Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase

Similar documents
Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

UREA CE Insert. 01 English - Ref.: 27. Ref.:27

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

Estimation of Serum Urea. BCH472 [Practical] 1

Enzymatic Assay of PYRUVATE KINASE (EC ) From Rabbit Liver

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

PhosFree TM Phosphate Assay Biochem Kit

liquicolor Design Verification

1.2 Synonyms There are several synonyms e.g. diaminomethanal, but in a medical context, this substance is always referred to as urea.

Date... Name... Group... Urine sample (Tube No 2)

Creatinine (serum, plasma)

Enzymatic Assay of CHOLINE KINASE (EC )

Enzymatic Assay of RIBOKINASE (EC )

Estimation of Serum Urea

Enzymatic Assay of PHOSPHODIESTERASE, 3':5'-CYCLIC NUCLEOTIDE Crude Complex

Enzymatic Assay of GLUCONATE KINASE (EC ) ß-NADPH = ß-Nicotinamide Adenine Dinucleotide Phosphate,

NIH Public Access Author Manuscript Biochem Cell Biol. Author manuscript; available in PMC 2010 January 11.

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

The Adiponectin Turbidimetric Immunoassay Reagent Kit

RUBISCO > 2 moles of 3-phosphoglycerate Mg +2

Enzymatic Assay of CREATININASE (EC ) From Pseudomonas species

Enzymatic Assay of NAD-PYROPHOSPHORYLASE (EC )

Enzymatic Assay of GUANYLATE KINASE (EC )

Blood Urea Nitrogen Enzymatic Kit Manual Catalog #:

liquicolor (AMP Buffer, IFCC) Design Verification

Enzymatic Assay of FRUCTOSE-6-PHOSPHATE KINASE, PYROPHOSPHATE DEPENDENT (EC ) from Mung Bean

Enzymatic Assay of URIDINE 5'-DIPHOSPHOGALACTOSE 4-EPIMERASE (EC )

LIPASE liquicolor. Design Verification. Multipurpose Reagent

Oxalate (urine, plasma)

DIRECT ENZYMATIC HbA1c ASSAY

Urea Nitrogen (BUN) detection Kit

RENAL FUNCTION ASSESSMENT ASSESSMENT OF GLOMERULAR FUNCTION ASSESSMENT OF TUBULAR FUNCTION

TECHNICAL BULLETIN. Sialic Acid Quantitation Kit. Catalog Number SIALICQ Storage Temperature 2 8 C

3 Linearity and Detection Limit Linearity Detection limit... 4

Creatinine Test Code: CRE

Micro Bioactivity Analyzer

Original Article. A novel method for measuring serum ornithine carbamoyltransferase. Introduction. Materials and methods. Abstract

NADH Assay Kit (Red) Catalog Number KA assays Version: 07. Intended for research use only.

TABLE OF CONTENTS GENERAL INFORMATION... 1

BIOO LIFE SCIENCE PRODUCTS

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Enzymatic Assay of GALACTOSYLTRANSFERASE (EC )

BCH 447. Estimation of Serum Urea

Evaluation of Cheongmeak DCS TM Reagents for Chemistry Analyzers

BASIC ENZYMOLOGY 1.1

Glycosyltransferase Activity Kit

Enzymatic Assay of PHOSPHORYLASE KINASE (EC )

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Estimation of Serum Urea and Urine Urea. Amal Alamri

LDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings

Triglycerides (as Glycerol) in Serum with the Gilford System 3500

2-Deoxyglucose (2DG) Uptake Measurement kit

A New Mass Screening Method of Detecting UDP-Galactose-4-Epimerase

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

J. Nutr. Sci. Vitaminol., 38, , Note. in Tissues

ab Sodium Assay Kit (Colorimetric)

ESS Method 310.2: Phosphorus, Total, Low Level (Persulfate Digestion)

Handheld Fluorometers

THE UNIVERSITY OF NEWCASTLE- DISCIPLINE OF MEDICAL BIOCHEMISTRY

kinetic methods for transaminase assay

Methods of Enzyme Assay. By: Amal Alamri

Clinical Chemistry (CHE 221)

KE-SIALIQ Sialic Acid Quantitation Kit. SialiQuant Sialic Acid Quantitation Kit

ab Urease Activity Assay Kit (Colorimetric)

CLINICAL CHEMISTRY REAGENTS. Product Profile

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry

Determination of hemoglobin is one of the most commonly

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

High-density Lipoprotein Cholesterol (HDL-C) Assay Kit

MATERIAL AND METHODS

1. Reagent Store at 2-8 ºC.

PublisherName : BioMed Central. PublisherLocation : London. PublisherImprintName : BioMed Central. ArticleID : 313. ArticleDOI : 10.

Blood Urea Nitrogen Enzymatic Kit Manual Catalog #:

Journal of Clinical Laboratory Instruments and Reagents, Vol. 34, No. 3 (June, 2011) Supplement

human Total Cathepsin B Catalog Number: DY2176

Ascorbic Acid Assay Kit

EXPERIMENT 3 ENZYMATIC QUANTITATION OF GLUCOSE

The Third Department of Internal Medicine, University of Tokyo Faculty of Medicine, Hongo, Tokyo 113

Mechanisms of dopamine and dobutamine interference in biochemical tests that use peroxide and peroxidase to generate chromophore

ab65656 Ascorbic Acid Assay Kit (Colorimetric)

CHAPTER 5 MEASUREMENT OF UREA AND URIC ACID IN BLOOD SERUM

Instructions for Use. Tg Antibody ELISA

D-FRUCTOSE and D-GLUCOSE

Fructose-6-Phosphate Colorimetric Assay Kit

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana

OxisResearch A Division of OXIS Health Products, Inc.

Research Article Derivative Spectrophotometric Method for Estimation of Metformin Hydrochloride in Bulk Drug and Dosage Form

Aspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #:

OraSure Technologies, Inc Eaton Avenue, Bethlehem, PA Phone: (610) (USA) Fax: (610)

Glutathione Assay Kit

Human Creatinine Urinary Detection Kit

DetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

A150 Product Number 31150

NITROGEN METABOLISM An Overview

Β-FRUCTOFURANOSIDASE ENZYME

Urinary Assay Using Micro Flow Injection Analysis

Glucose Detection Kit

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser

Transcription:

Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane, 1 Toshiaki Fukatsu, 1 Nobuo Nakashima, 1 Katsumi Tsuji, 2 Yoshihiro Soya, 2 Keizo Yoneda, 2 Shigeki Asano, 2 and Yoshihisa Kawamura 2 We describe a new kinetic assay for determining urea in serum or urine with use of urease (EC 3.5.1.5) and leucine dehydrogenase (EC 1.4.1.9). The latter enzyme is suitable for the kinetic assay of NH 4 because its K m value for NH 4 at ph 8.75 is large (500 mmol/l). Interference from endogenous NH 4 in serum or urine is obviated by subtraction of the assayed endogenous NH 4 value in a sample blank. For serum, within-assay CVs (n 10) were 0.39 0.58%; day-to-day CVs (n 10) were 1.56 2.30%. In urine, within-assay CVs (n 10) were 0.86 1.15%. Analytical recovery of urea (0.893 71.4 mmol/l) added to patients sera (urea 6.14 mmol/l) was 99.2 105.2%. The calibration curve for serum was linear through zero for urea concentrations up to 142.9 mmol/l and for urine up to 714.3 mmol/l. No influences of added ammonium ion, bilirubin, hemoglobin, ascorbic acid, or Intralipid were observed. The regression equations for this method (y) and conventional methods (x Determiner-LUN for serum assays, Serotec UUR-R for urine) were: y 1.016x 0.12 mmol/l (r 0.999, S yx 0.34 mmol/l, n 100) for sera, and y 1.070x 12.6 mmol/l (r 0.998, S yx 7.41 mmol/l, n 100) for urine. Urea nitrogen (UN) 3 in serum or urine has been measured by many methods [1, 2] that are based on enzymatic reaction with urease (EC 3.5.1.5) and glutamate dehydrogenase (EC 1.4.1.4). Endogenous NH 4 in serum and urine is a potential interference, because NH 4 produced from urea by urease is determined in the reaction system. The NH 4 in serum and urine has been eliminated with the glutamate dehydrogenase and 2-oxoglutaric acid before addition of urease. But in some urine that has a large amount of endogenous NH 4, the elimination of the NH 4 may not be complete, even in samples diluted 10 20-fold. Other approaches to avoiding interference from NH 4 have been described. To eliminate much of the NH 4 in urine, a recycling system (NADH7NAD ) with isocitrate dehydrogenase (EC 1.1.1.42) has been used [3]. Alternatively, ADP (produced from ATP by urease [4]) can be determined instead of NH 4. However, the cost of urea determination in these methods is very high; moreover, the former method was linear only to urea concentrations of 10 g/l, and the latter method was carried out at 376 nm, a wavelength unavailable on many automated analyzers. We investigated a new enzymatic kinetic assay of serum or urine UN by using leucine dehydrogenase (LED; EC 1.4.1.9) and urease [5, 6]. Interference from endogenous NH 4 in serum and urine is avoided by subtracting the endogenous NH 4 value assayed in a sample blank. Assay Principle Endogenous NH 4 in serum or urine is allowed to react with 2-ketoisohexanoic acid, NADH, and LED. The reaction rate at which NADH is oxidized to NAD depends on the amount of the endogenous NH 4 measured (I). (I) Sample R1 NH 4 2-ketoisohexanoic acid 1 Department of Clinical Laboratory, Nagoya University Hospital, 65 Tsurumai-cho, Showa-ku, Nagoya, 466, Japan. 2 Tsuruga Institute of Biotechnology, Toyobo Co., Ltd.,10 24 Toyo-cho, Tsuruga, 914, Japan. *Author for correspondence. Fax 81-52-744-2611. 3 Nonstandard abbreviations: UN, urea nitrogen; LED, leucine dehydrogenase; and Bicine, N,N-bis(2-hydroxyethyl)glycine. Received March 7, 1997; revision accepted May 7, 1997. NADH LED NAD l-leucine H 2 O Next, urease is added to the reaction system, and the oxidation rate of NADH to NAD by both the NH 4 produced from urea and the endogenous NH 4 is mea- 1932

Clinical Chemistry 43, No. 10, 1997 1933 sured (II). UN in the sample is calculated from the differences of oxidation rate between I and II. (II) Sample R1 R2 Urease Urea H 2 O 2H NH 4 CO 2 NH 4 2-ketoisohexanoic acid NADH LED NAD l-leucine H 2 O Materials and Methods Apparatus. This proposed method and the conventional (comparison) methods were performed with the Hitachi Model 7150 automated analyzer. Reagents. Urease (111 ku/g) and LED (45.1 ku/g; K m(nh4 ) 500 mmol/l [5]) were purchased from Toyobo. Analytical-grade ammonium chloride, l-ascorbic acid, urea, and NADH were from Wako Pure Chemical Industries. 2-Ketoisohexanoic acid sodium salt was from Nacalai Tesque (Kyoto, Japan); N,N-bis(2-hydroxyethyl)glycine (Bicine) from Dojindo Labs (Kumamoto, Japan); bilirubin from Sigma Chemical Co., and Intralipid 10% from KabiVitrum AB. Reagent 1 (R1) for the new method contained 2-ketoisohexanoic acid 3.0 mmol/l, -NADH 0.3 mmol/l, and LED 1.5 ku/l in 100 mmol/l Bicine buffer (ph 8.75). Reagent 2 (R2) contained urease, 70 ku/l, in R1. Calibrators were physiological saline, with and without 17.86 mmol/l UN. Assay procedure. UN in serum or urine was measured with the Rate-Analysis system B mode of the Hitachi 7150 automated analyzer, as shown Fig. 1. Serum (15 L) or urine (3 L) and R1 (300 L) were mixed, and the reaction rate was measured at measurement points 6 24, from 1 min to 5 min after the mixing (test I). Next, R2 (100 L) was added and the reaction rate was measured at points 30 50 (from 6 to 10 min) at 340 nm (test II). The lag time, to allow the urease reaction to proceed sufficiently, was set at 1 min (measurement points 25 30). A linear calibration curve based on the absorbance of the 0 (physiological saline) and 17.86 mmol/l UN calibrators was used to estimate the UN concentrations of the samples. Comparison method. The enzymatic method eliminated endogenous NH 4 in serum. The Determiner-LUN reagent kit from Kyowa Medex (Tokyo, Japan) was used for serum UN determination; another enzymatic method eliminated endogenous NH 4 in urine, the Serotec (Sapporo, Japan) UUR-R reagent kit for urine UN determination. Samples. L-Consera N (Nissui Pharmaceutical Co., Tokyo, Japan) was used as control serum. Patients serum and urine samples were from patients in Nagoya University hospital. Results optimization studies Optimization studies of this proposed method were carried out with the calibrator (UN 17.86 mmol/l), a patient s serum (UN 18.29 mmol/l), and L-consera N control serum (UN 5.82 mmol/l). Effects of ph. The effect of ph on the serum UN determination was examined in 100 mmol/l Bicine buffer at various ph values (8.00, 8.25, 8.50, 8.75, 9.00, and 9.25). With increasing ph values, the reaction rate (A/min) for assay of the calibrator increased, but the UN values of the patient s serum and L-consera N remained virtually unchanged (Fig. 2a). We chose to use Bicine buffer, ph 8.75. Effects of 2-ketoisohexanoic acid. We examined the effects of concentrations of 2-ketoisohexanoic acid on the serum UN determination. With increasing concentrations of 2-ketoisohexanoic acid (0.3, 0.5, 1.0, 3.0, 5.0, and 10.0 mmol/l), the reaction rate (A/min) of the calibrator increased, reaching almost the maximum at 5.0 10.0 mmol/l 2-ketoisohexanoic acid, whereas the UN values of the patient s serum and L-consera N were almost unchanged (Fig. 2b). We chose 10 mmol/l as the 2-ketoisohexanoic acid con- Fig. 1. Assay procedure.

1934 Morishita et al.: Urease leucine dehydrogenase assay of urea N in serum and urine Fig. 2. Effects of (a) ph, (b) 2-ketoisohexanoic acid, (c) LED activity, and (d) urease activity. The reaction rate (A/min; oxidation rate from NADH to NAD ) of UN calibrator (Œ, 17.86 mmol/l UN), patient s serum (E, 18.29 mmol/l UN), and L-consera N (F, 5.82 mmol/l UN) was measured under various conditions. centration in R1 and R2 to obtain the maximum reaction rate. Effects of LED activity. With increasing LED activity (0.1, 0.2, 0.5, 1.0, 2.0, and 4.0 ku/l), the rate of reaction (A/min) of the calibrator increased proportionally, whereas UN values of the patient s serum and L-consera N did not indicate much change (Fig. 2c). We added 1.5 ku/l LED to R1 and R2. Effect of urease activity. With increasing urease activity (5.0, 10.0, 20.0, 50.0, 70.0, and 100.0 ku/l), the reaction rate of the calibrator (A/min) increased, with the maximum occurring at 70 100 ku/l, whereas the UN values of the patient s serum and L-consera N did not show any marked change at urease of 50 ku/l or greater, as shown in Fig. 2d. We thus added 70 ku/l urease to R2. Time course. Typical time courses in this proposed kinetic assay of serum UN are shown for the calibrator, the patient s serum, and L-consera N in Fig. 3. The first reaction is the kinetic assay of the endogenous NH 4 Fig. 3. Time course of reaction. Calibrator (Œ, 17.86 mmol/l UN); patient s serum (E, 18.29 mmol/l UN); L-consera N (F, 5.82 mmol/l UN).

Clinical Chemistry 43, No. 10, 1997 1935 (points 6 24); the second reaction is the kinetic assay of both endogenous NH 4 and NH 4 produced from urea by urease (points 30 50). Within 1 min (from points 25 to 30), the urease reaction is ended. assay evaluation Precision. The serum UN assay samples were calibrator, L-consera N, and a patient s serum, which contained 17.91, 5.83, and 18.17 mmol/l UN, respectively. For the urine UN assay, we used a calibrator and two patients urines, with respective UN contents of 17.99, 93.68, and 575.49 mmol/l. Within-assay CVs were measured with n 10, and day-to-day CVs were determined from assays performed on 10 days (n 10). As shown in Table 1, serum UN within-assay CVs ranged from 0.39% to 0.58%; day-to-day CVs were 1.56 2.30%; in urine UN determinations, within-assay CVs (n 10) were 0.86 1.15%. Detection limit. We examined the detection limit of this UN determination by assaying physiological saline (NaCl 9 g/l, the 0 UN calibrator) 10 times. The result (mean SD) was 0.012 0.018 mmol/l. The detection limit, mean for physiological saline 3.0 SD, was 0.066 mmol/l. Table 1. Precision of the proposed UN assay. Calibrator L-Consera N Patient s serum Within-assay serum Mean, mmol/l 17.91 5.83 18.17 SD, mmol/l 0.095 0.034 0.071 CV, % 0.53 0.58 0.39 Day-to-day serum Mean, mmol/l 18.04 6.03 18.55 SD, mmol/l 0.282 0.139 0.406 CV, % 1.56 2.30 2.19 Calibrator Patient s urine Patient s urine Within-assay urine Mean, mmol/l 17.99 93.68 575.49 SD, mmol/l 0.206 1.044 4.929 CV, % 1.15 1.11 0.86 n 10 each. Analytical recovery. For UN additions of 0.89, 1.79, 2.68, 3.57, 7.14, 10.71, 17.86, 26.79, 35.71, 53.57, and 71.43 mmol/l to patients serum (UN 6.14 mmol/l), from 99.2% to 105.2% was recovered (mean 102.2%). Linearity of calibration curve. We examined the linearity of the calibration curve in duplicate with the UN calibrators. The calibration curve was straight up to at least 142.9 mmol/l for serum UN determinations and up to 714.3 mmol/l in urine UN determinations (Fig. 4). Interferences. Various substances were examined for their potential effects on this UN determination. One volume of examined substance was mixed with nine volumes of patient s serum (UN 16.29 mmol/l), L-consera N (UN 5.82 mmol/l), or a patient s urine (UN 324.1 mmol/l). Ammonium chloride was added to the patient s serum and L-consera N at concentrations of 0.056 0.56 mmol/l, and to the patient s urine at 55.6 166.7 mmol/l. Assay of UN was virtually unaffected (Fig. 5). Bilirubin, hemoglobin, ascorbic acid, and Intralipid were added to the patient s serum and L-consera N in the same way and assayed. We found no interference with serum UN determination from bilirubin up to 0.342 mmol/l, hemoglobin 0.059 mmol/l, ascorbic acid 1.14 mmol/l, and 50-folddiluted (0.2%) Intralipid. Correlation. Possible correlation between this proposed method (y) and the comparison methods (x) was examined (Fig. 6). The correlation between values obtained with our method and the Determiner LUN kit for 100 patients sera was: y 1.016x 0.12 mmol/l (r 0.999, S yx 0.34 mmol/l); comparison with the Serotec UUR-R kit for 100 patients urines gave: y 1.070x 12.6 mmol/l (r 0.998, S yx 7.41 mmol/l). Discussion After detailed optimization studies, we investigated the new enzymatic kinetic assay of serum or urine UN determination by LED and urease with the Hitachi 7150 Fig. 4. Calibration curve for serum UN (a) and urine UN (b) determinations.

1936 Morishita et al.: Urease leucine dehydrogenase assay of urea N in serum and urine Fig. 5. Interference of added NH 4 (as various concentrations of ammonium chloride) to samples of patient s serum (E, 16.29 mmol/l UN), L-consera N (F, 5.82 mmol/l UN), and patient s urine (, 324.1 mmol/l UN). The concentration of NH 4 shown indicates the concentration in the sample. Fig. 6. Correlation between this proposed method and conventional methods: comparison with (a) Determiner LUN kit for 100 patients sera, and (b) Serotec UUR-R kit for 100 patients urines. automated analyzer, which automatically calculated the difference between the endogenous NH 4 reaction (test I) and the reaction of both the endogenous NH 4 and the NH 4 produced from urea by urease (test II). Characteristic features of this method are: 1) Endogenous NH 4 in serum or urine does not affect the UN determination, so that urine, even when containing a high concentration of endogenous NH 4, need not be diluted before assaying. 2) This method yields linear results for large concentrations of UN up to 142.9 mmol/l in serum and up to 714.3 mmol/l in urine because the large K m value of LED for NH 4, 500 mmol/l, is suitable for the kinetic assay of NH 4. 3) No influences of added ammonium ion, bilirubin, hemoglobin, ascorbic acid, or Intralipid were observed. 4) Good precision, reasonable analytical recovery without interference, and good correlation with conventional methods were observed. The procedure takes 10 min and should be applicable to other discrete automated analyzers. The reagents could be prepared as liquid reagents. We therefore conclude this proposed method may be useful in routine clinical diagnosis. References 1. Hallett CJ, Cook JGH. Produced nicotinamide adenine dinucleotide-coupled reaction for emergency blood urea estimation. Clin Chim Acta 1971;35:33 7. 2. Sampson EJ, Baird MA. Chemical inhibition used in a kinetic urease glutamate dehydrogenase method for urea in serum. Clin Chem 1979;25:1721 9. 3. Marui Y, Nakano T, Hayashi T, Takagahara I. Method of terminating isocitrate dehydrogenase reaction in analytical system. US patent 4,742,001, 1988. 4. Fossati P. Enzymatic urea assay. US patent 4,608,335, 1986. 5. Tsuji K, Soya Y, Yoneda K, Asano S, Kawamura Y, Morishita Y, et al. A novel method for determination of urea nitrogen without interfering endogenous ammonia. Jpn J Clin Chem 1996;25(Suppl 3):29c. 6. Morishita Y, Fukatsu T, Nakane K, Tsuji K, Soga Y, Yoneta K, et al. Determination of urea-nitrogen in serum and urine by urease leucine dehydrogenase NADH system. Jpn J Clin Chem 1996; 25(Suppl 3):30c.