Dynamic simulation of stent deployment effects of design, material and coating

Similar documents
Computational analysis of mechanical stress-strain interaction of a bioresorbable scaffold with blood vessel

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

A computational study of stent performance by considering vessel anisotropy and residual stresses

Numerical Analysis of Coronary Stent for Diverse Materials

THE COMPREHENSIVE FINITE ELEMENT MODEL FOR STENTING: THE INFLUENCE OF STENT DESIGN ON THE OUTCOME AFTER CORONARY STENT PLACEMENT

Modeling of stent implantation in a human stenotic artery

Planar Titanium Stent Design

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Current DES designs and expansion capacity testing

Computational Structural Modelling of Coronary Stent Deployment: A Review

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion

FINITE ELEMENT ANALYSIS OF SHAPE MEMORY ALLOY NiTi STENT IN CORONARY ARTERY. Amatulraheem Al-abassi. A thesis presented to University of Guelph

Finite Element Methods to Analyze Helical Stent Expansion

Abstract. Introduction. Journal of Biomedical Engineering and Research

Modelling of Atherosclerotic Plaque for Use in a Computational Test-Bed for Stent Angioplasty

Computational Analysis on Commercially Available Stent Designs

A Study of the Mechanical Properties of a Real Coronary Stent Dynamical Behaviour of an Optimal Stent

Lessons for technique and stent choice

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

Nitinol biomedical devices: Design analysis

Technical Characteristics of coronary prostheses(stents)

Investigation of metallic surface area of coronary stents

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

Vascular stents: Coupling full 3-D with reducedorder structural models

Stents selection and optimal implantation: sizes, design, deployment Abbott Vascular. All rights reserved.

Computer Methods in Biomechanics and Biomedical Engineering. Stresses in peripheral arteries following stent placement: A finite element analysis

A comparison of balloon- and self-expanding stents

Is a Standardized Measurement of the Elastic Recoil of Coronary Stents under Vascular Conditions Necessary and Meaningful?

Measurement of Mechanical Properties of Coronary Stents according to the European Standard pren

EXAMINATION OF MECHANICAL AND MEDICAL APPLICATION PROPERTIES OF CORONARY STENTS

Experimental investigation of the stent artery interaction

Effect of Plaque Composition on Biomechanical Performance of a Carotid Stent: Computational Study

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

NUMERICAL ANALYSIS OF VASCULAR STENTS EXPLOITING SHAPE-MEMORY-ALLOY BEHAVIOR

Clinically proven. ordering info. Vascular Intervention // Peripheral Self-Expanding Stent System/0.018 /OTW. Pulsar-18

Design Optimisation for Stent Manufacture

Percutaneous Coronary Intervention: an Update for the Internist

Senior Honors Thesis Prospectus

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA)

Document issued on: January 13, 2005

PCI for Ostial Lesion

Coroflex ISAR. Sirolimus Eluting Polymer-Free Coronary Stent System. Vascular Systems

A COMPREHENSIVE ITERATIVE DESIGN, SIMULATION AND FINITE ELEMENT ANALYSIS OF BARE METAL STENT AND DISEASED ARTERY. A Thesis By.

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

Computational Analysis on Commercially Available Stent Designs

Renal artery stenting- A pictorial review of stent designs, indications and outcomes

Topology optimisation of hip prosthesis to reduce stress shielding

The titanium nitride oxide stent an alternative to DES. Adam de Belder MD FRCP Sussex Cardiac Centre BSUH on behalf of Hexacath

Bench test : Is it still useful? Do we need standardization?

Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery

CHALLENGES IN FEMORO-POPLITEAL STENTING

Longitudinal stent distortion in clinical practice. How common is the problem and does stent design matter?

Combine the best of both worlds

Finite-element analysis of NiTi wire deflection during orthodontic levelling treatment

Comparative study of the contact pressures in hip joint models with femoroacetabular impingment with different cephalic deformities

aneurysm initiation and fully developed aneurysms

An effort is made to analyse the stresses experienced by the human femur. In order

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Vascular V12. Covered Stent. The New Standard of Care

COMPLEX CASES: LEFT MAIN

Branch Pulmonary Artery Stenosis Long-term Outcomes of Endovascular Stenting

Ενδοςτεφανιαίεσ προκζςεισ (stents):παρόν και μζλλον. Σ Χατηθμιλτιάδθσ Αναπλθρωτισ Κακθγθτισ Πανεπιςτθμιακό Νοςοκομείο ΑΧΕΠΑ

Yukon Choice PC Translumina trust is what counts

Rheological, mechanical and failure properties of biological soft tissues at high strains and rates of deformation

Sites of Atherosclerosis In order of Frequency

Development Of DES Technology. HUO Yong,MD FACC Peking University First Hospital President-elect Chinese Society of Cardiology

COMPANY INTRODUCTION: PROFILE & PRODUCTS

Drug delivery to the vessel wall: Coated balloons and the role of the excipient

Assessing potential cracking zones in embankment dams

Yukon Chrome PC. Translumina trust is what counts. Sirolimus Eluting CoCr Coronary Stent System. 5-Years Randomized Clinical Follow-Up

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

CYPHER (Polymer-based Sirolimus-eluting DES) New Stent Platforms. Campbell Rogers, M.D. Chief Technology Officer

Transactions on Biomedicine and Health vol 3, 1996 WIT Press, ISSN

Balloon-expandable closed-cell stents were introduced

Cardiovascular stent design and vessel stresses:a finite element analysis

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Design and Dynamic Modeling of Flexible Rehabilitation Mechanical Glove

A Highly Flexible Slotted Tube Stent Design Coated with a-sic:h First Clinical Experiences

The BIO revolution: bioadsorbable stents. Federico Conrotto Cardiologia 2 Città della Salute e della Scienza di Torino

[ICESTM-2018] ISSN Impact Factor

Finite Element Analysis of Stent Deployment: Understanding Stent Fracture in Percutaneous Pulmonary Valve Implantation

Interventional Cardiology

Numerical predicting of contact and pressure sore of lower extremity parts caused by prosthetic and orthotic

What Stent to Use? JASVINDAR SINGH MD, FACC

MULTIVESSEL PCI. IN DRUG-ELUTING STENT RESTENOSIS DUE TO STENT FRACTURE, TREATED WITH REPEAT DES IMPLANTATION

A Two-layered Model for the Analysis of Arterial Rheology

Non-Destructive Analysis of Explanted Coronary Artery Stents

TESTING METHOD OF STENT S RADIAL FORCE

Laboratory tests for strength paramaters of brain aneurysms

Element Clinical Program Perseus Late Breaking News and the Platinum Study Design

Next Generation of DCB: Efficacy and tolerance of drug-coated hyper-compliant balloons (DCHCB) in peripheral arteries of swine

Aortic stents, types, selection, tricks in deployment.

Αγγειοπλαστική σε Eπαναστενωτικές Bλάβες

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Innovation by design. Technology that sets a new standard

IN-STENT RESTENOSIS. K.Boerlage-van Dijk CarVasZ 2014

Active Response of Smooth Muscle and Arterial Tissue to Stenting

Device and Clinical Program Highlights: SINOMED BuMA Stent

Post-conditioning. P a g e 1. To my Thesis Committee,

Transcription:

Journal of Physics: Conference Series OPEN ACCESS Dynamic simulation of stent deployment effects of design, material and coating To cite this article: A Schiavone et al 2013 J. Phys.: Conf. Ser. 451 012032 View the article online for updates and enhancements. Related content - Numerical fatigue life assessment of cardiovascular stents: A two-scale plasticity-damage model H A F A Santos, F Auricchio and M Conti - Computational design analysis for deployment of cardiovascular stents Sriram Tammareddi, Guangyong Sun and Qing Li - Vascular stents with submicrometer-scale surface patterning realized via titanium deep reactive ion etching Shannon C Gott, Benjamin A Jabola and Masaru P Rao This content was downloaded from IP address 46.3.195.206 on 20/11/2017 at 13:41

Dynamic simulation of stent deployment - effects of design, material and coating A Schiavone, L G Zhao and A A Abdel-Wahab Wolfson School of Mechanical and Manufacturing Engineering, Loughborough University, LE11 3TU, Loughborough, UK E-mail: a.schiavone@lboro.ac.uk Abstract. Dynamic finite-element simulations have been carried out to study the effects of cell design, material choice and drug eluting coating on the mechanical behaviour of stents during deployment. Four representative stent designs have been considered, i.e., Palmaz-Schatz, Cypher, Xience and Endeavor. The former two are made of stainless steel while the latter two made of Co-Cr alloy. Geometric model for each design was created using ProEngineer software, and then imported into Abaqus for simulation of the full process of stent deployment within a diseased artery. In all cases, the delivery system was based on the dynamic expansion of a polyurethane balloon under applied internal pressure. Results showed that the expansion is mainly governed by the design, in particular open-cell design (e.g. Endeavor) tends to have greater expansion than closed-cell design (e.g. Cypher). Dogboning effect was strong for slotted tube design (e.g. Palmaz-Schatz) but reduced significantly for sinusoidal design (e.g. Cypher). Under the same pressure, the maximum von Mises stress in the stent was higher for the open-cell designs and located mostly at the inner corners of each cell. For given deformation, stents made of Co-Cr alloys tend to experience higher stress level than those made of stainless steels, mainly due to the difference in material properties. For artery-plaque system, the maximum stress occurred on the stenosis and dogboning led to stress concentration at the ends of the plaque. The drug eluting coating affected the stent expansion by reducing the recoiling phenomenon considerably, but also raised the stress level on the stent due to property mismatch. 1. Introduction Finite-element method has been widely employed to assess the biomechanical behaviour of stents by modelling their deployment inside diseased arteries. Chua et al. [1] simulated the expansion of a Palmaz-Schatz stent inside an artery with stenotic plaque. The results demonstrated the capability of slotted-tube stents in resisting the recoil of blood vessel and preventing restenosis after expansion. Zahedmanesh et al. [2] modelled the deployment of an ACS Multi-Link RX DUET stent in a stenotic artery based on the digitalization of 3D angiography images of a human coronary artery. They suggested that direct application of pressure on the inner surface of stent may be used as an optimal modelling strategy for complex arterial geometries. Pericevic et al. [3] investigated the influence of stenotic plaque compositions (i.e., hypecellular, hypocellular and calcified) on stent deployment, and showed that a calcified plaque exhibited less stress level but higher resistance to expansion compared to other types of plaque. Lally et al. [4] performed a comparative study of two stents (i.e. S7 and NIR Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

stents) by simulating their deployment inside a stenotic artery. Their results showed that the S7 stent caused lower stress on the blood vessel wall compared to the NIR, which was consistent with clinical observations that the NIR stent has a higher restenosis rate than S7. Tambaca et al. [5] studied the mechanical behaviour of different fully expanded stent designs under uniform compression and bending forces. It was shown that the latest stent designs (i.e. Xience) exhibited good radial stiffness and bending flexibility. In addition, Gu et al. [6] studied the influence of an ultra-thin silicone coating during deployment of drug eluting stent in an artery with long fusiform aneurysm. It was shown that the coated stent needed an appreciatively 30% higher pressure to expand the artery than its bare metal version. On the other hand, Hopkins et al. [7] investigated the 2D delamination effect of the stent coating during deployment, which occurred mainly in the region of hinge with high plastic strain. The initiation of coating debonding depended on the coating thickness, the coating material and the curvature of the hinge. Currently, there is a lack of research in the modelling of the full deployment process of the latest stent designs, in particular the effects of stent materials and coatings on the expansion behaviour of stents. It is also not well understood how the deformation and stress of the artery-plaque system are influenced by different stent designs during deployment. In this paper, finite-element analyses have been carried out to simulate stent deployment for four representative stent designs (Palmaz-Schatz, Cypher, Xience and Endeavor). Comparisons have been made to assess the effects of cell design, material and coating on stent expansion, stress distribution and geometrical change during the deployment process inside a diseased artery. 2. Finite element simulation 2.1. Geometrical models Models of Palmaz-Schatz, Cypher, Xience and Endeavor stents were produced using ProEngineer software according to their real geometrical features. The length of all stents was chosen to be 10mm and the diameter at the crimpled state was fixed at 3mm. The strut thickness is 100µm, 140µm, 81µm and 91µm for Palmaz-Schatz, Cypher, Xience and Endeavor, respectively. To study the coating effect, a two-ring model was created for Xience stent, with a length of 4mm and a diameter of 3mm, coated with a thin layer of polymer (7.8µm). The balloon was modelled as a simple cylinder with an outer diameter that matched the inner diameter of the stent. The balloon was always 2mm longer than the stent (i.e., one millimetre longer at each end). The blood vessel was modelled as a cylinder with axisymmetric stenotic plaque inside. The artery has a length of 20mm and a wall thickness of 0.4mm, while the plaque has a length of 5mm and a maximum thickness of 0.58mm. The outer diameter was 5.4 mm and 5.0 mm for the artery and the plaque, respectively. 2.2. Materials The materials used to manufacture the stents are 316L stainless steel for Palmaz-Schatz and Cypher and Co-Cr alloy for Xience (type L605) and Endeavor (type F562). These materials were modelled with a bilinear elastic-plastic stress-strain relationship with parameters given in Chua et al. [1] and Pochrząst et al. [10]. The stent coating was made of Phosphorylcholine (PC) polymer with bilinear model parameters obtained from Hopkins et al. [7]. The balloon was modelled using a Mooney-Rivlin hyperelastic strain energy potential [1] and the blood vessel was described by 5-parameter hyperelastic strain energy potential and considered uncompressible [4]. Material properties are summarized in Tables 1, 2 and 3. 2.4. Load and constraint A uniform and linearly increasing pressure was applied to the inner surface of the balloon to simulate the inflation process, followed by deflation (pressure decreased to zero). The step time was chosen to be 25ms for inflation and 5ms for deflation. Implicit dynamic analyses have been used to simulate the 2

whole process using Abaqus. In all simulations, the ends of the balloon were fully constrained to avoid rigid body motion. Figure 1. Finite element mesh for Cypher stent and stenotic artery. Table 1. Parameters for stent and coating materials. Material ρ (kg/mm 3 ) E (MPa) ν S y (MPa) E t (MPa) SS316L 7.8 10-6 193000 0.27 207 692 Co-Cr L605 9.7 10-6 243000 0.3 476 680 Co-Cr F562 8.8 10-6 233000 0.35 414 744 PC-Polymer 1.2 10-6 240 0.5 16 7 Table 2. Mooney-Rivlin coefficients for polyurethane balloon. Material ρ (kg/mm 3 ) C 10 C 01 D 1 Polyurethane 1.07 10-6 1.03176 3.69266 0 Table 3. Coefficients of the hyperelastic model for the blood vessel. Material ρ (kg/mm 3 ) a 10 a 01 a 20 a 11 a 30 Artery Wall 1.066 10-6 18.90 2.75 590.42 857.18 0 Calcified Plaque 1.45 10-6 495.90 506.61 1193.53 3637.80 4737.25 2.3. Mesh Stents were meshed with tetrahedral elements (C3D10), and the number of elements was 58428, 44222, 46221 and 63076 for Palmaz-Schatz, Cypher, Xience and Endeavor, respectively. The balloons were meshed using hexahedral elements with reduced integration and hybrid formulation (type C3D8RH), and the number of elements was about 10000 for all simulations. The artery wall was also meshed using hexahedral elements with reduced integration and hybrid formulation (C3D8RH), while the plaque was meshed with tetrahedral elements with hybrid formulation (C3D10H). The artery and plaque system has 40247 elements in total. Figure 1 shows the generated mesh for the whole system (Cypher stent). Interaction between the interfaces (balloon-stent-artery/plaque) was modelled as 3

surface-to-surface hard contact. Mesh sensitivity study has confirmed the convergence of results for the current mesh. Figure 2. Radial expansion (in percentage) against inflation pressure for the four stents under free expansion. Figure 3. Diameter change and recoil effect for Palmaz-Schatz (PS), Cypher (C), Xience (X) and Endeavor (E) under free expansion. 2.5. Model validation The model was first validated by reproducing the Palmaz-Schatz expansion and obtained results are in close agreement with those in Ju et al. [8] and Chua et al. [9]. This was purely for model validation purposes, to ensure that the mesh, boundary/loading conditions and solvers were correctly adopted and the materials and contacts were also properly simulated. 3. Results and discussion 3.1. Free expansion Figure 2 shows the change of diameter against the pressure, applied inside the balloon, for all four stent designs during free expansion. The diameter change is mainly controlled by the radial stiffness of stent. Among the four designs, Endeavor is the most flexible stent, which expanded 143% of its original diameter at 0.8 MPa pressure, due to its open-cell design and peak-to-peak strut connection. Xience, with a spiral-cell design and peak-to-valley strut connection, and Cypher, with a closed-cell design and peak-to-peak strut connections, exhibited similar flexibility and expanded about 71% and 78% of their original diameter, respectively, under the same pressure. Palmaz-Schatz, with a slotted tube design, showed the highest stiffness with an expansion of only 45% at 0.8MPa. In fact, Palmaz- Schatz stent required a pressure of 1MPa in order to reach an expansion of 103%. Recoiling effect after deflation is compared in Figure 3 for the four stents. Palmaz-Schatz stent showed the lowest recoiling effect (4%) due to its high radial stiffness, followed by Xience (8%) and Endeavor (11%). In terms of dogboning, Xience exhibited the strongest negative dogboning effect (- 24%), followed by Endeavor (-34%), Palmaz-Schatz (-19%) and Cypher (-18%). Negative dogboning effect is usually appreciated, as it helps with the dilation of the stenotic plaque which is in direct contact with the middle section of the stent. In addition, Palmaz-Schatz had the highest foreshortening effect (6.4%), followed by Cypher (2.0%), Xience (1.0%) and Endeavor (0.4%). It is important to make foreshortening effect as low as possible to ensure a full coverage of the diseased part of the artery. 4

Figure 4. Change of diameter against inflation pressure for the four stents during expansion inside a stenotic artery. Figure 5. The von Mises stress distribution for Palmaz-Schatz (PS), Cypher (C) and Xience (X) and Endeavor (E) in a diseased artery at an inflation pressure of 1.6MPa. All four stents showed severe stress concentrations at the corners of struts. At the same deformation level, stents made of Co-Cr alloys tend to have higher stress level compared to those made of stainless steel, in line with material properties. After deflation, residual von Mises stresses were obtained for all simulated stents. For Cypher, residual stress was 58.1% (211MPa) of the stress at the peak pressure, while for Palmaz-Schatz it was 52.9% (230MPa). Xience and Endeavor have a residual stress level similar to that at the maximum inflation pressure, i.e., 84.7% (466MPa) and 90.6% (690MPa) of the peak stress, respectively. The high stresses in some regions also indicate a potential fracture during the deployment process, which has been also frequently reported by clinical studies. 3.2. Expansion inside a stenotic artery For stent expansion inside a diseased artery, the diameter change during the inflation and deflation for each stent is shown in Figure 4. All four stents had similar expansion behaviour with a maximum diameter of 5.35mm achieved at 1.7MPa pressure. Endeavor was the stent that expanded the artery slightly more than others while Cypher was the stent that expanded less. 5

Figure 6. Diameter change and recoil effect for coated and bare metal Xience stent. Figure 7. The von Mises stress distribution on bare metal Xience stent (X) and coated Xience stent (XC) at the end of inflation (i.e., peak pressure). The von Mises stress distribution within the stents is shown in Figure 5 at the same pressure 1.6MPa. The location of the maximum stress is consistent with the case of free expansion. Endeavor and Xience showed higher magnitudes (861 MPa and 810MPa respectively) compared to the Cypher and the Palmaz-Schatz (440MPa and 512MPa, respectively), which is also consistent with free expansion simulations. For dogboning effect, Xience exhibited the lowest level (1.8%), Endeavor showed also a very low level (3.7%), while Cypher and Palmaz-Schatz had higher levels (12% and 13%, respectively). This indicates the advantage of the latest stent designs which can expand the artery more uniformly and prevent damage to the healthy section of blood vessel. Furthermore, at the same diameter, Xience had the lowest foreshortening effect (1.2%), followed by Cypher and Endeavor (2.0% and 2.1%), whilst the highest value was ehibited by Palmaz-Schatz (6.2%). All these results are again consistent with free expansion simulations, showing that Xience tends to preserve its length better during deployment. 3.3. Effect of stent coating To study the influence of stent coating, expansion of 2-ring Xience stent (coated and uncoated) was simulated under an inflation pressure of 1MPa. We used Xience as an example study and we assumed the effects of coating were similar for other types of stent. Figure 6 shows the diameter change during expansion process. The maximum diameter reached 5.5mm in both simulations, but the coated stent recoiled only 7.3% while the bare metal one had a recoil effect of 12.7%. Results suggest that drug eluting coatings do not affect stent expansion (i.e. radial stiffness) but reduce recoil effect significantly. For both cases, the maximum von Mises stress was located at the peaks and the valleys of the sinusoidal struts, as shown in Figure 7, at the maximum pressure. The coated stent had a higher stress level (749MPa) compared to the bare metal stent (510MPa). After recoil, the residual stresses were found to be 691MPa and 448MPa for the coated and uncoated stents, respectively. The increase of stress level on the coated stent is due to property mismatch between the stent and the coating. 6

Figure 8. The von Mises stress distribution on stenotic blood vessel, expanded with Palmaz-Schatz (PS), Cypher (C), Xience (X) and Endeavor (E) stents, respectively, under a pressure of 1MPa. 3.4. Stresses on plaque/artery The von Mises stress distribution on the plaque/artery had a similar pattern for all stents (Figure 8), with the maximum stress located towards the ends of the plaque where the stent and the plaque were directly in contact. At a pressure of 1MPa, Xience and Endeavor caused higher stress magnitude on the plaque (18.7MPa and 16.1MPa, respectively) than Cypher (6.8MPa) and Palmaz-Schatz (6.1MPa). 4. Conclusions Effects of cell design, material choice and drug eluting coating on the expansion behaviour of stents have been studied using finite element simulations. Results showed that stent expansion is mainly controlled by its design. Stents with open-cell design (e.g. Endeavor) tend to have greater expansion than those with closed-cell design (e.g. Cypher). Among the four designs, Xience, with a spiral-cell design and peak-to-valley strut connection, seems to have the lowest dogboning and foreshortening effects. For given deformation, stents made of Co-Cr alloys tend to have higher stress level than those made of stainless steels due to the difference in material properties. The drug eluting coating affected the stent expansion by reducing the recoiling phenomenon considerably, but also raised the stress level on the stent due to material property mismatch. The von Mises stress distribution on the plaque/artery had similar pattern for all four stents, with stress concentrations located towards the ends of plaque where it experienced a strong contact with the stent due to the dogboning effect. References [1] Chua S N D, MacDonald B J and Hashimi M S J 2004 J. Mater. Process. Technol. 155-156 1772 [2] Zahedmanesh H, Kelly D J and Lally C 2010 J. Biomech. 43 2126 [3] Pericevic I, Lally C, Toner D and Kelly D J 2009 Med. Eng. Phys. 31 428 [4] Lally C, Dolan F and Prendergast P J 2005 J. Biomech. 38 1574 [5] Tambaca J, Canic S, Kosor M, Fish R D and Paniagua D 2011 Tex. Heart I. J. 38 491 [6] Gu L, Santra S, Mericle R A and Kumar A V 2004 J. Biomech. 38 1221 [7] Hopkins C G, McHugh P E and McGarry J P 2010 Annu. Rev. Biomed. Eng. 38 2263 [8] Ju F, Xia Z and Sasaki K 2008 J. Mech. Behav. Biomed. Mat. 1 86 [9] Chua S N D, McDonald B J and Hashimi M S J 2003 J. Mat. Process. Technol. 143-144 591 [10] Pochrząst M, Walke W and Kaczmarek M 2009 J. Achiev. Mater. Manuf. Eng. 32 340 7