UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD

Similar documents
MARATHON Cross-Linked Polyethylene

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

39 year old, 260 lb. park ranger. TKA in situ: 4 Years. First reported Metal-backed, non-modular PCA knee. 16% incidence. Modular metalbacked

TOTAL KNEE ARTHROPLASTY (TKA)

Over 20 Years of Proven Clinical Success. Zimmer Natural-Knee II System

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

The information contained in this document is intended for healthcare professionals only.

MicroSeal Total Hip Acetabular System. Like Nothing Else

HIGH FLEXION IN CONTEMPORARY TOTAL KNEE DESIGN: A PRECURSOR OF UHMWPE DAMAGE? A FINITE ELEMENT STUDY

The information contained in this document is intended for healthcare professionals only.

CURRICULUM VITAE CHRISTINE STEUBBEN HEIM Ward Drive Chesterland, OH Telephone:

Understanding Hip Implant Options

Optimum implant geometry

Instrumentation Options

Optimum implant geometry

CLASSIFICATION OF MOBILE BEARING KNEE DESIGN: MOBILITY AND CONSTRAINT

Regenerex Porous Titanium Construct. Knees Hips Extremities Cement and Accessories PMI Trauma Technology

Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc

Fixed Bearing Knees Function with Wear Resistance

Duraloc CONSTRAINED LINER

Bone Bangalore

Active Articulation E1

Section of total knee replacement. Total Knee Replacement System. Knieendoprothesen System. Système de prothèse totale de genou

The Leader in Orthopaedic Innovation

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

Peggers Super Summaries: THR and Revision Concepts

Optimum implant geometry

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Reaching new heights. Comprehensive. Efficient. Simple.

Zimmer Segmental System

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique

I want that next-generation knee

Sulzer Orthopedics Joint & Fracture Care

Consistency in U2 Knee System over Time

Ceramic Femoral Heads and Acetabular Cup Liners

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC

Gold standard of a TKA. Conflicting goals? POLYETHYLENE WEAR THE SOLUTION: MOBILE BEARING KNEES. MOBILE BEARING A totally new approach (1977)

Design Rationale and Surgical Technique. *smith&nephew TANDEM. Bipolar and Unipolar Hip System

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

Arcos Modular Femoral Revision System

RM Philosophy Innovation based on clinical evidence.

CPGAN #002. FTIR Quantification of Absorbed Radiation Dose in Polyethylene

G7 ACETABULAR SYSTEM. beautifully. efficient.

Precautionary Statement ( )

1Acetabular Reaming. 2Shell Sizing and Positioning. Trial Range of Motion. Trabecular Metal Natural Cup System

Stability without compromise. Epsilon Durasul Constrained Insert

StelKast Acetabular. Surgical Protocol and Product Specifications. Including ProForm, Provident, Provident Apical Threaded Hole & Systems

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

PINNACLE REVISION CUP SYSTEM

TriboFit Hip System. Issue 3 - August 2013 English

ACETABULAR CUP SURGICAL TECHNIQUE

Continuing the Tradition. VerSys Heritage Hip System

Massachusetts Institute of Technology Harvard Medical School Brigham and Women s Hospital VA Boston Healthcare System 2.782J/3.961J/BEH.

Reduction of In Vivo Oxidation induced by Lipid Absorption by Phospholipid Polymer Grafting on Orthopedic Bearings

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Exeter. Designed for Anatomic Restoration. Clinical Evidence Education

Progeny Hip Stem. Surgical Protocol and Product Specifications

Knee Replacement Implants

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

D E S I G N R A T I O N A L E

Total Knee Replacement

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

ArComXL Polyethylene: A New Way to Process Highly Crosslinked Polyethylene

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval

EMPOWERING LIFE THROUGH NATURAL MOTION

CC TRIO VERSAFITCUP. Surgical Technique. each to their own. Hip Knee Spine Navigation

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

REFLECTION Constrained Liner Introduction

Total Hip Replacement. Find out why the Anterior Approach may be right for you.

Revolution. Unicompartmental Knee System

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Versatility in a proven design. Trilogy Acetabular System

Luminus Flex Total Knee System (Fixed Type) The Innovative Technology for Knee Replacement

Furlong H-A.C. THR System

Comprehensive Reverse Shoulder System

Bi-Polar 22.2mm & 28mm System - Operative technique

BIOMECHANICAL MECHANISMS FOR DAMAGE: RETRIEVAL ANALYSIS AND COMPUTATIONAL WEAR PREDICTIONS IN TOTAL KNEE REPLACEMENTS

MULTIPOLAR BIPOLAR CUP. Surgical Technique

Identify the role of the skeletal system particularly in relation to maintaining an upright stance and protecting vital organs.

Featuring. Technology. Product Rationale

Dual Articular.

Bone Preservation Stem

CONSENSUS ORTHOPEDICS INC. CONSENSUS KNEE SYSTEM

EARLY CLINICAL RESULTS OF PRIMARY CEMENTLESS TOTAL KNEE ARTHROPLASTY

itotal Customized POSTERIOR-STABILIZED knee replacement system

SURGICAL TECHNIQUE. Protrusio Cage A COMPREHENSIVE ACETABULAR REVISION SYSTEM

Impingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C.

- October 2013 English

ConforMIS, Inc. 28 Crosby Drive Bedford, MA Phone: Fax:

Scan Bi-Polar 22/28. Operative Technique

Manza Cup HA SURGICAL TECHNIQUE.

Product Information & Procedure Wall Chart

Vanguard. Total Knee System

Clinical Study Comparison of Conventional Polyethylene Wear and Signs of Cup Failure in Two Similar Total Hip Designs

The Effect of Oxidative Aging on the Wear Performance of Highly Crosslinked Polyethylene Knee Inserts under Conditions of Severe Malalignment

Transcription:

UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 66th Annual Meeting February 4-9, 1999 Anaheim, California COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON HIP AND KNEE ARTHRITIS Prepared by: A. Seth Greenwald, D.Phil.(Oxon) Christopher J. Lyons, M.D. Jennifer R. Bohl, B.A. The following individuals are acknowledged for their contributions: Clive P. Duncan, M.D. Herbert Kaufer, M.D. Bernard N. Stulberg, M.D.

INTRODUCTION Ultra-high molecular weight polyethylene (UHMWPE) wear is an inevitable consequence of in vivo articulation. Its manifestation is determined by device, patient, and surgical variables which together define an orthopaedic triad, influencing the clinical longevity of hip and knee arthroplasty. Over the last three decades of arthroplasty use, patient profiles have substantially changed resulting in a greater service life demand for UHMWPE components. The following is an overview of some of the factors challenging this need for increased longevity. MATERIAL AND DESIGN FACTORS Bad Poly! The UHMWPE used in hip and knee components results from polymerization of ethylene gas into a fine resin powder of sub-micron and micron size distribution. The polymer is then consolidated using ram extrusion or compression molding techniques. Inadequate processing control can result in fusion defects arising from incomplete polymerization, voids and foreign body inclusions such as calcium sterate which ultimately contribute to the in vivo degradation of the final part. UHMWPE Sterilization Gamma irradiation in air has been the standard method of UHMWPE sterilization for many years in the orthopaedics industry. This process is also ascribed to improve the strength and wear performance through recombination of free radicals generated from the irradiation. These free radicals, however, will preferentially combine with any oxygen that may be present in the environment resulting in oxidation of the polymer. This leads to embrittlement over time and subsequent polymer degradation. Figure 1 is a cross-section of a three-year, gamma irradiated UHMWPE tibial component retrieval demonstrating a circumferential white band indicative of polymer embrittlement. Numerous fusion defects are also appreciated. Contemporary methodologies seek to overcome this problem by gamma irradiating in a depleted oxygen environment or through the use of EtO or gas plasma techniques. Figure 1 New Polys for Old? Previous attempts to improve UHMWPE performance have included carbon fiber reinforcement of the polymer mass (Poly-2) and, more recently, the reprocessing of conventional ultra-high molecular weight material to enhance mechanical properties (Hylamer). The former was withdrawn due to excessive inflammatory responses attributed to carbon fiber release while the latter continues to demonstrate mixed results in early reports. Figure 2, a retrieved Poly-2 tibial plateau demonstrates surface abrasion and pitting. Laboratory simulation demonstrates that UHMWPE resistance to wear is improved with increased cross-linking of the polymer chains. A number of Figure 2 thermal and chemical processing solutions have been advocated. One such approach involves storage of the components in an elevated temperature environment following gamma irradiation encouraging kinetic combination. Other techniques dramatically increase the radiation dose administered to the components followed by re-melting which changes the amorphous and crystalline regions of the polymer. These "new polys" must maintain a constancy of mechanical properties to ensure in vivo viability. Clearly, clinical experience will demonstrate the efficacy of these approaches.

Polymer Shelf Life As demonstrated in Figure 3, progressive changes in peak surface stress distributions following gamma sterilization in air and shelf storage over time increase the prospect of UHMWPE material compromise early after surgery. Figure 4 illustrates a tibial retrieval with a 9 year, 7 month shelf life that was removed 13 months after implantation. Gross delamination and pitting are noted. Evidence suggests that shelf dating of components may be a qualifying index in assisting improvements in component longevity. 8 Extension and 4 x BW (n=5) 6 Contact Area [mm 2 ] 4 2 2 4 6 3-Months 5-Years Figure 4 Duration of shelf storage after gamma irradiation of UHMWPE tibial inserts Sub Yield Post Yield > 4 MPa Figure 3 Influence of Polyethylene Thickness on Peak Surfaces Stress Component Design: Hip Liner Thickness: Thicker liners accommodate the annual progression of UHMWPE wear, a benefit to the longevity of component integrity in the younger patient. It has been demonstrated that peak contact stress attenuation is a further derived benefit (Figure 5) and that increasing thickness may also serve to accommodate shock absorption during high demand functional activities. Peak Surface Stress [MPa] 6 4 2 Acetabular Component 2 4 6 8 1 12 Polyethylene Thickness [mm] From Bartel et al., JBJS, 68A:141, 1986 Figure 5 Liner Conformity: Lack of conformity between a polyethylene liner and its acetabular shell as demonstrated in Figure 6 limits the surface area available for load transfer and concentrates surface and subsurface stresses in regions of contact increasing the prospect of permanent material deformation and backside wear contributing to component failure. Figure 6

Component Design: Hip (cont.) Head Size: The use of 32mm femoral heads results in increased volumetric wear (Figure 7) in comparison to the increased linear wear observed when 22mm heads are employed (Figure 8). Enhanced articular performance lies midway in the selection of 26 or 28mm head sizes. The Influence of Femoral Head Size on Wear Volume 2. The Influence of Femoral Head Size on Linear Wear Wear Volume [mm 2 ] 8 6 4 2 22mm 28mm 32mm 32mm Greatest Wear Volume (p<.1) Linear Wear [mm] 1.5 1..5 22mm 28mm 32mm 28mm Least Linear Wear (p<.1) Figure 7 Femoral Head Diameter [mm] From Livermore et al, JBJS, 72A:518, 199 Figure 8 Femoral Head Diameter [mm] From Livermore et al, JBJS, 72A:518, 199 Locking Mechanism Integrity: Deficient mechanical capture of the shell/liner composite in modular designs has contributed to early UHMWPE failure. Figure 9 demonstrates galling of the metal cup, liner separation, and polyethylene fracture one year following index surgery due to a poorly designed locking mechanism. Figure 9 Material Caveats: The use of titanium as an articulation counterface in hip and knee designs has resulted in accelerated wear of both the polymer and titanium surfaces contributing to early reports of lytic response and component loosening. The five-year retrieval shown in Figure 1 demonstrates an out of round geometry suggesting considerable loss of titanium substance. Figure 1

Component Design: Knee Conformity: The increased tibial-femoral conformity realized in PCL sacrificing knee plateaus serves to enhance UHMWPE service life by attenuation of peak contact stresses responsible for material damage. This is appreciated in the comparison shown in Figure 11 between PCL preserving and PCL sacrificing plateau geometries articulating against a common femoral component. Contact Area [mm2] 2 Extension and 4 x BW (n=3) Extension and 4 x BW (n=5) 3 PCL Preserving PCL Sacrificing Sub Yield Post Yield > 4 MPa Contact Area [mm 2 ] 5 15 2 25 6.7 8.7 1.7 12.7 15.7 Minimum Tibial Insert Thickness [mm] Sub Yield Post Yield > 4 MPa Figure 12 Figure 11 Plateau Thickness: Recent findings demonstrated in Figure 12 describe the relative insensitivity of peak surface stress distribution to thickness changes for a conformal knee design with tibial inserts greater than 6mm. Therefore, bone stock preservation, joint line restoration, PCL retention, patient weight and activity level should be the primary factors influencing component thickness selection. Third-Body Wear: The interaction of third-body particulate between articulation surfaces in hip and knee replacements consistently demonstrates catalysis of UHMWPE damage. Surface scratching of the metallic counterface resulting from these interactions further contributes to the damage process. Figure 13 shows an early retrieval of a cementless metal-backed tibial component demonstrating the effects of third-body entrapment. Figure 13 Locking Mechanism Integrity: Observed effects of UHMWPE insert micro-motion in retrieved modular tibial components suggest the presence of adhesive and abrasive wear which will influence component durability. This raises concerns regarding the effectiveness of current tray surface finish and capture mechanisms. The presence of adhesive film transfer is visualized in Figure 14 and is indicative of insert rotation. Figure 14

Component Design: Knee (cont.) Extension and 4 x BW (n=3) Femoral Component Finishing: The use of precision ground components enhances articulation conformity and improves surface stress distribution reducing the prospect of polymer wear. This is appreciated in Figure 15 which demonstrates the advantage of precision ground components through maximized contact area and minimized peak surface stress distributions. Contact Area [mm 2 ] 2 3 4 Belt Finished Precision Ground Sub Yield Post Yield > 4 MPa PATIENT FACTORS Figure 15 An increasing under 65-year old patient population demonstrating more active lifestyles and longer life expectancy directly leads to greater in vivo service life demands on UHMWPE components. Quantitative assessments of walking suggest that cyclic activity in joint replacement patients varies from less than 1 million cycles per year in the elderly to greater than 3 million in younger populations. Clinical observations between patient weight and damage in retrieved components point to the significant influence of this further variable. For every one pound of weight loss in the obese patient, three pounds are removed from the load acting on the patient s hip. Other factors of patient habitus, including individual musculoskeletal assessment, will further influence UHMWPE integrity. SURGICAL FACTORS In 1997, over 55, hip and knee arthroplasty procedures were carried out in North America. Sixty percent of these were performed by orthopaedic surgeons who do twenty or fewer joint replacements annually. Clinical outcomes are directly influenced by physician experience. The following surgical factors effect the loading across UHMWPE articulating surfaces, influencing polymer durability over time. Hip Factors Knee Factors maximization of femoral head coverage acetabular cup medialization restoration of femoral head offset avoidance of femoral neck impingement accurate component placement ligamentous balance optimization of alignment TAKE HOME MESSAGE It is an irony that as the millenium is approached, the UHMWPE solution for achieving low-friction arthroplasty advanced three decades ago by Sir John Charnley has, in itself, become a significant problem challenging the longevity of hip and knee replacement procedures. This exhibit describes a triad of device, patient, and surgical variables that influence the demanded longevity of UHMWPE. It is appreciated that polymer compromise is indeed a multi-factorial process. AAOS 1999