Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Similar documents
Calibration of two 90 Sr+ 90 Y dermatological applicators

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Evaluation of glasses containing cadmium for high dose dosimetry by the thermoluminescence technique

Mammography calibration qualities establishment in a Mo- Mo clinical system

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Good manufacturing practices (GMP) utilized on human blood irradiation process

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

In vivo dosimetry with thermoluminescent. dosimeters in external photon beam. radiotherapy

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

BRACHYTHERAPY WITH 125-IODINE SOURCES: TRANSPORT AND RADIATION PROTECTION

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

The need for standardization of dosimetry in experimental radiation biology

X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS

Applications in brachytherapy

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Infrastructure of the Brazilian Government in Radioprotection Education

"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Male Breast Cancer-Dosimetry Evaluation in Teletherapy

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Hand Dose in Nuclear Medicine Staff Members

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

METALS IN SOILS ADJACENT TO AVENUES OF HIGHLY DENSE TRAFFIC OF SÃO PAULO CITY, BRAZIL

Breast phantom with silicone implant for evaluation in conventional mammography

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY

STUDY OF DOSIMETRIC EFFECTS DUE TO METALLIC HETEROGENEITY OF TISSUE EXPANDERS IN POST-MASTECTOMY RADIOTHERAPY

PROSTATE BRACHYTHERAPY DOSIMETRY USING VOXEL PHANTOMS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT

Radiologic Units: What You Need to Know

Development of the Nation-Wide Dosimetric Monitoring Network in Ukraine

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Radiochromic film dosimetry in water phantoms

ENTRANCE SKIN DOSE MEASUREMENTS FOR PAEDIATRIC CHEST X-RAYS EXAMINATIONS IN BRAZIL

SENSORIAL EVALUATION OF BRAZILLIAN HONEY

The clinical use of OSLD

High-Dose Standardization Study For γ-rays radiation processing at NIM

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis

Protection of the contralateral breast during radiation therapy for breast cancer

Sang Hun Shin, Wook Jae Yoo, Kyoung Won Jang, Seunghyun Cho, 1 Kum Bae Kim, 2 and Bongsoo Lee 3*

DOSIMETRIC STUDY OF I-125 SEEDS USED IN PROSTATE BRACHYTHERAPY

Sulfur status in long distance runners

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Radiation Protection Services in R.Macedonia

The determination of timer error and its role in the administration of specified doses

The use of polyvinyl chloride films dyed with methyl red in radiation dosimetry

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Zakithi Msimang SAAPMB/SARPA 2011

THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Process Control Methods in Radiation Technologies

Y Physics Radiation Measurements and Monitoring

U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Major accidents in radiotherapy

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Dosimetry in Life Sciences

Radiation Monitoring Instruments

RADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S

Publishable JRP Summary Report for JRP HLT11 MetroMRT Metrology for Molecular Radiotherapy Background

Model 3500 Manual TLD Reader with WinREMS 3500-W-O-0602 Page 5-1 Operator's Manual

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Improvement of radiation dosimetry: contribution of the IAEA

EURADOS: Overview of actions and initiatives in radiation protection in medicine

9.7 Dosimetry Study on A p-n Junction Semiconductor Detector

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI)

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Study of neutron-dna interaction at the IPEN BNCT Research facility ABSTRACT

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Energy dependent response of Al 2 O 3 and its potential application in personal monitoring.

A NEW USER-FRIENDLY EXTREMITY DOSIMETRY SYSTEM DESIGN, COMPARISON, AND TESTING

Assessment of dosimetric parameters in mammography employing a homemade ionization chamber

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Transcription:

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters Patrícia de Lara Antonio 1*, Rogério M. V. Silva 2, Divanizia do Nascimento Souza 2 and Linda V. E. Caldas 1 1 Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, IPEN-CNEN/SP Av. Professor Lineu Prestes 2242, 05508-000, São Paulo, Brazil 2 Departamento de Física, Universidade Federal de Sergipe, UFS Av. Marechal Rondon s/nº, Jardim Rosa Elze, 49100-000, São Cristovão, Sergipe, Brazil Abstract. According to international recommendations, the sources used in procedures of brachytherapy have to be calibrated in order to confirm the absorbed dose rates. In order to follow the recommendations and to collaborate for the continuous use of these sources, a dosimetric system for their calibration was developed at the Calibration Laboratory at IPEN, and it was tested in some hospitals at the São Paulo city to train the users of clinical applicators, and to answer the possible questions that may appear in relation to the calibration procedure. These situations were necessary to verify the possibility of application of this dosimetric system as a mail system, because there are hospitals and clinics in almost all Brazilian states that still use clinical applicators. To send these sources to the Calibration Laboratory is difficult, and it requires authorization from the governmental authorities. The thermoluminescent dosimeters of the mail system were already characterized. This postal kit was tested to calibrate 90 Sr+ 90 Y clinical applicators; the absorbed dose rates obtained were considered satisfactory. Keywords: 90 Sr+ 90 Y clinical applicators, calibration, CaSO 4:Dy pellets, thermoluminescent dosimeters. 1. Introduction The need of calibration of 90 Sr+ 90 Y clinical applicators, special beta particle emitting sources used in betatherapy procedures for clinical or esthetical treatments, is already known for many years, by means of reports and recommendations of several authors [1-4]. This procedure is of great importance, because these sources are very old (some are from the 60 s), in addition to the necessity of the establishment of a quality control program for these sources, assuring that the sources are used correctly and the treatments of the patients will bring effective results. There sources are mainly used to treat keloids (dermatological/plane applicators) and pterigium (ophthalmic/concave applicators) problems. There are many techniques to calibrate clinical applicators, as for example the use of extrapolation chambers [5], radiochromic films [6] and thermoluminescent dosimeters (TLD) [7]. This last method presents some advantages, as easy handling and readout, and small dimensions. Oliveira and Caldas [8] studied the application of thin CaSO 4:Dy dosimeters in the calibration of 90 Sr+ 90 Y clinical applicators, and concluded that this material is very efficient for the determination of the absorbed dose rates of these kinds of sources. In some countries, the use of 90 Sr+ 90 Y clinical applicators was suspended, because they are not commercialized anymore. The use of beta sources was replaced by linear accelerators ou kilovoltage * Presenting author, E-mail: patrilan@ipen.br

X-ray systems (in the case of keloids) [9]. However, the 90 Sr+ 90 Y concave and plane sources are still very utilized in Brazil, in almost all federal states, for betatherapy in the treatments of superficial lesions (kelois and pterigium). Based on the principle of the studies [1-3,8], Antonio and Caldas [7] developed in 2011 a dosimetric system for calibration of 90 Sr+ 90 Y clinical applicators, using thin CaSO 4:Dy pellets produced at the Instituto de Pesquisas Energéticas e Nucleares (IPEN). They applied this system in hospitals of São Paulo city with the objective to establish a training program to the operators of these sources about how to use them, and mainly to eliminate possible doubts about their calibration. The purpose was to send this dosimetric system, in future, as a postal system to other Brazilian clinics and hospitals that use 90 Sr+ 90 Y applicators. As the calibrations performed in the study of Antonio and Caldas [7] presented satisfactory results between the absorbed dose rates obtained and the ones provided in the calibration certificates, for different clinical applicators, the objective of this work was to verify the usefulness of this system as a postal kit, using the usual mail system. The need of the establishment of a program for this postal kit is mainly due to the fact of the transport difficulty from the clinics and hospitals to the calibration laboratories. 2. Materials and Methods For the calibration of the 90 Sr+ 90 Y clinical applicators, thin CaSO 4:Dy thermoluminescent (TL) dosimeters, with dimensions of 6.0 mm of diameter and 0.2 mm of thickness, were utilized; these samples were produced at the Dosimetric Materials Laboratory of IPEN. The dosimetric system was composed by sixteen TL dosimeters, a clamp, a polymethymethacrylate (PMMA) support (with dimensions of 5.0 cm of diameter and 1.0 cm of thickness) to fix the TL pellets involved by a plastic film, with superficial density of 1.095 mg/cm 2 (for protection of the pellets during the irradiations), an aluminum support to storage the TL detectors (before irradiations), a dark small box for the TL detectors (after the irradiation procedure), two pairs of gloves, a chronometer and a pen. All components were positioned in a organizing box for the storage of the items of the system, and it may be seen in Fig. 1. Furthermore, an irradiation methodology, an irradiation procedure form, a data collection form and a revision sheet were sent together with the kit.

Chronometer Pen Box for the irradiated TL detectors Clamp Aluminum support Gloves PMMA support Figure 1 Items of the dosimetric postal kit, for the calibration of 90 Sr+ 90 Y clinical applicators. A 90 Sr+ 90 Y clinical applicator was utilized as a secondary standard system for the calibration of the other applicators, because it was calibrated at the primary standard laboratory National Institute of Standards and Technology (NIST). This applicator was called, in this work, NIST applicator. During the irradiations, the distance between each sample and source was null (in the same conditions used during the irradiations of the NIST applicator), and an adequate support to fix the source (vertically) on the pellet was utilized. In this work, the dosimetric postal system was sent to the Federal University of Sergipe (UFS), at Aracaju city (about 2,174 km of distance from São Paulo). The TL dosimeters were irradiated following the irradiation procedure form. Three 90 Sr+ 90 Y clinical applicators, all from Amersham International, England, with calibration certificates from The Radiochemical Centre, Amersham (manufacturer certificate) were calibrated using the postal kit. The description of these sources is presented in Table 1. Source Number Table 1 Description of the 90 Sr+ 90 Y clinical applicators used in this work. Clinical Applicator Model Nominal Activity (MBq) Absorbed Dose Rate (Gy/s) Calibration Date 1 Dermatological 1520 SIA5 74 0.018 ± 0.004 27.11.1973 2 Ophthalmic 928 SIA6 370 0.027 ± 0.008 14.01.1992 3 Ophthalmic 1522 SIA6 370 0.022 ± 0.017 27.11.1973 The measurements were taken after the postal system returned to IPEN, and they were analyzed using a Harshaw TLD reader model 3500, with light emission integrated in the temperature interval from 180 C to 350 C. After the measurements, the CaSO 4:Dy pellets were termally treated at 300 C during 3 h for posterior reutilization.

3. Results Prior the establishment of the dosimetric system, the TL dosimeters were characterized, and these results were reported by Antonio and Caldas [7]. 3.1. Development of the Dosimetric Postal Kit As the dosimetric postal system (or postal kit) was developed to attend the necessity of the Brazilian clinics and hospitals in relation to the calibration of their 90 Sr+ 90 Y clinical applicators, it was necessary to verify the viability and the effectiveness of this system [7]. After the conclusion that this system could be used as postal system, the irradiation procedure form already utilized by Antonio and Caldas [7], was refined based in the doubts presented by the source operators during the visits to the hospitals in São Paulo city. This form contains all steps necessary to calibrate a clinical applicator, since important recommendations about the protection of the source operators and who are allowed to perform the irradiations, until the step by step procedure during the irradiations to be realized in the own place where the clinical applicators are used. Furthermore, a data collection form was also developed (to be completed by the responsible of the sources with data about the source owners, the sources, and the irradiation procedure). An item revision sheet consisting of all the items of the postal system was also sent; there is a proper place to be completed by the source operators, that must return to the LCI. 3.2. Determination of the Absorbed Dose Rates For the calibration of the 90 Sr+ 90 Y clinical applicators of the UFS, a dose-response curve obtained for the NIST applicator [7], was utilized; it is presented in Fig. 2. 40 35 30 TL Response (uc) 25 20 15 10 5 0 4 6 8 10 12 14 16 18 20 22 Absorbed Dose (Gy) Figure 2 Dose-response curve of the thin CaSO 4:Dy detectors, using the NIST reference 90 Sr+ 90 Y applicator.

From the set of sixteen CaSO 4:Dy pellets, four were not irradiated, and they were used as control dosimeters; four samples were irradiated using each one of the three clinical applicators during different time intervals (Table 2), which were determined and used according to the absorbed dose rates provided at the calibration certificate of each source. The absorbed dose rate for each applicator was determined using the dose-response curve of the NIST applicator. After the irradiations made at the UFS and the evaluation of their measurements at the LCI (IPEN), a calibration certificate was emitted for each clinical applicator. The results obtained for the calibration of all three clinical applicators are shown in Table 2. The maximum relative deviation obtained in all the measurements was 8.1%, for the applicator 2. Table 2 Absorbed dose rates obtained for the three 90 Sr+ 90 Y clinical applicators of the UFS. Source Number Source Irradiation Time (min) Absorbed Dose Rate (Gy/s) Certificate This work Difference (%) 1 Dermatological 10 0.0070 ± 0.0014 0.0076 ± 0.0015-7.9 2 Ophthalmic 6 0.0167 ± 0.0050 0.0227 ± 0.0076-26.4 3 Ophthalmic 8 0.0087 ± 0.0026 0.0189 ± 0.0059-54.0 The values of absorbed dose rates obtained in this work were compared with the ones provided in the calibration certificate of each applicator. Applicator 1, dermatological, presented the minimum difference of -7.9%, which agrees with calibration results of other applicators [7]. The applicators 2 and 3, ophthalmic, presented high differences. 4. Conclusions The dosimetric postal system was put into practice in this work after having been tested in a previous study. Using this system, three 90 Sr+ 90 Y clinical applicators (one dermatological and two ophthalmic) had their absorbed dose rates obtained using TL detectors of CaSO 4:Dy. The result obtained in this work for the dermatological applicator demonstrated a difference between the absorbed dose rate of this work and that provided in its calibration certificate (corrected for the radioactive decay) acceptable in the point of view of other studies and the uncertainty presented in its certificate of 20%. In relation to the ophthalmic applicators, the results of the calibration and the differences obtained between the absorbed dose rates were higher than that obtained for the dermatological applicator (-26.4% and -54.0% to the applicator 2 and 3, respectively). The first difference obtained can be considered acceptable, because it is compatible with the uncertainty presented in the calibration made by the manufacturer, that is 30%. However, the second difference (-54%) was considered a very high value, and this may be due to factors such as small errors during the irradiation

of the samples, the format of the source, or the disposal of radioactive material on the source surface, which may not be homogeneous. This clinical applicator will be calibrated again. Acknowledgments The authors are thankful to the Brazilian agencies FAPESP, CNPq, CAPES and MCT (Project INCT for Radiation Metrology in Medicine), Brazil, for partial financial support. References [1] IAEA, International Atomic Energy Agency. Calibration of photon and beta ray sources used in brachytherapy. IAEA-TECDOC-1274, Vienna (2002). [2] Soares, C. G. Comparison of NIST and manufacturer calibrations of 90 Sr+ 90 Y ophthalmic applicators. Med. Phys. 22 (9), 1487-1493 (1995). [3] Kollard, R. P.; Dries, W. J. F.; van Kleffens, H. J.; Aalbers, A. H. L.; van der Marel, J.; Marijnissen, J. P. A.; Piessens, M.; Schaart, D. R.; de Vroome, H. Recommendations on detectors, measurement procedures and quality control of beta sources. Task group quality control of sealed beta sources in brachytherapy. Report 14 of the Netherlands Commission on Radiation Dosimetry (2004). [4] Holmes, S. M.; Micka, J. A.; DeWerd, L. A. Ophthalmic applicators: an overview of calibrations following the change to SI units. Med. Phys. 36 (5), 1473-1477 (2009). [5] Oliveira, M. L.; Caldas, L. V. E. A special mini-extrapolation chamber for calibration of 90 Sr+ 90 Y sources. Phys. Med. Biol. 50, 2929-2936 (2005). [6] Soares, C. G. A method for the calibration of concave 90 Sr+ 90 Y ophthalmic applicators. Phys. Med. Biol. 37 (4), 1005-1007 (1992). [7] Antonio, P. L.; Caldas, L. V. E. Application of a dosimetric system for calibration of 90 Sr+ 90 Y sources used in brachytherapy. Radiat. Meas. 46, 2025-2027 (2011). [8] Oliveira, M. L.; Caldas, L. V. E. Performance of thin CaSO 4:Dy pellets for calibration of a 90 Sr+ 90 Y source. Nucl. Instr. Meth. Phys. Res. A 580, 293-295 (2007). [9] Eaton, D. J.; Barber, E.; Ferguson, L.; Simpson, G. M.; Collis, C. H. Radiotherapy treatment of keloid scars with a kilovoltage X-ray parallel pair. Radiot. Oncol. 102 (3), 421-423 (2012).