Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Similar documents
Dental Extraoral X-ray Systems

Dental Intraoral X-ray Systems

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

icat Precise Policy and Procedure Manual

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

CT Quality Control Manual FAQs

Agenda: Dental Cone Beam Imaging

X-RAY REGULATORY GUIDE

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

Code of Practice for Radiation Protection in Dentistry. Code of Practice For Radiation Protection in Dentistry

Mammography Quality Control: A Refresher

IONISING RADIATION REGULATIONS 99

The IAC Standards and Guidelines for Dental CT Accreditation

Radiation Protection in Dental Intraoral Radiography Edition

STANDARD OF PRACTICE. Dental CT Scanners CONTENTS. April 2011

Prepublication Requirements

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION SAFETY DIRECTORATE

NJ CBCT Regulation FAQ & Assistance

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

ALBERTA REGULATION 182/2003 RADIATION PROTECTION REGULATION

Computed tomography Acceptance testing and dose measurements

Radiation Safety Characteristics of the NOMAD Portable X-ray System

OKLAHOMA STATE UNIVERSITY X-RAY MACHINE PROGRAM AND PI HANDBOOK For easy navigation, please use Adobe s bookmark feature.

CHAPTER Section 3 of P.L.1983, c.296 (C.45: ) is amended to read as follows:

The Radiation Health and Safety Act, 1985

25 TEXAS ADMINSTRATIVE CODE (TAC) (x) Certification of Mammography Systems and Mammography Machines Used for Interventional Breast Radiography

STATE OF MINNESOTA BOARD OF DENTISTRY GENERAL STATEMENT

CT Scanning Protocol For V2R Guided Surgery Solutions

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

The New Jersey Radiographic Quality Assurance Program at 5 Years

Scientific Affairs. Scientific Affairs

Florida Medicaid. Dental Services Coverage Policy. Agency for Health Care Administration

Bureau of Laboratory Quality Standards

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

organizations including the International Commission on Radiation Protection exposure dose management system to all the studies and

AAPM Annual Meeting. ACR Accreditation Update in CT

Legislative Counsel s Digest:

MANUFACTURING OPERATING STANDARDS AND RESTRICTIONS FOR COMMERCIAL CANNABIS OPERATIONS IN CITY OF CARSON

Bone Densitometry Equipment Operator

Alfred Health's CT quality control program

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

To Shield or Not to Shield? Lincoln L. Berland, M.D.

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

ACR MRI Accreditation: Medical Physicist Role in the Application Process

Radiation Safety Manual

Annex A TITLE 49. PROFESSIONAL AND VOCATIONAL STANDARDS PART I. DEPARTMENT OF STATE. Subpart A. PROFESSIONAL AND OCCUPATIONAL AFFAIRS

Hearing Conservation Program Regulations and Recommendations Summary

Medical Events. Florida Department of Health Division of Emergency Preparedness and Community Support Bureau of Radiation Control

Quality Assurance and Radiation Protection Manual for Non-Human Use Radiation Generating Equipment

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Routine Quality Assurance Cookbook

Proposed Radiation Safety Regulations: Submission form

PERSONNEL MONITORING AND DOSIMETRY POLICIES

CHAPTER Committee Substitute for Committee Substitute for Senate Bill No. 2760

Minnesota. Prescribing and Dispensing Profile. Research current through November 2015.

-Ray Producing Equipment

Steven Tate Nicholas

The College of Dental Surgeons of Saskatchewan Radiation and Imaging Standard

Table of Contents. Introduction 3. Background 4

Varian Edge Experience. Jinkoo Kim, Ph.D Henry Ford Health System

Procedure Number 310 TVA Safety Procedure Page 1 of 6 Hearing Conservation Revision 0 January 6, 2003

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

Module Rhodes

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Implementation Guide for the DoseControl Dosimetry System

Imaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:

UPPER MISSOURI DISTRICT HEALTH UNIT REGULATIONS FOR TANNING FACILITIES

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Chapter 9 TOBACCO AND SYNTHETIC NICOTINE CONTROL

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Establishing the quality management baseline in the use of computed tomography machines in Kenya

Implementation of a New Tomosynthesis Program: A Physicists Perspective

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Douglas J. Simpkin, Ph.D. Aurora St. Luke s Medical Center Milwaukee, Wisconsin. www.

Estimating Iodine Concentration from CT Number Enhancement

April 21, Patricia H. Glasrud Executive Director PHG/pmn encl.

X-Ray & CT Physics / Clinical CT

Title 32: PROFESSIONS AND OCCUPATIONS

Why is CT Dose of Interest?

CAREER INFORMATION WHO IS THE REGISTERED DENTAL HYGIENIST?

Guide to CRD BYLAW NO A BYLAW REGULATING TANNING FACILITIES

Public Health Service Food and Drug Administration Dallas District 4040 North Central Expressway Dallas, Texas

Hearing Conservation Program

NOTICE OF INTENT. Department of Health and Hospitals Board of Dentistry

1 Course Syllabus + Study Guide for Lecture and Laboratory

Policy / Drug and Alcohol-Free Workshops

University of Cincinnati. Quality Assurance. and. Radiation Protection Manual. for. Non-Human Use. Radiation Generating. Equipment

ACR Accreditation Toolkit for Validation Site Surveys

Vision, Hearing and Spinal Screening Programs. Implementing Changes in the Vision, Hearing and Spinal Screening Programs for School Year

San Diego Community College OSHA Standards

SRT-100 Skin cancer TReaTmenT Simplified

Overview of TG225 - Medical Physics Practice Guideline #1

SUBJECT: Effective Date: Policy Number: Controlled Substance and Prescription Drugs 11/30/

ER75 Electro-Acoustic Ear Simulator. Operating Manual

Introduction and Background

Transcription:

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems PROPOSED REVISIONS TO 4732.XXXX, 1.0 4732.#### DENTAL CONEBEAM COMPUTED TOMOGRAPHY (CBCT) X-RAY SYSTEMS; STATIONARY AND MOBILE. Subpart 1. Applicability. A registrant s x-ray system used for dental CBCT imaging must meet the applicable performance standards in Code of Federal Regulations, title 21, section 1020.30 to 1020.40, or successor requirements. X-ray Equipment Subp. 2. X-ray beam alignment. A registrant is responsible for the x-ray beam alignment provisions of this subpart. A. Limit the x-ray field so that it does not exceed each dimension of the image receptor by more than two percent of source-to-image distance (SID), when the x-ray beam is perpendicular to the image receptor; and B. Align the center of the x-ray field with the center of the image receptor to within 2% of SID. Subp. 3. Mechanical support of the tube housing assembly. A registrant using a dental CBCT x-ray system is responsible for the requirements of this subpart. Page 1

A. The tube housing assembly must remain stable during the exposure unless tube housing movement is a designed function of the dental CBCT x-ray system; and B. All position locking, holding, and centering devices must function as intended. Subp. 4. Radiation exposure control. A registrant is responsible for the radiation exposure control provisions of this subpart. An x-ray control must: Commented [JC(1]: Based on Texas administrative code. 3. Adding by a qualified operator. Consistent with both SSRCR and Michigan administrative code. 4. Adding subitem 2 consistent with both SSRCR and Michigan and Texas administrative code. A. be incorporated into each x-ray system so that an exposure can be terminated by a qualified operator at any time during the scan or series of scans of greater than 0.5 second duration. B. after premature termination of x-ray exposure, require resetting of the Commented [JC(2]: SSRCR F. 11. A vii (4) and 21cfr1020.33(f)(2)(ii) CBCT condition of operation before initiating another scan. C. bear the warning statement under 21 CFR 1020 which is legible and accessible to view: "WARNING This x-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions, and maintenance schedules are observed., or successor requirements. Subp. 5. Beam-on indicators. A registrant is responsible for the beam-on requirements Commented [JC(3]: Consistent with SSRCR, Part F, p. 50. of this subpart. Page 2

A. A visual indication that is observable at the CBCT x-ray control and gantry whenever x-rays are produced; and Commented [JC(4]: SSRCR F. 11. A v (1) and 21cfr1020.33(h)(1) B. A signal that is audible to the qualified operator that the exposure has terminated. Subp. 6. Technique factors. A registrant is responsible for the technique factor Commented [JC(5]: SSRCR, Part F, p. 50 requirements of this subpart. A. The technique factors on manual and automatic exposure control x-ray systems must be: Commented [JC(6]: From SSRCR, Part F, p. 50 Item A is consistent with FDA 1020.31 (a) and Texas Administrative Code (1) indicated; and (2) visible to a qualified operator before the exposure begins. B. The requirements of item A may be met by permanent markings on dental CBCT x-ray systems that have fixed technique factors. C. An electronic or written protocol must be available at the control panel for each CBCT examination and the protocol must include: (1) the anatomical regions to be imaged; (2) all orientations and views that will be displayed; (3) scanner settings including: (a) field of view (FOV); (b) resolution; Page 3

(c) time; (d) kvp; (e) ma/mas; and (4) protocols for adult and pediatric. Subp. 7. Equipment performance evaluation; testing requirements; frequency. A registrant using a dental CBCT x-ray system is responsible for the equipment performance evaluation testing requirements under subparts 7 and 8. A. A registrant must have equipment performance evaluation testing performed over all clinical ranges used by the registrant. B. Initial equipment performance evaluation testing must be performed at installation before first patient use by: (1) a qualified medical physicist for a dental CBCT x-ray system that operates at 20 ma or greater or 100kV or greater. (2) a qualified service provider for a dental CBCT x-ray system that operates below 20 ma and below 100 kv. C. Periodic equipment performance evaluation testing must be: (1) performed by a qualified medical physicist at intervals not to exceed 12 months (365 calendar days) from the date of the previous equipment performance evaluation for a dental CBCT x-ray system that operates at 20 ma or 100kV or greater. A registrant may have a Page 4

grace period of 30 calendar days to comply with the periodic equipment performance evaluation testing interval requirement under this item. (2) performed by a qualified service provider at intervals not to exceed 24 months (730 calendar days) from the date of the previous equipment performance evaluation for a dental CBCT x-ray system operates below 20 ma and below 100 kv. A registrant may have a grace period of 30 calendar days to comply with the periodic equipment performance evaluation testing interval requirement under this item. D. If a registrant s dental CBCT x-ray system fails to meet any of the equipment performance evaluation testing under subpart 8, then a registrant must: (1) not use the dental CBCT x-ray system; and Commented [JC(7]: From 4732.0520, subp. 1, item E. (2) have a qualified service provider calibrate the dental CBCT x-ray system so that the operating parameter complies with this part. Subp. 8. Equipment performance evaluation; dental CBCT x-ray system. System performance measures must be evaluated using a phantom. Page 5

A. A registrant s dental CBCT x-ray system must meet manufacturer's specifications and tolerance limits for the following equipment performance tests: (1) contrast scale; (2) mean CT number of water and reference materials; (3) linearity; (4) slice thickness; (5) image display and storage; (6) safety analysis of audible and visual patient safety equipment; and (7) daily quality control tests under subpart 9. B. The manufacturer specifications required under item A must be available for: (1) use by a qualified service provider or a qualified medical physicist; and (2) review by the commissioner at the time of inspection. C. When a manufacturer s equipment performance evaluation specifications are not available, a registrant must [Placeholder: Seeking input on tolerance limits if not available.] Subp. 9. Quality control. A registrant using a dental CBCT x-ray system routine [daily and/or periodic] quality control tests must conduct equipment performance measurements Page 6

according to the manufacturer s recommendations, tolerance limits, and follow the requirements of this subpart. A. A registrant and radiation safety officer must develop and comply with written warm-up procedures, quality control procedures, and tolerance limits to evaluate image quality based on the dental CBCT x-ray system manufacturer s specifications. Daily procedures must include quality control Commented [TP(8]: SSRCR: The registrant [licensee] shall follow the QC recommendations provided by the CBCT manufacturer. In the absence of manufacturer provided QC recommendations, the registrant [licensee] shall implement and document QC guidelines established by a QMP [QE] in accordance to nationally recognized guidelines or those recognized by the Agency. tests for: (1) CT number accuracy for water and other reference material; (2) image noise; (3) artifact assessment; and (4) proper function of audible and visual patient safety equipment. Subitems (1) and (2) must be performed using a CT phantom by a qualified operator. B. Periodic procedures must include quality control tests with the use of a CT phantom for: Commented [JC(9]: Consistent with IAC CBCT testing section 2.3.2. IAC specifies periodic QA tests and does not define periodic. (1) spatial resolution for high and low contrast objects; (2) image uniformity; (3) image display and storage devices; and (4) air calibration, if required. Page 7

C. A registrant must have a CT phantom for use with a dental CBCT x-ray system at the facility according to 21 CFR 1020.33. D. A qualified operator must complete x-ray system warm-up procedures and daily quality control testing before first patient of the day using a CT phantom as specified by the dental CBCT x-ray system manufacturer. E. When a manufacturer s specifications on quality control are not available, a registrant s periodic quality control tests must [Placeholder: Seeking input on tolerance limits if not available.] Commented [JC(10]: Consistent with IAC CBCT testing section 2.3.2. IAC specifies periodic QA tests and does not define periodic. F. If any quality control test results fail to meet the manufacturer s specifications a registrant must: (1) remedy any corrective actions; and (2) conduct verification tests. G. A registrant must maintain quality control procedures, tolerance limits, and manufacturer s recommendations for: (1) use by a qualified operator; and (2) review by the commissioner at the time of inspection. Shielding Subp. 10. Shielding requirements. A registrant operating a stationary or a mobile dental CBCT x-ray system is responsible for the requirements of this subpart. Commented [JC11]: Based on 4732.0365, subps. 1, 2 Consistent with SSRCR Page 8

A. Maintain the dose levels so that they do not exceed the limits under parts 4732.#### to 4732.####; B. A dental CBCT x-ray system that is installed on or after the effective date of Commented [JC(12]: MDH intends to update the internal reference to the proposed equivalent of parts 4732.0410 47320.430 - DOSE LEVELS. this subpart must have the x-ray control permanently mounted behind a protective barrier so that a qualified operator remains behind the protective barrier during the entire exposure. C. A dental CBCT x-ray system that is installed before the effective date of this subpart and does not incorporate structural shielding or a shielded position must have a radiation survey that identifies radiation levels at the operator position and the spaces surrounding the dental CBCT x-ray system room or area by December 31 of the year following the effective date of this subpart. D. Corrective actions must be made when a radiation survey identifies that existing operating conditions may result in controlled (restricted) areas receiving a dose in excess of 0.1 mgy per week or uncontrolled Commented [JC(13]: Design goals taken from NCRP 145, pg. 102 F.2.2 Controlled area is defined in 4731.0100 subp. 49 and uncontrolled area is not; however, restricted (subp.205) and unrestricted (subp.256) are defined in 4731.0100 (unrestricted) areas receiving a dose in excess of 0.02 mgy per week. E. An evaluation must be performed to verify that corrective actions taken by the registrant resulted in controlled and uncontrolled areas receiving less than the restricted and unrestricted limits in subitem (1). F. Placing a clear identified marked restricted area outlining a safe distance for qualified operators. Page 9

Subp. 11. Shielding plan. A registrant with a dental CBCT x-ray system must comply with the shielding plan requirements under part 4732.####. Conditions of Operation Commented [JC(14]: MDH intends to update the internal reference to the proposed equivalent of part 4732.0360 - Shielding Requirements. Subp. 12. Qualified operators; dental CBCT x-ray systems. Qualified operators of a dental CBCT x-ray system include: A. a dentist licensed under Minnesota Statutes, section 150A.06; B. a dental therapist licensed under Minnesota Statutes, section 150A.06; C. a dental hygienist licensed under Minnesota Statutes, section 150A.06; and D. a dental assistant licensed under Minnesota Statutes, section 150A.06. Subp. 13. Prohibited uses. A registrant must not expose an individual to the useful beam except for healing arts purposes. Exposure to the useful beam is prohibited for: A. training; Commented [JC(15]: Replaces 4732.0305, subp. 1, item A (Prohibited Uses). Based on SSRCR part F p. 18 MDH intends to move this to General Requirements for all registrants and add a comparable qualified service provider requirement. B. demonstration; or C. other non-healing arts purposes. Subp. 14. Ordering of diagnostic radiographic examinations. A registrant is exempt from the requirements of part 4732.#### if: A. the registrant has a written procedure for ordering dental examinations; Commented [JC(16]: Contains revisions to part 4732.0560, subpart 3, suggested by Board of Dentistry and agreed upon by Dental Focus Group. MDH intends to update the internal reference to the proposed equivalent of part 4732.0560 (ORDERING OF DIAGNOSTIC RADIOGRAPHIC OR THERAPEUTIC PROCEDURE) Page 10

B. the written procedure is authorized and signed by a dentist who is licensed under Minnesota Statutes, chapter 150A.; and C. the written procedure is available and on site for review by the commissioner. Subp. 15. Utilization data. A registrant performing hand-held dental imaging examinations must maintain utilization data, in electronic or written form, including: A. a patient identifier; B. the type of examination; C. the date the examination was performed; D. the x-ray operator who is operating the x-ray system; E. the name of all individuals who remain in the room during an x-ray examination; and F. the name of all student externs if the registrant is an externship site underpart 4732.####. Subp. 16. Off-site use of a mobile dental CBCT x-ray system. A registrant must document the date and location when a registrant s qualified operator uses a registrant s mobile dental CBCT x-ray system at a location that is not listed on the registrant s equipment registration. Page 11

Subp. 17. Qualified operator protection. A registrant is responsible for the qualified operator protection requirements in this subpart. A. A qualified operator must remain behind a protective barrier or remain outside clearly identified and marked restricted area; B. At all times during an exposure, a qualified operator must be able to: (1) view the patient; and (2) monitor all entrances to the adjacent rooms or areas to prevent unauthorized access. C. A qualified operator must limit the useful beam to the patient's area of clinical interest. Subp. 18. Records. Commented [JC(17]: There will be one Records provision applicable to all registrants. Page 12