Characteristics Analysis of High Energy External Radiotherapy Beams in Water

Similar documents
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Quality assurance in external radiotherapy

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

A new geometric and mechanical verification device for medical LINACs

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

INTRODUCTION. Material and Methods

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Slide 1. Slide 2. Slide 3. Introduction of INTRABEAM IORT. Disclosure. Contents. I have nothing to disclose.

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Radiochromic film dosimetry in water phantoms

Electron therapy Class 3: Clinical procedures

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

IORT with mobile linacs: the Italian experience

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

SRS Uncertainty: Linac and CyberKnife Uncertainties


Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Diode calibration for dose determination in total body irradiation

Accurate Accurate boost or Simply Accuboost

Eric E. Klein, Ph.D. Chair of TG-142

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Expectations of Physics Knowledge for Certification

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Hampton University Proton Therapy Institute

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

A comparative dosimetric analysis of the effect of heterogeneity corrections used in three treatment planning algorithms

Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Commissioning and Radiobiology of the INTRABEAM System

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Imaging of Radiation Dose Using Cherenkov Light

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution

Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology

TREATMENT PLANNING: CLINICAL ELECTRONS John A. Antolak, Ph.D. Prepared for the AAPM Therapy Physics Review Course Austin, TX July 19 20, 2014 PURPOSE

Neutron Measurements for Intensity Modulated Radiation Therapy

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

The clinical use of OSLD

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MP (CLINICAL ONCOLOGY) PART I EXAMINATION - AUGUST Time : p.m p.m.

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE

The Journey of Cyberknife Commissioning

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Dosimetry in digital mammography

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Transcription:

Volume 3, No 1/2016 Research Article Malays. j. med. biol. res. ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) Characteristics Analysis of High Energy External Radiotherapy Beams in Water M. A. Rahman 1*, M. Jahangir Alam 2, M. Akhtaruzzaman 3 1 Department of Arts & Sciences, BAUST, Saidpur, Nilphamari, BANGLADESH 2,3 Ahsania Mission Cancer & General Hospital, Uttara, Dhaka, BANGLADESH *Email for Correspondence: mdashrafur.as@aust.edu ABSTRACT High energy external radiotherapy beam is being used widely for cancer treatment. Biological effect of radiation is concerned with the evaluation of energy absorbed in the tissue. The study of photon and electron beam characteristics is necessary before calibration machine. The aim of this study was to analysis characteristics of depth dose of different energy beams in water to enhance the quality of the radiotherapy treatment planning. Beam is attenuated by the medium and the transmitted beam with less intensity causes lesser absorbed dose as the depth increases. Relative attenuation on certain points on the beam axis and certain percentage of doses on different depths for 4-15 MV photon beams and 4-18 MeV electron beams have been investigated. Depth dose characteristics of the beams do not show identical attributes as interaction of the beams with matters is mainly governed by beam quality. Attenuation and penetration parameters show variation with dosimetric parameters like field size due to scattering and source to surface distance due to inverse square law but the major parameter in interaction is energy. Detailed analysis of the beam characteristics helps to select appropriate beam for radiotherapy treatment when variety of beam energies available and hence to increase accuracy in radiotherapy treatment. Key words: Radiation, Photon Beam, Electron Beam, Absorbed dose, Percentage Depth Dose and Treatment Planning Highlight: Depth dose characteristics of external beam helps to select appropriate beam for radiotherapy treatment. 2/10/2016 Source of Support: None, No Conflict of Interest: Declared This article is is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License. Attribution-NonCommercial (CC BY-NC) license lets others remix, tweak, and build upon work non-commercially, and although the new works must also acknowledge & be non-commercial. INTRODUCTION Radiation therapy refers to the treatment of malignant disease by using high energy radiation to kill the malignant cells or change genes so that cells stop growing. The energy beams affect the cells along their path as they go through the body to get to the cancer, pass through the cancer, and then exit the body. The goal is to deliver enough radiation doses to the tumor to destroy it while still limiting the dose to surrounding tissues (P. Metcalf et al., 2004). High-energy photon and electron beams are the most common form of radiations that have been used for cancer treatment. Linear accelerator is the most commonly used device that can generate an electron beam at various energies together with photon beam. Selection of appropriate set of treatment planning parameters is vital for determination of dosimetric characteristics of all radiation beams. Data on the percentage depth-dose of diagnostic X-rays are important in evaluating patient dose from medical exposure (Besim Xhafa at al, 2014 and Kato H et al, 2004). Quality of a radiation beam is most usefully expressed in terms of its penetrating power in radiotherapy which is a function mainly of the mean photon energy, and may be fully described by its depth dose characteristics in water (G.Narayanasamy et al, 2015) but an increase in surface dose with field size is also noted due to electron scattering from intervening materials (Khan F.M, 2003). Copyright CC-BY-NC 2014, i-proclaim MJMBR Page 51

Rahman etal: Characteristics Analysis of High Energy External Radiotherapy Beams in Water (51-60) The fundamental physical quantity of interest for relating radiation treatment to its outcome is the absorbed dose. Basic data of dose distributions usually derive from measurements in phantoms, and then use in a dose calculation system devised to predict dose distribution in an actual patient (Ravikumar M at al, 2002.). These phantoms are tissue equivalent and are made of different materials and different methods (White DR, at al, 1977). Dosimetry is a very significant element of radiotherapy treatment as all the treatment planning is based on the data obtained during dosimetry. Optimization of treatment plan, and calculation of dose for certain plan is performed when radiation physicist have measured dosimetry data. Absorbed dose in the body is dependent on depth, field size, beam energy and source to surface distance (SSD). Measurement of absorbed dose is performed using water or any other equivalent media phantom, which is kept perpendicular to the path of beam. This measurement is expressed as Percent of dose which gives a unique value for a certain set of parameters like beam energy, depth, SSD and field size. Variation in this value can be noted by change in any of these parameters (Wambersie A, et al., 2002). Beam which traverses a medium, the fluency decreases with increasing the depth but its energy stay practically constant and the percentage depth dose curve in ionization merges with the PDD curve in dose. MATERIAL AND METHOD It was intended to explore characteristics of high energy photon and electron beams of diverse range because of widely used for treatment of cancer patient. Doses were measured in water as it is always assumed to be a better phantom for being very close to human body due to its density and number of electrons per gram and universally available with reproducible radiation properties. The special supplement of British Journal of Radiology (BJR 25) serves as a guiding protocol for radiotherapy practices. The data of 10 cm 10cm field size for different energies were analyzed and examined the depths of maximum dose and depths where dose fall to half of its maximum value, d 50. The relationship between these depths according to beam energy have been investigated specially the difference between these two depths was examined in view point of beam energy. The distance function between these two dose levels was then compared for different beam energies to analyze the average decrease in dose with depth. Depth dose curve consists of two regions, the initial build-up region to the dose maximum followed by the exponential decay region which represents simple exponential attenuation in the water phantom. The important points such as dose at a depth of 0 cm and dose at a depth of 10 cm were located on the depth dose. These values are different for different field sizes even for same energy and source-to-surface distance (SSD). Similarly different depths can also be explored which gives a certain percentage of doses like depth of the maximum value of absorbed dose; d max and depth of 50 % dose; d 50. The surface dose can give an indication of the beam spectrum, for it is mostly due to the very low energy components of the beam. It is a general rule for photon that surface dose decreases with increasing beam energy and for any given beam energy increases with degradation of the spectrum towards the low energies. Unlike photon beams, the surface dose for electron increases with the increase in beam energy. High surface dose can have detrimental effects on the skin of therapy patients so it is desirable to minimize the surface dose. Relative attenuation can also be checked by comparing depths of certain percentage of dose rather than using doses at certain depths. The surface doses of a certain energy and field size were examined. It increases with increase both in energy and field size independently. Attenuation, penetration and scattering of beam results in a unique value of absorbed dose are function of depth, field size, source to surface distance, beam energy and the absorbing material. MEPHISTO software was used to analysis the beam parameters. RESULTS AND DISCUSSION For Photon Absorbed dose is a quantity which is used to quantify the exposure of biological objects to ionizaing radiation. Dose distributions along the beam central axis gives information required for an accurate dose description inside the patient.the dose is described usually as Percentage Depth Dose(PDD) which depends on the depth, field size, energy and source to surface diatance(ssd). The depths of measurement are typically at d max and 10 cm for verification of compliance with machine specifications, in addition to other depths required by the particular treatment planing system used in the department(e. B. Podrosak, Radiation Oncology Physics, A Hand book for Teachers and Students, p-194). PDD values measured at 10 cm for three specified photon beams are presented in the Table-1 and it can be seen that it increases with the beam energy. Table.1: Measured and tabulated values for PDD(10) for photon energies in the range 4-15 MV Energy in MV PDD(Measured) PDD(Tabulated)[8] 4 63.19 63.0 6 68.08 67.5 15 76.18 77.0 Page 52 Malaysian Journal of Medical and Biological Research Volume 3 Number 1/2016

Volume 3, No 1/2016 ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) When radiation beam enters in medium attenuation takes place and percentage depth dose varies with depth due to the attenuation. A typical depth dose curves for the photon is shown in Fig.1 which shows the relation between percentage depth dose and depth in the medium. One important thing is that the maximum dose, d max is not at the surface but at some depths. This is because of the range of the secondary electron (Khan FM. 2003). The elcetroans that are excited at the surface, on average, will travel a distance before they deposit dose. Beyond the depth of maximum dose, the PDD varies exponetially with the depth. Fig.1: Cental axis depth dose distribution for different quality photon beams The figure also indicates that for all energies the dose increases rapidly with the first few millimeters and then gradually achieves its maximum value at the depth of peak dose. In the case of 4 MV, the PDD increases from 59% to 83% in the first 3 mm, reaches 96% at 6 mm and achieves its maximum value at 11mm depth. Similarly maximum values occur at 15 mm and 25 mm for 6 MV and 25 MV photon beams respectively. Higher energy beams have greater penetrating power thus deliver a high percentage depth dose.with the increasing of the beam energy, the size of the chamber build up cap becomes excessively large and it is difficult to calculate the dose in free space from such measurement. The material of the build up cap different from the phantom introduces uncertainty in the tissue air ratio (TAR) measurements. To overcome the problem, the concept of tissue phantom ratio (TPR) was introduced for use in megavoltage iso-centric setup (M. Jahingir Alam et al, 2007). It is a measure of effective attenuation coefficient describing the exponential decrease with the depth beyond the depth of maximum dose shown in the Fig. 2. Fig. 2: Variation of Tissue phantom ratio (TPR) with Depth Beam quality correction factor, Kq, is a beam quality specifying parameter. The uncertainity of various dosimetric factors tend to cancel out by the use of the factor. For high energy photons with beam quality, Kq is specified by the TPR 20,10(M Saiful Huq et al., 2001). Both the parameters vary with the increasing beam energy. The data are presented in the Table-2. Copyright CC-BY-NC 2014, i-proclaim MJMBR Page 53

Rahman etal: Characteristics Analysis of High Energy External Radiotherapy Beams in Water (51-60) Table 2: Measured values of TPR 20,10 and Kq for different photon beams Beam Energy in MV TPR 20,10 Kq 4 0.969 0.997 6 1.018 0.996 15 1.149 0.975 The variation can be seen better to view in the Fig.3 where beam quality factor, Kq decreases with the Tissue phantom ratio, TPR 20,10. Fig.3: Beam qualitycorrection factor,kq as a function of beam quality parameter, TPR 20,10 The depths of maximum dose compared with depths of 50% dose for different beam energies. The difference between both depths increases with increase in beam energy.table-3 shows that d 50-d max is increasing with beam energy but important is that there is a diverse mode of variation. Table 3: Depth of maximum dose, depth of 50% dose and average decrease in percent dose per cm for photon beams of different energies Energy in MV Depth of 100% Dose in cm Depth of 50% Dose in cm d 50-d max Average decrease in percent dose per cm 4 1.1 13.57 12.57 4.0 6 1.5 15.22 13.72 3.64 15 2.5 18.98 16.48 3.04 The difference can be better to view in the Fig.4, where the relative difference between d max and d 50 of photon beams is plotted. The gap between curves seems to increase with energy to indicate a greater penetration. Less energy beam attenuates more than beam of higher energy. This relative attenuation analysis is helpful to analyse the beam attenuation, penetration and its ability to deliver the dose in specific increament of depth. Fig.4: Depth of maximum dose and 50% dose with beam energy Page 54 Malaysian Journal of Medical and Biological Research Volume 3 Number 1/2016

Volume 3, No 1/2016 ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) The dose falls off in the region between d max and d 50 can be evaluated with the average decrease in the dose per cm. The average decrease between these two depths decreases with increase in beam energy. The percentage depth doses for 4MV, 6MV and 15MV photon beams reduce to 4%, 3.64% and 3.04% per cm respectively. The relationship among these are obvious but the dose decrease rate do not have linear relationship. The reason is the mode of interaction with matter. The attenuation progression of high energy beam differs quite significantly from that of low energy beam because it interacts with matter with different attributes. Fig.5: Average Decrease in photon dose beyond d max Figure 5 shows the average decrease in dose in the region between d max and d 50.Here beam energy is plotted against PDD per cm and it can be seen that dose decrease is a function of energy. Average fall-off of the dose decreases with beam energy. This may not be a precise approach affirming the dose fall-off because it does not decrease in a continuous manner in the region between d max and d 50 or decrease certain value after every cm. In comparing the dose build up with dose fall off, the dose at 0 cm(surface dose) is needed so that increase in dose per unit of thickness with the decrease in dose per unit thickness beyond the depth of maximum dose. Measured values of surface dose for three photon energies and different field sizes are presented in Table-4. The surface dose for 15 MV photon for 10cm x 10cm field size is 35.5% and the absorbed attain its maximum value on a depth of 2.5 cm. It exhibits an average increase of 26.5% dose per cm. Similarly, the other two-photon energy beams have relatively grater surface dose but a smaller value of d max. The surface dose will decrease while d max increases with increase in energy shown in Fig.6. This is not generalized numerically for all energy values due to the relative importance of various types of interactions between photon and matter. Table 4: Surface dose of 4, 6 and 15 MV Photon beams for different field sizes Field size (cm x cm) 4MV % 6MV % 15MV % 5x5 56.55 45.43 28.20 10x10 60.77 49.57 35.85 15x15 63.6 54.26 40.89 20x20 67.38 57.97 47.32 25x25 69.63 61.39 51.15 30x30 72.26 64.08 54.55 Fig. 6: Variation surface dose with d max Copyright CC-BY-NC 2014, i-proclaim MJMBR Page 55

Rahman etal: Characteristics Analysis of High Energy External Radiotherapy Beams in Water (51-60) A photon beam propagating through a phantom is not only affected by the inverse square law but also by the attenuation or scattering of the photon beam inside the phantom. Beam energy influences isodose curve shape near field borders. Greater lateral scatter associated with lower-energy beams causes the isodose curves outside the field to bulge out. Therefore, the absorbed dose in the medium outside the primary beam is greater for low energy beam than that for higher energy beam. It can observe from the Fig. 7 that PDD(10) decreases with the absorbed dose in a very short region between d max and d 90 for 4 and 6 MV photon beam. The rates of falling in PDD are 0.25 and 0.30 for 4 and 6 MV photon beams respectively where the rates of increasing in absorbed dose for the same energies are 0.05 and 0.037 respectively. This was absent in 15 MV photon beam shown in the Fig.7. Therefore, it obserbes that the rate of falling increases in PDD but the rates of increasing in absorbed dose decreases with the increase of beam energy in the region of low energy. Sama Saed Al-Ahbabi et,al., 2012 obserbed the same for 6 MVphoton beam. Fig7: Variation of PDD (10) with Absorbed Dose of different energies For Electron The measurement of percentage depth dose along the central beam axis given information required for an accurate dose description inside the patient was conducted for energies (4-18) MeV which varies with the beam energy. The general shape of the central axis depth dose curves for electron beams differs from that of photon beams are shown in the Fig.8 for a field size of 10 x10 cm 2. Page 56 Malaysian Journal of Medical and Biological Research Volume 3 Number 1/2016

Volume 3, No 1/2016 ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) Fig.8: Cental axis depth dose distribution for electron beams The important thing is that the maximum doses are at some depths but not at the surface. The depths are comparatively lower than those for photon beam energies. The change in depth dose curve for higher energies was because of large angular scattering of the electron beams (T. Arunkumar, et al., 2010). It can see in the figure that the surface dose of electron beams is much higher than that for photon beams. Table 5: Surface dose and rate of increase of surface dose for different photon and electron beams Beam Energy MeV Surface dose for photon beam (%) Surface dose for electron beam (%) Rate of increase of photon surface dose [mm -1 ] Rate of increase of electron surface dose [mm -1 ] 4 60.77 82.10 3.56 1.99 6 49.57 84.60 3.36 1.18 15 35.85 95.20 2.56 0.25 The percent surface dose for electron beams increases with the energy. This may occur due to the nature of electron scatter. The electrons are scattered more easily and through larger angles for lower energies. This causes the dose build-up more rapidly and over a shorter distance. The rate at which the dose increase from surface to maximum decreases with beam energy for both of the beams but it is less for electron beam than for photon beam shown in the Table 5. The ratio of the surface dose (D s) to maximum dose (D max) is lower for lower-energy electrons than for higher-energy electrons shown in the Table.6 Therefore, the factor (D s /D max) increases with the increase in beam energy which can be better to view in the Fig. 9. Table 6: Ratio of surface dose to maximum dose for different electron beam energies Energy MeV Ratio of surface dose to maximum dose 4 0.82 6 0.846 8 0.896 10 0.901 12 0.923 15 0.952 18 0.964 Copyright CC-BY-NC 2014, i-proclaim MJMBR Page 57

Rahman etal: Characteristics Analysis of High Energy External Radiotherapy Beams in Water (51-60) Fig. 9: Variation of D s / D max with the beam energy Table 7: Typical Dose Parameters of Electron Beams Energy MeV R 100 R 50 cm R 80 cm R 90 cm Ē MeV d 50-d max Average decrease in percent dose per cm 4 0.9 1.732 1.413 1.216 3.94 0.832 60 6 1.299 2.488 2.033 1.822 5.80 1.189 42.05 8 1.70 3.421 2.795 2.508 7.97 1.721 29.05 10 2.100 4.046 3.327 2.997 9.43 1.946 25.69 12 2.299 4.797 3.935 3.538 11.18 2.496 20.03 15 1.802 6.092 4.879 4.30 14.19 4.290 11.65 18 1.400 7.158 5.732 4.971 16.68 5.759 8.68 The depth in centimeters at which electrons deliver a dose to the 80% to 90% is equal to about one-third to one-fourth of the electron energy in MeV shown in the Table.7. It can also see that the depth of maximum dose does not follow a linear relationship with the energy. The variation can be seen better to view in the Fig.10. Fig.10: Variation of depths of maximum, 90%and 80% doses with beam energy Page 58 Malaysian Journal of Medical and Biological Research Volume 3 Number 1/2016

Volume 3, No 1/2016 ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) The dose falls off in the region between d max and d 50 can be evaluated with the average decrease in the dose per cm. The data are given in Table 7. The average decrease between these two depths decreases with increase in beam energy.percentage dose for 4 MeV electron reduces 60% per cm but the same is 8.68% for 18 MeV beam. The relationship between two energies is obvious but the dose decrease rates do not have linear relationship. The reason may be mode of interaction with matter. High energy beam interacts with different attributes with the matter and hence its attenuation progression differes quite significantly from that of low energy beams. The beam energy is plotted against percent dose decrease per cm in Fig.11 in the region between d max and d 50 and it can be seen that average dose decrease is a function of energy and it falls of sharply. It can also be seen that the dose decrease does not follow a certain value after every centemetre. The similar trend also be noticed in the region between depth of maximum dose and depth of 90 and 80 % doses in Fig.11. Fig.11: Average decrease in electron dose in the region beyond d max The output dose varies with the field size. It can see from the Fig.12 that output dose ratio for all electron beams reached its maximum at a certain field size (10 x 10 cm 2 ) with increasing field size and then decreases with further increasing in field size. The intensity of the incident beam starts to decrease soon after its emission and it is significant after interacting with phantom material on its way. Intensity decreases continuously even in small fractions of depth changes. Therefore, incident electron has lower energy than that of the electron inside the accelerator given in Table 7. Fig. 12: Output dose for various field sizes of the electron beam Copyright CC-BY-NC 2014, i-proclaim MJMBR Page 59

Rahman etal: Characteristics Analysis of High Energy External Radiotherapy Beams in Water (51-60) CONCLUSION Characteristics analysis of depth dose for both the beams in radiotherapy treatment planning is helpful in achieving an increased degree of accuracy. Knowledge of energy and geometrical influence of different dosimetric parameters is indispensible for absorbed dose calculation. The clinically useful energy range for electron beams is 4 to 18 MeV. At these energies, the electron beam can be used for treating superficial tumor less than 5 cm deep with a characteristically sharp drop off in dose beyond the tumor. It may be noted that energy of photon beam is the major element of uniqueness of absorbed dose at certain depth in tissue or equivalent material. Due to other dosimetric consideration like field size and SSD, doses can have different values at certain location but the spectral and point to point distribution of the dose is the exclusive property of the beam energy. The detail analysis of the depth doses provides portions of depth containing a range and behavior of absorbed dose in water. Relative attenuation between two depths or between two doses describes the way in which dose decreases or increases for certain energy. The tumor or the target volume is never a small point. Keeping in view the shape and volume of the target along with its surrounding tissues require specific dose distribution. A beam capable of delivering the dose closely matching with the desired distribution always select for radiotherapy. The outcome of the present analysis will be helpful in view point to choose the appropriate beam for cancer treatment where a variety of beam energies is available. ACKNOWLEDGEMENT The authors would like to acknowledge the Director, Ahsania Mission Cancer and General Hospital, Uttara, Dhaka, Bangladesh for giving opportunity one of the authors to work there. Thanks to all technical staff of the Hospital for cooperation throughout process. REFERENCES Besim Xhafa, Tatjana Mulaj, Gezim Hodolli and Gazmend Nafezi; Dose distribution of photon beam by Siemens linear accelerator, International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2014; 3, 67-70. British Institute of Radiology, Central axis depth dose data for use in radiotherapy, Brit. J Radiol.1996; Suppl. 25. E. B. Podrosak, Radiation Oncology Physics, A Hand book for Teachers and Students, p-194. G.Narayanasamy,W Cruz, N Papanikolaou and S Stathakis. Comparison between measured tissue phantom ratio values and calculated from PDD with and without peak scatter correction factor in a 6 MV beam. Int J Cancer Ther Oncol, 2015; 3(2):03024 Kato H, Fujii S, Suzuki S. Method of calculating percentage depth dose for diagnostic X-rays. Nippon Hoshasen Gijutsu Gakkai Zasshi 2004;60:1107 15. Khan FM. Physics of Radiation Therapy. 3rd Edition, USA: Lippincott Williams & Wilkins; 2003. M Saiful Huq, Pedro Andreo and Haijun Song, Comparison of the IAEA TRS-398 and AAPM TG-51 absorbed dose to water protocols in the dosimetry of high-energy photon and electron beams; Phys. Med. Biol. 2001, 46, 2985-3006. M. Jahangir Alam, K. S. Rabbani, G. A. Zakaria, S. M. Akram Hossain, Adnan Kimber and Tanvir Baig; A modified formula for defining tissue phantom ratio of Photon beams, Bangladesh Med Res Counc Bull 2007; 33, 92-97. P. Metcalf, T. Kron and P. Hoban, The physics of Radiotherapy X-rays from Linear Accelerator. Medical Physics publishing, 2004. Ravikumar M, Ravichandran R. Dose measurements in the build-up region for the photon beams from Clinac-1800 dual energy medical linear accelerator. Strahlenther Onkol 2002; 176:223 8. Sama Saed Al-Ahbabi et,al, A comparison of protocols for external beam radiotherapy beam calibrations, Applied Radiation and Isotopes; 2012, 70, 1331-1336. T. Arunkumar, Sanjay S. Shupe, M. Ravikumar, S. Sathiyan, and M. Ganesh J Med Phys. 2010 Oct-Dec; 35(4): 207 214. Wambersie A, Menzel HG, Gahbauer RA, Jones DT, Michael BD, Paretzke H. Biological weighting of absorbed dose in radiation therapy. Radiat Prot Dosimetry 2002; 99:445 52. White DR, Martin RJ, Darlison R. Epoxy resin based tissue substitutes. Br J Radiol 1977; 50:814 21. -- 0 -- Page 60 Malaysian Journal of Medical and Biological Research Volume 3 Number 1/2016