R/F. Clinical Experience Using the SONIALVISION safire II Utility of Tomosynthesis in Orthopedic Surgery

Similar documents
R/F. Use of Tomosynthesis at the Aizawa Hospital. 1. Introduction. 2. Use of Tomosynthesis at the Aizawa Hospital

R/F. Use of Tomosynthesis Radiography for Total Hip Arthroplasty (THA) at Iida Hospital. 1. Introduction. 2. Present Situation.

R/F. Experience of Using Slot Radiography Experimental Use for Lower Extremity Venography. 1. Introduction

BODY TOMOSYNTHESIS A. INTRODUCTION/CURRENT PRACTICE. Conventional radiography (X-ray)

R/F. Can T-smart Tomosynthesis Improve Diagnostic Accuracy on THA Component Stability? 1. Abstract

PACS: ERGONOMIC CONSIDERATIONS 1

2

Carpal rows injuries!

Vascular. Experiences Using an Angiography System with a Inch FPD - CT-Like Imaging of the Neck and Abdomen - 1.

Characterizing scaphoid nonunion deformity using 2-D and 3-D imaging techniques ten Berg, P.W.L.

TRIQUETRUM FRACTURE. The triquetrum bone is one of the small bones that make up the carpus.

The Kienböck disease and scaphoid fractures. Mariusz Bonczar

Clinical Application of Computed Radiography in Orthopedic Surgery

SCAHPO-LUNATE DISSOCIATION

Hand and wrist emergencies

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462

RAD. Experiences Using RADspeed Pro EDGE for Tomosynthesis. 2. Background to Adoption. 1. Profile of Imamura General Hospital

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

Metal Artefact Reduction in CT

Shoulder Position: Supine arm in the neutral position. Collateral arm above head Indication: fracture humerus, fracture scapula

RADIOGRAPHY OF THE WRIST

Ethan M. Braunstein, M.D. 1, Steven A. Goldstein, Ph.D. 2, Janet Ku, M.S. 2, Patrick Smith, M.D. 2, and Larry S. Matthews, M.D. 2

Scaphoid Fractures. Mohammed Alasmari. Orthopaedic Surgery Demonstrator Majmaah University

Functional Orthopedic Imaging Capturing Motion, Flow and Perfusion. Case Study Brochure Centre University Hospital Nancy.

This presentation is the intellectual property of the author. Contact them for permission to reprint and/or distribute.

This presentation is the intellectual property of the author. Contact them at for permission to reprint and/or distribute.

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

A new acute scaphoid fracture assessment method: a reliability study of the long axis measurement

Bone Densitometry Radiation dose: what you need to know

Dimensional stability in composite cone beam computed tomography

Prediction Of Pathological Fracture In The Proximal Femora With Bone Lesions Using A CT Scan-based Finite Element Method

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

Emerging Applications in Musculoskeletal CT Imaging

UvA-DARE (Digital Academic Repository) Treatment of osteochondral defects of the talus van Bergen, C.J.A. Link to publication

RADIOGRAPHY OF THE HAND, FINGERS & THUMB

RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

Preparing for Medical Physics Components of the ABR Core Examination

Introduction. The wrist contains eight small carpal bones, which as a group act as a flexible spacer between the forearm and hand.

Department of Surgery, Medical Centre Haaglanden, The Hague, the Netherlands 2. Department of Surgery, Gelre Hospitals, Apeldoorn, the Netherlands 3

Musculoskeletal Imaging What to order? Brian Cole, MD

Reproducibility of Three Dimensional Digital Preoperative Planning for the Osteosynthesis of Distal Radius Fractures

University of Groningen. Thoracolumbar spinal fractures Leferink, Vincentius Johannes Maria

X-ray (Radiography) - Bone

Orthopedic Hardware Imaging Part II: MRI v. Metal

Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study

Proceeding of the SEVC Southern European Veterinary Conference

Institutional review board approval was obtained prior to the start of this study.

Mayo Clinic Disorders of the Wrist

ORTHOSCAN MOBILE DI POSITIONING GUIDE

PERSPECTIVES OF THE TOOTH RESTORATION TECHNOLOGY BASED ON THE COMPUTED TOMOGRAPHY DATA

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Biomechanics of Two Reconstruction Techniques for Elbow Ulnar Collateral Ligament Insufficiency

Case Report A Case of Delayed Myelopathy Caused by Atlantoaxial Subluxation without Fracture

Initial Fixation Strength of Bio-absorbable Magnesium Screw

Translating Protocols Across Patient Size: Babies to Bariatric

OnTarget. OnTime. OnBudget. On-Site Extremity CT Exams Deliver Clinical Excellence and Superb Productivity.

Isolated Comminuted Fracture of the Trapezium: A Case Report and Review of the Literature

Morita 3D Accuitomo Scanning Protocol

CLINICAL PRESENTATION AND RADIOLOGY QUIZ QUESTION

3. Ulno lunate, Ulno triquetral ligament. Poirier: Between RSC &LRL. 5. Dorsal intercarpal ligament

Breast positioning system for full field digital mammography and digital breast tomosynthesis system

The Journal of the Korean Society of Fractures Vol.13, No.4, October, 2000

Volar Approach for Scaphoid Fractures Nader Paksima, DO, MPH Clinical Professor of Orthopaedic Surgery NYU/Hospital for Joint Diseases

COMMON CARPAL INJURIES IN ATHLETES Nicholas A. Bontempo, MD Orthopedic Associates of Hartford I HAVE NO CONFLICTS OR DISCLOSURES TO REPORT OUTLINE

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Pediatric cervical spine injuries with neurological deficits, treatment options, and potential for recovery

THE WRIST. At a glance. 1. Introduction

CLINICAL PAPER / ORTHOPEDIC

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

ORIGINAL PAPER. Department of Hand Surgery, Nagoya University School of Medicine ABSTRACT

PROPHECY. Preoperative Navigation Guides ANKLE CT SCAN PROTOCOL

Scaphoid imaging with digital tomosynthesis as an adjunct to Radiography: A Single Department's experience

DIAGNOSTIC IMAGING. OPTIMIZED.

Nuclear Associates

Posterolateral elbow dislocation with entrapment of the medial epicondyle in children: a case report Juan Rodríguez Martín* and Juan Pretell Mazzini

INVISION Total Ankle Replacement System with PROPHECY Preoperative Navigation Revision of a Failed Agility Total Ankle Replacement

Percutaneous Scaphoid Fixation: A Volar Approach

University of Groningen. Fracture of the distal radius Oskam, Jacob

Surgical Technique Carpal Fusion

Detection of clinically suspected scaphoid bone fractures using a dedicated cone-beam CT (CBCT). A retrospective study of 139 patients.

Fujifilm s Innovative Mammography AMULET Innovality Provides Optimal Tomosynthesis-Imaging for Diagnosis and Screening

CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report

Index. B Backslap technique depth assessment, 82, 83 diaphysis distal trocar, 82 83

Specialist Referral Service Willows Information Sheets. Limb deformity

The use of unidirectional porous β-tricarcium phosphate in surgery for calcaneal fractures: A report of four cases

.org. Tibia (Shinbone) Shaft Fractures. Anatomy. Types of Tibial Shaft Fractures

CT ARTHROGRAPHY It s not Always About the

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree

Common fracture & dislocation of the cervical spine. Theerachai Apivatthakakul Department of Orthopaedic Chiangmai University

Radiology. General radiology department. X-ray

X-Ray & CT Physics / Clinical CT

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

Extraction with Immediate Implant Placement and Ridge Preservation in the Posterior

Study of Lumbar Vertebrae with respect to the dimensions of the pedicle in South Indian population

Cover Page. The handle holds various files of this Leiden University dissertation.

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

AUA Guidelines for Imaging Known or Suspected Ureteral Calculi. Michael Ferrandino, MD Assoc Professor of Urology Duke University Medical Center

PEM GUIDE CHILDHOOD FRACTURES

Link to related CJSM article: ts Frequency_and.5.

Transcription:

R/F Clinical Experience Using the SONIALVISION safire II Utility of Tomosynthesis in Orthopedic Surgery Iwate Medical University Hospital, Central Department of Radiology Shouta Miura Mr. Shouta Miura 1. Introduction Iwate Medical University Hospital is located in Morioka City in the center of Iwate Prefecture, at the northern end of the Kitakami Basin. It is a large hospital with 39 medical departments offering 15 clinical beds and handles an average of 16 outpatients per day. The Iwate Private Medical School was established in 191. This became the Iwate Medical School in 1928, and finally Iwate Medical University in 1951. The hospital contributes to regional medical services based on its philosophy of Sound Spirit, Sound Medicine. Since 1956 it has added the Dental Hospital, Advanced Critical Care and Emergency Center, Hanamaki Onsen Hospital, and Memorial Heart Center. The School of Pharmacy was added in April 27. It is the only private medical university in the Tohoku region. (Fig. 1) In April 27, we introduced a Shimadzu SONIALVISION safire II system with a direct-conversion flat-panel detector (FPD) that can conduct tomosynthesis. Tomosynthesis is mainly used for orthopedic surgery at this hospital. This paper reports on basic investigations into metal artifacts in tomosynthesis of the scaphoid bone with Herbert screw fixation. It also introduces the effectiveness of the tomosynthesis image enhancement software that was introduced on a trial basis in October 28 on tomosynthesis images deemed to have clinical utility at this hospital. Fig. 1 Iwate Medical University Hospital (Mt. Iwate in the Background) 2. Basic Investigation into Metal Artifacts in Tomosynthesis 2.1 Background Scaphoid bones with Herbert screw fixation are traditionally evaluated using simple X-ray images and stress radiography, as artifacts make computed tomography (CT) inappropriate for these follow-up observations. As tomosynthesis is reported to be less affected by artifacts than CT 1), we evaluated the possibility of using it for follow-up observations of the scaphoid bone. Tomosynthesis artifacts vary due to the radiography parameters and image processing parameters. Positioning may also affect the area of interest. We conducted simulations of tomosynthesis on the scaphoid bone after screw fixation to determine the effects on artifacts of changing the radiography parameters and image processing parameters and the effects of the positional relationship of the metal in the X-ray tube scanning direction. 2.2 Method A water phantom containing a 3 mm-diameter metal rod representing a screw was positioned on the tabletop with the rod long axis parallel to the tabletop (Fig. 2). Tomosynthesis was conducted with radiography conditions 5 kv and 4 mas/pulse, 5 sec tomography time, and fixed 9-inch effective field size. FBP image reconstruction was used. The following parameters were changed: (1) Tomography angle (8, 2, 3, 4 ); (2) Inclination of metal rod long axis with respect to the X-ray tube scanning direction (, 3, 45, 6, 9 ); (3) Image reconstruction slice thickness (++, +, ±, -, --). The angles of (rod parallel to X-ray tube scanning direction), 45, and 9 were used for (2) inclination of metal rod long axis with respect to the X-ray tube scanning direction (Fig. 3). Parameter (3) image reconstruction slice thickness is an image processing parameter for frequency

band limitation 2) that is intended to reduce noise. Slice thickness (++) indicates the maximum slice thickness and (--) indicates the minimum slice thickness. Artifacts were analyzed using Image J image analysis software on DICOM data obtained from radiography. Data was read in a direction perpendicular to the metal rod long axis. It was used to create an artifact profile curve, with the pixel value along the vertical axis and pixel spacing along the horizontal axis (Fig. 4). Metal rod Fig. 2 Position of Metal Rod and Radiography Method Tube scanning direction X-ray tube Fig. 3 Angle of Metal Rod Long Axis Water Artifact Water phantom 45 9 Metal Water Artifact Tabletop FPD Metal 2.3 Results and Comments (1) Effect of tomography angle on artifacts The results show that the minimum artifact occurs at 4 tomography angle (Fig. 5). This tomography angle produces the minimum artifact because the projected information is acquired from multiple directions at high tomography angles. (2) Effect of inclination of metal rod long axis with respect to the X-ray tube scanning direction on artifacts The artifact effect decreases as the metal rod angle becomes closer to parallel to the X-ray tube scanning direction (Fig. 6). The inclination of the scanned X-rays increases as the metal rod angle deviates from, which increases the filter effect of the FBP method on both sides of the metal rod, resulting in undershoot artifacts. (3) Effect of image reconstruction slice thickness on artifacts Little artifact occurred at the greatest slice thickness (++) (Fig. 7). This is believed to result from reduced noise due to stronger artifact reduction processing for image reconstruction with greater slice thicknesses 2). However, if the slice thickness is set greater than the thickness of the radiography target, partial volume effects may result in indistinctness of the area of interest. Therefore care is required when determining the image reconstruction slice thickness for clinical applications. Consequently, to minimize artifacts in tomosynthesis of the scaphoid bone, this hospital positions the screw axis parallel with both the tube scanning direction and the tabletop and uses 4 tomosynthesis angle. The image reconstruction slice thickness parameter was set to the minimum thickness (--) to set a 1 mm slice pitch. Fig. 8 shows an actual tomosynthesis image of the scaphoid bone. The black artifact band apparent along the length of the screw does not affect on the evaluation of bone union at the margins of the scaphoid bone. As tomosynthesis permits the evaluation of screw fixation of the scaphoid bone, we do not currently conduct stress radiography. 6 5 4 8 2 3 4 Reading direction Pixel spacing 1 1 6 11 16 21 Fig. 4 Image J Reading Method and Profile Curve Fig. 5 Relationship Between Tomography Angle and Metal Artifact

6 3 5 45 6 9 4 1 1 6 11 16 21 Fig. 6 Relationship Between Inclination of Metal Rod Long Axis and Metal Artifact Fig. 9 Carpal Tunnel Radiograph ++ 6 + 5 ± -- 4 1 1 6 11 16 21 Fig. 7 Relationship Between Image Reconstruction Slice Thickness and Metal Artifact Fig. 1 Lateral Tomosynthesis of the Wrist Joint a b (a) Initial diagnosis (b) 2 months later: calcification disappeared Case 2 5-month-old female infant. Visited the hospital due to a congenital hip dislocation. The patient was cast-immobilized after a closed reduction of the hip dislocation. Tomosynthesis was performed, as observations were not possible by frontal general radiography of the hip joints due to interference by the cast. The tomosynthesis image clearly shows the femoral heads. It shows no displacement in the hip joints (Fig. 11). Fig. 8 Tomosynthesis Image of Scaphoid Bone with Herbert Screw Fixation 3. Clinical Cases Where Tomosynthesis Was Effective Case 1 29-year-old male. The patient visited the hospital complaining of unexplained pain in the right volar wrist during flexing for the past few months. Carpal tunnel radiography confirmed calcification at the ulnar side of the pisiform bone (Fig. 9), although it was not apparent in general radiographs from two directions. Lateral tomosynthesis of the wrist joint revealed a calcified deposit. Another tomosynthesis image taken two months later confirmed repair at the exfoliated area (Fig. 1). Fig. 11 Frontal Images of Infant Hip Joints (a) General radiograph (b) Tomosynthesis image a b Case 3 58-year-old female. A patient who had fallen down stairs at home was immediately hospitalized after CT confirmed a Jefferson fracture. The treatment for a Jefferson fracture is determined according to whether the transverse ligament is intact or broken. This is evaluated from the degree of lateral dislocation of the lateral mass of the atlas with respect to the lateral mass of the axis on both

sides 3). Frontal tomosynthesis of the cervical vertebrae was performed the day after hospitalization. It provided better evaluation of the lateral mass of the atlas and axis than general radiographs or CT (Fig. 12). Conservative treatment by Halo vest fixation was performed. Fig. 12 Frontal Tomosynthesis Image of the Cervical Vertebrae Lateral dislocation of the atlantal articular process Case 4 15-year-old male. Injured right hand in a fall from a bicycle. Visited our hospital one month later, after diagnosis of fracture of the scaphoid bone at another hospital. Tomosynthesis performed on arrival at the hospital confirmed a bone defect in the scaphoid waist. A bone graft with screw fixation and cast immobilization was performed two months after the injury. The fracture line in the scaphoid bone has become indistinct in the tomosynthesis image taken in the 16th week. This is assumed to result from progress of bone union (Fig. 13). a c 4. Tomosynthesis Image Enhancement Software In conventional tomosynthesis images, the graininess deteriorates at the specified sectional plane (center sectional plane, and approximately 2 mm above) in comparison with the other sectional planes. The phenomenon becomes more pronounced as the image reconstruction slice thickness becomes thinner. Shimadzu has installed trial image enhancement software to overcome this problem. The results are described below. 4.1 Improvement of Graininess at the Center Sectional Plane Due to the principle of operation of tomosynthesis, the minute sensitivity discrepancies between sensors are magnified at the center sectional plane (the virtual center of the linear trajectory) during image reconstruction calculations and this tends to cause deterioration in graininess. The new image enhancement software offers air calibration. Air calibration periodically conducts one normal tomosynthesis operation for each field size to acquire the optimal calibration data for each tomosynthesis scan trajectory in order to improve the graininess. Fig. 14 shows images before and after introduction of the software. The graininess of the overall image has clearly improved. This overcomes the problems at the thinnest image reconstruction slice thickness setting, and we currently use this setting without any problems. b Fig. 13 Frontal Tomosynthesis Image of the Scaphoid Bone (a) Preoperative: bone defect confirmed (arrow) (b) Week 8 postoperative (c) Week 16 postoperative: fracture line in scaphoid bone has become slightly indistinct (arrow) Fig. 14 Improvement in Graininess at the Tomosynthesis a b Center Sectional Plane (Knee Phantom) (a) Pre-modification (b) After air calibration 4.2 Enhanced Clarity of Reconstructed Images Grid movement control during tomosynthesis eliminates the need for the conventional moire elimination filter. It improves the horizontal resolution that was sacrificed when a moire elimination filter was used (Fig. 15).

4.3 Reduced X-Ray Exposure Dose The FPD high-gain mode achieves image quality equivalent to the conventional method using just one-half the radiography conditions. This results in reduced X-ray exposure dose. Direction of grid movement a b c Fig. 15 Enhanced Clarity of Reconstructed Images (a) Resolution chart (b) Pre-modification (c) With grid movement control 5. Conclusions At this hospital, tomosynthesis is mainly used for orthopedic surgery. We believe that the tomosynthesis images can confirm the diagnosis in many cases. Tomosynthesis is available on demand at this hospital, unlike the reservation system used for computed tomography (CT). Tomosynthesis can create an arbitrary sectional plane from a single scan operation. The examination is completed quickly, similar to general radiography. As the tomosynthesis function is installed on a fluoroscopy instrument, positioning is very easy under fluoroscopy. This achieves good reproducibility of the three-dimensional movements that are required for the scaphoid bone. Tomosynthesis images are useful for diagnosis, as FBP reconstructed images clearly reveal trabecular bone and the images can be further adjusted or reconstructed on a workstation to suit the diagnostic aim. The X-ray exposure dose is lower than computed tomography (CT) 4) and the newly developed image enhancement software further reduces the dose. Tomosynthesis is used together with plain bone X-rays in the field of orthopedic surgery. The opportunities to use functional imaging under fluoroscopy are also increasing. However, discrepancies occur between bone X-rays. We have many years' experience with CR images. How to create bone images of equivalent quality with the SONIALVISION safire II tomosynthesis function is an essential subject of future investigation. Computed tomography (CT) continues to rise in popularity. However, through its application to lung cancer diagnosis and the creation of 3D imaging systems, tomosynthesis is expected to form a new modality positioned between general radiography and CT. Acknowledgement I wish to express my gratitude to the clinical instruction I received from Dr. Katsuro Furumachi on the orthopedic surgery course at this university. References: 1) Shigeru Okamoto. Digital Tomosynthesis in the Thoracic and Orthopedic Fields by Direct-Conversion FPD. INNERVISION. 27; 22 (2): 14-16 2) Yukihiro Takumi. Development of a Digital Tomosynthesis Workstation. Shimadzu Review. 25; Vol. 61, 127-134 3) Osamu Kataoka, Shigeo Toya ed.: Clinical Study of the Upper Cervical Vertebrae. Nankodo. 2: 162 4) Hiroyasu Yano. The Effective Use of Tomosynthesis in Orthopedic Surgery -Follow-up after procedures using metal-. MEDICAL NOW. 26; No. 59, 11