CT Dose Reduction in Pediatric Patients

Similar documents
Radiation Dose To Pediatric Patients in Computed Tomography in Sudan

Utilization of a Patient Dose Tracking and Monitoring System to Modify CT Protocols and Lower Patient Doses

Seattle Children s Hospital Radiology Department. Statement regarding radiation exposure related to computed. tomography (CT) exams

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Debra Pennington, MD Director of Imaging Dell Children s Medical Center

Ask EuroSafe Imaging Tips & Tricks. CT Working Group. On the use of Diagnostic Reference Levels in CT

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

CT Radiation Risks and Dose Reduction

Justification, Optimization and Communication in Pediatric CT Imaging: Recent Improvements and Persistent Challenges Designated Emphasis in Nuclear

IMAGE GENTLY HOW CAN YOU HELP?

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

How to Develop CT Protocols for Children

Managing Patient Dose in Computed Tomography (CT) INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION

Dose Metrics in the Patient Report: Myopic or Foresighted?

Why is CT Dose of Interest?

Fetal Dose Calculations and Impact on Patient Care

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

THE UTILIZATION OF A DOSE MANAGEMENT SOLUTION TO LOWER RADIATION DOSES IN MEDICAL IMAGING

CT Quality Control Manual FAQs

Session 83X Dose Management: Patient and Staff Radiation Safety in Radiology

Translating Protocols Across Patient Size: Babies to Bariatric

Dianna Cody, PhD, DABR, FAAPM Professor & Clinical Operations Director Imaging Physics U.T. M.D. Anderson Cancer Center Houston, TX

Managing Patient Dose in Computed Tomography (CT)

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

8/1/2017. Financial Disclosures. Dose Tracking at MGH. How Dose Tracking Affected Protocol Optimization in a Tertiary Quaternary Healthcare Center

Aquilion ONE: Pediatric Imaging. Richard Mather, PhD. Senior Manager, CT Clinical Science Toshiba America Medical Systems, Inc.

Dosimetry in Medical Exposures: Trends, Challenges and Next Steps


Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Accounting for Imaging Dose

Introduction and Background

Washington State Hospital Association Safe Table Webcast 100K Children Campaign Safe Imaging September 15, 2014

Multi-stakeholder meeting on the optimised use of CT scanners

CT Head Dose Reduction Using Spiral Scanning Protocol

8/18/2011. Acknowledgements. Managing Pediatric CT Patient Doses INTRODUCTION

USING TEAM AUDITS TO IMPROVE IMAGE QUALITY: Our Institution s Experience

Optimizing radiation dose by varying age at pediatric temporal bone CT

Low-Dose CT Cancer Screening Program

The Increasing Use of CT and Its Risks

A Closer Look at Radiology s Role in Corporate Healthcare. Mark D. Jensen, CEO, Charlotte Radiology

Radiation Dose in Pediatric Imaging

Managing Radiation Risk in Pediatric CT Imaging

Steven Aaron Ross, M.D. Pediatric Radiologist El Paso Imaging Consultants El Paso Children s Hospital

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

Estimation of the Risk of Cancer Associated with Pediatric Cranial Computed Tomography

7/31/2017. Image Gently is. The Pediatric Radiologist Tailoring CT Protocols to the Patient's Age, Size, and Clinical Scenario.

Guidance for the Communication of Clinical and Imaging Observations and Procedure Details by Radiologist Assistants to Supervising Radiologists

The AAO- HNS s position statement on Point- of- Care Imaging in Otolaryngology states that the AAO- HNS,

CT and Dose at Affidea: What every referring doctor should know

Prepublication Requirements

Over the past 18 months several articles have appeared in the New York Times and other

Estimating Risk of Low Radiation Doses (from AAPM 2013) María Marteinsdóttir Nordic Trauma,

Outline. NCRP Scientific Committee 6-2

Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.

Optimization of kvp and mas for Pediatric Low-Dose Simulated Abdominal CT: Is It Best to Base Parameter Selection on Object Circumference?

A multicentric study on patient dose in multislice CT

Radiation Dosimetry for CT Protocols

Sujit Iyer M.D., Patrick Boswell, Shaheen Hussaini MD, Julie Sanchez M.D, Tory Meyer M.D.

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Re: PSM of the National Voluntary Consensus Standard for Patient Safety Measures, 2 nd Report

Radiation exposure from a CT scan A guide for parents

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Ionizing Radiation Exposure from Radiologic Imaging: The Issue and What Can We Do?

Electronic collimation and radiation protection in paediatric digital radiography: revival of the silver lining

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

Karyn A. Ledbetter, MD; Andrew K. Moriarity, MD; Safwan Halabi, MD Henry Ford Hospital, 2799 W. Grand Blvd, Detroit, MI, 48202

Tracking Doses in the Pediatric Population

What Have We Learned from Children? Donald P. Frush, MD

Medical Physicists Bring New Value to Patient-Centered Care

AMERICAN COLLEGE OF SURGEONS CRITICAL CARE REVIEW COURSE 2012 HOT TOPICS IN PEDIATRIC CRITICAL CARE

Overview of Medical Diagnostic and Interventional Imaging: Usage Patterns

Survey of patients CT radiation dose in Jiangsu Province

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011

Commentary ISSUES TO BE CONSIDERED

Radiation Dose Specification

Radiation Dose in X-Ray and CT Exams

An Assessment of Organ and Effective Dose of Patients who Undertake CT Examinations in two Teaching Hospitals of Mashhad&Isfahan

Quality Control and Patient Dosimetry on line for Computed Tomography

Radiation Units and Dosimetry 15 August Kalpana M. Kanal, Ph.D., DABR 1

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

People Exposed to More Radiation from Medical Exams

Class Time: 9/12,10-10,11-21 Days: Listed Saturdays 1p-4p Room:HS208. Office Phone: Home Phone:

Undue Exposure to Ionizing Radiation

Regional diagnostic reference levels and collective effective doses from CT scanners in India

Estimated Radiation Dose Associated With Low-Dose Chest CT of Average-Size Participants in the National Lung Screening Trial

Department of Radiology

Background Radiation in U.S. ~ msv/yr msv/yr ~0.02 ~0.02 msv msv/day /day (~2 m rem/day) mrem/day) NCRP 4

Ionizing Radiation Exposure from Radiologic Imaging

ASRT Position Statements

Radiation related cancer risk & benefit/risk assessment for screening procedures

Catalog Addendum

Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray

Scientific Affairs. Scientific Affairs

An audit of radiation dose of 4D CT in a radiotherapy department

Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode

Radiation Safety in the Catheterization Lab

Lung Cancer Screening

Transcription:

CT Dose Reduction in Pediatric Patients By Kelly Firestine, RT(R)(CT)(M) Executive Summary CT is an incredibly valuable imaging tool, but there are unique concerns with pediatric patients, including the increase of sensitivity to radiation, increase risk of cancer formation with the longer life expectancy, and larger radiation doses received when adult CT settings are used. With a grant from the AHRA and Toshiba Putting Patients First program, Memorial Hermann Healthcare System was able to better educate technologists and physicians which helped in reducing radiation dose to pediatric patients by as much as 60%. Standardizing CT protocols and properly training staff are very important. But to make improvements on a larger scale, strategies for tracking compliance and providing feedback must be used. Many clinical decisions are based upon information derived from imaging procedures, so ensuring optimal image quality is crucial to the patient, clinician, and the efficiency of the healthcare system. The role of our computed tomography (CT) departments in providing safe, quality service to achieve this goal is of paramount importance. Last year, the Memorial Hermann Healthcare System in Houston, TX performed over 14,000 pediatric CT scans. The Toshiba and AHRA Putting Patients First grant helped us further our focus on pediatric CT quality and safety. Using the grant to educate our technologists and physicians has helped us succeed in reducing radiation dose to our pediatric patients by as much as 60%. Soon, medical imaging, with CT scans as the largest contributor, will approach or potentially exceed background radiation as the single largest source of radiation for humans. 1 Research clearly indicates children are more sensitive to radiation and have a lifetime to manifest those changes. CT is an extremely valuable, diagnostic instrument. But this tool is only valuable if the exams are performed properly by technologists. On the contrary, exams performed improperly could potentially harm patients. Also, if there are no well defined standards for optimal imaging techniques, clinically important information may be missed or patients may suffer long term effects. Standardizing CT protocols and properly training staff are very important. But to make improvements on a larger scale, strategies for tracking compliance and providing feedback must be used. Radiation doses in pediatric imaging have become a very important topic, especially when performing pediatric CT. Although CT is an incredibly valuable imaging tool, a few unique concerns with pediatric patients are present. These concerns include the increase of sensitivity to radiation, increase risk of cancer formation with the longer life expectancy, and larger radiation doses received when adult CT settings are used. The American College of Radiology has noted, Because they have more rapidly dividing cells than adults and have longer life expectancy, the odds that children will develop cancers from x-ray radiation may be significantly higher than adults. 2 It has been estimated by the National Research Council s Committee on the Biological Effects of Ionizing Radiation those children less than 10 years of age are several times more sensitive to radiation than middle-aged adults. 3 We recognize unnecessary radiation may be delivered when CT scanner parameters are not appropriately adjusted for patient size. When a CT scan is performed on a child with the same technique factors that are 40

used for a typically-sized adult, the child receives a significantly larger effective dose than the full-sized patient. 4 Every institution should evaluate their CT practices and try to emphasize the importance of keeping the radiation doses as low as reasonably achievable (ALARA) while still maintaining good diagnostic image quality. By standardizing pediatric exposure techniques, placing more emphasis on proper education and training of CT staff, as well as regular auditing of compliance we have shown good image quality can be obtained with a reasonable reduction in radiation dose. Baseline Data A case series study of CT scans on children and adolescents found in the Memorial Hermann archives was retrospectively performed. From four separate months, random samplings of the pediatric CTs performed at each facility were documented. From this documentation, the following data were compared: facility, patient age and weight, type of study, kvp, mas, exposure time, pitch, the CTDI vol (volume computed tomography dose index, the average dose delivered to the scan volume for a specific examination), and dose length product (DLP) when available. The results were then compared to the recommended doses from the American College of Radiology, the Society for Pediatric Radiology, and the Image Gently TM campaign (an inititaive by the Alliance for Radiation Safety in Pediatric Imaging). This evaluation of our current practices in regards to pediatric CT showed us that the overexposure of children or small adults can easily go unrecognized by radiologist and technologist because the image quality is not compromised or can even be improved. We recognized a need for continuous self monitoring and system wide standardization regarding pediatric patient techniques in CT. Improvement Process Our pediatric dose reduction project followed this systematic approach: 1. Analysis of current practices 2. Standardization of protocols 3. Implementation of policies and procedures 4. Education and training 5. Measuring performance 6. Reporting results 7. Continued quality improvement The first part of the improvement process was to develop standardized pediatric CT protocols. A set of pre-programmed weight and age protocols was developed for pediatric exams by using the recommendations from the Image Gently TM campaign as a starting point. 5 Under the supervision of pediatric radiologists throughout the system and a professor of pediatric radiology for the University of Texas, the scan techniques were reduced in an incremental fashion to avoid unacceptable deterioration in image quality. Gradually, the protocols were adjusted and the CT images were checked against the image criteria given in the European Commission for Quality Criteria for Computed Tomography and against the American College of Radiology s Image Quality Guidelines. 6-8 Current scan protocols are now felt to be approaching the optimum and any further attempts at dose reduction would necessitate a more in depth analysis of image quality. Examples of some CT protocols are listed in Table 1. j TABLE 1. Project budget for VWCs EHR implementation. (Note: This budget does not reflect an endorsement of any vendor or system, it is only one example of an EHR implementation in a multi-facility ambulatory environment.) radiology management march/april 2011 41

CT Dose Reduction in Pediatric Patients Box 1. Pediatric CT dose and image quality audit form. These standardized protocols were programmed in all CT scanners and password protected to prevent any unauthorized changes. Along with these protocols, new polices were implemented such as saving the patient dose profile page with the study and when not to proceed with a study due to patient motion. Education The next piece of the improvement process was education. Not only do technologists performing the studies need to be educated, but also the physicians ordering the exams. First, an all day seminar was held for technologists. Representatives from each of the 21 CT departments were invited to learn more about radiation dose. Technologists spent the day learning about x-ray production, biological effects of radiation, and how to use proper scan techniques to achieve optimal image quality. To provide additional incentive to attend this very important seminar, technologists were 42

rewarded with continuing education credits from the ASRT for their participation. The seminar was also professionally recorded and set up to view as web based training for those unable to attend. An educational handout was also created for referring physicians to inform them of goals and give them resources when referring a child to one of our imaging centers. Compliance The final part of the improvement process was the compliance audit. The pediatric CT dose audit is performed monthly on pediatric CT abdomens and head. The exams are scored, using a pediatric CT dose and image quality audit form (see Box 1). This form assigns a certain weight for each category. This audit not only monitors CT exams for reduced dose, but also for image quality. The goal is to achieve the lowest radiation dose while still maintaining good image quality. The dose threshold for the audit is set at 20 mgy for CT abdomen. This threshold is based on the American College of Radiology s Pass/Fail criteria (Table 2). 9 The dose threshold for CT head is 70 mgy. This number was derived from the ACR s adult head threshold since no pediatric head threshold has been set by the ACR yet. Anything that exceeds these numbers is considered over-dose by ACR. These numbers are the national benchmark the audit is based on. The results of this audit are presented monthly to the Memorial Hermann Imaging Council which is comprised of all the radiology directors and led by the CEO of Memorial Hermann s retail healthcare businesses, which includes 21 outpatient imaging centers, 8 breast care centers, 10 ambulatory surgery centers, sports medicine and rehabilitation centers, and 19 diagnostic labs. The audit brings attention to areas in which improvement is needed. The audit also lets the technologists know they will be held accountable for quality on the j TABLE 2. ACR CT Accreditation Dose Pass/Fail Criteria and Reference Levels Examination studies they perform. Awareness of this very important issue has been kept on the forefront by this audit and feedback process. The audit shows we are diligent in our continued success in lowering pediatric CT dose. In the last year, not only has each department exceeded the audit score goal of 98%, but we have actually achieved 100% scores system wide for seven out of the last 12 months. We will continue to monitor monthly to ensure compliance is ongoing. Documented Improvement Outcomes Pediatric head CT protocols are chosen by patient s age due to differences in skull formation. Figure 1 demonstrates the monthly average CTDI vol (CT dose index) of all pediatric head CTs sampled before and after the implementation of the pediatric CT protocols. This graph demonstrates the decline in the CTDI vol for pediatric heads. A decrease is shown in dose for every age group, well below the dose audit threshold of 70mGy. Over the last year, the low pediatric CT head doses have leveled off and have been Pass/Fail Criteria CTDI vol (mgy) maintained throughout the year for each age group. A reduction in dose is shown over 50% for patients 0 to 5 years old and over 40% for patients 6 to 15 years old. Pediatric abdomen CT protocols are chosen by weight. Figure 2 demonstrates the quarterly average CTDI vol of all pediatric abdomen CTs sampled. The current dose for patients 0 to 27 kg has been reduced 67% when compared to the doses delivered before the implementation of the standardized pediatric protocols. Also for patients 28 to 45 kg the dose has been reduced 60% and 45% for patients 46kg and greater. Conclusion Reference Levels CTDI vol (mgy) Adult Head 80 75 Adult Abdomen 30 Pediatric Abdomen (5 year old) 20 CT scans help save the lives of children, but it is recognized that children are more sensitive to radiation. As a result, imaging departments must take steps to adjust imaging procedures to meet the special needs of children. This project has proven that good quality pediatric CT examinations can be obtained at a reasonable dose of radiation exposure, thus reducing the risk to children. The AHRA and Toshiba Putting Patients First Grant Not only has each department exceeded the audit score goal of 98%, but we have actually achieved 100% scores system wide. radiology management march/april 2011 43

CT Dose Reduction in Pediatric Patients CTDI vol mgy Over 50% 75 Dose Reduction 70 65 60 55 50 45 40 35 30 20 15 10 5 0 56 29 Figure 1 Monthly average CTDI vol Collective Pediatric CT Head Doses FY 2009/2010 66 33 33 32 32 0 18m 19m 5y 6y 10y 11y 15y Pre Audit 1st Quarter FY 2010 2nd Quarter 3rd Quarter 4th Quarter enabled us to educate our technologists and physicians, significantly contributing to the awareness of this important issue. By performing internal quality audits, CTDI vol mgy 27 24 21 18 15 12 9 6 17 Collective Pediatric CT Abdomen Doses FY2009/2010 15.3 8.43 8.98 6.96 7 6.29 5.58 6.3 6.1 72 41 41 39 16.17 73 50 10.28 9.4 49 70 mgy Threshold creating system standardization, educating staff, and monitoring conformity, a cycle of continuous quality improvement is ensured. 9.3 42 8.9 References 1 National Council on Radiation Protection and Measurements. Ionizing Radiation Exposure of the Population of the United States. Report No. 160. Bethesda, MD: NCRP, 2009. 2 One Size Does Not Fit All: Reducing Risks from Pediatric CT. ACR Bulletin 2001;57(2):20 23. 3 National Research Council, Committee on the Biological Effects of Ionizing Radiations. Health Effects of Exposure to Low Levels of Ionizing Radiation (BEIR V). Washington, DC: National Academy Press; 1990. 4 Paterson A, Frush DP, Donnelly L. Helical CT of the body: are settings adjusted for pediatric patients? AJR Am J Roentgenol. 2001; 176(Feb):297 301. 5 Cincinnati Children s Hospital Medical Center. A Practice Quality Improvement (PQI) Program in Computed Tomography (CT) Scans in Children. 2009. Available at: http:// spr.affiniscape.com/associations/5364/ig/ index.cfm?page=518. Accessed January 21, 2011. 6 Bongartz G, Golding SJ, Jurik AG, Leonardi M, van Meerten EvP. European Commission s Radiation Protection, EUR 16262 EN. 7 American College of Radiology. ACR Practice Guideline for the Performance of Computed Tomography (CT) of the Abdomen and Computed Tomography (CT) of the Pelvis. Revised 2006 (Res. 13,17,35). 8 American College of Radiology. ACR Practice Guideline for the Performance of Computed Tomography (CT) of the Extracranial Head and Neck in Adults and Children. Revised 2006 (Res. 12,17,35). 9 Amis ES, Butler PF, Applegate KE, et al. American College of Radiology white paper on radiation dose in medicine. J Am Coll Radiol. 2007;4:272 284. Kelly Firestine, RT(R)(CT)(M) is the CT clinical quality manager for the Memorial Hermann Healthcare System in Houston, TX. She is responsible for the clinical quality of more than 40 CT scanners and is very focused on reducing radiation dose in CT. Kelly can be reached at kelly.firestine@memorialhermann.org. 3 0 0 27kg 28 45kg 46kg + Pre Audit 1st Quarter FY 2010 2nd Quarter 3rd Quarter 4th Quarter Figure 2