The early results of joint-sparing proximal tibial replacement for primary bone tumours, using extracortical plate fixation

Similar documents
Intercalary Femur and Tibia Segmental Allografts Provide an Acceptable Alternative in Reconstructing Tumor Resections

Functional Outcome Study of Mega-Endoprosthetic Reconstruction in Limbs With Bone Tumour Surgery

Clinical Study The Use of Massive Endoprostheses for the Treatment of Bone Metastases

SYMPOSIUM: 2015 MEETINGS OF THE MUSCULOSKELETAL TUMOR SOCIETY AND THE

Below-knee amputation through a jointsparing proximal tibial replacement for recurrent tumour

Laura M. Fayad, MD. Associate Professor of Radiology, Orthopaedic Surgery & Oncology The Johns Hopkins University

North of England Bone and Soft Tissue Tumour Service

Giant Cell Tumour of the Distal Radius: Wide Resection and Reconstruction by Non-vascularised Proximal Fibular Autograft

Methods Used for Reconstruction in Aggressive Bone Tumours: An Early Experience

Custom endoprosthetic reconstruction for malignant bone disease in the humeral diaphysis

Total Knee Original System Primary Surgical Technique

Objectives. Limb salvage surgery. Age distribution bone cancer. Age distribution soft tissue sarcomas

Giant cell tumour of the proximal femur

Growth in the lower lim b following chem otherapy for a malignant prim ary bone tumour: a straight-line graph

Clinical Study Distal Femur Allograft Prosthetic Composite Reconstruction for Short Proximal Femur Segments following Tumor Resection

BONE TRANSPLANTATION IN LIMB SAVING SURGERIES: THE PHILIPPINE EXPERIENCE

Neurologic Damage. The most common neurologic injury following intramedullary tibial nailing is injury to the peroneal nerve.

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

CASE REPORT. Bone transport utilizing the PRECICE Intramedullary Nail for an infected nonunion in the distal femur

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

CASE REPORT. Antegrade tibia lengthening with the PRECICE Limb Lengthening technology

Metastatic Disease of the Proximal Femur

Biological Reconstruction after Excision of Juxta-articular Osteosarcoma around the Knee: A New Classification System

PROXIMAL TIBIAL PLATE

Arthrex Open Wedge Osteotomy Technique Designed in conjunction with:

Knee Reconstruction With Prosthesis and Muscle Flap After Total Arthrectomy

Complications of limb salvage surgery in childhood tumors and recommended solutions

Title: An intramedullary free vascularized fibular graft combined with pasteurized

Vascularised free fibular flap in bone resection and reconstruction *

Tibial Replacements. Orthopaedic Salvage System

Classification of failure of limb salvage after reconstructive surgery for bone tumours

Limb Salvage Surgery Using Whole Knee Joint Allograft Reconstruction in Osteosarcoma

JMSCR Vol 06 Issue 12 Page December 2018

TRK REVISION KNEE Surgical Technique

Chondroblastoma of bone

Effects of metaphyseal bone tumor removal with preservation of the epiphysis and knee arthroplasty

U2 PSA. Revision Knee. Surgical Protocol

Over a 25-year period we have treated 36 patients

Transfemoral Amputation

The other bone sarcomas

Clinical Study Comparison between Constrained and Semiconstrained Knee Allograft-Prosthesis Composite Reconstructions

.org. Tibia (Shinbone) Shaft Fractures. Anatomy. Types of Tibial Shaft Fractures

The results of total humeral replacement following excision for primary bone tumour

Periosteal osteosarcoma: a review of 17 cases with mean follow-up of 52 months

Lateral tibial condyle reconstruction by pedicled vascularized fibular head graft: Long-term result

LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

ARMS. Reconstruction of Large Femur and Tibia Defect with Free Vascularized Fibula Graft and Locking Plate INTRODUCTION.

Knee Revision. Portfolio

Surgical procedure. METS Modular Proximal Femur

Cementless Oxford unicompartmental knee replacement shows reduced radiolucency at one year

Fractures of the tibia shaft treated with locked intramedullary nail Retrospective clinical and radiographic assesment

SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY

PediLoc Extension Osteotomy Plate (PLEO)

Fractures of allografts used in limb preserving operations

Medical Practice for Sports Injuries and Disorders of the Knee

Intramedullary Tibial Preparation

TOTAL KNEE ARTHROPLASTY (TKA)

ACL Athletic Career. ACL Rupture - Warning Features Intensive pain Immediate swelling Locking Feel a Pop Dead leg Cannot continue to play

Microwave ablation of malignant extremity bone tumors

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique IMAGE TO COME

Wichita Fusion Nail Surgical Technique. David A. McQueen, MD

DePuy LPS Limb Preservation System

Author's response to reviews

LPS SYSTEM POCKET GUIDE

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

What Is the Outcome of Allograft and Intramedullary Free Fibula (Capanna Technique) in Pediatric and Adolescent Patients With Bone Tumors?

Reconstruction of Massive Femur Defect with Free Vascularized Fibula Graft Following Tumor Resection

Fibula bone grafting in infected gap non union: A prospective case series

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Result of extracorporeal irradiation and re-implantation for malignant bone tumors: A review of 30 patients

Wrist Joint Reconstruction With a Vascularized Fibula Free Flap Following Giant Cell Tumor Excision in the Distal Radius

2017 Resident Advanced Trauma Techniques Course COMPLICATIONS / CHALLENGES MALUNIONS/DEFORMITY

Mr Aslam Mohammed FRCS, FRCS (Orth) Consultant Orthopaedic Surgeon Specialising in Lower Limb Arthroplasty and Sports Injury

General Concepts. Growth Around the Knee. Topics. Evaluation

Ethan M. Braunstein, M.D. 1, Steven A. Goldstein, Ph.D. 2, Janet Ku, M.S. 2, Patrick Smith, M.D. 2, and Larry S. Matthews, M.D. 2

This publication is not intended for distribution in the USA.

Robert Botte, DVM, Diplomate ACVS Veterinary Surgical Service San Diego, California. Kyon Symposium 2010 Zurich

EVOS MINI with IM Nailing

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

We studied the CT and MR scans, and the

North of England Bone and Soft Tissue Tumour Service

Surgical procedure. METS Modular Proximal Tibia

Periarticular knee osteotomy

Femoral reconstruction by single, folded or double free vascularised fibular grafts

Knee Surgical Technique

The Birmingham Hook Plate Treatment Of Irreduceable Displaced Mallet Avulsion Fractures: A Technical Note

Total Hip Replacement in Diaphyseal Aclasis: A Case Report

Femur. Monoaxial Locking Plate System. Operative Technique. Distal Lateral Femur Universal Holes Targeting Instrumentation.

Bipolar Radial Head System

Total Knee Replacement

Physeal Fractures and Growth Arrest

The ipsilateral and contralateral fibulae have been

The gastrocnemius with soleus bi-muscle flap

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Resurfacing Distal Femur. Orthopaedic Salvage System

Longitudinal growth following treatm ent for osteosarcom a

4/28/2010. Fractures. Normal Bone and Normal Ossification Bone Terms. Epiphysis Epiphyseal Plate (physis) Metaphysis

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System.

QuadsTape System TM. For Quadriceps Tendon Reconstruction. Surgical Technique Manual

Transcription:

The early results of joint-sparing proximal tibial replacement for primary bone tumours, using extracortical plate fixation B. G. I. Spiegelberg, M. D. Sewell, W. J. S. Aston, G. W. Blunn, R. Pollock, S. R. Cannon, T. W. R. Briggs From The Royal National Orthopaedic Hospital, Middlesex, England This paper describes the preliminary results of a proximal tibial endoprosthesis which spares the knee joint and enables retention of the natural articulation by replacing part of the tibial metaphysis and diaphysis. In eight patients who had a primary malignant bone tumour of the proximal tibia, the distal stem, which had a hydroxyapatite-coated collar to improve fixation, was cemented into the medullary canal. The proximal end had hydroxyapatite-coated extracortical plates which were secured to the remaining proximal tibial metaphysis using cortical screws. The mean age of the patients at operation was 28.9 years (8 to 43) and the mean follow-up was for 35 months (4 to 48). The mean Musculoskeletal Tumour Society score was 79% (57% to 90%), the mean Oxford Knee score was 40 points of 48 (36 to 46) and the mean knee flexion was 112 (100 to 120 ). In one patient, revision to a below-knee amputation through the prosthesis was required because of recurrence of the tumour. Another patient sustained a periprosthetic fracture which healed with a painful malunion. This was revised to a further endoprosthesis which replaced the knee. In the remaining six patients the prosthesis allowed preservation of the knee joint with good function and no early evidence of loosening. Further follow-up is required to assess the longevity of these prostheses. B. G. I. Spiegelberg, MRCS, MBBS, BSc, Research Fellow M. D. Sewell, MRCS, MBBS, Specialist Registrar W. J. S. Aston, FRCS(Orth), MBBS, BSc, Specialist Registrar G. W. Blunn, PhD, Professor R. Pollock, FRCS(Orth), Consultant Orthopaedic S. R. Cannon, MCh(Orth), FRCS, Consultant Orthopaedic T. W. R. Briggs, MCh(Orth), FRCS, Consultant Orthopaedic The Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex HA7 4LP, UK. Correspondence should be sent to Mr B. G. I. Spiegelberg; e-mail: benspiegelberg@doctors.net.uk 2009 British Editorial Society of Bone and Joint Surgery doi:10.1302/0301-620x.91b10. 22076 $2.00 J Bone Joint Surg [Br] 2009;91-B:1373-7. Received 18 November 2008; Accepted after revision 8 June 2009 The proximal tibia is the second most common site for primary bone tumours after the distal femur. 1 Before 1977, the main treatment for such tumours was above-knee amputation. However, subsequent improvements in adjuvant therapy with improved cure rates, 2 and simultaneous advances in surgical technique and biomedical design have made limb salvage a viable option for these patients. 3 Limb salvage using a cemented proximal tibial endoprosthesis to replace both the affected portion of the tibia and knee joint, is an accepted method of filling the defect created by resection of the primary bone tumour. Initial results for this technique had a revision rate as high as 70% at ten years. 4 Those poor results were mainly due to infection, which was related to inadequacies of soft-tissue coverage. A change in practice to include mobilisation and coverage of the prosthesis with a gastrocnemius rotation flap reduced the incidence of infection from 36% to 12%. 4,5 The second problem was that proximal tibial replacements were prone to aseptic loosening, with one study reporting an incidence of almost 40% at 120 months; the authors found that implant survival was much poorer when > 60% of the tibia was resected. 6 However, this incidence has been reduced by the introduction of a rotating hinge knee (SMILES: Stanmore Modified Individualarised Lower Extremity System, Stanmore Implants Worldwide, Stanmore, United Kingdom) and a hydroxyapatite (HA) collar at the distal bone/ prosthesis interface. 4 Nevertheless, high loads still pass through the endoprosthesis, predisposing it to loosening. Preservation of the knee joint offers a reduction in the peak loads transferred at the bone-prosthesis interface. 7 A knee joint-sparing proximal tibial replacement has been developed with HAcoated extracortical plates which are screwed into the remnant of the tibia. The use of HAcoated extracortical plates follows experimental work in goats using various different designs of plate. 8 It was found in these studies that three extracortical plates with horizontal grooves provided better early stability, easier insertion and consistent long-term fixation. 4 Cobb et al 7 described this technique mainly for revision surgery for aseptic loosening of prostheses of the distal femur, proximal tibia and humerus. They found the triplate design incorporated well within a remodelled cortex VOL. 91-B, No. 10, OCTOBER 2009 1373

1374 B. G. I. SPIEGELBERG, M. D. SEWELL, W. J. S. ASTON, G. W. BLUNN, R. POLLOCK, S. R. CANNON, T. W. R. BRIGGS Hydroxyapatite-coated plateau Anteromedial plate Hydroxapatite-coated collar Locking bolts inserted anterior/lateral Titanium nitridecoated shaft Hydroxyapatite-coated collar Cemented stem Tapering 9 > 8 mm Fig. 1 Two x 10 mm high cutting fins Anterolateral plate Technical drawing of joint-sparing proximal tibial replacement (Stanmore Implants Worldwide). and achieved osseo-mechanical integration, with all patients regaining their original level of function within five months. 7 As a result of this success, HA-coated extracortical plates have been used in joint-sparing surgery of the distal femur with excellent osseointegration at the prosthesis-proximal bone interface and formation of new bone around the HA collar. The prosthesis allows preservation of the knee and achieves a good functional result. Formation of new bone and remodelling at the interface made the implant more secure. 9 We report the early outcome of eight patients who underwent joint sparing proximal tibial replacement for primary bone tumour. Patients and Methods Between November 2004 and July 2008, eight patients had knee-sparing proximal tibial endoprosthetic replacements for primary non-metastatic bone tumours of the proximal tibia. All patients met the criteria for limb salvage namely a biopsy proving malignant tumour of the proximal tibia requiring complete excision, no involvement of neurovascular structures, no invasion of the knee by the tumour, either clinically or on MRI, and the ability to resect the tibia with clear margins, leaving sufficient bone to allow safe fixation of the prosthesis. The mean horizontal transection point was 20.6 mm proximal (2 to 35) and 36.3 mm distal (10 to 80) to the tumour. There were two males and six females with a mean age of 38.9 years (8 to 43). Of these, five had high grade osteosarcoma, one a malignant fibrous histiocytoma, one an adamantinoma and one a Ewing s sarcoma. All patients received post-operative chemotherapy and none received radiotherapy. Implant design. Each implant was custom-made using CT imaging to ensure accurate dimensions, after proximal and distal transection points had been determined by inspection of the pre-operative MR scans. The implants were manufactured by Stanmore Implants Worldwide Ltd, Elstree, England (Fig. 1). Distal fixation was with a cemented intramedullary stem and an HA-coated collar, as this method had proved successful in distal femoral replacements 10 and may have contributed to the improved outcome at ten years of proximal tibial replacements with a fixed hinge, compared with a rotating-hinge design with an HA-collar. 11 Proximal fixation to the tibial metaphysis was by HA-coated extracortical plates, through which 4.5 mm cortical screws were placed in the medial and lateral cortices. The HA plates were not slotted as described previously 7 because they were too short to be able to incorporate any grooves. The mean width and length of the plates was 23.2 mm (13 to 30) and 17.8 mm (11 to 30) respectively. They were applied close to the joint line as the remaining proximal tibial bone was short, with the length of plate determined by the amount of retained proximal tibia. The mean length of the remaining metaphyseal bone was 26.3 mm (13 to 48) and the mean distance between the upper margin of the plates and the joint line was 8.5 mm (2 to 27, Fig. 2). On the proximal horizontal surface of the implant two HA-coated ridges 10 mm high were used to enhance fixation in the remaining medullary bone (Fig. 3). These ridges were positioned so that they abutted the internal cortices as closely as possible. Operative technique. Under tourniquet control the tibia was resected at the pre-determined proximal and distal levels after marking the anterior tibial cortex as a reference for correct rotational placement of the component. The distal part of the prosthesis was cemented and the proximal part fixed to the remaining tibial metaphysis with a degree of impaction, so that the internal ridges became press-fitted into the cancellous bone. Holes were drilled and 4.5 mm cortical screws were inserted through the medial and lateral plates. The two parts were secured with two locking bolts. A rotation flap of the medial head of gastrocnemius was mobilised to cover the implant and sutured into place. In five patients the attachment of the patellar tendon was sacrificed, and it was repaired directly on to the transposed medial head of gastrocnemius. In one patient an osteotomy was made, preserving the anterior cortex and tibial tuberosity. The proximal part of the prosthesis was fixed in the method described previously, but the anterior bone was wired to the prosthesis, the anterior part of which was also THE JOURNAL OF BONE AND JOINT SURGERY

THE EARLY RESULTS OF JOINT-SPARING PROXIMAL TIBIAL REPLACEMENT FOR PRIMARY BONE TUMOURS 1375 Fig. 3 Computer 3D reconstruction of a joint-sparing replacement housing tibial metaphyseal bone (Stanmore Implants Worldwide). Note two x 10 mm blades on the component surface to further improve metaphyseal fixation. Fig. 2a Fig. 2b a) Anteroposterior and b) lateral radiographs of a joint-sparing proximal tibial replacement 44 months post-operatively. HA-coated to allow for osseointegration of the tibial tuberosity. In the remaining two patients the transection point was distal to the tuberosity and therefore the patellar tendon was preserved. Post-operatively the limb was elevated and a suction drain was used. Antibiotics were given at the induction of anaesthesia and continued for three days post-operatively. The patients remained in an extension brace for six weeks, undertaking only passive flexion and extension, and were required to remain non-weight-bearing on the affected leg. After this progressive weight-bearing was commenced. In the patient in whom the tibial tuberosity was preserved and attached to the prosthesis, a cylinder cast was used to prevent knee movement until eight weeks post-operatively to allow for osseointegration to occur. Functional outcome. Patients were evaluated using the Musculoskeletal tumour society scoring system 12 for limb salvage and the revised Oxford Knee Score (OKS). 13 The former is a six-item scale evaluating pain, function, emotional acceptance, use of supports, walking ability and gait cosmetics in order to produce a score ranging from 0 to 30. A percentage can then be calculated to compare function with other limb-salvage techniques. The OKS is a 12-question form scoring 0 to 4 points per question, producing a score ranging from 0 to 48. The questions are specific to the knee and cover pain, function, use of supports, walking ability and gait. Results Implant function. The mean follow-up was for 35 months (4 to 48), with an overall implant survival of six of the eight devices; the endpoint was revision. One implant failed because the patient fell on the eighth post-operative day and sustained a fracture through the remnant of the tibial metaphysis. Open reduction and internal fixation with two cancellous screws was performed, but malunion with pain and limited mobility developed. At 19 months after the initial operation she required revision to excise the tibial metaphysis and insert a joint sacrificing proximal tibial replacement. The patient with an adamantinoma developed a recurrence at the distal bone-prosthesis interface. She had undergone previous intralesional curettage before referral. Further limb-salvage procedures were not considered possible and a below-knee amputation was performed through the prosthesis. The proximal component was retained and a small end cap was fashioned to fit over it and provide the skeletal structure around which a myocutaneous flap could be fashioned, thereby enabling the patient to retain the knee joint. This operation was performed 14 months after the initial surgery. Before the amputation, she was walking unaided and free of pain, and had been able to swim and cycle. 14 One patient complained of a prominent lateral metaphyseal screw, which was removed at six months. Another VOL. 91-B, No. 10, OCTOBER 2009

1376 B. G. I. SPIEGELBERG, M. D. SEWELL, W. J. S. ASTON, G. W. BLUNN, R. POLLOCK, S. R. CANNON, T. W. R. BRIGGS patient complained of pain at the ankle joint. It was felt that this was due to syndesmotic instability and a diastasis screw was inserted at 41 months, with resolution of the symptoms. Functional scores were made on the six patients who still had the joint-sparing prosthesis in situ. The mean musculoskeletal tumour society outcome score was 24 of 30 (79% (57% to 90%) and the mean OKS was 40 (36 to 46) at final follow-up. All patients reported difficulty kneeling down and standing up owing to pain experienced when pressure was applied to the operated area. The mean knee flexion was 112 (100 to 120 ). There was no statistical significance between the amount of physis remaining and functional outcome, although patient 2 had the least amount of remaining metaphysis (13 mm) and the worst musculoskeletal tumour society score (57%). Varus and valgus instability of the knee was not encountered as the collateral ligaments remained intact and rotational malalignment was avoided. Discussion Extracortical plate fixation has the advantage of allowing incorporation of the prosthesis into the load-bearing structure of the bone as it remodels. 15 It takes advantage of the biological phenomenon shown by callus at a fracture site or in a loose massive replacement, in which bone is laid down outside the cortex, away from the loosening prosthesis. Bone forms on the periosteal surface in a centripetal fashion, incorporating the plates into the new expanded cortex. 7 This process is aided by the use of plates of different lengths and tapering thickness which allow gradual transfer of the load. Ridges on the proximal surface of the implant to aid stability as does HA coating of the inside and outside of the plates, the plateau and collar of the implant. Minimal additional soft-tissue dissection was required at the proximal end of the tibia to obtain adequate exposure for fixation of the extracortical plates. This was achieved without evidence of disruption of the periosteal blood supply to the remaining bone which might have led to cortical necrosis. At 44 months, it was not possible to determine on plain radiographs whether there was evidence of extraosseous bone formation, in contrast with the findings of our distal femoral series. 9 The plates used in the tibial device were not slotted, as they were too short and this might partly explain the lack of extraosseous bone. Despite this, there was no radiological or clinical evidence of loosening, indicating good stable fixation at the proximal interface. The screws locking the extracortical plates from the medial and lateral sides provide strong fixation in the short-term. The smallest plates were 10 mm long, which allowed sufficient room to incorporate the cortical screws. The remaining metaphysis must be at least 2 mm longer than the plates so that they do not interfere with the function of the joint. In our series the maximum length of remaining metaphyseal bone was 48 mm. It is considered that these prostheses could be used when there is up to 60 mm of remaining metaphysis. In circumstances where there is a greater length of bone remaining a diaphyseal replacement is an option, as an intramedullary stem can be used to aid stability. The length of 60 mm has been defined as the cut-off, as the shortest stem is 50 mm and a further 10 mm is required to prevent the stem disrupting the articular surface. Another treatment option for limb-salvage is the use of grafts, and research using osteoarticular or intercalary allografts to fill the defect created after tumour resection has been performed. A series of osteoarticular allografts reported failure in five of seven grafts. 16 Results with intercalary grafting have been more favourable. A multicentre study performed by the European Musculoskeletal Oncological Society observed an 18.5% incidence of infection and a 21% prevalence of fracture of the graft in the tibia. 17 Delayed or nonunion was observed in 63% of the intercalary grafts, with overall function reported as excellent or good in 63%. 17 Another long-term follow-up study of intercalary allografts reviewed the results of 104 procedures, of which 38 were in the tibia. Infection occurred in 12%, nonunion in 30% and fracture in 17% and an additional 81 procedures were needed to obtain satisfactory results. Importantly, there was an overall success rate of 84% of patients, but in those who required adjuvant therapy, complications and failures occurred more frequently and the rate of success was reduced to 60%. 18 A more recent study of 59 intercalary grafts, of which 19 were at the tibia, reported a five-year survival of 79%, with infection and fracture rates of just 5% and 7%, respectively. 19 Similar results have been observed in patients who received free vascularised fibular grafts, with four of ten reconstructions resulting in failure, 20 including two stress fractures and two nonunions, one of which developed local recurrence and the other an infection requiring above-knee amputation. The functional outcome using the Musculoskeletal Tumour Society score was excellent in six patients, good in two and poor in two. The main problems associated with grafting procedures, are the increased complications, mainly infection, fracture and nonunion. Furthermore, with grafts the time to full weight-bearing is considerably longer than with an endoprosthesis. In our series, one implant failed and one patient required an amputation for tumour recurrence. In the remaining six patients the joint-sparing prosthesis has comparable results to the joint-sacrificing prostheses analysed by Grimer et al. 4 Comparing their results, the mean range of flexion in our patients was 112, compared to 104, and our overall functional score was 79% compared to 77%. We feel our preliminary results are promising, with preservation of the natural joint, at least in the short term. Our patients need longer follow-up with regard to their susceptibility to aseptic loosening, which is a known midterm problem in proximal tibial replacements. 6 THE JOURNAL OF BONE AND JOINT SURGERY

THE EARLY RESULTS OF JOINT-SPARING PROXIMAL TIBIAL REPLACEMENT FOR PRIMARY BONE TUMOURS 1377 No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. References 1. Unni KK. Dahlin s bone tumors: general aspects and data on 11,087 cases. Fifth ed. Philadelphia: Lippincott Wiliams & Wilkins, 1996. 2. Bramwell VH, Burgers M, Sneath R, et al. A comparison of two short intensive adjuvant chemotherapy regimens in operable osteosarcoma of limbs in children and young adults. J Clin Oncol 1992;10:1579-91. 3. Simon MA. Limb salvage for osteosarcoma. J Bone Joint Surg [Am] 1988;70-A:307-10. 4. Grimer RJ, Carter SR, Tillman RM, et al. Endoprosthetic replacement of the proximal tibia. J Bone Joint Surg [Br] 1999;81-B:488-94. 5. Malawer MM, McHale KA. Limb sparing surgery for high-grade malignant tumours of the proximal tibia: surgical technique and a method of extensor mechanism reconstruction. Clin Orthop 1989;239:231-48. 6. Unwin PS, Cannon SR, Grimer RJ, et al. Aseptic loosening in cemented custom-made replacements for bone tumours of the lower limb. J Bone Joint Surg [Br] 1996;78-B:5-13. 7. Cobb JP, Ashwood N, Robbins G, et al. Triplate fixation: a new technique in limbsalvage surgery. J Bone Joint Surg [Br] 2005;87-B:534-9. 8. Coathup MJ, Cobb JP, Walker PS, Blunn GW. Plate fixation of prostheses after segmental resection for bone tumours. J Orthop Res 2000;18:865-72. 9. Gupta A, Pollock R, Cannon SR, et al. A knee sparing distal femoral endoprosthesis using hydroxyapatite-coated extracortical plates: preliminary results. J Bone Joint Surg [Br] 2006;88-B:1367-72. 10. Unwin PS, Walker PS, Briggs TWR. Rotating hinged, hydroxyapatite coated versus fixed hinged, uncoated distal femoral replacements: a study of 402 cases. In: Procs Tenth International Symposium of the International Society of Limb Salvage, 1999:15. 11. Myers GJC, Abudu AT, Carter SR, Tillman RM, Grimer RJ. The long-term results of endoprosthetic replacement of the proximal tibia for bone tumours. J Bone Joint Surg [Br] 2007;89-B:1632-7. 12. Enneking WF, Dunham W, Gebhardt MC, Malawer M, Pritchard DJ. A system for the functional evaluation of reconstructive procedures after surgical treatment of tumours of the musculoskeletal system. Clin Orthop 1993;286:241-6. 13. Murray DW, Fitzpatrick R, Rogers K, et al. The use of the Oxford hip and knee scores. J Bone Joint Surg [Br] 2007;89-B:101-4. 14. Spiegelberg BGI, Sewell MD, Coltman T. Below-knee amputation through a jointsparing proximal tibial replacement for recurrent tumour. J Bone Joint Surg [Br] 2009;91-B:815-19. 15. Coathup MJ, Bates P, Cool P. Osseo-mechanical induction of extra-cortical plates with reference to their surface properties and geometric designs. Biomaterials 1999;20:793-800. 16. Rodl RW, Ozaki T, Hoffmann C, et al. Osteoarticular allograft in surgery for highgrade malignant tumours of bone. J Bone Joint Surg [Br] 2000;82-B:1006-10. 17. Donati D, Capanna R, Campanacci D, et al. The use of massive bone allografts for intercalary reconstruction and arthrodeses after tumour resection: a multicentric European study. Chir Organi Mov 1993;78:81-94. 18. Ortiz-Cruz E, Gebhardt M, Jennings C, Springfield DS, Mankin HJ. The results of transplantation of intercalary allografts after resection of tumours. J Bone Joint Surg [Am] 1997;79-A:97-106. 19. Musculo DL, Ayerza MA, Aponte-Tinao L, Ranalletta M, Abalo E. Intercalary femur and tibia segmental allografts provide an acceptable alternative in reconstructing tumour resections. Clin Orthop 2004;426:97-102. 20. Hsu RW, Wood MB, Sim FH, Chao EY. Free vascularised fibular grafting for reconstruction after tumour resection. J Bone Joint Surg [Br] 1997;79-B:36-42. VOL. 91-B, No. 10, OCTOBER 2009