TITLE: Computerized Tomography of Projection Ultrasound. CONTRACTING ORGANIZATION: The Catholic University of America Washington, DC 20064

Similar documents
Keith 0. Plowden, Ph.D. Leonard Derogatis, Ph.D. U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

CONTRACTING ORGANIZATION: North Eastern Ohio Universities Rootstown OH 44202

CONTRACTING ORGANIZATION: University of California Lawrence Berkeley National Laboratory Berkeley, CA 94720

TITLE: Dietary Genistein and Prostate Cancer Chemoprevention

// Award Number: DAMD TITLE: Markers of Breast Cancer Risk in Women with Benign Breast Disease PRINCIPAL INVESTIGATOR:

TITLE: Neural Protein Synuclein (SNCG) in Breast Cancer Progression

TITLE: New Advanced Technology to Improve Prediction and Prevention of Type 1 Diabetes

AD (Leave blank) TITLE: Proteomic analyses of nipple fluid for early detection of breast cancer

TITLE: Evaluation of DNA Methylation as a Target for Intraductal Therapy for Ductal Carcinoma in Situ of the Breast

REPORT DOCUMENTATION PAGE OMB No

TITLE: 64Cu-DOTA-trastuzumab PET imaging in women with HER2 overexpressing breast cancer

TITLE: Outcomes of Screening Mammography in Elderly Women

CONTRACTING ORGANIZATION: Sloan-Kettering Institute for Cancer Research New York, NY 10021

CONTRACTING ORGANIZATION: New York University School of Medicine New York, New York 10016

TITLE: Dynamic Functional Mammoscopy: A Non-Ionizing Imaging Technique Enhancing Early Detection of Breast Cancer

TITLE: Hierarchical Nonlinear Mixed Effects Modeling: Defining Post-Radiation Therapy Relapse in Prostate Cancer Patients

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Do the Effects of Exercise on Breast Cancer Vary with Environment?

Fort Detrick, Maryland

AWARD NUMBER: W81XWH TITLE: Treatment of Pain and Autonomic Dysreflexia in Spinal Cord Injury with Deep Brain Stimulation

TITLE: A PSCA Promoter Based Avian Retroviral Transgene Model of Normal and Malignant Prostate

TITLE: Identification of New Serum Biomarkers for Early Breast Cancer Diagnosis and Prognosis Using Lipid Microarrays.

TITLE: Oxidative Stress, DNA Repair and Prostate Cancer Risk. PRINCIPAL INVESTIGATOR: Hua Zhao, Ph.D.

Jane Teas, Ph.D. University of South Carolina Columbia, South Carolina 29208

Award Number: W81XWH

TITLE: Long Term Outcomes of BRCA1/BRCA2 Mutation Testing. CONTRACTING ORGANIZATION: Georgetown University Washington, DC

Award Number: W81XWH TITLE: Global Positioning Systems (GPS) Technology to Study Vector-Pathogen-Host Interactions

TITLE: Identification of Chromosomes Alterations in Primary Breast Cancer Using Premature Chromosome Condensation

A Randomized Placebo-Controlled Trial of Citalopram for Anxiety Disorders Following Traumatic Brain Injury

Award Number: W81XWH

TITLE: Enhancing the Anti-Tumor Activity of ErbB Blockers with Histone Deaccetylase (HDAC) Inhibition in Prostate Cancer Cell Lines

U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Role of Estrogen Metabolism in the Initiation of Prostate Cancer: Biomarkers of Susceptibility and Early Detection

CONTRACTING ORGANIZATION: Western Institute For Biomedical Research Salt Lake City, UT

TITLE: Phase I and II Trial of Huanglian, A Novel Botanical Against Breast Cancer That Enhances Taxol Activity. Gary K. Schwartz, M.D.

Approved for Public Release; Distribution Unlimited

AWARD NUMBER: W81XWH TITLE: Gulf War Illness as a Brain Autoimmune Disorder. PRINCIPAL INVESTIGATOR: Apostolos Georgopoulos, MD, PhD

Detection of Prostate Cancer Progression by Serum DNA Integrity

TITLE: Reduction of Radiation- or Chemotherapy-Induced Toxicity by Specific Expression of Anti-Apoptotic Molecules in Normal Cells

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

CONTRACTING ORGANIZATION: Regents of the University of Michigan Ann Arbor, MI 48109

CONTRACTING ORGANIZATION: Evanston Northwestern Healthcare Research Institute

U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Inhibitors of Histone Deacetylases for Radiosensitization of Prostate Cancer

Deborah Watkins-Bruner, Ph.D. Fox Chase Cancer Center Philadelphia, Pennsylvania 19111

TITLE: The Prospective Army Coronary Calcium (PAAC) Study

Award Number: W81XWH TITLE: Regenerative Stem Cell Therapy for Breast Cancer Bone Metastasis

AD (Leave blank) Award Number: W81XWH TITLE: Targeting Autophagy for the Treatment of TSC and LAM. PRINCIPAL INVESTIGATOR: Elizabeth Henske

CONTRACTING ORGANIZATION: Oregon Health & Science University Portland OR 97239

CONTRACTING ORGANIZATION: Johns Hopkins Bloomberg School of Public Health

Approved for public release; distribution unlimited

Biomarker Analysis on Ductal Lavage Fluid as a Tool for Breast Cancer Early Detection

TITLE: Computerized Tailored Interventions for Behavioral Sequelae of Post-Traumatic Stress Disorder in Veterans

TITLE: A Tissue Engineering Approach to Study the Progression of Breast Tumor Metastasis in Bone

CONTRACTING ORGANIZATION: Cold Spring Harbor Laboratory Cold Spring Harbor, NY 11724

TITLE: Effect of Reminder Telephone Calls on Mammography Compliance in High Risk

TITLE: Genes Associated with Food Allergy and Eosinophilic Esophagitis

TITLE: Autocrine and Paracrine Control of Breast Cancer Growth by Sex Hormone-Binding Globulin

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

Approved for Public Release; Distribution Unlimited

Marcia Boehmke Jean K. Brown, Ph.D. Ruth McCorkle M. Tish Knobf Bill Wu. The State University of New York Amherst, NY

TITLE: Targeting the Immune System s Natural Response to Cell Death to Improve Therapeutic Response in Breast Cancers

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

Award Number: W81XWH TITLE: Dietary Fish Oil in Reducing Bone Metastasis of Breast Cancer

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

Targeting Stromal Recruitment by Prostate Cancer Cells

TITLE: Homeostatic and Circadian Abnormalities in Sleep and Arousal in Gulf War Syndrome

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Adipose Estrogen and Increased Breast Cancer Risk in Obesity: Regulation by Leptin and Insulin

TITLE: Post-Traumatic Headache and Psychological Health: Mindfulness Training for Mild Traumatic Brain Injury

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Maximizing Energy After Traumatic Brain Injury: A Novel Intervention

K I. Award Number: DAMD TITLE: Ethnic and Environmental Influence on Vitamin D Requirement in Military Personnel

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Exploring Early Detection Methods: Using the Intraductal Approach to Predict Breast Cancer

CONTRACTING ORGANIZATION: Vanderbilt University Medical Center Nashville, TN

AWARD NUMBER: W81XWH TITLE: Androgen Deprivation Therapy and Cognitive Impairment. PRINCIPAL INVESTIGATOR: Robert N. Pechnick, Ph.D.

CONTRACTING ORGANIZATION: Cleveland Clinic Foundation Cleveland, OH 44195

TITLE: Characterization of the Mechanisms of IGF-I-Mediated Stress-Activated Protein Kinase Activation in Human Breast Cancer Cell MCF-7

Expression and Promoter Methylation of P16INK4A During Estrogen-Induced Mammary Carcinogenesis in the ACI Rat. Omaha, NE

CONTRACTING ORGANIZATION: Mount Sinai School of Medicine New York, New York

TITLE: Prazosin for Treatment of Patients With PTSD and Comorbid Alcohol Dependence

TITLE: Short-Term Exercise and Prostate Cancer Prevention in African American Men. CONTRACTING ORGANIZATION: Howard University Washington DC 20059

La Jolla, CA Approved for Public Release; Distribution Unlimited

TITLE: A Phase II Trial of Androgen Suppression and Radiation Therapy with Samarium-1 53 in Localized, High-Risk, Prostate Cancer

TITLE: Targeted Eradication of Prostate Cancer Mediated by Engineered Mesenchymal Stem Cells

TITLE: Induction of Ephs/Ephrins-Mediated Tumor Cells-Endothelial Cells Repulsion as an Anti-Cancer Therapeutic Approach

TITLE: Effect of COX-2 (PGE2) and IL-6 on Prostate Cancer Bone Mets

CONTRACTING ORGANIZATION: Mount Sinai School of Medicine, New York, NY 10029

TITLE: MicroRNA in Prostate Cancer Racial Disparities and Aggressiveness

TITLE: Interchromosomal Associations that Alter NF1 Gene Expression Can Modify Clinical Manifestations of Neurofibromatosis 1. Palo Alto, CA 94304

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

TITLE: Role of ADAM15 in Tumor/Endothelial Interactions Prostate Cancer Regression

Sphingosine-1-Phosphate Receptor Subtype 3: A Novel Therapeutic Target of K-Ras Mutant Driven Non-Small Cell Lung Carcinoma

CONTRACTING ORGANIZATION: Fred Hutchinson Cancer Research Center Seattle, WA 98109

Award Number: W81XWH TITLE: Characterizing an EMT Signature in Breast Cancer. PRINCIPAL INVESTIGATOR: Melanie C.

Page 1 AWARD NUMBER: W81XWH TITLE: A Novel Pleiotropic Anti-Inflammatory Drug to Reduce ARDS Incidence. PRINCIPAL INVESTIGATOR: Gary Nieman

Transcription:

AD Award Number: DAMD17-01-1-0197 TITLE: Computerized Tomography of Projection Ultrasound PRINCIPAL INVESTIGATOR: Chu-Chuan Liu Yue Wang, Ph.D. CONTRACTING ORGANIZATION: The Catholic University of America Washington, DC 20064 REPORT DATE: September 2003 TYPE OF REPORT: Annual Siommary PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland 21702-5012 DISTRIBUTION STATEMENT: Approved for Public Release; Distribution Unlimited The views, opinions and/or findings contained in this report are those of the author(s) and should not be construed as an official Department of the Army position, policy or decision unless so designated by other documentation. 20040414 162

REPORT DOCUMENTATION PAGE Form Approved 0MB No. 074-0188 Public reporting burden for Ihls collection of information Is estimated to average 1 hour per response, Including the time for reviewing instoictions, searching existing data sources, gathering and maintaining the data needed, and completing and reviewing this collection of infonratlon. Send comments regarding this burden estimate or any other aspect of this collection of Information, Including suggestions for reducing this burden to Washington Headquarters Services, Directorate for Infonratlon Operations and Reports, 1215 Jefferson Davis Highvray, Suite 1204, Arlington, VA 22202-4302, and to the Office of Management and Budget, Papenwork Reduction Project (0704-018B), Washington, DC 20503 1. AGENCY USE ONLY (Leave blank) 2. REPORT DATE September 2003 4. TITLE AND SUBT/TLE Computerized Tomography of Projection Ultrasound 3. REPORT TYPE AND DA TES COVERED Annual Summary (30 Aug 2002 29 Aug 2003) 5. FUNDING NUMBERS DAMD17-01-1-0197 6. AUTHOR(S) Chu-Chuan Liu Yue Wang, Ph.D. 7. PERFORMING ORGANIZATION NAME(S) AND ADDRESSIESI The Catholic University of America Washington, DC 20064 8. PERFORMING ORGANIZATION REPORT NUMBER E-Mail: 1 OLiu@cua. edu 9. SPONSORING / MONITORING AGENCY NAME(S) AND ADDRESS(ES) U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland 21702-5012 10. SPONSORING / MONITORING AGENCY REPORT NUMBER 11. SUPPLEMENTARY NOTES Original contains color plates: All DTIC reproductions will be in.black and white. 12a. DISTRIBUTION / A VAILABILITY STA TEMENT Approved for Public Release; Distribution Unlimited 12b. DISTRIBUTION CODE 13. ABSTRACT (Maximum 200 Words) The second year of this training/research project focused on ultrasound tomography system simulation, integrating, and the tomography image reconstruction. A micro-stepping motor has been purchased and integrated in the transmission ultrasound laboratory system (Imperium Acoustocam) and form a Transmission Ultrasound Computerized Tomography (TUCT) prototype system. The stepping motor is capable of rotating at a 1.8-degree per step with 200 steps per revolution up to a 1600 steps per revolution with 0.225 degrees resolution. Based on the prototype system, the first experiment of UTCT images has been performed using a silicone phantom. The image profiles were acquired by the Imperium Acoustocam. These image intensity profiles were then converted into ultrasound attenuation and used to reconstruct the ultrasound attenuation tomography images through filtered backprojection reconstruction algorithm. Since the CMOS acquisition is not perfectly linear with 1100 chip, the newly developed 1300 chip will be used to perform the study in the future study. In addition, a scattering component will be added in the filtering algorithm to compensate for the attenuation signal. 14. SUBJECT TERMS Transmission ultrasound, ultrasound tomography, medical imaging, and breast cancer diagnosis computed tomography. 15. NUMBER OF PAGES 7 16. PRICE CODE 17. SECURITY CLASSIFICATION OF REPORT Unclassified 18. SECURITY CLASSIFICATION OF THIS PAGE Unclassified 19. SECURITY CLASSIFICATION OFABSTRACT Unclassified 20. LIMITATION OFABSTRACT Unlimited NSN 7540-01-280-5500 Standard Form 298 (Rev. 2-89) Prescribed by ANSI Std. Z39-18 298-102

Table of Contents Cover SF298 Table of Contents Introduction 1 Body 1 Key Research Accomplishments 4 Reportable Outcomes 4 Conclusions 4 References 4 Appendices

1. INTRODUCTION The second year of this trainmg/research project focused on ultrasound tomography system simulation, integrating, and the tomography image reconstruction. A micro-stepping motor has been purchased and integrated in the transmission ultrasound laboratory system (Imperium Acoustocam) and form a Transmission Ultrasound Computerized Tomography (TUCT) prototype system. The stepping motor is capable of rotating at a 1.8-degree per step with 200 steps per revolution up to a 1600 steps per revolution with 0.225 degrees resolution. Based on the prototype system, the first experiment of UTCT images has been performed using a silicone phantom. The image profiles were acquired by the Imperium Acoustocam. These image mtensity profiles were then converted into ultrasound attenuation and used to reconstruct the ultrasound attenuation tomography images through filtered backprojection reconstruction algorithm. The image intensity profiles, attenuation profiles and TUCT images of the testing phantom will be demonstrated in the following sections of this report. Since the CMOS acquisition is not perfectly linear with 1100 chip, the newly developed 1300 chip will be used to perform the study in the future study. In addition, a scattering component will be added m the filtering algorithm to compensate for the attenuation signal. 2. RESEARCH ACTIVITIES 2.1 System calibration and synchronization of the stepping motor with PC In our laboratory, located in the Georgetown University Medical Center (GUMC), the TUCT system consists of two sources of equipment: an transmission ultrasound prototype (Acoustocam by Imperium Inc. - Silver Sprmg, MD) and an stepping motor system (MS-1 by Stepperworld - Los Angels, CA). The Acoustocam has several major components: (1) A 5MHz ultrasound transducer and generator, (2) a compound acoustic lens, (3) a sensing CMOS array, and (4) two display monitors. The stepping motor system consists of a micro-stepping motor and a controller with power supply. The assembly of the two modules for the ultrasound tomography is shown in Figure 1 and Figure 3. The imaging system must be operated m Windows NT environment but the micro-stepping controller must be operated m Windows 95, 98 or ME environment. In order to solve this problem, a laptop was used with Windows ME to control the stepping motor and the control panel is shown m Figure 2. This stepping motor has several settings for the angle increment: full-step (1.8-degree), half-step (0.9- degree), quarter-step (0.45-degree) and 1/8 step (0.225-degree) for the control of the operation. p.i.i^i»iiimii'ji iii i.^iiiiii.iij,i.iiipamiiiiiptiiaiiiii!bi I PoioJtolPwl MtcraStepSilB PucucuWiorrMolli Senumcseifttl 37BH(8e«) i VB ij Dynmfc ' PnsKta 0000 Rote j»j, B 92 <ft FWMon "OH 1 f-ccw ' SmgMtep 1 1» IrSiflun J Start / Slop 1 Figure 1. Overall system configuration Figure 2. Micro-stepping control panel 1

Stepping motor controller Stepping motor power supply The Compound acoustic lens The Camera Stepping motor 5 MHz transducer Water tank Figure 3. System components of the prototype transmission ultrasound system. 2.2 Development for the TUCT Prototype A silicone phantom was made and was used for the first experiment. The phantom was mounted on the platform of the rotating rod and was immersed in the water. The stepping motor was then operated at VA step to provide the 0.45-degree increment of the rotation in the experiment. A total of 400 projections of the target were generated for 180 degrees rotation. The TUCT prototype collected the projection images from the CMOS sensing array. With these 400 projection images, we sample a single data profile from all projection views to form a sinogram. Each sinogram contains the same slice of the whole target at different angles. Hence we reconstructed an image of the target at corresponding slice based on the sinogram. For example, the red line in Figure 4(a) represents the desired slice that we would like to reconstruct form the projections. First, the bilinear interpolation method was employed to remove the inactive pixels as shown in Figure 4(b). Then, the line profiles from the 400 projection images were sampled to form an ultrasound intensity sinogram as shown in Figure 5 (a). And its corresponding ultrasound attenuation sinogram is shown Figure 6(a). By applying the parallelbeam backprojection method, the result intensity and attenuation images are shown in Figure 5. (b) and Figure 6(b), respectively. (a) Original projection image (b) After fixing inactive pixel Figure 4. One of the projection images of the phantom

r %-#.f (a) Original sinogram (b) Result image Figure 5. (a) The sinogram of the red Ime (b) The reconstructed result m 100 150 2fn (a) Extraction of the attenuation information (b) Ultrasound attenuation image without post-processing Figure 6. (a) The sinogram of the red line profile, (b) The reconstructed result. 2.2 A Brief Theoretical Background of the Ultrasound Tomography Let us assume that the origmal ultrasound intensity, lo, is attenuated while the energy is passmg through the material with a distance, z. I = lq&xip[-jga^dz]...(1) where / = Intensity at depth z IQ = Intensity at the source «2 = Attenuation coefficient In this experiment, the distance of the transducer to the CMOS sensing array is fixed. Thus, we can extract the attenuation information by the fouowmg equation:,...=,og(^)...,> The value on the each pixel of the attenuation sinogram, converted from mtensity sinogram using eq. (2), represents the Ime integral of the ultrasound transmission. The filter backprojection was then used to generate the attenuation image a^c^.y) This is demonstrated in the Figure 6(a) and 6(b). Figure 5(b), however, it does not have a physical meaning.

KEY RESEARCH ACCOMPLISHMENTS: Integrating an Acoustocam and a micro-stepping motor system to from a TUCT prototype system. A micro-stepping motor has been integrated a transmission ultrasound system to from a TUCT system that can generate ultrasound projection image at any angle. Test of the laboratory TUCT system: altered electronic component, designed phantoms and evaluated the laboratory system. Successfully getting prototype ultrasound computerized tomography images with a silicone phantom. REPORTABLE OUTCOMES: Poster presentation m the U.S. Army Medical Research and Materiel Command's (USAMRMC) Era of Hope 2002 DoD Breast Cancer Research Program Meeting in Orlando, Florida, September 25-28,2002. Abstract submitted to the 2004 SPIE Medical Imaging Conference. CONCLUSIONS: Base on our transmission ultrasound device and the micro-stepping motor we successfully generated the UCT images. The future work will be concentrated on two directions: "phantom design" and "filter design". We have tried several materials such as Jell-0 and vegetable gelatin (extracted for seaweed) but failed. One reason that we select silicone to test is because we could make phantom in any shape, any size and even we could insert object in it. But the critical angle of the silicone surface is too small so that the ultrasound waves may significantly reflected without much transmission in several angles. We plan to modify the shape of the silicone phantom or find a better material to make the breast phantom. The Ram-Lak filter and the Shepp-Logan filter may not be perfect for the ultrasound attenuation profiles acquired by Imperium 1100 chip. Therefore, we attempt to develop a better filter that could compensate for the ultrasound scattermg and non-linearity of the sensor array (1100). In addition, we plan to use Imperium 1300 or 1400 chip to perform the study. These chips have a better dynamic range. REFERENCES 1. A. C. Kak and M. Slaney, " Principles of Computerized Tomographic Imaging ", IEEE Press, 1988. 2. MS-1 MicTostcppingMotorfromhttp://www.stepperworld.com/ 3. Peter Fish, " Physics and Instrumentation of Diagnostic Medical Ultrasound", John Wiley & Sons Ltd., NY, 1990. 4. Jerrold T. Bushberg, 'The Essential Physics of Medical Imaging," Williams & Wilkins, 1994.