Pulse Oximeter Accuracy Study

Similar documents
Pulse Oximeter Accuracy Study

Testing Masimo rainbow Technology with vpad-a1 1. Background

Biomedical Instrumentation D. The Photoplethysmogram

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

Appendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires:

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Muscle OxyGen monitor. The Science Behind Moxy

Infinity MCable -Masimo SET

PULSE oximetry is a noninvasive technology for continuous

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood

Nonin Medical OEM Pulse Oximetry Solutions. Superior Performance for Challenging Patient Conditions

RESPONSE OF 10 PULSE OXIMETERS TO AN IN VITRO TEST SYSTEM

Trust Nonin Oximeters, Even For Your Sickest Patients

Monitoring Patients for Respiratory Depression: outside of the ICU. James D. Harrell, RCP San Diego Patient Safety Council November 14, 2014

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

Permanent City Research Online URL:

ALCO Regulations. Protocol pg. 47

Anesthesia Monitoring. D. J. McMahon rev cewood

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014

Accuracy of Pulse Oximetry in Children

TABLE OF CONTENTS. Physician's Manual. I. Device Description 2. II. Intended Use 5. III. Contraindications 5. IV. Warnings and Precautions 5

" Annex SpO 2 SPIROMETER D A T O S P I R MU2 *USER'S MANUAL* Rev.: 2.02

Consumer Near Infrared Spectroscopy as a Training and Recovery Tool. Jon Tarkington

3. Which of the following would be inconsistent with respiratory alkalosis? A. ph = 7.57 B. PaCO = 30 mm Hg C. ph = 7.63 D.

SUBJECT INFORMATION AND CONSENT FORM

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

Pulse Oximetry Test: Summary Report

MedStar Health considers Continuous Home Pulse Oximetry medically necessary for the following indications:

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Fundamentals of Pulse Oximetry

Clinical Policy: Oxygen Therapy in the Home Reference Number: CP.MP.485

By utilizing VirtuOx IDTF and our VPOD line of Overnight Pulse Oximeters your center will:

Empowering the RT with New Noninvasive Monitoring Capabilities

Home Pulse Oximetry for Infants and Children

They are updated regularly as new NICE guidance is published. To view the latest version of this NICE Pathway see:

Pulse oximetry in the accident and emergency department

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

CMS-50E Instructions by Cooper Medical Supplies (These instructions are to supplement the manufacturer s user manual not to replace it!

JAMES R. KNIGHT

Competitive Intelligence Report Technology & Patents

Defibrillators TEC-8300K series. It s all seamless

Parts Catalog Issued: May 12, Pulse Oximetry Equipment

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.

Pancreas: digestive enzymes Liver: bile-helps to digest fat, stored in gallblader; breaks down toxins, makes some parts of blood

Too Good to be True? Three Ways Cheap Pulse Oximeters Could be Short-changing Your Patients and Profits

All motion is not created equal.

Endorsements. Review Board. Pulse Oximetry and CO Oximetry 6/13/2007. Oximetry and CO-Oximetry for EMTs and

Medical Test Supply. User Manual. Wrist Pulse Oximeter OX-610A

HOME HEALTH FOUNDATION CLINICAL TEAM COMPETENCY Clinical Bag Technique. EMPLOYEE STAFF POSITION Date

Critical Care Monitoring. Assessing the Adequacy of Tissue Oxygenation. Tissue Oxygenation - Step 1. Tissue Oxygenation

Measuring body temperature, blood pressure, pulse, respiratory rate and oxygen saturation

Finger Pulse Oximeter. User Manual. Einstein Associates LLC Stafford Centre Drive Suite B Stafford, TX

Non-Invasive Monitoring

Is Hypoxia common despite oxygen treatment in the Acute Medical ward?

Study Of Effects Of Varying Durations Of Pre-Oxygenation. J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh

OXIMETRY TEST INSTRUCTIONS (to be performed by a regulated health professional)

Zacurate CMS 500DL. Finger Pulse Oximeter. User Manual

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology

Traceability of breath alcohol concentrations in Romania

Respiratory Physiology & Neurobiology

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana

Chapter 12 - Vital_Signs_and_Monitoring_Devices

Airway Clearance Devices

505 Recovery rates and training (Pulse Oximeter)

ISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164

uppression of Liver Blood Flow by Carbon Dioxide Pneumoperitoneum can be Improved by Prostaglandin E1 in Pigs

ihealth PO3 Fingertip Pulse Oximeter OPERATION GUIDE INDEX

Choosing Mindray SpO2

COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO :1998

Auto HbA1C ANALYZER HbA1LC HPLC 3660

Maintaining adequate patient tissue oxygenation

COPD. Ebrahim Razi, Hossein Akbari

Cover Page. The handle holds various files of this Leiden University dissertation

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM

Instructions to User

INTERNATIONAL STANDARD

Supplementary Online Content

The role of pulse oximetry in the accident and emergency department

Chapter 26. Assisting With Oxygen Needs. Elsevier items and derived items 2014, 2010 by Mosby, an imprint of Elsevier Inc. All rights reserved.

Health Tech Symposium Fall, Dan Sommers P.E. EMT-P

Capnography 101. James A Temple BA, NRP, CCP

Assessment of respiratory outcome among patients with lower respiratory tract disorders

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor

It costs you nothing, but gains everything for your patient!

RESULTS: The population had an age range of 46-99, with an average age of 71.5 years. The population was 44% female and 56% male.

Mechanical Ventilation. Assessing the Adequacy of Tissue Oxygenation. Tissue Oxygenation - Step 1. Tissue Oxygenation

Wearable Pulse Oximetry Measurements on the Torso, Arms, and Legs: A Proof of Concept

End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia

sounds are distant with inspiratory crackles. He sits on the edge of his chair, leaning forward, with both hands on his

THE RESPIRATORY SYSTEM Unit Quiz Preparation KEY

TECHNOCARE MEDISYSTEMS

April 18, Dear Mr. Lei Chen:

Dr. Puntarica Suwanprathes. Version 2007

Transcription:

Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709 Printed : 2008/12/24

This document is owned or controlled by Nihon Kohden and is protected by copyright law. Use of this document is permitted for personal, non-commercial purposes only. Modification, reproduction, transmission or distribution of this document is prohibited without prior, written authorization. NIHON KOHDEN CORPORATION. All rights reserved. Information in this document is subject to change without notice.

Nihon Kohden commissioned a study to evaluate Nihon Kohden pulse oximeter and probes at the Hypoxia Research Laboratory, University of California, San Francisco. The following document is a summarized version of the report from that study, created by Nihon Kohden Corporation. This document has been reviewed and approved by the investigators. This document is owned and controlled by Nihon Kohden and is protected by copyright law. Use of this document is permitted for personal, non-commercial purposes only. Modification, reproduction, transmission or distribution of this document is prohibited without prior written authorization. NIHON KOHDEN CORPORATION. All rights reserved. Information in this document is true accurate to the best of our knowledge at the time of publication.

SpO2 Accuracy Validation of Nihon Kohden Pulse Oximeters SpO2 Accuracy And Specifications Pulse oximeters measure oxyhemoblobin saturation (SpO2) from the signals of the red and infrared light transmitted through the tissue. The theoretical formula for calculating the SpO2 value from transmitted light signals is not completely established and each SpO2 manufacturer adopts their empirically developed calculation method. Therefore, the SpO2 accuracy depends on two factors of a) the validity of the calculating formula (the calibration curve) and b) deviations from the calculating formula. For this reason, ISO 9919, the International Standard of pulse oximeter equipment, requires each manufacturer to conduct controlled desaturation studies on human subjects and validate the SpO2 accuracy in comparison to gold standard (SpO2 value) which is intended to include the validation of those two factors previously mentioned. In the ISO 9919, the SpO2 value determined by analyzing arterial blood samples with a Hemoximeter is described as reference, and SpO2 accuracy of the pulse oximeter equipment is stated in terms of the RMS (root-mean-square) difference between measured value (SpO2) and reference value (SpO2). This RMS difference means that two-thirds of the SpO2 values measured by a pulse oximieter can be expected to fall within the range of RMS. All pulse oximeters and patient monitors with Nihon Kohden SpO2 technology are designed and manufactured to meet the same specifications for SpO2 accuracy. Nihon Kohden SpO2 accuracy is specified below. These specifications exceed the requirements of ISO9919 as noted. Nihon Kohden Pulse Oximeters Table 1. Accuracy criteria for SpO2 measurement SaO2 range SaO2 accuracy (RMS) 80 100% 2% or less 70 80% 3% or less ISO 9919: 2005(E) 70 100% 4% or less Summary of Invasive Controlled Desaturation Study on Healthy Volunteers Hypoxia Research Laboratory (University of California, San Francisco) conducted the study using Nihon Kohden pulse oximeters, OLV-3100, as a third party and reported the result. 1 SpO2 Monitoring Pulse Oximeter Accuracy Study

During this study, each subject was induced hypoxia to different arterial oxyhemoglobin saturation levels between 70% to 100% by controlling the inspired oxygen concentration, the SpO2 readings of pulse oximeters were recorded simultaneously with drawing arterial blood samples, and statistical analysis was performed between the SpO2 readings and SpO2 reference values from the Hemoximeters. The SpO2 reference values are determined from the readings of the two Hemoximeters. Exclusion criteria for this statistical analysis were predefined as follows: - Data which drifted 4% or more during the 10 second plateau reading. - Data where the two Hemoximeter values disagreed to 2% or more. A summary for the test result is shown below. The test results show that all tested probes meet the accuracy specifications. Table 2. SpO2 accuracy (RMS difference) studied by Hypoxia Research Laboratory, University of California, San Francisco Probe type SaO2 range 70 80% 80 100% Finger Probe TL-201T 1.62% 1.07% Multi-site Y Probe TL-260T 1.83% 1.56% Adult Disposable Probe TL-271T 2.05% 1.39% Finger Probe TL-631T 2.79% 1.45% Details are given on succeeding pages, Pulse Oximeter Report for Nihon Kohden Corporation, March 12-13 and April 16, 2008 (excerpted version) studied by Hypoxia Research Laboratory, University of California, San Francisco. October 20, 2008 Nihon Kohden Corporation Monitoring Technology Center SpO2 Monitoring Pulse Oximeter Accuracy Study 2

3 SpO2 Monitoring Pulse Oximeter Accuracy Study This page is intentionally left blank.

Pulse Oximter Report for Nihon Kohden March 12-13 and April 16, 2008 Hypoxia Research Laboratory University of California, San Francisco John R. Feiner, M.D. Phillip E. Bickler, Ph.D. M.D. John W. Severinghaus, M.D. SpO2 Monitoring Pulse Oximeter Accuracy Study 4

Methods The current study for Nihon Nihon Kohden performed March 12-13 and April 16, 2008, included 16 subjects-5 women and 11 men. A total of four oximeter probes, designated, TL201T, TL260T, TL271T, and TL631T. All were on model OLV-3100 oximeters. A radial arterial cannula was placed in either the left or right wrist of each subject. Blood gas analysis to determine oxyhemoglobin saturation was performed on an two OSM 3 multi-wavelength oximeters (Hemoximeter, Radiometer, Copenhagen). No subject was anemic (Hemoglobin < 10 gm dl -1 ). Each subject had control data taken at the beginning of each experiment, with two control blood samples drawn while breathing room air. Hypoxia was induced to different levels of oxyhemoglobin saturation (between 70-100%) by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide. Each plateau level of oxyhemoglobin saturation was maintained for at least 30 seconds and until pulse oximeters readings had stabilized. Two arterial blood samples were then obtained, approximately 30 seconds apart. A total of 24 samples were obtained per subject. Data were recorded Nihon Kohden and provided for analysis. Figure 1. Probes and Pulse Oximeter OLV-3100 used for study 5 SpO2 Monitoring Pulse Oximeter Accuracy Study

Data Analysis Data analysis was performed on Excel. Data was excluded if Hemoximeter values disagreed to 2 points or more, or if there was failure to achieve stable pulse oximeter measurements. Subject #7 was excluded because of an apparent upper respiratory infection, and Subject #9 because she was unable to complete the study. In the figures, a different marker is used for each study subject. Tables of mean, standard deviation, count and root mean square are provided for each oximeter s bias in the following ranges of SpO2 (Hemoximeter): 70-80%, 80-100%. Root mean square error (RMS error) is calculated as follows: RMSError = (SpO2 -SaO2) 2 n Table 3. Subject data Subject Ethnicity Gender Age Skin Subject #1 Caucasian M 30 Medium Subject #2 African M 23 Dark Subject #3 Asian M 25 Light Subject #4 Caucasian F 29 Medium Subject #5 Caucasian M 22 Light Subject #6 Caucasian M 24 Light Subject #7 Caucasian M 33 Light Subject #8 Caucasian F 30 Light Subject #9 Caucasian F 34 Light Subject #10 Caucasian M 28 Light Subject #11 Caucasian M 22 Light Subject #12 India F 25 Dark Subject #13 India M 25 Medium Subject #14 African M 27 Dark Subject #15 Mexican/Caucasian F 23 Medium Subject #16 India M 21 Med/Dark SpO2 Monitoring Pulse Oximeter Accuracy Study 6

TL-201T Finger Probe SpO2 accuracy Table 4. Result for TL-201T Bias (SpO2-SaO2) Analysis Probe TL-201T Hemoximeter SaO2 Range 70 80% 80 100% Count 158 464 Mean -0.47% -0.46% Standard Deviation 1.56% 0.97% Root Mean Square (RMS) 1.62% 1.07% Figure 2. TL-201T Bias (SpO2-SaO2) Pulse Oximete:OLV-3100 Hemoximete:OSM 3 Data supply: Hypoxia Research Laboratory, University of California, San Francisco 7 SpO2 Monitoring Pulse Oximeter Accuracy Study

TL-260T Multi-site Y Probe SpO2 accuracy Table 5. Result for TL-260T Bias (SpO2-SaO2) Analysis Probe TL-260T Hemoximeter SaO2 Range 70 80% 80 100% Count 158 464 Mean 0.08% -1.01% Standard Deviation 1.84% 1.18% Root Mean Square (RMS) 1.83% 1.56% Figure 3. TL-260T Bias (SpO2-SaO2) Pulse Oximeter:OLV-3100 Hemoximeter:OSM 3 Data supply: Hypoxia Research Laboratory, University of California, San Francisco SpO2 Monitoring Pulse Oximeter Accuracy Study 8

TL-271T Adult Disposable Probe SpO2 accuracy Table 6. Result for TL-271 Bias (SpO2-SaO2) Analysis Probe TL-271T Hemoximeter SaO2 Range 70 80% 80 100% Count 158 464 Mean 0.84% 0.53% Standard Deviation 1.88% 1.29% Root Mean Square (RMS) 2.05% 1.39% Figure 4. TL-271T Bias (SpO2-SaO2) Pulse Oximeter:OLV-3100 Hemoximeter:OSM 3 Data supply: Hypoxia Research Laboratory, University of California, San Francisco 9 SpO2 Monitoring Pulse Oximeter Accuracy Study

TL-631T Finger Probe SpO2 accuracy Table 7. Result for TL-631T Bias (SpO2-SaO2) Analysis Probe TL-631T Hemoximeter SaO2 Range 70 80% 80 100% Count 158 464 Mean -1.81% -0.76% Standard Deviation 2.13% 1.24% Root Mean Square (RMS) 2.79% 1.45% Figure 5. TL-631T Bias (SpO2-SaO2) Pulse Oximeter:OLV-3100 Hemoximeter:OSM 3 Data supply: Hypoxia Research Laboratory, University of California, San Francisco SpO2 Monitoring Pulse Oximeter Accuracy Study 10

First Edition December 19th, 2008 SpO2 Monitoring Pulse Oximeter Accuracy Study Patient Monitoring Technical Library NIHON KOHDEN CORPORATION 1-31-4 Nishiochiai, Shinjuku-ku, Tokyo 161-8560, Japan Phone +81 (3) 5996-8036 Fax +81 (3) 5996-8100

SpO2 Monitoring Pulse Oximeter Accuracy Study NIHON KOHDEN CORPORATION 1-31-4 Nishiochiai, Shinjuku-ku, Tokyo 161-8560, Japan Phone +81 (3) 5996-8036 Fax +81 (3) 5996-8100 http://www.nihonkohden.co.jp/