Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Similar documents
LYDIA NELSON, M.S., MT(ASCP), SH, SAMUEL CHARACHE, M.D., SANDRA WINGRELD, MT(ASCP), AND EDWARD KEYSER, MT(ASCP)

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

3. Blood Cell Histograms:

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Detection of important abnormalities of the differential

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

CELL-DYN Strength in Technology, Proven Reliability. Optical WBC Technology. Patented M.A.P.S.S. Differential. Multiple Technologies

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Hematology 101. Cindy Rogers, MT(ASCP) Diagnostics System Specialist

Interpreting Hematology Scatter-Plots; One Cancer Center s Keys to Seeing the BIG Picture

CELL-DYN 3700 Strength in Technology, Proven Reliability

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Performance evaluation of Celltac G: a new automated hematology analyzer

Collect and label sample according to standard protocols. Gently invert tube 8-10 times immediately after draw. DO NOT SHAKE. Do not centrifuge.

A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter

10/30/2015. XN Case Studies: Every Picture Tells a Story

XN series. Case interpretation. Gebruikersdag Vlaanderen- 6 oktober 2016

ADVANCED HAEMATOLOGY BATTLE OF THE BANDS. Dennis B. DeNicola, DVM, PhD, DACVP IDEXX Laboratories, Inc. Westbrook, Maine, USA BACKGROUND

Sysmex Journal International Vol.9 No.2 (1999)

ADx Bone Marrow Report. Patient Information Referring Physician Specimen Information

Evaluation of haematology analyzer CELL-DYN 3700 SL

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Detection and Classification of Acute Leukemia by the Coulter STKS Hematology Analyzer

XN-SERIES. XN Technology and Case Studies

Lavender Top Management SUCCESS BEYOND FINANCES

HISTOLOGY VIRTUAL LABORATORY BLOOD AND LYMPHATICS SYSTEM

Flow cytometry leukocyte differential : a critical appraisal

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens

Guide to the 1-3 Minute Blood Film Microscopic Review: Why and How?

HumaCount 5D. Outperforming 5-part Hematology System. CoreLab DX. Hematology

Hematopathology Case Study

Differential Blood Smear H3

Ordering Physician CLIENT,CLIENT. Collected REVISED REPORT

Interpreting the CBC. Robert Miller PA Assistant Professor of Clinical Pediatrics and Family Medicine USC Keck School of Medicine Retired

EDUCATIONAL COMMENTARY DIFFERENTIATING IMMATURE PERIPHERAL BLOOD CELLS

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

Proper Slide Preparation

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Pseudobasophilia in Pediatric Age Group

Complete Blood Count (CBC) Assist.Prof. Filiz BAKAR ATEŞ

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

Validation of Criteria for Smear Review Following Automated Blood Cell Analysis in Ain Shams University Laboratory

XN-SERIES. Hematology Case Studies: Every Picture Tells a Story

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential

NEW YORK STATE CYTOHEMATOLOGY PROFICIENCY TEST PROGRAM Glass Slide - November 2016

Impact of Integrating Rumke Statistics to Assist with Choosing Between Automated Hematology Analyzer Differentials vs Manual Differentials

Hematology & Coagulation Practicum Objectives CLS - 647

2008 CAP TODAY Q & A

DR SUDHIR MEHTA MD,MNAMS,FICP. Senior Professor & Head Medical Unit SMS Medical College & Hospital Jaipur

Analytic Inaccuracy Resulting from Hematology Specimen Characteristics

EDUCATIONAL COMMENTARY BLOOD CELL IDENTIFICATION

PointCare NOW TM Technical Note

Combining. and New Diagnostic. to Help Clinicians Achieve. Patient Outcomes at. per Healthcare Encounter

The White Blood Cell Differential. Evaluation of Rapid Impression Scanning Versus the Routine Manual Count

Performance Evaluation of the CellaVision DM96 System WBC Differentials by Automated Digital Image Analysis Supported by an Artificial Neural Network

2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228

Kathleen Finnegan MS MT(ASCP)SHCM

Clinical Utility. Clinical Utility of the Automated IG-Count Xavier Troussard

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

HEMATOLOGIC MORPHOLOGY- AECOM HEMATOLOGY COURSE

Collected: , PM Sent: , PM Received: , PM Preliminary: , PM. Notification Status: COMPREHENSIVE DIAGNOSIS

Differential Blood Smear H3

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana

INFECTION/ INFLAMMATION

3/31/2017 OBJECTIVES CASE STUDY #1 MANUAL REVIEW. Hematology Case Studies: Every Picture Tells a Story

The fluid medium (blood) is a highly specialized connective tissue that consists of various blood cells (formed elements) suspended in a fluid matrix

Clinical use of flow cytometry

Customer Information Literature List White Blood Cells

Beyond the CBC Report: Extended Laboratory Testing in the Evaluation for Hematologic Neoplasia Disclosure

Lymphoma Tumor Board Quiz! Laboratory Hematology: Basic Cell Morphology

Automated Hematology Analyzer. Differentiate with the XT-4000i

Standard Operating Procedure

Hematopathology Case Study

Validation of the Efficacy of a Practical Method for Neutrophils Isolation from Peripheral Blood

Myeloproliferative Disorders - D Savage - 9 Jan 2002

LEANING Away Waste and Improving Patient Care with Advances in Hematology

Blood Cell Identification Graded

Bone Marrow Pathology. Part 1. R.S. Riley, M.D., Ph.D.

Blood Cell Identification Graded

EDUCATIONAL COMMENTARY DISTINGUISHING MORPHOLOGIC LOOK-ALIKES

VETERINARY HEMATOLOGY ATLAS OF COMMON DOMESTIC AND NON-DOMESTIC SPECIES COPYRIGHTED MATERIAL SECOND EDITION

Blood DLC, Retic count, PCV, Hb and ESR. Dr. Tamara Alqudah

We noted a problem with the use of the hematology

Cbc with differential

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

The white blood cell differential (WBC-diff) technologies

Morfologia normale e patologica

Biology 218 Human Anatomy. Adapted form Martini Human Anatomy 7th ed. Chapter 20 The Cardiovascular System: Blood

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation

Notes for the 2 nd histology lab

Contents. Section Editor David Blomberg, MD

g/dl g/dl

MECHANISMS OF HUMAN DISEASE: LABORATORY SESSIONS LYMPHOMA. April 16, 2008

Continuing Education Questions

Transcription:

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities JOANNE CORNBLEET, M.D. AND SALLY KESSINGER, MT(ASCP)SH The authors evaluated the Coulter 15 S-Plus three-part differential and compared it with the total of component cells obtained by manual slide differential. The patient population (N = 989) was strikingly abnormal, in that 109 samples contained nucleated red blood cells, 118 samples contained one or more abnormal cells (blasts, promyelocytes, or lymphoma cells), and 95 samples had >10% monocytes. The GRAN and LYMPH fractions were accurately measured in the specimens that did not display region (R) flags. The MONO fraction correlated poorly with the manual differential and underestimated monocytes when they were increased. Not all samples with small numbers of abnormal cells or nucleated red blood cells were flagged by the three-part differential. Specimens with eosinophilia and reactive lymphocytosis likewise did not always show R-flags. The following findings (prevalence = 23.9%) were arbitrarily defined to constitute an "abnormal" manual differential: nrbc > 0%, blasts > 0%, promyel > 0%, lymphoma cells > 0%, baso > 5%, myel > 5%, meta > 5%, eos > 10%, reactive lymphs > 10%. Sensitivity of R- flagging for these abnormalities was 81.7% and specificity 73.3%. The predictive value of an R-flagged ("positive") result was 48.3% and of an unflagged ("negative") result, 93.0%. Efficiency (accuracy) was calculated to be 75.5%. (Key words: Leukocyte sizing differential; Automated differential; Three-part differential) Am J Clin Pathol 1985; 84: 620-626 AUTOMATION of leukocyte differentials has been achieved by flow cytometry and pattern-recognition methods. Both technics require expensive instruments and often cannot identify abnormal cells. Thus, these instruments provide screening differentials, flagging a certain proportion of suspect samples for further review. Recently, a three-part screening differential based on leukocyte volume analysis has been added to Coulter S- Plus series instruments. Hemolysis of red blood cells (RBCs) is accomplished with a weaker lysing reagent; this reagent leaves the white blood cells (WBCs) intact but shrinks the WBC cytoplasm. Lymphocytes are reduced to one-half their original volume, while granulocytes, because of their cytoplasmic granules, shrink to two-thirds original size. 2 ' 3 Monocytes and other mononuclear cells are reduced to a volume in between that of lymphocytes and granulocytes. Received November 16, 1984; received revised manuscript and accepted for publication April 29, 1985. Address reprint requests to Dr. Cornbleet: Department of Pathology, Stanford University Medical Center, Stanford, California 94305. Core Clinical Laboratories and Department of Pathology, Stanford University Medical Center, Stanford, California The resulting WBC sizing histogram is seen in Figure 1. The instrument sizes approximately 20,000 WBCs and determines the percentage of cells contained in the lymphocyte (LYMPH), mononuclear (MONO), and granulocyte (GRAN) fractions. 2 ' 3 The LYMPH fraction appears in the region between approximately 35-90 fl and is claimed to contain mature lymhocytes, atypical lymphocytes, and lymphoma cells. 2 ' 3 The GRAN fraction, in the area between 160-450 fl, contains polymorphonuclears, bands, metamyelocytes, eosinophils, and basophils. 2 ' 3 The MONO fraction is said to contain monocytes, blasts, promyelocytes, and myelocytes. 23 The divisions between the fractions are individualized for each sample. However, if these valleys are not located within the expected region, an R-flag is displayed on the data terminal next to the result for the affected fraction. These R-flags indicate an abnormal WBC volume distribution, and, because the three-part differential may not be valid, these flagged results are not printed on the report form. When the WBC distribution is severely distorted, or when the WBC count is less than 1.0 or greater than 0 X 10 3 /ML, the fraction percentages are not computed at all. Backlighting, which shows as a darkened color behind the parameter on the data terminal, also serves to identify potential abnormal specimens and erroneous results. Backlighting of the WBC count indicates potential interference at 35 fl, which can be caused by nucleated RBCs, clumped platelets, giant platelets, megakaryocyte fragments, cryoglobulin, fibrin strands, or unlysed RBCs. The MONO fraction also shows backlighting if its absolute value is greater than 1.5 X 10 3 //tl as a warning that a significant number of mononuclears may be present. Ideal performance of the three-part differential would include accurate categorization and estimation of normal cells while R-flagging the samples with abnormal cells. To assess how closely these goals are met, we evaluated the performance of the Coulter S-Plus IV with three-part differential as compared with the conventional manual 620

Vol. 84 No. 5 EVALUATION OF LEUKOCYTE SIZING DIFFERENTIAL 621 REL. NO. LVMPH Table 1. Comparison of White Counts on Samples with Backlighting N IV-D Mean WBC IV ZBI &t 35 R1, 22 ijzzv ffizk MONONUCLEAR 90 R2 160 R3 GRANULOCYTE 450 FL R4 All samples nrbc > 5 nrbc <. 5 107 20 87 P < 0.05 for IV-D minus IV equals 0. t P > 0.05 for IV-D minus ZBI equals 0. t P < 0.05 for IV minus ZBI equals 0. 14.1* 18.9*,f 13.0* 13.3 17.41: 12.4 18.1 LYMPH MONO LYMPH MONOCYTE REACT LYM BLAST LYMPHOMA PROMYEL MYEL GRAN POLY BAND META EOS BASO FIG. 1. Illustration of three-part differential histogram and component cells of the fractions. Volumes along x-axis are approximate points or "regions" where valleys are expected. When valleys are not found, an R-flag is placed next to the affected fraction(s). slide differential in a population with a high prevalence of abnormalities. Materials and Methods The Coulter S-Plus IV with three-part differential (IV-D) was installed and operated according to the manufacturer's instructions. The instrument was calibrated to reference manual methods for WBC and hemoglobin (Hgb), to historic moving averages for other RBC parameters, and to the Clay-Adams Ultraflo for platelets. Calibration of the leukocyte sizing differential was performed by the manufacturer. Since no assayed commercial control existed for this instrument at the time of evaluation, the IV-D was compared daily with a Coulter S-Plus IV in operation in the laboratory. This S-Plus IV was calibrated in a similar manner and controlled by using moving averages, 4C+, Technicon TQC platelet-rich plasma (self-assayed), and by cross-referencing to manual WBC and Hgb methods. Samples were collected in K 3 EDTA Vacutainers (Becton Dickinson) or K2EDTA Microvettes (Sarstadt) and processed between one to five hours of collection. Results from the IV-D three-part differential were compared with 100- or 200-cell manual slide differentials, with manual fractions consisting of the cells stated by the manufacturer (Fig. I). 2 When the manual and instrument differentials were discrepant, using the 2 SD limits Rumke, 7 both were repeated. Repeat manual differentials were performed on 200 WBCs, and these new results were used for data analysis. When the WBC count was less than 1.5 X [0 3 /fil, a buffy coat slide was prepared for manual differential. Sorting and tabulating data, as well as regression analysis and scatter plots, were performed using Minitab, a computerized statistical spreadsheet. 9 Regression was performed using the IV-D results as the independent variable (because of increased precision), then solving the resulting equation for the IV-D as the dependent variable. Statistical significance was assessed using a two-sided paired /-test or by testing for differences between two proportions. 10 Graphic illustrations were constructed on a Macintosh computer. Results Precision Measurement precision estimated on 14 samples by the method of duplicates showed coefficients of variation (CVs) of 1.1% for the GRAN fraction (mean 84.2%), 3.6% for the LYMPH fraction (mean 11.0%), and 12.3% for the MONO fraction (mean 4.8%). Since the IV-D uses a different lyse reagent than previous S-Plus instruments, assessment of WBC and Hgb precision is also important. Using three levels of 4C+ control over 20-30 days, the TOTAL (989) MICROTAINERS ICU HEME GNC XRT REMAINDER RANDOM DAYS (663) ^U=MMI*MlM=l=!=MIM=lllllilN?=l=lMlM= :i;;i;! ' i::i::: f MMi T :ip3 3* * mm^ * 1"!!» t i i i i i 20 30 40 SO 60 70 80 % wilk R-FLA6GIN6 FIG. 2. Percent of samples that show one or more R-flags, according to various patient categories.

622 CORNBLEET AND KESSINGER A.J.C.P. November 1985 80. % 60. '22*2 *2 * * ** + 80.+ <V 60.+ * * * * * 64fci 2*676972 29487*473 2522t/<47* * 265tt><8534* *3585m8774 * * **2*4+8/t8222 2282576*354* * * 2582/5763** 22475J3 ** 2 * 22***6* * 22ir* * 40.+ 20.+ * 24>T**42 343 22* *2**>38543* * *387>7584*2 * 44 + 8 44tf^676* 4++9<+953 * 4+J/++34 u- 40.+ 20.+ ** 2 > V *** + Zs ** rz /. r * 742 + 20. 40. 60. 80. % LYMPH FRACTION. MANUAL DIFFERENTIAL 0. 20. 40. 60. 80. % GRAN FRACTION. MANUAL DIFFERENTIAL 80.+ 60 40.+ 20.+ 0. 20. 40. 60. 80. % GRAN FRACTION. MANUAL DIFFERENTIAL 0. 20. 40. 60. 80. % GRAN FRACTION. MANUAL DIFFERENTIAL FIG. 3. Scattergrams of three-part differential results (Y-axis) versus manual differential (x-axis). Starred points are individual values; numerals correspond to multiple results plotted at the same point; and pluses represent greater than ten identical results. A. IV-D LYMPH fraction versus sum of component cells in manual differential in unflagged group of samples. B. IV-D GRAN fraction versus sum of component cells in manual differential in unflagged group of samples. C. IV-D GRAN fraction versus sum of component cells in manual differential in samples when GRAN fraction is R-flagged but computed. D. IV-D GRAN fraction versus sum of component cells in manual differential when GRAN fraction is unflagged but LYMPH and/or MONO fraction is R-flagged. E. IV-D MONO fraction versus sum of component cells in manual differential in unflagged group of samples. F. IV-D MONO fraction versus percent monocytes in manual differential in unflagged group of samples. CV for WBC ranged from 1.0 to 2.5% and Hgb from 0.5 to 1.0%. Comparison of Hemogram Parameters Comparison of WBC and Hgb, as well as RBC parameters and platelets, showed excellent agreement between the S-Plus IV-D and the S-Plus IV. Regression analysis on 535 paired samples gave slopes ranging from 0.98 to 1.01, intercepts less than 2% of the mean of the data set, and correlation coefficients all greater than 0.99. However, samples with backlighting of the WBC count due to interference at the lower counting threshold gave higher WBC values on the IV-D (Table 1). This positive bias

Vol. 84. No. 5 EVALUATION OF LEUKOCYTE SIZING DIFFERENTIAL 623 40.+ 16.0+ a. 12.0+ it 8.0+ 4.0+.0+ 0. 10. 20. 30. 40. % MONONUCLEAR FRACTION. MANUAL DIFFERENTIAL existed whether or not significant numbers of nrbc were present. The samples without nrbc that showed a greater uplifting of the S-Plus IV-D histogram at the lower WBC threshold were most often from patients with giant platelets (myeloproliferative syndrome) or from patients with RBCs more resistant to lysis (sickle cell disease, severe liver disease with acanthocytes, and neonates). Thus the WBC count on the S-Plus IV-D may be more susceptible to interference by giant platelets and unlysed RBCs because of the weaker lysing reagent and a decreased lower counting threshold. In 20 cases with NRBC s 5/100 WBC, the WBC count was compared with a ZBI with a lower threshold setting of 20 fl. As expected, the IV (with lower threshold of 45 fl) had a lower total WBC count for these samples, as all of the smaller nucleated RBCs were not included. However, the IV-D (with lower threshold of 35 fl) actually had a higher total WBC count than the ZBI. Review of individual patient cases revealed that many of the samples FIG. 3. (Continued) 10.0 12.0 14.0 16.0 % MONOCYTES. MANUAL DIFFERENTIAL 18.0 20.0 with nrbc also contained giant platelets or RBCs more resistant to lysis. Thus, it was impossible to assess whether or not the IV-D included all of the nrbc in the WBC count. Flagging Rate A total of 989 samples analyzed on the IV-D were compared with the manual differential. These consisted of 13% from hematology clinic, 6% from oncology clinic, 8% from radiation therapy clinic, 8% from intensive care units, 11% in Microvettes (mostly neonates), and 54% random patients. The all-overflaggingrate (samples with any R-flag) in the study was 39%, resulting in 387flaggedand 602 unflagged specimens. Regionflaggingby patient category is shown in Figure 2. The Microvettes samples (mostly neonates) have a very highflaggingrate, which, on inspection of the histograms and review of cell morphology on the LYMPH MONO GRAN GRAN (R-FLAG) GRAN* (OTHER R-FLAG) Table 2. Comparison of Percentages in the IV-D Differential versus the Sum of the Constituent Cells in the Manual Differential N Y* X BIAS by.x 3y-x r s*, 602 602 602 219 42 21.3f 5.4 73.2 55.8 58.4 21.0 7.4 71.6 47.9 55.2 0.3-2.0 1.6 7.9 3.2 0.96 1.16 0.97 0.98 0.95 l.i -3.1 3.9 8.9 5.8 0.95 0.53 0.94 0.82 0.96 4.5 4.0 5.6 16.0 8.4 d> 10% 7(1.2%) 14 (2.3%) 30 (5.0%) 108 (49.3%) 6(14.3%) Y - mean of IV-D fraction: X = mean of constituent cells in manual differential; Bias = Y X; Regression: Y = by., [X] + a,.,; r = correlation coefficient; S*, = Standard error of regression; d individual Y X differences. t All values expressed as percent of total white blood cells. % GRAN fraction not flagged, but LYMPH and/or MONO fraction

624 CORNBLEET AND K.ESSINGER A.J.C.P. November 1985 MONOS > 10* ^ ^ ^ ^ m 0 LYMPH MCNO GRAN 0 2 A 6 g 10 12 M 16 IB NUMBER SPECIMENS WITH DIFFERENCES > I OX FIG. 4. Causes of individual differences of more than 10% between IV-D and manual differential in unflagged sample group. smear, resulted from larger lymphocytes entering the R 2 region. Other sample categories had a similar flagging rate to the random patients. To ensure that ourflaggingrate was not biased through purposeful selection of abnormal samples, all samples received in the laboratory on three afternoons were run on the IV-D. The 683 random samples processed on these three days showed a lowered rate ofr-flagging(23%). MONO fraction backlighting (MONO > 1.5 X 10 3 /VL) can also serve to flag the presence of abnormal cells. However, only ten samples were found in our study to have MONO fraction backlighting without the presence of an R-flag. Thus, this indicator was not considered further in this study. Quantitative Accuracy Regression analysis of the IV-D fraction percentages versus the manual differential results is shown in Table 2 and scatter plots in Figure 3. For the 602 unflagged samples (unflagged in any R-region), the LYMPH and GRAN percentages agreed well with the corresponding cell totals in the manual differentials. Regression coefficients and scatter plots (Figs. 3A and B) were as expected for a line of identity; bias was small, and only a few sample pairs showed absolute differences of more than 10%. Regression analysis also was performed on 219 samples with R-flags on the GRAN fraction. These flagged samples had a lower mean total of manual granulocytes (47.9%) than the unflagged group (71.6%). A higher positive bias was reflected by the higher least-squares intercept. Of note was the increase in scatter of the results (Fig. 3C), reflected by the increased standard error of regression and lower correlation coefficient. Nearly one-half of the samples showed absolute differences of more than 10% from the manual differential. Thus, the % GRAN is not accurate when the result is R-flagged. Forty-two cases occurred where the LYMPH and/or MONO fraction wasflagged,but the GRAN fraction unflagged. Results for this group (Table 2 and Fig. 3D) were comparable to that for the completely unflagged samples (Fig. 3B). Results for the MONO fraction were less satisfactory. Correlation coefficient was 0.53. Considering that the mean value for the manually calculated MONO fraction was 7.4%, the bias of-2.0% was high, as was the standard error of regression of 4.0%. The scatter plot (Fig. 3E) also showed increased variability and suggested that IV-D often may give lower values than the manual differential. Ninety-five unflagged samples with high monocytes (> 10%) were included in our study. These samples had a mean of 12.4% monocytes, but the mean of the IV-D fraction was only 7.4% (P < 0.001). In Figure 5F, the percent in the IV-D MONO fraction is plotted versus the percent monocytes. The correlation coefficient is 0.15, and no linear relationship is evident. Thus, the MONO fraction does not reflect the percent of monocytes present in the sample and should be thought of as a remainder fraction rather than a measure of mononuclear cells. Of the unflagged 602 samples, only 9 samples for the LYMPH fraction, 11 samples for the MONO fraction, and 26 samples for the GRAN fraction had differences of more than 10% from the expected results. Causes for these discrepancies are shown in Figure 4. Unflagged lymphoid blasts that the IV-D sized in the LYMPH fraction but were tallied in the manual differential as part of the MONO fraction, and myelocytes that sized in the GRAN fraction, but were tallied as belonging to the MONO fraction, accounted for eight discrepant cases. Twelve more discrepancies resulted from high percentages of monocytes, a portion of which were included by the IV-D in the GRAN fraction. But the largest number (28) of discrepancies occurred when the WBC count was low. In these cases, the IV-D counted a higher number in the GRAN fraction and less in the MONO and LYMPH fraction than the manual differential. We speculate that the manual differential, rather than the IV-D, may be erroneous when the WBC count is low because of nonrandom distribution on the wedge smear. A study comparing the IV-D differentials with flow cytometry, which also classifies large numbers of cells, could resolve this issue. Causes of R-Flagging Comparison of the proportion of flagged and unflagged samples with various abnormalities is seen in Figure 5. Significantly higher (P < 0.001) proportions of the following abnormalities were seen in the flagged group: nrbcs > 0%, blasts > 0%, promyelocytes > 0%, basophils > 5%, myelocytes > 5%, metamyelocytes > 5%, monocytes > 10%. However, not all instances of these abnormalities wereflagged.the proportion of the following ab-

Vol. 84 No. 5 EVALUATION OF LEUKOCYTE SIZING DIFFERENTIAL 625 wee > on so -r BUST > cm PROWL KM irrphcrw >M MSO.SH mtl >5X rtta > 5» RAWS > I OS EC6 > 101 SEACLVT1 > I0» nc«o> ion FLAGGED (N-387) E3 UNFLAGGED(N-602).V.VAVA'AV.V.VA'.VAVA'AVAVWAVAV. it b 30 30 «tsrmpusuhti FINDING FlG. 5. Proportion of samples with various abnormalities in R-flagged and unflagged sample groups. Starred categories indicate a statistically significant difference (P < 0.05). 40-- 0 FLAGGED 30-- UNFLAGGED 20-- ^ 0 ^ EL JBL N - 118 LU M 2 3 4 5 6-10 11-20 >20 % ABNORMAL CELLS IN MANUAL DIFFERENTIAL FIG. 7. Flagging of specimens with various numbers of abnormal cells. normalities was not higher (P > 0.05) in the R-flagged samples: lymphoma cells > 0%, bands > 10%, eosinophils > 10%, reactive lymphocytes > 10%. Flagging of Increased Eosinophils and Reactive Lymphs The rate of R-flagging of eosinophils and reactive lymphocytes is dependent on the percent present, as shown in Figure 6. Although our total number of cases with > 10% eosinophils (30) or reactive lymphocytes (14) is small, it is evident that many cases of eosinophilia and reactive lymphocytosis will not be detected. Flagging of Abnormal Cells R-flagging of various percentages of abnormal cells (blasts, promyelocytes, or lymphoma cells) is shown in Figure 7. A high proportion (13/41 or 32%) of samples with 1-5% abnormal cells were not flagged by the IV-D. In addition, 6/77 or 7% of samples with larger numbers of abnormal cells (small-sized lymphoblasts or lymphoma cells) were not flagged. Flagging ofnrbc Samples with nrbc frequently cause WBC backlighting because of interference at the lower counting threshold for WBC. Also, it is possible that the presence of nrbc would be linked to other WBC abnormalities that would result in R-flags. As seen in Figure 8, combination of R- flagging with WBC backlighting added to the ability of the IV-D to detect nrbcs. However, for the 53 cases of one to two nrbcs, 19 (36%) samples still were not detected. R-Flag Sensitivity and Specificity The following findings were considered to constitute an "abnormal" differential, that is, one that should be flagged by the three-part differential screen for manual review: NRBC > 0%; blasts > 0%; promyelocytes > 0%; Oil EOS-RA00ED EOS-UNFLA86ED B REACLVM-RAMED REACLVM-UNFLA6CED R-FLAGGED EH WBC BACKLIGHTED N-IM is.. FIG. m IO-14* is-ten ixt X CELLS IN MftNURL IFFEHENTIIL 6. Flagging of specimens with various degrees of eosinophilia and reactive lymphocytosis. 4 5 8-10 11-20 >20 nrbc/1oowbc FlG. 8. Ragging of specimens with various numbers of nucleated red blood cells.

626 CORNBLEET AND KESSINGER A.J.C.P. November 1985 "Abnormal" diff* Table 3. Predictive Value of R-Flagging in Three-Part Differential Flagged (positive) Sensitivity (True pos) 187(81.7%) Unflagged (negative) (False neg) 42(18.3%) Specificity (False pos) (true neg) Normal" diff 200(26.3%) 560(73.7%) 760 Total 387 602 989 Total Prevalence 229 (23.9%) Predictive value or positive: True pos/total flagged = 48.3%; Predictive value of negative: True neg/total unflagged =- 93.0%; Efficiency (Accuracy): (True pos + True neg)/total = 75.5%. "Abnormal" differentials are defined in text. lymphoma cells > 0%; basophils > 5%; myelocytes > 5%; metamyelocytes > 5%; eosinophils > 10%; Reactivelymphs > 10%. Sensitivity and specificity for detecting these abnormalities are shown in Table 3. For our prevalence rate of 23.9%, the probability of abnormality with a flagged result is 48.3%, and the probability of no abnormality with an unflagged answer is 93.0%. For a population with a lower prevalence of these abnormalities, the predictive value of aflagged (positive) result would be lower and that of an unflagged (negative) result would be higher. Discussion The differential leukocyte count is a labor-intensive and expensive laboratory test.' Although limited in effectiveness for screening or casefinding, 8 it is of use in detecting or following acute infectious processes, hematologic disorders, and allergies, and in monitoring therapies with hematologic side effects. 1 A variety of decisions can be made from a screening differential that accurately enumerates total neutrophils and lymphocytes. 5 Our evaluation shows that the Coulter S-Plus three-part differential is capable of accurate enumeration of these cells, when the sample is not R-flagged. However, in a predominantly inpatient population, such as represented in our study, a high proportion of samples may be R-flagged. In addition, a screening differential shouldflagsamples with abnormal cells or increased numbers of normal cells. Our study shows that, based on R-flag criteria alone, this is not always true for small numbers of blasts, lymphoma cells, or nucleated RBCs, or increased numbers of reactive lymphocytes or eosinophils. The presence of abnormal cells or increase in reactive lymphocytes may be linked to other criteria in the complete blood count (CBC) (e.g., leukocytosis, neutropenia, lymphocytosis, or low platelet count) that would identify the samples requiring manual review. However, such linkage is not likely for increased eosinophils, and a manual differential or total eosinophil count must be performed when eosinophilia is suspected. The Coulter leukocyte sizing differential will not separate bands from segmented neutrophils orflagan increase in bands. However, counting of bands in a 100-cell manual differential is limited by poor reproducibility. 4 In a recent study, cytochemical determination of total neutrophils was found to be a more sensitive test for the diagnosis of acute appendicitis than the manual determination of bands. 6 Similarly, the determination of total granulocytes on the IV-D, which examines large numbers of cells, may prove to be a more precise indicator of acute infectious or inflammatory conditions than the manual enumeration of bands. It may indeed be true that all "abnormals" as arbitrarily defined in our study could have been identified if we had imposed high and low value limitflagsfor all of the CBC parameters in addition to the R-flags. However, in an inpatient environment, as opposed to an outpatient screening situation, requiring the complete CBC to be within normal range would lead to an unacceptably high rate of specimens needing blood smear examination. In a population such as ours, with a high prevalence of abnormalities, the best use of the Coulter S-Plus three-part differential will be for patient follow-up rather than for screening. Once it has been established that the patient is free of abnormal cells, the three-part differential can replace the manual differential in monitoring the percentage or absolute number of granulocytes. Acknowledgments. The authors thank Coulter Electronics, Inc., for their support during this evaluation. They also thank all the technologists in the Stanford Hospital Hematology Laboratory for their help. References 1. Connelly DP, McClain MP, Crowson TW, Benson ES: The use of the differential leukocyte count for inpatient casefinding. Hum Pathol 1982; 13:294-300 2. Coulter Electronics: The subpopulation differential flags and alerts. Hematology Analyzer 1984; 5:1-2 3. Coulter Electronics: Product Reference Manual: S-Plus IV with 3- Part Differential and Auto Transfer. Hialeah, Coulter Electronics, 1983 4. Dutcher TF: Automated leukocyte differentials: A review and prospectus. Lab Med 1983; 14:483-488 5. Koepke J A: Standardization of the manual differential leukocyte count. Lab Med 1980; 11:371-375 6. Marchand A, Van Lente F, Galen RS: The assessment of laboratory tests in the diagnosis of acute appendicitis. Am J Clin Pathol 1983; 80:369-374 7. Rumke CL: The statistically expected variability in differential leukocyte counting, Differential leukocyte counting. Edited by Koepke J. Aspen, College of American Pathologists, 1977, pp 39-45 8. Rich EC, Crowson TW, Connelly DP: Effectiveness of differential leukocyte count in case finding in the ambulatory care setting. JAMA 1983; 249:633-636 9. Ryan TA, Joiner BL, Ryan BF: Minitab student handbook. Boston, Duxbury Press, 1976 10. Walpole RE, Myers RH: Probability and statistics for engineers and scientists. New York, McMillan, 1972