Policy for Participation in Proficiency Testing Activities

Similar documents
Proficiency Testing Policy

Proficiency Testings (PTs) and Interlaboratory Comparisons (ILCs) for the Department of Calibration Laboratories

PROFICIENCY TESTING AND OTHER COMPARISON PROGRAMME REQUIREMENTS FOR CALIBRATION LABORATORIES

LAS-Q008. Proficiency Testing Requirements for Testing and Calibration Laboratories

SNAS POLICY ON PARTICIPATION IN PROFICIENCY TESTING

Developments in IAF/PAC/ILAC/APLAC/ISO. Anil Jauhri CEO, NABCB

IAF Mandatory Document. Requirements for the Migration to ISO 45001:2018 from OHSAS 18001:2007 (IAF MD 21:2018)

TG Approved By: SANAS Page 1 of 6

Raad voor Accreditatie (Dutch Accreditation Council RvA) Migration from BS OHSAS 18001:2007 to ISO 45001:2018

UKAS Guidance on the Application of ISO/IEC Dealing with Expressions of Opinions and Interpretations

ISO and the Delivery of Competent NDT Services

PCN-ISO GEN Requirements for the Qualification of personnel for Limited Applications of Non-destructive Testing

NA-019 LEVEL 3 QUALIFICATION RECOGNITION FOR AS3669 AND EN 4179 / NAS 410

ttabl National Accreditation Board for Testing and Calibration Laboratories Department of Science &Technology, India CERI'IFICATE OF ACCREDITATION.

LAB5 EDITION 2 NOVEMBER 2012

Test Report To AS/NZS :1998 in conjunction with AS/NZS :2003. Report No. T140218_SA

NANDTB-F012 Rev 0 Audit Checklist

Water Microbiology Proficiency Test Scheme. Overview & Description

A118 Laboratory Proficiency Testing Plan Revision 1.4 March 6, 2014

NA LEVEL 3 TECHNICAL QUALIFICATION RECOGNITION

TRANSITION OVERVIEW. OHSAS 1800:2007 to ISO 45001:2018

Non-destructive testing Penetrant testing. Part 2: Testing of penetrant materials

NA-019 Level 3 Qualification Recognition for EN 4179 / NAS 410 and AS3669

IAS Presentation at Kuwait Labs nd Kuwait Conference for Laboratory Quality and Accreditation Empowering Laboratory Personnel

About NATA and GLP recognition. Information on the OECD Principles of Good Laboratory Practice compliance monitoring program

OHSAS Project Group. Implementation Guidance for migrating from OHSAS 18001:2007 to ISO 45001:2018 CONTENTS 1.0 INTRODUCTION

a practical guide ISO 13485:2016 Medical devices Advice from ISO/TC 210

CERTIFICATION OF PERSONNEL FOR ULTRASONIC TESTING OF CASTINGS

Comparability and quality of experimental data under different quality systems. S. Caroli Istituto Superiore di Sanità Rome

NMISA A discussion on Force ILC's and methods to improve results. 13 October 2015 Speaker: S.T.S. Dlamini Author: C.H. Gouws

Magnetic Testing (MT) Level 2

LEAF Marque Assurance Programme

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD

ISO 13485:2016 MEDICAL DEVICES QMS TRANSITION GUIDE

APMP-APLAC joint PT (APLAC T109) Measurement of Cadmium in Milk Powder

OCCUPATIONAL THERAPISTS BOARD. MANUAL for Accreditation of Continuing Professional Development (CPD) Program Provider. for

ISO INTERNATIONAL STANDARD. Non-destructive testing Qualification of personnel for limited applications of non-destructive testing

Sterilization of health care products Radiation. Part 2: Establishing the sterilization dose

SEG s Final Draft THE ORGANISATION OF THE ISLAMIC CONFERENCE (OIC) GUIDELINES FOR THE HALAL ACCREDITATION BODY ACCREDITING HALAL CERTIFICATION BODIES

The use of QCMD proficiency testing panels in clinical virology.

The techniques of continually improving the quality system for ISO accredited laboratories

Advanced Radiographic Testing (RT) Level 2.9

Trust Policy 218 Ionising Radiation Safety Policy

Iso Audit Checklist

Drug Treatment Centre Board

Introduction to Quality Management Systems

Role of Accreditation in Halal Food Industry مركز االمارات العالمي لالعتماد

AUDIT TOOL FOR SELF INSPECTION OF COMPLIANCE WITH QUALITY MANAGEMENT SYSTEM FOR PATIENT GROUP DIRECTIONS

National Voluntary Laboratory Accreditation Program

This document is a preview generated by EVS

IPC Athletics. Classification Rules and Regulations

Iso Audit Checklist

Airmid Health Group Limited t/a Airmid Health Laboratory

Ultrasonic Testing (UT) Level 2

ISO INTERNATIONAL STANDARD. Steel products Employer's qualification system for non-destructive testing (NDT) personnel

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ACOUSTICS, ULTRASOUND AND VIBRATION METROLOGY

ISO/TR TECHNICAL REPORT. Non-destructive testing Guidelines for NDT personnel training organizations

Jun06 Rev A

UNITED ARAB EMIRATES

Non-destructive Testing ISO/IEC Application Document. Issue Date: October 2017

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

ESYD POLICY RELEVANT TO PARTICIPATION IN PROFICIENCY TESTING SCHEMES AND INTERLABORATORY COMPARISONS

ISO IDF 211 INTERNATIONAL STANDARD

Eddy Current Testing Level 1

Primary Health Networks

a practical guide Medical Devices Advice from ISO/TC 210 This is a free 11 page sample. Access the full version online.

Draft for comments only Not to be cited as East African Standard

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Procedure for Registration / Renewal of Third Party Hazmat/Industrial Tank Cleaning Contractors

Certificate of Accreditation to ISO/IEC 17025:2005

Medical Bureau of Road Safety

CLIENT PROCEDURE FOR ANNUAL APPROVAL OF SHIP REPAIR COMPANIES

Testing Railway Safety Critical Workers for Drugs and Alcohol

Alpha Analytical Services Ltd

Accreditation Criteria

OIE Reference Laboratory Reports Activities

IPC Athletics Classification Rules and Regulations

INTERNATIONAL STANDARD

Form for Regular Reporting of Regional Instrument Centres

Quality Assurance Documents

Approval of Dosimetry Services in Ireland Guidelines for Applicants

13 th Meeting Forensic Science Technical Advisory Committee (FSTAC)

GUIDELINES OF IMPORTS FOR THE TRADE EMIRATES CONFORMITY ASSESSMENT SCHEME (ECAS) HALAL CERTIFICATION - MIDDLE EAST

This document is a preview generated by EVS

Medical Bureau of Road Safety

Test report KIBING SPECIAL GLASS ZHANGZHOU CO., LTD CHENG AN ROAD, KANGMEI TOWNSHIP, DONGSHAN COUNTY ZHANGZHOU,FUJIAN, CHINA

Dairygold Co-Op Society Limited Dairygold Feed Laboratory

NHS public health functions agreement Service specification No.32 Human papillomavirus immunisation programme for men who have sex with men

Chemotherapy Training and Assessment Policy. For Medical Prescribers and Pharmacy Verifiers

TECHNICAL GUIDANCE FOR THE ACCREDITATION OF ANDROLOGY LABORATORIES

ttabl National Accreditation Board for Testing and Calibration Laboratories Department of Science &Technology, India CERI'IFICATE OF ACCREDITATION.

Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects

This document is a preview generated by EVS

Darwin Marine Supply Base HSEQ Quality Management Plan

TECHNICAL SPECIFICATION

भ रत य व म नपत तन प र ध करण

INTERNATIONAL STANDARD

TERMS AND CONDITIONS NATIONAL GOOD LABORATORY PRACTICE (GLP) COMPLIANCE MONITORING AUTHORITY

ISO INTERNATIONAL STANDARD. Steel products Employer's qualification system for non-destructive testing (NDT) personnel

Transcription:

National Accreditation Board for Testing and Calibration Laboratories (NABL) NABL 163 Policy for Participation in Proficiency Testing Activities ISSUE NO. : 08 ISSUE DATE : 07-Mar-2019 AMENDMENT NO. : -- AMENDMENT DATE :--

AMENDMENT SHEET Sl Page No. Clause No. Date of Amendment Amendment Reasons Signature QM Signature CEO 1 2 3 4 5 6 7 8 9 10 Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 1/7

CONTENTS Sl. No. Title Page Amendment sheet 1 Contents 2 1. Introduction 3 2. Scope 3 3. NABL policy 4 4. Annexure A Proficiency Testing Plan (Form 18) 7 Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 2/7

1. INTRODUCTION Proficiency Testing is one of the important tools to determine the technical competence of the testing / medical / calibration laboratories. According to ISO/IEC 17025:2017, a laboratory shall have quality control procedures for monitoring the validity of tests and calibrations undertaken. This monitoring may include the participation in interlaboratory comparisons or proficiency testing programs. Other means may include the regular use of reference materials, or replicate tests or calibrations using the same or different methods. By these mechanisms a laboratory can provide evidence of its competence to its clients, interested parties and the accreditation body. According to ISO 15189:2012, medical laboratories seek confirmation for confidence in their results through participation in suitable inter-laboratory comparisons. This document derives NABL policy on participation in Proficiency testing activities to conform to the requirements of ILAC-P9:06/2014, ILAC Policy for Participation in Proficiency Testing activities. 2. SCOPE This document stipulates the minimum requirements of PT participation during accreditation process of testing, calibration and medical laboratories (Conformity Assessment Bodies - CABs). This also gives information about the measures against poor performance of CABs as defined in Proficiency Testing program. Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 3/7

3. NABL POLICY 3.1. Applicant CAB shall successfully participate in at least one PT program prior to gaining accreditation in each discipline applied. 3.2. For applicable tests and calibrations in the CAB s scope of accreditation, the CAB shall seek participation in one or more of the following PT activity options. They are listed in descending order of preference, with the most desirable at the top of the list: I. Scheme delivered by an Accredited PT provider II. Scheme delivered by PTP which is accepted by NABL by other means like approval, funded, policies or guidance document; III. Scheme delivered by an Applicant PT provider; However, where above formal PT Programs are not available / scheduled or not appropriate, alternatively: a. The applicant testing/ medical laboratory shall participate in suitable inter-laboratory comparisons with sufficient number of accredited laboratories, as per ISO/IEC 17043:2010 Conformity assessment General requirements for proficiency testing. b. The applicant calibration laboratory shall participate in suitable inter- laboratory comparisons, as per NABL 164 Guidelines for Inter-laboratory Comparison for Calibration Laboratories where formal PT Programs are not available. Note: There are areas of testing and calibration for which suitable PT / ILC is not practical. In such cases, NABL and the laboratory shall discuss and agree on suitable alternative means by which performance can be assessed and monitored. This would need to be considered as part of the planned PT and/or related activities. 3.3. Accredited CAB shall have 4-year plan for Proficiency Testing participation which shall cover all the accredited groups as practicable (e.g. Food & agriculture, pressure & vacuum, high voltage, building material, etc.) under each discipline of accreditation. In Medical testing, accredited laboratories shall participate in at least one Proficiency Testing / External Quality Assurance Scheme in a year per discipline, as appropriate. This plan shall be applicable strictly considering the above options for PT given under section 3.2 of this document. 3.4. Successful participation in PT is necessary for significant enhancement of scope such as addition of discipline Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 4/7

3.5. CAB shall take appropriate corrective actions in case of poor performance in PT programs within one month period. In two consecutive events of poor performance in PT participation/ unsatisfactory corrective actions leads to the scope reduction as per NABL 216. 3.6. Apart from the above defined minimum requirements, CAB shall also abide by the PT participation required by the respective Regulators, NABL & regional bodies like APAC. 3.7. Accredited CAB shall submit 4 year PT plan as per Form 18 (Annexure A) which will be reviewed by NABL for its suitability in relation to the scope of accreditation. In case of non-suitability of the PT plan submitted by the CAB, NABL may guide in identifying the required PT coverage. 3.8. NABL provides guidance to the CABs regarding the proficiency testing programs through its website, by giving the list of NABL accredited / applicant PT Providers with their accredited PT scheme details. NABL has also given reference to some international organization providing PT program. This helps the CABs in participating in other programs, in case where no national programs are available. 3.9. The PT program should be conducted in accordance with ISO/ IEC 17043:2010. All techniques covered in ISO/IEC 17043:2010 shall be acceptable for the purpose of Proficiency Testing. 3.10. Successful participation in national / international PT programs enhances the confidence in the competence of CABs. 3.11. At the time of on-site assessment, assessment team shall verify whether the CABs have participated in recognized Proficiency Testing Program as per the PT plans (Form 18) and whether the necessary corrective action were taken, if in terms of z-score, En values etc. the CAB performance has not been satisfactory. Note: For calibration laboratories, measurement audits will be considered to be a complimentary effort to Inter-Laboratory Proficiency Testing. The calibration laboratories will be provided with artifacts with their characterization by National Physical Laboratory, New Delhi or nodal laboratories appointed by NABL for the same objective or any other NMI s who are signatory to BIPM. 3.12 NABL shall nominate its accredited/applicant CABs for participation in APAC and other international Proficiency Testing program, as and when they are available. Following sequential preference shall be given to the laboratories: Accredited CABs not having participated in any other APAC PT program. CABs having unsatisfactory performance in a past PT program. Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 5/7

CAB with long period of accreditation Applicant CAB NABL shall co-ordinate all activities related to PT program with APAC and the participating CAB. Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 6/7

Annexure A Proficiency Testing Plan Form 18 CAB ID CAB Name Field of Accreditation Testing / Calibration / Medical 4 - Year Period of Participation Accredited Discipline(s) Group(s) under Discipline Proficiency Testing Participation Plans (Year wise) Results Remarks (for NABL use only) Note: CAB should ensure that PT plan covers the major parameters in each group under the relevant discipline For NABL use only Reviewed and Approved by: Issue No: 08 Issue Date: 07-Mar-2019 Amend No: -- Amend Date:-- Page No: 7/7

(NABL) NABL House Plot No. 45, Sector 44, Gurgaon - 122002, Haryana Tel. no.: 91-124-4679700 (30 lines) Fax: 91-124-4679799 Website: www.nabl-india.org