Volumetric Modulated Arc Therapy - Patient Specific QA

Similar documents
Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Unrivaled, End-to-End

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

IMRT QA: Point Dose Measurements or 2D Array?

The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Leila E. A. Nichol Royal Surrey County Hospital

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Verification of Advanced Radiotherapy Techniques

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

EPID dosimetry for pretreatment quality assurance with two commercial systems

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Small field diode dosimetry

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

EORTC Member Facility Questionnaire

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Agenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

DELIVERED PATIENT DOSE

Patient QA Recommended Publications

Effective and efficient radiotherapy dosimetry audit: Where to next?

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

ph fax

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Follow this and additional works at:

Pre-treatment verification for IMRT / IGRT and Rotational therapy

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Future upcoming technologies and what audit needs to address

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Normal tissue dose in pediatric VMAT Piotr Zygmanski

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

PMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.

Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

Statistical process control analysis for patient-specific IMRT and VMAT QA

A review of RTTQA audit. Karen Venables

Dosisverifikation mit Delta 4 Discover während der Behandlung

Implementation of advanced RT Techniques

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

biij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

IMAGE-GUIDED RADIATION THERAPY

PGY-1. Resident Review Session Schedule

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

The impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

Online Care, Online Control, Online Confidence

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Delivery Quality Assurance with a High Resolution Liquid Filled Ion Chamber Array for Robotic Radiosurgery

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Quantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Limits of Precision and Accuracy of Radiation Delivery Systems

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

A national dosimetric audit of VMAT and Tomotherapy in the UK

IMRT pre-treatment verification basic approaches

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

The Journey of Cyberknife Commissioning

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

TomoTherapy Patient Quality Control With Dosimetry Check Version 5 Release 8. Math Resolutions, LLC 5975 Gales Lane Columbia, MD USA 21045

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

Active Scanning Beam 3 Checking Delivery/Dosimetry

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Efficient Dosimetry for Proton Therapy

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

ORIGINAL ARTICLE Oncology & Hematology INTRODUCTION

Many vendors are beginning to allow couch motion during radiation delivery.

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Hampton University Proton Therapy Institute

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Transcription:

Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA

VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck in MapPhan phantom ArcCheck Delta4 Fluence measurement 2D ion chamber array MatriXX mounted on Gantry Portal dosimetry 2D ion chamber array MatriXX in MULTICube phantom PTW 729 in Octavius phantom

Which system do you use/intend to use in your clinic for VMAT QA 0% 0% 0% 0% 0% A. Film & ion chamber B. Ion chamber array (MatriXX( MatriXX,, PTW 729) C. Diode array (mapcheck( mapcheck, ArcCheck,, Delta4) D. Portal dosimetry E. others 10

Film & Ion Chamber A stack of solid water phantom with a total thickness of 15cm. A 0.6cc farmer chamber was inserted at the depth of 10cm. A film was sandwiched 1cm above the ion chamber.

Prostate Case Plan Measurement Ion Chamber measurement is 1.3% below the predicted value

H&N Case

H&N Case: absolute dose comparison measurement plan The Gamma analysis passing rate in this case is 96.4% The measured ion chamber dose was 0.2% more than the planned one

2D Diode Array A MapCheck device inserted into a MapPhan solid water phantom

MapCheck in MapPhan QA result With 3% and 3mm criteria, the gamma analysis passing rate is 96.5% Dose threshold in gamma analysis is 10%

Dose Rate Dependency 1.005 relative dose response 1.000 0.995 0.990 0.985 0.980 0.975 Mapcheck diode MatriXX ion chamber 0.00 50.00 100.00 150.00 200.00 250.00 300.00 350.00 400.00 450.00 500.00 dose rate (MU/minute) The variation of dose response with the dose rate can reach up to 2.5% for MapCheck diode.

Calibrate MapCheck with medium dose rate Dose calibration was performed at 224 MU/Min Gamma passing rate increased to 98.8%

Angular dose response of MapCheck diode -90 0 detector array & phantom Couch -180 180 90 P er ce n t E r r o r (% ) 30.0 25.0 20.0 15.0 10.0 5.0 0.0-5.0-10.0-180 -120-60 0 60 120 180 Gantry Angle phantom setup MapCheck Diode MatriXX ion chamber p e r c e n t e r r o r (% ) 30.0 25.0 20.0 15.0 10.0 5.0 0.0-5.0-10.0-180 -120-60 0 60 120 180 Gantry Angle Significant variation of the dose response can be found when the incidence beam angle is parallel to the detector plane. Each diode in MapCheck may have different angular dose response curve.

Arc Check Sun Nuclear 1386 diode detectors arranged in cylindrical geometry Measures entrance and exit dose The QA can be done in composite or per control point

Central Cavity Options ArcCHECK features a versatile central cavity May be used to accommodate different detectors and inserts MultiPlug (left) also accommodates EBT film cassette and multiple detector locations

SNC Patient Software Software enables composite and control point analysis DICOM RT Dose is imported and ArcCHECK software then extracts 3D dose corresponding to detector locations, and performs a comparison All data files from ArcCHECK are open format for easy export Raw data also available for export

ArcCheck QA result Gamma analysis passing rate is 95.2% for this case.

Virtual Inclinometer

3DVH Option The most advanced 3D patient dose and DVH tools available Uses existing ArcCHECK measurements Supports VMAT and RapidArc deliveries No secondary dose calculation algorithm 3D dose and DVH analysis on patient geometry (not phantom geometry)

Scandidos Delta 4 A cylinder-shaped plastic phantom with 2 imbedded orthogonal crossing detector planes. 1069 diode detectors Dose is recorded in 2 planes and a 3D dose is reconstructed for comparison with the QA plan.

Composite-level analysis of SmartArc plan with Delta 4

MatriXX & MULTICube phantom MatriXX system has 1020 vented ion chambers with an active area of 24.4cm x 24.4cm. The size of each ion chamber is 4.5cm (diameter) x 5cm (height). The raw pixel size is 7.62mm, which can be interpolated down to 1mm. Laser is used to align system. Dose from both coronal and sagittal planes can be measured depending on the phantom.

MatriXX data conversion linear Cubic Spline The raw image has only 1024 pixels with 7.62mm dose grid. The data set can be interpolated to 1mm dose grid.

Results: Prostate Case measurement plan With 3% and 3mm criteria, the passing rate is 99% for this coronal slice

MatriXX QA: H&N Case measurement plan With 3% and 3mm criteria, the passing rate is 98.2% for this coronal onal slice No dose threshold was applied in gamma analysis. Note the passing g rate reduces to 97.1% if a 10% threshold is applied.

Angular variation of dose response 0 10.0 6.0-90 detector array & phantom Couch 90 percent error (%) 2.0-2.0-6.0-180 180 phantom setup -10.0-180 -120-60 0 60 120 180 Gantry Angle Up to 8% dose variation can be found when the photon beam is parallel allel to the detector plane

MatriXX Evolution Input Inclinometer The gantry angle for each frame is recorded during the VMAT delivery. A angular dose response correction is applied to all the data based on the gantry angle information.

With and Without Angular response correction Plan Measurement Without Correction A two-arc case: H&N Measurement With Correction The measured data without angular dose response correction is slightly colder than the one with correction.

With and Without Angular response correction Without Correction 81.9% With Correction 96.2% 3% and 3mm passing criteria for Gamma analysis.

0% 0% 0% 0% 0% Which of the following cannot be measured using MatriXX 2D ion chamber array A. Dose distribution on a sagittal plane B. Absolute dose at detector center C. Dose distribution on a coronal plane D. Dose distribution on a transverse plane E. Radiation fluence of a VMAT plan 10

Answer A. Dose distribution on a sagittal plane B. Absolute dose at detector center C. Dose distribution on a coronal plane D. Dose distribution on a transverse plane E. Radiation fluence of a VMAT plan

Portal Dosimetry Portal Dosimetry Process Created verification plan for each patient Calibrated panel prior to collecting data Data measured in units of CU Definition of Calibrated Unit (CU) Field Size =10 x 10 cm2 SDD = 105 cm Deliver 100 MU to panel Set reading = 1 CU 1 CU is numerically approximate to 1 Gy

Portal Dosimetry RapidArc (Prostate)

Portal Dosimetry RapidArc (Prostate)

Portal Dosimetry RapidArc (HN)

Portal Dosimetry RapidArc (HN)

Which QA system has the highest resolution 0% 0% 0% 0% 0% A. portal dosimetry using EPID B. film C. MatriXX 2D ion chamber array D. ArcCheck E. Delta4 10

Answer A. portal dosimetry using EPID B. film C. MatriXX 2D ion chamber array D. ArcCheck E. Delta4

Summary VMAT QA using film and ion chamber is easy to setup and perform. However, there is no instantaneous result and It is time-consuming for data analysis. Diode-array and ion-chamber chamber-array array based devices can provide instantaneous data analysis and can be used for VMAT patient specific QA once the angular dose response and dose rate dependence is considered during data analysis/collection. Portal dosimetry can provide an easy and robust way for patient specific QA. The only issue with this technique is that information ion on gantry angle cannot be checked independently. Users need to know limit of each QA technique/device for best patient specific QA result.

Acknowledgement David M. Shepard, Ph.D. Muhammad K. N. Afghan, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Vivek Mehta, M.D. David Housley Grace Kim Igor Gomola, Ph.D. Gerry Vantellingen Jie Shi, Ph.D. Kris Lyle

Thank you!