associated with Femoroacetabular Impingement

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

Direct MR Arthrography of the Shoulder at 3 Tesla: Optimization of Gadolinium Concentration

Validation of MRI Classification System for Tibial Stress Injuries

Musculoskeletal MR Protocols

CAN SOFT TISSUES STRUCTURES DIFFERENTIATE BETWEEN DYSPLASIA AND CAM-FAI OF THE HIP?

Musculoskeletal Imaging at 3T with Simultaneous Use of Multipurpose Loop Coils

Comparative study of imaging at 3.0 T versus 1.5 T of the knee

Variant origin of the bifurcated long head of biceps tendon - the case for supraspinatus.

Musculoskeletal Imaging Clinical Perspective

Original Report. The Reverse Segond Fracture: Association with a Tear of the Posterior Cruciate Ligament and Medial Meniscus

Shoulder MR arthrography

Meniscus T2 Relaxation Time at Various Stages of Knee Joint Degeneration

The Meniscal Roots: Gross Anatomic Correlation with 3-T MRI Findings

MR Arthrography of the Labral Capsular Ligamentous Complex in the Shoulder: Imaging Variations and Pitfalls

Orthopedic Hardware Imaging Part II: MRI v. Metal

FILED: QUEENS COUNTY CLERK 12/15/ :11 PM INDEX NO /2015 NYSCEF DOC. NO. 54 RECEIVED NYSCEF: 12/15/2017

ACR MRI Accreditation: Medical Physicist Role in the Application Process

JMSCR Vol 05 Issue 01 Page January

Comparison of radiation doses of various approaches of MR arthrograms with fluoroscopic guided contrast injection

Magnetic resonance imaging of the shoulder. Sensitivity, specificity, and predictive value

Accuracy of SPECT bone scintigraphy in diagnosis of meniscal tears ABSTRACT

The Temporomandibular joint: Anatomy, Mechanics, Pathology. Aditya Bahel, DO

MR Arthrography of Shoulder

4 MR and MR Arthrography

Musculoskeletal Imaging Original Research

ORIGINAL ARTICLE. Anastasia Fotiadou & Antonios Drevelegas & Michelangelo Nasuto & Giuseppe Guglielmi

Department of Orthopaedic Surgery, Tampere University Hospital, Tampere, Finland 3

Comparisons of the Various Partial- Thickness Rotator Cuff Tears on MR Arthrography and Arthroscopic Correlation

Anterior shoulder instability: Evaluation using MR arthrography.

Role of Magnetic Resonance Imaging in Internal Derangement of Shoulder

Injury, Int. J. Care Injured 44 S3 (2013) S26 S32. Contents lists available at SciVerse ScienceDirect. Injury

Infraclavicular brachial plexus blocks have been designed

醫用磁振學 MRM 肌肉骨骼磁振造影簡介 肌肉骨骼磁振造影. 本週課程內容 General Technical Considerations 肌肉骨骼磁振造影簡介 盧家鋒助理教授國立陽明大學生物醫學影像暨放射科學系

High b Value Diffusion-Weighted MRI for Detecting Pancreatic Adenocarcinoma: Preliminary Results

Meniscal tears on 3T MR: Patterns, pearls and pitfalls

Index. Note: Page numbers of article titles are in boldface type.

MRI of Osteochondral Defects of the Lateral Femoral Condyle: Incidence and Pattern of Injury After Transient Lateral Dislocation of the Patella

PROPELLER Technique to Improve Image Quality of MRI of the Shoulder

AMSER Case of the Month January 2019

HAGL lesion of the shoulder

Variant origin of the bifurcated long head of biceps tendon - the case for supraspinatus.

Superior Labrum Anterior Posterior lesions: ultrasound evaluation

MRI Appearance of Posterior Cruciate Ligament Tears

Normal Anatomy and Strains of the Deep Musculotendinous Junction of the Proximal Rectus Femoris: MRI Features

Sonographic Findings of Adductor Insertion Avulsion Syndrome With Magnetic Resonance Imaging Correlation

Case Report Bilateral Congenital Agenesis of the Long Head of the Biceps Tendon: The Beginning

Clinical Correlation of a New Practical MRI Method for Assessing Cervical Spinal Canal Compression

Magnetic resonance imaging evaluation of lateral ankle ligaments and peroneal tendons in a group of asymptomatic patients

Hill-Sachs Lesion on MR Arthrography of the Shoulder: Relationship with Bankart Lesion on Arthroscopy and Frequency of Shoulder Dislocations

FAI syndrome with or without labral tear.

MR Arthrography of the Shoulder - A Beginner's Guide

Transcription:

Magee and Williams MRI for Detection of Labral Tears Musculoskeletal Imaging Clinical Observations C M E D E N T U R I C L I M G I N G JR 2006; 187:1448 1452 0361 803X/06/1876 1448 merican Roentgen Ray Society Y O Thomas H. Magee 1 David Williams Magee TH, Williams D Keywords: labral tears, MRI, shoulder DOI:10.2214/JR.05.0338 F Received February 27, 2005; accepted after revision June 13, 2005. 1 Both authors: Department of Radiology, Neuroskeletal Imaging, 255 N Sykes Creek Pkwy., Merritt Island, FL 32953. ddress correspondence to T. H. Magee (tmageerad@cfl.rr.com). Sensitivity and Specificity in Detection of Labral Tears with 3.0-T MRI of the Shoulder OBJECTIVE. MRI of the shoulder has been found to be sensitive and specific for detection of labral tears at 1.5 T or lower field strength compared with arthroscopy, whereas 3.0-T MRI of the shoulder has not been specifically assessed. This study assesses the sensitivity and specificity of MRI at 3.0 T for labral tears compared with arthroscopy. CONCLUSION. MRI of the shoulder at 3.0 T is very sensitive and specific compared with arthroscopy in detection of superior, anterior, and posterior labral tears. RI of the shoulder has been found M to be accurate in the diagnosis of labral tears. In previous studies, conventional MR sensitivity in detection of labral tears has ranged from 44% to 93% sensitivity compared with arthroscopy [1, 2]. Two recent studies have assessed conventional MRI evaluation of the glenoid labrum using a 0.2-T extremity MR system. Shellock et al. [3] found a sensitivity of 89% and a specificity of 100% for detection of labral tears compared with arthroscopy on a 0.2-T MR system. Zlatkin et al. [4] found a sensitivity of 55% and a specificity of 100% for detection of labral tears compared with arthroscopy on a 0.2-T MR system. Shoulder MR arthrography has been reported to be more sensitive and specific for detection of labral tears compared with conventional shoulder MRI in several previous studies performed at 1.5-T or lower field strength. pplegate et al. [5] found shoulder MR arthrography to be highly accurate for assessing the glenoid labrum in patients with chronic labral tears. Magee et al. [6] found shoulder MR arthrography to be more sensitive for detection of labral tears than conventional shoulder MRI. Palmer and Caslowitz [7] showed a 92% sensitivity and specificity for MR arthrograms in detection of labral lesions. Chandnani et al. [1] found shoulder MR arthrography to be more accurate than conventional shoulder MR examination in detection of labral lesions. Jee et al. [8] found MR arthrography to be highly accurate in detection of superior labral anteroposterior (SLP) tears. None of these studies was performed on a 3.0-T MR scanner. To our knowledge, the sensitivity and specificity of 3.0-T MRI of the shoulder for detection of labral tears compared with arthroscopy has not been specifically assessed. We undertook a retrospective review of 100 shoulder MRI examinations to determine the sensitivity and specificity of shoulder imaging at 3.0 T for superior, anterior, and posterior labral tears at 3.0 T. Materials and Methods Two experienced musculoskeletal radiologists retrospectively reviewed 3.0-T MR scans of the shoulder in 100 consecutive patients. MRI examinations were assessed for the presence or absence of superior, anterior, or posterior labral tears. Other findings (such as supraspinatus tendon tears) were not included in this retrospective review. Interpretations were performed by consensus review. ny MR arthrography performed in addition to conventional MRI was not included in the retrospective consensus review. MRI examinations were performed between October and December 2004. The age range of the 100 patients was 14 71 years (mean age, 42 years). The 100 patients all presented with shoulder pain, instability, or both. There was no selection bias for placing patients on the 3.0-T MR scanner because our practice consists exclusively of 3.0-T MR scanners. ll patients underwent MRI of the shoulder in oblique coronal, oblique sagittal, and axial planes on a 3.0-T Signa scanner (GE Healthcare). Oblique coronal and sagittal fast spin-echo T1- weighted (TR/TE, 550/10; number of excitations [NEX], 2), oblique coronal and sagittal fast spinecho intermediate-weighted (3,850/55; NEX, 4), and fast spin-echo intermediate-weighted axial (3,250/55; NEX, 3) sequences with a field of view 1448 JR:187, December 2006

MRI for Detection of Labral Tears Fig. 1 47-year-old man with shoulder pain. Coronal fast spin-echo intermediateweighted (TR/TE, 3,850/55) MR image shows findings consistent with superior labral anteroposterior (SLP) tear (arrow). Patient had surgically proven SLP tear. Fig. 2 38-year-old man with shoulder instability. xial fast spin-echo intermediateweighted (TR/TE, 3,250/55) MR image shows findings consistent with anterior labral tear (arrow). Patient had surgically proven anterior labral tear. Fig. 3 35-year-old man with shoulder pain and instability. xial fast spin-echo intermediate-weighted (TR/TE, 3,250/55) MR image shows findings consistent with posterior labral tear (arrow). Patient had surgically proven posterior labral tear. Fig. 4 28-year-old man with shoulder instability. xial fast spin-echo intermediateweighted (TR/TE, 3,250/55) MR image shows findings consistent with anterior labral tear (arrow). Patient had surgically proven anterior labral tear. JR:187, December 2006 1449

Magee and Williams Fig. 5 37-year-old woman with shoulder pain. Coronal fast spinecho intermediateweighted (TR/TE, 3,850/55) MR image shows findings consistent with superior labral anteroposterior (SLP) tear (arrow). Patient had surgically proven SLP tear. of 14 cm on all images were used. Slice thickness was 4 mm with a 10% interslice gap on all sequences except the fast spin-echo intermediateweighted axial sequence, which had a 3-mm slice thickness. The echo-train length was 10 on all T2- weighted and proton-density sequences and 3 on the T1-weighted sequences. The bandwidth was 31.25 khz on all sequences. The imaging time for the coronal and sagittal T2-weighted sequences was 4 minutes 43 seconds. The imaging time for the axial intermediate-weighted sequences was 3 minutes 26 seconds, whereas the imaging time for the T1-weighted sequences was 2 minutes 28 seconds. The matrix for all intermediate-weighted images was 320 320, and the matrix on all T1- weighted images was 320 256. three-channel phased-array shoulder coil was used. ll 100 patients had their MRI examinations retrospectively interpreted by consensus of two reviewers. The reviewers were blinded to the results of arthroscopy at the time of consensus review. Consensus was achieved when both reviewers agreed that a superior, anterior, posterior, or combined labral tear was present or not present on an MRI examination. Retrospective MR interpretations were then correlated with results in those patients who underwent arthroscopy (n = 67). The surgeons were aware of prospective MR interpretations before the patients underwent arthroscopy. The MR criterion used for diagnosis of a labral tear was an abnormality of the glenoid labrum morphology, signal intensity, or both. ll arthroscopies were performed within 4 weeks of the MRI examination. Fig. 6 19-year-old man with shoulder pain and instability. Patient had surgically proven anterior labral tear., xial fast spin-echo intermediate-weighted (TR/TE, 3,250/55) MR image shows findings that could not be definitively described as anterior labral tear (arrow) by either MR reviewer. B, xial fat-saturated T1-weighted MR arthrogram (550/10) shows findings consistent with displaced anterior labral tear (arrow). This was confirmed at arthroscopy. MR arthrography images were not included in consensus retrospective MR review. Displaced labral tear could be definitively seen only at MR arthrography in this case. B 1450 JR:187, December 2006

MRI for Detection of Labral Tears Fig. 7 24-year-old man with shoulder pain and instability. xial fast spinecho intermediateweighted (TR/TE, 3,250/55) MR image shows findings that could not be definitively described as anterior labral tear (arrow) by either MR reviewer. Patient had surgically proven anterior labral tear. Results Of the 67 patients who went on to arthroscopy, 42 patients had a total of 46 labral tears. Twenty-one patients had SLP tears. Eighteen patients had anterior labral tears, and seven patients had posterior labral tears. Three patients had both a SLP tear and an anterior labral tear, and one patient had a SLP tear and a posterior labral tear. Nineteen of the 21 SLP tears seen at arthroscopy were seen on retrospective consensus MRI review. Sixteen of the 18 anterior labral tears and six of the seven posterior labral tears seen at arthroscopy were seen on retrospective consensus MRI review. In this study, MRI sensitivity for detection of SLP tears was 90% (19 of 21 full-thickness supraspinatus tendon tears seen at arthroscopy were seen on MRI) and specificity was 100%. MRI sensitivity for detection of anterior labral tears was 89% (16 of 18 anterior labral tears seen at arthroscopy were seen on MRI) and specificity was 100%. MRI sensitivity for detection of posterior labral tears was 86% (six of seven posterior labral tears seen at arthroscopy were seen on MRI) and specificity was 100% (Figs. 1 8). Fig. 8 42-year-old woman with shoulder pain. Patient had surgically proven superior labral anteroposterior (SLP) tear., xial fast spin-echo intermediate-weighted (TR/TE, 3,250/55) MR image shows findings that could not be definitively described as anterior labral tear (arrow) by either MR reviewer. Even in retrospect after arthroscopy results were known, this had appearance of sublabral foramen to both MR reviewers. B, Coronal fast spin-echo intermediate-weighted (3,850/55) MR image shows findings both MR reviewers considered to be appearance of sublabral foramen (arrow). B JR:187, December 2006 1451

Magee and Williams Discussion MRI of the shoulder at 3.0 T is highly sensitive and specific for the detection of labral tears compared with arthroscopy. Previous studies have shown MRI at 1.5-T field strength or less to be sensitive for detection of labral tears [1, 2]. To our knowledge, shoulder imaging at 3.0 T has not been specifically assessed. In this study, 3.0-T MRI without intraarticular contrast material was highly sensitive and specific in the detection of SLP, anterior labral, and posterior labral tears with sensitivities at or above those achieved with 1.5-T MRI. In addition, there were no false-positive readings of labral tears at MRI that were not seen at arthroscopy. The consensus interpretations were highly specific. Two SLP tears, two anterior labral tears, and one posterior labral tear were not seen on prospective MRI consensus review but were seen at arthroscopy. In one case at arthroscopy, a SLP tear was described and an anchor was placed. On consensus MRI review, this was thought to represent a sublabral foramen. Even when arthroscopy results were known and MR images were retrospectively reviewed, it was thought this represented a sublabral foramen (Fig. 8). One surgically proven displaced anterior labral tear was not seen on retrospective consensus MRI review of conventional shoulder MR images. In this case the patient also had an MR arthrogram. The displaced anterior labral tear was seen on the MR arthrogram but not on conventional MR images. The MR arthrography images were not included in retrospective MRI consensus review (Fig. 6). The other SLP tear, anterior labral tear, and posterior labral tear were not seen on prospective MRI consensus review but were seen at arthroscopy. These three tears were all débrided by the orthopedic surgeon. The three tears were not seen even on retrospective review after arthroscopy results were known. It is possible these were nondisplaced tears that may have been visible if MR arthrography was performed. However, in each of the three cases, the orthopedic surgeons indicated that surgery was performed for reasons other than a labral tear. prospective MRI review of a labral tear in these cases would not have affected the decision to perform arthroscopy. MRI at 3.0 T allows a higher signal-tonoise ratio (SNR) compared with 1.5-T MRI. The spin-spin relaxation time, T2, remains fairly constant at different field strengths. However, the spin-lattice relaxation time, T1, increases as the field strength increases. Therefore, at 3.0 T, the TR must be longer than on 1.5-T MR scanners to maximize the SNR gain. t 3.0 T, the TR must be longer to attain the same type of contrast on T1-weighted images as seen on a 1.5-T scanner. lso on 3.0-T MR scanners, the TE must be slightly shorter than on 1.5-T MR scanners to account for decreased T2 relaxation time [9]. The parameters used on our shoulder imaging have a slightly increased TR and a slightly decreased TE on all sequences compared with parameters previously used on our 1.5-T MR scanner. This was done to optimize our SNR on the 3.0-T MR scanner. Limitations of this study include its retrospective nature and the fact that MR images were interpreted by consensus rather than independently. Most of our referrals came from orthopedic surgeons. These patients have a high prevalence of positive findings on MRI. The surgeons were aware of prospective MRI interpretations before arthroscopy. This may have biased their arthroscopy results. In conclusion, MRI of the shoulder at 3.0 T is highly sensitive, specific, and accurate in the diagnosis of labral tears compared with arthroscopy. References 1. Chandnani VP, Yeager TD, DeBerardino T, et al. Glenoid labral tears: prospective evaluation with MR imaging, MR arthrography, and CT arthrography. JR 1993; 161:1229 1235 2. Gusmer PJ, Porter HG, Schatz J, et al. Labral injuries: accuracy of detection with unenhanced MR imaging of the shoulder. Radiology 1996; 200:519 524 3. Shellock FG, Bert JM, Fritts HM, et al. Evaluation of the rotator cuff and glenoid labrum using a 0.2- Tesla extremity magnetic resonance (MR) system: MR results compared to surgical findings. J Magn Reson Imaging 2001; 14:763 770 4. Zlatkin MB, Hoffman C, Shellock FG. ssessment of the rotator cuff and glenoid labrum using an extremity MR system: MR results compared to surgical findings from a multi-center study. J Magn Reson Imaging 2004; 19:623 631 5. pplegate GR, Hewitt M, Snyder SJ, et al. Chronic labral tears: value of magnetic resonance arthrography in evaluating the glenoid labrum and labral bicipital complex. rthroscopy 2004; 20:959 963 6. Magee T, Williams D, Mani N. Shoulder MR arthrography: which patient group benefits most? JR 2004; 183: 969 974 7. Palmer WE, Caslowitz PL. nterior shoulder instability: diagnostic criteria determined from prospective analysis of 121 MR arthrograms. Radiology 1995; 197: 819 825 8. Jee WH, McCauley TR, Katz LD, et al. Superior labral anterior posterior (SLP) lesions of the glenoid labrum: reliability and accuracy of MR arthrography for diagnosis. Radiology 2001; 218:127 132 9. Gold GE, Han E, Stainsby J, et al. Musculoskeletal MRI at 3.0 T: relaxation times and image contrast. JR 2004; 183:343 351 1452 JR:187, December 2006