patients: When to do calibration?

Similar documents
A pilot assessment of the FloTrac TM cardiac output monitoring system

Minimally invasive cardiac output monitors

Minimally invasive cardiac output monitors

Frequently Asked Questions & Features and Benefits

Assessing Preload Responsiveness Using Arterial Pressure Based Technologies. Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC

Radboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children?

Chapter 3. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery

Perioperative monitoring of cardiac

Fluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration

Pulse Pressure Analysis

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery

PulsioFlex Patient focused flexibility

Tailored Volume Resuscitation in the Critically Ill is Achievable. Objectives. Clinical Case 2/16/2018

MONITORIA EN CUIDADO INTENSIVO DEL PACIENTE CON ENFERMEDAD HEPATICA DIEGO FERNANDO BAUTISTA R. INTERNISTA INTENSIVISTA

Hemodynamic Monitoring in Critically ill Patients in Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory

Button et al. using a thermistor-tipped catheter. The CO is calculated using an algorithm based on the area under the systolic part of the pressure wa

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for

PCV and PAOP Old habits die hard!

Sepsis Wave II Webinar Series. Sepsis Reassessment

Measurement of the cardiac output (CO) and

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor

THE MEASUREMENT of cardiac output (CO) is a parameter

Disclaimer. Improving MET-based patient care using treatment algorithms. Michael R. Pinsky, MD, Dr hc. Different Environments Demand Different Rules

Cardiac Output Monitoring - 6

Definition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill

Received: 7 Jun 2006 Revisions requested: 28 Jun 2006 Revisions received: 30 Aug 2006 Accepted: 21 Nov 2006 Published: 21 Nov 2006

The Comparison of Stroke Volume Variation and Arterial Pressure Based Cardiac Output with Standard Hemodynamic Measurements during Cardiac Surgery

Functional Hemodynamic Monitoring and Management A practical Approach

Invasive Cardiac Output Monitoring and Pulse Contour Analysis. Harshad B. Ranchod Paediatric Intensivist Chris Hani Baragwanath Hospital COPICON 2011

Goal-directed vs Flow-guidedresponsive

Monitoring of cardiac index (CI) after surgery for congenital heart

Cardiac Output Monitoring: Expense justified by outcome?

An algorithmic approach to the very high risk surgical patient

Fluid Resuscitation and Monitoring in Sepsis. Deepa Gotur, MD, FCCP Anne Rain T. Brown, PharmD, BCPS

ORIGINAL ARTICLE. Abstract. Introduction. Materials and Methods

The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients,

One monitor for the whole patient pathway

BMC Anesthesiology. Open Access. Abstract

Noninvasive continuous cardiac output monitoring in perioperative and intensive care medicine

Transpulmonary thermodilution in septic shock & ARDS

EVOLUCIÓN DE LA MONITORIZACIÓN CARDIOVASCULAR EN LA UCI

Impedance Cardiography (ICG) Method, Technology and Validity

Chapter 2. Performance of three minimally invasive cardiac output monitoring systems

Comparison of cardiac output derived from FloTrac TM /Vigileo TM and impedance cardiography during major abdominal surgery

Dr. F Javier Belda Dept. Anesthesiology and Critical Care Hospital Clinico Universitario Valencia (Spain) Pulsion MAB

DESIGNER RESUSCITATION: TITRATING TO TISSUE NEEDS

Evaluation of Stroke Volume Variation Obtained by the FloTrac /Vigileo System to Guide Preoperative Fluid Therapy in Patients Undergoing Brain Surgery

Prof. Dr. Iman Riad Mohamed Abdel Aal

Comparison of Calibrated and Uncalibrated Arterial Pressure Based Cardiac Output Monitors During Orthotopic Liver Transplantation

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter

ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE

What you need. When you need it. EV1000 Clinical Platform

Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis

CLINICAL PRACTICE Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery

The cornerstone of treating patients with hypotension,

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE

Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit

Continuous monitoring of cardiac output: why and how

Assessment of a smartphone app (Capstesia) for measuring pulse pressure variation: agreement between two methods A Cross-sectional study

CARDIAC OUTPUT Monitoring ANDY CAMPBELL JOURNAL CLUB NOV 2011

CARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery: pulse-contour analyses vs pulmonary artery thermodilution

Hemodynamic monitoring beyond cardiac output

Resuscitation 83 (2012) Contents lists available at ScienceDirect. Resuscitation

British Journal of Anaesthesia 101 (6): (2008) doi: /bja/aen277 Advance Access publication October 12, 2008

Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states

Hemodynamic Monitoring To Guide Volume Resuscitation

Fluid responsiveness and extravascular lung water

Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients

Nurse Driven Fluid Optimization Using Dynamic Assessments

PiCCO technology. Hemodynamic monitoring at the highest level

6/5/2014. Sepsis Management and Hemodynamics. 2004: International group of experts,

Making the Case For Less Invasive Flow Based Parameters: APCO + SVV. Patricia A. Meehan, RN, MS, CCRN (a) Education Consultant Edwards Lifesciences

Jarisch A. Kreislauffragen, Dünser et al. Critical Care 2013, 17:326 Sunday, March 30, 14

Prediction of fluid responsiveness in patients during cardiac surgery

Swan Ganz catheter: Does it still have a role? Daniel De Backer Department of Intensive Care Erasme University Hospital Brussels, Belgium

Sepsis: Identification and Management in an Acute Care Setting

The value of arterial pressure waveform cardiac output measurements in the radial and femoral artery in major cardiac surgery patients

Establishing goals of volume management in critically ill patients with renal failure

L: Cardiovascular. Saskatchewan Association of Licensed Practical Nurses, Competency Profile for LPNs, 3rd Ed. 107

The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults

PREVENT COMPLICATIONS IN MAJOR SURGERY

Perioperative Goal- Protocol Summary

Perioperative Goal- Protocol Summary

Admission of patient CVICU and hemodynamic monitoring

FLUID RESUSCITATION AND MONITORING IN SEPSIS PROTOCOLIZED VS USUAL CARE DEEPA BANGALORE GOTUR MD, FCCP ASSISTANT PROFESSOR, WEILL CORNELL MEDICAL

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations

How can the PiCCO improve protocolized care?

IV fluid administration in sepsis. Dr David Inwald Consultant in PICU St Mary s Hospital, London CATS, London

Hemodynamic Monitoring Pressure or Volumes? Antonio Pesenti University of Milan Italy

Actualités sur le remplissage peropératoire. Philippe Van der Linden MD, PhD

EFFECT OF EARLY VASOPRESSIN VS NOREPINEPHRINE ON KIDNEY FAILURE IN PATIENTS WITH SEPTIC SHOCK. Alexandria Rydz

3/14/2017. Disclosures. None. Outline. Fluid Management and Hemodynamic Assessment Paul Marik, MD, FCCP, FCCM

Respiratory systolic variation test in acutely impaired cardiac function for predicting volume responsiveness in pigs

Perioperative Fluid Management in ERPs

Sepsis Management: Past, Present, and Future

Accuracy of an autocalibrated pulse contour analysis in cardiac surgery patients: a bi-center clinical trial

Surviving Sepsis Campaign. Guidelines for Management of Severe Sepsis/Septic Shock. An Overview

At present, cardiac output is usually determined

Frequently asked questions and answers:

Bioreactance-based passive leg raising test can predict fluid responsiveness in elderly patients with septic shock

Transcription:

Acta Clin Croat 2011; 50:267-272 review Constant cardiac output MONITORING using the PiCCO AND LiDCO methods VERSUS PAK in septic patients: When to do calibration? Mladen Širanović, Josip Kovač, Aleksandar Gopčević, Mijo Kelečić, Marinko Vučić, Nataša Kovač and Bojan Rode University Department of Anesthesiology, Intensive Care and Resuscitation, Sestre milosrdnice University Hospital Center, Zagreb, Croatia Summary The accuracy of cardiac output measurement by two most widely used methods of less invasive hemodynamic monitoring and by the standard technique of thermodilution with pulmonary catheter was assessed. The measurements were carried out in septic surgical patients immediately after and between system calibrations. Study results showed satisfactory compatibility of measurements performed by the two methods and by pulmonary catheter in both phases, thus system calibration being recommendable in hemodynamically unstable septic patients. Key words: Cardiac output physiology; Monitoring, physiologic methods; Sepsis; Intensive care units Introduction Comparative cardiac output measurements by the standard Pulmonary Artery Cardiac Output (PAK- CO) technique using pulmonary catheter for continuous cardiac output monitoring and two novel real-time methods, Pulse Contour Cardiac Output (PiCCO) and Lithium Diluted Cardiac Output (LiDCO), were performed to assess compatibility of measurements in septic patients. In addition, we compared the level of compatibility immediately after and between continuous method calibrations, and assessed the need of calibration. Target use of the novel methods, and in some cases substitution of hemodynamic monitoring by use of pulmonary catheter with less invasive continuous methods appear to be justified by the less invasiveness of these methods and thus the lower rate Correspondence to: Mladen Širanović, MD, University Department of Anesthesiology, Intensive Care and Resuscitation, Sestre milosrdnice University Hospital Center, Vinogradska c. 29, HR-10000 Zagreb, Croatia E-mail: mladen0906@gmail.com Received June 8, 2009, accepted July 25, 2011 of complications. LiDCO and PiCCO are less invasive methods of hemodynamic monitoring based on cardiac output calculation according to the area under arterial curve, i.e. pulse contour methods. Both methods are calibrated with previous dilution (LiDCO) and thermodilution (PiCCO) technique 1-3. Upon calibration, Pulse CO software calculates cardiac output per pulse 4. Methods A total of 269 comparative cardiac output measurements using PAK and PiCCO, 203 consecutive measurements with PAK and LiDCO with 18 comparative measurements in the phase of PiCCO system calibration, and 21 measurements in the phase of LiDCO system calibration were performed in 11 septic patients aged 43-59 without previous history of cardiovascular disease indicators. Continuous measurements were done in the patient hemodynamically stable stage, and at the time of PiCCO system and LiDCO system calibration. Patients belonged to the group of surgical patients with severe sepsis defined Acta Clin Croat, Vol. 50, No. 2, 2011 267

according to the Sepsis Surviving Campaign (SSC), showing signs of the septic inflammatory response syndrome (SIRS) with organ failure and microbiologically proved sepsis. Disease severity was estimated by the Sepsis Organ Failure Assessment (SOFA) and Acute Physiology And Chronic Health Evaluation II (APACHE II) scores. The SOFA score was 11 and the APACHE II score 24. Cardiac output was determined by use of the PiCCO plus-pulsion technology (Pulsion, Germany) with Pulsion catheter Ø 1.7 mm, length 20 cm, inserted in femoral artery. System calibration was done with 15 ml saline, temperature <7 ºC, administered via central venous catheter placed in internal jugular vein. The measurement was repeated five times and the was taken for calibration 5. In the LiDCO method, the LiDCO Cardiac Sensor System with lithium electrode attached on radial artery was employed. Calibration was done by single administration of 0.3 mmol lithium chloride and washing with 20 ml saline 6,7. Continuous output monitoring by Q-vue PAK and Abbott Q-vue monitor served as reference monitoring. The Q-vue PAK 7.5 Fr was placed in pulmonary artery and its location corrected by radiological and clinical testing. In both cases, the PiCCO and LiDCO systems were analog plugged on the Dräger Infinity basic monitor for invasive pressure monitoring. Measurement results and numerical data were compared and processed by the Bland Altman method using the MedCalc software and by descriptive statistics including Wilcoxon method of ranked pairs using the Statistica 6 software 8. Results Upon statistical processing, measurement data were classified in Tables 1-4, reporting descriptive statistics of the data obtained on cardiac output measurement by use of the PAK-CO, PiCCO and LiDCO methods. Each table is followed by statistical analysis of the respective data (measurement) by Wilcoxon pair test and Bland Altman diagram of comparison of two methods. Table 1 shows results of statistical analysis of 269 pairs of cardiac output measurement by the PAK and PiCCO methods. Bland Altman analysis, used for comparison of measurement compatibility between the two methods, is shown in Figure 1. Table 2 shows descriptive statistics of measurements by the same methods immediately after PiC- CO system calibration. Figure 2 shows Bland Altman statistical analysis for pairs of PAK-CO and PiCCO measurements. In the next step, measurements done with lithium dilution method by calibration pulse contour technique and PAK-CO by the reference technique of pulmonary catheter measurement in the phase of calibration and between calibrations were compared. Table 1. Pairs of cardiac output measurement by PAK-CO and PiCCO pairs 269 PAK-CO PiCCO error of arithmetic 6.80 3.10 4.35 4.30 0.65 0.04 6.12 2.2o 3.73 3.57 0.64 0.03 Statistical analysis of the pairs of measurement by Wilcoxon pair test: PAK-CO and PiCCO Sample size = 269 Lowest = 3.1000 Highest = 6.8000 = 4.3000 95% CI for median = 4.1000 to 4.4000 Variable: PiCCO Sample size = 269 Lowest = 2.2000 Highest = 6.1200 = 3.5700 95% CI for median = 3.4900 to 3.6500 positive differences = 50 negative differences = 217 Large sample test statistics Z = 10.102348 Two-tailed probability P<0.0001 268 Acta Clin Croat, Vol. 50, No. 2, 2011

2.0 1.8 1.6 1.4 1.15 Ratio of PAK-CO and PiCCO upon calibration 1.10 1.05 1.2 1.0 1.00 0.8 0.6 0.4 3 4 5 6 7 Mean PAK-CO and PiCCO Fig. 1. Comparative measurement PAK-CO vs. PiCCO. Table 3 shows descriptive statistics for 203 pairs of simultaneous PAK-CO and LiDCO measurements. Figure 3 shows statistical analysis of measurements by Bland Altman method. Table 4 shows numerical results obtained by descriptive statistics for pairs of cardiac output measurements upon LiDCO system calibration with lithium chloride in comparison with measurements done by PAK. Figure 4 shows Bland Altman statistical analysis of pairs of cardiac output measurements by PAK-CO and LiDCO immediately upon calibration. 0.95 0.90 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2 Mean PAK-CO and PiCCO upon calibration Fig. 2. PAK-O vs. PiCCO upon calibration. Accordingly, Bland Altman statistical analysis of the results obtained on cardiac output measurement by PAK and by the new methods of PiCCO and LiDCO yielded good compatibility between the two study methods and PAK in all phases of measurement, upon calibration and between calibrations. A low, clinically acceptable limit of compatibility was recorded between the two methods and PAK. Graphic presentation shows that the results of measurement obtained by different methods compared according to from the (±1.96 SD) were grouped, falling in-between the limits, thus indicating good Table 2. Pairs of PAK-CO and PiCCO measurements immediately after PiCCO system calibration pairs 18 error of arithmetic PAK-CO PiCCO calibration 5.00 3.90 4.31 4.30 0.35 0.08 5.00 3.70 4.11 4.00 0.40 0.09 Statistical analysis of pairs of measurements by Wilcoxon test: PAK-CO and PiCCO immediately upon calibration Sample size = 18 Lowest = 3.9000 Highest = 5.0000 = 4.3000 95% CI for median = 4.0000 to 4.5309 Variable: PiCCO immediately upon calibration Sample size = 18 Lowest = 3.7000 Highest = 5.0000 = 4.0000 95% CI for median = 3.7691 to 4.2926 positive differences = 1 negative differences = 16 Smaller total of ranks = 16.50 Two-tailed probability P=0.0026 Acta Clin Croat, Vol. 50, No. 2, 2011 269

Table 3. Pairs of cardiac output PAK-CO and LiDCO measurements pairs 203 PAK-CO LiDCO error of arithmetic 9.60 6.50 8.06 8.00 0.54 0.03 9.20 6.20 7.81 7.80 0.59 0.04 Statistical analysis of the pairs of measurements by Wilcoxon test: PAK-CO and LiDCO Sample size = 203 Lowest = 6.5000 Highest = 9.6000 = 8.0000 95% CI for median = 7.9000 to 8.0000 Variable: LiDCO-CO Sample size = 203 Lowest = 6.2000 Highest = 9.2000 = 7.8000 95% CI for median = 7.7000 to 7.8000 positive differences = 40 negative differences = 158 Large sample test statistics Z = 8.478611 Two-tailed probability P<0.0001 compatibility of the methods compared. Based on the Wilcoxon rank method and pair comparison, PAK to LiDCO comparison yielded better compatibility of results immediately upon calibration, whereas PAK to PiCCO comparison showed better result compatibility between calibrations (P<0.001 both). Discussion Cardiovascular instability is a common cause of morbidity and mortality in critically ill patients. Appropriate choice of therapy requires thorough understanding of the pathophysiological impairments, which can frequently be revealed by use of advanced Table 4. Pairs of cardiac output measurements by PAK-CO and LiDCO immediately upon calibration pairs 21 PAK CO LiDCO calibration error of arithmetic 9.0 6.50 8.03 8.10 0.53 0.11 8.80 6.80 7.89 7.90 0.45 0.09 Statistical analysis of the pairs of measurement by Wilcoxon test: PAK-CO and LiDCO immediately upon calibration Sample size = 21 Lowest = 6.5000 Highest = 9.0000 = 8.1000 95% CI for median = 7.9000 to 8.3000 Variable: LiDCO_upon calibration Sample size = 21 Lowest = 6.8000 Highest = 8.8000 = 7.9000 95% CI for median = 7.8000 to 8.0171 positive differences = 2 negative differences = 17 Smaller total of ranks = 17.50 Two-tailed probability P=0.0008 270 Acta Clin Croat, Vol. 50, No. 2, 2011

1.10 1.05 1.00 0.95 0.90 Ratio of LiDCO-CO and PA-CO PAK_COupon 0.85 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 Fig. 3. PAK-CO vs LiDCO. hemodynamic monitoring. The methods applied cannot influence the outcome, but do represent a tool for active intervention, depending on the data collected. With this in mind, we tried to compare cardiac output measurement by use of the two novel pulse contour methods, PiCCO and LiDCO, with the reference PAK-CO method as a gold standard, immediately upon calibration and between calibrations. The results obtained were consistent with literature reports comparing the two methods (PiCCO and LiDCO) with standard measurement by the PAK thermodilution method 9-11. However, there are not many studies comparing these methods in septic patients. In comparison with PAK, both methods showed good compatibility in the results of cardiac output measurement immediately upon calibration and between cali- 0.5 0.4 0.3 0.2 0.1 0.0-0.1-0.2-0.3-0.4 Fig. 4. PAK-CO vs. LiDCO Mean LiDCO-CO and PAK-CO Ratio PAK-CO-LiDCO upon calibration 6.5 7.0 7.5 8.0 8.5 9.0 9.5 Mean PAK-CO-LiDCO-CO upon calibration brations. A recent consensus conference on the efficacy and utility of the new generation of less invasive hemodynamic monitoring in critically ill patients pointed to the paucity of data and studies in the field 12. Most validation studies of pulse contour methods were conducted in patients undergoing cardiac surgery, whereas data on septic patients are rather insufficient 13,14. Therefore, our study included septic patients having undergone non-cardiac surgery and free from any history of cardiac disease, thus eliminating the impact of cardiac disease on measurement results. Good compatibility and the trend of cardiac output hemodynamic measurement by the study methods enables a simpler and less invasive approach as well as faster acquisition of data, along with a lower risk of complications for the patient. Although thermodilution measurement by pulmonary catheter remains the gold standard of hemodynamic monitoring, these new methods, even though inadequately evaluated as yet, have gradually taken their place in daily routine. Conclusion In this preliminary study, the LiDCO and PiCCO methods of cardiac output measurement were compared with the reference PAK technique of continuous cardiac output monitoring and showed a good level of compatibility in septic patients. Therefore, LiDCO and PiCCO as less invasive methods, thus associated with a lower rate of complications, could replace PAK in continuous real-time monitoring of cardiac output in septic patients. When using either of these methods in hemodynamic monitoring of hemodynamically stable septic patients, calibration can be done at longer intervals, thus decreasing the risk of complications associated with manipulation and reducing the cost of treatment, while having a beneficial effect on the quality of patient management. References 1. Heller LB, Fisher M, Phanselter N. Continuous intraoperative cardiac output determination with arterial pulse analysis Anesth. Analg 2002;93:SCA 1-SCA 112. 2. Rhodes A, Suderland R. Arterial pulse power analysis. The LiDCO plus System. Update in intensive care and emergency medicine. Berlin-Heildelberg: Springer-Verlag, 2005,42:183-92. Acta Clin Croat, Vol. 50, No. 2, 2011 271

3. Linton R, Band D, O Brien T, et al. Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med 1997;25:1796-800. 4. Aronson A, Heller L, Jayakar D, Jeevanandam V. Continuous intraoperative cardiac output determination with arterial pulse analysis (Pulse CO). The Society of Cardiovascular Anesthesiologists Annual Meeting, New York, NY, May 2002. 5. Hamilton T, Lyne M, Jeesen ME. Pulse CO a less invasive technique to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg 2002;74:S1408-12. 6. Pinsky MR, Payen D. Functional hemodynamic monitoring. Berlin, Heidelberg: Springer-Verlag, 2004, 42:135-53. 7. Jonas M, Bruce R, KNIGHT J, KELLY F Comparison of cardiac output measurements using a continuous arterial waveform analysis monitor with an indicator dilution technique (LiDCO) in intensive care Presentation at the ICCM. Sydney: October, 2001. 8. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1(8476):307-10. 9. Orme RM, Pigott DV, Mihm FG. Measurement of cardiac output by transpulmonary arterial thermodilution using a long radial artery catheter. A comparison with intermittent pulmonary artery thermodilution. Anaesthesia 2004;59:590-4. 10. Della Rocca G, Costa MG, Coccia C. Continuous and intermittent cardiac output measurement; pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002;88:350-6. 11. Linton RA,YOUNG LE,MARLIN DJ et al. Cardiac output measurement by lithium dilution and trans pulmonary thermodilution in patients in a paediatric intensive care unit. Intensive Care Med 2000, 26 (10):507-11. 12. Antonelli M, Azoulay E, Bonten M, Chastre J, Citerio G, Conti G, De Backer D, Lemaire F, Gerlach H, Groeneveld J, Hedenstierna G, Macrae D, Mancebo J, Maggiore SM, Mebazaa A, Metnitz P, Pugin J, Wernerman J, Zhang H (2008) Year in review in Intensive Care Medicine, 2007. II. Haemodynamics, pneumonia, infections and sepsis, invasive and non-invasive mechanical ventilation, acute respiratory distress syndrome. Intensive Care Med 34:405-422 13. 3Godje O, Hoke K, Goetz AE, Felbinger TW, Reuter DA. Reliability of a new algorithm for continuous cardiac output determination by pulse contour analysis during hemodynamic instability. Crit Care Med 2002;30:52-8. 14. Button D, Weibel L, Reuthenbuch O, Genoni M, Zollinger A. Clinical evaluation of the Flo-Trac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 2007;99:329-36. Sažetak STALNI NADZOR SRČANOG IZBAČAJA METODAMA PiCCO I LiDCO PREMA PAK-u U SEPTIČNIH BOLESNIKA: KALIBRIRATI ILI NE? M. Širanović, J. Kovač, A. Gopčević, M. Kelečić, M. Vučić, N. Kovač i B. Rode U ovom radu uspoređivala se točnost mjerenja srčanog izbačaja pomoću dviju najčešće upotrebljavanih metoda manje invazivnog hemodinamskog nadzora i standardnom tehnikom termodilucije plućnim kateterom. Mjerenje je provedeno kod septičnih kirurških bolesnika u razdoblju neposredno nakon i između kalibracija sustava. Rezultati su pokazali zadovoljavajuću podudarnost mjerenja u obje faze primjenom obiju metoda i pomoću plućnog katetera, pa se kalibriranje sustava preporuča u hemodinamski nestabilnih septičnih bolesnika. Ključne riječi: Srčani izbačaj fiziologija; Nadzor, fiziološki metode; Sepsa; Jedinice intenzivne skrbi 272 Acta Clin Croat, Vol. 50, No. 2, 2011