The Facts about Biphasic Defibrillation

Similar documents
SMART Biphasic. A Proven, Fixed, Low-Energy Defibrillation Waveform

Biphasic Clinical Summaries

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

Helping you save more lives. Philips HeartStart Resuscitation Solutions Leading the way in quality resuscitation

3Replacement Transducer Cables

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Early defibrillation programs for the workplace. Give your workplace the power to save a life

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

HeartStart. XL+ cardiac resuscitation. Ready. to save lives. Not available for sale in the U.S.

Making lifesaving faster, easier, better

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Emergency Care Clinical Data Transmission Networks

Ordinary person, extraordinary moment. Philips HeartStart OnSite Defibrillator

Early Defibrillation. Dr. M. Ravishankar

Cardiac Resuscitation. HeartStart FRx. Lead the way. to save a life Philips HeartStart FRx defibrillator with Life Guidance

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

AutoPulse Plus. Maximizes the likelihood of. Shock Success

Maximises the likelihood of. Shock Success

Triphasic Waveforms Are Superior to Biphasic Waveforms for Transthoracic Defibrillation Experimental Studies

Setting a standard in external defibrillation: Low energy biphasic waveforms. Richard M. Luceri, MD, FACC Editor

Sudden cardiac arrest associated with ventricular fibrillation

Journal of the American College of Cardiology Vol. 36, No. 3, by the American College of Cardiology ISSN /00/$20.

Defibrillation. Learning outcomes. Introduction. Mechanism of defibrillation. Factors affecting defibrillation. success. Transthoracic impedance

Defibrillators TEC-8300K series. Improving quality of resuscitation

1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium

Cardiac Resuscitation. HeartStart OnSite. Lead the way. to save a life Philips HeartStart OnSite defibrillator

Part 4: The Automated External Defibrillator

OTHER FEATURES SMART CPR

In settings of cardiac arrest and resuscitation,

Cardiopulmonary Resuscitation in Adults

Cardio First Angel EVERY LIFE COUNTS!

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS

Advanced Resuscitation - Adult

ICD: Basics, Programming and Trouble-shooting

The Importance of CPR in Sudden Cardiac Arrest

Improving quality of resuscitation. Defibrillators TEC-5600 series

THE FOLLOWING QUESTIONS RELATE TO THE RESUSCITATION COUNCIL (UK) RESUSCITATION GUIDELINES 2005

DSED: Is It Real? Brent Myers, MD MPH FACEP CMO and EVP of Medical Operations, Evolution Health Associate CMO, American Medical Response

Automated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if

The Automated Defibrillator: A Biomedical Engineering Success Story. Dr. James A. Smith

Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation

Advanced Resuscitation - Adolescent

One intelligent solution

Pediatric Cardiac Arrest General

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS

CHAPTER House Bill No. 1429

Cardiac Electrical Therapies. By Omar AL-Rawajfah, PhD, RN

Adult Advanced Cardiovascular Life Support 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular

Analysis Algorithm Overview

Cardiac resuscitation. HeartStart FRx. Lead the way to save a life

Power to see the big picture

QUICK SETUP GUIDE SECULIFE DFBASE

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

Wake County EMS System Peer Review/Clinical Data/System Performance

CPR Guidelines - Update

Electromagnetic Interference between External Defibrillator and Cardiac Resynchronization Therapy-Pacemaker (CRT-P) Devices

ADVANCED LIFE SUPPORT

A A J Adgey, M S Spence, S J Walsh

Resuscitation Guidelines update. Dr. Luis García-Castrillo Riesgo EuSEM Vice president

The ALS Algorithm and Post Resuscitation Care

Implantable Defibrillators Market by Product Type [Transvenous Implantable Cardioverter-Defibrillator (T-ICDs), Subcutaneous Implantable Cardioverter

Chain of Survival. Highlights of 2010 American Heart Guidelines CPR

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS

A simple solution for your complex patients

SUMMARY OF MAJOR CHANGES 2010 AHA GUIDELINES FOR CPR & ECC

Engineering involved in Cardiac Arrest Management Sahar Sorek

DEFIBRILLATORS CATALOG

New and Future Trends in EMS. Ron Brown, MD, FACEP Paramedic Lecture Series 2018

How to Perform CPR on a Cardiac Arrest Victim (Adult Victim Only)

Singapore DEFIBRILLATION. Guidelines 2006

Lesson 4-3: Cardiac Emergencies. CARDIAC EMERGENCIES Angina, AMI, CHF and AED

TENNESSEE Project ADAM. Preventing Sudden Cardiac Death (SCD): Implementing your school s Public Access to Defibrillation (PAD) program.

1. Types of Cardiac Arrhythmia

Use of Automated External Defibrillators (AED s) Frequently Asked Questions

Shock-induced termination of cardiac arrhythmias

What cardioversion protocol for ventricular fibrillation should be followed for patients who arrest shortly post-cardiac surgery?

Neurostimulation Systems

IMPACT OF THE 2005 AHA GUIDELINES ON RESUSCITATION OUTCOMES Ronna Zaremski, RN, MSN, CCRN

Developments in Cardiopulmonary Resuscitation Guidelines

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Protocol. This trial protocol has been provided by the authors to give readers additional information about their work.

One intelligent solution

Visualization of Defibrillation Simulation Using Multiple Transthoracic Electrodes

TITLE: Dose of Electrical Current for Biphasic Defibrillators for Synchronous Cardioversion in Patients with Tachyarrhythmia: Guidelines

Adrenaline 1mg in 10mL (1:10,000) Pre-filled syringe 3 Amiodarone 300mg/10mL Pre-filled syringe 5

Efficacy and safety of the reciprocal pulse defibrillator current waveform

MARS Ambulatory ECG Analysis The power to assess and predict

NOISE DETECTION ALGORITHM

Defibrillators TEC-8300K series. It s all seamless

Lecture. ALS Algorithm

A Bill Regular Session, 2005 HOUSE BILL 1231

AHA Scientific Statement

Hanna K. Al-Makhamreh, M.D., FACC Interventional Cardiologist

Biomedical Instrumentation

Transcription:

The Facts about Biphasic Defibrillation Introduction In reference to the SMART Biphasic waveform, Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, published by the American Heart Association, states The growing body of evidence is now considered sufficient to support a Class IIa recommendation for this low energy, BTE [biphasic truncated exponential] waveform. A Class IIa recommendation is defined as standard of care, is considered intervention of choice by majority of experts, and is supported by good to very good evidence. 1 At the same time, the AHA issued a similar recommendation for the general practice of low-energy biphasic defibrillation, but cautioned that at this time no studies have reported experience with other biphasic waveforms in long-duration VF in out-ofhospital arrest. When such data becomes available, it will need to be assessed by the same evidence-evaluation process as used for the biphasic AED and this guidelines process. 1

Not All Biphasic Waveforms Are the Same Different waveforms are designed to work at different energies. Consequently, an appropriate energy dose for one biphasic waveform may be inappropriate for a different biphasic waveform. There is already evidence to suggest that a waveform designed for low-energy defibrillation is less effective if applied at high energies. Tang demonstrated good resuscitation performance for the SMART Biphasic waveform, but found that more shocks were required at 200 J than at 150 J. 2 Conversely, a waveform designed for high-energy defibrillation may not perform effectively at lower energies. The same study showed poorer resuscitation performance for a 200 µf biphasic waveform at 200 J compared to a 100 µf biphasic waveform (SMART Biphasic) at 200 J. 3 Higgins showed that the 200 µf biphasic waveform performed better at 200 J than it did at 130 J. 4 Consequently, to determine the proper dosing for a given waveform, it is necessary to refer to the manufacturer's recommendations and the clinical literature. The recommendations for one biphasic waveform should not be arbitrarily applied to a different biphasic waveform. It would be irresponsible to use a waveform designed for high energy with a low-energy protocol simply to satisfy the letter of the current AHA recommendation. In the Guidelines 2000, the AHA states Energy levels vary with the type of device and type of waveform used. 1 Referring to studies involving the SMART Biphasic waveform, Guidelines 2000 states, This research indicates that repetitive lowerenergy biphasic waveform shocks (repeated shocks at 200 J) have equivalent or higher success for eventual termination of VF than defibrillators that increase the current (200, 300, 360J) with successive shocks (escalating). 1 Energy and Current The way the energy is delivered makes a big difference in the efficacy of the waveform. Electric current has been demonstrated to be the variable most highly correlated with defibrillation efficacy. The SMART Biphasic waveform uses a 100 µf capacitor to store the energy inside the AED; the Medtronic ADAPTIV biphasic waveform uses a 200 µf capacitor. The energy stored on the capacitor is given by the equation: E = ½ C V 2 The voltage and the current involved with defibrillating a patient are related to patient impedance by the equation: V = I R For the Medtronic biphasic waveform to attain levels of current similar to those of the SMART Biphasic waveform, it must apply the same voltage across the patient's chest. This means that to attain similar current levels, the Medtronic waveform must store twice as much energy on the capacitor and deliver much more energy to the patient; the graph below demonstrates this relationship. This is the key reason that the Medtronic biphasic waveform requires high-energy doses, whereas the SMART Biphasic waveform does not. Current (amps) 80 60 40 20 0 360J Monophasic Peak Current Levels Low Impedance (50 ohms) 150 J SM A RT Biphasic Medtronic 200J Biphasic Medtronic 300J Biphasic Medtronic 360J Biphasic 2

Special Circumstances It has been asserted that a patient may need more than 150 J with a biphasic truncated exponential waveform when conditions like heart attacks, high-impedance, delays before the first shock, and inaccurate electrode pad placement are present. This is not true for the HeartStart SMART Biphasic waveform, as the evidence presented in the following sections clearly indicates. On the other hand, the evidence indicates that more energy may be required for BTE waveforms other than SMART Biphasic. Heart Attacks Medtronic references two animal studies using their waveform to support their claim that a patient may require more than 200 J for cardiac arrests caused by heart attack (myocardial infarction). The SMART Biphasic waveform has been tested in real-world clinical settings with human heart attack victims and has proven its effectiveness at terminating ventricular fibrillation (VF). In a prospective, randomized, out-of-hospital study, the SMART Biphasic waveform demonstrated a firstshock efficacy of 96% versus 59% for monophasic waveforms, and 98% efficacy with three shocks as opposed to 69% for monophasic waveforms. 5 Fifty-one percent of the patients treated with SMART Biphasic had suffered a heart attack. The published evidence clearly indicates that the SMART Biphasic waveform does not require more than 150 J for heart attack victims. High-Impedance Patients High-impedance patients do not pose a problem with the lowenergy SMART Biphasic waveform. Using a patented method, SMART Biphasic technology automatically measures the patient's impedance and adjusts the waveform dynamically during each shock, to optimize the waveform for each shock on each patient. As demonstrated in published, peer-reviewed clinical literature, the SMART Biphasic waveform is as effective at defibrillating patients with high impedance (greater than 100 ohms) as it is with low-impedance patients. 5 The bottom line is that the SMART Biphasic waveform does not require more than 150J for high-impedance patients. Delay before the First Shock The SMART Biphasic waveform is the only biphasic waveform to have extensive, peer-reviewed, published emergency resuscitation data for long-duration VF. In a randomized out-ofhospital study comparing the low-energy SMART Biphasic waveform to high-energy escalating monophasic waveforms, the average call-to-first-shock time was 8.9 +/- 3.0 minutes. Of the 54 patients treated with the SMART Biphasic waveform, 100% were defibrillated - 96% on the first shock and 98% with three or fewer shocks. Of these patients, 76% experienced a return of spontaneous circulation (ROSC), versus only 55% of the patients treated with high-energy monophasic waveforms. 5 In a post-market, out-of-hospital study of 100 VF patients defibrillated with the SMART Biphasic waveform, the authors concluded, Higher energy is not clinically warranted with this waveform. 6 The SMART Biphasic waveform does not require more than 150 J when there are delays before the first shock. Inaccurate Electrode Pad Placement It has also been claimed that higher-energy escalating shocks are more forgiving of inaccurate electrode pad placement. This is a purely speculative argument with no basis in scientific evidence. However, common sense would suggest that if a given biphasic waveform needs more energy when pads are located properly, that waveform would also need more energy when pads are placed sub-optimally. Once again, real-world data demonstrate high efficacy with the SMART Biphasic waveform in out-of-hospital studies. 5,6 These studies included hundreds of AED users with a range of experience and training. 3

Commitment to SMART Biphasic All HeartStart AED products use the 150 J SMART Biphasic waveform. The HeartStart XL, HeartStart XLT, and HeartStart MRx provide a manual mode of operation for defibrillation and cardioversion as well as an AED mode. These products provide selectable energy settings from 2 to 200 J in the manual mode of operation and a constant 150 J in the AED mode. Some waveforms may need more than 200 J for defibrillation, but the patented SMART Biphasic waveform does not. Published clinical evidence indicates that the SMART Biphasic waveform does not require more than 150 J to effectively defibrillate, even if the patient has experienced a heart attack or has a higher-than-normal impedance, or if there have been delays before the first shock is delivered. Published clinical evidence also indicates that there is increased dysfunction associated with high-energy shocks. 7,8,9,10,11 With this in mind, why would you want to deliver more energy if you don't have to? With the SMART Biphasic waveform, you don't have to deliver more than 150 J. The SMART Biphasic waveform has demonstrated its efficacy in numerous studies and, as a result, has the evidence necessary for conformance with the AHA Class IIa recommendation for low-energy biphasic defibrillation. 4

References 1 Cummins RO, et al. Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care (ECC), I-63, I-5, I-90. 2 Tang W, et al. Effects of low- and higher-energy biphasic waveform defibrillation on success of resuscitation and post-resuscitation myocardial dysfunction after prolonged cardiac arrest. Circulation (supplement)1999:100(18):i-662 (abstract). 3 Tang W, et al. Low capacitance biphasic waveform shocks improve immediate resuscitation after prolonged cardiac arrest. Circulation (supplement)1999:100(18):i-663 (abstract). 4 Higgins SL, et al. A comparison of biphasic and monophasic shocks for external defibrillation. PreHospital Emergency Care 2000; 4:305-313. 5 Schneider T, et al. Multicenter, randomized, controlled trial of 150-J biphasic shocks compared with 200- and 360-J monophasic shocks in the resuscitation of out-of-hospital cardiac arrest victims. Circulation 2000;102:1780-1787. 6 Gliner BE, et al. Treatment of out-of-hospital cardiac arrest with a low-energy impedance-compensating biphasic waveform automatic external defibrillator. Biomedical Instrumentation & Technology 1998;32:631-644. 7 Weaver, et al. Ventricular defibrillation - A comparative trial using 175-J and 320-J shocks. New England Journal of Medicine 1982:307:1101-1106. 8 Xie J, et al. High-energy defibrillation increases the severity of postresuscitation myocardial dysfunction. Circulation. 1997;96:683-688. 9 Bardy, GH, et al. Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation. Circulation 1996;94:2507-2514. 10 Tang W, et al. Defibrillation with low-energy biphasic waveform reduces the severity of post-resuscitation myocardial dysfunction after prolonged cardiac arrest. Journal of Critical Care Medicine 1999;27:A43 (abstract). 11 Tokano T, et al. Effect of Ventricular Shock Strength On Cardiac Hemodynamics. Journal of Cardiovascular Electrophysiology 1998;9:791-797. 5

Philips Medical Systems is part of Royal Philips Electronics UNITED STATES: Philips Medical Systems Cardiac and Monitoring Systems 3000 Minuteman Road Andover, MA 01810 (800) 934-7372 CANADA: Philips Medical Systems Canada 281 Hillmount Road Markham, ON L6C 2S3 (800) 291-6743 EUROPE, MIDDLE EAST AND AFRICA: Philips Medizinsysteme Böblingen GmbH Cardiac and Monitoring Systems Hewlett-Packard Str. 2 71034 Böblingen Germany Fax: (+49) 7031 463 1552 LATIN HEADQUARTERS: Philips Medical Systems 1550 Sawgrass Corporate Parkway #300 Sunrise, FL 33323 Tel: (954) 835-2600 Fax: (954) 835-2626 ASIA PACIFIC HEADQUARTERS: Philips Medical Systems 30/F Hopewell Centre 17 Kennedy Road Wanchai Hong Kong Tel: (852) 2821 5888 Fax: (852) 2527 6727 www.medical.philips.com Philips Electronics North America Corporation, 2003 All rights are reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder. Philips Electronics North America Corporation reserves the right to make changes in specifications or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Published August, 2003, Edition 2