Investigating the recheck rules for urine analysis in children

Similar documents
It s not just water! What is Urinalysis?

GENERAL URINE EXAMINATION (URINE ANALYSIS)

PRINCIPLE OF URINALYSIS

LABORATORY 5: The Complete Urinalysis

Application Note. Light Microscopic Analysis of Urine ZEISS Primo Star and ZEISS Axio Lab.A1

2008 CAP TODAY Q & A

A. SAP is the D-Lab's name for a specific set of serum biochemical tests.

LABORATORY 3: Microscopic Urinalysis

Urinalysis and Body Fluids CRg. Urine Casts. Microscopic Sediment Casts. Unit 2; Session 6

Squamous epithelial cells in urine 0-5

Michelle Moy, MAd Ed, MT(ASCP)SC Program Director Clinical Laboratory Science Program Loyola University Chicago, Illinois

Microscopic Examination of Urine

BASIC METABOLIC PANEL

Andrea Tessari Microbiology Unit, Hospital of Rovigo, ULSS 18 Rovigo (Italy)

Automated Urinalysis Evaluation of the Sysmex UF-50

Microscopic Sediment Miscellaneous

Light yellow to dark golden yellow Clear ph range Specific gravity Sediments

Microscopic Sediment Epithelial Cells

Poor Predictive Ability of Urinalysis and Microscopic Examination to Detect Urinary Tract Infection

Analytic Performance of the iq200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers

Evaluation of frequency of abnormal Urine R.E tests in Pathology Laboratory

A. History Urinalysis is the oldest lab test still being performed today

Journal of Global Pharma Technology

Role of Quantitative MA Testing in Everyday Practice

MODULE 5: HEMATURIA LEARNING OBJECTIVES DEFINITION. KEY WORDS: Hematuria, Cystoscopy, Urine Cytology, UTI, bladder cancer

MLS Continuing Education Conference November PACE Session # Urinary Casts: The Importance of Laboratory Identification

SediVue Dx Urine Sediment Analyzer. Revolutionary urinalysis. Load-and-go efficiency. High-resolution digital images.

Evaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing

Taking a dip into urinalysis

References Required document for Laboratory Accreditation by the College of American Pathologists.

URINANLYSIS. Pre-Lab Guide

How to interpret your urine sample results

CLINITEK Novus Automated Urine Chemistry Analyzer

Assisting in the Analysis of Urine. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.

Introduction to Clinical Diagnosis Nephrology

Dipstick Urinalysis as a Test for Microhematuria and Occult Bladder Cancer

Surveillance report Published: 7 July 2016 nice.org.uk. NICE All rights reserved.

Disorders of the kidney. Urine analysis. Nephrotic and nephritic syndrome.

PHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE

Diabetic Nephropathy

URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST. Špela Borštnar UREX 2015, Ljubljana, Slovenia

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

September Focus: In-House vs. Reference Lab Testing, Part II

How to interpret your urine sample results

SediVue Dx Urine Sediment Analyser. Fresh samples / Revolutionary technology / A new standard of care. The Complete Diagnostic Solution

Role of Quantitative MA Testing in Everyday Practice

Dipstick Testing of Urine Can It Replace Urine Microscopy?

The Minimum Diagnostic Database: Urinalysis

SCBM343- Urinalysis. Associate Professor Dr. Wannee Jiraungkoorskul. Department of Pathobiology, Faculty of Science, Mahidol University

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter

Overview of glomerular diseases

Children s Services Medical Guideline

Urinary System. Analyze the Anatomy and Physiology of the urinary system

Detail PRINCIPLE: Body fluids other than blood and urine will be analyzed according to their site of origin, and the providers specific orders.

Multiple Choice Questions and Answers

Separation of Plasma and Serum and Their Proteins from Whole Blood

AN APPROACH TO HEMATURIA. Dr Saima Ali

CONCORD INTERNAL MEDICINE CHRONIC KIDNEY DISEASE PROTOCOL. Revised May 30, 2012

Paediatrica Indonesiana. Urine dipstick test for diagnosing urinary tract infection

Analysis of regulatory T cell subsets in the peripheral blood of immunoglobulin A nephropathy (IgAN) patients

Urinary System Part of the Excretory System

URINARY CRYSTALS. by Geoffrey K. Dube and Robert S. Brown

Clinical Test Report. of DUS10 (Urine Reagent Strips) Effective Date: April DFI Lab. Dong-Ai Hospital Medical Center: Clinical Pathology

Hematuria. Ramzi El-Baroudy (ESPNT)

Non-protein nitrogenous substances (NPN)

instrument. When 13C-UBT positive value is greater than or equal to / - 0.4, the the subject can be 1. Data and methods details are as follows:

Significance of Anti-C1q Antibodies in Patients with Systemic Lupus Erythematosus as A Marker of Disease Activity and Lupus Nephritis

BCH472 [Practical] 1

H(a)ematuria. FX Keeley Consultant Urologist Bristol Urological Institute

idc 20 AUTOMATED LIQUID-BASED CYTOLOGY SLIDE PREPARATION SYSTEM

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

우리나라의학교소변검사프로그램영남대학교의과대학소아청소년과

Peritoneal Dialysis. Choosing your logo. V2.0 logos. information. you can trust. Certified Member. The Information Standard

URINARY TRACT INFECTIONS 3 rd Y Med Students. Prof. Dr. Asem Shehabi Faculty of Medicine, University of Jordan

Urinalysis Review and Case Studies

BIOCHEMISTRY OF BLOOD

MEDICAL HISTORY. 23-Jan-2018 to 23-Jan VCA Miller-Robertson Animal Hospital 8807 Melrose Ave, Los Angeles, CA (310)

Urinalysis (Testing Urine Samples)

Pediatric GU Dysfunction

By Sandipkumar Kanazariya. Tuesday, December 11,

Dr P Sigwadi 30 May 2012

The Medical Laboratory

Color: Gray/Yellow. 5/7/2018 L Hematology results from IDEXX VetLab In-clinic Laboratory Requisition ID: 0 Posted Final Test Result Reference Range

Urine Sediment Photomicrographs/Photographs

URINALYSIS/URINE CHEMISTRIES

International Journal of Food Nutrition and Safety, 2012, 1(2): International Journal of Food Nutrition and Safety

Literature List Body Fluids

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

Preanalytics of urine sediment examination: effect of relative centrifugal force, tube type, volume of sample and supernatant removal

URINARY TRACT INFECTIONS

Pathology and Laboratory

The Urinary System. Lab Exercise 38. Objectives. Introduction

Chapter 17 Worksheet Code It

Study of culture and sensitivity pattern of urinary tract infection in febrile preschool children in a tertiary care hospital

Topic Objectives. Physical Examination LECTURE MODULE 3; PHYSICAL EXAMINATION OF URINE. Appearance. Odor Specific Gravity Volume

Urine analysis. By Dr. Gouse Mohiddin Shaik

Transcription:

Investigating the recheck rules for urine analysis in children Y.M. He*, S.W. Yao*, Y.J. Huang, B.S. Liang and H.Y. Liu Clinical Laboratory, Guangzhou Women and Children s Medical Center, Guangzhou Medical University, Guangzhou, China *These authors contributed equally to this study. Corresponding author: H.Y. Liu E-mail: haiyingliud@163.com Genet. Mol. Res. 15 (2): gmr.15027349 Received July 30, 2015 Accepted December 7, 2015 Published April 25, 2016 DOI http://dx.doi.org/10.4238/gmr.15027349 ABSTRACT. The aim of this study was to establish recheck rules of urinalysis in children by investigating the concordance rate of the results obtained using the LabUMat urine dry chemistry analyzer (referred to as dry chemistry) and the UriSed tangible composition analyzer with that of the microscopic examination. First, 1040 urine samples from children (mean age 6.5 years) were analyzed using LabUMat and UriSed analyzers, and subsequently subjected to microscopic examination. The missed detection rate was evaluated and recheck rules were established to avoid missed diagnoses of abnormal renal function. Finally, clinical validations of the recheck rules were performed on 200 additional specimens. Among the samples used to investigate the recheck rules, the samples with positive microscopic examination results accounted for 58.65% of the total, while the samples with negative results accounted for 41.35%. Of the positive samples, a major portion (>50%) were RBC positive. The samples that were WBC positive and CAST positive accounted for 23.08 and 7.69%, respectively. The concordance rate was 87.5% and the missed detection rate was 2.9%. For the validation of the recheck rules in 200 urine samples, the concordance rate was 87.5% and the missed detection rate was 2.4%. When the detection of

Y.M. He et al. 2 occult blood, WBC, and protein by dry chemistry, and the detection of RBC, WBC, and CAST by the UriSed analyzer are inconsistent, or the differences between them greater than 2 levels, recheck by microscopic examination is suggested. Key words: Automatic urine analyzer; Microscopic examination; Recheck rules INTRODUCTION Since urine specimens are easy to obtain, preserve, and are non-invasive, urine samples are used for routine clinical testing. However, as the number of laboratory specimens increased, the traditional manual microscopic examination methods for urine sediments were unable to meet the current clinical demands. In line with advancements in medical test technology, automatic analyzers used to detect urine-formed elements are considered important for routine urine testing, because these tests are rapid and less labor intensive. As such, automated urine analyzers have become increasingly popular among the healthcare professionals. However, due to urine characteristics including complexity, diversity, propensity to change in children (Gu, 2009), and technological limitations, current clinical laboratory automated urine analyzers cannot completely replace microscopic examinations (Li et al., 2007). Recheck guidelines for urinalysis in children are rarely reported. Therefore, due to the changing urine characteristics in children, there is a need to explore suitable recheck criteria for the use of automated urine analysis. MATERIAL AND METHODS Sample source In total, 1040 urine samples (640 males and 400 females) from outpatients and inpatients (without regard to clinical department) were random collected. Patient age ranged from 1 month to 13 years with an average age of 6.5 years. Another 200 urine samples were randomly collected for investigation of recheck rules. Instruments and materials Instruments and materials included the LabUMat urine dry chemistry analyzer (referred to as dry chemistry) (77 Elektronika, Budapest, Hungary), UriSed tangible composition analyzer (77 Elektronika), test strips and sediment supporting plate, Japan Olympus Binocular Microscope (Tokyo, Japan), and a standard laboratory centrifuge. Automated analysis Each day, 20 to 30 fresh urine specimens were randomly selected. The LabUMat automated urine analyzer was used for testing dry chemical and physical compositions of the urine. Original test results, which included urine dry chemical occult blood (BLD), leukocytes (LEU), protein (PRO), urinary sediment red blood cells (RBC), white blood cells (WBC), and CAST (urinary casts or tubes) were saved. Automated analyses of all samples were completed within 2 h.

Recheck rules for urine analysis in children 3 Microscopic examination The samples, which were analyzed by the automated analyzer, were subjected to microscopic examinations carried out by two experienced competent technicians in a doubleblind manner. In accordance with the 3rd Edition of National Clinical Laboratory Procedures (Ye et al., 2006), the mean of the two testing results was used as the result of the microscopic examination. Positive standards for microscopic examination of urinary sediment used herein were as follows: RBC > 3/hydroxylysyl pyridinoline (HP); WBC > 5/HP; and CAST > 1/ lysyl pyridinoline (Ye et al., 2006). The levels of the results detected by the three methods are shown in Table 1. Table 1. Results of the three detection methods for each level. Level RBC or BLD WBC or LEU Urised (number/µl) Dry chemistry Microscope (number/µl) UriSed (number/µl) Dry chemistry Microscope (number/µl) Level 1 0.00-5.00-0.00-3.00 0.00-9.00-0.00-5.00 Level 2 5.01-40.00 1+ 3.01-10.00 9.01-45.00 1+ 5.01-15.00 Level 3 40.01-80.00 2+ 10.01-20.00 45.01-100.00 2+ 15.01-30.00 Level 4 80.01-150.00 3+ 20.01-40.00 100.01-150.00 3+ 30.01-50.00 Level 5 >150 4+ >40.00 >150 4+ >50.00 RESULTS Quality control of artificial microscopic detection The criterion of microscopic examination was compared by a chief technician and competent technicians in clinical tests every week and the concordance was greater than 95%. Assessment of the automated urine analyzer Based on microscopic examination standards for RBC, WBC, and CAST, the sensitivity, specificity, and concordance rate detected by the LabUMat automated urine analyzer were assessed and are shown in Table 2. Among the samples used to investigate the recheck rules, the samples with positive microscopic examination results accounted for 58.65% of the total, while the samples with negative results accounted for 41.35%. Of the positive samples, a major portion (>50%) were RBC positive. The samples that were WBC positive and CAST positive accounted for 23.08 and 7.69%, respectively. The concordance rate was 87.5% and the missed detection rate was 2.9%. The microscopy recheck criteria of urine for children were as follows: if the results of the BLD, LEU, PRO by dry chemical analyzer, and the RBC, WBC, and CAST by UriSed sediment analyzer were inconsistent, or if there were two or more levels of difference between the results, they should be rechecked. Additionally, when the instruments prompt a review of the pictures, they should be rechecked. Complete recheck rules for urine analysis are shown in Table 3. Validation of recheck rules About 200 urine samples were randomly selected to be verified; the compliance rate was 87.5% and the missing rate was 2.4%.

Y.M. He et al. 4 Table 2. Sensitivity, specificity, and concordance rate of RBC, WBC, and CAST detected using an automated urine analyzer. Threshold value Sensitivity (%) Specificity (%) Coincidence rate (%) RBC 5.00 94.23 84.62 96.15 WBC 9.00 73.08 98.72 92.31 CAST 2.00 75.00 90.63 89.42 Table 3. Automated urine analyzer recheck criteria. Rule number LabUMat automatic analyzer recheck criteria Number of cases 1 LabUMat dry chemical BLD positive, UriSed RBC positive 400 2 LabUMat dry chemical BLD positive, UriSed RBC positive 60 3 LabUMat dry chemical BLD negative, UriSed RBC positive 10 4 LabUMat dry chemical LEU negative, UriSed WBC positive 110 5 LabUMat dry chemical BLD positive, UriSed RBC negative 130 6 LabUMat dry chemical LEU positive, UriSed WBC negative 30 7 LabUMat dry chemical PRO positive, UriSed CAST positive 110 8 LabUMat dry chemical PRO positive, UriSed CAST negative 170 9 LabUMat dry chemical PRO negative, UriSed CAST positive 90 10 LEU, BLD, PRO and corresponding the WBC, RBC, CAST a corresponding order of magnitude difference level 2 above 350 11 Picture prompt review instrument 120 DISCUSSION There are many limitations of existing tangible component analysis technology of urine from children including: asymptomatic proteinuria or hematuria, and low molecular weight proteinuria not forming tubes in the urine due to child hyperactivity; child urine secretions are fewer than those of adults; and child urine sample complexity, diversity, and propensity to change in vitro. The pathological changes of occult nephritis in children are not the same as adults, and can be relatively mild, especially in simple hematuria, which may be related to the young age of children and hence the early stage of disease. Therefore, it is necessary to set recheck rules of child urinalysis to avoid false negatives and missed diagnoses of child urinary system diseases and reduce illness delay, which would aid in improving work quality and efficiency, as well as meet clinical needs. The LabUMat automatic tangible component analyzer is a new automated urine analyzer with the following benefits: it provides a clear panoramic view of urine samples through microscope imaging technology; it utilizes neural network control technology; and it has intelligent automatic recognition functions, which together provides results that are intuitive and visible. Additional tangible components that cannot be accurately identified by the instrument can be identified and amended manually to obtain accurate and reliable results. In the present study, dry chemistry was used to detect BLD, LEU, PRO, as well as tangible components such as RBC, WBC, and CAST via an automated urine analyzer. These were detected in accordance with rules for urine sediment microscopic examination, which was used as the reference method. With these guidelines, the missing rate was 2.9%, which meets the clinically acceptable false-negative rate of less than 5% (Barnes et al. 2005). Of the 61 false-negative samples, most were from children with nephrotic syndrome, purpura kidney disease, or lupus erythematosus, whereas few were from children with influenza, hemangiomas, or peritonitis. More specifically, 3 to 10 RBC false-negative cases were found in one HP, and microscopic examination revealed that majority were due to the RBC profile

Recheck rules for urine analysis in children 5 or crystallization; 5 to 15 WBC false-negative cases were found in one HP, and microscopic examination found that most of them were caused by WBC deformation, too many epithelial cells, or other components; CAST false-negative cases were primarily from patients with fever, diarrhea and vomiting, or peritonitis. Recheck results were in positive critical value. There were few false negatives due to the amorphous effect, which cannot be recognized by the instrument. In false-positive samples, many were misidentified as pathological tubes due to cloudy urine, amorphous crystals, bacteria, or impurities-bacterial-cell adhered by mucus. Those misidentified as RBC were due to calcium oxalate crystals, urate, phosphate, yeast-like fungi, or fat droplets. Interference factors of WBC detection were relatively fewer than those of the RBC, and there was a high concordance rate with the microscopic examination. However, when epithelial cells are present simultaneously with WBC, they can affect each other, causing identification errors in the instrument, i.e., the instrument is unable to differentiate between small round epithelial cells that are similar to WBC in size, or WBC and deformed WBC that are similar to the epithelial cells. Since the LabUMat automatic tangible component analyzer provides a real-time panoramic view, manual correction can be performed via the user interface directly, greatly reducing the rate of re-examination, which can increase quality and improve efficiency. Automated detection of urine tangible components is important for diagnosis and prognosis of urinary system diseases, especially when the specimen is not suitable for screening or direct microscopic examination (Cong, 2011). Since urine testing was affected by many factors in this study, in the following circumstances, a recheck is highly recommended: if the BLD, LEU, or PRO detected by dry chemistry does not match the RBC, WBC, or CAST detected by UriSed tangible component analyzer, or there are differences that are greater than two levels between them and the instrument prompts review of the pictures; if the specimens of urinary system diseases need re-examination to prevent missed diagnosis; and lastly, if there is a large difference between current and prior results. For severe hematuria, pyuria, or severe crystallized urine, automated detection should be performed after sample dilution to obtain accurate and reliable results. The missing rate reported herein was slightly different from those previously reported in the literature (Chen et al., 2011; Yang et al., 2012), although it was similar to others (Ma et al., 2011; Li et al., 2014), which may be related to the source of the samples, rule settings, and different instruments. The samples in this study were from children, and pathological changes of occult nephritis in children are relatively mild and are not the same as adults, so clinical symptoms were not obvious. Therefore, laboratories should develop recheck rules that are applicable to their own circumstances, and continue to perform clinical validation and adjustment, which can greatly improve the accuracy of the results and work efficiency. Conflicts of interest The authors declare no conflict of interest. REFERENCES Barnes PW, McFadden SL, Machin SJ and Simson E (2005). The International Consensus Group for Hematology review: suggested criteria for action following automated CBC and WBC differential analysis. Lab. Hematol. 11: 83-90. http://dx.doi.org/10.1532/lh96.05019

Y.M. He et al. 6 Chen Y, Cheng M, Li W, Du J, et al. (2011). Development and clinical application of microscopy review criteria of automated urine chemistry and sedimental analysis. Chin. J. Lab. Med. 34: 501-506. Cong YF (2011). Examination of the urine sediment and establishment of the criterion for microscopy review. Chin. J. Lab. Med. 34: 481-483. Gu KL (2009). Difficulties and doubts of examination of urine formed elements. Chin. J. Lab. Med. 32: 605-608. Li CB, Fan J, Dong ZJ, Hong W, et al. (2014). Development and application of microscopy review criteria of urinalysis combination of dry chemistry and flow cytometry. Chin. J. Gen. Pract. 13: 116-119. Li W, Wang WH and Liu C (2007). Clinical applications of urine visible ingredients by three urine sediment detective methods. J. Lab. Med. 28: 342-344. Ma JL, Lu YJ, Li XC, Wang SJ, et al. (2011). Establishment and evaluation of the review criteria of automatic urine analysis workstations. Chin. J. Lab. Med. 35: 810-814. Yang XM, Li GD and Zhao YL (2012). Comparison of the results of two different automatic urinalysis systems and construction of re-examination rules. J. Lab. Med. 33: 671-675. Ye YW, Wang YS and Shen ZY (2006). National Clinical Laboratory Procedures: Southeast University Press, Nanjing, 275-276.