Aina Blood Monitoring System

Similar documents
Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

State of the Art of HbA1c Measurement

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Why do we need SD LipidoCare?

Determination of hemoglobin is one of the most commonly

3. System Accuracy Evaluation LUX Hgb test 1. Samples

Smart Diagnostics for Point-of-Care!

1 PROTOCOL. Comparison Study Summary

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology

Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing. Vincent Delatour, PhD

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Should HbA1C measured by POC instruments be used for diagnosis of diabetes? Sverre Sandberg, Norwegian Quality Improvement of Primary Care

Mission Hb accurately detects both

Dubrovnik 2014 Sverre Sandberg, Norwegian Quality Improvement of Primary Care Laboratories Noklus, Bergen, Norway

The Integrated Cardiovascular Clinical Network CHSA

Selecting a Hemoglobin A1C Point of Care Instrument. Objectives. Disclosure / Conflict of Interest 3/11/2015

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Performance of three portable blood glucose monitoring devices used in a veterinary application

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

HbA1c (Human) ELISA Kit

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA

NICO /14 Nipro Diagnostics, Inc.

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

MQ Comparison of glucometers using heparin whole blood

Insulin (Porcine/Canine) ELISA

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

RayBio Human Granzyme B ELISA Kit

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

Multi User Model. Product Overview

25 OH Vitamin D Rapid Test

EQAS. Hemoglobin Program (BC80) Cycle 12: December 2014 December 2015 Sample No: 1 Sample Date: 17 Dec 14. Exceptions. Customer Information

High-density Lipoprotein Cholesterol (HDL-C) Assay Kit

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

ISO 15197:2013 / EN ISO 15197:2015 Clinical Validation Study List for Blood Glucose Monitoring System

Apolipoprotein A-1 ELISA

FinTest IgG4 Screen 20 ELISA KIT

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual

EliKine Free Thyroxine (ft4) ELISA Kit

APOB (Human) ELISA Kit

Rapid Hemoglobin A1c Testing for Evaluation of Glucose Control

Vitamin A-E Serum HPLC Analysis Kit

Venous Blood Based Biomarker Data from the China Health and Retirement Longitudinal Study (CHARLS)

Professional Procedure Guide (90821B 04/2008)

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

TSH Receptor Autoantibody ELISA

Rat Hemoglobin A1c (HbA1c) Kit Instructions

RayBio Human ENA-78 ELISA Kit

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

Human LDL ELISA Kit. Innovative Research, Inc.

Parvovirus B19 IgM Human ELISA Kit

MQ Comparison of glucometers using whole blood

RayBio Human Thyroglobulin ELISA Kit

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere HIV Combo WHO reference number: PQDx

System Accuracy Evaluation LUX Chol HDL test

COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

References Required document for Laboratory Accreditation by the College of American Pathologists.

MQ Comparison of glucometers using whole blood

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

Relationship between glucose meter error and glycemic control efficacy

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers

A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Chymotrypsin ELISA Kit

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

2 Screen Islet Cell Autoantibody Human ELISA

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

HbA1c3 PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

Campylobacter Antigen ELISA Kit

Diabetes Monitoring: Easy as A-1-C An educational webinar sponsored by Siemens Healthcare Diagnostics

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

METHOD VALIDATION: WHY, HOW AND WHEN?

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.

Transcription:

Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale in Europe, India, Singapore, and Malaysia. When used by clinicians to manage chronic diseases, the Aina Blood Monitoring System is: Safe and reliable manufactured under the highest quality standards in a ISO 13485:2012 certified environment. Clinically proven clinically validated at numerous reputed clinical sites across the world. Proactive and personalized integrated with a patient-facing disease management platform. Revision 2.2 Last Updated : 27 July, 2017

Table of Content HbA1c Intended use........... 1 Test Principle........... 1 Specifications.............. 1 Clinical Studies............ 1-3 I. Narayana Health Hospitals (Bangalore, India)... 1 II. Narayana Health Hospitals (Bangalore, India)... 1 III. Study against Siemens DCA Vantage (Bangalore, India)..... 2 IV. Khoo Teck Puat Hospital (Singapore)....... 2 V. CARE Hospitals (Hyderabad, India)...... 2 VI. Dr. Mohan s Diabetes Specialities Center (Chennai, India)...... 2 Blood Glucose Intended use........... 4-5 Test Principle........... 4 Specifications.............. 4 Clinical Studies............ 4-5 I. Narayana Health Hospitals (Bangalore, India)... 4 II. Dr. Mohan s Diabetes Specialities Center (Chennai, India)..... 5 Hemoglobin Intended use........... 6 Test Principle........... 6 Specifications.............. 6 Clinical Studies............ 6 I. Narayana Health Hospitals (Bangalore, India)... 6 II. Khoo Teck Puat Hospital (Singapore)..... 6 Lipids Intended use........... 7-9 Test Principle........... 7 Specifications.............. 7 Clinical Studies............ 7-9 I. Narayana Health Hospitals (Bangalore, India)... 7 II. CARE Hospitals (Hyderabad, India)...... 8

HbA1c Intended Use The Aina HbA1c Monitoring System is intended to be used for the quantitative measurement of Glycosylated Hemoglobin (HbA1c) levels in capillary fingerstick and venous whole blood samples. This system is intended for clinical laboratory and point-of-care use to monitor long term glycemic control of persons previously diagnosed with diabetes. This test can also be used as an aid for screening or diagnosis of diabetes. Test Principle The Aina HbA1c Test utilizes the boronate affinity method. The Aina HbA1c Test Kit consists of test strips, reagents, wash buffers, capillary tubes for sample collection, and pipette tips. The reagent contains a lysing agent and a blue boronic acid conjugate. When blood is added to the reagent, the erythrocytes are lysed and all hemoglobin precipitates. The boronic acid conjugates binds to the glycosylated hemoglobin. An aliquot of the reaction mixture is applied to the test strip and all the precipitated hemoglobin, conjugate-bound and unbound, remains on top of the filter. Any unbound boronate is removed with the wash buffer. Specifications Measuring range: 4 to 15% HbA1c Supported hemoglobin range: 10 to 20 g/dl Test time: 3 minutes Operating temperature: 18 to 40 C Blood volume: 5 μl (whole blood capillary or venous) Clinical studies I. Narayana Health Hospitals (Bangalore, India) Reference method: Variant II Turbo (Bio-Rad) Number of subjects: 100 y = 0.993x + 0.073, R 2 = 0.972 II. Narayana Health Hospitals (Bangalore, India) Reference method: Alere Afinion AS100 Number of subjects: 63 All samples measured on the Aina HbA1c Monitoring System were found to be within 10% of the reference and showed great linearity (R 2 = 0.976). y = 1.005x - 0.183, R 2 = 0.976 Bio-rad Variant II Turbo (HbA1c%) Figure 1. Comparison of performance between Aina Device HbA1c and Bio-Rad Variant II Turbo HbA1c Alere Afinion (HbA1c%) Precision Evaluation A precision evaluation was performed using venous whole blood samples and showed a CV under 3% at all HbA1c levels. Level 1 Level 2 Level 3 Level 4 Figure 2. Comparison of performance between Aina Device HbA1c and Alere Afinion AS100 HbA1c Precision Evaluation A precision evaluation was performed using venous whole blood samples and showed a CV under 4% at all HbA1c levels. HbA1c Mean (%) SD (%HbA1c) 5.3 0.11 6.2 0.15 9.0 0.24 11.4 0.29 Control Level Mean Value Within Run Total CV (%) 2.16 2.34 2.66 2.56 L1 5.68 2.53 1.96-3.58 3.92 3.19-6.00 L2 13.08 2.53 1.95-3.58 2.85 2.50-3.77 Page 1

III. Study against the Siemens DCA Vantage (Bangalore, India) Reference method: DCA Vantage (Siemens) Number of subjects: 30 A total of 30 samples were included in the study. Measurements were taken in duplicates for all samples. All samples measured on the Aina HbA1c Monitoring System were found to be within 10% of the reference and showed great linearity (R 2 = 0.971). V. CARE Hospitals (Hyderabad) Reference method: Unicel DxC (Beckman Coulter) Number of subjects: 100 The Aina HbA1c Monitoring System showed great linearity versus the laboratory reference (R 2 = 0.963). In addition, 96% of the samples were within 10% of the reference. y = 0.948x + 0.308, R 2 = 0.971 Beckman Coulter UniCel DxC (HbA1c%) Siemens DCA Vantage (HbA1c%) Figure 3. Comparison of performance between Aina Device HbA1c and DCA Vantage (Siemens) HbA1c IV. Khoo Teck Puat Hospital (Singapore) Reference method: c501 (Cobas) Number of subjects: 41 The Aina HbA1c Monitoring System showed great linearity versus the laboratory reference (R2 = 0.982). In addition, 92.7% of the samples were within 10% of the reference. y = 1.138x + 0.949, R 2 = 0.982 Figure 5. Comparison of performance between Aina Device HbA1c and UniCel DxC (Beckman Coulter) HbA1c VI. Dr. Mohan s Diabetes Specialities Center (Chennai, India) Reference method: Variant II Turbo (Bio-Rad) Number of subjects: 131 The Aina HbA1c Monitoring System showed excellent correlation and agreement when compared to the laboratory gold standard for both capillary and venous blood samples. Performed Using Capillary Whole Blood Samples y = 1.033x - 0.153, R 2 = 0.951 Cobas c501 (HbA1c%) Figure 4. Comparison of performance between Aina Device HbA1c and c501 (Cobas) HbA1c Bio-Rad Variant II Turbo (HbA1c%) Figure 6. Performed Using Capillary Whole Blood Samples Page 2

The Aina HbA1c Monitoring System versus the gold standard revealed a correlation of 0.974 and a kappa agreement of 0.98. 94.5% of individual measurements measured on Aina HbA1c Monitoring system were within 10% of reference method (Variant II Turbo, Bio-rad). 98.3% of individual measurements measured on Aina HbA1c Monitoring system were within 15% of reference method (Variant II Turbo, Bio-rad). Performed Using Venous Blood Samples y = 1.02x - 0.143, R 2 = 0.962 Bio-Rad Variant II Turbo (HbA1c%) Figure 7. Performed Using Venous Blood Samples The Aina HbA1c Monitoring System versus the gold standard revealed a correlation of 0.984 and a kappa agreement of 0.988. 97.6% of individual measurements measured on Aina HbA1c Monitoring system were within 10% of reference method (Variant II Turbo, Bio-rad). 99.2% of individual measurements measured on Aina HbA1c Monitoring system were within 15% of reference method (Variant II Turbo, Bio-rad). Page 3

Blood Glucose Intended Use The Aina Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The Aina Blood Glucose Monitoring System is intended for both over-the-counter (OTC) for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes controls or for point-of-care (POC) use by a trained health care professional. It is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. Test Principle The test employs glucose oxidase, peroxidase and the chromogen 3,3,5,5 - Tetramethylbenzidine along with non reactive ingredients to produce a colour change that is directly proportional to the amount of D-glucose in the blood sample. Specifications Measuring range: 10 to 500 mg/dl (0.55 to 27.7 mmo/l) Supported hematrocit (PCV) range: 25 to 55% Test time: Approximately 5 seconds Operating temperature: 5 to 45 C Blood volume: At least 2 μl (whole blood capillary) Clinical studies I. Narayana Health Hospitals (Bangalore, India) Reference method: 2300D STAT Plus (YSI) Number of subjects: 126 Table 2: System accuracy results for glucose concentration 5.55 mmol/l ( 100 mg/dl) Within ± 5% Within ± 10% Within ± 15% A comparison against capillary whole blood using the YSI 2300D STAT Plus analyser produced the following regression: Y(mg/dL) = 0.9659x - 0.9925 with R 2 = 0.9941. 269/552 48.7 452/552 81.9 528/552 95.7 y = 0.9659x + 0.9925, R 2 = 0.9941 Table 3: System accuracy results for all combined glucose concentration System accuracy results for glucose concentrations between 38.9 mg/dl and 486 mg/dl Total number of samples 48.7 Total number of passing samples 731 System accuracy 97.5% A total of 126 subjects participated in this study. Aina Blood Glucose Monitoring System vs YSI 2300D Stat Plus met the EN ISO 15197:2013 standard accuracy requirements. YSI 2300 Analyzer (mg/dl) Figure 8. Comparison of performance between Aina Device Blood Glucose and YSI 2300 Analyzer Table 1: System accuracy results for glucose concentration < 5.55 mmol/l (<100 mg/dl) Within ± 0.28 mmol/l (Within ± 5 mg/dl) Within ± 0.56 mmol/l (Within ± 10 mg/dl) Within ± 0.83 mmol/l (Within ± 15 mg/dl) 1. More than 95% (97.5%) of the measure glucose values fell within +/- 15 mg/dl of the average measured values of the reference measurement procedure at glucose concentrations < 100 mg/dl (<5.55 mmol/l) or within +/- 15% at glucose concentrations >= 100 mg/dl (>= 5.55 mmol/l). 2. More than 99% (100%) of individual glucose measured value fell within zones of A and B of the Consensus Error Grid. 117/204 181/204 204/204 57.4 88.7 100.0 Page 4

Precision Evaluation y = 0.95x + 6.876, R 2 = 0.949 The repeatability obtained with the blood samples is shown in the following table. The table lists the pooled standard deviation and pooled CV% with 95% confidence intervals for the five levels of glucose tested (n=100). No outliers were detected and excluded from data analysis. At glucose concentrations of 49.6, 85.8, 130.2, 203.1 and 306.2 mg/dl coefficients of variation (CVs) of 1.6, 3.1, 2.7, 2.8 and 3.1% were obtained respectively, indicating a high degree of precision. At all glucose levels tested the coefficient of variation was below 4%. Level 1 Level 2 Level 3 Level 4 Level 5 Grand Mean (mg/dl) 49.6 85.8 130.2 203.1 306.2 Accu-chek Active Meter (mg/dl) Pooled SD (mg/dl) 0.8 2.6 3.5 5.7 9.5 Figure 9. Comparison of performance between Aina Device Blood Glucose and Accu-chek Active Meter (mg/dl) Pooled CV% 0.5-0.9 1.6 1.1-2.0 2.0-3.0 3.1 2.3-3.4 2.5-4.0 2.7 2.0-3.0 4.7-6.5 2.8 2.3-3.2 7.6-10.7 3.1 2.5-3.5 Performed Using Venous Whole Blood Samples A comparison of the Aina Blood Glucose Monitoring System versus the Beckman Coulter AU2700 analyzer was performed using venous whole blood samples. Testing was performed in duplicates for a total of 210 samples. II. Dr. Mohan s Specialties Center (Chennai, India) Reference method: Accu-chek Active Blood Glucose Meter (Roche Diagnostics) and AU2700 (Beckman Coulter) Number of subjects: 129 Accuracy Study Performed Using Capillary Whole Blood Samples This evaluation showed a good correlation against the reference method, with the following regression: Y(mg/dL) = 0.943x + 8.501 with R 2 = 0.96, and satisfied the EN ISO 15197:2013 standard accuracy requirements, as illustrated in the table and Consensus Error Grid below. Bin Within ± 5% Number of samples 108 %of samples 51 A comparison of the Aina Blood Glucose Monitoring System versus the Roche Diagnostics Accu-chek Active Blood Glucose Meter was performed using capillary (fingerstick) whole blood samples. Testing was performed in duplicates for a total of 218 samples. Within ± 10% Within ± 15% Total samples 169 199 210 80 95 This evaluation showed a good correlation against the reference method, with the following regression: Y(mg/dL) = 0.95x + 6.876 with R 2 = 0.949, and satisfied the EN ISO 15197:2013 standard accuracy requirements, as illustrated in the table and Consensus Error Grid below. y = 0.943x + 8.501, R 2 = 0.96 Bin Within ± 5% Within ± 10% Within ± 15% Number of samples 86 167 207 %of samples 39 77 95 Total samples 218 Beckman Coulter AU2700 analyzer Figure 10. Comparison of performance between Aina Device Blood Glucose and Beckman Coulter AU2700 analyzer Page 5

Hemoglobin Intended Use The Aina Hemoglobin Monitoring System is intended to be used for the quantitative measurement of hemoglobin levels in capillary fingerstick and venous whole blood samples. This system is intended for clinical laboratory and point-of-care use to monitor anemia. Test Principle Erythrocytes in the specimen are lysed to release hemoglobin. The hemoglobin is converted to methemoglobin. The intensity of the color produced from this reaction is proportional to the hemoglobin concentration. Specifications Measuring range: 7 to 23 g/dl Test time: 30 seconds Operating temperature: 15 to 30 C Blood volume: 10 μl (whole blood capillary or venous) Clinical studies I. Narayana Health Hospitals (Bangalore, India) Reference method: LH750 (Beckman Coulter) Number of subjects: 71 The Aina Hemoglobin Monitoring System showed great linearity versus the laboratory reference (R 2 = 0.97). In addition, 94.4% of the samples were within 10% of the reference and 97.2% were within 15% of the reference. 20 y = 0.972x + 0.292, R 2 = 0.97 II. Khoo Teck Puat Hospital (Singapore) Reference method: XN (Sysmex) Number of subjects: 42 The Aina Hemoglobin Monitoring System showed great linearity versus the laboratory reference (R 2 = 0.964). In addition, 97.6% of the samples were within 10% of the reference. 20 16 y = 0.971x + 0.196, R 2 = 0.964 Aina Device (g/dl) 16 12 8 Aina Device (g/dl) 12 8 4 4 0 4 8 12 16 20 0 4 8 12 16 20 Beckman Coulter LH750 (g/dl) Figure 11. Comparison of performance between Aina Device Hemoglobin and Beckman Coulter LH750r Sysmex XN (g/dl) Figure 12. Comparison of performance between Aina Device Hemoglobin and Sysmex XN Precision Evaluation A precision evaluation was performed with control solutions for 20 days and showed a CV under 5% at all Hb concentrations. Control Level Mean Value Within Run Total L1 7.92 3.15 2.58-4.03 4.13 3.69-5.08 L2 13.77 1.04 0.85-1.33 1.88 1.61-2.46 L3 17.46 0.75 0.62-0.96 1.33 1.13-1.75 Page 6

Lipids Intended Use The Aina Lipids System is intended to be used for quantitative measurement of total cholesterol, HDL cholesterol and triglycerides in capillary fingerstick and venous whole blood samples. This testing system is intended to measure lipids for the diagnosis and treatment of disorders involving excess cholesterol in the blood or for lipid and lipoprotein metabolism disorders. This system is intended for professional use only. Test Principle Test results are based on the instrument reading light reflected off a test strip that has changed color after blood has been placed on it. The darker the color, the higher the analyte concentration. The instrument converts this reading into a result that it displays. This procedure is based on the Trinder Method for the determination of lipids. Clinical studies Specifications Measuring range: Types Total Cholesterol HDL Cholesterol Triglycerides Measuring range (in mg/dl) 100 to 300 25 to 85 50 to 450 Supported hematocrit (PCV) range: Types Total Cholesterol & Triglycerides HDL Cholesterol Test time: Approximately 2 minutes Operating temperature: 10 to 40 C Blood volume: 15 μl (whole blood capillary or venous) Measuring range (in mmol/l) 2.59 to 7.76 0.65 to 2.20 0.56 to 5.08 PCV range 30-50% 33-49% I. Narayana Health Hospitals (Bangalore, India) Reference method: Dimension RxL Max (Siemens) Number of subjects: 42 Total Cholesterol 350 y = 1.01x + 3.645, R 2 = 0.973 A precision evaluation was performed with control solutions over 15 days, with 2 runs per day and using 2 Aina Devices, and showed a CV under 5% at all Total Cholesterol concentrations. Control Level 300 L1 174.5 7.6 4.38 250 200 150 100 L2 237.7 HDL Cholesterol 11.7 4.91 50 y = 1.033x + 1.464, R 2 = 0.969 0 50 100 150 200 250 300 350 Siemens Dimension RxL Max (mg/dl) Figure 13. Comparison of performance between Aina Device Total Cholesterol and Siemens Dimension RxL Max Precision Evaluation A precision evaluation was performed with venous whole blood samples with 3 operators in a single day and showed a CV under 5% at all Total Cholesterol concentrations. 90 80 70 60 50 40 30 20 10 137.2 4.5 4.29 0 10 20 30 40 50 60 70 80 Siemens Dimension RxL Max (mg/dl) Figure 14. Comparison of performance between Aina Device HDL Cholesterol and Siemens Dimension RxL Max 208.4 7.0 3.36 Page 7

Precision Evaluation A precision evaluation was performed with venous whole blood samples with 3 operators in a single day and showed a CV under 5% at all HDL Cholesterol concentrations. 88.2 4.0 4.56 343.5 16.1 4.67 32.4 1.4 4.41 A precision evaluation was performed with control solutions over 15 days, with 2 runs per day, and showed a CV under 6.5% at all Triglycerides concentrations. 50.6 1.8 3.51 A precision evaluation was performed with control solutions over 15 days, with 2 runs per day and using 2 Aina Devices, and showed a CV under 5.5% at all HDL Cholesterol concentrations. 127.5 194.3 5.8 12.3 4.58 6.35 Control Level L1 L2 27.0 39.5 1.2 2.0 4.45 5.08 II. CARE Hospitals (Hyderabad, India) Reference method: Unicel DxC (Beckman Coulter) Number of subjects: 100 Triglycerides Total Cholesterol 350 y = 0.998x + 0.041, R 2 = 0.97 350 y = 0.922x + 13.327, R 2 = 0.898 300 300 250 200 150 100 250 200 150 100 50 50 0 50 100 150 200 250 300 Siemens Dimension RxL Max (mg/dl) Figure 15. Comparison of performance between Aina Device Triglycerides and Siemens Dimension RxL Max 0 50 100 150 200 250 300 350 Beckman Coulter UniCel Dxc (mg/dl) Figure 16. Comparison of performance between Aina Device Total Cholesterol and Beckman Coulter UniCel Dxc Precision Evaluation A precision evaluation was performed with venous whole blood samples with 3 operators in a single day and showed a CV under 5% at all Triglycerides concentrations. The Aina Lipids Monitoring System for Total Cholesterol showed great linearity versus the laboratory reference (R 2 = 0.95). In addition, 100% of the samples were within 20% of the reference. Page 8

HDL Cholesterol Triglycerides 30 450 y = 0.998x + 6.84, R 2 = 0.966 Bias Aina vs Unicel DcC (mg/dl) 20 10 0-10 -20 25 50 75 100 400 350 300 250 200 150 100 50-30 Beckman Coulter UniCel Dxc (mg/dl) 0 50 100 150 200 250 300 350 400 450 Beckman Coulter UniCel Dxc (mg/dl) Figure 17. Comparison of performance between Bias Aina Vs. Unicel DcC and Beckman Coulter UniCel Dxc The Aina Lipids Monitoring System for HDL Cholesterol 99% of the samples were within 15 mg/dl of the reference. Figure 18. Comparison of performance between Aina Device Triglycerides and Beckman Coulter UniCel Dxc The Aina Lipids Monitoring System for Triglycerides showed great linearity versus the laboratory reference (R 2 = 0.97). In addition, 95% of the samples were within 15 mg/dl or 15% of the reference. Page 9