Cannabinoid Quantitation Using an Agilent 6430 LC/MS/MS

Similar documents
Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Validation of a Benzodiazepine and Z-Drug Method Using an Agilent 6430 LC/MS/MS

Rapid and Robust Detection of THC and Its Metabolites in Blood

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Application Note. Author. Abstract. Introduction. Food Safety

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Determination of Bath Salts (Pyrovalerone Analogs) in Biological Samples

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Supporting Information

Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Determination of Gamma-Hydroxy-Butyrate (GHB) in Biological Samples

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Application. Detection of Cannabinoids in Oral Fluid Using Inert Source GC/MS. Introduction. Authors. Abstract. Forensic Toxicology

[application note] Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Application Note. Abstract. Authors. Pharmaceutical

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

Two-Dimensional LC/MS/MS to Reduce Ion Suppression in the Determination of Cannabinoids in Blood Plasma

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Determination of Amantadine Residues in Chicken by LCMS-8040

Author. Introduction. Abstract

Extraction of 11-nor-9-carboxy-tetrahydrocannabinol from Hydrolyzed Urine by ISOLUTE. SLE+ Prior to GC/MS Analysis

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Authors. Abstract. Introduction. Environmental

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Mycotoxin Analysis in Infant Formula Using Captiva EMR Lipid Cleanup and LC/MS/MS

Liquid Liquid Extraction of Gamma Hydroxybutyric Acid (GHB) in Blood and Urine for GC-MS analysis

LC/MS/MS of Buprenorphine and Norbuprenorphine in Whole Blood Using Agilent Bond Elut Plexa PCX and an Agilent Poroshell 120 Column

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Detection of Cotinine and 3- hydroxycotine in Smokers Urine

Determination of Opiates and Metabolites in Blood Using Electrospray LC/MS. Application Note

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Determination of Benzodiazepines in Urine by CE-MS/MS

Extraction of Synthetic and Naturally Occurring Cannabinoids in Urine Using SPE and LC-MS/MS

Determination of Multi-Residue Tetracyclines and their Metabolites in Milk by High Performance Liquid Chromatography - Tandem Mass Spectrometry

Dienes Derivatization MaxSpec Kit

Identification of Imidacloprid Metabolites in Onions Using High Resolution Accurate Mass Spectrometry (LC/Q-TOF MS) and Accurate Mass Tools

Determination of 44 Pesticides in Foodstuffs by LC/MS/MS Application

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Application Note. Authors. Abstract. Food

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Authors. Abstract. Introduction. Forensics

Determination of Amphetamine and Derivatives in Urine

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Modernizing the Forensic Lab with LC-MS/MS Technology

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Determination of red blood cell fatty acid profiles in clinical research

Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Vitamin D3 and related compounds by ESI and APCI

Rapid Hydrolysis of Benzodiazepines in Urine. Alicia Zook 1 and Crystal Xander B.S. 2. Cedar Crest College, Allentown, PA 1

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

Quantitative Determination of Drugs of Abuse in Human Whole Blood by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Analyze Biogenic Amines in Salmon and Shrimp by Capillary Electrophoresis-Tandem MS

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Simultaneous Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Multiclass Mycotoxin Analysis in Cheese Using Agilent Captiva EMR Lipid Cleanup and LC/MS/MS

Transcription:

Cannabinoid Quantitation Using an Agilent 643 LC/MS/MS Application Note Forensics Authors Jason Hudson, Ph.D., James Hutchings, Ph.D., and Rebecca Wagner, Ph.D. Virginia Department of Forensic Science Curt Harper, Ph.D. Pat Friel Agilent Technologies, Inc. Abstract A method was developed for the quantitation of cannabinoids with LC/MS/MS using an Agilent 643 Triple Quadrupole LC/MS system. The method displays excellent accuracy and precision using a linear calibration for the analysis of THC, Carboxy-THC, and OH-THC in whole blood. Sufficient resolution and peak shape for the targets can be achieved within an overall run time of 12.5 minutes. Additional validation studies confirmed that this method meets all criteria required for routine analysis of cannabinoids in whole blood.

Introduction Cannabinoids are analyzed in urine, oral fluid, and blood in many forensic toxicology laboratories. Quantitative analysis of cannabinoids in blood, specifically THC, is necessary for meaningful toxicological interpretation in the investigation of Driving Under the Influence of Drugs (DUID) cases. The quantitation and confirmation of cannabinoids for DUID cases constitutes a significant portion of the workload for many forensic toxicology laboratories worldwide. Cannabinoid analysis in blood has been driven by advances in GC/MS technology, notably 2D-GC/MS [1] and GC/MS/MS with column backflushing [2]. During the last decade, the role and progress of liquid chromatography triple quadrupole mass spectrometry (LC/MS/MS) in forensic and clinical toxicology has been assessed several times by leading experts in the field. This technique is becoming increasingly useful in routine toxicological analysis given its accuracy and sensitivity [3]. This application addresses the development of an LC/MS/MS method for the quantitation of THC and its metabolites. The primary pharmacologically active natural cannabinoid is D9-THC, and the primary (inactive) metabolite is 11-nor-D9-tetrahydrocannabinol-9-carboxylic acid (carboxy-thc). Another (active) metabolite of forensic interest is 11-hydoxy- D9-tetrahydrocannbinol (OH-THC). A mathematical model using THC and carboxy-thc concentrations has been explored as a means of estimating time of last use of cannabis [4]. Recent research has also shown that a glucuronide conjugate of carboxy-thc is present in relatively high concentrations in the blood of cannabis users. Sample preparation methods that expose carboxy-thc glucuronide to basic ph result in its conversion to carboxy-thc and elevate the apparent carboxy-thc concentration by approximately a factor of two [5]. The LC/MS/MS method described within this application note has the advantage of simpler sample preparation, without derivatization, compared to standard GC/MS and GC/MS/MS methods. Liquid-liquid extraction at acidic ph does not result in artifactual elevation of carboxy-thc concentration due to hydrolysis of carboxy-thc glucuronide. Validation studies were conducted using the SWGTOX guidelines [6] in conjunction with the Virginia Department of Forensic Science validation guidelines. This method was determined to meet all criteria for the qualitative and quantitative analysis of cannabinoids [7]. Experimental Equipment and Instrumentation Agilent 643 LC/MS/MS system Agilent 126 Infinity LC with an Agilent Poroshell 12 EC-18, 2.1 75 mm, 2.7 µm column Agilent 126 Automatic Liquid Sampler Autosampler vials with inserts Agilent MassHunter Optimizer Software Zymark TurboVap Evaporator Screw capped extraction tubes with 12 ml or greater capacity Kimble/Chase tapered glass tubes for evaporation and reconstitution (p/n 73785-5) Glass Pasteur pipe ts Pipets for accurate dispensing of volumes 1 µl to 25 µl, and 1 ml to 1 ml Test tube rocker or rotator Centrifuge Materials Water, Type 1 or HPLC grade Acetonitrile, Optima grade or higher Methanol, HPLC grade or higher Formic acid, eluent additive for LC/MS ~ 98% Hexane, Optima grade or higher Ethyl acetate, O ptima grade or higher Glacial acetic acid, ACS plus grade or higher Blank blood The calibrators, controls, and internal standards for this method were purchased from Cerilliant and Grace-Alltech and are listed in Table 1. The calibrators are prepared using a different source than controls. Mobile phase solutions.1% formic acid in water (mobile phase A).1% formic acid in acetonitrile (mobile phase B) 2

Table 1. Target THC Carboxy-THC OH-THC Targets and Corresponding Internal Standards Internal standard THC-d3 Carboxy-THC-d3 OH-THC-d3 Sample Preparation Blood samples were prepared according to the procedure detailed in Figure 1. Whole blood samples were extracted using 9:1 hexane:ethyl acetate with 1% acetic acid. Extracts were reconstituted in 5:5 acetonitrile/water then separated with an Agilent 126 Infinity LC with an Agilent Poroshell 12 EC-18, 2.1 75 mm, 2.7 µm column. Internal Standard and Calibrator Preparation Working internal standard solution (1 µg/ml): Pipette 1 µl of the.1 mg/ml (or 1 µl of 1. mg/ml) stock solution of deuterated standards into a 1-mL volumetric flask and qs to volume with methanol. Working internal standard solution (.1 µg/ml): Pipette 1 ml of the 1 µg/ml working internal standard solution of deuterated standards into a 1-mL volumetric flask and qs to volume with methanol. The samples eluted in a gradient of.1% formic acid in acetonitrile and.1% formic acid in water. Once separated, the liquid phase was analyzed using an Agilent 643 Triple Quadrupole LC/MS in positive ion mode. Chromatography and transition ions were identified and optimized by Agilent MassHunter Optimizer Software, and used to determine the quantifier and qualifier transitions. 1 ml sample + Water + Acetic acid Rotate and + Hexane:Ethyl acetate centrifuge Transfer Evaporate Reconstitute Centrifuge Instrument Working standard solution (1/5 µg/ml): Pipette 25 µl/125 µl of the 1. mg/ml stock solution standards (THC, OH-THC/Carboxy-THC) into a 25-mL volumetric flask and qs to volume with methanol. Figure 1. Table 2. Sample preparation procedure. Preparation of Calibrators Working standard solution (.1/.5 µg/ml): Pipette 1 ml of the 1/5 µg/ml working standard solution into a 1-mL volumetric flask and qs to volume with methanol. To prepare the calibration curve, pipette the following volumes of the 1/5 or the.1/.5 µg/ml working cannabinoid standard solution into appropriately labeled 16 125 mm screw cap test tubes. To eliminate a solvent effect, calibrators and controls may be dried under nitrogen prior to the addition of blank blood. Add 1 ml blank blood to obtain the final concentrations listed in Table 2. Amount of 1/5 µg/ml stock solution (µl) Amount of.1/.5 µg/ml stock solution (µl) Final concentration of cannabinoids (mg/l) 1.1/.5 5.5/.25 25.25/.125 1.1/.5 5.5/.25 25.25/.125 1.1/.5 3

Procedure 1. Label clean 16 125 mm screw cap tubes appropriately with calibrators, controls, and case sample IDs. 2. Prepare calibrators and controls. 3. Add 1. ml case specimens/blank blood to the appropriately labeled tubes. 4. Add 1 µl of the.1 µg/ml internal standard working solution to each tube. 5. Add 2 ml of water and vortex briefly. 6. Add 8 µl of 1% acetic acid and vortex. 7. Add 8. ml of 9:1 hexane:ethyl acetate solution; cap and rotate tubes for 3 minutes. 8. Centrifuge at approximately 2,8 rpm for 15 minutes to achieve separation. Transfer organic (upper) layer to appropriately labeled tubes. 9. Evaporate samples to dryness at approximately 4 5 C under nitrogen. 1. Reconstitute samples in 1 µl of 5:5 acetonitrile:water. Centrifuge at approximately 2,8 rpm for 15 minutes. Transfer to autosampler vials with inserts for LC/MS/MS analysis. 11. Run the calibrators, controls, and samples in a Worklist, according to the LC/MS/MS method described in Table 3. Table 3. Instrument Conditions MSD parameters Parameter Value (+) Ionization Polarity ESI Positive Gas Temperature 35 C Gas Flow Nebulizer Pressure Capillary 1 L/min 4 psi 4, V LC parameters Injection volume 1. µl Column Agilent Poroshell 12 EC-18, 2.1 75 mm, 2.7 µm Column thermostat 4. C Needle wash 5. seconds Mobile phase A.1 % formic acid in water Mobile phase B.1 % formic acid in acetonitrile Flow rate.5 ml/min Gradient Initial 4% B 1. minutes 4% B 7. minutes 95% B 1. minutes 95% B 1.5 minutes 4% B Stop time 1.5 minutes Post time 2. minutes 4

Table 3. Instrument Conditions (continued) Acquisition time segments Index Start time Scan type Ion mode Div valve Delta EMV Store 1 MRM ESI To waste No 2 4 MRM ESI To MS 4 Yes 3 6.15 MRM ESI To MS 4 Yes 4 8 MRM ESI To waste No Time segment 2 Compound ISTD? Precursor ion MS1 resolution Product ion MS2 resolution Dwell time (ms) Fragmentor (V) THC-COOH-d3 Yes 348.2 Unit 33.2 Unit 5 125 12 THC-COOH-d3 Yes 348.2 Unit 32.2 Unit 5 125 16 THC-COOH No 345.2 Unit 299.2 Unit 5 125 16 THC-COOH No 345.2 Unit 193.1 Unit 5 125 24 OH-THC-d3 Yes 334.2 Unit 316.2 Unit 5 12 8 OH-THC-d3 Yes 334.2 Unit 196.1 Unit 5 12 2 OH-THC No 331.2 Unit 313.2 Unit 5 15 8 OH-THC No 331.2 Unit 193.1 Unit 5 15 2 CE (V) Time segment 3 Compound ISTD? Precursor ion MS1 resolution Product ion MS2 resolution Dwell time (ms) Fragmentor (V) THC-d3 Yes 318.2 Unit 196.1 Unit 1 12 2 THC-d3 Yes 318.2 Unit 123 Unit 1 12 32 THC No 315.2 Unit 193.1 Unit 1 12 2 THC No 315.2 Unit 123 Unit 1 12 32 CE (V) 5

Results and Discussion The method achieved acceptable resolution of target compounds in an overall run time of 12.5 minutes. Figures 2 5 illustrate the resolution and MRM transitions achieved with this method. Peak shape is excellent with no significant tailing or other chromatographic abnormalities. 1 5 1. OH-THC 23.9.8 THC-COOH.7.6.5.4.3.2 THC.1 4.1 4.3 4.5 4.7 4.9 5.1 5.3 5.5 5.7 5.9 6.1 6.3 6.5 6.7 6.9 7.1 7.3 7.5 7.7 7.9 Figure 2. Qualitative chromatographic retention of analytes. 1 5 +MRM (331.2 & 313.2) 1 2 331.2 & 313.2, 331.2 & 193.1 Ratio = 1.3 (1.6%) Counts 1..8.6.4.2 4.895 min 1.8263 ng/ml 377,95 4.2 4.4 4.6 4.8 5. 5.2 5.4 5.6 5.8 Relative abundance (%) 1..8.6.4.2 4.2 4.4 4.6 4.8 5. 5.2 5.4 5.6 5.8 Figure 3. MRM of OH-THC. Counts 1 4 +MRM (315.2 & 193.1) 2.5 2. 1.5 1..5 Figure 4. MRM of THC. 6.829 min 1.13 ng/ml 78,67 Relative abundance (%) 315.2 & 193.1, 315.2 & 123. 1 2 Ratio = 69.9 (98.7%) 6. 6.2 6.4 6.6 6.8 7. 7.2 7.4 7.6 6. 6.2 6.4 6.6 6.8 7. 7.2 7.4 7.6 1..8.6.4.2 6

Counts 1 5 +MRM (345.2 & 299.2).8.6.4.2 5.2 min 491.719 ng/ml 31,131 Relative abundance (%) 345.2 & 299.9, 345.2 & 193.1 1 2 Ratio = 46.5 (91.1%) 1..8.6.4.2 4.2 4.6 5. 5.4 5.8 4.2 4.6 5. 5.4 5.8 Figure 5. MRM of Carboxy-THC. Seven calibrators were analyzed with every batch and used to assess the linearity of the instrumental response. Figures 6 8 show sample calibration curves derived from the method. Calibrations were linear from 1 1 ng/ml for THC and OH-THC, and 5 5 ng/ml for carboxy-thc. The R 2 coefficients in this study were greater than.992 ±.7 for all target compounds. Relative responses 1 1 OH-THC 1..9.8.7.6.5.4.3.2.1 Figure 6. y =.15764*x.5866 R 2 =.99954335 1 2 3 4 5 6 7 8 9 1 Relative concentration Calibration Curve for OH-THC. Validation studies were conducted using the SWGTOX guidelines in conjunction with the Virginia Department of Forensic Science validation guidelines [7]. This method was determined to meet all criteria for the qualitative and quantitative analysis of cannabinoids [8]. Items assessed during this validation study included linearity and calibration model fit, precision and accuracy using pooled and spiked whole blood samples, sensitivity (Limit of Detection (LOD) and Limit of Quantitation (LOQ)) interferences, robustness, carryover, dilution integrity, stability, and ion suppression/enhancement. As a result of this validation study, it was determined that matrix effects of this method were minimal. For THC, carboxy-thc and OH-THC LODs were 1, 2.5, and 1 ng/ml, and the LOQs were 1, 5, and 1 ng/ml, within acceptable ranges for these compounds. At the default collision cell accelerator voltage of 7 V, there was no evidence of cross talk between MRM transitions for target compounds and isotopically-labeled ISTDs. A more detailed explanation and results of this validation study can be found in "Validation of a Cannabinoid Quantitation Method Using an Agilent 643 LC/MS/MS" [7]. Relative responses 1 1 THC y =.97367*x.76167.9 R 2 =.999452.8.7.6.5.4.3.2.1 Relative responses 1 1 THC-COOH 2.8 y =.56318*x.115548 R 2 =.999575 2.4 2. 1.6 1.2.8.4 1 2 3 4 5 6 7 8 9 1 Relative concentration 5 1 15 2 25 3 35 4 45 5 Relative concentration Figure 7. Calibration Curve for THC. Figure 8. Calibration Curve for THC-COOH. 7

Conclusion This method development provides a rapid and sensitive technique for the detection and quantitation of cannabinoids by LC/MS/MS. Sample preparation for LC/MS/MS analysis is more streamlined and does not require derivitazation as opposed to traditional GC/MS analysis. Validation studies detailed in "Validation of a Cannabinoid Quantitation Method Using an Agilent 643 LC/MS/MS" indicate that the method meets all criteria required for the routine analysis of cannabinoids in whole blood [7]. References 1. C. Moore, S. Rana, C. Coulter, F. Feyerherm, H. Prest, Application of two-dimensional gas chromatography with electron capture chemical ionization mass spectrometry to the detection of 11-nor-Delta9-tetrahydrocannabinol-9-carboxylic acid (THC-COOH) in hair, J. Anal. Toxicol., 26 Apr; 3(3):171-7. 2. S. Bauman, Rapid and Robust Detection of THC and Its Metabolites in Blood, Agilent Application Note 599-8456EN, July 5, 211. 3. O. Quintela, D.J. Crouch, The determination of cannabinoids using liquid chromatography with mass spectrometric detection, Methods Mol. Biol., 212;92:75-9. 4. E.L. Karschner, D.M. Schwope, E.W. Schwilke, R.S. Goodwin, D.L. Kelly, D.A. Gorelick, M.A. Huestis, Predictive model accuracy in estimating last D9- tetrahydrocannabinol (THC) intake from plasma and whole blood cannabinoid concentrations in chronic, daily cannabis smokers administered subchronic oral THC, Drug Alcohol Depend., 212 Oct 1;125(3):313-9. 5. S. König, B. Aebi, S. Lanz, M. Gasser, W. Weinmann, On-line SPE LC-MS/MS for the quantification of D9-tetrahydrocannabinol (THC) and its two major metabolites in human peripheral blood by liquid chromatography tandem mass spectrometry, Anal. Bioanal. Chem., 211 Apr;4(1):9-16. 6. Standard Practices for Method Validation in Forensic Toxicology, SWGTOX, Doc 3, Revision 1, May 2, 213. 7. J. Hudson PhD., J. Hutchings PhD., C. Harper PhD., R. Wagner PhD. Validation of a Cannabinoid Quantitation Method Using an Agilent 643 LC/MS/MS, Agilent Application Note 5991-2554EN, June 213. 8. http://www.dfs.virginia.gov/manuals/toxicology/procedures/22-d1%2toxicology%2procedures%2m anual.pdf For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 213 Printed in the USA June 2, 213 5991-2521EN