A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Similar documents
A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Dosimetric Characterization of a Prototype Nanotechnology Microbeam Radiation Therapy Device using Gafchromic EBT2 Film

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Verification of micro-beam irradiation

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Water equivalent PRESAGE for synchrotron radiation therapy dosimetry

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Y FILMS DOSIMETR Nederland België / Belgique

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Limits of Precision and Accuracy of Radiation Delivery Systems

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Hampton University Proton Therapy Institute

Radiochromic film dosimetry in water phantoms

Small field diode dosimetry

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned


Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers


Advanced film dosimetry for a UK brachytherapy audit

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Activity report from JCOG physics group

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Elekta Synergy Digital accelerator for advanced IGRT

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosisverifikation mit Delta 4 Discover während der Behandlung

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

IORT with mobile linacs: the Italian experience

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Small field dosimetry for the small animal radiotherapy research platform (SARRP)

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

A new geometric and mechanical verification device for medical LINACs

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

IN VIVO IMAGING Proton Beam Range Verification With PET/CT

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

IMRT QA: Point Dose Measurements or 2D Array?

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Unrivaled, End-to-End

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Diode calibration for dose determination in total body irradiation

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Safety issues related to the Synchrotron Stereotactic Radiation Therapy project at the ESRF

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The Journey of Cyberknife Commissioning

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

EORTC Member Facility Questionnaire

Basic dosimetric data for treatment planning system

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Commissioning and Radiobiology of the INTRABEAM System

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

RPC s Credentialing Programs for Clinical Trials

Dosimetry and QA of proton and heavier ion beams

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Quality assurance in external radiotherapy

Leila E. A. Nichol Royal Surrey County Hospital

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams.

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

Efficient Dosimetry for Proton Therapy

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

I. Equipments for external beam radiotherapy

Transcription:

Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital of Shanghai Jiaotong University, Shanghai 200233, China Abstract Synchrotron radiation (SR) represents a unique and innovative anti-cancer treatment due to its unique physical features, including high flux density, and tunable and collimated radiation generation. The aim of this work is to assess the dosimetric properties of SR in Shanghai Synchrotron Radiation Facility (SSRF) for potential applications to clinical radiation oncology. The experiments were performed with 34 and 50 kev X-rays on the BL13W biomedical beamline of SSRF and the 6 MV X-rays from ARTISTE linac for the dosimetry study. The percentage depth dose (PDD) and the surface dose of the SR X-rays and the 6 MV photon beams were performed in solid water phantom with Gafchromic EBT3 films. All curves are normalized to the maximum calculated dose. The depth of full dose buildup is about 10 µm deeper for the monoenergetic X-ray beams of 34 and 50 kev. The beam transmits through the phantom, with a linear attenuation coefficient. The profile in the horizontal plane shows that the dose distribution is uniform within the facula, while the vertical profile shows a Gaussian distribution of the dose. The penumbra is less than 0.2 mm in the horizontal profile. Gafchromic EBT film may be a useful and convenient tool for dose measurement and quality control for the high space and density resolution. It is therefore important to gain a thorough understanding about the physical features of SR before this novel technology can be applied to clinical practice. Key words Radiation therapy, Synchrotron radiation, Physical features, Dosimetry 1 Introduction Radiation therapy, along with surgery and chemotherapy, is the primary option for cancer treatment. Radiotherapy has been used to treat 60% 70% of cancer patients. The goal of radiation therapy is to deliver the maximum dose to a tumor and render the minimum radiation damage to surrounding normal tissues. The existing techniques of radiotherapy, including conventional MV X-ray radiation therapy, have some limitations in terms of physical and radiobiological aspects. Therefore, it is imperative to develop new technologies to maximize the dose to a tumor and minimize radiation damage to surrounding normal tissues. Recent advances in synchrotron radiation (SR) have offered a new option to radiotherapy. The Shanghai Synchrotron Radiation Facility (SSRF), a third-generation synchrotron radiation light source available for the users since Supported by research grants from Shanghai Jiaotong University (No.YG2012ZD02), Science and Technology Commission of Shanghai (No.2JC1407400) and National Natural Science Foundation of China (Nos.81272506, and 61227017). * Corresponding author. E-mail address: shen_fu@hotmail.com Received date: 2013-04-05 060102-1 2009, provides an invaluable tool for cancer treatment. Two SR-based techniques (SRT) have been reported [1,2] : Stereotactic synchrotron radiotherapy (SSRT) and microbeam radiation therapy (MRT). The first patient was admitted for SSRT in late June 2012 at European Synchrotron Radiation Facility (ESRF), while MRT has been in pre-clinical radiosurgery dedicated to treating brain tumors initially [3]. As the biomedical beamline of SSRF is similar to the ESRF beamlines for developments in SSRT and MRT, in terms of beam intensity and stability, it shall be possible to deliver enough dose to a tumor target within a short period of time, for potential treatments of aggressive tumors with hypofractionated irradiations like SSRT in ESRF. Therefore, it is necessary to gain a thorough understanding about the parameters related to radiation therapy on the SSRF beamline, including the skin dose, dose rate and X-ray spectrum, before it can be applied to clinical practice. The purpose of the present study

was to investigate the physical features of SSRF-based synchrotron radiotherapy. 2 Materials and methods 2.1 General data The experiments were performed on the BL13W biomedical beamline of SSRF, which runs beam bunches of 3.5 GeV electrons circulating a 432-m circumference, and 34 and 50 kev X-rays were selected from the synchrotron radiations generated from a wiggler. Photon beams from a 6 MV ARTISTE linac (Siemens Corp.) were used, too, for the dosimetry study. Dosimetric verifications of the SR X- rays and the 6-MV photon beams were performed at different depths in the solid water (PTW Solid Water Phantom). radiochromic film. It was found that the film is a useful and convenient tool for dose measurement and the quality control of SR X-rays. 3.1 PDD of SSRF In radiotherapy, PDD is the key parameter for calculating the prescribed dose and the tissue or tumor absorption dose. A PDD curve relates the absorbed dose with medium depth along the axis of the beam. The doses are expressed in percentages of the maximum dose, referred to as D max. The doses are measured generally in water or water equivalent plastic in an ionization chamber, as water is similar to human tissue in radiation scattering and absorption. 2.2 Film dosimetry Medical radiation dosimetry can be performed with radiochromic films capable of producing a permanent visible color change upon irradiation. In this work, Gafchromic EBT3 radiochromic film (International Specialty Products, NJ, USA) was used for radiationinduced auto-developing X-ray analysis. The film has high spatial resolution (5 µm), high sensitivity and good uniformity of response (<1.5%). Also, it can be used for wide dose range (0.01 Gy to 40 Gy) measurements with X-rays in wide range energies (from kev to MeV). All irradiated films were scanned by an EPSON 10000XL scanner (EPSON, JAPAN), and analyzed by FilmQA-PRO software (International Specialty Products, NJ, USA). A calibration curve was established using the 6 MV X-rays. In the calibration, a 13-mm thick PMMA plate was used to cover the films backed by a 5-cm thick tough-water phantom material. 3 Results and discussion For feasibility study of clinical SSRT trials on the BL13W beamline at SSRT, it is essential to know its specific features, i.e. the dosimetric characteristics of linearity, reproducibility, uniformity, penumbra, surface dose and percent depth dose (PDD), which were investigated experimentally with the Fig. 1 Central axis depth dose in surface of the solid water phantom. Fig.2 Absolute transmission of 34 kev SR X-rays and conventional 6 MV X-rays in the solid water phantom. Low energy X-rays have a high surface dose with a shallow build-up depth. Fig.1 shows the central axis depth dose, normalized to the maximum dose calculated, in surface of the solid water phantom, obtained from X-ray film measurements with 34 kev or 50 kev X-rays on the BL13W beamline. The depth of full dose buildup is about 10 µm deeper for the monoenergetic X-ray beams of 34 and 50 kev. The results are similar to those obtained at ESRF [4]. The beam transmits through the phantom, with a linear attenuation coefficient [5]. This is consistent with the predicted results by simulations [6]. In Fig.2, the PDD of 34 kev X-rays is compared with X-rays from the 6 MV linac. The dose 060102-2

attenuation with 34 kev X-rays in the phantom is much quicker than that with the 6 MV X-rays. 3.2 Facula features Synchrotron radiation has better reproducibility and uniformity. We irradiated two films in different hours with 34 kev X-ray of the same parameters, and the difference in the absorbed dose was less than 4%. Fig.3 shows the facula size. It is about 4 cm in length and 4 mm in height. The profile in the horizontal plane (Fig.4) shows that the dose distribution is uniform within the facula, while the vertical profile (Fig.5) shows a Gaussian distribution of the dose. radiation therapy. For the 34 and 50 kev SR X-rays, the penumbra is so small that organ at risk (OAR) surrounding the tumor can be protected. Synchrotron X-rays are of extraordinarily high flux density and small divergence, with sharply defined beam edges at depths in the body. Compared with conventional radiotherapy, SR-based radiotherapy is able to deliver a higher fraction dose to the target within a shorter time. In conventional radiation therapy, tumor movements during irradiation consequently result in missing of the treated tumor. Because of the high flux density of synchrotron X-rays, the tolerance of tumor motion can be well within the limits of accuracy in SR-based radiotherapy, which ensures higher therapeutic effects than conventional irradiation. 3.3 Field expansion Fig.3 Very small facula of the 34 kev SR X-ray. Fig.4 A profile in the horizontal plane, showing good uniformity of the 34 kev SR X-ray. The facula center is of the brightest, indicating non-uniformity of the dose in the radiation field, hence the necessity of intensity modulated radiation therapy technique in the vertical plane. The X-ray beam available at the BL13W beamline has a maximum height of 4 mm, but the size of a tumor is usually larger than facula. A scanning method can expand the irradiation field. A tumor volume can be outlined according to CT images. Plumbum masks are made in the beam view to define the incoming beam to the tumor shape. Fig.6 shows an expanded irradiation field. Fig.7 is the profile. If two consecutive faculas are interlaced, the dose is overlapped and the high dose may cause injury to normal tissues. If the consecutive faculas are separated, the dose is not enough, and low dose may increase the risk of tumor recurrence. How to adjoin the facula is an important problem for SR-based radiation therapy at SSRF. Fig.5 A profile in the vertical plane, showing a maximum facula width of only 4 mm for the 34 kev X-ray. In X-ray therapy, metal or plastic blocks are used to improve the dose distribution. As shown in Figs.4 and 5, the dose curves are very sharp at the facula edges. The penumbra is less than 0.2 mm in the horizontal profile (Fig.4). Penumbra has a major impact on uniformity of isodose distributions in Fig.6 Field expansion. Fig.7 A profile in the vertical plane for the expanded irradiation field. 060102-3

3.4 Dose calculation and treatment planning A treatment planning system for ESRF ID17 clinical trials is on the way by collaboration with the CHU- INSERM team and the company DOSIsoft [7] for SSRT application only. A realistic simulation of this distribution is necessary for treatment planning and evaluating the experimental results. In SRT, tumors are irradiated with very high doses. High flux synchrotron X-rays allow for extremely rapid irradiation, devoid of collimation due to micro-movement fluctuations of the target. The absolute dose to the target needs to be calculated and delivered accurately. Monte Carlo based calculations have been carried out for SRT to calculate the dose distribution that is finally needed for treatment planning [8 10]. These data allow a description of the physical dose deposition, but the reliable prediction of the biological effect has not been achieved. Especially in terms of the size of the treatment volume, the knowledge is limited to small volumes in the brain. Accordingly, a real optimization of the treatment parameters is not yet feasible. To set up a treatmentplanning program for human patients, additional data for normal tissues and tumors are necessary [7]. 3.5 Dose profile evaluation For SRT, dosimetry tests should be done before and/or after patient treatment with respect to dosimetric films, TLDs or gel dosimetry for patient safety. Dosimetry of SRT beams should record the total absorbed dose to the tissue. The calculation result should be compared with the dosimetric and microdosimetric measurements using ionization chambers, MOSFET detectors, TLD or radiochromic films [11 15], knowing that the measurement readout ensures the irradiation quality. To get better agreement with the experimental data and predict a better dose in therapeutic application, it is necessary to further refine the MC simulation of SRT [4]. In addition, systematic analysis of the therapeutic gain for different microdosimetric patterns on the basis of the beam energy, beam shape and overall field size is still evolving for SRT of targets in experimental animals. 3.6 Quality assurance for SSRF In radiotherapy, quality assurance (QA) plays an important role in the workflow. SRT is a technique designed to deliver radiation therapy very precisely to tumors. The technology used in SRT allows external beam radiation to be delivered with pinpoint accuracy. However, delivery of treatment in an accurate and consistent manner is by no means easy to achieve, since the radiation therapy process is a complex interweaving of a number of related tasks for designing and delivering radiation treatment. QA of the radiation therapy equipment is primarily an ongoing evaluation of functional performance characteristics. Accurate radiotherapy determines the therapeutic effect, and high accuracy is an obligation for prescribing a precise dose. QA is a prerequisite for the treatment setting. QA in radiotherapy ensures consistency of the medical prescription, and safe fulfillment of the radiotherapy related prescription. For patient treatment, the accuracy of the location and dose should be checked frequently. Although the synchrotron accelerator has very high accuracy and precision, the mechanical isocenter should be checked aperiodicly. The center of the facula may shift after several years. 4 Conclusion Synchrotron radiation has many physical and biological advantages for radiation therapy, and related medical research plays an important role in radiation therapy. SR therapy represents a great potential radiation source to be applied in the treatment of cancer. The results of our trial show that the physical features of SSRF are similar to those of the light source abroad. We believe that SSRF will be applied in radiation oncology in the near future. However, the transition from animal experiments to clinical therapy in human patients appears to be more complex. Further dosimetric studies are necessary to better predict the dose in clinical applications. A dose measurement within a different tissue density phantom with other methods is required. The film can measure the continuous dose distribution in the plane with the high space and density resolution, which may be a useful and convenient tool for dose measurement and 060102-4

quality control. With such knowledge, it can be anticipated that advancements in the conventional tools will lead to the betterment of medical practice [16]. Acknowledgements The authors thank Professors Zhao Zhentang and Xiao Tiqiao with SSRF for their support to the research. References 1 Curtis H J. Radiat Res Suppl, 1967, 7: 250 257. 2 Baker C P, Curtis H J, Zeman W, et al. Radiat Res, 1961, 15: 489 495. 3 Serduc R, Christen T, Laissue J, et al. Phys Med Biol, 2008, 53: 3609 3622. 4 Siegbahn E A, Bräuer-Krisch E, Stepanek J, et al. Nucl Instrum Meth A, 2005, 548: 54 58. 5 Dilmanian F A, Romanelli P, Zhong Z, et al. Eur J Radiol, 2008, 68S: S129 S136. 6 Dilmanian F A, Zhong Z, Bacarian T, et al. Proc Natl Acad Sci USA, 2006, 103: 9709 9714. 7 Dilmaniana F A, Qu Y, Liu S, et al. Nucl Instrum Meth A, 2005, 548: 30 37. 8 Masami T, Yumiko O, Naoto Y, et al. Eur J Radiol, 2008, 68S: S114 S117. 9 Orion I, Rosenfeld A B, Dilmanian, F A. et al. Phys Med Biol, 2000, 45: 2497 2508. 10 Stepanek J, Blattmann H, Laissue J A, et al. Med Phys, 2000, 27: 1664 1675. 11 Bräuer-Krisch E, Bravin A, Lerch M, et al. Med Phys, 2003, 30: 583 589. 12 Ptaszkiewicza M, Braurer-Kirsch E, Klosowski M, et al. Radiat Meas, 2008, 43: 990 993. 13 Siegbahn EA, Bräuer-Krisch E, Bravin A, et al. Med Phys, 2009, 36: 1128 11137. 14 Brauer-Krisch E, Bravin A, Lerch M, et al. Med Phys, 2003, 30: 583 589. 15 Nariyama N, Ohigashi T, Umetani K, et al. Appl Radiat Isotopes, 2009, 67: 155 159. 16 Suortti P and Thomlinson W. Phys Med Biol, 2003, 48: R1 R35. 060102-5