Diagnostic approach to heart disease

Similar documents
PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING

Appendix II: ECHOCARDIOGRAPHY ANALYSIS

Adult Echocardiography Examination Content Outline

Cardiac ultrasound protocols

Doppler Basic & Hemodynamic Calculations

Introduction to Biomedical Imaging

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Pediatric Echocardiography Examination Content Outline

5 Working With Measurements

TEE Outside of the Cardiac OR

ASCeXAM / ReASCE. Practice Board Exam Questions Monday Morning

found that some patients without stenotic lesions had blood velocity or pressure measurement across the

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

Data Collected: June 17, Reported: June 30, Survey Dates 05/24/ /07/2010

The Doppler Examination. Katie Twomley, MD Wake Forest Baptist Health - Lexington

Fig.1 Normal appearance of RV in SAX:

cardiac imaging planes planning basic cardiac & aortic views for MR

Assessment of LV systolic function

Echocardiography in Adult Congenital Heart Disease

Ultrasound 10/1/2014. Basic Echocardiography for the Internist. Mechanical (sector) transducer Piezoelectric crystal moved through a sector sweep

Concepts of Imaging and Knobology

Tissue Doppler Imaging

Tissue Doppler Imaging in Congenital Heart Disease

Physical Principles of Ultrasound

Uncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency

Introduction to Fetal Doppler Echocardiography

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

COMPREHENSIVE EVALUATION OF FETAL HEART R. GOWDAMARAJAN MD

Anatomy & Physiology

Department of Pediatrics and Child Health, Kurume University School of Medicine, Kurume, 830 Japan. Received for publication October 26, 1992

Ultrasound Physics & Terminology

Basic Approach to the Echocardiographic Evaluation of Ventricular Diastolic Function

Imaging Guide Echocardiography

PRACTICAL ECHOCARDIOGRAPHY IN THE ADULT with Doppler and color-doppler flow imaging

Basic Physics of Ultrasound in Transesophageal Echocardiography

Certificate in Clinician Performed Ultrasound (CCPU) Syllabus. Rapid Cardiac Echo (RCE)

HISTORY. Question: What category of heart disease is suggested by the fact that a murmur was heard at birth?

Right Ventricle Steven J. Lester MD, FACC, FRCP(C), FASE Mayo Clinic, Arizona

Background & Indications Probe Selection

Heart and Lungs. LUNG Coronal section demonstrates relationship of pulmonary parenchyma to heart and chest wall.

ISUOG Basic Training. Obtaining & Interpreting Heart Views Correctly Alfred Abuhamad, USA. Basic training. Editable text here

ULTRASOUND OF THE FETAL HEART

Making Sense of Cardiac Views and Imaging Characteristics for 13 Congenital Heart Defects (CHDs)

Standardising echocardiography and images. Version 2, 13/04/15

Basic Ultrasound Physics Board Review Questions

Practical Echocardiography: ECHOES in the REAL WORLD Know When to Hold Em and When to Fold Em

Normal TTE/TEE Examinations

Echocardiography Conference

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

ECHOCARDIOGRAPHY CHAPTER INTRODUCTION

Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Methods Forty patients representing all consecutive patients referred to. our laboratory on randomly selected days were chosen for

Principles of echocardiography for the anesthesiologist.

The Normal Echocardiogram

B-Mode measurements protocols:

Pulsed Wave Doppler and Color Flow Doppler Evaluation in Healthy Dogs and Dogs with Cardiac Disease

Journal of American Science 2014;10(9) Congenital Heart Disease in Pediatric with Down's Syndrome

Diastolic Function: What the Sonographer Needs to Know. Echocardiographic Assessment of Diastolic Function: Basic Concepts 2/8/2012

ECHOCARDIOGRAPHY DATA REPORT FORM

Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016

Doppler-echocardiographic findings in a patient with persisting right ventricular sinusoids

Left atrial function. Aliakbar Arvandi MD

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

Uptofate Study Summary

Pulmonary arteriovenous fistula

Congenital heart disease. By Dr Saima Ali Professor of pediatrics

Adel Hasanin Ahmed 1

MITRAL STENOSIS. Joanne Cusack

British Society of Echocardiography

Carlos Eduardo Suaide Silva, Luiz Darcy Cortez Ferreira, Luciana Braz Peixoto, Claudia Gianini Monaco, Manuel Adán Gil, Juarez Ortiz

Terminology Tissue Appearance

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

AIMI-HF PROCEDURE MANUAL TECHNICAL GUIDE FOR ECHOCARDIOGRAPHY. MHI Core Laboratory E. O Meara - J.C. Tardif J. Vincent, G. Grenier, C.

Quantitation of right ventricular dimensions and function

Doppler echocardiographic evaluation of the normal

Pulmonary Hypertension: Echocardiographic Evaluation of Pulmonary Hypertension and Right Ventricular Function. Irmina Gradus-Pizlo, MD

Gilles HANTON, BVSc, DVM, DABT, ERT GH Toxconsulting Brussels, Belgium

Ultrasound Principles cycle Frequency Wavelength Period Velocity

Optimising your Doppler settings for an accurate PI. Alison McGuinness Mid Yorks Hospitals

HISTORY. Question: What type of heart disease is suggested by this history? CHIEF COMPLAINT: Decreasing exercise tolerance.

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

Nancy Goldman Cutler, MD Beaumont Children s Hospital Royal Oak, Mi

VMS Quick Reference Guide

Congenital Heart Defects

Introduction to TEE using Heartworks Echocardiography Simulator

ECHOCARDIOGRAPHIC APPROACH TO CONGENITAL HEART DISEASE: THE UNOPERATED ADULT

Doppler Color Flow Imaging #4

Case # 1. Page: 8. DUKE: Adams

PROSTHETIC VALVE BOARD REVIEW

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Echocardiography in Congenital Heart Disease

HAND held echocardiography a dangerous toy?

Introduction & Physics of ED Ultrasound. Objectives. What? - Limited Studies. Who? - ED Docs

Supplement (videos)

Cardiology Fellowship Manual. Goals & Objectives -Cardiac Imaging- 1 P a g e

Echocardiography in Intensive Care: The Basics. Part I

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy.

Vevo 2100 System Cardio Measurements. Dieter Fuchs, PhD FUJIFILM VisualSonics, Inc.

CONGENITAL HEART DISEASE (CHD)

'circular shunt'1. CASE 1 Shortly after birth a 36-hour-old, full-term infant girl showed cyanosis and dyspnoea. Physical

Transcription:

Diagnostic approach to heart disease Initial work up History Physical exam Chest radiographs ECG Special studies Echocardiography Cardiac catheterization

Echocardiography principles Technique of producing images of the heart by means of reflected sound waves (echoes) Humans can hear sound waves with frequencies between 20 Hz and 20 Kilohertz (khz), frequencies higher than this range are termed ultrasound

Amplitude Principles of ultrasound Sound is a disturbance of a physical medium Propagated as a wave that represents local oscillations in air pressure Sound properties (physical characteristics) Frequency-the number of oscillations per unit of time, high frequency=shorter wavelengths; low frequency=longer wavelength Wavelength-distance between successive troughs or peaks (mm) wavelength Cycles per second Hz

Echocardiography- image creation summary Ultrasound beam strikes a structure and some of it is reflected back to the transducer Transducers usually use pulse wave US, in which the crystal turns on for a short time and then waits for a longer period for the reflected waves to return When the reflected US strikes the crystal it oscillates the crystal which is turned into a voltage detected by the machine

Echocardiography- image creation summary (continued) By using computer technology an US machine can detect numerous structures at once, creating a two-dimensional image. The machine can calculate the distance between the transducer and the structure by measuring the time it takes for the US beam to travel to the structure of interest and return to the transducer.

Types of echocardiography M-mode (motion mode) Two dimensional (2-D) Doppler mode

Two dimensional echo (2DE) Most often performed and easiest to understand Utilizing an ultrasound beam moving in a sector, 2DE creates a pie or fan shaped image Rapid repetitive interrogation by the transduce results in appearance of real time motion Displays anatomic and functional characteristics that are more anatomically intuitive than M-mode imaging.

Two dimensional echo Many potentially useful planes 4 chamber refers to visualization of the RA, RV, LA, LV 5 chamber adds the proximal aorta Long axis=sagital image planes that parallel the longitudinal axis Short axis=transverse images (horizontal)

Echocardiographic views RV=right ventricle LV=left ventricle RA=right atrium LA=left atrium AO=aorta

Echocardiographic views Short axis views Long axis views RV=right ventricle LV=left ventricle RA=right atrium RV=right ventricle LA=left atrium AO=aorta

Short axis 2DE loops

Long axis 2DE loop

Left apical views

Left apical views

Two dimensional echo Advantages Ability to detect motion perpendicular to beam, display size and shape of structures Disadvantages Slower frame rate than M-mode results in lower image quality

M-mode echocardiography M-mode imaging utilizes only a single crystal to produce a single beam ( ice pick ) which is pulsed at a high frequency at the heart The information from successive US pulses is displayed before the preceding lines have decayed and the result is an image in which the x-axis is time and the y-axis is distance from the transducer

M-mode echocardiography Advantages Rapid sampling rate results in high resolution Most accurate measurement of intracardiac dimensions and structures and accurate timing of cardiac events (valve closure, wall motion, etc.) Disadvantages The one dimensional image makes it difficult to appreciate spatial relationships and 3-D shapes of objects.

M-mode echocardiography M-mode cursor

M-mode uses Ventricular lumen size and wall thickness Shortening fraction SF= LVIDd- LVIDs x 100 LVIDd Index of contractility Significantly altered by changes in afterload

M-mode echocardiography

M-mode echocardiography M-mode cursor

M-mode echocardiography

Doppler echocardiography When a propagating wave is reflected from a moving object, a shift in frequency results. This frequency shift is related in a mathmatical manner to the velocity of the moving object and is known as the Doppler principle. The computer can use this formula and the Doppler shift to calculate direction and velocity of moving parts (i.e. red blood cells or cardiac wall).

velocity Doppler echocardiography A positive shift means flow is moving toward the transducer and a negative shift means blood is moving away. Positive shift Negative shift

Doppler Echocardiography Pulsed wave Doppler Color Flow and Spectral Assessment of blood flow in a specific region, but cannot assess high velocity flows Continuous wave Doppler Assessment of blood flow continuously along the cursor line Capable of measuring high velocity flows

Doppler echocardiography As the intercept angle, theta, deviates from 0, velocity will be underestimated The 4 and 5 chamber 2DE views optimize the beam angle (for assessment of the MV, TV, and AOV) to minimize this error.

Tricuspid valve Mitral valve

Aliasing phenomenon Nyquist limit=maximal unambiguous velocity measurement Nyquist limit is exceeded when: Doppler shift frequency>pulse repetition frequency/2 Therefore, the higher the blood flow velocity and the greater the depth, the more likely aliasing will occur with PW Doppler

Pulsed wave Doppler Advantages: Interrogation of blood velocity at a specific site Disadvantages: As depth of interrogation increases, and blood velocity increases, the Nyquist limit may be exceeded resulting in aliasing, and the flow pattern is ambiguous.

Aortic flow Pulmonary artery flow

Aliasing Pulse Wave, MV flow LA Doppler cursor

Aliasing- color flow Doppler

Continuous wave Doppler Advantages: Two crystals within the probe are constantly sending and receiving data so high velocity flow can be accurately measured. Disadvantages: Flow velocity is measured along the entire Doppler beam, a specific region cannot be individually interrogated.

Continuous wave Doppler

Continuous wave Doppler

Continuous wave Doppler

Modified Bernoulli Equation Pressure gradient=4v 2

Modified Bernoulli Equation Pressure gradient=4v 2 PG=4(5) 2 PG=4(25) PG=100

Special Echocardiographic techniques Contrast echocardiography Gas bubbles in agitated saline are hyperechoic and may help locate abnormal flow patterns such as shunts Bubbles are too large to cross capillary beds Most typically used for right to left shunts: Patent foramen ovale, Tetrology of Fallot, Pulmonic stenosis and atrial septal defect, or R to L patent ductus arteriosus

Contrast echocardiography