Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Similar documents
Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE

Update on MQSA and Mammography Accreditation

Update on the ACR FFDM QC Manual

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)

HISTORY OF MQSA AND ACR

Implementation of a New Tomosynthesis Program: A Physicists Perspective

AAPM Annual Meeting. ACR Accreditation Update in CT

Mammography Quality Control: A Refresher

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

CT Quality Control Manual FAQs

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

NONE. Disclosures. Accreditation Update

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017)

Overview of the 2005 IOM Report Improving Breast Imaging Quality Standards

Multicenter Trial, Phase I Assessment of 2-D FFDM Versus Combo of 2-D and 3-D Tomosynthesis in Breast Cancer Screening

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

The Radiology Aspects

MAMMOGRAPHY QC: MORE THAN PHANTOM IMAGES. Stephanie Schofield RTR QC Technologist Nova Scotia Health Authority

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

ACR Accreditation Toolkit for Validation Site Surveys

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

FDA Executive Summary

ACR MRI Accreditation: Medical Physicist Role in the Application Process

MQSA Physicist Qualifications. Jon J. Erickson, Ph.D., DABR

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Challenges to Delivery of High Quality Mammography

Low Dose CT Lung Screening: What is Technically Required?

Prepublication Requirements

9/3/2015. New Mammographic Modality Training. Mammographer. Qualified Radiologic Technologist. Tomosynthesis Training

MRI FAQs. After reading these documents and checking your protocols, you can apply online here:

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

Please follow all instructions carefully.

History of MQSA ACR creates testing procedures Oct Mammography Quality Standards Act Dec 1993 Interim Regulations Oct All mammog

The Mammography Examination

Steven LaFontaine Position: Certification: State of California Approved Physicist List: ACR Accreditation Programs Qualified For: Education:

Radiation Physics Consultants, Inc. Medical Physicist. Mammography Qualifications, Continuing Experience and Continuing Education

Hong Kong College of Radiologists Mammography Statement

Breast Tomosynthesis. What is breast tomosynthesis?

Steven Tate Nicholas

POSITIONING ACR REQUIREMENTS IMAGE REVIEW CATEGORIES DEFICIENCIES IN POSITIONING ACR REQUIREMENTS 3/28/2016 NUMBER 1 REASON FOR ACR FAILURE

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

Mammography. What is Mammography?

25 TEXAS ADMINSTRATIVE CODE (TAC) (x) Certification of Mammography Systems and Mammography Machines Used for Interventional Breast Radiography

Debbie Childs RDMS, RVT Sonographer Murphy Medical Center Murphy, NC

Breast Imaging Essentials

Medical Diagnostic Imaging

Mammography. What is Mammography? What are some common uses of the procedure?

EFOMP School for Medical Physics Experts Prague 2015

Dosimetry in digital mammography

The power and promise of breast tomosynthesis is here. Selenia Dimensions system with Acquisition Workstation 5000

Curriculum Vitae Cal R. Schmidt, MS DABR. Contact Info: Phone: (612) Fax: (612)

Computed tomography Acceptance testing and dose measurements

Mammographic Accreditation Phantom

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

Agenda: Dental Cone Beam Imaging

Introduction and Background

Radiographic techniques in screen-film mammography

Accreditation Case Review: Mammography and Stereotactic Biopsy

Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care.

Nuts and Bolts of MRI Physics Tests or Selling Preparation H. Robert A. Bell, PhD. Learning Objectives

ACR Ultrasound Accreditation

ALARA in mammography screening. Hilde Bosmans et al.

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

Digital Breast Tomosynthesis from a first idea to clinical routine

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

IAEA TECDOC 1517 Quality Control in Mammography Software

WOMEN S HEALTH SOLUTIONS. The power and promise of breast tomosynthesis is here. Selenia Dimensions system with Acquisition Workstation 8000

Dental Intraoral X-ray Systems

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Introduction 1. Executive Summary 5

Breast Tomosynthesis An additional screening tool in the fight against breast cancer

Contrast Enhanced Spectral Mammography (CESM) Updates

Contrast-Enhanced Digital Mammography

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company

Hands-On Ultrasound Physics and Quality Control Workshop. Outline

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

The American College of Radiology Breast MRI Accreditation Program: Frequently Asked Questions (Updated: October 4, 2017)

Automated Practical System Qualification, Validation and Reporting acc. ASTM E

ACR Accreditation Update in MRI

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI

HEALTHCARE AI DEVELOPMENT CYCLE

Patient Dosimetry in Mammography and Tomosynthesis:

"Optimal Dose Techniques and Image Quality: Can We Have Both?" Lorusso, J. R., Fitzgeorge, L., Lorusso, D., & Lorusso, E. 2014

ACR Breast MRI Accreditation Program - DRAFT

Quality assurance in external radiotherapy

ACCREDITATION CASE REVIEW 2: US AND US-GUIDED BIOPSY CRITERIA

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski

Screening Mammograms: Questions and Answers

Bone Densitometry Equipment Operator

organizations including the International Commission on Radiation Protection exposure dose management system to all the studies and

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Corporate Medical Policy

Transcription:

ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview New ACR Activities Requirements Today What s New For Tomorrow ACR: Recognized by FDA and CMS 1987 Mammography accreditation 1987 Radiation oncology 1995 Ultrasound 1996 Stereotactic breast biopsy 1997 MRI 1998 Ultrasound guided breast biopsy 1999 Nuclear medicine 2002 CT and PET 2011 Breast MRI 3 1

ACR Breast Imaging Centers of Excellence BICOE A center must be fully accredited in: Mammography by ACR (or FDA-approved state accrediting body) Stereotactic Breast Biopsy by the ACR Breast Ultrasound by the ACR (including the Ultrasound-Guided Breast Biopsy module) Breast MRI (Effective Jan 1, 2016) Adding Breast MR to BICOE Requirements Effective January 1, 2016 Existing BICOE facilities have two years to become accredited in Breast MR, or have their associated locations become accredited Facilities applying for BICOE designation after January 1, 2016 must have Breast MR accreditation, or be associated with a facility this is accredited 5 ~13% 2

ACR Breast Imaging Centers of Excellence 1/10/14 1123 Total # Breast Imaging Centers of Excellence (home facilities) # BICOE Color 0 1-10 11-20 21-30 31-40 41-50 51-60 61-70 71-80 81-90 Online ACR MAP Submission Went Live January 20, 2014 for Mammo ACRedit: Main Database where MAP account Triad: Is a separate system that handles/stores uploaded images Images kept 30-60 days in case they are appealed. Then deleted. Clear Canvas Software - Image Viewer Software Future: Going to all web applications Online ACR MAP Submission Login Screen - DMAP vs. MAP Numbers will always stay separate but login will get you to see both. 3

TRIAD Select how they want to upload the images TRIAD - Web Client Accepts DICOM, JPEG, TIFF, BMP Select Images for ACR review by uploading from their PC Fatty Dense Phantom Facility exports files into TRIAD MQSA - Who s Who The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) The Accreditation Bodies: (ACR, TX, IA, AR) The Inspectors: States 4

% of Final Reports Written that Pass on 1st Attempt US Mammography Facilities and Units (October 1 each year) In 2000 12,956 units at 9933 facilities 1.3 units/facility As of 7/1/14 13,612 units at 8713 facilities 1% increase in units/12% drop in facilities since 2000 95% are digital ACR Mammography Accreditation Program Pass Rates 100 90 80 70 60 50 40 30 20 10 0 1993 1994 1995 1996 1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 Year %Initial Pass %Renewal Pass %Total Pass MAP REASONS FOR UNIT FAILURES - 1st ATTEMPT 2013 5% 1% 2% 92% Clinical + Phantom Clinical Only Phantom (no clinical) Only Dose 5

Introduction 4 Accrediting Bodies (AB s) ~30 FDA approved units Before You Begin Qualifications Initial Master s or Bachelor s Pathway s Board Certification 8 hours of training in mammography (e.g. digital) Continuing Experience 2 Facilities & 6 Units over a 24-month Period 15 CME s in mammography in a 36-month period Introduction ACR Accreditation Program 6

Introduction ACR Manufacturer Accreditation Program Introduction ACR Accreditation Program Introduction MQSA 7

Introduction Golden Rules Must use manufacturer s QC procedures Mandate action limits Manufacturers QC may refer to Monitor & Printer Manufacturers QC Multimodality Workstations may have own separate QC Printers may have their own QC Various failures may result in stopping clinical imaging until failure can be corrected Introduction Golden Rules Clinical Tips Always get latest version of ACR Summary Forms Verify you re using correct Mfr QC Manual Record the correct Mfr QC Manual on your report Read the Mfr QC Manual - make sure you perform all tests Always seem to be updates or changed manuals ACR FFDM QC Manual Project Goals: Keep in mind Mammo has MQSA Regulation Account for all past, present, and future FFDM systems Reasonable and appropriate for mass implementation (~13,000 units) 8

The ACR DM Phantom Phantom Prototype Design Principles Based on existing ACR Accreditation Phantom Similar imaging and scoring to current SFM phantom Build on experience of QC techs and physicists at ~8,700 US facilities who already know how to use and score the existing phantom (~24,000 Techs) The ACR DM Phantom Phantom Prototype Design Principles Can be used on both SFM & FFDM Total attenuation matched to current SFM phantom Similar thickness Similar total dose Permits testing of the MQSA 3.0 mgy dose limit (Single CC view) Proposed Scoring Changes Eliminate subtraction for artifacts Add Fail for artifacts New pass/fail criteria from 4,3,3 To: 2,3,2 **But, objects are the same (effective) size as SFM Phantom 9

Wax Insert Specifications with Virtual Placement Grid ID Tag Specs 12.98 cm (+ 0.00, - 0.04 cm) 1.5 cm Virtual Box: height = 0.5 cm, length = 1.8 cm Location of Virtual Box ID Tag 0.49 cm 2.0 cm 6.98 cm (+ 0.00, - 0.04 cm) 2.0 cm 2.0 cm 0.49 cm 0.49 cm 2.0 cm each Notes: Test objects to be centered on their respective placement grid locations. 0.49 cm 0.49 cm perimeter around test object placement grid. 0.635 cm (1/4 inch) radius on corners of wax insert. Fiber Placement specs Fiber specifications Fiber Length = 1.0 cm + 0.1 cm Fiber Fiber Fiber Diameter = See Table 45 o + 5 o 45 o + 5 o Speck Placement & Specs Mass Placement & Specs 1. Specks to be placed at points on 1. Mass pre-cut sphere diameter = 5/8 inch star and middle of star 2. Mass placement to be within + 0.1 cm of 2. Speck Size (spherical) = See Table center of virtual grid 3. Center speck placement to be within + 0.1 cm of center of virtual grid 4. Distance from center speck to center of speck on perimeter = 0.5 cm + 0.1 cm The ACR FFDM Phantom Prototype Image of Entire Phantom Serial Number *Note: Gray dot in lower left corner of wax insert is an artifact due to a bubble in wax insert. 10

Wax Insert Expanded view of Wax Insert Pass Criteria: Equivalent to SFM Phantom: 2 Fibers, 3 Specks, 2 Masses 4 Fibers, 3 Specks, 3 Masses Pass Serial Number 11

Effects of Thickness Equalization New FFDM phantom equalizes attenuation inside and outside wax insert. This permits evaluation of artifacts over entire phantom area with same WW and WL used to score test objects. Technologist QC Tests Test Number Name (# of Test Elements) Minimum Frequency Required Corrective Action 1 ACR DM Phantom Image Quality (3-4) Weekly Before Clinical Use 2 Visual Checklist (1) Monthly As noted on form 3 Acquisition Workstation (AW) Monitor QC (3) Monthly 30 Days or Before Use for Severe Artifacts 4 Radiologist Workstation (RW) Monitor QC (4) Monthly 30 Days or Before Use for Severe Artifacts 5 Film Printer QC (4) Monthly Before Clinical Use 6 Viewbox Cleanliness (if app) (1) Monthly Before Clinical Use 7 Facility QC Review (1) Quarterly Not Applicable 8 Compression Force (1) Semiannual Before Clinical Use 9 Manufacturer Detector Calibration (If Applicable) Per Mfr Recommendation Before Clinical Use Optional - Repeat Analysis As Needed Within 30 days after analysis Optional System QC for Radiologist As Needed 30 Days or Before Use for Severe Artifacts Optional Radiologist Image Quality Feedback As Needed Not applicable Management Forms ACR DM Phantom Technique Summary AW & RW Monitor QC Summary Film Printer Procedure Summary Corrective Action Log Facility Equipment Inventory Form Mammography System QC Summary Checklist Display Device QC Summary Checklist Tech Tests 12

Tech Tests Tech Tests Medical Physicists QC Tests Test Number Name (# of Test Elements) Minimum Frequency Required Corrective Action 1 Mammography Equipment Evaluation and MQSA Req MEE Only Before Clinical Use 2 ACR DM Phantom Image Quality (5) Annual Before Clinical Use 3 Spatial Resolution (1) Annual Before Clinical Use 4 Automatic Exposure Control System Performance (1) Annual Before Clinical Use 5 Average Glandular Dose (1) Annual Before Clinical Use 6 Unit Checklist (1) Annual Before Clinical Use 7 Computed Radiography (If Applicable) (3) Annual Before Clinical Use 8 Acquisition Workstation (AW) Monitor QC (5) Annual Before Clinical Use 9 Radiologist Workstation (RW) Monitor QC (9) Annual Before Clinical Use 10 Film Printer QC (5) Annual Before Clinical Use 11 Evaluation of Site s Technologist QC Program (1) Annual Within 30 Days 12 Evaluation of Off-Site Technologist QC Program (If App) (1) Annual Within 30 Days MEE or Troubleshooting Test Forms Layer) Assessment MEE or Troubleshooting Beam Quality (Half-Value Before Clin or 30 D kvp Accuracy and Reproducibility MEE or Troubleshooting Before Clin or 30 D Collimation Assessment MEE or Troubleshooting Before Clin or 30 D Ghost Image Evaluation (Troubleshooting only) Troubleshooting Before Clinical Viewbox Luminance (Troubleshooting only) Troubleshooting ** Summary Report Forms Medical Physicist DM QC Summary Technique Chart (Clinical & Phantom) Medical Physicist Summary Letter for the Radiologist 13

Facility Name Address Address Address MAP ID Unit# (00000-00) X-Ray Unit Manufacturer X-Ray Unit Model X-Ray Unit Control Serial # X-ray Unit Date of Manufacture Room ID Survey Date Report Date Medical Physicist Signature Unit Mfr Telephone 000-000-0000 CR Unit Mfr CR Unit Model Date of Installation February 1, 2015 CR Unit Serial # Survey Type: Mammography Equipment Evaluation (MEE) - Acceptance Testing Routine Annual Survey Equipment Tested: DM Unit AW Monitor RW Monitor Viewbox Printer Other MP Oversight Level: Medical Physicist On-Site Medical Physicist Oversight Unit Description: Digital radiography (DR) Computed Radiography (CR) Tomosynthesis (DBT) Unit Type: Diagnostic & Screening Diagnostic Only Screening Only 1. Mammography Equip Eval (MEE) 1. ACR DM Phantom Image Quality 2. ACR DM Phantom Image Quality 2. Visual Checklist 3. Spatial Resolution 3. AW Monitor QC 4. AEC System Performance 4. RW Monitor QC 5. Average Glandular Dose 5. Film Printer QC 6. Unit Checklist 6. Viewbox Cleanliness 7. Computed Radiography (if app) 7. Facility QC Review 8. AW Workstation Monitor QC 8. Compression Force CA 9. Mfr Detector Calibration (if app) Optional - Repeat Analysis 9. RW Monitor QC Fiber score 10. Film Printer QC Mass score 11. Evaluation Tech QC Program Speck group score 12. Off-Site Tech QC Program (if app) SNR Beam Quality (HVL) Assessment kvp Accuracy and Reproducibility Collimation Assessment Ghost Image Evaluation Viewbox Lumuminance Imaging Facility 1891 Preston White Dr Overall Pass/Fail/NA Unit S/N January 1, 2015 Room 1 December 1, 2015 December 2, 2015 Mary Smith, PhD CR Mfr CR Model CR S/N ("Pass" means all components of test passes; "Fail" means any or all components fail; if "CA" checked, see Corrective Action Summary) Facility Name Address Address Address MAP ID Unit# (00000-00) X-Ray Unit Manufacturer X-Ray Unit Model X-Ray Unit Control Serial # X-ray Unit Date of Manufacture Reston, VA 20191 Overall Pass/Fail/NA Overall 12345-xx Unit Model Dates tested if not on same day as unit Room ID Survey Date Report Date Medical Physicist Signature Overall Pass/Fail/NA ACR DM Phantom Summary CNR Value Unit Mfr Telephone 000-000-0000 CR Unit Mfr CR Unit Model Date of Installation February 1, 2015 CR Unit Serial # Survey Type: Mammography Equipment Evaluation (MEE) - Acceptance Testing Routine Annual Survey Equipment Tested: DM Unit AW Monitor RW Monitor Viewbox Printer Other MP Oversight Level: Medical Physicist On-Site Medical Physicist Oversight Unit Description: Digital radiography (DR) Computed Radiography (CR) Tomosynthesis (DBT) Unit Type: Diagnostic & Screening Diagnostic Only Screening Only 1. Mammography Equip Eval (MEE) 1. ACR DM Phantom Image Quality 2. ACR DM Phantom Image Quality 2. Visual Checklist 3. Spatial Resolution 3. AW Monitor QC 4. AEC System Performance 4. RW Monitor QC 5. Average Glandular Dose 5. Film Printer QC 6. Unit Checklist 6. Viewbox Cleanliness 7. Computed Radiography (if app) 7. Facility QC Review 8. AW Workstation Monitor QC 8. Compression Force 9. Mfr Detector Calibration (if app) Optional - Repeat Analysis 9. RW Monitor QC Fiber score 10. Film Printer QC Mass score 11. Evaluation Tech QC Program Speck group score 12. Off-Site Tech QC Program (if app) SNR Beam Quality (HVL) Assessment kvp Accuracy and Reproducibility Collimation Assessment Ghost Image Evaluation Viewbox Lumuminance Imaging Facility 1891 Preston White Dr Reston, VA 20191 Overall Pass/Fail/NA Unit S/N January 1, 2015 CA Room 1 CNR 2.0 2.0 3.0 45.2 3.45 1.75 December 1, 2015 December 2, 2015 Mary Smith, PhD CR Mfr CR Model CR S/N ("Pass" means all components of test passes; "Fail" means any or all components fail; if "CA" checked, see Corrective Action Summary) Overall Pass/Fail/NA Overall 12345-xx Unit Model Dates tested if not on same day as unit Overall Pass/Fail/NA 45.2 Limit 2.0 2.0 3.0 40.0 2.0 ACR DM Phantom Summary Value 2.0 2.0 3.0 3.45 1.75 Limit 2.0 2.0 3.0 40.0 2.0 MP Tests MP Tests Medical Physicist's DM QC Test Summary Medical Physicist's QC Tests Medical Physicists Tests CA Tech QC Evaluation CA MEE or Troubleshooting Pass/Fail/NA CA AGD (mgy) 3.0 Medical Physicist's Section - DRAFT 1 of 2 6. MP TESTS - Forms-DRAFT_2014-07-15.xlsx MP Tests Medical Physicist's DM QC Test Summary Medical Physicist's QC Tests Medical Physicists Tests CA Tech QC Evaluation CA MEE or Troubleshooting Pass/Fail/NA CA AGD (mgy) 3.0 Medical Physicist's Section - DRAFT 1 of 2 6. MP TESTS - Forms-DRAFT_2014-07-15.xlsx 14

MP Tests MP Tests MP Tests 15

Challenges Accounting for, and incorporating, all the different current & future FFDM technologies Handling offsite equipment Ensuring all necessary tests are included, meaningful, and relevant for an accreditation program What s Next 2 Steps Draft being sent to manufacturers for preliminary feedback Final draft to be sent to FDA from ACR to apply for alternative standard under current regulations Alternative standard will allow facilities to use this instead of the manufacturer s manuals Potential for ACR QC Manual to be basis for new MQSA Regulations Preemptive Questions Cost of phantom? Don t know. Reason to believe it will be affordable. Implementation and roll-out? ACR to develop a plan to include some form of training. 16

End of Presentation Questions? AAPM 2014 Berns SAM s Questions In digital mammography, who mandates the pass/fail criteria for site QC? 6% 1. The American College of Radiology 24% 24% 29% 18% 2. The FDA 3. The FFDM unit manufacturer 4. MITA 5. MQSA 17

Answer: 3 - The FFDM unit manufacturer References MQSA Regulations 900.12(e)(6) http://www.fda.gov/cdrh/mammograp HY/frmamcom2.html#s90012 To meet the FDA requirement for continuing experience, how many mammography facilities and mammography unit surveys must be performed within the previous 24 months? 5% 1. 6 Facilities and 2 Mammography Units 15% 0% 25% 10% 2. 4 Facilities and 6 Mammography Units 3. 2 Facilities and 6 Mammography Units 4. 1 Facilities and 6 Mammography Units 5. 2 Facilities and 12 Mammography Units Answer: 3 - The FFDM unit manufacturer References MQSA Regulations 900.12(e)(6) http://www.fda.gov/cdrh/mammograp HY/frmamcom2.html#s90012 18

Answer 3: 2 Facilities and 6 Mammography Units References http://www.fda.gov/cdrh/mammography/robohelp/ START.HTM 900.12(a)(3)(iii)(B): Continuing experience. Following the second anniversary date of the end of the calendar quarter in which the requirements of paragraphs (a)(3)(i) and (a)(3)(ii) of this section were completed or of April 28, 1999, whichever is later, the medical physicist shall have surveyed at least two mammography facilities and a total of at least six mammography units during the 24 months immediately preceding the date of the facility s annual MQSA inspection or the last day of the calendar quarter preceding the inspection or any date in between the two. The facility shall choose one of these dates to determine the 24-month period. No more than one survey of a specific facility within a 10-month period or a specific unit within a period of 60 days can be counted towards this requirement. For FFDM, the exposure for a single CC view of the ACR phantom shall not exceed: 12% 1. 0.75 mgy/exposure 8% 12% 15% 15% 2. 1.25 mgy/exposure 3. 2.00 mgy/exposure 4. 3.00 mgy/exposure 5. 4.00 mgy/exposure Answer: 4 3.00 mgy/exposure References http://www.fda.gov/radiation- EmittingProducts/MammographyQualityStandardsActandProgram/ Guidance/PolicyGuidanceHelpSystem/ucm052690.htm 900.12(e)(5)(vi): Dosimetry. The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed 3.0 milligray (mgy) (0.3 rad) per exposure. The dose shall be determined with technique factors and conditions used clinically for a standard breast. 19

AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report 20