KATALYST. Bipolar Radial Head System. Surgical Technique. orthopedics. KATALYST English. PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

Similar documents
Bipolar Radial Head System

Integra. Katalyst Bipolar Radial Head System SURGICAL TECHNIQUE

Integra. Modular Radial Head System SURGICAL TECHNIQUE

the shape, the size, the fit Ascension Modular Radial Head

Integra. Modular Radial Head System SURGICAL TECHNIQUE

Integra. Solustaple Standard staple SURGICAL TECHNIQUE. Uni-Clip Compression staple. Products for sale in Europe, Middle-East and Africa only.

Rehabilitation after Total Elbow Arthroplasty

modular RADIAL HEAD E VOLVE

rhead System Extended stems Operative technique

rhead System Radial Head Arthroplasty Operative technique

E-CENTRIX. Ulnar Head Replacement SURGICAL TECHNIQUE

Integra. surgical technique. Advansys Midfoot Plating System. eng. D.L.P. Dorsal Lisfranc Plate. M.L.P. Medial Lisfranc Plate

UDHT08.1.qxd:UDHT /03/08 17:14 Page 1. Surgical. Technique. Elbow Prosthesis. RHS Radial Head System.

Integra. Silicone PIP Implant SURGICAL TECHNIQUE

UNI-Elbow System Radio-Capitellar Replacement. Operative technique

Locking Radial Head Plates

Ascension. Silicone MCP surgical technique. surgical technique Ascension Silicone MCP

Modular Ulnar Head surgical technique. Transforming Extremities

Integra. Titan Modular Shoulder System, 2.5

Elbow Elbow Anatomy. Flexion extension. Pronation Supination. Anatomy. Anatomy. Romina Astifidis, MS., PT., CHT

Radial Head Fractures Save or Replace?

Integra. DigiFuse Cannulated Intramedullary Fusion System SURGICAL TECHNIQUE

Surgical. Technique CRF II. Radial Head Prosthesis

EXTERNAL FIXATION SYSTEM

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Case Presentation: Comminuted Radial Head Fracture

Radial Head Prosthesis System

LCP Distal Humerus Plates

SURGICAL TECHNIQUE GUIDE

Functional Anatomy of the Elbow

Recurrent subluxation or dislocation after surgical

Integra. Thumb Joint Replacement PATIENT INFORMATION. with PyroCarbon implants

Technique Guide Small Bone Fusion System

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

Ascension PIP Post-Operative therapy protocol

Conventus CAGE PH Surgical Techniques

Visualize, stabilize, mobilize. Wristore * Distal Radius Fracture Fixator Abbreviated Surgical Technique

Case Presentation: Comminuted Fractures of the Proximal Ulna 11/28/2017. Disclosures. Surgical Strategy. Implant Choice. Melvin P.

ACETABULAR CUP SURGICAL TECHNIQUE

WEEKEND 2 Elbow. Elbow Range of Motion Assessment

Anterior Elbow Capsulodesis

Silicone PIP, MCP & MCP-X (PreFlex)

RADIAL HEAD PROSTHESIS SYSTEM For primary and revision joint replacement of the radial head

Surgical Technique. DISCLOSURE: This device is not approved for sale in the U.S.A. Customer Service:

2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Optimum implant geometry

Fractures and dislocations around elbow in adult

MANAGEMENT OF INTRAARTICULAR FRACTURES OF ELBOW JOINT. By Dr B. Anudeep M. S. orthopaedics Final yr pg

Elbow dislocations represent 10% to 25% of all injuries. Elbow Fracture-Dislocations. The Role of Hinged External Fixation

Terrible triad of the elbow

Integra. Salto Talaris Total Ankle Prosthesis PATIENT INFORMATION

Simply Complete. VA-LCP Elbow Plating System.

Total Knee Original System Primary Surgical Technique

Integra. Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

Wright Medical Technology, Inc Cherry Road Memphis, TN

Elbow Anatomy, Growth and Physical Exam. Donna M. Pacicca, MD Section of Sports Medicine Division of Orthopaedic Surgery Children s Mercy Hospital

TIBIAXYS ANKLE FUSION

Integra. Freedom Wrist Arthroplasty System SURGICAL TECHNIQUE

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

A Patient s Guide to Adult Radial Head (Elbow) Fractures

System. Humeral Nail. Surgical Technique

Surgical Technique. Forearm Fracture Solutions

Integra. Because we are committed to limiting uncertainty, we provide a specific design dedicated to Pediatric procedures.

Bone Preservation Stem

MIS Cemented Tibial Component

large qwix Positioning and fixation screws Surgical Technique lower large qwix English PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

Surgical Technique Guide

Integra. Silicone MCP Joint Prosthesis SURGICAL TECHNIQUE

The Bio-Modular Choice Shoulder System,

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

U2 PSA. Revision Knee. Surgical Protocol

AcUMEDr. Locking Proximal Humeral Plate. PoLARUSr PHPt

THE P.F.C. SIGMA FEMORAL ADAPTER. Surgical Technique

Osteology of the Elbow and Forearm Complex. The ability to perform many activities of daily living (ADL) depends upon the elbow.

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique IMAGE TO COME

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

MICRONAIL. Intramedullary Distal Radius System SURGICAL TECHNIQUE

Progeny Hip Stem. Surgical Protocol and Product Specifications

SURGICAL TECHNIQUE. Alpine Cementless Hip Stem

Adam J. Seidl, MD Assistant Professor University of Colorado School of Medicine Shoulder & Elbow Surgery Division of Sports Medicine and Shoulder

Bi-Polar 22.2mm & 28mm System - Operative technique

Elbow. Chapter 2 LISTEN. Mechanism of Injury (If Applicable) Pain

Forearm Fracture Solutions. Product Overview

WristMotion Wrist Hemiarthroplasty System Instructions for Use

A Locking IM Rod that won't back out. Simple and straight to the point! SURGICAL TECHNIQUE

Integra. Solutions for Hand and Plastic Surgeons. Products for sale in Europe, Middle-East and Africa only

Hand and wrist emergencies

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

THE ELBOW. The elbow is a commonly injured joint in both children and adults.

Pinit Plate Small Bone Fusion System Bone Plate & Screw System

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

Transcription:

KATALYST Bipolar Radial Head System KATALYST English Surgical Technique orthopedics UPPER extremity PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY

KATALYST Introduction Description The Katalyst Telescoping Bipolar Radial Head implant restores the support and bearing surface of the radial head in the face of fracture, arthritis or failed prior radial head implantation. The Katalyst system offers three unique features Telescoping shaft allows correct length and tension of the soft tissues. Bipolar neck provides optimal load transfer to the humerus and decrease the potential for capitellar wear. Anatomic curvature of the head recreates normal anatomy. These features allow the surgeon to recreate appropriate tension in situ with an implant that articulates with the capitellum and sigmoid notch in an anatomic fashion. In addition, the modular design allows for radial head replacement without surgical disruption of the lateral ligament complex from the humerus. Indications Comminuted radial fracture with or without associated elbow instability Comminuted radial fracture with associated failure of the interosseous membrane; the Essex-Lopresti lesion Post-traumatic or primary osteoarthritis involving the radiocapitellar or proximal radioulnar joint Rheumatoid arthritis involving the radiocapitellar joint or proximal radioulnar joints Revision of failed radial head implant Contraindications Inadequate bone and surrounding soft-tissues to accept and support the implant Neurological, metabolic or other medical conditions that might compromise the bone, soft tissues or proprioception about the elbow joint Open physes in the radius, ulna and distal humerus Infection 2 Surgical Technique KATALYST PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST AND AFRICA ONLY

Integra as the manufacturer of this device, does not practice medicine and does not recommend this or any other surgical technique for use on a specific patient. The surgeon who performs any implant procedure is responsible for determining and using the appropriate techniques for implanting the device in each patient. Surgical technique Step 1 Assessment of elbow stability Valgus stability, posterolateral stability and longitudinal stability are determined using commonly accepted practices such as preoperative radiographs, intraoperative inspection, and stress radiographs. Fig. 1 Surgical approach After the skin incision is performed (Fig. 1), the radial head is exposed via a modified lateral approach designed to preserve the lateral ligaments. The approach splits the thick common extensor tendon origin in the midline. Step 2 Exposure The appropriate plane can be identified by palpating the lateral condylar ridge and radial head. A full-thickness incision is made through the common tendon and ligament origin along the longitudinal axis of the radial head and neck (Fig. 2). The origin of the radial wrist extensors and the joint capsule may be elevated from the anterolateral aspect of the distal humerus as needed for exposure. Extend the split of the lateral complex distally to the metaphysis of the radial neck. In the atraumatic setting, the exposure can be performed without destabilizing the elbow since the posterior half of the extensor origin and lateral ligament complex remain intact. Fig. 2 NOTE: When the radial head has fractured in association with a dislocation of the elbow, the lateral complex will often be stripped from its origin on the lateral epicondyle. This should be noted during the exposure, and a plan should be formulated to repair or reconstruct the complex at the end of the case. (Text on file) Surgical Technique KATALYST PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST AND AFRICA ONLY 3

Step 3 Radial head resection Radial head resection at the proximal articular margin at this level will prevent impingement between the cut end of the proximal radius and the sigmoid notch. An additional 3-4 mm of the proximal radius is resected for the implant. The appropriate location for resection can be determined by using the Cutting Template (Fig. 3); this location corresponds with the disection of the head using the 1mm (4.5 ) shim. Fig. 3 NOTE: Once the cut in the neck is made, measure the distance from the coronoid to the cut edge of the proximal radius and re-create that distance with the head and neck (see Step 7). 3-4 mm additional cut Step 4 Opening intramedullary canal The Starter Awl is used to open the intramedullary canal of the proximal canal. (Fig. 4) radial head Step 5 Broaching The combination Broach with End Cutting Mill is used to create a smooth cut end of the proximal radius intramedullary canal that is perpendicular to the long axis of the radius (Fig. 5) It is recommended to start with the 6.5mm Broach prior to using the 7.5mm Broach. Step 6 Choosing trial head Once the canal is broached up to the appropriate size, the corresponding Trial Stem is placed into the intramedullary canal with the black line on the side of the trial stem facing the surgeon. The Broaches and trial stems are color coded for convenience. The appropriate diameter Trial Head is chosen based on pre-operative templating and/or reassembly of the fracture components using the Head Sizing Fixture as a template (Fig. 6). In cases where there is a question between 2 head sizes, the smaller implant is chosen. Fig. 4 Fig. 5 Fig. 6 4 Surgical Technique KATALYST PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST AND AFRICA ONLY

Step 7 Using spacers to adjust separation Once the Trial Head is snapped in place on the stem, the two components are extended to re-create the appropriate distance from the coronoid, to the edge of the proximal radius as determined in Step 3. This separation is achieved by inserting the First Spacer which will position the implant in the first extended position. For each additional extension required, 2mm Spacers are used (Fig. 7). Fig. 7 NOTE: The forearm must be in 90 of flexion and at full pronation when determining the appropriate tension via the spacer forks. Step 8 Appropriate separation Determining the appropriate separation in the head-neck construct is more difficult when the lateral complex has torn. This can be accomplished by provisional reduction of the tendon and ligament origin to the humerus with the ulno humeral joint reduced by flexion and pronation of the forearm. Length can also be checked by evaluating the radial head implant relative to the measurements between the coronoid and the cut edge of the proximal radius (as determined in Step 3). The proximal margin of the implant should line up with the articular surface of the coronoid when viewed from the lateral side. Step 9 Locking the stem Once the appropriate tension is achieved, the Trial Stem is locked in place by turning the Set Screw in a clockwise fashion for 3/4 of a turn (Fig. 8). Fig. 8 NOTE: The set screw in the trial is captured in place and cannot be removed. Surgical Technique KATALYST PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST AND AFRICA ONLY 5

Step 10 Implant alignment Intraoperative radiographs may be used to confirm appropriate size and position of the implant. The stem should be aligned with the longitudinal axis of the radius. The head should match the capitellum and seat in a congruent fashion against the sigmoid notch. The medial joint space of the ulnohumeral articulation should be symmetrical and not converge on the anteroposterior projection (Fig. 9).; if so, the elbow may have been forced into a valgus position by an overstuffed implant (Fig. 10). Move the elbow through a full arc of flexion and fore arm rotation. A lack of motion may be additional evidence of an oversized or over-distracted trial implant. Fig. 9 / 10 Fig. 11 Step 11 Stem insertion The Trial Implant is removed and the Permanent Implant of the same size is placed. Prior to placing the Stem in the surgical field, the Set Screw is removed from the side of the Stem. The Stem is fully inserted into the canal, again ensuring that the black line on the implant faces the surgeon (Fig. 11). Step 12 Attaching radial head The appropriate sized Head is placed over the spherical end of the Stem. There is a hex feature on both the Head and the Stem. By placing one Hex Driver in each of these two holes and pinching the two Hex Drivers together, the Head will snap in place on the stem. Pliers are provided to assist with pinching the two drivers together when snapping the Head onto the Stem. Fig. 12 NOTE: DO NOT DISASSEMBLE Once the head has been assembled to the stem, disassembly of these components will adversely affect retention strength between them and is not recommended. 6 Surgical Technique KATALYST PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST AND AFRICA ONLY

Step 13 Set screw locking Once in the correct position, the Set Screw is placed back into the stem, taking care not to over-torque the screw. Finally, the Spacers are removed from the implant. Repeat radiographs and intra-operative stress testing are now performed. Step 14 Repairing If the posterior half of the lateral complex is intact, the anterior half is repaired to the intact posterior limb. If the lateral complex was avulsed or released, it is repaired to the lateral epicondyle. The skin is closed and the elbow is placed in a posterior splint. Step 15 Final radiograph In cases of unstable elbow fracture dislocations it is recommended that a lateral radiograph of the splinted elbow be obtained in the splint to ensure that a concentric reduction of the elbow has been maintained. Postoperative care Elbow motion and forearm rotation are begun in the early postoperative period. Rehabilitation is guided by intraoperative stability. Serial follow-up radiographs should be examined for evidence of implant malposition or wear. Sclerosis in the bone at the implant-bone interface is evidence of appropriate load transfer through the implant. NOTE: REVISION OR REMOVAL To disengage the head from the stem, place one hex driver in the head component and a second hex driver in the stem component and lever the hex driver in the head away from the driver in the stem until the head is disassembled. Reusing a head component that has been previously assembled to a stem is not recommended as the retention force will be compromised. Instrument tray A B C D E Pliers Cutting Template Head Sizing Fixture Starter Awl 7.5 mm Stem Broach A B C G D E F H L-P F G 6.5 mm Stem Broach 7.5 Trial Stem Q R H I 6.5 Trial Stem 24 mm Trial Head I J K J 21 mm Trial Head K 18 mm Trial Head O 2 mm Spacer L First Spacer (1 mm) P 2 mm Spacer M 2 mm Spacer Q 2.5 mm Hex Driver N 2 mm Spacer R 2.5 mm Hex Driver Surgical Technique KATALYST PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST AND AFRICA ONLY 7

KATALYST Bipolar Radial Head System Features In situ telescoping shaft utilizes a First Spacer (1mm) and then spacers increase in 2 mm increments up to 10 mm. Two hex drivers are included in the instrument set to facilitate manipulating, assembling and disassembling both trials and implants. Cutting template tool available (Radial Head Sizing Block) to assist in accurate resection measurement for radial head implant. These features allow the surgeon to recreate appropriate tension in situ in the supporting ligaments of the elbow with an implant that articulates with the capitellum and sigmoid notch in an anatomic fashion. In addition, the modular design allows for radial head replacement without surgical disruption of the lateral ligament complex from the humerus. Component materials Telescoping implant with Multiple Head and Stem Sizes Instrument set Head and poly Assembly: Cobalt Chrome with polyethylene liner Stem Assembly: Stainless Steel 24 mm 21 mm 18 mm Part N Description Size 22-1418 Head and Poly Assembly 18 mm 22-1421 Head and Poly Assembly 21 mm 22-1424 Head and Poly Assembly 24 mm 7.5 mm 6.5 mm 22-1665 Stem Assembly 65 mm 22-1675 Stem Assembly 75 mm 22-1001 Instrument Set Distributed by EC REP Integra LifeSciences Services (France) SAS Sales & Marketing EMEA Immeuble Séquoia 2 97 allée Alexandre Borodine Parc technologique de la Porte des Alpes 69800 Saint Priest FRANCE +33 (0)4 37 47 59 00 fax +33 (0)4 37 47 59 99 emea.info@integra-ls.com www.integra-ls.com Customer Service International: +33 (0)4 37 47 59 50 +33 (0)4 37 47 59 25 (Fax) csemea@integra-ls.com France: +33 (0)4 37 47 59 10 +33 (0)4 37 47 59 29 (Fax) custsvcfrance@integra-ls.com Benelux: +32 (0)2 257 4130 +32 (0)2 253 2466 (Fax) custsvcbenelux@integra-ls.com Switzerland: +41 (0)2 27 21 23 30 +41 (0)2 27 21 23 99 (Fax) custsvcsuisse@integra-ls.com Integra LifeSciences Corporation 4900 Charlemar Drive Cincinnati, OH, 45227 USA Availability of these products might vary from a given country or region to another, as a result of specific local regulatory approval or clearance requirements for sale in such country or region. Always refer to the appropriate instructions for use for complete clinical instructions. Non contractual document. The manufacturer reserves the right, without prior notice, to modify the products in order to improve their quality. WARNING: Applicable laws restrict these products to sale by or on the order of a physician. Katalyst and the Integra wave logo are trademarks of Integra LifeSciences Corporation. 2009 Integra LifeSciences Corporation. All rights reserved. ILS 08-07-065-01-09 PRODUCTS FOR SALE IN EUROPE, MIDDLE-EAST and AFRICA ONLY